throbber
Portal Imaging Technology:
`Past, Present, and Future
`
`made, more development needs to be directed towards
`making portal imaging convenient and reliable. Image
`quality must be improved further, to improve the robust-
`ness of image registration techniques and more thought
`must be given to integrating and automating the various
`steps in the image registration process. Otherwise, too
`much time will have to be devoted to these tasks.
`Finally, and most importantly, users will have to decide
`what is the best way of using EPlDs clinically. Much
`development is required before the full potential of this
`exciting technology can be realized.
`Copyright © 1995 by W. B. Saunders Company
`
`Peter Munro
`
`Many different electronic portal imaging devices (EP|Ds)
`have been developed to improve geometric accuracy in
`radiation therapy. This article describes the two types of
`EPlDs that have become available commercia|ly—the
`television camera-based EPID and the matrix ion cham-
`ber EP|D—as well as describing the amorphous silicon
`array, a device that may become available in the future
`for portal imaging. in addition, the various image regis-
`tration techniques that identify geometric errors from
`the portal images are described. These include interac-
`tive techniques, landmark-based techniques, contrast-
`based techniques, and hybrid techniques. Although great
`improvements in portal imaging technology have been
`
`he goal of radiation therapy is to deliver a
`prescribed radiation dose to the target volume
`accurately while sparing the surrounding normal and
`critical tissues. Experimental and clinical evidence
`shows that small changes in dose of 7% to 1. % can
`reduce local tumor control signiIicanLly,"3 or increase
`the rate of normal tissue complications.4 As a result,
`recommendations by the International Commission
`on Radiation Units (ICRU) suggest that the accu-
`racy in dose delivery be i‘5‘.’/o.'Z'3 Such accuracy can be
`achieved only if field placement is precise during the
`entire course of radiation treatment, so that the
`treatment beams irradiate only the prescribed re-
`gions.
`Unfortunately, the geometric accuracy of radia-
`tion treatments can be compromised resulting in the
`irradiation of regions other than those prescribed. A
`number of studies“? have shown that discrepancies
`in field placement occur frequently, especially for
`complicated treatment setups. Furthermore, these
`geometric discrepancies can also influence the out-
`come of treatment.8"3‘ *5 Fortunately, frequent moni-
`toring of patient positioning can reduce the fre-
`quency of discrepancies in field placement.7‘9-“H9 As a
`result, several studies have suggested that patient
`positioning be checked daily.19’20
`
` (cid:14) (cid:14)
`
`Trends in radiation treatments are increasing the
`need for accurate patient positioning.
`\/Vith the
`integration of computed tomography (CT) and, on
`occasion, magnetic resonance imaging (MRI) data
`into the treatment planning process, with the devel-
`opment of affordable three-dimensional treatment
`planning work stations, and with the advent of
`programmable multi—leaf collimators for field shap-
`ing, radiation treatment portals have become more
`highly tailored with smaller margins around the
`target volumes. In addition, new treatment tech-
`niques such as dynamic beam modulation are increas-
`ing the need for routine monitoring of patient
`positioning even further. Because of these trends in
`radiation therapy, much effort has been devoted to
`developing more convenient methods to image the
`patient during radiation treatment,
`the process
`known as portal imaging. This article examines the
`histoiy of portal imaging developments, describes
`some of the devices (known as electronic portal
`imaging devices or EPIDSJ that are currently avail-
`able commercially, discusses one of the imaging
`devices that may become available in the future, and
`examines some of the image processing and image
`registration techniques that are essential for EPIDs
`to be useful in the clinic.
`
`From t/zzDe,tnzrtnzmt_r cf/l/[radical Bin};/yxirx and Oncology, I/nioerxity Ly"
`I/Vestern Ontario, and London Regional Cancer Centre, London, Ontario,
`Canada.
`Addrm reprint requert: to Peter Munro, PhD, London Regional Cancer
`Centre, 790 Cammissinrmry Ro1zd,EzLrt, Lnmiun, 0m.‘z1r1'oN6/l 4L6, Canada.
`Supported by the Ontario Cancer Treatment r1ndRe5mrch Foundotzon.
`Copyright © 1995 by WB. Saundm Company
`I053-4296/ 95/ 0502~0005$05.00/ 0
`
`History
`
`There is a long history of imaging with high—energy
`radiation beams. As early as 1904, radiographs of
`human hands, mice, purses, and other objects had
`been made using radium as the radiation source.“
`However, because of the high energy of the radium
`gamma rays, the images suffered from low contrast,
`
`Ser/Lilian in Rachofiorz Oncology, V015, Nu 2 (April), 1995' M) 115-133
`
`1 1 5
`
`Page 1 of 19
`
`Elekta Exhibit 1006
`
`

`
`Peter Kl/Izmra
`
`which made them unsuited for the diagnostic applica-
`tions that were of interest at the time. One of the
`
`l958, several artielcsm” described the use of a
`television-roentgen (TVR) system for monitoring
`the position of patients during pendulum therapy.
`The patients lay on the treatment couch while the
`gantry-mounted TVR system (consisting of a X—ray
`image intensifier and a TV camera) rotated around
`the patient during the treatment with a 200-kV(p)
`x—ray beam. The video signal was sent to a monitor
`located in the control room of the treatment Ina-
`
`this TVR
`chine. The major limitation was that
`system had a field of View of only 5.0 inches (2.5
`inches at the patient). Independently, a somewhat
`similar system was being developed by Andrews et
`al” using a TV camera—based device called a ‘johns
`Hopkins screen intensifier”3‘ as the image receptor.
`It consisted of a fluorescent screen viewed by an
`image orthicon TV camera using a complicated
`series of relay mirrors and lenses known as a Schmidt
`optical system. These efforts resulted in some of the
`earliest megavoltage images, using a 2-MeV Van de
`Graaff generator as the X-ray source. Not surpris-
`ingly, the authors found that normal tissue contrast
`was insullicient and they had to use air or mercury
`contrast materials to visual anatomic structures. In
`
`
`
`first references to imaging in radiation therapy was
`by Nielsen and Jensen” who, in l942, described a
`rotation therapy technique for treating cancer of the
`esophagus. In the treatment, the patient sat upright
`in a rotating chair while a stationary x—ray beam was
`directed horizontally towards the patient. The thera-
`peutic radiation (l80 kV(p) x-ray beam) exiting the
`patient hit a fluorescent screen that was viewed by an
`observer who was looking through a lead glass
`window. Not only did the observer view the treat-
`ment in real time, but corrections to the position of
`the beam were made remotely during the treatment.
`Thus,
`this treatment may have been one of the
`earliest examples of dynamic conformal radiation
`therapy! By l95l, a similar treatment for cancer of
`the esophagus had been developed by Hare et 21123
`using a Q—MeV Van de Graafl‘ generator, They
`described the use of film radiography to ensure
`accurate positioning of the patient before commence-
`ment of the rotation therapy. They showed a great
`deal of sophistication in their portal lilrn activities,
`investigating the utility of introducing air into the
`bladder and rectum as a contrast agent as well as
`using double-exposure radiographs to visualize
`anatomy outside of the treatment field. Surprisingly,
`during the 1950s most of the scientific interest in
`imaging with megavoltage beams dealt with diagnos-
`tic applications. Several articles by Tuddenham et
`al2”*v25 promoted the use of megavoltage x—ray beams
`for chest radiography to optimize exposure in both
`the mediastinum and the lung regions and to mini-
`mize the visibility of overlying “osseous” structures.
`In 1960, Perryman et al26 described cobalt 60 radiog-
`raphy. The technique used Kodak type AA industrial
`film (Eastman Kodak, Rochester, NY) placed in a
`cassette where the standard intensiiying screens had
`been replaced by two 0.01-inch lead sheets. The only
`drawback was that Iilrn development took approxi-
`mately 30 minutes to complete. Not only were the
`radiographs used for treatment localization but also
`to give diagnostic information “regarding the extent
`and location of soft tissue lesions.” Finally, in 1962,
`Springer et al” suggested an improvement to cobalt
`60 radiography where two fluorescent screens were
`placed between the lead sheets and the film. This
`modification reduced the exposure time and, it was
`claimed, increased the contrast of the final radio-
`graphs.
`About this time, non-film imaging methods started
`to be introduced into radiation therapy clinics. In
`
`1962, another article“ described an imaging system
`that used an imaging device similar to the johns
`Hopkins screen intensifier. The field of view at the
`detector was 25 cm in diameter and the system was
`attached to a 30—MV betatron. Again, only high-
`contrast objects such as tungsten, gold, and lead
`markers produced sufficient contrast to be visualized
`by the system. The investigators showed great inge-
`nuity in introducing high—contrast, high-density ob-
`jects into various body orifices such as the esophagus.
`However, the limited image quality and the need for
`contrast agents reduced the utility of all of these
`imaging systems and they never came into wide
`spread clinical use.
`important developments for
`One of the most
`portal imaging was First described by Swain and
`Steckcl in L96633 and was refined further by Marks
`and Haus and their colleagues?/"35 The method used
`a slow, wide-latitude film that was placed in card-
`board film holders and was exposed for the entire
`duration of the treatment. Unlike portal film tech-
`niques common at the time, the patient was not
`moved between the exposure of the film and the
`treatment and the cardboard film~ho1ders were much
`more comfortable to lie on than the film-screen
`
`cassettes that were typically modified for therapy
`verification, Not only was much less effort required to
`treat the patient but, in addition, a record of the
`
`Page 2 of 19
`
`

`
`Portal Imaging Technology
`
`117
`
`entirc treatment was available from the film. Be-
`cause of the efibrts of Marks and Haus, a film
`(eventually known as Kodak XV-2 film), which was
`compatible with the 90—second lilm processors37 that
`had been introduced into hospitals in 1965, became
`available for therapy verification. This was a major
`development for portal imaging. Not only did Marks
`and Haus and their colleagues develop a convenient
`method ofverifying patient positioning during radia-
`tion treatment but they also showed in a series of
`studies7’” the importance of routine verification in
`reducing errors in patient positioning and in improv-
`ing local control. As a result, use of portal films
`became routine and eventually generated the de-
`mand for even more convenient methods of therapy
`verification that has since lead to the development of
`EPIDS.
`
`Commercially Available Imaging
`Devices
`
`Many different devices have been examined since the
`early 1980s as alternatives to film. These devices can
`be divided into two categories: scanning systems,
`where the radiation detector subtends only a small
`
`fraction of the radiation beam and must be scanned
`
`underneath the patient to form the image, and area
`systems, where the detector subtends the entire
`radiation beam. Examples of these devices include
`scanning diode arrays,33‘”‘0 scanning scintillator ar-
`rays,“ storage pl1osphors,”‘2”“ coded aperture ar-
`rays,“ matrix ion chamber systems,‘*5”*7 and TV
`camera—based systerr1s.”‘E*54 lnsullicient space is avail-
`able to discuss all of these systems and readers are
`referred to two comprehensive reviews of portal
`imaging devices for further details.55=55 The following
`discussion concentrates on the matrix ion chamber
`
`and the TV camera-based EPIDS, which are both
`available commercially, as well as the amorphous
`silicon array,57‘72 which is an imaging device that may
`become available commercially in the future.
`
`Matrix Ion Chamber
`
`A schematic of the matrix ion chamber device, which
`was originally developed by Meertens and van Ilerk
`and their colleagues‘*59”‘7 is shown in Fig l. The device
`consists of two sets of electrodes that are oriented
`
`perpendicularly to each other separated by a 0.8—mm
`gap, which is filled with a fluid (2,2,4-trimethylpen-
`tane) that is ionized when the device is irradiated.
`
`
`
`Processor
`2)] +256 KMemory
`
`Accelerator
`Sync Pulse
`
`Analog
`Inputs.
`Mulllplexar
`Control
`4.:
`
`Dedicated
`Control
`Proi:e5sor
`
`125 lol
`Multiplexer
`
`Control
`Electronics
`
`User
`
`Contro Unit
`
`Megavoltage”Camera" Cassette
`
`the matrix ion chamber. The device consists ofa set of 256 signal electrodes and a set of
`Figure 1. Schematic diagram 0
`256 high-voltage electrodes oriented perpendicularly to each other and separated by a 0.8—mm gap filled with a fluid called
`2,2,4 trimethylpentane. VVhen irradiated, the fluid is ionized and generates signals in the signal electrodes. The active area
`of the device is 32.5 X 32.5 cm and its overall dimensions are approximately 60 X 60 X 5 cm including the readout
`electronics, which are immediately adjacent to the ion chambers. The device is controlled by a computer located in the
`control area of the linear accelerator. (Reprinted with permission.46)
`
`Page 3 of 19
`
`

`
`Peter Fl/Iunro
`
`The electrode spacing is 1.27 mm and, since each set
`of electrodes consists of 256 electrodes, the active
`area of the matrix ion chamber array is 32.5 cm on a
`side. One set of electrodes is connected to 256
`electrometers and the other set of electrodes is
`
`
`
`connected to a high-voltage supply that can apply a
`300-V potential to each electrode individually. The
`matrix ion chamber array is read out by applying a
`high voltage to each of the high—voltage electrodes in
`succession (for approximately 20 milliseconds) and
`measuring the signal generated in each of the 256
`signal electrodes. This procedure takes 5.5 seconds to
`read out an image. In addition, a fast (lower resolu-
`tion) scanning mode is available that scans the array
`in 1.5 seconds by applying the high voltage for a
`l0—millisecond period to two high—voltage electrodes
`at a time. The fast acquisition mode is useful for
`acquiring double—exposure images.
`The most obvious advantage of the matrix ion
`chamber is its compact size, which makes the device
`a convenient replacement for film cassettes. Another
`advantage is geometric reliability—images acquired
`with this EPID have no geometric distortions. Fur-
`thermore, unlike other scanning EPIDS, the matrix
`ion chamber has no moving parts, reducing the
`likelihood ofmeehanical problems.
`The major limitation of most EPIDS that use a
`scanning radiation detector, such as the matrix ion
`chamber, is quantum utilization. Ideally, an image
`receptor should use all of the available radiation
`efficiently (even for megavoltage imaging) because
`this will improve image quality. Clearly, this is not
`the case for the matrix ion chamber, where only one
`high-voltage electrode (out of 256) is active at any
`one time. However, the physics of signal generation
`in the 2,2,4 trimethylpentane improves the quantum
`utilization of the matrix ion chamber considerably.
`The signal measured by the matrix ion chamber
`depends on the rate of formation and the rate of
`recombination of the ion pairs that are generated in
`the ionizing fluid. Even when no high voltage is
`applied to the electrodes, the rate of recombination
`of the ion pairs generated in the 2,2,4 trimethylpen-
`tane is relatively slow. Therefore, the concentration
`of ion pairs can increase over a period of time until an
`equilibrium is reached between ion-pair formation,
`which depends on the dose rate at the matrix ion
`chamber, and ion-pair recombination, which de-
`pends on the probability of ions encountering each
`other and is proportional to the square of the ion-pair
`concentration. The signal measured by any electrode
`of the matrix ion chamber does not depend greatly
`
`on the dose rate during the 20-millisecond period
`when the high voltage is applied but on the previous
`irradiation history of the electrode. Calculations
`have shown that after 0.5 second, a latent image has
`been form ed over the entire irradiated region of the
`matrix ion chamber and that irradiating for a longer
`time will not increase the size of the signal, ie, will not
`improve image quality. These observations have both
`positive and negative implications. The measured
`signal is six to seven times greater than would be
`expected if no charge integration occurred in the
`2,2,4 trimethylpentane. However, the effective pe-
`riod of the charge integration (~05 second) is still
`short compared with the total image acquisition time
`of 5.5 seconds. Therefore, a large fraction of the
`radiation that interacts with the matrix ion chamber
`
`does not generate any measurable signal. For this
`reason,
`the matrix ion chamber requires higher
`doses to generate images than other portal imaging
`devices.
`
`An example of an image acquired with the matrix
`ion chamber EPID is shown in Fig 2. One ofthe most
`noticeable characteristics of this image is how exten-
`sively the raw signals have to be processed before
`yielding a usable image.
`(The artifacts in Fig 2
`represent an extreme case because these images
`were acquired using a prototype device.) Because
`spurious signals are generated in the electrometers
`and ion chambers, and because the sensitivities of
`each ion chamber can vary,
`the device must be
`calibrated routinely. In addition, because the matrix
`ion chamber is a scanning EPID, it is more suscep-
`tible to artifacts if the dose rate of the accelerator
`
`changes during image acquisition. Thus, the radia-
`tion beam has to stabilize for some period (typically
`1.0 second) after startup before image acquisition
`can begin.
`
`TV Camera—Based EPIDs
`
`The design of TV camera—based EPIDS, which is
`shown in Fig 3, uses technologies that have long been
`used for other applications. The x—ray detector con—
`sists of a metal plate to which is attached a gadolin-
`ium oxysulphide (Gd2O2S) screen similar to the
`screens used in diagnostic radiology. VVhen irradi~
`ated, high—energy electrons generated in the metal
`plate and the Gd2OgS screen are converted into light.
`The light that diffuses through the screen and exits
`the rear surface of the x-ray detector is viewed by a
`TV camera using a 45° mirror. The video signal from
`the camera is digitized and the digitized image can
`
`Page 4 of 19
`
`

`
`Portal Imaging Technology
`
`1 19
`
`
`
`Figure 2. Image of an Alderson Rando phantom acquired using the matrix ion chamber EPID and an 8-MV x-ray beam:
`(A) before and (B) after corrections for variations in electrometer offset and sensitivity as well as variations in electrode
`shape and fluid thidmess. (Reprinted with permission."°)
`
`be viewed on a monitor located in the control area of
`the accelerator.
`
`This design has one major advantage. The x-ray
`detector subtends the entire area of the radiation
`beam so all of the radiation that exits from the
`
`patient has the potential to generate a signal in the
`EPID. However, the design suffers from one major
`limitation and this is the light collection efliciency of
`the optical chain. Figure 4 shows the problem sche-
`matically. Because the light is highly scattered within
`. the phosphor screen, the light is emitted from the
`rear of the screen in all directions with equal probabil-
`
`ity. Only those light photons that are emitted within
`a small cone subtended by the lens of the camera can
`generate a signal in the TV camera; typically only
`0.1% to 0.01% of the light emitted by the phosphor
`screen reaches the TV camera. This poor light
`collection efficiency can reduce image quality in two
`ways. Firstly, if an x-ray photon interacts in the x-ray
`detector and none of the light generated by this
`interaction reaches the TV camera, then no measur-
`able signal is produced. Seoondly,
`if only a small
`signal is produced in the TV camera then noise
`generated by the preamplifier and other electronics
`
`imaging Imlon
`. m.
`
`
`
`Figure 3. Schematic dia-
`gram of the TVcamera—baxed
`EP1Ds. These devices use a
`metal plate/phosphor screen
`as the x-ray detector. light
`emitted by the phosphor
`screen is viewed by a TV cam-
`era using a 45° mirror.
`
`Page 5 of 19
`
`

`
`I 20
`
`Peter ;’l'Iunra
`
`
`
`acce lance cone
`0 tholens
`
`Figure 4. Schematic diagram drawn approximately to
`scale showing the light collection efliciency of TV camera-
`based EPIDS. The narrow cone shows how only a small
`frauion of the light emitted from the phosphor screen
`reaches the TV camera.
`
`may be quite large compared with the small signal.
`As a result, much ofthc effort in the development of
`TV camcra—based EPIDS has been to improve the
`light collection of the optical chain, by increasing the
`light output of the phosphor scrccn,5'-73 using a large
`aperture lens that collects more of the light,“-55~7‘ or
`using TV cameras with high light-quantum efficien-
`cics.5'
`
`Care has to be taken when using these ap-
`proaches. \Vhile Munro et 2115‘ have shovm experimen-
`tally that using thick (-l00 mg/cm“) phosphor screens
`improves overall image quality, the current trend in
`the manufacture of TV camera—based EPTDs is
`
`towards the use of thinner (I50 mg/cm2) screens.
`This is because the thicker scrccns suffer from loss of
`
`spatial resolution and blemishes that generate visu-
`allydistracting flaws that have proven to be unaccept-
`able to users. Large aperture lenses suffer from
`decreased spatial resolution because of spherical
`aberrations (light rays reaching the edges of the lens
`do not focus to the same point as those reaching the
`center) and the spatial resolution of these lenses
`decreases from the center to the edge of the lens.
`Large aperture lenses also suffer from vignetting,
`which results in images that are brighter at the
`center of the lens than the edge. This change in
`image brightness complicates the display of portal
`
`images because level and windowing to display one
`part of the image will obscure other parts of the
`image that have different brightness levels. Thus,
`manufacturers have had to devise software or hard-
`
`ware correction schemes to compensate for lcns
`vignctting. Finally, large aperture lenses can gener-
`ate distortions, such as pin cushion or barrel distor-
`tion, which cause straight lines to appear curved in
`the image, especially at the edges of the Field of view.
`Care must be taken to ensure that these geometric
`distortions are negligible, otherwise the portal image
`will give a geometrically inaccurate representation of
`the patient's position.
`An example of an image acquired by the Infimed
`T.V. camera-based EPID (Infimed, Liverpool, NY)
`while using a special video digitizer (not available on
`the commercial system) is shown in Fig 5. The image
`was acquired using a 2-monitor unit irradiation and a
`6 MV x-ray beam. The TV camera. of the lnfimed
`EPID has a shading correction circuit that corrects
`for gradual changes in brightness caused by lens
`Vignetting and so the effect of lens vignetting is not
`visible in Fig 5A. No corrections have been applied to
`the image apart from changing the display contrast.
`Note that, although this EPID generates images that
`are free of artifacts such as lines or point blemishes,
`Fig 5 is not free from confounding structures. The
`tennis racket of the treatment couch is visible and
`can obscure some anatomic structures.
`
`Practical Considerations
`
`The above discussions have focussed on the factors
`
`that influence image quality. However, there are a
`number of other considerations that can also influ-
`
`ence the clinical utility of EPIDS. These include (1)
`the size of the mechanical assembly; (2) the field of
`view of the device; and, (3) the speed of the image
`acquisition. Table l summarizes some of the charac-
`teristics of the commercially available EPIDS.
`The aim of EPIDS is to improve the geometric
`accuracy of radiation treatments. Therefore, it is
`essential that the EPII)s do not make the setup ofthe
`patient more awkward, which could lead to an
`increase rather than decrease in field-placement
`errors. Photographs of EPIDS from different manu-
`facturers, whilc retracted and while deployed, are
`shown in Fig 6. Clearly, widely different designs have
`been used for the EPID assemblies. The mechanical
`
`assemblies are either retractable (Fig 6A,B,G,H),
`dcmountablc (Fig 6C,D), or partly dcmountable and
`partly retractable (Fig 6E,F). These differences
`
`Page 6 of 19
`
`

`
`Portal Imaging Technology
`
`121
`
`
`
`Figure 5. Image of the pelvic region of a patient acquired using the Infimed EPID, a 6-MV x-ray beam, and a 2-monitor
`unit irradiation: (A) before and (B) after window and leveling to optimize the display contrast. The TV camera of the
`Infimed EPID uses a shading correction circuit that corrects for lens vignctting during image acquisition.
`
`change how the devices are used. For instance: the
`gantry has to be set to 0° to attach and deploy the
`demountable EPID (Fig 6C,D); two of the EPID
`designs permit the detector-to-isocenter distance to
`be adjusted to acoomodate thicker patients or in-
`crease the efiective field ofvicw (Fig 6E,F,G,H); and,
`irradiation of the external electronics is a concern for
`
`one of the EPID designs (Fig 6G,H).
`One limitation of all currently available EPIDs is
`their small effective field of view. Whereas the sizes
`
`‘of the x—ray detectors are similar to that of standard
`(35 X 43 cm) film cassettes, the EPIDs are deployed
`at much larger isocenter-to-detector distances than
`film cassettes, thereby reducing their effective field
`of view. Having a large field of view is obviously
`important if one is imaging a large radiation field
`such as the mantle fields. However, Table I shows
`that one does not have to be treating with such large
`fields to reach the limits of the current EPIDs. Most
`
`EPIDs have fields ofview that cover an area approxi-
`mately 1/§ that ofa film cassette (which typically could
`be placed almost in Contact with the patient). It is not
`clear whether EPIDs with larger fields of View will
`become available soon because this would require
`that the current EPIDs be redesigned completely. It
`remains to be seen whether the advantages ofEPlDs,
`such as quantitative image analysis, can make up for
`their limited field ofview.
`
`Surprisingly, the characteristic that identifies each
`of the diflerent EPIDs most uniquely is not image
`quality, field of view, or physical size of the radiation
`detector, but the speed of image acquisition. More-
`over,
`it is this characteristic that has the largest
`influence on how these devices are used in the clinic.
`
`For instance, the Varian EPID (Varian Oncology
`Systems, Palo Alto, CA), which requires a long time
`to acquire an image,
`is unlikely to be used to
`intervene in patient positioning or to generate a
`movie of patient motion during treatment. In con-
`trast, the Siemens EPID (Siemens Medical Systems,
`Oncology Care Systems, Concord, CA) has been
`used extensively to generate movies of patient mo-
`tion during treatment,”-75 but requires that the user
`manually turn the radiation beam off‘when acquiring
`double-exposure images. Somewhere in between in
`acquisition rate is the Infimed EPID, which accumu-
`lates signals generated by a short (0.5 to l.0 second)
`irradiation and displays the image immediately after
`the irradiation has been completed, or generates a
`sequence of images every 0.5 to 1.0 second during a
`treatment. Thus,
`the Infimed EPID is very well
`suited to acquiring double-exposure images while
`minimizing the radiation exposure the patient. Al-
`though the products from the different manufactur-
`ers are becoming similar, distinctions remain be-
`cause of the different philosophies that initiated their
`
`Page 7 of 19
`
`

`
`122
`
`Peter 11/Iunm
`
`Table 1. The Characteristics of the Four Commercially Available EPIDS
`
`Supplier
`
`Iryimea.’/ GE
`
`Name
`
`Type
`Detector
`
`TV Camera
`
`Mechanical assembly
`
`Mounting
`
`Collision interlock
`
`Field ofview (cm)
`
`Detector to isocentre
`distance
`
`lsocentre accuracy
`Prototype descriptions
`
`Theraview/Target-
`View
`TV camera—based
`1.5 mm brass
`plate + 4-00
`mg/cm? Gd2O2S
`screen
`Plurnhicon or Chal-
`nicon
`Partly retractable and
`partly demountable
`Any accelerator (GE,
`Varian, Scandi-
`tronix)
`yes (customer must
`connect to accel-
`erator motion
`interlocks)
`Adjustable
`30.8 dia. Varian (iso-
`center)
`31.5 dia. Scanditronix
`31.7 dia. GE.
`40 X 40 (detector)
`30-60 cm (Varian)
`27-78 cm (Scanditro—
`nix)
`26-67 cm
`:5 m m
`Ref:')l
`
`Philips
`SR1 100
`
`Sicmem
`
`Varizzn
`
`Be amyiewph”
`
`PortalVision
`
`TV camera—based
`1.5 mm steel
`plate + 411
`mg/cm? GdgO2S
`screen
`CCD
`
`Demountable
`
`TV camera—based
`l—2mm plate + 160
`mg/cm?’ GdQOgS
`scrccn
`
`Newyicon
`
`Matrix ion chamber
`1.0 mm stainless steel
`plate + 0.8 mm
`2,2,4-trimethyl-
`pentane
`n/a
`
`Fully retractable
`
`Fully retractable
`
`Philips only
`
`Siemens only
`
`Any accelerator (cus-
`tomer must attach)
`
`Yes
`
`No (interlock prevents
`collision during
`deployment only)
`
`yes
`
`Fixed
`19 X 24- (isocenter)
`
`Fixed
`24 X 30 (isocenter)
`
`Adjustable
`25 X 25 (isocenter)
`
`30 X 38 (detector)
`60 cm
`
`35 X 4-4 (detector)
`39 cm
`
`325 X 32.5 (detector)
`5-80 cm
`
`ilmm
`Ref 19
`
`:2 mm
`Ref 53
`
`:5 mm
`Ref 46, 47
`
`
`
`NOTE. Some important considerations are how well the collision detection rnecliaiiisms ofthe (le\/ices wor

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket