throbber
Int J Radiation Oncology Biol Phys , Vol 45, No 3, pp 773–789, 1999
`Copyright © 1999 Elsevier Science Inc
`Printed in the USA All rights reserved
`0360-3016/99/$–see front matter
`
`PII S0360-3016(99)00118-2
`
`PHYSICS CONTRIBUTION
`
`A RADIOGRAPHIC AND TOMOGRAPHIC IMAGING SYSTEM INTEGRATED
`INTO A MEDICAL LINEAR ACCELERATOR FOR LOCALIZATION OF BONE
`AND SOFT-TISSUE TARGETS
`
`DAVID A. JAFFRAY, PH.D.,*† DOUGLAS G. DRAKE, B.S.,*† MICHEL MOREAU, PH.D.,*
`ALVARO A. MARTINEZ, M.D.,* AND JOHN W. WONG, PH.D.*†
`
`*Department of Radiation Oncology, William Beaumont Hospital, Royal Oak, MI; and †Department of Physics,
`Oakland University, Rochester, MI
`
`Purpose: Dose escalation in conformal radiation therapy requires accurate field placement. Electronic portal
`imaging devices are used to verify field placement but are limited by the low subject contrast of bony anatomy
`at megavoltage (MV) energies, the large imaging dose, and the small size of the radiation fields. In this article,
`we describe the in-house modification of a medical linear accelerator to provide radiographic and tomographic
`localization of bone and soft-tissue targets in the reference frame of the accelerator. This system separates the
`verification of beam delivery (machine settings, field shaping) from patient and target localization.
`Materials and Methods: A kilovoltage (kV) x-ray source is mounted on the drum assembly of an Elekta SL-20
`medical linear accelerator, maintaining the same isocenter as the treatment beam with the central axis at 90° to
`the treatment beam axis. The x-ray tube is powered by a high-frequency generator and can be retracted to the
`drum-face. Two CCD-based fluoroscopic imaging systems are mounted on the accelerator to collect MV and kV
`radiographic images. The system is also capable of cone-beam tomographic imaging at both MV and kV energies.
`The gain stages of the two imaging systems have been modeled to assess imaging performance. The contrast-
`resolution of the kV and MV systems was measured using a contrast-detail (C-D) phantom. The dosimetric
`advantage of using the kV imaging system over the MV system for the detection of bone-like objects is quantified
`for a specific imaging geometry using a C-D phantom. Accurate guidance of the treatment beam requires
`registration of the imaging and treatment coordinate systems. The mechanical characteristics of the treatment
`and imaging gantries are examined to determine a localizing precision assuming an unambiguous object. MV and
`kV radiographs of patients receiving radiation therapy are acquired to demonstrate the radiographic perfor-
`mance of the system. The tomographic performance is demonstrated on phantoms using both the MV and the
`kV imaging system, and the visibility of soft-tissue targets is assessed.
`Results and Discussion: Characterization of the gains in the two systems demonstrates that the MV system is
`x-ray quantum noise-limited at very low spatial frequencies; this is not the case for the kV system. The estimates
`of gain used in the model are validated by measurements of the total gain in each system. Contrast-detail
`measurements demonstrate that the MV system is capable of detecting subject contrasts of less than 0.1% (at 6
`and 18 MV). A comparison of the kV and MV contrast-detail performance indicates that equivalent bony object
`detection can be achieved with the kV system at significantly lower doses (factors of 40 and 90 lower than for 6
`and 18 MV, respectively). The tomographic performance of the system is promising; soft-tissue visibility is
`demonstrated at relatively low imaging doses (3 cGy) using four laboratory rats.
`Conclusions: We have integrated a kV radiographic and tomographic imaging system with a medical linear
`accelerator to allow localization of bone and soft-tissue structures in the reference frame of the accelerator.
`Modeling and experiments have demonstrated the feasibility of acquiring high-quality radiographic and tomo-
`graphic images at acceptable imaging doses. Full integration of the kV and MV imaging systems with the
`treatment machine will allow on-line radiographic and tomographic guidance of field placement. © 1999
`Elsevier Science Inc.
`
`Portal imaging, Conebeam computed tomography, Kilovoltage, Megavoltage.
`
`Oral Presentation at the 1997 ASTRO Meeting, Orlando, FL—
`October 1997
`Reprint requests to: D. Jaffray, Ph.D., Dept. of Radiation Oncol-
`ogy, William Beaumont Hospital, Royal Oak, MI 48073. Tel: (248)
`551-7024; Fax: (248) 551-0089; E-mail: djaffray@beaumont.edu.
`Acknowledgements: We would like to acknowledge Mr. Rob
`Cooke, Mr. Kevin Brown, and Dr. Di Yan for valuable discus-
`sions; Mr. John Musselwhite for his engineering support; and, Mr.
`Daniel Bilsky, Mr. Walter Jendhoff, Mr. Bill Porteners, Mr. Gabe
`
`Blosser, and Mr. John Kuchar for their technical assistance
`throughout the development of the dual-beam system. We thank
`Dr. Peter Munro for the long-term loan of the contrast-detail
`phantom. Initial interactions with Dr. Paul Cho and Dr. Cheng Pan
`are greatly appreciated. This development would not have been
`possible without the technical support provided by Elekta Oncol-
`ogy Systems. This project is funded in-part by the National Cancer
`Institute (CA-66074).
`Accepted for publication 25 March 1999.
`
`773
`
`Page 1 of 17
`
`Elekta Exhibit 1010
`
`

`
`774
`
`I. J. Radiation Oncology c Biology c Physics
`
`Volume 45, Number 3, 1999
`
`INTRODUCTION
`
`The escalation of tumor dose in radiotherapy promises
`increased probability of disease control (1). The proximity
`of the target to normal tissues makes dose escalation chal-
`lenging. Safe pursuit of these higher doses has spurred the
`development of conformal therapy techniques. These tech-
`niques attempt to conform the radiation field to a well-
`specified target volume. The success of the conformal ther-
`apy approach requires (i) accurate characterization of the
`uncertainties in field placement to create reasonable margins
`for use in the planning process, and (ii) verification of the
`treatment delivery.
`Numerous studies of field placement error have been
`reported in the literature over the past 10 years (2–5). These
`studies have demonstrated that on-line megavoltage portal
`imaging is useful in measuring field placement uncertainty.
`Many of these studies also report on the poor quality of the
`portal images and how the quality of the images hampers
`the detection of field placement errors (5).
`The primary reason for the poor quality of the megavolt-
`age portal images is the intrinsically low subject contrast of
`bony anatomy at megavoltage energies (6). The dramatic
`drop in contrast with increasing x-ray energy requires that
`the noise introduced by the imaging system be extremely
`low. In the case of conventional verification/localization
`film, the low contrasts are masked by film noise and the
`fixed display contrast of the film (7). Fluoroscopic systems
`suffer from poor light collection and electronics noise in the
`light sensor and readout electronics (8). While many inves-
`tigators continue to investigate methods of reducing noise in
`these systems, we have taken a more direct approach: to
`increase the subject contrast by using a kilovoltage (kV)
`x-ray source to localize the placement of the field.
`Using a kV x-ray source to determine field placement is
`not a novel concept. Johns et al. (9) report the design of a
`60Co treatment unit with a kV x-ray tube in the head to
`guide field placement. Investigators have reported on the
`addition of an x-ray tube to the head of an accelerator to
`generate kV radiographs of the patient in treatment position
`(10 –12). Other investigators have demonstrated that a low-
`energy x-ray beam can be produced by a low atomic num-
`ber, transmission target in the head of the accelerator (13).
`In 1992, Suit et al. (14) urged the community to develop
`better techniques for verifying field placement, proposing
`the addition of a kV source to the accelerator. In a previous
`report, we described the development of a prototype digital
`imaging system for kV localization on a medical linear
`accelerator (15).
`Beyond the obvious advantage of improved image quality
`for portal imaging, the development of an on-line kV im-
`aging system can provide other advantages. The dose de-
`livered in a kV exposure is significantly lower than that
`required for a megavoltage image. This reduction in dose
`allows more frequent imaging with open-fields, and the
`acquisition of images at non-treatment gantry angles, which
`may improve the precision of localization. In this article, we
`
`report on the characteristics and performance of an MV
`imaging system and a kV imaging system that have been
`integrated with a medical linear accelerator for radiographic
`localization. The dosimetric advantage of kV radiographs
`for localization is examined and the potential for tomo-
`graphic guidance is demonstrated with the system using
`cone-beam computed tomography (CB-CT). The combined
`system consisting of the medical accelerator and the two
`on-line imaging systems (kV and MV) is referred to as the
`dual-beam system (DBS).
`
`MATERIALS AND METHODS
`
`Dual-beam system
`Overview. Photographs of the DBS are shown in Fig. 1.
`Two fluoroscopic imager assemblies are attached to the
`accelerator; one detects the megavoltage (MV) treatment
`beam, the other detects the kV beam projected at 90° to the
`treatment beam axis. An Elekta SL-20 (Elekta Oncology
`Systems, Crawley, UK) linear accelerator forms the basis of
`the system. This accelerator is computer controlled and
`produces 6 and 18 MV photon beams. Field shaping is
`performed with an 80 leaf collimator. The SL-20 is a
`drum-based accelerator, making the installation of a retract-
`able kV x-ray source relatively straightforward (Fig. 1d).
`A 600,000 heat unit (HU) x-ray tube (Eureka Rad-92 in
`Varian Sapphire housing, 0.6 and 1.2 mm focal spots, 12.5°
`rotating anode; Varian X-ray Tube Products, Arlington
`Heights, IL, USA) with a manual collimation system has
`been installed on the SL-20 with the focal spot located 100
`cm from the machine isocenter. The tube is supported by 3
`hardened steel shafts (nominal 2.5” diameter) that retract
`into the accelerator’s drum structure (Fig. 1d). The shafts
`are supported by three bearing assemblies set in an alumi-
`num mount. The aluminum mount is attached to the face of
`the accelerator drum structure at the position corresponding
`to a 90° gantry angle. Adjustments are built into the assem-
`bly for focal spot alignment. The total weight of the x-ray
`tube and mechanical components is approximately 90 kg.
`Under this load, and at full extension, the predicted deflec-
`tion in focal spot position was approximately 0.2 cm. The
`x-ray tube is powered by a 40 kW high-frequency, radio-
`graphic generator (Innerscan Inc., Chicago, IL, USA). The
`exposure-to-charge ratio in air for this tube is measured
`to be 5.7 mR/mAs at isocenter (100 cm) for a 90 kVp
`spectrum. The tube spectrum has been modeled based on
`measured first and second HVLs of aluminum (Al) (HVL1
`5 9.5 mm) (16).
`5 4.0 mm; HVL2
`The MV radiographic imager assembly is shown sche-
`matically in Fig. 2 and is similar in design to that reported
`by Atari et al. (17). Briefly, a phosphor screen (Gd2O2S:Tb,
`Fuji HR-16 [back], 165 mg/cm2) mounted on a 1.5-mm
`stainless steel plate is used for primary x-ray detection. The
`light emitted by the phosphor screen is reflected by three
`front-surface mirrors and focussed by a lens (f/0.95 operat-
`ing at f/1.4, 50 mm) on to a 512 3 512 back-thinned
`charge-coupled device (CCD) (Sensor SI502A, Scientific
`
`Page 2 of 17
`
`

`
`Imaging systems integrated into a linear accelerator c D. A. JAFFRAY et al.
`
`775
`
`Fig. 1. Photographs of the dual-beam system. The dual-beam system was constructed on a Elekta SL-20 medical linear
`accelerator. The design allows normal operation of the accelerator with the imagers removed and the kV x-ray tube
`retracted (a). The imager can be easily attached for kV localization. The kV tube is extended manually to reach the
`source plane (b). The focal spot of the x-ray tube is at 100 cm SAD providing the same imaging geometry as that of
`the MV source. This provides a large aperture for conebeam CT with the system (c). The drum structure of the SL-20
`simplifies the addition of the retractable arm of the x-ray tube. Three hardened-steel shafts support the x-ray tube (d).
`
`Imaging Technologies [SITE] Inc., Beaverton, OR, USA).
`Back-thinning provides a quantum efficiency of 78% at 550
`nm (peak in the Gd2O2S:Tb spectrum). The CCD sensor is
`housed in a Photometrics CH-250 camera housing (Photo-
`metrics Inc., Tucson, AZ, USA). The sensor is cooled to
`240°C to minimize the dark current collected in the sensor
`during read-out. The sensor has a full-well capacity of
`280,000 electrons, corresponding to the full-range of the
`12-bit digitizer. A shutter is installed to minimize the effects
`of light leaks in the imager housing. The camera is read out
`under computer control at 500 kpixels/s. The long read-out
`period (0.5 s) of the camera prevents operation in a fluoro-
`scopic mode.
`The kV imager assembly is identical to the MV assembly
`with the exception of one less mirror in the optical path and
`the replacement of the 1.5 mm stainless steel plate with a
`thin Al plate (1/32”) for support of the phosphor screen. The
`kV imager assembly is a refined design of the original MV
`assembly; the third mirror has been eliminated through
`modifications to the mounting assembly. The presence of
`the third mirror does not significantly alter the performance
`
`of the MV system. In the case of the megavoltage detector,
`the stainless steel plate increases the interaction probability
`for the MV X rays and reduces the signal generated by
`scattered X rays exiting the patient. The imager housing
`(Fig. 1b) for the MV system has been modified to allow
`retraction by ;34 cm. The kV imager housing is non-
`retractable and is attached and removed before and after an
`imaging procedure.
`An in-depth characterization of the imaging performance
`of the kV and MV radiographic systems is on-going and is
`not the focus of this article.
`Control system. The DBS is slaved to the operation of the
`linear accelerator. A personal computer (Pentium-based)
`controls the x-ray generator and the acquisition of images
`and also monitors the gantry angle reported by the acceler-
`ator. A schematic of the control system is shown in Fig. 3.
`The generator is controlled via an RS-232 port, allowing
`full-control of the generator’s parameters and control of the
`exposure. The cameras are read via two ISA control cards in
`the PC. The gantry angle is monitored via the accelerator’s
`analog monitoring channel; the voltage reading is digitized
`
`Page 3 of 17
`
`

`
`776
`
`I. J. Radiation Oncology c Biology c Physics
`
`Volume 45, Number 3, 1999
`
`Fig. 2. Schematic of the imaging systems used to acquire the MV and kV radiographs. Both systems are based on a
`cooled CCD camera for collecting light emitted by a 165 mg/cm2 Gd2O2S:Tb screen. The kV and MV systems are
`identical with the exception of an additional mirror in the MV assembly and the use of different metal plates. The kV
`assembly is a newer generation of the MV design. The housing for the MV assembly has been altered to permit
`retraction of the screen and large mirror allowing the couch to lower to the floor completely.
`
`to 12 bits by an analog-to-digital board in the control PC.
`The gantry angle can be determined to within 60.1 degrees
`with this mechanism.
`
`System performance
`Mechanical characteristics of the system. The DBS was
`constructed to measure the patient position with respect to
`the reference frame of the medical linear accelerator. The
`localization accuracy and precision that can be achieved
`with this system will depend on the mechanical stability or
`
`Fig. 3. The kV and MV imagers are controlled via a PC. This
`figure shows the mechanisms used to control and monitor the
`different components of the system. The gantry angle of the
`accelerator is monitored via an analog voltage line provided
`through the accelerator’s interface. The images acquired by the
`system are stored in a distributed database. The same system is
`used to acquire the many projections used for the conebeam
`reconstructions.
`
`rigidity of the system’s components. To estimate the preci-
`sion with which high-contrast objects can be located, mea-
`surements were made of the range of flex for each of the
`four components of the DBS as the gantry was rotated
`through 360°. These four components include: the kV x-ray
`source, the MV x-ray source, the MV imager assembly, and
`kV imager assembly.
`The magnitude of accelerator drum wobble and eccen-
`tricity was determined by directing a small (,1 mm diam-
`eter) laser beam onto the face of the drum. The gantry was
`rotated through 360° and the path of the laser spot on the
`drum was plotted. The laser was then adjusted to the geo-
`metric centroid of the path and the process was repeated.
`After a few iterations, the maximum extent of the path was
`determined to be the magnitude of eccentricity at the face of
`the drum. The measurements were repeated with a rigid
`mechanical extension securely fastened to the face of the
`drum. The extension moved the plane of measurement 1 m
`out from the face of the drum (isocenter lies 1.25 m out from
`the face of the drum). This experiment, in conjunction with
`the results from the face of the gantry, allowed us to
`approximate the magnitude of accelerator (drum) wobble at
`the plane of the x-ray source orbit (isocenter).
`The run-out of the gantry (motion of the entire drum
`assembly along the axis of rotation) was also measured.
`This was done with the use of a dial gauge attached to the
`front face of the accelerator (near the axis of rotation). The
`tip of the dial gauge was placed in contact with the end of
`the accelerator’s couch. The tip of the gauge was positioned
`on the gantry’s axis of rotation through an iterative proce-
`dure similar to that described in the previous paragraph. The
`gantry was then rotated through 360° and the relative move-
`ment between the end of the couch and the drum face was
`
`Page 4 of 17
`
`

`
`Imaging systems integrated into a linear accelerator c D. A. JAFFRAY et al.
`
`777
`
`Table 1. Mechanical characteristics of the DBS components
`
`Table 2. Stages in imaging system and their corresponding gains
`
`Component
`
`Range
`(mm)
`
`Tangential
`Axial
`Radial
`Drum
`
`MV
`source
`
`MV
`imager
`
`kV
`source
`
`kV
`imager
`
`1.5
`1.1
`2.0
`0.6
`
`4.0
`1.1
`5.0
`0.6
`
`4.0
`1.1
`5.0
`0.6
`
`1.5
`1.1
`1.5
`0.6
`
`The measured range of motion of each of the system’s four
`components. To establish a conservative estimate for localizing
`precision, it is assumed that each of these motions are completely
`independent. The ability to correct for these motions depends on
`their reproducibility as a function of gantry angle; preliminary
`investigations suggest that such corrections could be applied reli-
`ably.
`
`recorded. Care was taken to verify that the couch was rigid
`and immobile as the gantry rotated.
`To estimate the range of flex of the two x-ray source
`assemblies and the two imager assemblies, a solid state laser
`was mounted rigidly on the face of the drum and its beam
`(after passing through a cylindrical lens) was directed onto
`each of the four system components in-turn. As the gantry
`was rotated, the intersection of the line projected by the
`laser with the component was recorded on a small strip of
`paper attached securely to the surface of the component.
`The range of relative movement between the projected laser
`beam and the structure was taken to be due to flex in the
`component. The laser (and projected line) was then rotated
`and the measurement was repeated in the orthogonal dimen-
`sion (Table 1). It is important to note that the results of this
`measurement present a combination of the component’s flex
`and any shearing of the assembly. In addition, they do not
`determine the path of the component but provide a reason-
`able estimate for the range of the non-rigid motion. A more
`complete characterization of component flex is on-going, in
`which the movements of each component are characterized
`in three separate dimensions (radial, longitudinal, and an-
`gular) as a function of the nominal gantry angle (18).
`With these estimates of component flex, the geometric
`precision of the DBS for three-dimensional localization was
`estimated assuming an orthogonal-pair imaging application,
`which uses either the MV or kV system. The calculations
`were performed assuming the three-dimensional point of
`interest is near machine isocenter. Using the uncertainties in
`the viewing geometry introduced by the mechanical non-
`rigidity of the system,
`it was possible to calculate the
`localizing precision using the following method. At a given
`gantry angle, the range of angular motion of the source (kV
`or MV) and the range of angular motion of its correspond-
`ing detector assembly define a truncated sector in the iso-
`center plane. In an orthogonal view, the same ranges of
`motion define a second truncated sector in the isocenter
`plane. The extent of the area represented by the intersection
`of these two truncated sectors is taken as a measure of the
`localizing precision. Small ranges of radial motion (toward
`
`Stage in
`imaging
`system
`
`Gain
`(quanta exiting
`stage/quanta
`entering stage)
`
`1
`
`2
`
`3
`
`4
`
`5
`
`6
`
`g1 (interacting X ray/
`incident X ray)
`g2 (optical quanta
`produced/interacting
`X ray)
`g3 (fraction of optical
`quanta escaping
`screen)
`g4 (fraction of optical
`quanta reflected by
`mirrors while en
`route to lens)
`g5 (fraction of optical
`quanta collected by
`lens)
`g6 (quantum efficiency
`of CCD)
`
`Detector and X ray energy
`
`Megavoltage
`imager
`(6 MV)
`
`Kilovoltage
`imager
`(90 kVp)
`
`0.025
`
`0.67
`
`1.96 3 104
`
`3.05 3 103
`
`0.7
`
`0.7
`
`0.83
`
`0.88
`
`1.36 3 10-4
`
`1.42 3 10-4
`
`0.78
`
`0.78
`
`Listing of the gain stages in the imaging system. Scattering
`stages have not been included in this analysis. The values used for
`each stage were determined directly or taken from the literature.
`
`or away from isocenter) of the components have negligible
`effects on the shape and location of the intersection defined
`here, and, therefore, have been ignored in this analysis.
`These calculations are performed to indicate the worst-case
`condition in which no corrections for mechanical flex have
`been applied.
`
`Radiographic imager performance: kilovoltage and
`megavoltage
`The kV and MV imaging systems operate on the same
`principle: the interaction of an x-ray within the phosphor
`screen (or metal plate/phosphor screen) produces light,
`which is collected by a high-speed lens and focussed onto
`an efficient, low-noise CCD. Each stage in the two systems
`has been characterized to examine the transfer of signal
`through the imaging chain. Table 2 lists the six stages
`included in the analysis. The gain or interaction probability
`for each stage has been estimated or taken from the litera-
`ture. From these values, it can be determined whether the
`two systems are operating near their theoretical limit of
`performance, as dictated by the noise associated with x-ray
`quanta used to form the image. This analysis is restricted to
`low-spatial frequencies and does not consider the small
`amount of additive noise introduced by the CCD camera.
`The gain estimates for the first stage (g1) were determined
`using Monte Carlo techniques (19) for the MV detector at 6
`MV (20) and using a spectral model (90 kVp) for the
`kilovoltage detector (16). The conversion of x-ray energy to
`light was determined from the Monte Carlo generated (MV)
`or analytical (kV) estimates of energy deposition in the
`phosphor layer. For the kV detector, K-fluorescence and
`re-absorption was approximated using the results of Chan et
`
`Page 5 of 17
`
`

`
`778
`
`I. J. Radiation Oncology c Biology c Physics
`
`Volume 45, Number 3, 1999
`
`Table 3. Measured and calculated gains for kV and MV imaging
`systems
`
`Detector and X-ray Energy
`
`Kilovoltage
`imager with
`90 kVp spectrum
`
`Megavoltage
`imager with
`6 MV spectrum
`
`0.125
`
`0.139
`
`1.11
`0.19
`
`0.036
`
`0.030
`
`0.83
`1.43
`
`Gain
`
`Measured (e2 detected in
`CCD/incident X ray)
`Calculated e2 detected in
`CCD/incident X ray)
`Calculated/measured
`Measured* (e2 detected in
`CCD/interacting X ray)
`
`The total gain of the two imaging systems for 6 MV and 90 kVp
`x-ray beams. The total gain of the system has also been calculated
`and comparison is made to measurement. Excellent agreement
`between the measured and calculated values was found.
`* Note: This estimate is based on the measured gain and the
`estimates of g1 presented in Table 2.
`
`al. (21) The number of optical quanta was calculated as-
`suming a conversion efficiency of 15% for the Gd2O2S:Tb
`screen and an average optical photon energy of 2.3 eV (550
`nm, green light, peak in emission spectrum). The probabil-
`ity of light escape from the phosphor was taken to be 0.7;
`this is the average of the results reported by Bissonnette et
`al. (22) and Ginzburg and Dick (23) for a Lanex Fast (back)
`screen (133 mg/cm2). The fraction of light reaching the
`CCD was determined by the reflectance of the front-surface
`mirrors (BV-2, Optical Coating Laboratories Inc., Santa
`Rosa, CA, USA) (g4) and the collection efficiency of the
`lens (g5) (8). The quantum absorption efficiency of the CCD
`at 550 nm was measured to be 0.78 (Calibration performed
`by Photometrics Inc., Tuscon, AZ, USA). The electronic
`gain of the camera electronics was measured to be 75.7 e2
`and 82.0 e2 per analog-to-digital converter unit (ADCU) at
`a gain setting of 1X for the MV imager and kV imager,
`respectively. From these gains, the quantum accounting
`diagrams (QADs) for the two systems are generated (24).
`These diagrams indicate the relative magnitude of quantum
`noise introduced at each stage in the system. Ideally, the
`great majority of the quantum noise should be introduced by
`the x-ray fluence itself. If this is the case, the system is
`considered “x-ray quantum-noise limited.”
`To verify that the estimates for gain are correct, the total
`gain of each system (electrons/incident x-ray) was measured
`under controlled exposure conditions. The fluence was es-
`timated using Rogers’ (25) dose-to-fluence conversion ta-
`bles (for 6 MV) and the Tucker and Barnes (16) model for
`simulating kV spectra (for 90 kVp). The total gains for both
`systems are plotted at the final stage in the QAD and are
`included in Table 3.
`
`Contrast-detail performance
`The contrast-detail (C-D) performance of both the kV and
`the MV radiographic imaging systems were measured (26).
`
`Transmission images of an aluminum C-D phantom (1.29
`cm thick) in 27 cm of water (lucite container) were acquired
`at a fixed imaging geometry (87 cm SSD, 160 cm SDD, 123
`cm from source to Al plate). The 27 cm water bath was
`included to simulate the scatter and absorption in a clinical
`imaging condition. Two MV images were acquired with 10
`monitor units at both 6 MV and 18 MV x-ray energies.
`Single kV images were acquired at 90 kVp using 10, 20, 30,
`40, 60, and 120 mAs exposures. The maximum dose for the
`5 1.5 cm) for
`megavoltage exposures were 13.6 cGy (dmax
`5 3 cm) for 18 MV. The
`6 MV and 13.7 cGy (dmax
`maximum dose for the kilovoltage exposures (correspond-
`ing to the entrance dose) determined from in-air measure-
`ments was 0.097, 0.195, 0.29, 0.39, 0.58, and 1.17 cGy for
`the six exposure levels. The in-air exposures were converted
`to dose assuming an approximate value of 0.9 for the
`exposure-to-dose conversion factor (fmed) and a measured
`back-scatter factor of 1.42 for 90 kVp (16 cm 3 18 cm field
`size at isocenter). Six observers were asked to locate the
`minimum observable diameter for each contrast level (depth
`of water-filled aluminum well). Each observer was allowed
`to vary the ambient room light level and was allowed to
`vary the display contrast to their liking. The display mag-
`nification was fixed. The average minimum detectable hole
`diameter was calculated across all six observers. At low
`contrasts and low doses, some observers could not detect a
`minimum hole diameter (i.e., no hole was large enough); if
`2 or more of the 6 observers were unable to detect anything
`at a given contrast level, that contrast level was removed
`from the analysis.
`For the 6 and 18 MV beams, the subject contrasts were
`determined from measured attenuation coefficients for wa-
`ter, aluminum, and lucite. For the 90 kVp images, the model
`of the x-ray spectrum was used to calculate the subject
`contrasts for each well in the phantom. The subject contrasts
`are calculated using the ‘difference over the mean’ tech-
`nique (27).
`The contrast-detail performance of the megavoltage and
`kilovoltage systems were then compared to demonstrate the
`dosimetric advantage of kV radiography over MV radiog-
`raphy in detecting bony structures. The contrast-detail
`curves for the MV and kV systems were compared on a
`common scale: depth of water-filled aluminum well. The
`MV and kV doses that produced equivalent object detection
`were determined for both 6 and 18 MV using intermediate
`contrasts and hole diameters. The use of the aluminum
`phantom (as opposed to bone) for this measurement is
`reasonable; the ratio of linear attenuation coefficients for Al
`to bone is constant within 610% over x-ray energies rang-
`ing from 10 keV to 20 MeV (28). From this comparison, the
`increase in dose (at depth of maximum dose) required at 6
`and 18 MV to achieve the same contrast visibility of a
`bone-like structure at 90 kVp was determined. It should be
`kept in mind that, in this study, three factors have a signif-
`icant influence on visibility: (i) the change in subject con-
`trast of the Al (nominally bone-equivalent) phantom with
`x-ray energy; (ii) the change in transmitted x-ray fluence;
`
`Page 6 of 17
`
`

`
`Imaging systems integrated into a linear accelerator c D. A. JAFFRAY et al.
`
`779
`
`and (iii) the difference in detector performance between the
`kV and MV imaging systems. Strictly speaking, the dosi-
`metric advantage determined here only applies to these two
`imaging systems and this specific imaging task.
`The clinical operation of the two imaging systems (mega-
`voltage and kilovoltage) was demonstrated on patients re-
`ceiving radiation therapy. The number of monitor units used
`to acquire each MV image is reported with each image (see
`captions). The exposure (in-air) for each kV radiographic
`exposure was measured using a PMX-3 electrometer (RTI
`Electronics AB, Molndal, Sweden) and an energy compen-
`sated radiation diode (R-25). (The dose is reported in the
`figure caption for each image.) For some fields, both kV and
`MV images were acquired. These have been fused to dem-
`onstrate a potential use of the DBS for producing kV open-
`field double exposures.
`Conebeam tomographic imaging. In addition to radio-
`graphic localization, the DBS has been designed to allow
`conebeam tomographic localization of soft-tissue structures
`in the reference frame of the accelerator. The high-fre-
`quency x-ray generator has a high duty cycle (20%) that
`allows multiple (;200) radiographs to be acquired in rapid
`succession (1 per second). To generate a CB-CT dataset, a
`series of radiographic exposures are acquired at regular
`angular intervals as the accelerator gantry is rotated through
`a specified range (typically 180° or 360°). The control
`system operates the camera’s shutter and read-out mecha-
`nisms in synchrony with the firing of the x-ray generator;
`these operations are slaved to the accelerator’s gantry angle.
`The accelerator gantry rotates continuously at a relatively
`low speed (;30°/min) with the control system monitoring
`the gantry angle. When the gantry angle is within a pre-set
`tolerance (60.1°) of the desired angle, the camera shutter is
`opened, the generator delivers a short exposure (50 ms), and
`the image is read out and stored to disk. Minor variations in
`gantry speed do not affect the acquisition, and the short
`exposure period (50 ms) introduces negligible blurring in
`the radiographs at these gantry speeds.
`The energy of the x-ray spectrum used for the conebeam
`acquisitions is typically 120 kVp and each image is pro-
`duced with approximately 3 mAs of charge. A 180 projec-
`tion scan would deliver a total charge of 540 mAs to the
`anode. In addition to the projection images, 20 bright-field
`images and 20 dark-field images are taken following acqui-
`sition. Each projection image is corrected by subtracting the
`dark-field signal (light leaks, digital offset, and dark-current
`in the CCD) and by dividing out the non-uniformities in the
`bright-field. No corrections for uncertainties in the scanning
`geometry due to flex or warping were applied. No scatter
`compensation algorithms were used in the reconstruction of
`these images.
`The 180 images are transferred via network to a work-
`station for reconstruction. A volumetric CB-CT dataset is
`reconstructed on a 70 MHz SparcStation 20 using a filtered
`back-projection algorithm based on that first described by
`Feldkamp et al. (29,30). The filtering process takes app

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket