throbber
Dual-Echo Interleaved Echo-Planar Imaging of the Brain
`
`Glenn S. Slavin, Kim Butts, John N. Rydberg, Clifford R. Jack, Stephen J. Riederer
`
`An interleaved echo-planar imaging (EPI) technique is de-
`scribed that provides images from 20 sections of the brain at
`two echo times (27 and 84 ms) in 1:M. Six echoes per image
`per repetition are collected in 24 repetitions of the pulse
`sequence. MR images of the brain obtained from five volun-
`teers using the dual-echo EPI sequence, fast spin-echo (FSE),
`and conventional dual-echo spin-echo were evaluated quali-
`tatively for diagnostic use and quantitatively for relative sig-
`nal-to-noise ratio (SNR), contrast, and contrast-to-noise
`ratios (CNR).
`Key words: echo-planar imaging; brain; MR; fast MR imaging.
`
`INTRODUCTION
`As introduced over 10 years ago, the two-dimensional
`Fourier-transform multi-slice spin-echo
`technique
`continues to be the standard procedure for clinical
`magnetic resonance imaging (MRI) of the brain (1).
`This is due to the high quality of the images in terms of
`signal-to-noise ratio and tissue contrast. Furthermore,
`conventional spin-echo (SE) is capable of providing
`both early and late echo images from one acquisition.
`The importance of acquiring long TR short TE images
`lies in the ability to provide pathology/cerebrospinal
`fluid (CSF) contrast as a complement to the pathology/
`gray matterlwhite matter contrast inherent in long TR
`long TE (T,-weighted) images. Nevertheless, it is well
`recognized that these high quality images are obtained
`at the cost of a relatively long scan time, typically on
`the order of 8 min for 256 X 192 resolution images
`using a 2500-ms TR. The purpose of this work was to
`develop a fast scan technique for providing long TR
`multiple echo images from a large volume such as the
`brain within an acquisition time arbitrarily selected to
`be 1 min long. The expectation is that such a sequence
`could be useful in providing MR images of patients for
`whom long acquisition times are problematic.
`A number of fast imaging techniques have been devel-
`oped in an attempt to reduce acquisition time while
`maintaining the overall quality of conventional T,-
`weighted spin-echo (T,SE) imaging. These include vari-
`ants of rapid acquisition with relaxation enhancement
`(RARE) imaging (2,3), combined gradient- and spin-echo
`(GRASE) imaging (4), and echo-planar imaging (EPI) (5).
`Hybrid RARE, or “fast” or “turbo” spin-echo (FSE),
`uses multiple repetitions of trains of individually phase-
`
`MRM 33264-270 (1995)
`From the Magnetic Resonance Laboratory, Department of Diagnostic Ra-
`diology, Mayo Clinic, Rochester, Minnesota
`Address correspondence to: S. J. Riederer, Ph.D., MR Laboratory, Mayo
`Clinic, 200 1st Street SW, Rochester, MN 55905.
`Received August 5, 1994; revised October 26. 1994; accepted October 26,
`1994.
`This work was supported in part by NIH Grant R01 CA37993 and by General
`Electric Medical Systems.
`0740-3194/95 $3.00
`Copyright 0 1995 by Williams & Wilkins
`All rights of reproduction in any form reserved.
`
`encoded RF-refocused spin-echoes. Although it is theo-
`retically possible to reduce the T,SE acquisition time by
`a factor equal to the number of echoes acquired per
`repetition, increasing the length of the echo trains can
`decrease the number of sections imaged per TR. Dual-
`echo contrast can be achieved in several ways. First, with
`appropriate phase-encoding assignments, the early and
`late echoes can be acquired from two separate scans.
`This, however, defeats some of the advantage of time
`reduction provided by the RARE technique. Alterna-
`tively, a single scan can be devised in which the first half
`of the collected data is incorporated into the early-echo
`image and the second half of the data is incorporated into
`the late-echo image (6). However, this method also limits
`the number of sections available for imaging relative to
`conventional SE, thereby compromising reduced scan
`time to achieve coverage of the targeted volume. A mod-
`ified version of FSE has been developed to generate dual-
`echo images by view sharing (7). In clinical studies, this
`method has been used to produce dual-echo images at 40
`and 80 ms of 18 sections in 2 min. Because some of the
`measurements are used in more than one image, the
`shared-echo approach in general has better efficiency for
`image formation at multiple echo times than a standard
`FSE approach.
`Although FSE imaging of the brain has in general dem-
`onstrated contrast and overall image quality similar to
`conventional SE imaging (8, 9), it has been shown to be
`subject to several complications. These include blurring
`of early echo or spin density-weighted images (lo), edge
`enhancement and ringing in T,-weighted images, and
`perhaps most significantly, decreased sensitivity to mag-
`netic susceptibility effects (9). Furthermore, FSE appears
`more prone to magnetization transfer effects than does
`conventional SE (11).
`GRASE imaging, which utilizes short gradient-recalled
`echo trains centered about each of a number of RF-refo-
`cused spin-echoes, has also been proposed as a means to
`reduce imaging time while maintaining SE contrast.
`While GRASE appears to produce contrast comparable
`with conventional SE, because it is closely related to
`FSE, it too can be penalized in terms of fewer imaged
`sections compared with conventional SE. Recently, a
`variant of GRASE has been presented (12) in which im-
`ages corresponding to two echo times are generated for
`each section. With this technique as described, only the
`late echo is generated in the true GRASE fashion; with
`only one refocusing pulse, the early echo is acquired in a
`manner identical to that of interleaved EPI and similar to
`that presented in this work.
`As originally described, EPI collects the entire raw
`image data set by sampling the free induction decay after
`a single excitation, although variants can be used (13)
`with a refocusing pulse to provide later echo times as
`desired for T,-weighting. Recently, dual-echo single-shot
`EPI, using three 90° pulses to generate a spin-echo
`
`264
`
`General Electric Co. 1035 - Page 1
`
`

`
`Dual-Echo Interleaved EPI of the Brain
`
`265
`
`(TI31 = 50-80 ms) and a stimulated echo (TE2 = 200-
`500 ms) has been demonstrated (14). Such single-shot
`methods, which can generate images in as little as tens of
`milliseconds, necessitate sampling bandwidths of sev-
`eral hundred kHz. In addition to the reduced signal-to-
`noise ratio (SNR) resulting from the high bandwidth,
`single-shot EPI is also susceptible to resolution limita-
`tions and off-resonance artifacts as a consequence of the
`extended readout duration (15). These resolution and
`off-resonance limitations can potentially be addressed
`with specialized gradient systems.
`In an attempt to exploit the EPI methodology while
`simultaneously addressing some of its intrinsic limita-
`tions, several methods of “interleaved” or “multi-shot”
`EPI acquisition have been explored (16-21). These meth-
`ods acquire multiple gradient-refocused echoes in each
`of several repetitions. In fact, one such variant (19, 21)
`has recently been shown to provide diagnostic quality
`images of the liver with acquisition times of less than 20
`s. Given these results and the considerations described
`above, the specific purpose of this work was to design a
`multi-slice dual-echo interleaved EPI technique for rapid
`MR imaging of the brain.
`
`METHODS
`Motivation
`To facilitate a comparison between the fast-scan se-
`quences discussed in the Introduction and the dual-echo
`EPI technique presented here, a summary of pulse se-
`quence characteristics is presented in Table 1. It should
`be emphasized that the values presented in this table are
`simply representative of those used in research or clini-
`
`Table 1
`Summary of Fast-Scan Pulse Sequence Characteristics
`
`cal settings. A certain degree of freedom is available to
`the user in choosing values for most of the listed param-
`eters. These choices will inevitably affect the number of
`imaged sections as well as the total acquisition time.
`The first two columns in Table 1 list the pulse se-
`quences and their corresponding references. The first
`entry, conventional dual-echo SE, is considered the stan-
`dard technique against which the others are ultimately
`compared. The third column lists the duration of data
`acquisition within one repetition. For techniques that
`involve repeated 180’ pulses (e.g., FSE and GRASE), this
`duration is considered to extend from the start of the first
`echo to the end of the last echo. For techniques that
`utilize gradient-recalled echo trains with only one 180’
`pulse per TE (e.g., EPI, SE), the duration of data acquisi-
`tion is the duration of the echo train. Column 5 shows a
`second basis for comparison. The inter-echo time is ef-
`fectively the duration of data acquisition (column 3) di-
`vided by the number of echoes collected (column 4)
`during that time. It can be interpreted as one measure of
`a technique’s efficiency of data collection. The values
`listed for number of echoes per repetition per image
`(column 4) sometimes called the echo train length, or
`ETL, as well as the TR (column 7) are taken either from
`the cited reference or from clinical protocols. The acqui-
`sition times in column 8 are those resulting from the
`aforementioned parameters. Two values are shown for
`each technique: those not in parentheses for equal fields-
`of-view (FOV) in the readout and phase-encoding direc-
`tions and those in parentheses if a three-fourths FOV is
`used in the phase-encoding direction (usually the pa-
`tient’s right/left).
`
`1
`
`2
`
`Technique
`
`Reference
`
`Standard
`Dual-echo SE
`
`One image per section
`Single-shot EPI
`Multi-shot EPI
`Single-echo GFIASE
`Single-echo FSE
`
`(1 1
`
`(5)
`(19, 21)
`(4)
`(2)
`
`3
`
`4
`
`z2::: Echoes per
`
`repetition
`acquisition per image
`(msec)
`8 + 16b
`
`1
`
`5
`
`6
`
`7
`
`Inter-echo
`time
`(msec)
`
`Bandwidth Sections imaged
`per scan
`(256 X 192)
`(kH.4
`
`8
`
`~
`
`9
`
`Acquisition ST lr
`
`time per
`scana
`
`dual-echo
`sectionsa
`
`516, ?ab
`
`21 (TR = 2300)
`
`7:49 (552)
`
`7:49 (5x52)
`
`100
`80
`188
`127
`
`128
`16
`24
`8
`
`0.25-1 .o
`5
`7.8
`17
`
`250-500
`232
`232
`216
`
`arbitraw
`12 (TR = 2000)
`13 (TR = 2500)
`15 (m = 2600)
`
`100mdimage
`(0:18)‘
`0:23 (0:18)
`1 :08 (052)
`
`(1 :03)‘
`1:12 (056)”
`3:21 (2:34)‘
`
`(1 4)
`
`73
`
`4
`4
`7,9b
`6
`
`1.5
`
`17
`20
`2.6, 9.ab
`5
`
`Two images per section
`110 + 110
`Dual-echo single-shot
`EPI
`59 + 59
`(6)
`Dual-echo FSE
`128
`(7)
`Shared-echo FSE
`26 + 88
`(1 2)
`Dual-echo GFIASE
`30 + 30
`Dual-echo interleaved This work
`EPI
`See text for details. The technique proposed here, dual-echo interleaved EPI, is the last entry in the table.
`a Acquisition times for the 3/4 FOV case are shown in parentheses.
`Values for early and late echoes, respectively.
`‘ One acquisition for the early echo plus one acquisition for the late echo (7R = 3400 ms). Also, employs 3/4 k-space sampling.
`128 x 128 resolution.
`‘ One acquisition ( W T R = 4/2200) for the early echo plus one acquisition (ETUTR = 8/3500) for the late echo.
`One acquisition for the early echo plus one acquisition for the late echo (7R = 4000 ms).
`g Employs 1/2 k-space sampling.
`rR = 3000 ms.
`
`255
`
`arbitravg
`
`300md2 img
`
`216
`216
`50, 216.7b
`? 32
`
`20 (TR = 3500)
`18 (TR = 2600)
`20 (TR = 3500)
`20 (TR = 2600)
`
`2 5 5 (2:13)
`2:55 (2:13)
`2:30 (1 :54)h
`2:lO (1 :38)
`1:41 (1:17)
`1:41 (1:17)
`1 :26 (1 :05) 1 :26 (1 :05)
`
`General Electric Co. 1035 - Page 2
`
`

`
`266
`
`As stated previously, the purpose of this work was to
`develop an RF-refocused interleaved EPI pulse sequence
`that could provide both early and late echo images, e.g.,
`at 30 and 80 ms echo times, for multiple sections from a
`single 1-min imaging acquisition. Specific design consid-
`erations included the ability to provide high resolution,
`heavily T,-weighted, multiple-section images of the
`brain. It was felt that these requirements could be met by
`using 256 X 192 resolution and a repetition time greater
`than 2000 ms. The first echo time (TE1) was targeted to
`be short enough that CSF would be no brighter than the
`surrounding brain parenchyma. The second echo time
`(TE2) was targeted to be sufficiently long that both CSF
`and typical pathology would be bright compared with
`brain parenchyma but not so long that SNR would be
`severely compromised. To provide adequate image qual-
`ity, additional RF and gradient waveforms were incorpo-
`rated for both fat suppression as well as spatial presatu-
`ration of flowing spins. The ultimate goal was to use the
`above specifications to image the entire brain in a single
`acquisition. From our experience, adequate coverage can
`be achieved by imaging 20 sections with a 5-mm thick-
`ness and 2.5-mm gap. The sequence was implemented on
`a standard 1.5 T GE Signa system with 1 G/cm maximum
`gradient amplitudes and 600 ps rise time from zero to full
`scale.
`The last column in Table 1 gives approximations for
`the total scan time necessary to generate dual-echo im-
`ages from at least 20 sections. These scan times are a
`consequence of selecting parameters such as ETL and TR
`which allow appropriate weighting, adequate coverage,
`and acceptable quality of the early and late echo images
`while attempting to minimize the acquisition period.
`These times are illustrative and may be slightly longer or
`shorter depending on the actual choice of parameters.
`Given the specific design considerations, it was con-
`cluded that the dual-echo interleaved EPI sequence, the
`last entry in Table 1, was a plausible candidate.
`
`Pulse Sequence
`The dual-echo EPI pulse sequence is shown in Fig. 1.
`Data acquisition periods in the form of gradient-recalled
`echo trains are placed around each of the two RF-refo-
`
`-
`
`G,
`
`RF
`
`I
`80
`
`I
`100
`
`I
`0
`
`I
`2 0
`
`I
`40
`
`I
`6 0
`Time (rnsec)
`FIG. 1. Dual-echo interleaved EPI pulse sequence diagram. Posi-
`tions of the sliding echo trains for the 1st and 12th of the 24
`waveform are depicted with solid and
`repetitions of the G,,
`dotted lines, respectively. Although not shown, the phase-encod-
`ing blips of the Gphase waveform slide in synchrony with their
`respective echo trains.
`
`Slavin et al.
`
`cused echo times. By targeting echo times near 30 and 80
`ms, the maximum duration available for each gradient
`echo train is approximately 40 ms. This estimate is used
`in determining the maximum possible number of echoes
`per echo train, i.e., the echo train length (ETL). With a
`receiver bandwidth of -+32 kHz, the read window dura-
`tion for an individual 256-point echo is about 4 ms.
`Including ramp times, the inter-echo time is 5 ms. Addi-
`tionally, the technique of “sliding echo trains” (17, 21-
`24) is incorporated into both acquisition periods to
`smooth out magnitude and phase transitions in the raw
`data. Because the total sliding period is approximately
`equal to the duration of a single read window, each of the
`two data acquisition periods must allow room for ETL+ 1
`echoes. As a consequence of these constraints, an ETL of
`six, which provides two 30-ms data acquisition periods,
`was selected. Therefore, to achieve 256 X 192 resolution,
`32 repetitions of the sequence are required, with six
`echoes acquired per image per repetition. For this work,
`a three-quarters FOV in the phase-encoding direction
`was used, permitting the number of repetitions to be
`reduced to 24.
`The phase-encoding procedure is illustrated in Fig. 2,
`again using the following parameters: y-resolution of 192,
`ETL of 6, and 24 repetitions (144 total views). After the
`90’ excitation and first 180’ refocusing pulses, six lines
`of data (corresponding to views 1,25,49,73,97, and 121)
`are acquired as in conventional interleaved EPI for the
`early echo image (Fig. 2a). The second 180’ pulse inverts
`the position along both k, and k,, and six lines of data
`(corresponding to views 24, 48, 72, 96,120, and 144) are
`acquired for the late echo image (Fig. 2b). Thus, k, is
`traversed in the same direction, top to bottom, for both
`echo trains. During subsequent repetitions, however, the
`echo train for the first data set must step in the -k,
`direction (downward in Fig. 2a), while the echo train for
`the second data set must step in the +k, direction (up-
`ward in Fig. 2b). For example, on the second repetition,
`views 2,26,50, 74,98, and 122 are collected for the early
`echo image while views 23,47, 71, 95, 119, and 143 are
`collected for the late echo image.
`The echo time of the first image (TE1) is defined as the
`time after the 90’ excitation at which the k, = 0 line is
`measured. For RF-refocusing, this typically occurs at ex-
`actly twice the delay time between the goo and 180’
`pulses. In Fig. 2, the six groups of k, lines (24 lines per
`group) sampled by the six echoes of each echo train are
`shown to exactly straddle ky = 0. A problem with this for
`the bandwidth and resolution used in this work is that
`measuring three groups of ky lines prior to refocusing
`forced TE1 to be no smaller than 37 ms. This echo time
`resulted in images with unacceptably high CSF signal
`relative to surrounding brain parenchyma. Accordingly,
`to reduce TE1, fewer ky lines were measured prior to the
`refocusing time, resulting in asymmetric filling of
`k-space along the k, direction. For this work, k, = 0 was
`straddled by the second group of k, lines. Specifically,
`phase-encodings +AkJ2 and -AkJ2
`are acquired on
`views 36 and 37 for the early echo and views 108 and 109
`for the late echo, where Ak,, is the sampling interval used
`along k,.
`
`General Electric Co. 1035 - Page 3
`
`

`
`Dual-Echo Interleaved EPI of the Brain
`
`267
`
`semndmu
`
`Rep 1
`Rep 2
`Rep 3
`
`I
`4 ......................................................................
`
`----------__ - - - - - - _ _ _ _ _ _ _ kx 73+72 4 i
`- - - _ _ _ _ _ _ _ _ _ _
`----_-__---_
`I
`,
`
`- ....................................................................
`I
`b k x
`
`t
`
`a.
`
`t
`
`b.
`
`FIG. 2. Schematic of k-space tra-
`jectories for dual-echo interleaved
`EPI. Three of the 24 repetitions are
`shown. Each repetition after data
`collection ends for the first echo (0
`in part (a)), the readout position is
`inverted along both k, and k,, by the
`second 180” pulse. This new posi-
`tion determines where data collec-
`tion begins for the second echo (0
`in part (b)).
`
`To implement the sliding echo trains with this phase-
`encoding scheme, on successive repetitions the first echo
`train is preceded by a small cumulative time delay while
`the second echo train is provided an equally cumulative
`time advance. In Fig. 1, this corresponds to the simulta-
`neous sliding of the first and second echo trains to the
`right and left, respectively. As mentioned previously, the
`purpose of the sliding echo trains is to allow continuous
`sampling of the T,* decay curves, thereby providing
`smooth magnitude and phase modulations over k-space.
`The difference in the sliding directions compensates for
`filling k-space “backwards” in the second image.
`
`In Vivo Imaging
`Multi-section MR images of the brain of five healthy
`volunteers were obtained using the dual-echo EPI se-
`quence and compared with FSE and SE images of the
`same sections. The dual-echo EPI sequence was per-
`formed with a TElITE2ITR of 27/84/2500 and 256 X 192
`resolution. Axial sections, 5 mm thick with 2.5 mm sep-
`aration, were acquired using a 22-cm FOV and a 314 FOV
`in the phase-encoding direction. Two dummy repeti-
`tions, played out prior to the 24 repetitions of data col-
`lection, were included to set the spins close to a steady
`state. Fat saturation was used to reduce the intensity of
`chemical shift misregistration (15), while inferior spatial
`presaturation (25) and first-order gradient moment null-
`
`ing (26) on the late echo in the slice-select direction were
`used to minimize artifacts due to flowing blood and CSF.
`For comparison, FSE and conventional dual-echo SE
`were performed with parameters substantially matching
`those of the EPI (Table 2). The FSE used two acquisitions
`with different echo train lengths at a 516 lcHz band-
`width: TE/TRIETL of 17/2500/4 for the early echo and
`85/2500/8 for the late echo. The SE sequence used a
`TElITE2ITR of 27/84/2500 with variable bandwidth:
`216 lcHz for the first echo and +8 lcHz for the second.
`Both FSE and SE utilized inferior spatial presaturation,
`5-mm sections with 2.5-mm separation, and 22-cm FOV
`with 314 FOV in the phase-encoding direction.
`Dual-echo images from 20 sections were obtained us-
`ing EPI, FSE, and SE in 1:05,3:15, and 6:30, respectively.
`Although the FSE technique with a 17-ms TE is typi-
`cally used at our institution for the early echo, the FSE
`was also performed using a 34-ms TE, as this echo time
`more closely matched the TE1 used for SE and EPI.
`
`Evaluation
`Images were evaluated based on signal-to-noise (SNR],
`contrast and contrast-to-noise ratio (CNR) calculations
`and overall image quality as determined using radiolog-
`ical criteria. Mean signal measurements, from three sec-
`tions per technique, were made in uniform regions of
`both gray and white matter. The regions of interest were
`
`Table 2
`Scan Parameters
`
`Sequence
`
`SE
`
`FSE
`
`TEITR
`2712500
`8412500
`
`1712500
`8512500
`3412500
`
`ETL
`
`1
`1
`
`4
`8
`4
`
`BW
`(W
`216
`-C8
`
`216
`216
`216
`
`Sections
`
`Imaging options
`
`Sat-I
`Sat-I
`
`Sat-I
`Sat-I
`Sat-I
`
`20
`2oa
`20
`
`EPI
`
`232
`6
`2712500
`Sat-I, Fat Sat
`232
`8412500
`6
`Sat-I, Fat Sat, GMN
`NOTE: Three-fourths FOV in the phase-encoding direction and 5 mm section thickness are used for all techniques. Sat-I = inferior spatial presaturation; Fat
`Sat = spectral saturation of fat; GMN = first order gradient moment nulling in slice-select direction.
`a Two acquisitions were necessary to achieve 20 sections at this TR.
`This Scan was performed to facilitate comparison with the first echo of the SE and EPI scans.
`
`Acquisition time
`
`} 6:30
`] 3:15
`1 :35
`1 :40b
`} 1:05
`
`General Electric Co. 1035 - Page 4
`
`

`
`268
`
`chosen to be in identical locations for both early and late
`echo images for each of the three techniques. Standard
`deviations of the noise were measured in regions of the
`image outside the head along the phase-encoding direc-
`tion, excluding areas corrupted by pulsatile flow artifact.
`Gray matter SNR was defined as gray matter signal di-
`vided by the standard deviation of the noise, and white
`matter SNR was defined similarly using the white matter
`signal. Gray mattedwhite matter contrast was defined as
`the gray matter signal minus the white matter signal
`divided by the white matter signal. Contrast-to-noise ra-
`tios were calculated as the signal difference between gray
`and white matter divided by the standard deviation of
`the noise. For each technique, mean contrast and CNR
`with associated standard deviations were determined by
`averaging the contrast and CNR values from the five
`volunteers. The results were then normalized to those
`from the SE images. Image quality was assessed radiolog-
`ically on the basis of anatomic resolution and contrast as
`well as prominence of any artifact.
`
`RESULTS
`Early and late echo images of the SE (a and b), FSE (c and
`d), and EPI (e and fl sequences hom one section of one
`volunteer are shown in Fig. 3.
`Table 3 shows the results of SNR, contrast, and CNR
`measurements for the three techniques. The EPI and
`FSE SNR measurements for gray and white matter ap-
`proximately scale with bandwidth relative to SE. Both
`EPI and FSE images exhibit significantly higher aver-
`age gray matterlwhite matter (GM/WM) contrast on the
`early echo, by a factor exceeding 1.50, relative to con-
`ventional SE. EPI, FSE, and SE all show similar average
`
`Slavin et al.
`
`contrast on the second echo. Average CNR from the
`early EPI echo is similar to the SE values due primarily
`to the increased contrast. On the other hand, average
`CNR from the late EPI echo is lower, by a factor of
`0.52 2 0.10, relative to SE.
`Radiologically, the images in all studies were felt to
`have adequate contrast characteristics in the early and
`late echoes, as targeted. Several EPI images showed arti-
`fact due to magnetic susceptibility effects near regions of
`air-tissue interfaces. These artifacts occurred predomi-
`nantly at sinuses and at the skull base but also appeared
`in isolated locations near the surface of the brain. In
`addition, ghosting artifacts due to pulsatile flow of both
`blood and CSF seemed to be more apparent with EPI than
`with FSE or SE. Once again, these artifacts were most
`prominent at the more inferior locations in the head.
`Clearly, many of these radiological criteria can only be
`better assessed in prospective patient studies.
`
`DISCUSSION
`This work was undertaken to develop an interleaved EPI
`pulse sequence aimed at providing clinically useful dual-
`echo images of the whole brain within a 1-min scan time.
`We feel that the results of this study demonstrate that
`early and late echo images obtained with the new dual-
`echo EPI sequence are of promising quality and, due to
`the short acquisition time, are especially suited for pa-
`tients uncooperative for standard imaging procedures.
`Results from a preliminary clinical study performed with
`the dual-echo EPI sequence are shown in Fig. 4.
`The EPI results generated in this work show G M I W
`contrast similar to FSE in both echoes and significantly
`higher than SE in the early echo. The cause of this in-
`
`FIG. 3. Early and late echo images
`from one section of the head of a
`healthy volunteer. (a, b) Conven-
`tional spin-echo. (c, d) Fast spin-
`echo. (e, 9 EPI.
`
`General Electric Co. 1035 - Page 5
`
`

`
`Dual-Echo Interleaved EPI of the Brain
`
`269
`
`Table 3
`Relative SNR and Gray MatterMlhite Matter (GMMIM) Contrast and CNR for EPI and FSE versus SE
`Relative
`white matter
`SNR
`1 .o
`1 .o
`
`Sequence
`
`TEITR
`
`SE
`
`27/2500
`84/2500
`
`Bandwidth
`(kH4
`
`216
`28
`
`Relative gray
`matter S N R
`1 .o
`1 .o
`
`Relative GM/
`WM contrast
`1 .o
`1 .o
`
`FSE
`
`17/2500
`34/2500
`85/2500
`
`216
`216
`216
`
`EPI
`
`232
`27/2500
`2 32
`84/2500
`Measurements are averages over five volunteers.
`
`1.02 2 0.03
`0.93 2 0.02
`0.79 2 0.04
`
`0.65 2 0.02
`0.46 2 0.04
`
`0.97 2 0.03
`0.88 2 0.02
`0.80 2 0.04
`
`0.62 2 0.02
`0.45 2 0.03
`
`1.53 2 0.16
`1.64 2 0.02
`0.96 2 0.16
`
`1.53 2 0.20
`1.15? 0.17
`
`Relative GM/
`WM CNR
`1 .o
`1 .o
`1.44 * 0.04
`
`1.48 2 0.11
`
`0.76 2 0.13
`
`0.95 2 0.12
`0.52 2 0.10
`
`severe SNR penalty is on the second echo, where EPI
`uses a ?32 kHz bandwidth compared with 2 8 kHz for
`conventional SE. This fourfold increase in bandwidth
`results in a twofold reduction in SNR. It is conceivable
`that this problem of SNR loss with the EPI sequence can
`be addressed with improved receiver coils.
`Magnetic susceptibility effects can manifest them-
`selves as several types of artifacts. Because data is col-
`lected along a train of gradient-recalled echoes, signal
`loss due to spin dephasing can be expected in regions of
`susceptibility differences. Furthermore, due to the rela-
`tively long (30 ms) data acquisition periods, large phase
`differences can accumulate between tissues with differ-
`ent magnetic susceptibilities. These phase differences
`give rise to spatial shifts in the phase-encoding direction
`which can cause overlapping of structures in the image.
`The resulting artifacts appear as streaks of very high
`signal intensity. Nevertheless, the increased sensitivity
`to susceptibility effects that cause artifacts in some EPI
`sections may, in other sections, prove advantageous with
`respect to FSE. The short inter-echo spacing in FSE has
`been implicated in the reduced sensitivity of FSE to
`susceptibility variations (3, 27). Because EPI would be
`expected to have a sensitivity to magnetic susceptibility
`somewhere between that of SE and gradient-recalled
`echo imaging, iron-containing structures and hemor-
`rhagic lesions may be better depicted with EPI than with
`FSE. However, this has yet to be shown in a patient
`series.
`As suggested by column 9 of Table 1, the interleaved
`EPI sequence can be considered one sample along a
`continuum of pulse sequences which can be ordered
`according to the acquisition time required to perform
`whole-brain T,-weighted imaging. For the shortest
`times-several hundred milliseconds per section (or
`images can be
`up to 10 s for the entire brain)-the
`prone to off-resonance effects and resolution limits as
`well as reduced SNR because of the required high
`bandwidth. These effects become less severe as times
`are increased to the 40-s to 1 min range, where either
`averaging of single-shot or implementation of multi-
`shot echo-planar methods can be employed. Incorpo-
`ration of RF-refocusing for some or all echoes reduces
`the sensitivity to off-resonance but typically causes an
`increased inter-echo time and increased scan time.
`Finally, conventional spin-echo requires the longest
`
`d
`C
`FIG. 4. Dual-echo EPI (a, b) and conventional spin-echo (c, d)
`images of the same section of the head of a 38-year-old male
`diagnosed with multiple sclerosis. The patient was scanned with
`the conventional SE sequence before the EPI. The EPI used the
`parameters listed in Table 2, while the SE had the following minor
`differences: TEl/TE2/TR = 30/80/2200, 20 x 20 cm FOV, and
`first-order gradient moment nulling in the section-select direction.
`FSE was not used. Scan times for the SE and EPI were 7:29 and
`1 :05, respectively. Qualitatively, the pathology is visible (arrows) in
`the EPI images (a, b) to a degree similar to that in the SE images
`(c, d). Slight section misalignment resulted from interscan patient
`movement.
`
`crease in contrast requires further study. Despite the
`improved contrast, the EPI images demonstrate a reduced
`CNR relative to SE and FSE. Decreased SNR, which is
`reflected in CNR, is expected with the EPI sequence due
`primarily to the higher sampling bandwidth. The most
`
`General Electric Co. 1035 - Page 6
`
`

`
`270
`
`Slavin et al.
`
`acquisition time, 5 min or more, but has few of these
`limitations. The operating point along this continuum
`will likely depend on user preference.
`In summary, we have demonstrated the ability of an
`interleaved RF-refocused echo-planar pulse sequence to
`produce multiple-section dual-echo images of the whole
`brain in a 1-min long acquisition period, approximately
`six to seven times shorter than that for conventional
`spin-echo.
`
`REFERENCES
`1. M. Brant-Zawadski, D. Norman, T.H. Newton, Magnetic
`resonance imaging of the brain: the optimal screening tech-
`nique. Radiology 152, 71-77 (1984).
`2. J. Hennig, A. Naureth, H. Friedburg, RARE imaging: a fast
`imaging method for clinical MR. Magn. Reson. Med. 3,823-
`833 (1986).
`3. P. S. Melki, R. V. Mukern, L. P. Panych, F. A. Jolesz, Com-
`paring the FAISE method with conventional dual-echo se-
`quences. J. Magn. Reson. Imaging 1, 319-326 (1991).
`4. D. A. Feinberg, K. Oshio, GRASE (Gradient- and Spin-Echo)
`M R imaging: a new fast clinical imaging technique. Radiol-
`ogy 181, 597-602 (1991).
`5. P. Mansfield, Multi-planar image formation using NMR spin
`echoes. J. Phys. C10, L55-L58 (1977).
`6. P. S. Melki, F. A. Jolesz, R. V. Mulkern, Partial RF echo-
`planar imaging with the FAISE method. II. Contrast equiv-
`alence with spin-echo sequences. Magn. Reson. Med. 26,
`342-354 (1992).
`7. B. A. Johnson, E. K. Fram, B. P. Drayer, B. L. Dean, P. J.
`Keller, R. Jacobowitz, Evaluation of shared-view acquisition
`using repeated echoes (SHARE): a dual-echo fast spin-echo
`MR technique. AJNR 15,667-673 (1994).
`8. K. M. Jones, R. V. Mulkern, M. T. Mantello, P. S. Melki, S. S.
`Ahn, P. D. Barnes, F. A. Jolesz, Brain hemorrhage: evalua-
`tion with fast spin-echo and conventional spin-echo images.
`Radiology 182(1), 53-58 (1992).
`9. A. M. Norbash, G. H. Glover, D. R. Enzmann, Intracerebral
`lesion contrast with spin-echo and fast spin-echo pulse se-
`quences. Radiology 185, 661-665 (1992).
`10. R. V. Mulkern, S . T. S. Wong, C. Winalski, F. A. Jolesz, Con-
`trast manipulation and artifact assessment of 2D and 3D
`RARE sequences. Magn. Reson. Imaging 8, 557-566 (1990).
`11. R. T. Constable, A. W. Anderson, J. Zhong, J. C. Gore, Fac-
`tors influencing contrast in. fast spin echo M R imaging.
`Magn. Reson. Imaging 10,497-511 (1992).
`12. D. A. Feinberg, B. Kiefer, A. W. Litt, Dual contrast GRASE
`
`(gradient-spin echo) imaging using mixed bandwidths.
`Magn. Reson. Med. 31(4), 461-464 (1994).
`13. M. S. Cohen, R. M. Weisskoff, Ultra-fast imaging. Magn. Re-
`son. Imaging 9, 1-37 (1991).
`14. P. Bornert, D. Jense

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket