throbber
Bioprosthesis Diameter
`23 mm
`26 mm
`
`Frame Height (Profile)
`14.3 mm
`16.1 mm
`
`
`
`The following table identifies the bioprosthesis size that
`should be used based on native valve annulus size, as
`measured by transesophageal echocardiography (TEE).
`
`Native Valve Annulus Size
`(Tissue Annulus Diameter)
`18-22 mm
`21-25 mm
`
`Bioprosthesis Diameter
`
`23 mm
`26 mm
`
` 
`
` Ascendra Balloon Catheter – Model 9100BCL23 for 23
`mm valve procedure and 9100BCL26 for 26 mm valve
`procedure (Figure 2)
`The Ascendra Balloon Catheter is used for delivery of the
`Edwards SAPIEN Transcatheter Heart Valve. The balloon
`catheter has radiopaque markers for visualization under
`fluoroscopy and a balloon for deployment of the
`bioprosthesis. The system also comes with a loader that is
`used to cover the bioprosthesis during delivery. An extension
`tubing is supplied for use with the balloon catheter during
`inflation.
`
`Figure 2. Ascendra Balloon Catheter
`
`
`
`
`2.0 Indications
`The Edwards SAPIEN™ Transcatheter Heart Valve, model
`9000TFX, sizes 23mm and 26mm, is indicated for
`transapical delivery in patients with severe symptomatic
`native aortic valve stenosis who have been determined by
`two cardiac surgeons to be at high risk for surgical aortic
`valve replacement, not suitable for transfemoral delivery per
`heart team decision, and in whom existing co-morbidities
`would not preclude the expected benefit from correction of
`the aortic stenosis.
`The Ascendra Balloon Catheter is indicated for the
`transapical delivery of the Edwards SAPIEN Transcatheter
`Heart Valve.
`
`1
`
`
`
`
`Edwards SAPIEN
`Transcatheter Heart Valve with
`the Ascendra Balloon Catheter
`Instructions for Use
`Caution: Federal (USA) law restricts this device to
`sale by or on the order of a physician.
`Transapical Approach
`Implantation of the transcatheter heart valve should be
`performed only by physicians who have received
`Edwards Lifesciences training. The implanting physician
`should be experienced in balloon aortic valvuloplasty.
`
`Please verify that you have the latest version of the
`instructions for use prior to using the device by visiting
`http://THVIFU.edwards.com or by calling 1.800.822.9837.
`In order to access the instructions for use, an IFU Code
`will be required.
`STERILE: The bioprosthesis is supplied sterilized with
`gluteraldehyde solution. The delivery system is supplied
`sterilized with ethylene oxide gas.
`
`1.0 Device Description
`• Edwards SAPIEN Transcatheter Heart Valve – Model
`9000TFX (Figure 1)
`The Edwards SAPIEN transcatheter heart valve
`(bioprosthesis) is comprised of a balloon-expandable,
`radiopaque, stainless steel (316 L) frame, three bovine
`pericardial tissue leaflets, and a polyethylene terephthalate
`(PET) fabric. The bioprosthesis is treated according to the
`Carpentier-Edwards ThermaFix process, packaged, and
`terminally sterilized in glutaraldehyde
`Figure 1. Edwards SAPIEN Transcatheter Heart Valve
`
`
`
`_______________________
`Edwards Lifesciences, the stylized E logo, Edwards,
`Ascendra, Carpentier-Edwards, PARTNER, ThermaFix and
`Edwards SAPIEN are trademarks of Edwards Lifesciences
`Corporation.
`
`
`
`001
`
`ENDOHEART AG, EX. 2006 Page 1
`EDWARDS LIFESCIENCES CORPORATION (PETITIONER) v. ENDOHEART AG (PATENT OWNER)
`Case No.: IPR2016-00299, U.S Patent No. 8,182,530
`
`

`
`
`
`
`3.0 Contraindications
`The bioprosthesis and delivery system are contraindicated in
`patients who cannot tolerate an anticoagulation/antiplatelet
`regimen or who have active bacterial endocarditis or other
`active infections.
`4.0 Warnings
` Observation of the pacing lead throughout the procedure
`is essential to avoid the potential risk of pacing lead
`perforation.
` There is an increased risk of stroke in transcatheter
`aortic valve replacement procedures, as compared to
`balloon aortic valvuloplasty or other standard
`treatments.
` The devices are designed, intended, and distributed for
`single use only. Do not re-sterilize or reuse the
`devices. There are no data to support the sterility, non-
`pyrogenicity, and functionality of the devices after
`reprocessing.
`Incorrect sizing of the bioprosthesis may lead to
`paravalvular leak, migration, embolization and/or annular
`rupture.
` Accelerated deterioration of the bioprosthesis may occur
`in patients with an altered calcium
`metabolism.Bioprosthesis must remain hydrated at all
`times and cannot be exposed to solutions other than its
`shipping storage solution and sterile physiologic rinsing
`solution. Bioprosthesis leaflets mishandled or damaged
`during any part of the procedure will require replacement
`of the bioprosthesis.
` Caution should be exercised in implanting a
`bioprosthesis in patients with clinically significant
`coronary artery disease.
` Patients with pre-existing mitral valve devices should be
`carefully assessed prior to implantation of the
`bioprosthesis to ensure proper bioprosthesis positioning
`and deployment.
` Patients presenting with combination AV low flow, low
`gradient should undergo additional evaluation to
`establish the degree of aortic stenosis.
` Do not use the bioprosthesis if the tamper evident seal is
`broken, the storage solution does not completely cover
`the bioprosthesis, the temperature indicator has been
`activated, or the bioprosthesis is damaged, or the
`expiration date has elapsed.
` Do not mishandle the Ascendra Balloon Catheter or use
`it if the packaging or any components are not sterile,
`have been opened or are damaged (e.g. kinked or
`stretched), or the expiration date has elapsed.
` Use of excessive contrast media may lead to renal
`failure. Measure the patient’s creatinine level prior to
`the procedure. Contrast media usage should be
`monitored.
` Patient injury could occur if the delivery system is not un-
`flexed prior to removal.
` Care should be exercised in patients with
`hypersensitivities to chromium, nickel, molybdenum,
`manganese, copper, silicon, and/or polymeric materials
` The procedure should be conducted under fluoroscopic
`
`guidance. Some fluoroscopically guided procedures are
`associated with a risk of radiation injury to the skin.
`These injuries may be painful, disfiguring, and long-
`lasting.
` The safety and efficacy of the transapical procedure has
`not been evaluated in for only those patient populations
`where the transfemoral procedure delivery is not suitable.
`
`5.0 Precautions
` Long-term durability has not been established for the
`bioprosthesis. Regular medical follow-up is advised to
`evaluate bioprosthesis performance.
` Glutaraldehyde may cause irritation of the skin, eyes,
`nose and throat. Avoid prolonged or repeated exposure
`to, or breathing of, the solution. Use only with adequate
`ventilation. If skin contact occurs, immediately flush the
`affected area with water; in the event of contact with
`eyes, seek immediate medical attention. For more
`information about glutaraldehyde exposure, refer to
`Material Safety Data Sheet available from
`Edwards Lifesciences.
` To maintain proper valve leaflet coaptation, do not
`overinflate the deployment balloon.
` Appropriate antibiotic prophylaxis is recommended post-
`procedure in patients at risk for prosthetic valve infection
`and endocarditis.
` Bioprosthetic valve recipients should be maintained on
`anticoagulant and antiplatelet therapy (e.g. clopidogrel or
`ticlopidine [75 mg/day]) for 6 months post procedure and
`aspirin (75-100 mg/day) for life, except when
`contraindicated, as determined by their physician.
` The safety of the bioprosthesis implantation has not been
`established in patients who have:
`o
`Pre-existing prosthetic heart valve in the
`aortic position
`Severe ventricular dysfunction with ejection
`fraction <20%
`o Hypertrophic cardiomyopathy with or without
`obstruction (HOCM)
`Safety, effectiveness, and durability have not been
`established for valve-in-valve procedures.
`Safety and effectiveness have not been established for
`patients with the following characteristics/comorbidities:
`o Non-calcified aortic annulus
`o Congenital unicuspid or congenital bicuspid
`aortic valve
`o Mixed aortic valve disease (aortic stenosis
`and aortic regurgitation with predominant
`aortic regurgitation >3+)
`Pre-existing prosthetic heart valve or
`prosthetic ring in any position
`Severe mitral annular calcification (MAC),
`severe (>3+) mitral insufficiency, or Gorelin
`syndrome
`Blood dyscrasias defined as: leukopenia
`(WBC<3000 mm3), acute anemia (Hb <9
`mg%), thrombocytopenia (platelet count
`<50,000 cells/mm3), or history of bleeding
`diathesis or coagulopathy
`o Hypertrophic cardiomyopathy with or without
`
`o
`
`
`
`
`
`o
`
`o
`
`o
`
`2
`
`
`
`002
`
`ENDOHEART AG, EX. 2006 Page 2
`EDWARDS LIFESCIENCES CORPORATION (PETITIONER) v. ENDOHEART AG (PATENT OWNER)
`Case No.: IPR2016-00299, U.S Patent No. 8,182,530
`
`

`
`• Abnormal lab values (including electrolyte imbalance)
`• Hypertension or hypotension
`• Allergic reaction to anesthesia or to contrast media
`• Hematoma
`• Syncope
`• Pain or changes at the access site
`• Exercise intolerance or weakness
`• Inflammation
`• Angina
`• Heart murmur
`• Fever
`• Mechanical failure of delivery system and/or
`accessories
`
`
`Additional potential risks specifically associated with the use
`of the bioprosthesis include, but may not be limited to the
`following:
`
`
`Cardiac arrest
`
`Cardiogenic shock
`
`Emergency cardiac surgery
`
`Cardiac failure or low cardiac output
`
`Coronary flow obstruction/transvalvular flow
`disturbance
`
`
`
` 
`
`
`
` 
`
`
`
` 
`
`
`
`
`
` 
`
` 
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`Injury at the site of ventricular access that may require
`repair
`
`Device thrombosis requiring intervention
`
`Valve thrombosis
`
`Device embolization
`
`Device migration or malposition requiring intervention
`
`Valve deployment in unintended location
`
`Valve stenosis
`
`Structural valve deterioration (wear, fracture,
`calcification, leaflet tear/tearing from the stent posts,
`leaflet retraction, stent creep, suture line disruption of
`components of a prosthetic valve, thickening, stenosis)
`
`Device degeneration
`
`Paravalvular or transvalvular leak
`
`Injury to the mitral valve
`
`Valve regurgitation
`
`Hemolysis
`
`Device explants
`
`Nonstructural dysfunction
`
`Non-emergent reoperation
`
`All listed risks may include symptoms associated with the
`above mentioned medical conditions.
`
`3
`
`
`
`o
`
`o
`
`o
`
`obstruction (HOCM)
`Echocardiographic evidence of intracardiac
`mass, thrombus, or vegetation
`A known hypersensitivity or contraindication
`to aspirin, heparin, ticlopidine (Ticlid), or
`clopidogrel (Plavix), or sensitivity to contrast
`media, which cannot be adequately
`premedicated
`o Native aortic annulus size <18 mm or >25
`mm as measured by echocardiogram
`Significant aortic disease, including
`abdominal aortic or thoracic aneurysm
`defined as maximal luminal diameter 5 cm or
`greater; marked tortuosity (hyperacute bend),
`aortic arch atheroma (especially if thick [> 5
`mm], protruding, or ulcerated) or narrowing
`(especially with calcification and surface
`irregularities) of the abdominal or thoracic
`aorta, severe “unfolding” and tortuosity of the
`thoracic aorta
`Bulky calcified aortic valve leaflets in close
`proximity to coronary ostia
`
`o
`
`
`6.0 Potential Adverse Events
`Potential risks associated with the overall procedure
`including potential access complications associated with
`standard cardiac catheterization for the transapical access
`procedure, balloon valvuloplasty, and the potential risks of
`local and/or general anesthesia
`
`
`• Death
`• Stroke/transient ischemic attack clusters or neurological
`deficit
`• Paralysis
`• Permanent disability
`• Respiratory insufficiency or respiratory failure
`• Hemorrhage requiring transfusion or intervention
`• Cardiovascular injury including perforation or dissection
`of vessels, ventricle, myocardium or valvular
`structures that may require intervention
`• Pericardial effusion or cardiac tamponade
`• Embolization including air, calcific valve material or
`thrombus
`• Infection including septicemia and endocarditis
`• Heart failure
`• Myocardial infarction
`• Renal insufficiency or renal failure
`• Conduction system injury (defect) which may require a
`permanent pacemaker
`• Arrhythmia
`• Retroperitoneal bleed
`• Femoral AV fistula or pseudoaneurysm
`• Reoperation
`• Peripheral ischemia or nerve injury
`• Restenosis
`• Pulmonary edema
`• Pleural effusion
`• Bleeding
`• Anemia
`
`003
`
`ENDOHEART AG, EX. 2006 Page 3
`EDWARDS LIFESCIENCES CORPORATION (PETITIONER) v. ENDOHEART AG (PATENT OWNER)
`Case No.: IPR2016-00299, U.S Patent No. 8,182,530
`
`

`
`7.0 Directions for Use
`7.1 Required Equipment
` Standard cardiac catheterization lab equipment
` Fluoroscopy (fixed, mobile or semi-mobile fluoroscopy
`systems appropriate for use in percutaneous coronary
`interventions)
` Transesophageal or transthoracic echocardiography
`capabilities
` Exchange length 0.035 inch (0.89 mm) soft, standard and
`extra-stiff guidewires
` Temporary pacemaker (PM) and pacing lead
` Sterile rinsing basins, physiological saline, heparinized
`saline, and 15% diluted radiopaque contrast medium
` 20 cc or larger luer-lock syringe
` 60 cc or larger luer-lock syringe
` High-pressure 3-way stopcock
` Edwards SAPIEN Transcatheter Heart Valve
` Ascendra Balloon Catheter
` 20 mm balloon valvuloplasty catheter such as Ascendra
`Balloon Aortic Valvuloplasty Catheter Model 9100BAVC
` Ascendra Introducer Sheath Set Model 9100IS
` Crimper Model 9100CR23 for 23 mm valve procedure
`and Model 9100CR26 for 26 mm valve procedure
`Inflation device provided by Edwards Lifesciences for this
`application
`7.2 Bioprosthesis Handling and Preparation
`Follow sterile technique during device preparation and
`implantation.
`7.2.1 Bioprosthesis Rinsing Procedure
`The bioprosthesis is packaged sterile in a plastic jar with a
`screw-cap closure and seal. Before opening, carefully
`examine the jar for evidence of damage (e.g. a cracked jar
`or lid, leakage, or broken or missing seals).
`CAUTION: Bioprostheses from containers found to be
`damaged, leaking, without adequate sterilant, or missing
`intact seals must not be used for implantation.
`
`
`
`Step Procedure
`1
`Set up two (2) sterile bowls with at least 500 mL of sterile
`physiological saline to thoroughly rinse the
`glutaraldehyde sterilant from the bioprosthesis.
`The bioprosthesis is contained in the jar within a holder.
`Carefully remove the bioprosthesis/holder assembly from
`the jar without touching the tissue. The holder is tagged
`with the bioprosthesis’ serial identification number.
`Inspect the bioprosthesis for any signs of damage to the
`frame or tissue.
`
`2
`
`3
`
`Rinse the bioprosthesis as follows:
`Place the bioprosthesis in the first bowl of sterile
`physiological saline. Be sure the saline solution
`completely covers the bioprosthesis and holder. With the
`bioprosthesis and holder submerged, slowly agitate (to
`gently swirl the bioprosthesis and holder) back and forth
`for a minimum of 1 minute. Transfer the bioprosthesis
`and holder to the second rinsing bowl of physiological
`saline and gently agitate for at least 1 more minute.
`Ensure the rinse solution in the first bowl is not used.
`The bioprosthesis should be left in the final rinse solution
`until needed to prevent the tissue from drying.
`CAUTION: Do not allow the bioprosthesis to come in
`contact with the bottom or sides of the rinse bowl
`during agitation or swirling of the bioprosthesis.
`Care must be taken to ensure that the identification
`tag does not come in contact with the tissue and
`damage it. No other objects should be placed in the
`rinse bowls. The bioprosthesis should be kept
`hydrated throughout the rest of the preparation
`procedure to prevent the tissue from drying.
`
`
`7.2.2 Prepare Transapical Procedure Components
`
`2
`
`3
`
`4
`
`5
`
`Step Procedure
`1
`Refer to Ascendra Introducer Sheath Set and
`Crimper instructions for use on device preparation
`and handling.
`Remove the balloon cover from the Ascendra
`Balloon Catheter
`Loosen the pusher nut and slide the pusher as far
`proximal as possible. Rotate the pusher nut to
`secure the pusher. Slide the loader cap, washers,
`and seal as far proximal as possible.
`CAUTION: Overtightening the pusher nut may
`result in improper balloon inflation.
`Prime and flush the guidewire lumen of the balloon
`catheter with heparinized saline.
`Insert an extra stiff guidewire (0.035” [0.89 mm] and
`≥ 100 cm long) in the guidewire lumen, leaving a 2 to
`3 cm segment of the guidewire protruding from the
`distal tip.
`Flush the balloon catheter with heparinized saline
`through the flush port.
`Attach extension tubing to balloon inflation port.
`Prepare a 60 mL or larger luer-lock syringe with
`diluted contrast medium (15:85 contrast to
`heparinized saline) and attach it to the balloon
`extension tubing.
`Completely fill the inflation device provided by
`Edwards Lifesciences with diluted contrast medium
`and attach to the balloon extension tubing.
`Close stopcock to inflation device. De-air the balloon
`catheter.
`Close the stopcock to the syringe. Insert the balloon
`into the balloon gauge located on the Crimper. Inflate
`the balloon and verify its diameter fits the gauge with
`minimal friction. While gently pulling and pushing the
`balloon, verify that the balloon moves with some
`resistance within the balloon gauge. If the balloon
`does not reach the correct diameter when fully
`inflated, add or discard some of the inflation solution
`in the inflation device provided by Edwards
`Lifesciences until the correct diameter is reached.
`
`6
`
`7
`8
`
`9
`
`10
`
`11
`
`4
`
`
`
`004
`
`ENDOHEART AG, EX. 2006 Page 4
`EDWARDS LIFESCIENCES CORPORATION (PETITIONER) v. ENDOHEART AG (PATENT OWNER)
`Case No.: IPR2016-00299, U.S Patent No. 8,182,530
`
`

`
`12
`
`Step Procedure
`The inflation device must remain connected to the
`balloon catheter throughout the rest of the
`procedure.
`
`Note: Correct balloon sizing is critical to successful
`valve deployment and valve function.
`Close stopcock to the balloon catheter and remove
`any remaining diluted contrast medium in the
`inflation device to the syringe. Lock the inflation
`device provided by Edwards Lifesciences.
`Close the stopcock to the syringe and verify the
`balloon is sized appropriately with the balloon gauge.
`Remove the syringe.
`Unlock inflation device provided by Edwards
`Lifesciences and deflate the balloon while creating a
`three-wing fold configuration, and ensure no diluted
`contrast medium is left behind. Lock the inflation
`device provided by Edwards Lifesciences.
`
`13
`
`14
`
`
`7.2.3 Mount and Crimp the Bioprosthesis on the Balloon
`Catheter
`
`Step
`1
`
`2
`
`3
`
`4
`
`5
`
`6
`
`7
`
`Procedure
`Remove the bioprosthesis from the holder and
`gently place the bioprosthesis into the crimper
`aperture.
`Gradually crimp the bioprosthesis to a diameter of
`approximately 12 mm.
`Remove the bioprosthesis from the crimper and
`place it on the balloon catheter with the inflow
`(fabric cuff end) of the bioprosthesis towards the
`proximal end of the balloon catheter. Center
`bioprosthesis between the radiopaque markers.
`Place the bioprosthesis back in the crimper
`aperture, and completely crimp until it fits inside the
`crimp gauge.
`CAUTION: The physician must verify correct
`mounting/orientation of the bioprosthesis prior
`to its implantation.
`Loosen the pusher nut and advance the pusher to
`align the pusher tip with the proximal end of the
`crimped bioprosthesis. Rotate the pusher nut to
`secure the pusher in place.
`CAUTION: Overtightening the pusher nut may
`result in improper balloon inflation.
`Flush the loader with sterile heparinized saline and
`slide the threaded end of the loader over the
`crimped bioprosthesis.
`Slide the washers and seal on the balloon catheter
`shaft distally to the pusher. Insert into loader.
`Ensure washers and seal are flat against each
`other within the loader to prevent leakage. Slide
`loader cap distally over balloon catheter so it sits
`flat against the washers and seal and rotate the
`loader cap onto the base of loader. Check that the
`thread is not exposed. This indicates that the loader
`cap and seal are fully engaged around the pusher
`tubing. Do not overtighten the loader cap.
`Note: The loader must fully cover the bioprosthesis.
`
`8
`
`9
`
`Re-flush the balloon catheter through the flush port
`and close stopcock to the balloon catheter.
`Note: Keep bioprosthesis hydrated until ready for
`implantation.
`Remove guidewire and flush guidewire lumen.
`
`Valvuloplasty and Bioprosthesis Delivery
`7.3
`Valvuloplasty and bioprosthesis delivery should be
`performed under general anesthesia with hemodynamic
`monitoring in a catheterization lab/hybrid operating room
`with fluoroscopic and echocardiographic imaging
`capabilities.
`Administer heparin to maintain the ACT at ≥ 250 sec.
`CAUTION: Use of excessive contrast media may lead to
`renal failure. Measure the patient’s creatinine level prior
`to the procedure. Contrast media usage should be
`monitored.
`7.3.1 Baseline Parameters
`
`2
`
`Step Procedure
`1
`Perform a supra-aortic angiogram with the projection of the
`native aortic valve perpendicular to the view.
`Evaluate the height between the inferior aspect of the annulus
`and the inferior aspects of the lowest coronary ostium for
`subsequent prosthetic aortic valve implantation.
`Introduce a pacemaker (PM) lead until its distal end is
`positioned in the right ventricle.
`Set the stimulation parameters, and test pacing.
`
`3
`
`4
`
`
`7.3.2 Valvuloplasty
`Refer to Ascendra Balloon Aortic Valvuloplasty Catheter
`Instructions for Use (IFU) for information on device
`preparation and handling.
`Note: Rapid ventricular pacing should be performed when
`using the Ascendra Balloon Aortic Valvuloplasty Catheter for
`valvuloplasty prior to aortic transcatheter valve implantation.
`After placement of the balloon at the intended site, begin
`rapid ventricular pacing. Once the blood pressure has
`decreased to 50 mmHg or below, balloon inflation can
`commence.
`CAUTION: Prosthetic valve implantation should not be
`carried out if the balloon cannot be fully inflated during
`valvuloplasty.
`
`7.3.3 Bioprosthesis Delivery
`
`Step Procedure
`Insert the introducer sheath. Refer to the Ascendra
`1
`Introducer Sheath Set IFU for additional information on
`device preparation and handling.
`
`2
`
`3
`
`Insert loader into the sheath until it locks. Tap lightly on
`the loader and loosen the loader cap to de-air. Tighten
`cap until loader is sealed and catheter can move with
`minimal resistance. Check that the thread is not
`exposed. This indicates that the loader cap and seal are
`fully engaged around the pusher tubing. Do not
`overtighten.
`Cross the native aortic valve and position the
`bioprosthesis within the diseased valve.
`
`5
`
`
`
`005
`
`ENDOHEART AG, EX. 2006 Page 5
`EDWARDS LIFESCIENCES CORPORATION (PETITIONER) v. ENDOHEART AG (PATENT OWNER)
`Case No.: IPR2016-00299, U.S Patent No. 8,182,530
`
`

`
`4
`
`5
`
`6
`
`7
`
`8
`
`9
`
`Loosen the pusher nut and retract the pusher, leaving
`the bioprosthesis in position. Rotate the pusher nut to
`secure the pusher. Verify that the pusher is completely
`off of the balloon before it is inflated and the
`bioprosthesis is deployed.
`CAUTION: The pusher must be pulled back for
`proper balloon inflation and bioprosthesis
`deployment.
`Position the mid-point of the bioprosthesis at the plane
`of the hinge points of the native valve leaflets.
`Verify the correct location of the bioprosthesis with
`respect to the calcified valve.
`Begin bioprosthesis deployment:
` Unlock the inflation device.
` Begin rapid pacing; once arterial blood pressure has
`decreased to 50 mmHg or below, balloon inflation
`
`can commence.
` Deploy the bioprosthesis by inflating the balloon with
`the entire volume in the inflation device. When the
`balloon catheter has been completely deflated, turn
`off the pacemaker.
`If deflection was used, straighten the catheter tip.
`
` Retract the balloon catheter into the introducer
`sheath.
`Disengage loader from sheath and remove balloon
`catheter.
`Remove sheath when the ACT level is appropriate (e.g.
`reaches < 150 sec). Close apical access site.
`
`
`8.0 How Supplied
`STERILE: The bioprosthesis is supplied sterilized with
`glutaraldehyde solution. The balloon catheter is supplied
`sterilized with ethylene oxide gas.
`8.1
`Storage
`The bioprosthesis must be stored between 10 C-25 C
`(50 F-77 F).Each jar is shipped in an enclosure containing
`a temperature indicator to detect exposure of the
`bioprosthesis to extreme temperature.
`The Ascendra Balloon Catheter should be stored in a cool,
`dry place.
`
`9.0 MR Safety
`
` MR Conditional
`Non-clinical testing has demonstrated that the Edwards
`SAPIEN THV (implant) is MR Conditional. It can be scanned
`safely under the following conditions:
`
`
`• Static magnetic field of 1.5 Tesla (T) or 3 Tesla.
`• Spatial gradient field of 2500 Gauss/cm or less.
`• Maximum whole-body-averaged specific absorption rate
`(SAR) of 2 W/kg for 15 minutes of scanning.
`• Normal mode operation, as defined in IEC 60601-2-33
`Ed. 3.0, of the MR system.
`In non-clinical testing and analysis, the implant was
`determined to produce a temperature rise of less than 1.1 °C
`above background for a whole body SAR of 2.0 W/kg for 15
`minutes of MR scanning in a 1.5 T cylindrical whole body
`
`MR system, assessed using a GE Signa whole body coil and
`a phantom designed to simulate human tissue. The phantom
`average SAR calculated using calorimetry was 2.2 W/kg and
`local background SAR at the site of the implant was 5.6
`W/kg. The measured rise above background was 0.7 °C for
`a whole body SAR of 2 W/kg in a 3.0 T cylindrical bore
`whole body MR system, assessed using a GE Signa HDx
`whole body active shield MR scanner with software version
`14/LX/MR and a phantom designed to simulate human
`tissue. The phantom average SAR calculated using
`calorimetry was 2.9 W/kg and local background SAR at the
`site of the implant was 8.4 W/kg
`The image artifact extended as far as 15 mm from the
`implant for spin echo images and 40 mm for gradient images
`when scanned in non-clinical testing in a 3.0 T GE Signa
`HDx MR system. The implant has not been evaluated in MR
`systems other than 1.5 or 3.0 T.
`
`10.0 Patient Information
`A patient registration form is provided with each
`transcatheter heart valve. After implantation, all requested
`information should be completed on this form. The serial
`number may be found on the package and on the
`identification tag attached to the transcatheter heart valve.
`The original form should be returned to the Edwards
`Lifesciences address indicated on the form and provide the
`temporary identification card to the patient prior to discharge.
`
`11.0 Recovered Clinical
`Bioprosthesis
`The explanted bioprosthesis should be placed into a suitable
`histological fixative such as 10% formalin or 2%
`glutaraldehyde and returned to the company. Refrigeration is
`not necessary under these circumstances. Contact
`Edwards Lifesciences to request an Explant Kit.
`Disposal of Used Devices
`Used devices may be disposed of in the same manner that
`hospital waste and biohazardous materials are handled.
`There are no special risks related to the disposal of these
`devices.
`
`12.0 Clinical Studies
`
`The Placement of Aortic Transcatheter Valves
`(PARTNER) trial, a prospective, randomized-
`controlled, multi-center pivotal trial, evaluated the
`safety and effectiveness of the Edwards SAPIEN
`Transcatheter Heart Valve via transfemoral and
`transapical delivery in a stratified population of high-
`risk and inoperable patients with severe symptomatic
`native aortic stenosis. Patients were stratified into two
`cohorts based on their risk of operability for standard
`aortic valve replacement surgery – those who were
`considered high surgical risk were eligible for Cohort
`A, while inoperable patients were eligible for Cohort B
`due to coexisting conditions that resulted in the
`probability of death or irreversible morbidity exceeding
`50%.
`
`
`
`6
`
`
`
`
`
`006
`
`ENDOHEART AG, EX. 2006 Page 6
`EDWARDS LIFESCIENCES CORPORATION (PETITIONER) v. ENDOHEART AG (PATENT OWNER)
`Case No.: IPR2016-00299, U.S Patent No. 8,182,530
`
`

`
`Study Design – Cohort A
`
`This was a randomized study with the primary
`objective of ascertaining if TAVR is non-inferior to
`AVR surgery with respect to 12-month survival
`outcomes in high-risk surgical patients. Other
`objectives were focused on characterizing the benefit
`to risk ratio of TAVR relative to AVR.
`
`Patients in Cohort A were first evaluated for vascular
`access to determine whether their peripheral arteries
`could accommodate the 22 or 24 French sheaths
`required for the transfemoral TAVR approach to
`deliver the 23 mm or 26 mm Edwards SAPIEN valve
`sizes. Those patients who could accommodate these
`sheaths were then randomized 1:1 between
`transfemoral TAVR and surgical AVR. Those patients
`whose arteries could not accommodate these sheaths
`were randomized 1:1 between transapical TAVR and
`surgical AVR.
`
`The primary study endpoint was based on a pooled
`transapical and transfemoral analysis, and was
`defined as freedom from all-cause mortality at one
`year for the high-risk cohort. All patients were
`followed for at least 1 year, and cross-over from the
`surgical AVR group to the TAVR group was not
`permitted, except when findings or events during the
`assigned procedure prevented the planned
`treatment. Prespecified secondary endpoints included
`the following: time from randomization to the first
`occurrence of a Major Adverse Cardiac and
`Cerebrovascular Event (MACCE) within one year for
`which MACCE definition was comprised of death, MI,
`stroke, and renal failure as defined by protocol, total
`hospital days through one year, NYHA functional
`class at one year, and 6-minute walk test at one year.
`Additional prespecified efficacy endpoints were
`measured at 30 days, six months, and one year for
`the following: functional improvement from baseline
`as measured per (1) NYHA functional classification,
`(2) EOA, and (3) 6-minute walk test, freedom from
`MACCE, improved Quality of Life (QoL), and
`improved valve function demonstrated by an
`improvement in EOA.
`
`Study Design – Cohort B
`
`This was a randomized study with the primary
`objective of ascertaining if TAVR is superior to
`standard therapy in a control group for inoperable
`patients with respect to 12-month survival outcomes.
`Other objectives were focused on characterizing the
`benefit to risk ratio of TAVR relative to the standard
`therapy control group.
`
`Patients in Cohort B were also evaluated for vascular
`access and those meeting the criteria were
`randomized 1:1 to either transfemoral delivery of the
`Edwards SAPIEN valve or to a control group. Patients
`in the control group were treated with medication
`and/or balloon valvuloplasty. Patients in Cohort B who
`
`did not meet the criteria for vascular access were not
`eligible for the trial.
`
`Study Results – Cohort A
`
`A total of 699 (657 in the As-Treated [AT] population)
`high-risk patients with severe aortic stenosis were
`enrolled at 26 centers (23 in the United States) and
`assigned to TAVR (344 patients) or AVR (313
`patients) with baseline characteristics described in
`Table 1. Among the TAVR patents, 240 were treated
`using transfemoral access and 104 were treated
`using transapical access. Severe aortic stenosis was
`defined as a mean gradient > 40 mmHg, jet velocity >
`4.0 m per sec, or an initial aortic valve area (AVA) of
`0.8 cm2. The primary endpoint for the high-risk cohort
`was freedom from all-cause mortality at one year.
`Clinical outcomes of TAVR (transfemoral and
`transapical) as compared to AVR are summarized in
`Tables 5, 6, and 7. At day 365, the Kaplan-Meier
`estimate of all-cause death was 23.7% in the TAVR
`group, as compared to 25.2% in the AVR group. The
`estimated difference between these treatment groups
`is -1.5% with a one-sided lower 95% confidence
`interval of -4.0%, which is greater than the pre-
`specified margin of -7.5%. The non-inferiority p-value
`for this difference is 0.0037, indicating that TAVR is
`non-inferior to AVR with respect to all-cause death
`[Figure 3]. Pre-specified secondary endpoints
`included valve performance [Figures 6 and 7] and
`NYHA functional class [Figure 8]. When interpreting
`NYHA results, consider that the evaluation was
`unblinded. As with other heart valve trials, the
`patients are aware of their treatment group.
`Accordingly there is the potential for bias in the NYHA
`v

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket