throbber
SCOTT A. MIROWITZ, MD, CONSULTING EDITOR
`
`MAGNETIC
`RESONANCE
`IMAGING
`CLINICS OF NORTH AMERICA
`
`Physics of MR Imaging
`
`J. PAUL FINN, MD, GUEST EDITOR
`
`VOLUME 7
`
`•
`
`NUMBER4
`
`•
`
`NOVEMBER 1999
`
`W.B. SAUNDERS COMPANY
`A Division of Harcourt Brace & Company
`PHILADELPHIA
`LONDON
`TORONTO
`
`MONTREAL
`
`SYDNEY
`
`TOKYO
`
`General Electric Co. 1005 - Page 1
`
`

`
`W.B. SAUNDERS COM PANY
`A Division of HarcoHrt Brace & Company
`
`The Curtis Center • Independence Square West • Philadelphia, Pennsylvania 19106
`http://www .wbsaunders.com
`
`MRI CLINICS OF NORTH AMERICA
`November 1999
`Bditor: Barton Dudlick
`
`Volume 7, Number 4
`ISSN 1064-9689
`Production Editor: Mkhael Pittard
`
`Copyright © 1999 by W.8. Saunders Company. All rights reserve d. No part of this publication may
`in any form or by any means, electronic or mechanical, including
`be reproduced or transmjtted
`photocopy, recording, or any information retrieval system, without permission from the Publisher .
`
`The appearance of the code at the top of the first page of an article in this periodical indicates the
`consent of the W.B. Saunders Company that copies of the article may be made for personal or
`interna l use, or for the personal or internal use of specific clients, for those registered with the Copyright
`Clearance Center, Inc. (222 Rosewood Drive, Danvers, MA 01923; (978) 750-8400; wunv.copyright.com).
`This consent is given on the condition that the copier pay the stated per-copy fee for tltat article
`through the Copyright Clearance Center, Inc. for copying beyond that permitted by Sections 107 or
`108 of the US Copydght Law. This consent does not exten~ to other kinds of copying, such as
`copying for genera l distribution, for advertising or promotional purposes, for creating new collective
`works, or for resale. Absence of the code indicates that the material may not be processed through
`the Copyright Clearance Center, Inc.
`
`The ideas and opinions exp ressed in Magnetic Resonance Imaging Clinics of North America. do not
`necessarily reflect those of the Publisher. The Publisher does not assume any responsibility for any
`injury and I or damage to persons or property arising out of or related to any use of the material
`contained in this periodical. The reader is advised to check the appropriate medical literature and
`the product information currently provided by the manufacturer of each drug to be administered to
`verify the dosage, the method and duration of administration, or contraindications. It is the responsi(cid:173)
`bi_Uty of the treating physician or other health care professional, relyi ng on independent experience
`and knowledge of the patient, to determine drug dosages and the best treatment for the patient.
`Mention of any product in this issue should not be construed as endorsement by the cont ribut ors,
`editors, or the Publisher of the product or manufacturers' claims.
`
`Magnetic Resonance Imaging Clinics of North America (ISSN 1064-9689) is published quarterly by the
`W.B. Saunders Company. Corporate and editorial offices: The Curt is Center, Independence Square
`West, Philadelphia, PA 19106-3399. Accounting and circulation offices: 6277 Sea Harbor Drive,
`Orlando, FL 32887-4800. Periodicals postage paid at Orlando, FL 32862, and addit ional mailing
`offices. Subscription prices are $136.00 per year (US individuals), $173.00 per year (US institutions),
`$153.00 per year (Canad ian individuals) , $209.00 per year (Canadian institutions), $183.00 per year
`(foreign individuals), and $209.00 per year (foreign institutions). Foreign air speed delivery
`is
`included in all Clinics subscription prices. All prices are subject to change without notice. POSTMAS(cid:173)
`TER: Send address changes to Magnetic Resonance Imaging Clinics of Nort/1 America, W.B. Saunders
`Company, Per iodicals Fulfillment, Orlando, FL 32887-4800. Cus tomer Service: 1-800-654-2452 (US).
`From outside of the US, call 1-407-345-4000.
`
`Magnetic Resonance Imaging Clinics of North America is covered in the RSNA Index of Imaging Literature,
`Index Medic11S, MEDLINE, and EMBASE!Excerta Medica.
`
`Printed in the United States of America.
`
`General Electric Co. 1005 - Page 2
`
`

`
`PHYSICS OF MR IMAGING
`
`1064-9689/99 $8.00 + .00
`
`OVERVIEW OF MR IMAGING PULSE
`SEQUENCES
`
`John P. Mugler ill, PhD
`
`The tremendous contrast flexibility of MR
`imaging results from the literally unlimited
`variety of pulse-sequence
`techniques. Even
`though MR imag ing is, in many ways, a ma(cid:173)
`ture imaging modality, pulse sequences con(cid:173)
`tinue to be developed that provide significant
`improvements for existing applications or im(cid:173)
`portant new capabilities. Nonetheless, every
`pulse sequence must accomplish two basic
`tasks. First, transverse magnetization must be
`created using one or more radiofrequency
`(RF) pulses, and this magnetization must then
`be encoded using gradient pulses so that the
`spatial positions of the tissues can be deter(cid:173)
`mined to form the image. Second, the desired
`contrast between
`tissues must be created
`based on the time of application and duration
`of the various RF and gradient pulses. In
`this article, after a review of general pu lse(cid:173)
`sequence properties, the features of the tech(cid:173)
`niques most commonly used for clinical MR
`imaging are discussed. In particu lar, the man(cid:173)
`ner in which each technique generates and
`encodes transverse magnetization
`and the
`typical forms of image contrast available from
`each technique are emphasized.
`
`GENERALPULS&SEQUENCE
`PROPERTIES
`
`The Pulse-Sequence Timing Diagram
`
`the im(cid:173)
`A useful method for conveying
`portant characteristics of a pulse sequence is
`a graphical representation called the pulse(cid:173)
`seq uence timing diagram. These diagrams
`will be used extensively when specific tech(cid:173)
`niques are described. The major features of
`timing diagrams will be explained using a
`single spin-echo pulse sequence as an exam(cid:173)
`ple (Fig. 1).
`In general, the timing diagram depicts the
`temporal sequence of the RF, gradient, and
`data-sampling events in the pulse sequence.
`The first line, labeled RF, illustrates the time
`of application and the waveforms for the RF
`pulses, in this case 90° (excitation) and 180°
`(refocusing) pulses. The next three lines de(cid:173)
`pict the time of application and the wave(cid:173)
`forms for gradients applied along three mutu(cid:173)
`ally orthogonal axes. These are labeled Gss,
`GrEJ and GRo, for the section-select, phase(cid:173)
`encoding, and readout gradients, respectively,
`although other labels, for example Gx, G~ Gz,
`
`From the Departm ents of Radiology and Biomedi cal Engin eering, University of Virginia School of Medi cine, Charlottes(cid:173)
`ville, Virginia
`
`MRI CLINICS OF NORTH AMERICA
`
`VOLUME 7 • NUMBER 4 • NOVEMBER 1999
`
`661
`
`General Electric Co. 1005 - Page 3
`
`

`
`662
`
`MUGLER
`
`RF
`
`Gss
`
`Figure 1. Pulse-sequence timing for a single spin-echo technique. RF =
`radiofrequency pulses; Gss = section-select gradient; GPe = phase(cid:173)
`encoding gradient; GRo = readout gradient; DAO = data acquisition; N20
`, = number of times that the basic pulse-sequence timing is repeated,
`which, in this case, equals the number of in-plane phase-encoding lines;
`TR = repetition time; and TE = echo time.
`
`\
`
`can be used to indicate the assignment of the
`gradients for a particular section orientation.
`For the two-dimensional (20) acquisition de(cid:173)
`picted in Figure 1, section-select gradients are
`applied in synchrony with the RF pulses, thus
`spatially localizing their effects, or in other
`words selecting the section of interest. The
`other gradient waveforms, applied along the
`phase-encoding and readout axes, spatially
`encode the magnetization within this section
`along the remaining two directions. The gra(cid:173)
`dient waveform on the phase-encoding axis,
`containing a series of closely spaced hori(cid:173)
`zontal lines, is the graphical representation of
`a gradient table: the strength of the gradient
`is stepped through a series of values as the
`basic pulse-sequence
`timing is repeated
`to
`collect the data required to form the image.
`With each repetition, another line of data is
`collected. The fifth line in the timing diagram
`shows when the MR signals occur. Although
`some pulse sequences generate a number of
`distinct signals, typically only the signals of
`interest are depicted. The last line of the tim(cid:173)
`ing diagram, labeled OAQ for data acquisi(cid:173)
`tion, illustrates the time period during which
`the data are collected. This line is sometimes
`labeled ADC, for analog-to-digital convertor.
`
`It is not uncommon for either or both of the
`signal and DAQ lines to be omitted. Finally,
`the square brackets around all of the timing
`events with the N 20 label in the lower right
`comer indicate that these events are repeated
`N20 times.
`
`Modes of Spatial Encoding
`
`MR images are typically acquired either as
`a set of 20 sections or as a three-dimensional
`(30) volume (Fig. 2). In a 30 acquisition, an
`additiona l phase-encoding gradient table is
`applied along the third dimension to yield a
`set of contiguous sections within the volume .
`These sections are often referred to as parti(cid:173)
`tions to distinguish them from image sections
`derived from a fundamentally 20 acquisition.
`The advantages of a 30 acquisition include
`the availability of thin, contiguous sections
`that can be used to obtain high -resolution
`images of arbitrary orientations through post(cid:173)
`processing11, 72. 112, and, for pulse-sequence pa(cid:173)
`rameter values that are equivalent in the 20
`and 30 acquisitions, an increase in the signal(cid:173)
`to-noise ratio (SNR) by a factor equal to the
`square root of the number of partitions. 106 A
`
`General Electric Co. 1005 - Page 4
`
`

`
`OVERVIEW OF MR IMAGING PULSE SEQUENCES
`
`663
`
`20 sections
`
`30 single-slab
`
`30 multi-slab
`
`Figure 2. The three basic modes of spatial encoding used in MR imaging: two-dimensional
`(20) sections, three-dimensional (30) single-slab, and 30 multislab. This example illustrates
`the possibilities for acquiring axial images of the brain. Acquisitions using 20 sections or 30
`multislab often employ a gap between the sections or slabs, as shown, to increase the
`anatomical coverage and to decrease crosstalk (significant degree of interference) effects,
`although the use of contiguous sections or slabs (i.e., no gaps) is appropriate for some
`applications.
`
`hybrid of the 20 and 30 acquisition modes,
`called multislab
`30,
`is also sometimes
`133 For this acquisition a set of slabs is
`used. 103
`•
`acquired, and each slab in tum is phase en(cid:173)
`coded along the third dimension to yield a
`set of contiguous partitions (Fig. 2). With ref(cid:173)
`erence to this mode, single -volume 30 acqui(cid:173)
`sition is sometimes referred to as single-slab
`30.
`The 20 mode of acquisition can be subdi(cid:173)
`vided into two important forms: sequential
`section and multisection. 19 In a sequential(cid:173)
`section acquisition, which is analogous to the
`operation of conventional computed
`tomog(cid:173)
`raphy (CT), all of the data required for a
`given image are collected before proceeding
`to the next image. A multisection acquisition,
`however, collects only a portion of the re(cid:173)
`quired data for a given image, typically one
`or several phase-encoding
`lines, before col(cid:173)
`lecting the corresponding data for each of the
`other images. This process is repeated until
`all of the data for all of the images have been
`
`collected (Fig. 3). Obviously, a given image
`can be acquired more rapidly using the se(cid:173)
`quential-section method, and this method is
`typically used when short acquisition times
`are critical, for example, in some implementa (cid:173)
`tions of MR angiography or for imaging rap (cid:173)
`idly moving structures. On the other hand, a
`multisection
`acquisition
`is typically used
`when a relatively long time period (on the
`order of 1 second) is required between the
`interrogations of a particular section to permit
`Tl recovery to occur and thereby generate the
`desired image contrast.
`Another practical aspect of sequential-section
`and multisection acquisitions is the temporal
`order in which the data for the sections are
`collected. This may be important, for exam(cid:173)
`ple, if there is blood flow perpendicular to the
`sections, 2• 7 or if there is a significant degree of
`(crosstalk) between sections. 71• 74
`interference
`The most common acquisition orders are con(cid:173)
`secutive, beginning at either end of the image
`set, and interleaved, for example collecting all
`
`General Electric Co. 1005 - Page 5
`
`

`
`SECTION 1
`
`SECTION 2
`
`SECTION 3
`
`664
`
`MU CLER
`
`DATA
`SEGMENT'
`
`A
`
`2
`
`3
`
`SEQUENCE REPE:TIONEJ
`
`SEQUENCE
`
`REPE:TIO~Ej ~ ~ EJ § g Ej ~ ~
`
`Figure 3. The order of data acqu isition for 20 multisect ion and sequential-sect ion imaging. A
`hypothetica l case is illustrated ' where three images are to be acqu ired, and the k-space data
`corresponding to each image is com posed of three segments . One data segment of an image section
`is acquired during each repetition of the pulse sequence , thus nine repetitions are required to collect
`the data for all three images. A, The k-space data layout corresponding to the three images. B, The
`data ac,quisition order for 20 multisection imaging. C, The data acquisi tion orde r for 20 sequential(cid:173)
`section imaging.
`
`odd-numbered sections followed by all even(cid:173)
`numbered sections, although arbitrary acqui(cid:173)
`sition orders can be used. As a final point,
`the sequential-section and multisection con(cid:173)
`cepts are also applicable to multislab 30 ac(cid:173)
`quisitions.
`Acquisitiorn modes other than the 20 and
`30 forms described above are sometimes
`used. For example, a thin column of tissue
`can be imaged by using either two RF pulses
`that excite intersecting planes 30 or a single 20
`RF pulse .109 Applications of this mode include
`analysis of cardiac function 110 and rapid dilfu(cid:173)
`sion imaging. 8 1
`
`Steady-Statei Versus Transient
`Acquisitions :
`
`Many of the MR pulse sequences in clinical
`use operate in the steady-state mode in which
`severa l repeti .tions of the basic pulse-sequence
`
`timing are required to collect the necessary
`data (see Fig. 1). These repetitions are spaced
`uniformly in time, so that the amplitude and
`phase of the magnetization when the data are
`collected are the same in all repetitions, aside
`from the effects of the phase-encoding gradi(cid:173)
`ent. Anything that disrupts this steady state
`may result in undesired variations in the am(cid:173)
`plitude or phase of the magnetization and, as
`a consequence, may produce image artifacts.
`Such disruptions
`include heart rate-depen(cid:173)
`in the repetition
`time (TR)
`dent variations
`during electrocardiogram (ECG)-triggered ac(cid:173)
`quisitions.
`In contrast with a steady -state acquisition,
`some pulse sequences operate in a transient
`mode in which the amplitude or phase of the
`magnetization varies throughout
`the cou1·se
`of data collection. Such methods of acquisi(cid:173)
`tion are commonly referred to as single-shot
`methods.
`Several important pulse sequences operate
`
`General Electric Co. 1005 - Page 6
`
`

`
`OVERVIEW OF MR lMAGlNG PULSE SEQUENCES
`
`665
`
`in a mode that is a hybrid between the pure
`steady -state and transient acquisitions.
`In
`these techniques,
`the basic pulse-sequence
`timing is repeated several times, just as in a
`steady-state acquisition. During each repeti(cid:173)
`tion, however, the data collection for a given
`image spans a period during which the am(cid:173)
`plitude or phase of the magnetization varies
`substantially, for example because of relax(cid:173)
`ation or the application of RF pulse:s. Hence,
`the data that are collected reflect these varia(cid:173)
`tions. These acquisition methods are often re(cid:173)
`ferred
`to as segmented methods. 25 When
`pulse sequences are described in this article,
`the temporal mode of the acquisition is desig(cid:173)
`single-shot .. or seg(cid:173)
`nated as steady-state,
`mented.
`The total time required to genera 1te an im(cid:173)
`age or set of images with a selected contrast
`behavior is often one of the most important
`characteristics of the pulse sequence. For most
`steady-state and segmented
`techniques,
`the
`imaging time can be calculated from
`Imaging Time = TR-NREl'·NAc
`where NREP is the number of times that the
`basic pulse-sequence
`timing is repeated
`to
`permit collection of the data required to form
`the image, and N AC is the number of times
`that the complete data collection process is
`repeated to perform averaging for reduction
`of the noise level. The term NAo (number of
`acquisitions), is also commonly denoted NEX,
`for number of excitations. For many steady(cid:173)
`state pulse sequences, NREP is equal to N20 N30 ,
`where N20 is the number of in-plallle phase(cid:173)
`encoding lines, and NJo is the number of par(cid:173)
`titions within the volume for single-slab 30
`or within each subvolume for multislab 30.
`For 20 acquisitions, N30
`is set equal
`to 1.
`Segmented acquisitions typically acquire sev(cid:173)
`eral phase-encoding
`lines for each repetition
`of the basic pulse-sequence timing, and hence
`the value of NREP must be reduced by an
`appropriate factor. For example, if the num(cid:173)
`ber of phase -encoding lines acquired per rep(cid:173)
`etition is denoted ETL (for echo train length),
`then NREP becomes (N2oNJo)/ ETL.
`A further distinction concerning the tempo(cid:173)
`ral characteristics of a pulse sequence can be
`made by whether the tasks of spatial encod-
`
`(1)
`
`ing and contrast generation are intimately
`coupled or separable. In many steady-state
`pulse sequences, the same set of RF and gra(cid:173)
`dient pulses that generate the contrast also
`spatially encode the magnetization. On the
`other hand, the contrast generation in many
`transient pulse sequences is performed by a
`series of RF and gradient pulses that is dis(cid:173)
`tinct from those that perform spatial encod (cid:173)
`ing. Pulse sequences that use a separate con(cid:173)
`trast preparation
`are often referred
`to as
`49• 93 Of
`magnetization -prepared
`techniques.
`course, the contrast can also be generated in
`a hybrid fashion, using the combined effects
`of a magnetization preparation and the por(cid:173)
`tion of the pulse sequence that performs spa(cid:173)
`tial encoding.
`
`The Effects of Radiofrequency
`Pulses on Magnetization
`
`To facilitate discussion of the various pulse
`sequences, it is useful to review the ways in
`which an RF pulse can affect the magnetiza(cid:173)
`tion. 130 Consider an RF pulse with a flip angle
`e applied to a longitudinal component of the
`magnetization (M2 ). Just after this RF pulse,
`a fraction of the original magnetization
`(Mzcos0) remains along the longitudinal axis,
`and transverse magnetization
`(of amplitude
`M2sin0) is also created. Note that transverse
`magnetization
`is created for any value of 0
`except for an integer multiple of 180°.
`The effects of an RF pulse on a transverse
`component of the magnetization
`(MXY) are
`more complex. Just after the RF pulse, a frac(cid:173)
`tion of the original magnetization [Mxycos2(0/
`2)] remains
`in the
`transverse
`plane, un(cid:173)
`affected by the application
`of the pulse.
`Another
`portion
`of
`the magnetization
`(Mxvsin 2(0 /2)] remains
`in the
`transverse
`plane, but the phase state of this magnetiza(cid:173)
`tion is reversed (conjugated) by the applica(cid:173)
`tion of the pulse. It is this latter portion of
`the magnetization
`that can evolve to form a
`spin echo. Note that phase-reversed
`trans(cid:173)
`verse magnetization
`is created for any value
`of O except for an integer multiple of 360°. In
`addition
`to producing
`two transverse com-
`
`General Electric Co. 1005 - Page 7
`
`

`
`666
`
`MUGLER
`
`ponents, an RF pulse applied to transverse
`magnetization
`also results in two distinct
`components of longitudinal magnetization.
`Both components have the same amplitude
`[Mxv(sin0)/2], and they both retain the phase
`inform ation of the transverse magnetization.
`The difference between
`th ese two compo(cid:173)
`nents is that the phase information
`in one
`component
`is reversed, analogous
`to the
`phase-reversed transverse magnetization, and
`the phase information in the other component
`is not. If the phase-reversed
`longitudinal
`magnetization
`is converted
`into transverse
`magnetization by a subsequent RF pulse, it
`can evolve to form a stimu lated echo.37
`Radiofrequen cy pulses are often given a
`qualitative label based on their desired effect.
`For example, pulses that have the purpose of
`creating transverse magnetization are gener(cid:173)
`ally called excitation pulses, and pul ses that
`have the purpose of reversing the phase state
`of transverse magnetization
`are generally
`called refocusing pulses. Inversion pulses are
`intended
`to convert positive
`longitudinal
`magnetization
`into negative
`longitudinal
`magnetization without creating any trans(cid:173)
`verse component. The flip angle of such a
`pulse is thus 180°. Note, how ever, as can be
`deduced
`from the relationships discussed
`previously, that any excitation pulse poten(cid:173)
`tially performs some degree of refocusing,
`and any refocusing pulse (except integer mul(cid:173)
`tiples of 180°) potentially performs some de(cid:173)
`gree of excitation. This behavior leads to a
`large number of higher-order echoes when
`more than a small number of RF pulses are
`applied in a time period that is short com(cid:173)
`pared with the tissue relaxation times.58 In
`this regard, pulse seq uences that employ
`many closely spaced RF pul ses must be care(cid:173)
`fully designed to ensure that only the desired
`echo signals contribute to the image.
`
`k-Space Trajectories and Filters
`
`The MR signals recorded from the patient
`are actually the spatia l-frequency space, or k(cid:173)
`space, description of the image. Therefore,
`the desired image is formed by a Fourier
`transform of these k-space data. An important
`
`characteristic of many pulse sequences is the
`temporal order, or arrangement, used to col(cid:173)
`lect the k-space data. 9• 80•
`124 This order of data
`collection is commonly referred to as the k(cid:173)
`space trajectory. Foir example, in conventional
`spin-echo imaging, the k-space data are typi(cid:173)
`ca 11 y collected sequentially,
`line by line,
`moving from one ,extreme of k-space to the
`opposite extreme. In contrast, fast imaging
`techniques often use other types of k-space
`trajectories, such as spirals. 91 In principle, any
`form of k-space trajectory can be used if two
`conditions are met. First, the region of k-space
`that is covered must be sufficiently large to
`describe the anatomy of interest with the de(cid:173)
`sired spati al resolultion. Second, the minimum
`spacing between data samples must be suffi(cid:173)
`ciently small
`to prevent aliasing
`(wrap(cid:173)
`around) artifacts. Major factors affecting the
`selection of the k-space traje_ctory include the
`speed of image acquisition, ·the desired con(cid:173)
`trast, and the level of artifacts.
`The relationship of the k-space trajectory
`to the image contrast behavior and potential
`artifacts can be understood by reviewing
`some of the basic properties of k-space. The
`data in the central portion of k-space, that is,
`the low spatiaJ-fre,quency data, represent the
`gross structure and contrast
`in the image.
`Conversely, the data in the outer portions
`of k-space, that is, the high spatial-frequency
`data, represent the detailed structure and con(cid:173)
`trast in the image. The position of the k-space
`data relative to the center corresponds to the
`orientation of the structures that the data rep(cid:173)
`resent in the image. These principles are illus(cid:173)
`trated in Figure 4. For example, in Figure 4C
`only k-space data for high-spatial frequencies
`are present. Thus, only features with rapidly
`changing intensity values (that is, edges, par(cid:173)
`ticularly in the vicinity of the bright subcuta(cid:173)
`neous fat) are expected to be seen in the cor(cid:173)
`responding image .. Further, only k-space data
`in the vicinity of the horizontal axis are in(cid:173)
`cluded. Therefore, only edges oriented paral(cid:173)
`lel to the vertical direction are visible.
`Recall, as discussed previously, that the am(cid:173)
`plitude and phase: of the magnetization gen(cid:173)
`erally vary thrornghout the course of data
`collection for a single-shot or segmented ac(cid:173)
`quisition and may also vary in a steady-
`
`General Electric Co. 1005 - Page 8
`
`

`
`OVERVIEW OF MR IMAGING PULSE SEQUENCES
`
`667
`
`~lgure 4. The relationship between the data location in k-space and the resulting features in the
`image. The k-space data (upper row) and corresponding images (lower row) are shown for (A) the
`complete 256 x 256 data set, (B) the central 64 x 64 data points, (C) two 96 x 32 data blocks
`centered on the horizontal axis, anc! (D) two 32 x 96 data blocks centered on the vertical axis. The
`k-space data is displayed using a k>garithmic intensity scale to provide an improved visualization of
`the high spatial-frequency compone,nts.
`
`state acquisition because of effects such as
`motion. The extent of these variations de(cid:173)
`pends on the details of the pulse sequence
`and the physical properties, such as the relax(cid:173)
`ation times, of the anatomy. Co:nsidering
`these potential signal variations and the rela(cid:173)
`tionships between k-space and image, it is
`obvious that the relative amplitude .and phase
`of the data may vary as a function of the
`location in k-space. This variation as a func(cid:173)
`tion of spatial frequency is often ca:Ued the k(cid:173)
`space filter of the acquisition. Depcmding on
`the form of this filter, as determined by the
`k-space trajectory, image artifacts such as
`blurring, edge enhancement, or gho,sting may
`result. Figure 5 shows an example ,of k-space
`filter effects that could be caused by T2 relax(cid:173)
`ation in a segmented acquisition. In Figure
`SB, the k-space filter attenuates high spatial(cid:173)
`frequency components along one direction.
`Because these data correspond
`to detailed
`structure, blurring is seen in the correspond(cid:173)
`ing image along the same directio,n. On the
`other hand, the k-space filter in Figure SC
`enhances the high spatia l-frequency compo(cid:173)
`nents, and the corresponding image demon(cid:173)
`strates edge enhancement.
`
`Criteria for Pulse Sequence
`Evaluation
`
`To choose the best pulse sequence for a
`particular imaging application, it is necessary
`to evaluate the pertinent features of the avail(cid:173)
`able techniques. The major features of impor(cid:173)
`tance include (1) the speed of image acquisi(cid:173)
`tion; (2) the SNR, which is affected by the
`signa l levels and the acquisition (receiver)
`bandwidth; (3) the available image contrasts,
`for example Tl, T2, T2*, proton-density, or
`diffusion weighting; (4) the sensitivity to off(cid:173)
`resonance signals caused by static-field inho(cid:173)
`mogeneities, magnetic-susceptibility
`inter(cid:173)
`faces, or chemical shift; (5) the sensitivity to
`flow or motion; (6) the sensitivity to hardware
`performance; and (7) the complexity of the
`reconstruction algorithms . The descriptions of
`the various pulse sequences highlight the fea(cid:173)
`tures that are a particular advantage or disad(cid:173)
`vantage of each technique.
`
`Pulse Sequence Acronyms
`
`It is very common for MR techniques to
`be referred to by acronyms that are typically
`
`General Electric Co. 1005 - Page 9
`
`

`
`668
`
`MUGLER
`
`A
`
`B
`
`c
`
`SPATl !AL FREQUENCY
`
`SPATIAL FREQU ENCY
`
`SPATIAL FREQU ENCY
`
`Figure 5. Simulation of k-space filter effects. The filter functions (upper row), filtered k·space data
`(middle row), and the corresponding images (bottom row) are shown for (A) an ideal, uniform k(cid:173)
`space filte r, and (B) an attenuation or (C) an enhancement of the high spatial-frequency components
`along one axis of k-space. The images in (B) and (C) were generated by applying the respective,
`simulated k-space filters to the experimentally measured k-space data for (A). The k-spaoe data is
`displayed! using a logarithmic intensity scale to provide an improved visualization of the high spatial(cid:173)
`frequency components.
`
`derived from major structural or operational
`features of the pulse sequence . This discus(cid:173)
`sion of the techniques most commonly used
`for clinical MR imaging uses generic terms,
`when possible, to describe the features and
`functions of each technique but also provides
`some of the more common acronyms. 28 Note,
`hovvever, that although pulse-sequence acro(cid:173)
`nyms can serve as useful abbreviations, they
`can also be obstacles for the novice MR user
`who is trying to learn the array of available
`techniques and how they work. This is partic(cid:173)
`ularly true for many of the acronyms used by
`the equipment manufacturers, because totally
`
`different names are often used for basically
`identical techniques. Most clinically applica(cid:173)
`ble pulse -sequence
`techniques are available
`on the MR platforms of all manufacturers,
`and, even though the names are different, the
`underlying physical principles of any given
`technique are the same, regardless of the im(cid:173)
`aging platform.
`
`SPIN-ECHO-BASED PULSE
`SEQUENCES
`
`The review of spin -echo (SE) imaging be(cid:173)
`gins with the basic single- and double-SE
`
`General Electric Co. 1005 - Page 10
`
`

`
`OVERVTEW OF MR IMAGING PULSE SEQUENCES
`
`669
`
`puJse sequences that are the foundation for
`many of the more complex SE-based meth(cid:173)
`ods. Next, a family of rapid SE techniques are
`presented and followed by a discussion of the
`inversion-recovery contrast preparation
`that
`is used with a variety of SE-based methods.
`Finally, a class of ultra-fast SE techniques is
`described.
`
`Convention al Spin-Echo Imaging
`
`Only a few years ago, most clinical MR
`imaging was performed using what is now
`called conventional SE imaging . The word
`conven tional was added to the name recently
`to distinguish this form of SE-based imaging,
`in which one phase-encoding line of k-space
`data is collected per image with each excita(cid:173)
`tion RF pulse, from the more rapid methods
`that collect several phase-encoding lines with
`each excitation pulse. Even with the advent
`of faster SE-based techniques, however, con(cid:173)
`ventional SE is still widely used beca~se of
`its well understood contrast characteristics
`and its resistance to image artifacts from a
`variety of sources such as RF or static-field
`inhomogeneities.
`The timing diagram for one of the work(cid:173)
`horses of conventional SE imaging, the dou-
`
`ble-echo puJse sequence, is shown in Figure
`6. As illustrated in the figure, the RF pulse
`scheme includes an excitation RF pulse, typi(cid:173)
`cally using a flip angle of 90°, followed by
`one or more refocusing RF pulses, typically
`using a flip angle of 180°. The important fea(cid:173)
`ture of the spatial -encoding scheme is that all
`echoes collected in any given repetition are
`encoded the same. Thus, for multiple refocus(cid:173)
`ing RF pulses, the echo signals recorded after
`each pulse are used to generate images with
`different T2 weightings. That is, each echo
`time (TE) corresponds
`to a separate image.
`Conventional SE imaging is a steady -state
`technique and is used almost exclusively to
`perform 20 multisection acquisitions. Its typi(cid:173)
`cal k-space trajectory is simp le, collecting data
`sequentially, line by line, moving from one
`extreme of k-space to the opposite extreme.
`Aside from the effects of motion, no k-space
`filtering is inherent to a conventional SE ac(cid:173)
`quisition.
`Although the most common implementa(cid:173)
`tions employ a flip angle of 90° for the excita(cid:173)
`tion RF pulse, flip angles other than 90° are
`sometimes used. 6, 21, 89 For example, in a sin(cid:173)
`gle-echo Tl-weighted pulse sequence the ex(cid:173)
`citation flip angle can be decreased to yield
`increased Tl weighting at the expense of de(cid:173)
`creased SNR. In a double-echo T2-weighted
`
`1so·
`go·
`
`1so·
`
`Figu re 6. Pulse-sequence timing for a double spin-echo technique. RF
`= radiofrequency pulses; Gss = section-select gradient ; GPe = phase(cid:173)
`encoding gradient; GRo = readout gradient; DAO = data acquisition; and
`N20 = number of in-plane phase-encoding lines.
`
`General Electric Co. 1005 - Page 11
`
`

`
`670
`
`MUGLER
`
`pulse sequence the excitation flip angle can
`be decreased to gener

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket