throbber
MAGNETIC RESONANCE IN MEDICINE 3, 823-833 (1986)
`
`RARE Imaging: A Fast Imaging Method for Clinical MR
`
`J. HENNIG,*A. NAUERTH,? AND H. FRIEDBURG*
`
`*Department of Diagnostic Radiology, University Hospital, University of Freiburg, Hugstetter Strasse 55,
`West Germany, and TBruker Medizintechnik GmbH, POB 67, 7512 Rheinstetten-4/West Germany
`
`Received May 6, 1985; revised April 14, 1986
`
`Based on the principles of echo imaging, we present a method to acquire sufficient data
`for a 256 X 256 image in from 2 to 40 s. The image contrast is dominated by the transverse
`relaxation time T2. Sampling all projections for 2D FT image reconstruction in one (or a
`few) echo trains leads to image artifacts due to the different T2 weighting ofthe echo. These
`artifacts cannot be described by a simple smearing out of the image in the phase direction.
`Proper distribution of the phase-encoding steps on the echoes can be used to minimize
`artifacts and even lead to resolution enhancement. In spite of the short data acquisition
`times, the signal amplitudes of structures with long T2 are nearly the same as those in a
`conventional 2D FT experiment. Our method, therefore, is an ideal screening technique
`for lesions with long T2. 0 1986 Academic h ~ , Inc.
`
`Conventional imaging techniques used in MRI take several minutes for a multislice
`and/or multiecho 256 X 256 image. The use of these time-consuming methods causes
`several problems in routine clinical work. These well known problems include patient
`discomfort and positioning-especially when one is looking for small lesions and
`image artifacts due to patient movement.
`It would therefore be of great importance to have an imaging method that allows
`a more rapid data acquisition. Although one is willing to pay some price in image
`quality for such a faster image, the image has to fulfill several conditions to be of any
`help in the problems mentioned above. The resolution should not be worse than that
`of a normal image, the image contrast should be variable and good enough to distinguish
`normal tissue from lesions, and of course the signal-to-noise ratio should be good
`enough to deliver an acceptable image quality. Above that, the imaging method should
`use only rffield strength, gradient field strength, and rising times which lay well within
`the limits given by the national health organizations ( I ) .
`There are several methods that lead to faster imaging. The standard menu of most
`of today's clinical MRI systems allows the fast acquisition of a low-resolution (128
`X 256) image by simply reducing the repetition time of a single-slice single-echo ex-
`periment. Since in such an experiment practically all the magnetization is very heavily
`saturated, the resulting image has very low S/N and contrast. Better results can be
`achieved by performing the experiment using excitation pulses with smaller flip angle
`(2-4). Although this leads to much better SIN and contrast, these methods still lack
`sensitivity for structures With long relaxation times, which are the ones most sought
`after in clinical MRI, especially in head investigations.
`The method we pursued is based on the principle of echo imaging, first introduced
`
`823
`
`0740-3194/86 $3.00
`Copyright 0 1986 by Academic Press, Inc.
`All rights of reproduction in any form resewed.
`
`General Electric Co. 1034 - Page 1
`
`

`
`824
`
`HENNIG, NAUERTH, AND FRIEDBURG
`
`to MRI by Mansfield (5-7). Long echo trains are generated whereby each echo is
`encoded differently for spatial coordinates (Fig. 1). Applied to 2D FT, where one axis
`of the imaged plane is represented by the phase information of the echoes, this means,
`that all other sources affecting the phase of the echoes have to be dealt with adequately.
`Disastrous phase errors can be introduced by H I , Ho, and switched gradient field
`inhomogeneities. Sampling data very fast minimizes magnetic field effects, as has been
`shown by Mansfield (5-7). Although a MR-real-time movie could be demonstrated
`with this method, its routine application is restricted since the gradient field strength
`necessary lies outside the possibilities of routine MRI systems.
`Several methods to overcome inhomogeneities by self-compensating spin-echo se-
`quences have been proposed (8-Z0), but, due to the fact that only some inhomogeneities
`were accounted for, only a modest approximation to a clinical fast imaging sequence
`could be achieved (10, ZZ). The problem can be solved by respecting the CPMG
`conditions regarding the phase of the magnetization not only in respect to the phase
`of the transmitter signals but to the switched gradient fields as well (12). This means,
`all gradient effects are compensated within one echo cycle, so that no accumulative
`effects can arise. There are different ways this principle can be incorporated into a
`working pulse program (13, Z8). This enables the data acquisition with only a few-
`in the limiting case one-excitations of the spin system.
`
`IMAGE QUALITY
`The data flow in the 2D FT image reconstruction algorithm can be characterized
`by
`
`F T
`FT
`E + F - E l - S.
`
`1
`
`FIG. 1. Principle of echoimaging. (a-d) show 4 echo trains of a conventional 2D FI experiment. Each
`echotrain carries different phase encoding cp. In the echo imaging experiment only one echo train is sampled,
`the phase encoding is different for each echo.
`
`General Electric Co. 1034 - Page 2
`
`

`
`RARE IMAGING
`
`825
`
`Echoes E are sampled in the time domain, their Fourier transforms Fare rearranged
`to give the “echoes” E’, which yield the shape function S after the second FT. In a
`conventional 2D FT experiment all projections necessary for image reconstruction
`are read out at the same time after excitation. Contrary to that, each projection in an
`echoimaging experiment is read out at a different time and therefore experiences a
`different attenuation due to T2 relaxation. To evaluate the quality of the image mea-
`sured by this method one has to know how this affects the image contrast.
`An easy approach to this problem is to consider first a conventional 2D FT exper-
`iment performed on a point-like object located at the center of the gradient system.
`Due to its location this object experiences no influence due to gradients whatsoever
`and all echoes E are characterized by frequency zero and an amplitude that is pro-
`portional to the product of the spin density with the value of the T2 relaxation function
`at the echo readout time. The Fourier transform yields identical delta functions at the
`origin. Rearrangement of the data matrix according to the 2D FT algorithm yields
`“echoes” E‘, all of which are zero except one, which again is characterized by frequency
`zero. The second Fourier transform yields the set of shape functions S, all of which
`are zero except the one at the center, which again is given by a delta function. The
`resulting image shows intensity only in the voxel at the center.
`If an echoimaging experiment is performed on this trivial case, all echoes are again
`characterized by frequency zero. The amplitude, however, is now different in all pro-
`jections due to T2 relaxation. This means, that after the first Fourier transform one
`gets a set of delta functions with different amplitudes. The result after rearrangement
`and second Fourier transform strongly depends on the pattern of the data sampling.
`If the dephasing steps are distributed stochastically on the echoes of the echotrain, the
`result is a noise-like pattern for the “echo” E‘, the corresponding image consists of a
`point in the center due to the fact, that all values of E‘ are greater zero, plus many
`artifacts randomly distributed along the central line of the image.
`A more rational approach would be a sequential distribution of the phase-encoding
`steps on the echo train. In this case the “echo” E’ looks like a discrete representation
`of the T2 relaxation function of the object. The time interval between two sampling
`points is given by the echo spacing 2*7,. The resulting shape function consists of a
`Lorentzian line, whose width is determined by the relaxation time T2. The image
`appears to be smeared out in the phase direction, the smearing being more severe for
`short T2 and/or long echo spacing 2*7,.
`If an echo imaging experiment is performed that delivers coherent echoes in the
`manner described in the previous chapter, then the outcome of the experiment is not
`altered by the addition of gradients. This means, that independent of its location, a
`point-like object always yields an image, which is smeared out somewhat in the phase
`direction (14).
`The complementary case is that of an image with infinite extension. Here E’ is
`given by a delta function whose amplitude is solely defined by the amplitude of the
`projection with zero dephasing, since this is the only one with nonzero intensity. The
`resulting image is a grey plane with homogeneous intensity, the intensity value is given
`by the product of the spin density with the value of the relaxation function at the
`readout time of that particular echo. In general, objects with short relaxation times
`will appear darker than those with long relaxation times at equal spin density.
`
`General Electric Co. 1034 - Page 3
`
`

`
`826
`
`HENNIG, NAUERTH, AND FRIEDBURG
`
`0
`
`90'
`
`-90'
`
`0
`
`c 'f
`
`t
`-180'
`
`-90'
`
`180'
`=-180°
`FIG. 2. The "echo" E' as yielded after the first half of the 2D FT algorithm. The horizontal scale gives
`the relative phaseshift of neighboring voxels due to phase encoding. For clarity two cycles of E' are shown.
`In an echo imaging experiment E' is multiplied by the relaxation function. Three possibilities differing in
`the sequence of phase-encoding steps are represented by (a-c).
`
`A somewhat more realistic situation arises if one looks at objects with finite extension
`inside the image frame and characterized by one relaxation function. The result of an
`echo imaging experiment for such a case is shown qualitatively in Figs. 2 and 3. Figure
`2 shows the "echo" E' as a result of the first Fourier transform and rearrangement of
`the data matrix in a conventional 2D FT experiment. E' is cyclic in the sense that a
`phase difference of - 180" between neighboring points yields the same result as a phase
`difference of 180" according to the aliasing principle in discreet Fourier transformation.
`For the sake of the following argument, two cycles of E' are shown. As described
`above, this "echo" has to be multiplied by the discreet representation of the relaxation
`
`FIG. 3. The result of experiments 2a-c. The resolution of the final shape functions S (right) strongly
`depend on the way of phase encoding. S can appear even to be resolution enhanced as compared to the
`result of a conventional 2D FT (d). The effects are exaggerated in these purely qualitative drawings.
`
`General Electric Co. 1034 - Page 4
`
`

`
`RARE IMAGING
`
`827
`
`function, if an echo imaging experiment is performed, where the phase-encoding steps
`are distributed sequentially. The outcome of this experiment depends not only on the
`fact that the phase-encoding steps are distributed sequentially on the echoes, but also
`on which phase is attributed to the first echo, that is, where the experiment starts in
`the phase cycle. Three cases are shown in Figs. 2a-c. Figure 3 shows the final result
`after the second Fourier-transform. Figure 3d shows the shape function as yielded in
`a conventional 2D FT experiment. Figures 3a-c show the strong dependence of the
`result on the exact manner of distributing the phase-encoding steps. It is particularly
`interesting to note, that the shape can appear to be sharpened (Fig. 3a), contrary to
`what one would expect by a simplistic generalization of the argument for point-like
`objects.
`This sharpening can be rationalized if E' is separated into an ascending part and a
`descending part. Multiplying the ascending part with the monotonous descending
`relaxation function leads to a flatter slope and therefore to a sharpening after Fourier
`transformation. Multiplication of the descending part (shown in dotted line in Figs.
`3a-c) yields an even steeper slope and broadening after Fourier transformation. The
`shape function S is the sum of both parts.
`According to the basic properties of Fourier transformation the center (= zero fre-
`quency) amplitude of the shape function is given by the integral of E' after multipli-
`cation with the discreet representation of the relaxation function. It is therefore de-
`pendent on T2 and 2*7, as well as on the exact shape of the object and the manner
`of phase encoding.
`The slope of the discreet representation of the relaxation function depends not only
`on T2, but also on the echo spacing 2*7,. Shorter echo spacing leads to a more gradual
`slope and therefore to a less pronounced effect on the amplitude of the shape function
`as well as its distortion.
`Due to technical restrictions, 2*7, cannot be made arbitrarily small in a multiecho
`experiment. An artificial shortening of the sampling steps for the relaxation function
`can be achieved if the experiment is performed with n excitations instead of one and
`the phase-encoding steps are interleaved (15). Discontinuities arising from the fact
`that by simple repetition of the multiecho experiment under variation of the phase-
`encoding groups of n projections are sampled at identical readout times, can be removed
`by one of the following methods:
`(1) Prolongation of the times between 90" and the first 180" pulse by 7Jn in con-
`secutive echo trains.
`(2) Substitution of the 90" pulse by a 90"-~,/n-180" pulse sequence.
`(3) Substitution of the 90" pulse by a 90" pulse followed by an orthogonal spin-
`locking pulse of duration 7Jn.
`For an explicit calculation for a realistic case, the image representing spin density has
`first to be separated into regions with identical T2. E' has to be calculated for all those
`shape functions by inverse Fourier transformation and the multiplication according
`to the phase-encoding scheme carried out using the appropriate discreet representation
`of the respective relaxation functions. Fourier transformation and addition of the
`results for all different T2's leads to the echo image.
`Of course such a calculation is impracticable for a clinical case. Some insight can
`
`General Electric Co. 1034 - Page 5
`
`

`
`828
`
`HENNIG, NAUERTH, AND FREDBURG
`
`however be gained by looking at the result of the explicit calculation for objects with
`Lorentzian spindensity distribution along the phase direction, which is given in the
`Appendix. Figures 4a-d show the influence of T2 and the width of a structure on the
`signal intensity for a singleexcitation echo image and one that used 16 excitations in
`the manner described above. It is clear, that the relaxation constant is the main de-
`tennining parameter for the intensity attenuation (Figs. 4a, b). We therefore call this
`method RARE imaging for Rapid Acquisition with Relaxation Enhancement.
`Figure 4b shows that the T2 contrast is much less pronounced if one samples the
`data in 16 echotrains with 16 echoes each, leading to a good presentation of even
`short T2 structures. The influence of the width of the structure is shown in Figs.
`4c-d. Sharp structures are more attenuated than structures with medium width. Not
`shown in the graphs is the sharp decline of the signal intensity for very broad structures
`especially when the dephasing zero point is put into late echoes.
`A very interesting point to note is given by the second vertical scale in Figs. 3 and
`4. It gives the signal sampled per unit of time as compared to a conventional 2D
`experiment and therefore can be used as a measure of the efficiency of the imaging
`method. For structures with long T2, this figure of merit can be as high as 100, clearly
`contradicting the common misconception, that faster imaging for a given resolution
`is only attainable at the cost of low signal to noise.
`
`Fu
`50
`
`25
`
`c
`
`LE1
`
`a
`32
`
`d
`
`1000
`
`2000 als-’)
`
`FIG. 4. Calculated relative intensities for Lorentzian shapes in a 256 X 256 RARE image. The intensity
`units Id are scaled to I equal to the spin density. Fwre of merit F,,, is given as the quotient of S i
`sampled per unit of time for RARE image divided by the same number for a conventional 2D FT image
`with same resolution. (a) I vs T2 for a singleshot imase; (b) I vs T2 for a 16cxcitation image; (c) I vs u for
`a singleshot image; (d) I vs u for 16-excitation image where u is the linewidth of the Lorentzian shape in
`Hz for a total image size of 25 kHz. y ~ o gives the acho number with dephasing gradient 0.
`
`
`
`General Electric Co. 1034 - Page 6
`
`

`
`RARE IMAGING
`
`829
`
`RESULTS
`The validity of the arguments presented in the previous chapter is demonstrated
`by the graphs shown in Fig. 5, where the relative signal intensity is plotted vs the
`relaxation time for a human patient head slice. Although an attempt to attribute the
`size of structures has not been made quantitatively, it is obvious, that the sharper
`structures of the skin are more attenuated than more extended structures inside the
`head for equal T2. The images, from which the figures are taken, are shown in Figs.
`6a-c. T2 values were calculated from 12 echoes in a conventional multiecho 2D fl
`sequence.
`A comparison of the single acquisition image (Fig. 5b) with a RARE image taken
`with 16 acquisitions (Fig. 5c) shows, that the T2 contrast is much stronger in the single
`acquisition image. It is obvious as well, that the single acquisition image has sufficient
`resolution to be a valuable tool for patient positioning. More than that, due to its T2
`character it is especially suited to detect lesions with long T2.
`From a clinical point of view it is most important to note that the RARE sequence
`delivers images which are practically free from artifacts if an appropriate phase-encoding
`scheme is chosen. This is markedly different from the situation one encounters when
`echo imaging is applied to a back-projection imaging scheme (16).
`
`DISCUSSION
`Since its implementation in October 1984, we have examined about 1 100 patients
`with RARE imaging in our department. We have found, that it is an indispensable
`tool for exact patient positioning. Above that, more than 90% of lesions in the head,
`which were lesions with long T2, were clearly shown in the RARE imaging alone.
`Multiple sclerosis plaques are shown clearly in the RARE image. Combined with
`multislice techniques, a head scan with 8 slices can be performed in less than 2 min,
`leading to very short overall investigation times per patient. Due to the fact that the
`RARE sequence, which has proven to be most useful in head investigations and which
`is represented by Fig. 6c, gives only low T2 contrast between medium T2 values (200
`ms and more), fat around the spinal chord is not always clearly distinguishable from
`tumors and therefore different sequences giving sharper contrast have to be tailored
`
`I ':"
`
`5
`5
`1 0
`FIG. 5. Rel. intensities I=, and figures of merit FM vs T2 for RARE images shown in FIG. 6. (a) Single-
`shot experiment; (b) 16-excitation experiment. Open circles correspond to sharp structures.
`
`10
`
`General Electric Co. 1034 - Page 7
`
`

`
`8 30
`
`HENNIG, NAUERTH, AND FRIEDBURG
`
`FIG. 6. Conventional 2D FT image (a) compared to a single-shot RARE image (b) and a 16-excitation
`RARE image (c) of a patient with recurrent pituitary adenoma with infiltration of the brain tissue. Imaging
`parameters for the conventional image were 1800 ms for the repetition time, echo readout time 33 ms.
`
`for these applications. The single acquisition experiment (Fig. 6b) promises to be most
`useful in these cases (18).
`In the abdomen, where the prediction of relaxation times is most difficult and short
`T2 are not uncommon, the choice of the appropriate RARE sequence is often difficult-
`as is the choice of the appropriate conventional 2D FT experiment. RARE can be
`used here as a fast means to check for the T2 sensitivity of a suspected lesion. According
`to the contrast of the RARE image one can choose the 2D FT sequence most likely
`to yield a good presentation of the lesion with more confidence, so that the overall
`efficiency and speed of an examination is increased even if the RARE image gives
`only poor lesion contrast.
`The high signal sampling efficiency of RARE promises to make it a good method
`for 3D FT. Apart from that, it can be used to acquire images with very high SIN in
`reasonable time by averaging. Both possibilities are studied at the moment in our
`department.
`
`EXPERIMENTAL
`
`All our experiments were performed on a 0.23-T system Bruker BNT 1 100. Patients
`were investigated using a 60-cm body coil, a 50-cm body coil, 28 and 22 cm head
`coils, and various surface coils for special cases. No special setup procedures were
`
`General Electric Co. 1034 - Page 8
`
`

`
`RARE IMAGING
`
`83 1
`
`FIG. 6-Continued.
`
`General Electric Co. 1034 - Page 9
`
`

`
`832
`
`HENNIG, NAUERTH, AND FRIEDBURG
`
`necessary for imaging with our method, allowing us to apply it as part of our routine
`measurement menu.
`Slice width was 8 mm standard some investigations were performed with down to
`2 mm width. 2D FT and RARE imaging were taken without patient dislocation to
`allow the direct comparison of images.
`
`APPENDIX
`
`Explicit Evaluation of the Shape Function of a Lorentzian
`Image with RARE Imaging
`
`A structure with Lorentzian spin-density distribution oriented along the phase-
`encoding gradient and symmetrical around the gradient field zero point yields in a
`2D FT imaging sequence an echo for the second Fourier transformation which can
`be described as
`
`E = Moexp(a - cp)
`E = Moexp(-a - cp)
`
`for
`
`for
`
`cp < 0
`
`cp > 0
`
`where cp, the time coordinate in a normal 1D F'T experiment, is given by the dephasing
`gradient as described by Ernst (1 7). Fourier transformation yields the well-known
`function
`
`where w is the spatial coordinate in frequency units. To calculate the lineshape according
`to a RARE imaging experiment, one has to multiply the echo function E with the
`relaxation function. The zero point of both functions needs not to be identical.
`The time variable of the relaxation function has to be transformed into the time
`domain of the phase-encoding gradient. This can be done easily if one takes into
`account that the dwell time of the phase-encoding gradient steps has to be identified
`with the echo spacing 27,. Due to the symmetry of the 2D lT algorithm, the dwell
`time of the phase encoding can be set equal to the dwell time dw of the data sampling
`relaxation by a factor 27,/dW for a single shot experiment or 27,/(n - dw) if one samples
`for a geometrically square image. This leads to a compression of the timescale for the
`data in n acquisitions in the interleaving manner described above. This yields
`
`E' = Moexp - 27e
`
`)exp(a(cp -
`
`( n.dw.T2
`
`General Electric Co. 1034 - Page 10
`
`

`
`Fourier transformation gives
`
`RARE IMAGING
`
`833
`
`-
`a -
`
`exp(- a - p0).
`- iu
`
`MO
`27,
`n-dw+ T2
`(po represents the dephasing zero point, that is, the center of the “echo” E’.
`Neglecting the first-order phase distortion factor exp(-iw(po) this can be described
`as the sum of three functions. The real part of the first term gives a Lorentzian line
`sharpened by 27J(n-dw* T2) compared to (I) and the real part of the second term
`gives a Lorentzian line broadened by the same term. The amplitude of both functions
`is attenuated by exp((-2r,cpo/(n-dw* T2)). The third term gives the loss of the signal
`due to the cutoff of the echo at cp = 0. If (po = 0 the first and third terms cancel and
`one gets the familiar solution of the Fourier transform of an FID. For sharp structures,
`where a < 27&
`dw T2) the amplitude of the signal is directly proportional to MO T2,
`giving an image identical in appearance to the product of a spin-density image with
`a T2 image as calculated from a multiecho sequence.
`
`REFERENCES
`1. T. F. BUDmGER, IEEE Trans. Nwl. Ski. 26,2812 (1979).
`J. C. WATERTON, H. G. LOVE, X. P. ZHU, 1. ISHERWOOD, AND D. RoWUNDS, Br.
`2. J. P. R. JENKINS,
`Heart J. 53,91 (1985).
`3. P. VAN DUK, Roceedings Fourth Annual Meeting of the Society for Magnetic Resonance in Medicine,
`1985, p. 572.
`4. A. WE, J. FRAIM, D. MATTHAEI, W. HAEMCKE, AND K. D. MERDOLDT, Ro~xdqs, Fourth
`Annual Meeting of the Society for Magnetic Resonance in Medicine, 1985, p. 980.
`5. P. MANSFIELD AND A. A. MAuDSLEy, J. Magn. Reson. n, 101 (1977).
`6. R. J. OWRIDGE, P. MANSFIELD, M. DOYLE, AND R. E. COPELAND. Br. J. Radiol. 55,729 (1982).
`7. P. MANSFIELD, Br. Med. Bull. 40, 187 (1984).
`8. W. A. EDELSTrm, U.S. Patents Nos. 4 451788,4 431 %8,4 471 306.
`9. Z.H.CHO,H.S.KIM,H.B.SONG,ANDJ.CUMMINGS,PTOCIEEE~O,~~~~(~~~~).
`10. E. M. HAACKE, F. H. BEARDEN, AND J. R. CLAYTON, Proceedings Fourth Annual Meeting of the
`Society for Magnetic Resonance in Medicine, 1985, p. 978.
`11. C. M. J. VAN UUEN, Magn. Reson. Med. 18,268 ( 1984).
`12. S. M E ~ M
`AND D. GILL, Rev. Ski. Instrum. 29,688 (1958).
`13. B. STROEBEL AND D. RATZEL, U.S. Patent pending.
`14. A. MEHLKOPF, P. VAN DER MEULEN, AND J. SMIDT, Mugn. Reson. Med. 1,295 (1984).
`15. 1. R. YOUNG AND M. BURL, UK Patent Awl. GB 2056078 A.
`16. L. D. HALL AND S. SUKUMAR, Magn. Reson. Med. 18, 179 (1984).
`17. A. KUMAR, D. WELTI, AND R. ERNST, J. Magn. Reson. 1669 (1975).
`18. J. HENNIG, H. FRIEDBURG, AND B. STROEBEL, J. Compul. Assist. Tomw. 10,375 (1986).
`
`General Electric Co. 1034 - Page 11

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket