throbber
Trials@uspto.gov
`571-272-7822
`
`
`
`
`
`
`
`
`
`
`
` Paper 54
`
`
` Date: September 12, 2016
`
`
`
`
`UNITED STATES PATENT AND TRADEMARK OFFICE
`____________
`
`BEFORE THE PATENT TRIAL AND APPEAL BOARD
`____________
`
`PRAXAIR DISTRIBUTION, INC.,
`Petitioner,
`
`v.
`
`INO THERAPEUTICS LLC,
`Patent Owner.
`
`
`
`Case IPR2015-00884
`Patent 8,291,904 B2
`
`
`
`Before KEN B. BARRETT, MICHAEL J. FITZPATRICK, and
`SCOTT A. DANIELS, Administrative Patent Judges.
`
`
`FITZPATRICK, Administrative Patent Judge.
`
`
`
`FINAL WRITTEN DECISION
`35 U.S.C. § 318(a)
`
`
`
`
`
`

`
`IPR2015-00884
`Patent 8,291,904 B2
`
`I.
`
`INTRODUCTION
`
`Petitioner, Praxair Distribution, Inc., filed a Petition to institute an
`
`inter partes review of claims 1–16 of U.S. Patent No. 8,291,904 B2 (Ex.
`
`1001, “the ’904 patent”) pursuant to 35 U.S.C. § 311(a). Paper 1 (“Pet.”).
`
`Patent Owner, INO Therapeutics LLC, filed a Preliminary Response
`
`pursuant to 35 U.S.C. § 313. Paper 9 (“Prelim. Resp.”). In a September 22,
`
`2015, Decision, we granted the Petition, instituting trial on all claims on the
`
`following grounds:
`
`claims 1–8 and 11–16 as obvious over Bathe (Ex. 1005)1, Peters
`
`(Ex. 1004)2, Paoli (Ex. 1006)3, and IR Standard (Ex. 1007)4;
`
`claims 3 and 4 as obvious over Bathe, Peters, Paoli,
`
`IR Standard, and Lebel (Ex. 1008)5; and
`
`claims 9 and 10 as obvious over Bathe, Peters, Paoli,
`
`IR Standard, and Durkan (Ex. 1010).6
`
`Paper 14 (“Inst. Dec.”).
`
`After institution, Patent Owner filed a Patent Owner Response (Paper
`
`30, “PO Resp.”) to which Petitioner filed a Reply (Paper 40, “Pet. Reply”).
`
`
`
`1 U.S. Patent No. 5,558,083, issued Sept. 24, 1996.
`2 U.S. Patent No. 7,114,510 B2, issued Oct. 3, 2006.
`3 French Patent Application Publication No. 2,917,804, published Dec. 26,
`2008.
`4 ISO/IEEE 11073-30300, “Health informatics -- Point-of-care medical
`device communication -- Part 30300: Transport profile -- Infrared
`wireless,” ISO, IEEE, (Dec. 15, 2004).
`5 U.S. Patent No. 6,811,533 B2, issued Nov. 2, 2004.
`6 U.S. Patent No. 4,462,398, issued July 31, 1984.
`2
`
`

`
`IPR2015-00884
`Patent 8,291,904 B2
`
`Also, Patent Owner filed a Motion to Exclude. See Paper 44; see also Paper
`
`48 (Petitioner’s Opposition); Paper 49 (Patent Owner’s Reply).
`
`A hearing for oral arguments was held on May 16, 2016, and a
`
`transcript of the hearing is included in the record. Paper 53.
`
`As discussed below, Petitioner has not shown by a preponderance of
`
`the evidence that any of the challenged claims is unpatentable.
`
`A. The ’904 Patent
`
`The ’904 patent relates to the administration of a therapy gas, such as
`
`nitric oxide (NO), to a patient. Ex. 1001, 1:14–16. In a background section,
`
`the ’904 patent states that there was a need “to ensure that patient
`
`information contained within [a] computerized system matches the gas that
`
`is to be delivered” to the patient and “also a need for such an integrated
`
`device that does not rely on repeated manual set-ups or connections and
`
`which can also track individual patient usage accurately and simply.” Id. at
`
`1:40–45.
`
`The ’904 patent describes a gas delivery system comprising a valve
`
`assembly having a valve and circuit in communication with a control module
`
`to control administration of the therapy gas to a patient. Id. at 5:59–6:4.
`
`Administration of the therapy gas to the patient is controlled by controlling
`
`delivery of the gas from the gas source (i.e., a cylinder to which the valve
`
`assembly is mounted) to a medical device for introducing gas to a patient
`
`(e.g., a ventilator, nasal cannula, endotracheal tube, or face mask). Id.
`
`Figures 2 and 3 are reproduced below.
`
`3
`
`

`
`IPR2015-00884
`Patent 8,291,904 B2
`
`
`
`
`
`Figure 2 shows valve assembly 100 attached to gas source (cylinder) 50 via
`
`attachment portion 102. Ex. 1001, 6:26–29. The valve assembly includes
`
`inlet 104, outlet 106, valve 107, data input 108, and actuator 114 with cap
`
`112 mounted thereto, as well as a circuit that is not shown in Figure 2. Id. at
`
`6:26–37. Figure 3 shows the assembly valve partially disassembled, thus
`
`revealing circuit 150 within the actuator. Id. at 6:30–35.
`
`
`
`
`
`4
`
`

`
`IPR2015-00884
`Patent 8,291,904 B2
`
`Figure 4 is reproduced below.
`
`
`
`Figure 4 shows a block diagram of circuit 150 having valve processor 122,
`
`valve memory 134, valve transceiver 120, valve display 132, reset 128,
`
`power source 130, timer 124,7 and open/close sensor 126. Ex. 1001, 6:41–
`
`54. Gas data, such as gas composition and concentration, can be input to
`
`memory 134 in various ways such as programmed by the gas supplier or
`
`scanned from a bar code on the gas source. Id. at 7:5–21. The valve
`
`assembly is configured to communicate with the control module via wireless
`
`optical line-of-sight transmission between the valve transceiver and a CPU
`
`transceiver of the control module. Id. at 8:41–48.
`
`
`
`
`
`
`
`7 Figure 4 mislabels the timer as 134. It should be labelled 124. Ex. 1001,
`6:45.
`
`5
`
`

`
`IPR2015-00884
`Patent 8,291,904 B2
`
`Figure 9 is reproduced below.
`
`
`
`Figure 9 shows control module 200 that is physically separate from, but in
`
`close proximity to, the valve assembly 100. Ex. 1001, 10:36–48. A
`
`depiction of the previously mentioned wireless optical line-of-sight
`
`transmission between control module 200 and valve assembly 100 is labeled
`
`300. Id. at 6:7, 8:24–27. The control module is ultimately responsible for
`
`delivery and regulation of a desired gas to a ventilator and patient, and it
`
`requests data from circuit 150 within valve assembly 100 at pre-determined
`
`intervals to facilitate the appropriate gas delivery to the patient. Id. at 8:41–
`
`57, 9:62–10:4.
`
`
`
`
`
`6
`
`

`
`IPR2015-00884
`Patent 8,291,904 B2
`
`B. The Challenged Claims
`
`Petitioner challenges all claims of the ’904 patent, i.e., claims 1–16.
`
`Claim 1 is the sole independent claim. It is illustrative and reproduced
`
`below.
`
`1. A valve assembly to deliver a gas comprising NO
`from a gas container containing the gas comprising NO, the
`valve assembly comprising:
`
`a valve attachable to the gas container containing the gas
`comprising NO, the valve including an inlet and an outlet in
`fluid communication and a valve actuator to open or close the
`valve to allow the gas comprising NO through the valve to a
`control module;
`
`a circuit supported within the valve assembly and
`disposed between the actuator and a cap, the circuit including:
`
`a valve memory to store gas data comprising gas
`concentration in the gas container and
`
`a valve processor and a valve transceiver in
`communication with the valve memory to send wireless
`optical line-of-sight signals to communicate the gas data
`to the control module that controls gas delivery to a
`subject; and
`
`a data input disposed on the actuator and in
`communication with said valve memory, to permit a user to
`enter the gas data into the valve memory.
`
`
`
`
`
`7
`
`

`
`IPR2015-00884
`Patent 8,291,904 B2
`
`II. ANALYSIS
`
`A. Claim Construction
`
`“A claim in an unexpired patent shall be given its broadest reasonable
`
`construction in light of the specification of the patent in which it appears.”
`
`37 C.F.R. § 42.100(b). Pursuant to that standard, the claim language should
`
`be read in light of the specification, as it would be interpreted by one of
`
`ordinary skill in the art. In re Suitco Surface, Inc., 603 F.3d 1255, 1260
`
`(Fed. Cir. 2010). Thus, we generally give claim terms their ordinary and
`
`customary meaning. See In re Translogic Tech., Inc., 504 F.3d 1249, 1257
`
`(Fed. Cir. 2007) (“The ordinary and customary meaning is the meaning that
`
`the term would have to a person of ordinary skill in the art in question.”)
`
`(internal quotation marks omitted).
`
`Neither party has proposed expressly construing any term of the
`
`challenged claims. Pet. 8–9; PO Resp. 6. We discern no reason to construe
`
`expressly any term for purposes of this Decision.
`
`B. Petitioner Ground 1
`
`Petitioner asserts that claims 1–8 and 11–16 are unpatentable because
`
`they would have been obvious over Bathe, Peters, Paoli, and IR Standard.
`
`Pet. 10. For the reasons explained below, Petitioner has not met its burden
`
`of persuasion in proving that any claim is unpatentable as asserted.
`
`1. Overview of Bathe
`
`Bathe identifies the same first named inventor as the challenged ’904
`
`patent. Ex. 1005, at [75]. Bathes discloses a “nitric oxide delivery system
`
`that is useable with various means of administering the NO . . . such as a
`
`8
`
`

`
`IPR2015-00884
`Patent 8,291,904 B2
`
`ventilator or with spontaneous ventilation where the NO is introduced by
`
`means of a gas proportioning device that provides a continuous flow to the
`
`patient.” Id. at 2:14–19.
`
`Figure 1 of Bathe is reproduced below.
`
`
`
`Figure 1 shows a schematic view of the Bathe system having flow sensor 26
`
`and flow transducer 46, which determine the flow of gas in the system, and
`
`CPU 56 with input device 58, which provides for an operator to select a
`
`desired concentration of NO to the patient. With flow and operator input
`
`information, the CPU calculates the desired flow to provide the selected NO
`
`concentration and, via feedback loop shown above in Figure 1, adjusts the
`
`desired gas concentration and flow via signals sent to valves 14, 18, 20, and
`
`24. Ex. 1005, 6:5–20. Another input to CPU 56 is the NO concentration in
`
`supply cylinder 10. Id. at 6:5–6. Bathe states the following about that input:
`
`The NO sensor 65 senses the concentration of NO
`in the supply cylinder 10 so that the user can verify
`
`9
`
`

`
`IPR2015-00884
`Patent 8,291,904 B2
`
`that the proper supply is being utilized or,
`alternatively, the CPU 56 may use that input to
`adjust the system to adapt for any concentrations of
`NO in the supply within certain limits.
`
`Ex. 1005, 6:6–11. In other words, the CPU knows the gas flow and NO
`
`concentration from supply 10, as well as the actual flow of gas administered
`
`to the patient from the delivery device by transducer 46 and gas sensing
`
`bench 52, and, knowing the desired NO concentration set by the user via
`
`input 58, the CPU can adjust valves 14, 18, 20, and 24 to bring the actual gas
`
`flow and NO concentration to the patient into accord with the user’s desired
`
`input level. Id. at 6:43–53.
`
`2. Overview of Peters
`
`Peters discloses a valve with a “smart” handle for use with gas
`
`cylinders involving the use of gases administered to patients in medical
`
`treatments. Ex. 1004, 1:16–17, 1:34–35. The “valve records all the
`
`treatment information and makes the information readily accessible for use
`
`in tracking and invoicing.” Id. at 1:35–37.
`
`Figure 1 of Peters is reproduced below.
`
`10
`
`

`
`IPR2015-00884
`Patent 8,291,904 B2
`
`
`
`Figure 1 illustrates an exploded view of valve 10 having valve body 14
`
`supporting valve handle 16 and gas inlet port 18 for connecting to and
`
`communicating with a gas cylinder (not shown). Within the handle are
`
`several electronic components, namely, processor 23, timer 21, memory 22
`
`and data port 22', sensor 28, battery 25 and display 26. Id. at 2:58–64.
`
`Figure 2b is reproduced below.
`
`Figure 2b shows a block diagram of the electronic components just
`
`mentioned. Peters explains that the memory configuration is established by
`
`
`
`11
`
`

`
`IPR2015-00884
`Patent 8,291,904 B2
`
`initial parameters such as the following: born-on date (date when cylinder
`
`was filled); cylinder serial number; gas lot number; set the timers (which
`
`may include a calendar timer and an event timer); and clear-the-log registers.
`
`Id. at 5:43–56.
`
`“When the valve handle 16 is turned to open or close the valve, the
`
`proximity sensor 28 triggers the processor 23 to instruct the memory device
`
`22 to log the event, including date, time, and whether the event was an
`
`opening or a closing of the valve.” Id. at 6:21–25. “Thus, as the handle 16
`
`is rotated to open the valve 10 in order to provide gas treatments to patients,
`
`the memory device 22 in the handle 16 records the number and duration of
`
`the treatments.” Id. at 6:29–32. Also, Peters teaches that data recorded in
`
`the memory can be downloaded using a wand reader via data port 22', or
`
`handle 16 can “include a transmitter to transmit the data to a remote
`
`recording device at intervals or on command, as desired.” Id. at 6:47–7:4.
`
`3. Overview of Paoli
`
`Paoli relates to a connection system for a valve to a gas bottle or
`
`cylinder. Ex. 1006, 017. The described connection system includes a
`
`mechanism whereby valve “opening may take place only if the type of gas
`
`contained in the bottle 10 corresponds to the type of gas intended to supply
`
`the circuit 1 used through the valve 20, so as to avoid any risk of error in the
`
`connection of the bottle to the valve.” Id. at 019.
`
`12
`
`

`
`IPR2015-00884
`Patent 8,291,904 B2
`
`Figure 1 of Paoli is reproduced below.
`
`
`
`Figure 1 shows a block diagram illustrative of control module 300 for
`
`controlling valve 20. The control module receives input signal IDb, which is
`
`the identification of gas type being supplied from the bottle, and compares
`
`this with input data IDv, which is the desired type of gas for the procedure
`
`that is stored in memory 200. Id. Paoli explains that “the control module
`
`300 comprises means 310 for comparing the identification data IDb and IDv
`
`and means 320 for transmitting a control signal to the valve 20, capable of
`
`emitting a signal for opening the valve in case of a positive comparison.” Id.
`
`In another embodiment, Paoli discloses that the type of gas in bottle
`
`10 (IDb) can be acquired from information carrier 120, such as an RFID tag
`
`on the bottle, which would be read by sensor 110 when the valve is attached
`
`to the bottle. Id. at 020.
`
`4. Overview of IR Standard
`
`IR Standard is a protocol promulgated by the Institute of Electrical
`
`and Electronics Engineers (IEEE) as an international standard for short-
`
`range infrared (IR) wireless communication for medical devices used at or
`
`13
`
`

`
`IPR2015-00884
`Patent 8,291,904 B2
`
`near a patient. Ex. 1007, Abstract. IR Standard describes wireless
`
`communication standards with a goal of “[f]acilitat[ing] the efficient
`
`exchange of vital signs and medical device data, acquired at the point-of-
`
`care, in all health care environments.” Id. at vi. IR Standard further
`
`explains that such “standards are especially targeted at acute and continuing
`
`care devices, such as patient monitors, ventilators, infusion pumps, ECG
`
`devices, etc.” Id. IR Standard illustrates an IR communication system
`
`including an IR transceiver in order to retrofit a previously hard wired cable-
`
`communicating system. Id. at 39–40.
`
`5. Petitioner Has Not Shown That Claims 1–8 and 11–16 Would
`Have Been Obvious over Bathe, Peters, Paoli, and IR Standard
`
`All of the challenged claims are directed to a valve assembly with a
`
`circuit having:
`
`a valve memory to store gas data comprising gas
`concentration in the gas container and
`
`in
`transceiver
`a valve processor and a valve
`communication with the valve memory to send wireless optical
`line-of-sight signals to communicate the gas data to the control
`module that controls gas delivery to a subject.
`
`Peters discloses a memory within the handle of its smart valve but it
`
`does not store gas data comprising gas concentration in the container. It
`
`stores born-on date (date when cylinder was filled); cylinder serial number;
`
`gas lot number; set the timers (which may include a calendar timer and an
`
`event timer); and clear-the-log registers. Ex. 1004, 5:43–56. It also stores
`
`when and for how long the valve is opened or closed. Id. at 6:21–25. Paoli
`
`discloses a memory for storing the type, but not concentration, of a gas in a
`
`supply cylinder (i.e., IDb data). Ex. 1006, 020–021. Bathe teaches
`
`14
`
`

`
`IPR2015-00884
`Patent 8,291,904 B2
`
`obtaining the NO concentration actually being delivered to a ventilator in
`
`order to provide feedback for a CPU to adjust as needed administration of
`
`NO a patient is receiving. Ex. 1005, 6:5–20. IR Standard describes wireless
`
`communication standards in medical devices. Ex. 1007, vi.
`
`Although it may not be anticipated, a claimed invention is nonetheless
`
`unpatentable “if the differences between the claimed invention and the prior
`
`art are such that the claimed invention as a whole would have been obvious
`
`before the effective filing date of the claimed invention to a person having
`
`ordinary skill in the art to which the claimed invention pertains.” 35 U.S.C.
`
`§ 103(a).8 In assessing obviousness, “the scope and content of the prior art
`
`are to be determined; differences between the prior art and the claims at
`
`issue are to be ascertained; and the level of ordinary skill in the pertinent art
`
`resolved.” Graham v. John Deere Co., 383 U.S. 1, 17 (U.S. 1966).
`
`Additionally, secondary considerations such as “commercial success, long
`
`felt but unsolved needs, failure of others, etc., might be utilized to give light
`
`to the circumstances surrounding the origin of the subject matter sought to
`
`be patented. As indicia of obviousness or nonobviousness, these inquiries
`
`may have relevancy.” Id. at 17–18.
`
`Each of Petitioner’s grounds relies on multiple references. In
`
`particular, with respect to Ground 1, the Petition asserts that a “person of
`
`ordinary skill in the art would have been motivated to combine the ’083
`
`Patent [Bathe], the ’510 Patent [Peters], the FR ’804 Publication [Paoli], and
`
`
`
`8 The Leahy-Smith America Invents Act (“AIA”), Pub. L. No. 112-29, took
`effect on March 18, 2013. As the application from which the ’904 patent
`issued was filed before that date, our citations to 35 U.S.C. § 103 are to its
`pre-AIA version.
`
`15
`
`

`
`IPR2015-00884
`Patent 8,291,904 B2
`
`the IR Standard to predictably result in an improved nitric oxide delivery
`
`system incorporating the advantageous aspects of each reference.” Pet. 19.
`
`For example, and with respect to the valve memory limitation, Petitioner
`
`argues Ground 1 as follows:
`
`Claim 1 requires the circuit to include a “valve memory to store
`gas data comprising gas concentration in the gas container.” As
`discussed above, the ’510 Patent [Peters] discloses a valve
`memory 22, disposed between the actuator and the cap. (Ex.
`1004 at 2:58–61, 3:3–5, Ex. 1002 ¶ 23.) The ’510 Patent
`discloses that gas data can be stored in the valve memory to
`indicate information about the gas in the cylinder. (See Ex. 1004
`at 5:43–6:12.) This information can be used for gas therapy.
`(Ex. 1004 at 7:36–47.) The FR ’804 Publication [Paoli] also
`discloses a memory for storing data (i.e., IDb data) about the gas
`in the cylinder. (Ex. 1006 at 20–21.) Finally, the ’083 Patent
`[Bathe] teaches that one of the gas data characteristics that can
`be stored and used to trigger alarms is gas concentration. (Ex.
`1005 at 5:60–6:4.) The ’083 Patent also teaches that the actual
`concentration of the gas in the cylinder can be used to “verify
`that the proper supply is being utilized.” (Ex. 1005 at 6:5–8.)
`Accordingly, the combination of references in this Ground
`discloses that the valve memory can store gas data indicative of
`the concentration of gas in the container, as required by this
`limitation. (Ex. 1002 ¶¶ 101, 107.)
`
`Pet. 28.
`
`A claimed invention, however, “is not proved obvious merely by
`
`demonstrating that each of its elements was, independently, known in the
`
`prior art.” KSR Int’l Co. v. Teleflex Inc., 550 U.S. 398, 418 (2007). To
`
`prove obviousness, there must have been, at the time of invention, “an
`
`apparent reason to combine the known elements in the fashion claimed by
`
`the patent at issue.” Id. Such reason cannot be provided by the very patent
`
`being challenged as obvious. Id. at 421 (“A factfinder should be aware, of
`
`16
`
`

`
`IPR2015-00884
`Patent 8,291,904 B2
`
`course, of the distortion caused by hindsight bias and must be cautious of
`
`arguments reliant upon ex post reasoning.”); Cheese Sys. v. Tetra Pak
`
`Cheese & Powder Sys., 725 F.3d 1341, 1352 (Fed. Cir. 2013) (“Obviousness
`
`cannot be based on the hindsight combination of components selectively
`
`culled from the prior art to fit the parameters of the patented invention.”)
`
`(quotation marks omitted).
`
`The Petition includes a section titled “Motivation to Combine Prior
`
`Art” that runs approximately five pages. Pet. 19–25. It is difficult to discern
`
`precisely what Petitioner’s proffered reasons are for why a person of
`
`ordinary skill in the art would have made a combination that falls within the
`
`scope of the challenged claims.
`
`Petitioner offers, in part, the following reasoning:
`
`The ’083 Patent [Bathe] teaches a basic NO delivery system with
`various component parts such as a gas cylinder 10, valves 14, 18,
`20, and 24, and a control CPU 56 for controlling the flow and
`concentration of NO gas delivery to a patient (see Ex. 1005 at
`6:20-28). The ’510 Patent [Peters] discloses an advanced or
`smart handle and valve attached to a gas cylinder for storing and
`transmitting information about the gas to a remote module. A
`person of skill in the art would have known to combine the
`control module disclosed [in] the ’083 Patent with the smart
`handle and valve disclosed in the ’510 Patent to obtain both the
`benefits of the smart handle of the ’510 Patent and the benefits
`of the delivery system of the ’083 Patent. (Ex. 1002 ¶ 100.)
`
`Pet. 20. This proffered reasoning, which proposes combining prior art
`
`teachings merely for the purposes of obtaining the individual benefits of
`
`each, is not persuasive. “Although common sense directs one to look with
`
`care at a patent [] that claims as innovation the combination of two known
`
`devices according to their established functions, it can be important to
`
`17
`
`

`
`IPR2015-00884
`Patent 8,291,904 B2
`
`identify a reason that would have prompted a person of ordinary skill in the
`
`relevant field to combine the elements in the way the claimed new invention
`
`does.” KSR, 550 U.S. at 418. “This is so because inventions in most, if not
`
`all, instances rely upon building blocks long since uncovered, and claimed
`
`discoveries almost of necessity will be combinations of what, in some sense,
`
`is already known.” Id. at 418–19. It is important—indeed, we determine
`
`that it is required in the context of the challenged claims and the relied-upon
`
`prior art teachings before us—for Petitioner to identify a reason that would
`
`have prompted a person of ordinary skill to combine the prior art in the
`
`fashion claimed by Patent Owner.
`
`The above block-quoted Petitioner argument cites to declaration
`
`testimony of Dr. Stone. But, the cited testimony is similarly insufficient. It
`
`states:
`
`100. The ‘083 Patent [Bathe] teaches a nitric oxide
`delivery system with various components including a gas
`cylinder, several valves and a CPU for controlling therapy, and a
`ventilator for delivering gas to a patient. The ‘510 Patent [Peters]
`teaches a valve assembly for gas delivery that could easily be
`incorporated into the delivery system described in the ‘083
`Patent by installing the ‘510 Patent’s valve on the cylinder of the
`‘083 Patent. When so incorporated, the valve of the ‘510 Patent
`is placed in fluid communication with the fluid circuit of the ‘083
`Patent. The valve from the ‘510 Patent also adds its “smart
`handle” features to the ‘083 Patent’s system, including a
`processor, a transceiver, and a memory device disposed in the
`valve handle. These “smart handle” features provide the same
`benefits described in the ‘510 Patent to the system of the ‘083
`Patent.
`
`Ex. 1002 ¶ 100. Dr. Stone’s reasoning is unpersuasive. First, Dr. Stone
`
`testifies that the smart valve of Peters could be incorporated into Bathe
`
`18
`
`

`
`IPR2015-00884
`Patent 8,291,904 B2
`
`without saying why a person of ordinary skill in the art would do so in the
`
`first instance. Ex. 1002 ¶ 100. He then describes, albeit vaguely, the
`
`benefits of making the combination. Id.
`
`In his concluding sentence, Dr. Stone testifies that incorporating the
`
`smart valve of Peters into Bathe would provide Bathe with “the same
`
`benefits described in” Peters. Id. Peters describes the benefits of its smart
`
`valve as providing the ability to “record[] the number and duration of the
`
`treatments,” to “read or download[]” the recorded information, and to
`
`“generate[] reports to keep a record of the treatments on the patients, for
`
`record keeping, for billing the patients, and for checking the billing [a
`
`healthcare provider] receives from its [therapeutic gas] supplier.” Ex. 1004,
`
`6:21–63.
`
`Petitioner has not provided an adequate reason to combine the
`
`teachings in the manner Petitioner asserts renders the claims unpatentable.
`
`Even if we were to assume that the described benefits of each reference
`
`provided a reason to combine their teachings—as Dr. Stone’s testimony
`
`implies—the result would not be the combination Petitioner asserts. Rather,
`
`the result would be the Bathe NO delivery system remaining controlled by a
`
`CPU that operates based on real time feedback from transducers in which
`
`the introduced smart valve would record, as Peters teaches, the number and
`
`duration of treatments for record-keeping purposes, not for controlling gas
`
`delivery.
`
`Relying on the same paragraph 100 of Dr. Stone’s declaration,
`
`Petitioner offers the following additional reason to combine the prior art:
`
`[S]ince some advantages of the ’510 Patent [Peters] involve
`tracking gas data in the cylinders (e.g., “assign patient ID to
`
`19
`
`

`
`IPR2015-00884
`Patent 8,291,904 B2
`
`cylinders” and “identify and control cylinders for blinded clinical
`trials,” see Ex. 1004 at 7:36-47), a person of skill in the art would
`have understood that the valve memory 22 disclosed in the ’510
`Patent could be used to store gas data, and in turn, that data could
`be transmitted to a control module, as disclosed in the ’083
`Patent, to control gas therapy. (Ex. 1002 ¶ 100.)
`
`Pet. 20–21. In its Reply, Petitioner expands on this reasoning, quoting the
`
`Abstract of Peters as disclosing “that data stored in its valve memory can be
`
`used ‘for inventory control, and for other record-keeping and control
`
`functions.’” Reply 14 (quoting Ex. 1004, Abstract) (Petitioner’s emphasis).
`
`Petitioner argues that this reference to “control functions” refers, not to
`
`inventory control, but rather to controlling the administration of therapeutic
`
`gas to a patient. Id. at 14–15 (citing Ex. 2020, 124:16–125:6, 128:23–
`
`129:1).
`
`We reach a different factual finding. The relevant sentence from the
`
`Abstract reads in entirety as follows: “The log of the events and the
`
`corresponding dates and times may be used to prepare invoices for billing
`
`gas treatments, for inventory control, and for other record-keeping and
`
`control functions.” Ex. 1004, Abstract. By considering the subject of the
`
`sentence—the log of the events and the corresponding dates and times—it
`
`becomes clear that Petitioner’s interpretation that “control functions” refers
`
`to controlling the administration of therapeutic gas to a patient is
`
`unsupportable. Indeed, Peters repeatedly refers to using its smart handle
`
`only for post-treatment uses such as billing and inventory control and the
`
`like. For example, the “Summary” of Peters provides in full:
`
`The present invention provides a valve with a smart handle
`for the gas bottle (or cylinder). This valve records all the
`treatment information and makes the information readily
`
`20
`
`

`
`IPR2015-00884
`Patent 8,291,904 B2
`
`accessible for use in tracking and invoicing. It permits the
`vendor to invoice the user for total treatment time and to provide
`users, such as hospitals or clinics, the information to bill
`individual patients. It also provides both the vendor and the user
`with data which is useful for trend analysis and inventory control.
`
`The valve handle includes sensors for sensing the opening
`and closing of the valve, a timer for timing the duration over
`which the valve is opened, and an electronic memory device
`which records the pertinent information. The information
`recorded by the memory device may include the cylinder fill
`date, the lot batch number, cylinder number, the patient's name,
`the number of times the valve is opened, and the date, time, and
`duration of each opening of the valve, as well as additional
`information, if desired.
`
`The data then can be readily transferred from the memory
`device to a device that generates reports or invoices.
`
`Ex. 1004, 1:34–54; see also id. at 6:56–63 (describing the printing of reports
`
`for billing and inventory purposes); 7:9–15 (describing the generation of
`
`reports “to track treatments, do billings, and to control inventory”). Peters
`
`does not teach using its smart valve for controlling the administration of
`
`therapeutic gas to a patient.
`
`Petitioner’s challenge lacks an adequate reason, not only to
`
`incorporate the smart valve of Peters into Bathe, but also to: (1) repurpose
`
`the incorporated valve by using (a) its “memory to store gas data comprising
`
`gas concentration in the gas container” and (b) its processor and transmitter
`
`“to send wireless optical line-of-sight signals to communicate the gas data to
`
`the control module that controls gas delivery to a subject” and (2) modify the
`
`Bathe system, such that its CPU would receive the wirelessly transmitted gas
`
`concentration data, even though the CPU already receives (and uses) flow
`
`and operator input data to provide a desired NO concentration. See Ex.
`
`21
`
`

`
`IPR2015-00884
`Patent 8,291,904 B2
`
`1005, 6:5–20. Accordingly, Petitioner has not demonstrated that any of the
`
`challenged claims is unpatentable over Bathe, Peters, Paoli, and IR Standard.
`
`C. Petitioner Grounds 2 and 3
`
`Petitioner’s Grounds 2 and 3, challenging claims 3 and 4 further in
`
`view of Lebel and claims 9 and 10 further in view of Durkan, respectively,
`
`suffer from the same failure as its Ground 1. Accordingly, Petitioner has not
`
`demonstrated that any of the challenged claims is unpatentable over the
`
`asserted combination of references.
`
`III. PATENT OWNER’S MOTION TO EXLCUDE
`
`Patent Owner filed a Motion to Exclude Exhibits 1002 and 2020, i.e.,
`
`Dr. Stone’s declaration and deposition testimony. Paper 44. Because we do
`
`not rely on any of Dr. Stone’s testimony in a manner adverse to Patent
`
`Owner, we dismiss the Motion as moot.
`
`IV. CONCLUSION
`
`Petitioner has not shown by a preponderance of the evidence that any
`
`claim of the ’904 patent is unpatentable.
`
`V. ORDER
`
`Accordingly, it is
`
`ORDERED that claims 1–16 of U.S. Patent No. 8,291,904 B2 are not
`
`unpatentable on the record presented;
`
`FURTHER ORDERED that Patent Owner’s Motion to Exclude is
`
`dismissed as moot; and
`
`FURTHER ORDERED that, because this Decision is final, a party to
`
`22
`
`

`
`IPR2015-00884
`Patent 8,291,904 B2
`
`the proceeding seeking judicial review of the Decision must comply with the
`
`notice and service requirements of 37 C.F.R. § 90.2.
`
`
`
`
`
`23
`
`

`
`24
`
`IPR2015-00884
`Patent 8,291,904 B2
`
`Petitioner:
`
`Sanjay Murthy
`Sanjay.murthy@morganlewis.com
`
`Margaux Nair
`Margaux.nair@klgates.com
`
`Maria Doukas
`Maria.doukas@morganlewis.com
`
`Benjamin Weed
`Benjamin.weed.ptab@klgates.com
`
`
`
`Patent Owner:
`
`Bob Steinberg
`Bob.steinberg@lw.com
`
`Daniel Brown
`Daniel.brown@lw.com
`
`David Callahan
`David.callahan@lw.com
`
`Kenneth Schuler
`Kenneth.schuler@lw.com
`
`Marc Zubick
`Marc.zubick@lw.com

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket