throbber
Trials@uspto.gov
`571-272-7822
`
`
`
`
`Paper 54
`Entered: September 15, 2016
`
`
`
`UNITED STATES PATENT AND TRADEMARK OFFICE
`____________
`
`BEFORE THE PATENT TRIAL AND APPEAL BOARD
`____________
`
`PRAXAIR DISTRIBUTION, INC.,
`Petitioner,
`
`v.
`
`INO THERAPEUTICS, LLC,
`Patent Owner.
`____________
`
`Case IPR2015-00891
`Patent 8,573,210 B2
`
`
`
`Before KEN B. BARRETT, MICHAEL J. FITZPATRICK, and
`SCOTT A. DANIELS, Administrative Patent Judges.
`
`DANIELS, Administrative Patent Judge.
`
`
`
`FINAL WRITTEN DECISION
`35 U.S.C. § 318(a)
`
`
`

`
`IPR2015-00891
`Patent 8,573,210 B2
`
`Praxair Distribution, Inc. (“Praxair”) filed a Petition to institute an
`
`inter partes review of claims 1–16 of U.S. Patent No. 8,573,210 B2 (Ex.
`
`1001, “the ’210 patent”). Paper 1 (“Pet.”).1 We instituted trial for claims 1–
`
`16 of the ’210 patent on certain grounds of unpatentability asserted in the
`
`Petition. Paper 14 (“Decision to Institute” or “Inst. Dec.”). After institution
`
`of trial, Patent Owner, INO Therapeutics, LLC, (“Patent Owner” or “INO”),
`
`filed a Patent Owner Response. Paper 30 (“PO Resp.”). Praxair timely filed
`
`a Reply. Paper 40 (“Reply”).
`
`A consolidated hearing for this proceeding as well as IPR2015-00884,
`
`IPR2015-00888, IPR2015-00889, and IPR2015-00893, each involving
`
`Praxair and INO but different patents, was held on May 16, 2016. The
`
`transcript of the consolidated hearing has been entered into the record.
`
`Paper 53. (“Tr.”).
`
`Also before the Board is INO’s Motion to Exclude Evidence (Paper
`
`44), Praxair’s Opposition to the same (Paper 48) and INO’s Reply (Paper
`
`49).
`
`After considering the evidence and arguments of counsel and for the
`
`reasons set forth below, we determine that Praxair has not met its burden of
`
`showing that claims 1–16 of the ’210 patent are unpatentable.
`
`I.
`
`BACKGROUND
`
`A. The ’210 Patent (Ex. 1001)
`
`The ’210 patent (Ex. 1001), titled “Nitric Oxide Delivery Device,”
`
`relates generally to a gas delivery device used in a gas delivery system, and a
`
`method for administering therapy gas, such as nitric oxide (NO), to a
`
`
`1 The ’210 Patent is the subject of litigation captioned INO Therapeutics
`LLC et al v. Praxair Distribution Inc. et al., 1:15-cv-00170-GMS currently
`pending in the District Court for the District of Delaware.
`
`2
`
`
`
`

`
`IPR2015-00891
`Patent 8,573,210 B2
`
`medical patient. Ex. 1001, 1:14–27, Fig. 1. In the Background section, it
`
`states that “[t]here is a need for a gas delivery device that integrates a
`
`computerized system to ensure that patient information contained within the
`
`computerized system matches the gas that is to be delivered by the gas
`
`delivery device” to the patient, and “also a need for such an integrated
`
`device that does not rely on repeated manual set-ups or connections and
`
`which can also track individual patient usage accurately and simply.” Id. at
`
`1:40–46.
`
`The ’210 patent describes specifically gas delivery system 10
`
`including a supply cylinder of gas, i.e. gas source 50, and valve assembly
`
`100 for connecting gas source 50 to a gas delivery circuit including
`
`ventilator 400. Id. at 6:5–15.
`
`
`
`3
`
`

`
`IPR2015-00891
`Patent 8,573,210 B2
`
`Figure 1 of the ’210 patent, above, is a diagrammatic representation of
`
`gas delivery system 10 for delivering therapeutic gas to a patient.
`
`The ’210 patent is particularly drawn to valve assembly 100 which, by
`
`opening and closing, allows gas to flow from gas source 50, generally a
`
`cylinder containing NO gas for example, to the gas delivery system. Id. at
`
`2:10–20. Figures 2 and 3 are reproduced below depicting valve assembly
`
`100 having manual actuator 114, valve 107 and circuit 150 disposed in the
`
`actuator for communicating with a control module to control administration
`
`of the therapy gas to a patient. Id. at 5:60–6:5.
`
`
`
`
`
`
`
`
`
`4
`
`

`
`IPR2015-00891
`Patent 8,573,210 B2
`
`Figure 2, above, depicts valve assembly 100 and actuator 114 of the
`
`gas delivery device in communication via valve 107 with gas source 50.
`
`Figure 3, above, illustrates an exploded view of actuator 114 and circuit 150.
`
`Administration of therapy gas to the patient is regulated by a control
`
`module that controls delivery from gas source 50 (i.e., cylinder or tank to
`
`which the valve assembly is mounted) to a medical device for introducing
`
`gas to a patient (e.g., a ventilator, nasal cannula, endotracheal tube, face
`
`mask, etc.). Id. Circuit 150 of valve assembly 100, shown diagrammatically
`
`below, is disposed in actuator 114 and communicates, for example via a
`
`wireless link by valve transceiver 120, with the control module. Id. at 6:20–
`
`25.
`
`
`
`Figure 4 is a block diagram depicting various components of circuit
`
`
`
`150.
`
`Figure 4 depicts various components of circuit 150 including, inter
`
`alia, valve processor 122, memory 134, valve transceiver 120, power source
`
`5
`
`

`
`IPR2015-00891
`Patent 8,573,210 B2
`130, and valve display 132.2 Memory 134 stores the gas data for the
`
`particular gas source to which the valve assembly is attached. Gas data,
`
`such as gas composition and concentrations, can be input to memory 134 in
`
`various ways such as programmed by the gas supplier or scanned from a bar
`
`code on the gas source itself. Id. at 7:5–10. Valve display 132 allows a
`
`user, via window 113 on actuator 114, to view information regarding valve
`
`operation such as open or close, as monitored by open/close sensor 126, and
`
`the time duration which the valve was open for an event. Id. at 7:30–38.
`
`Valve transceiver 120 communicates with the control module that is
`
`physically separate, but in relatively close proximity to the valve assembly,
`
`via an optical wireless line-of-sight signal “during a pre-determined interval
`
`in response to a signal from the control module CPU transceiver 220.” Id. at
`
`8:24–27, 10:36–48, Figs. 7–9.
`
`Control module 200, shown below in Figure 9, is ultimately
`
`responsible for delivery and regulation of a desired gas to a ventilator and
`
`patient, and requests data from valve assembly circuit 150 at pre-determined
`
`intervals to facilitate the appropriate gas delivery to the patient. Id. at 8:41–
`
`57, 9:62–10:4.
`
`
`2 Timer component is apparently mislabeled as 134, and recited in the
`specification as reference number 124.
`
`6
`
`

`
`IPR2015-00891
`Patent 8,573,210 B2
`
`Figure 9 illustrates gas source 50 and valve assembly 100 in
`
`communication with control module 200 via a transceiver mounted on cover
`
`
`
`portion 225.
`
`B. Illustrative Claim
`
`Of the challenged claims, claim 1 is independent. Each of dependent
`
`claims 2–11 relate expressly to a nitric oxide delivery device and depend
`
`directly or indirectly from claim 1. Claims 12–16 are method claims that
`
`depend directly or indirectly from the delivery device recited in claims 1 and
`
`6. Claim 1 illustrates the claimed subject matter and is reproduced below:
`
`
`
`1. A nitric oxide delivery device comprising:
`a control module to deliver a gas comprising NO to
`a patient in an amount effective to treat or prevent hypoxic
`respiratory failure; and
`
`7
`
`

`
`IPR2015-00891
`Patent 8,573,210 B2
`
`a valve assembly to deliver the gas comprising NO
`from a gas container containing the gas comprising NO to
`the control module, the valve assembly comprising:
`a valve attachable to the gas container
`containing the gas comprising NO, the valve
`including an
`inlet and an outlet
`in
`fluid
`communication and a valve actuator to open or
`close the valve to allow the gas comprising
`NO through the valve to the control module; and
`a circuit supported within the valve assembly
`and disposed between the actuator and a cap, the
`circuit including:
`a valve memory to store gas data
`comprising one or more of gas identification,
`gas expiration date and gas concentration in
`the gas container and
`a valve processor and a valve
`transceiver in communication with the valve
`memory to send and receive wireless optical
`line-of-sight signals to communicate the gas
`data to the control module and to verify one
`or more of the correct gas, the correct gas
`concentration and that the gas is not expired.
`
`
`C. The Asserted Grounds of Unpatentability
`
`We instituted a trial on Praxair’s challenges to claims 1–16 of the ’210
`
`patent as obvious under 35 U.S.C. § 103 over the combination of the
`
`references listed below and as further supported by the Declaration of Dr.
`
`Robert T. Stone (Ex. 1002).
`
`8
`
`

`
`IPR2015-00891
`Patent 8,573,210 B2
`
`References
`
`Bathe3, Peters4, FR ’8045, INOMAX label6, and
`IR Standard7
`Bathe, Peters, FR ’804, INOMAX label, IR
`Standard, and Lebel ’5338
`Bathe, Peters, FR ’804, INOMAX label, IR
`Standard, and Durkan ’3989
`
`
`Basis Claims
`Challenged
`§ 103 1–9, and 12–16
`
`§ 103 4–5
`
`§ 103 10–11
`
`II.
`
`CLAIM CONSTRUCTION
`
`Neither party proposes express constructions for any term. See Pet.
`
`8–9; PO Resp. 7. We maintain our earlier determination that no term
`
`requires express construction. Inst. Dec. 8.
`
`III. OBVIOUSNESS OVER BATHE, PETERS, FR ’804, INOMAX
`LABEL, THE IR STANDARD, AND LEBEL OR DURKAN
`
`Praxair asserts that claims 1–9 and 12–16 would have been obvious
`
`over Bathe, Peters, FR ’804, INOMAX Label and the IR Standard and
`
`Bathe, Peters, FR ’804, INOMAX Label, the IR Standard, and either Lebel
`
`or Durkan for claims 4–5 and 10–11 respectively. Pet. 9–10. A patent is
`
`invalid for obviousness:
`
`if the differences between the subject matter sought to be
`patented and the prior art are such that the subject matter as a
`whole would have been obvious at the time the invention was
`
`
`3 Ex. 1005, U.S. Patent No. 5,558,083 (Sept. 24, 1996)
`4 Ex. 1004, U.S. Patent No. 7,114,510 B2 (Oct. 3, 2006)
`5 Ex. 1006, FR Pub. No.: 2 917 804 (Dec. 26, 2008)
`6 Ex. 1014, FINAL PRINTED LABELING, CENTER FOR DRUG EVALUATION AND
`RESEARCH, Appl’n. No.: NDA 20845 (2000).
`7 Ex. 1007, Health informatics — Point-of-care medical device
`communication — Part 30300: Transport profile — Infrared wireless,
`ISO/IEEE 11073–30300 (2004)
`8 Ex. 1008, U.S. Patent No. 6,811,533 B2 (Nov. 2, 2004).
`9 Ex. 1010, U.S. Patent No. 4,462,398 (July 31, 1984).
`
`9
`
`

`
`IPR2015-00891
`Patent 8,573,210 B2
`
`made to a person having ordinary skill in the art to which said
`subject matter pertains.
`
`35 U.S.C. § 103. Obviousness is a question of law based on underlying
`
`factual findings: (1) the scope and content of the prior art; (2) the differences
`
`between the claims and the prior art; (3) the level of ordinary skill in the art;
`
`and (4) objective indicia of nonobviousness. See Graham v. John Deere Co.
`
`of Kansas City, 383 U.S. 1, 17–18 (1966). As the party challenging the
`
`patentability of the claims at issue, Praxair bears the burden of proving
`
`obviousness by a preponderance of the evidence. See 35 U.S.C. § 316(e).
`
`We instituted trial on claims 1–16 based on Praxair’s arguments and
`
`accorded some weight initially to the testimony of Dr. Stone with respect to
`
`a motivation to combine based on asserted safety considerations in NO
`
`delivery systems. See Inst. Dec. 18–20. Now, after considering the
`
`evidence and arguments of counsel fully developed on the record before us,
`
`for the reasons set forth below, we determine that Praxair has not met its
`
`burden of showing that claims 1–16 are unpatentable by a preponderance of
`
`the evidence.
`
`A. Scope and Content of the Prior Art
`
`Praxair relies on the combination of Bathe, Peters, FR ’804, the IR
`
`Standard, the INOMAX Label, and either Lebel or Durkan as obviousness
`
`references. We discuss each reference below.
`
`1.
`
`Bathe
`
`Bathe discloses a nitric oxide (NO) delivery system for use with a
`
`medical ventilation device. Ex. 1005, 1:1–11. Figure 1 of Bathe is
`
`reproduced below.
`
`10
`
`

`
`IPR2015-00891
`Patent 8,573,210 B2
`
`
`
`Figure 1 is a depiction of schematic view, partially in block diagram form, of
`
`apparatus in accordance with an embodiment of Bathe. Id., 3:33–35.
`
`Figure 1 shows a schematic view of the Bathe system having flow
`
`sensor 26 and flow transducer 46, which determine the flow of gas in the
`
`system, and CPU 56 with input device 58, which provides for an operator to
`
`select a desired concentration of NO to the patient. Id. at 5:60–6:4. With
`
`flow and operator input information, the CPU calculates the desired flow to
`
`provide the selected NO concentration and, in the feedback loop shown
`
`above in Figure 1, adjusts the desired gas concentration and flow via signals
`
`sent to valves 14, 18, 20, and 24. Id. at 6:5–20. Another input to CPU 56 is
`
`the NO concentration in supply cylinder 10. Id. at 6:5–6. Bathe explains
`
`that
`
`[t]he NO sensor 65 senses the concentration of NO in the supply
`cylinder 10 so that the user can verify that the proper supply is
`being utilized or, alternatively, the CPU 56 may use that input to
`
`11
`
`

`
`IPR2015-00891
`Patent 8,573,210 B2
`
`adjust the system to adapt for any concentrations of NO in the
`supply within certain limits.
`
`Id. at 6:6–11. In other words, CPU 56 knows the NO concentration from
`
`supply 10, as well as the actual flow of gas administered to the patient from
`
`the delivery device by transducer 46 and gas sensing bench 52, and then
`
`knowing the desired NO concentration set by the user via input 58, CPU 56
`
`adjusts valves 14, 18, 20, and 24 to bring the actual gas flow and NO
`
`concentration to the patient into accordance with the user’s desired input
`
`level. Id. at 6:43–53.
`
`2.
`
`Peters
`
`Peters discloses, as shown in Figure 1 below, valve 10 with “smart”
`
`handle 16 having a memory module and circuit to log data such as opening
`
`and closing time for the valve. Ex. 1004, Abst., 2:43–51.
`
`
`
`
`
`12
`
`

`
`IPR2015-00891
`Patent 8,573,210 B2
`
`
`
`Figure 1 of Peters illustrates an exploded view of valve 10 having
`
`valve body 14 supporting valve handle 16 and gas inlet port 18 for
`
`connecting to and communicating with a gas cylinder (not shown). Id. at
`
`1:58–60; 2:43–51. Inside handle 16 are several electronic devices, namely,
`
`processor 23, timer 21, memory 22 and data port 22’, sensor 28, battery 25
`
`and display 26. Id. at 2:58–64. With respect to the electronics, the ’510
`
`patent explains that memory configuration is established by initial
`
`parameters such as:
`
`Born on date (date when cylinder was filled)[;]
`Cylinder serial number[;]
`Gas lot number[;]
`Set the timers (which may include a calendar timer and an event
`timer) [;] and
`Clear the log registers[.]
`
`Id. at 5:43–53; see also id. at 5:54–56 (“Additional area may be available for
`
`recording specific notes or information relative to a specific treatment or
`
`lot.”). When gas is dispensed through valve 10 during operation, sensor 28
`
`prompts processor 23 to log the event, including parameters such as date,
`
`time, and opening or closing of the valve and “[t]hus, as the handle 16 is
`
`rotated to open the valve 10 in order to provide gas treatments to patients,
`
`the memory device 22 in the handle 16 records the number and duration of
`
`the treatments.” Id. at 6:21–32. Also, Peters teaches that data recorded in
`
`the memory can be downloaded using a wand reader via data port 22’ or
`
`handle 16 can “include a transmitter to transmit the data to a remote
`
`recording device at intervals or on command, as desired.” Id. at 6:47– 7:4.
`
`13
`
`

`
`IPR2015-00891
`Patent 8,573,210 B2
`
`3.
`
`FR ’804
`
`FR ’804 relates to a connection system for a valve to a gas bottle or
`
`cylinder. Ex. 1006, 1.10 The described connection system includes a
`
`mechanism whereby valve “opening may take place only if the type of gas
`
`contained in the bottle 10 corresponds to the type of gas intended to supply
`
`the circuit 1 used through the valve 20, so as to avoid any risk of error in the
`
`connection of the bottle to the valve.” Id. at 3. Observing Figure 1 of FR
`
`’804 as reproduced below, control module 300 communicating with valve 20
`
`receives input signal IDb, the identification of gas type being supplied from
`
`the bottle 10, and compares this with input data IDv, the desired type of gas
`
`for the procedure that is stored in memory 200. Id. at 3.
`
`
`
`
`10 We refer to the top numbered pages of the English translation of FR ’804.
`
`14
`
`

`
`IPR2015-00891
`Patent 8,573,210 B2
`
`Figure 1 of FR ’804 is a block diagram illustrative of control module 300 for
`
`controlling valve 20. Once IDb and IDv are input to control module 300, FR
`
`’804 explains that “the control module 300 comprises means 310 for
`
`comparing the identification data IDb and IDv and means 320 for
`
`transmitting a control signal to the valve 20, capable of emitting a signal for
`
`opening the valve in case of a positive comparison.” Id. at 3.
`
`In another embodiment, FR ’804 also discloses that the type of gas
`
`(IDb) in bottle 10 can be input from information carrier 120, such as an
`
`RFID tag on bottle 10, that is read by sensor 110 connected to control
`
`module 300 when valve 20 and bottle 10 are connected. Id. at 4, Fig. 2.
`
`4.
`
`The IR Standard
`
`The IR Standard is a protocol promulgated by the Institute of
`
`Electrical and Electronics Engineers (IEEE) as an international standard for
`
`short-range infrared wireless communication for medical devices used at or
`
`near a patient. Ex. 1007, Abstract (ii). The IR Standard purports to describe
`
`wireless communication standards to “[f]acilitate the efficient exchange of
`
`vital signs and medical device data, acquired at the point-of-care, in all
`
`health care environments.” Id. at vi. This reference further explains that
`
`such “standards are especially targeted at acute and continuing care devices,
`
`such as patient monitors, ventilators, infusion pumps, ECG devices, etc.” Id.
`
`The IR Standard further illustrates an IR communication system including
`
`an IR transceiver in order to retrofit a previously hard wired cable-
`
`communicating system. Id. at 39–40.
`
`5.
`
`INOMAX Label
`
`The INOMAX label reference is a U.S. Food and Drug
`
`Administration (FDA) publication approving labeling for Patent Owner’s
`
`INOmax drug generally used in conjunction with NO therapies for newborns
`
`15
`
`

`
`IPR2015-00891
`Patent 8,573,210 B2
`
`with hypoxic respiratory failure. Ex. 1014, 2. The INOMAX label also
`
`indicates that the drug was used with ventilators administering NO to
`
`neonatal patients. Id. at 6. The INOMAX label provides guidance and
`
`monitoring protocols and states that
`
`[t]he nitric oxide delivery systems used in the clinical trials
`provided operator-determined concentrations of nitric oxide in
`the breathing gas, and the concentration was constant throughout
`the respiratory cycle. . . . In the ventilated neonate, precise
`monitoring of inspired nitric oxide and N02 should be instituted,
`using a properly calibrated analysis device with alarms. This
`system should be calibrated using a precisely defined calibration
`mixture of nitric oxide and nitrogen dioxide, such as INOcal™ .
`Sample gas for analysis should be drawn before the Y-piece,
`proximal to the patient. Oxygen levels should also be measured.
`
`Id. The INOMAX label indicates that the INOmax™ drug is supplied in
`
`aluminum cylinders. Id.
`
`6.
`
`Lebel
`
`Lebel discloses a protocol for an RF telemetry communication system
`
`for medical devices. Ex. 1008, Abst. More specifically, Lebel teaches
`
`optimizing power consumption in communication devices for instance
`
`between a surgically implanted device in a human body, such as an insulin
`
`pump, and an external control or monitoring device by sending
`
`communications intermittently to conserve battery power. Id. at 2:18–35.
`
`Lebel states that the time between communications can occur for example
`
`“no more than 15 seconds apart, more preferably no more than 10 seconds
`
`apart, and even more preferably no more than 5 seconds apart, and most
`
`preferably no more than about 2 seconds apart.” Id. at 25:5–9. Lebel further
`
`explains that power conservation in such medical devices is a balance in
`
`which “desire to minimize power drain in the implantable device is balanced
`
`16
`
`

`
`IPR2015-00891
`Patent 8,573,210 B2
`
`with a desire to have the implantable device respond quickly to commands
`
`transmitted by the external communication device.” Id. at 25:60–63.
`
`7.
`
`Durkan
`
`Durkan discloses a gas supply apparatus and method for providing
`
`respirating gas to a medical patient. Ex. 1010, 1:6–8. Particularly, Durkan
`
`discloses a gas delivery system having a circuit 100 (Figure 5) including a
`
`visual signal such as an illuminated LED 92 that provides visual
`
`confirmation that a valve is in a position such that a gas pulse is being
`
`supplied. Id. at 11:43–46. When the valve is in a position that gas is not
`
`being supplied, the LED will likewise not be illuminated. Id. at 47–51, Fig.
`
`5. Durkan further discloses a dual gas supply for its respiratory system and
`
`that the control means disclosed in Figure 5, including alarm circuit 100, can
`
`be used with the dual gas supply embodiment to indicate by illuminating
`
`LED 92 when both gas supply valves are open and supplying gas to the
`
`system. Id. at 7:29–54, 13:57–65, 14:8–12.
`
`B. The Level of Ordinary Skill in the Art
`
`Praxair asserts that a person of ordinary skill in the art would have had
`
`“a bachelor’s degree in electrical engineering, computer science, computer
`
`engineering, or the equivalent, and would have had at least two years’
`
`experience in biomedical engineering designing medical gas delivery or
`
`monitoring systems.” Pet. 8 (citing Ex. 1002 ¶¶ 17–18). INO does not
`
`expressly dispute the general level of skill in the art proffered by Praxair but
`
`adds through its declarant, Mr. Warren P. Heim, that one of skill in the art
`
`would “know Food and Drug Administration (FDA) Current Good
`
`Manufacturing Practices.” Ex. 2021 ¶ 107.
`
`
`
`17
`
`

`
`IPR2015-00891
`Patent 8,573,210 B2
`
`C. Claims 1–16 over Bathe, Peters, FR ’804, the INOMAX Label, the IR
`Standard and either Label or Durkin
`
`1. Whether One of Ordinary Skill in the Art Would Have Combined
`and/or Modified the Prior Art to Achieve the Claimed Invention
`
`INO asserts that Praxair “has failed to meet its burden of proving that
`
`a POSA would have been motivated to combine the cited prior art references
`
`in the precise manner the claims require.” PO Resp. 38. INO argues that (1)
`
`a person of ordinary skill in the art was unaware of the problem solved by
`
`the ’209 patent; (2) a person of ordinary skill in the art would not have
`
`combined Bathe and Peters; (3) a person of ordinary skill in the art would
`
`not have combined Bathe and Peters with FR ’804; and (4) nothing in the IR
`
`Standard, Lebel, Durkan or INOMAX Label cures the improper asserted
`
`combination of Bathe, Peters, and FR ’804. Id. at 39–55.
`
`INO’s first argument is not persuasive because the evidence indicates
`
`that a person of ordinary skill in the art would have been aware that there
`
`was a risk to providing a patient with the incorrect gas or gas concentration.
`
`For example, the ’210 patent states in the Background section that
`
`[g]as delivery devices are often utilized by hospitals to deliver
`the necessary gas to patients in need. It is important when
`administering gas therapy to these patients to verify the correct
`type of gas and the correct concentration are being used. It is
`also important to verify dosage information and administration.
`
`Ex 1001, 1:22–27 (emphasis added). Bathe also discloses in the Background
`
`of the Invention that it was critical to control the concentration of gas to a
`
`patient:
`
`There is, of course, a need for that concentration to be precisely
`metered to the patient since an excess of NO can be harmful to
`the patient.
`
`18
`
`

`
`IPR2015-00891
`Patent 8,573,210 B2
`
`Ex. 1005, 1:37–39. The Federal Drug Administration (“FDA”) also
`
`recognized the problem of administering incorrect gas, for example, FDA
`
`guidance as of January 24, 2000 promulgated
`
`use of a standard gas-specific fitting (as well as the use of nitric
`oxide gas analyzer) will control the risk of incorrect drug
`administration resulting from use of the incorrect compressed
`gas.
`
`Ex. 1012, 8, § 3.1.1.i. Based on this evidence, we find that the broad
`
`problem of verifying gas concentration was known to those of ordinary skill
`
`in the art. INO cites to Novartis Pharm. Corp. v. Watson Labs., Inc., 611
`
`Fed. App’x. 988, 995 (Fed. Cir. 2015) in support of its position. PO Resp.
`
`40. Novartis is distinguishable because in that case the patent owner solved
`
`an entirely unknown problem, not a previously solved problem. Novartis,
`
`611 Fed. App’x. 988, 997 (The Federal Circuit upholding the district court’s
`
`conclusion of non-obviousness, in part, because of the finding that
`
`“rivastigmine was not known to be susceptible to oxidative degradation.”).
`
`INO’s argument is more properly stated as: the problem of
`
`verification of supplying the correct gas to a patient had already been
`
`solved. Therefore, as FDA approval could be obtained based on the current
`
`state of the art, there was no incentive or motivation to find a different way.
`
`PO Resp. 40–41. We are persuaded that such an argument is directed not to
`
`a brand new and novel problem solved by INO, but instead, as discussed
`
`below, supports INO’s position that Praxair has failed to present persuasive
`
`reasons with rational underpinnings supporting the combination of prior art
`
`references.
`
`After reviewing the record in total, the arguments and evidence
`
`presented by both parties and the prior art itself, we find that Praxair has not
`
`shown a sufficient reason for a person of ordinary skill in the art to have
`
`19
`
`

`
`IPR2015-00891
`Patent 8,573,210 B2
`
`made the proposed combination of Bathe, Peters, FR ’804, the INOMAX
`
`Label, and the IR Standard. PO Resp. 36–53. INO points out, correctly, that
`
`Praxair has not shown that a person of ordinary skill in the art would have
`
`been motivated to combine Peters’s smart valve having a memory for
`
`compiling data for billing and invoicing functions, with Bathe. Id.
`
`Moreover, we are not persuaded that Praxair’s inability to provide sufficient
`
`motivation for combining Peters and Bathe is remedied by the introduction
`
`of FR ’804 and the INOMAX Label, particularly given that Bathe already
`
`included an in-line sensing unit to sense and verify the gas concentration
`
`from the cylinder as mandated by FDA guidance. Id. at 46–47, see also Pet.
`
`Reply 11–12.
`
`Bathe teaches a gas cylinder supplying NO gas to a gas delivery
`
`system and has in-line sensor 65 to detect NO concentration in the delivery
`
`system prior to delivery to a patient. Ex. 1005, 6:5–11. Peters’s smart valve
`
`with a memory, teaches delivering data such as on/off times for the valve, to
`
`a hand-held or laptop computer for compiling usage and billing reports.
`
`Ex. 1004, 6:17–57. Praxair argues that motivation to combine the references
`
`springs from the contention that Peters’s valve “can control ‘the flow of gas
`
`from the cylinder 12 to a ventilator or other gas dispensing device,’ such as
`
`the gas delivery system of [Bathe].” Pet. 20 (citing Ex. 1004, 2:52–55).
`
`Making a similar assertion, Praxair’s expert Dr. Stone, states that Peters’s
`
`“valve assembly for gas delivery [] could easily be incorporated into the
`
`delivery system described in [Bathe] by installing [Peters] valve on the
`
`cylinder of [Bathe].11 Ex. 1002 ¶ 107. INO correctly points out that this
`
`
`11 For the purpose of consistency in our Decision we have replaced
`Petitioner’s quoted use of “the ’083 patent” with the first named inventor
`“[Bathe],” and similarly replaced “the ’510 patent” with “[Peters].”
`
`20
`
`

`
`IPR2015-00891
`Patent 8,573,210 B2
`
`argument and testimony that Peters’s valve could work on Bathe’s cylinder
`
`does not explain why or how a person of skill in the art would combine
`
`Bathe and Peters, only that they could. PO Resp. 43–44. That Bathe could
`
`be modified by simply placing Peters’s valve on the cylinder to control gas
`
`flow from the cylinder, does not, alone, satisfy the requirements for
`
`combining the prior art. See InTouch Techs., Inc. v. VGO Commc’ns, Inc.,
`
`751 F.3d 1327, 1352 (Fed. Cir. 2014) (Experts “testimony primarily
`
`consisted of conclusory references to her belief that one of ordinary skill in
`
`the art could combine these references, not that they would have been
`
`motivated to do so.”).
`
`Petitioner’s challenge lacks an adequate motivation, not only to
`
`incorporate the smart valve of Peters into Bathe, but also to: (1) repurpose
`
`the incorporated valve by using (a) its “memory to store gas data comprising
`
`one or more of gas identification, gas expiration date and gas concentration”
`
`and (b) its processor and transmitter “to send and receive wireless optical
`
`line-of-sight signals to communicate the gas data to the control module that
`
`controls gas delivery to a subject” and (2) modify the Bathe system, such
`
`that its CPU would receive the wirelessly transmitted gas concentration data,
`
`even though the CPU already receives (and uses) flow and operator input
`
`data to provide a desired NO concentration. See Ex. 1005, 6:5–20.
`
`2. Whether the Combination of Prior Art References is Supported by
`Articulated Reasoning and Evidentiary Underpinning
`
`The Supreme Court requires an expansive and flexible approach in
`
`determining whether a patented invention would have been obvious at the
`
`time it was made. See KSR Int’l Co. v. Teleflex Inc., 550 U.S. 398, 415
`
`(2007). The existence of a reason for a person of ordinary skill in the art to
`
`modify a prior art reference is a question of fact. See In re Constr. Equip.
`
`21
`
`

`
`IPR2015-00891
`Patent 8,573,210 B2
`
`Co., 665 F.3d 1254, 1255 (Fed. Cir. 2011). In an obviousness analysis, some
`
`kind of reason must be shown as to why a person of ordinary skill would
`
`have thought of combining or modifying the prior art to achieve the patented
`
`invention. See Innogenetics, N.V. v. Abbott Labs., 512 F.3d 1363, 1374
`
`(Fed. Cir. 2008). A reason to combine or modify the prior art may be found
`
`explicitly or implicitly in market forces; design incentives; the “‘interrelated
`
`teachings of multiple patents’”; “‘any need or problem known in the field of
`
`endeavor at the time of invention and addressed by the patent’”; and “the
`
`background knowledge, creativity, and common sense of the person of
`
`ordinary skill.” Perfect Web Techs., Inc. v. InfoUSA, Inc., 587 F.3d 1324,
`
`1328–29 (Fed. Cir. 2009) (quoting KSR, 550 U.S. at 418–21).
`
`The closest Praxair comes to providing a reason that one of ordinary
`
`skill in the art would have been motivated to combine these references is that
`
`doing so allegedly would “allow the user to better link the gas information
`
`with patient treatments.” Pet. 22. This assertion however, leaves several
`
`unanswered questions in its wake. First, it is unexplained why a direct
`
`“link” between the cylinder gas information and patient treatment control
`
`module would have been perceived as necessary or better based on any
`
`express or even implicit teachings of either Bathe or Peters or the knowledge
`
`of a person of ordinary skill. Second, Praxair fails to explain why or a how a
`
`person of ordinary skill in the art would have “linked” Peters’s data to
`
`patient treatments, as Peters only communicates certain data, such as valve
`
`opening and closing times, to a computer for billing and invoicing purposes,
`
`and neither Peters’s communicated data nor computer is used to determine
`
`any physiological aspect of patient treatment. See Ex. 1004 5:43–6:15.
`
`(“The hospital or other user, as well as the distributor, may later download
`
`22
`
`

`
`IPR2015-00891
`Patent 8,573,210 B2
`
`the data from the memory device 22 to be used for record keeping and
`
`billing.”).
`
`Dr. Stone further opines that “[a] person of skill in the art would have
`
`read [Bathe] and been motivated to apply subsequent improvements to
`
`component parts of that system, such as the improved valve of [Peters].”
`
`Ex. 1002 at ¶ 112 (citing Ex. 1005, 6:11–15). Dr. Stone does not, however,
`
`point to any persuasive evidence or disclosure in Bathe, or provide sufficient
`
`explanation or reasoning to support this statement. See id. In particular, Dr.
`
`Stone does not describe what the “improvements” would have been

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket