throbber
The Sensitivity of Low Flip Angle RARE Imaging
`
`David C. Alsop
`
`It is demonstrated that the stability of the Carr-Purcell-Mei-
`boom-Gill (CPMG) sequence reflects the existence of a steady
`state solution to the Bloch equations in the absence of T2 and
`T, decay. The steady state theory is then used to evaluate the
`performance of low flip angle RARE imaging sequences with
`both constant and optimally varied refocusing flip angles. The
`theory is experimentally verified in phantoms and then opti-
`mized, single shot, low flip angle RARE is used to obtain
`artifact-free images from the brain of a normal volunteer.
`Key words: RARE; fast imaging; magnetic resonance imaging.
`
`INTRODUCTION
`RARE imaging (1) is a method to increase the speed of
`spin-echo imaging by acquiring a series of spin echoes
`with different phase encodings after each excitation. Be-
`cause the technique offers either shorter scan times or
`higher signal-to-noise ratio than standard spin-echo im-
`aging while producing similar contrast (2, 3), RARE im-
`aging has become a clinical success. The faster speed of
`RARE imaging has allowed increased flexibility in choos-
`ing TR for optimal contrast and has improved the feasi-
`bility of inversion recovery preparation (4). In clinical
`applications, usually fewer than 16 echoes are employed,
`although in specialized applications where long T2 fluid
`is being imaged, such as MR cholangiography (5), longer
`echo trains are feasible.
`It is straightforward to further increase the speed of
`RARE imaging by acquiring a larger number of spin ech-
`oes but this approach eventually leads to several prob-
`lems. One major concern is RF power deposition within
`the subject. A long series of compact 180’ pulses can
`readily introduce more RF heating than is acceptable
`according to normal safety guidelines, particularly if
`multiple slices are being acquired. A long train of 180°
`pulses also increases the interaction between neighbor-
`ing slices and can lead to undesirable loss of contrast and
`signal. At the power levels employed in long echo train
`RARE imaging, magnetization transfer saturation (6) is
`quite significant and can increase the effective slice in-
`teraction. Finally, unless all the echoes are acquired
`within a time on the order of T2, significant distortion of
`the point spread function of the image will occur (I,?’, 8).
`This requirement places a strong restriction on the length
`of an echo train.
`
`MRM 37~176-184 (1997)
`From the Department of Radiology, University of Pennsylvania Medical
`Center, Philadelphia, Pennsylvania.
`Address correspondence to: David Alsop, Ph.D., Department of Radiology,
`University of Pennsylvania Medical Center, 3400 Spruce Street, Philadel-
`phia, PA 19104-4283.
`Received July 16. 1996; revised September 24, 1996; accepted September
`26, 1996.
`This research was supported in part by a Biomedical Engineering Research
`Grant from the Whitaker Foundation.
`0740-31 94/97 $3.00
`Copyright 0 1997 by Williams & Wilkins
`All rights of reproduction in any form reserved.
`
`These limitations of RARE imaging were recognized by
`one of its inventors (9) and the use of reduced flip angles
`for the refocusing pulses was proposed as a solution.
`When these reduced flip angle pulses are used, the echo
`amplitude becomes a complex combination of stimulated
`echoes and spin echoes. It was demonstrated with nu-
`merical simulations and experiments in long T2 speci-
`mens that the spin echo amplitude approaches a tempo-
`rary steady state which then slowly decays due to T2 and
`TI relaxation. Because the “steady state” is actually only
`temporary, it will be referred to as a pseudosteady state.
`The empirically determined pseudosteady state echo am-
`plitude was well approximated by the sine of half the
`refocusing flip angle. The use of a reduced flip angle
`dramatically decreases the power deposited by the se-
`quence and also reduces the slice interaction.
`A method for optimizing the flip angles of the first few
`refocusing pulses to achieve a constant amplitude echo
`train was later presented by LeRoux and Hinks (10). They
`demonstrated that beginning the RF pulse train with
`higher amplitude pulses that slowly decrease and ap-
`proach a constant, or asymptotic, flip angle can produce
`a constant echo amplitude from the very first echo. Al-
`though this work was primarily focused on stabilizing
`the signal when large flip angle, slice selective pulses
`were employed, the approach also eliminated the pri-
`mary drawback of constant flip angle, low flip angle
`RARE: the large number of echoes which have to be
`discarded until the pseudosteady state is reached. In
`addition, the optimized flip angle approach produced
`pseudosteady state echo amplitudes that were consider-
`ably higher than when a corresponding constant flip
`angle RF train was employed.
`The dependence of the pseudosteady state echo ampli-
`tude on the flip angles of the early pulses in the RF pulse
`train raises two questions whose answers will determine
`the quality and utility of low flip angle RARE imaging:
`What is the highest pseudosteady state echo amplitude
`attainable for a given asymptotic flip angle and what RF
`pulse trains can achieve this amplitude? Below an ana-
`lytical solution for the pseudosteady state spin echo train
`is derived that provides an upper limit for the attainable
`pseudosteady state echo amplitude. The solution is then
`employed to assess the efficiency of different RF pulse
`trains for producing magnetization in the pseudosteady
`state condition and to evaluate the utility of low flip
`angle RARE imaging. Finally the conclusions are evalu-
`ated experimentally in both phantoms and a normal vol-
`unteer.
`
`THEORY
`Pseudosteady State Solution
`A steady state solution to the Bloch equation has been
`derived for the short TR gradient echo sequence by fol-
`lowing the evolution of the magnetization from one ex-
`176
`
`General Electric Co. 1004 - Page 1
`
`

`
`The Sensitivity o/ Low Flip Angle RARE Imaging
`
`citation to the next and requiring that the net change is
`zero (11). A similar method will be used below for the
`spin echo pseudosteady state. Because the spin echo
`pseudosteady state is not stable against T, and T2 decay,
`terms involving T,, T, and consequently M, in the Bloch
`equations will be ignored. Because RF pulses cause only
`rotation, the effect of the RF pulse can be expressed in
`terms of three Euler angles: a rotation around the z axis of
`angle p followed by a rotation around the y axis of angle
`a, and finally a rotation around the z axis of angle 6. This
`can be written in matrix form (12) as
`
`i;
`
`cos2- exp i (p + 6)
`a
`
`sin a
`exp ip
`
`I
`
`cy
`-sin'- exp -i(p - 6) -sin a exp i6
`2
`a
`cosz- exp -i(p + 6) - sin a exp -i6
`2
`sin cy
`2 ___ exp -ip
`
`cos a
`
`where M, is the complex transverse magnetization given
`by
`
`Mt = M, + iMy
`121
`Define the magnetization as equal to M at echo 1. The
`magnetization experiences a phase rotation of 412 fol-
`lowed by the RF pulse and a second rotation of 412 before
`echo 2. The phase rotation, 4, is caused by applied slice
`select and readout gradients as well as frequency offset
`from the transmitter frequency. This can be written as
`
`131
`
`is equal to
`The steady state requirement demands that
`M. This system of equations can be solved directly, see
`Appendix, to yield
`M*, = -exp i(p - 6)M,
`
`[41
`
`i s i n a e x p i ( y ) s i n ( P + S + + )
`i sina exp i ( 7 ) sin (
`)
`
`P + S + +
`
`P - 6
`
`177
`
`M, =
`
`(1 - coscy)
`
`Equation [4] determines the phase of the echo as a func-
`tion of the phase of the RF pulse and is a statement of the
`Meiboom-Gill condition (13) for the stability of a spin
`echo train. Equation [5] determines the fraction of the
`magnetization along the longitudinal direction. If one
`assumes that all of the equilibrium magnetization, M,,,
`has been placed in the pseudosteady state condition,
`then the transverse magnetization is given by, see Ap-
`pendix,
`
`-7 'I i
`
`(1 - cosa)
`
`\
`
`+
`The echo amplitude is given by averaging this expression
`over 4. This average can be easily performed numerically
`but by making use of special functions and integral tables
`(14). one can also find an analytical solution. see Appen-
`dix,
`
`Me,,,, = +-i Mo exp -i
`
`1
`
`where P is a Legendre function.
`In Fig. 1, the theoretical pseudosteady state function,
`Eq. [7], is compared with the empirical formula for the
`amplitude when a constant flip angle RF pulse train is
`employed (9). The theoretical echo amplitude is larger
`than the constant flip angle echo amplitude for all flip
`angles. This discrepancy arises from the major unjusti-
`fied assumption in the theoretical derivation: that all of
`the equilibrium longitudinal magnetization is transferred
`to the pseudosteady state. There is an inefficient transi-
`tion to the pseudosteady state when a constant flip angle
`RF pulse train is used. It is possible to derive the echo
`amplitude produced by a constant flip angle RF pulse
`train by assuming that all magnetization not initially
`aligned with the pseudosteady state condition eventually
`dissipates. The 90° excitation pulse produces a purely
`transverse magnetization. Assuming the phase of the ex-
`citation pulse is chosen to satisfy the Meiboom-Gill con-
`dition (13), Eq. [4], the component of the excited magne-
`tization along the pseudosteady state solution
`is
`determined by the angle between the pseudosteady state
`solution, M"", and the transverse plane.
`
`General Electric Co. 1004 - Page 2
`
`

`
`178
`
`Alsop
`
`best to consider optimization of efficiency and stabiliza-
`tion of the echo amplitude simultaneously.
`The method of Hennig (9) for calculating the echo
`amplitudes of an arbitrary RF pulse train was imple-
`mented. Following LeRoux and Hinks (lo), the method
`was modified so that the flip angle of each subsequent RF
`pulse could be determined by specification of the desired
`amplitude of the echo produced by the pulse. To make
`possible the specification of other than constant echo
`amplitudes, an iterative determination of the flip angle
`was used in place of the exact solution developed by
`LeRoux and Hinks for the constant echo amplitude case.
`Beginning with the first echo, an initial guess for the flip
`angle was chosen. The amplitude of the echo was deter-
`mined and then the guess for the flip angle was revised
`based on an iterative root finding algorithm. For simplic-
`ity only nonselective pulses were considered in the cal-
`culations. Figure 1 summarizes the results of the calcu-
`lations. The theoretical pseudosteady state solution
`provides an upper bound to the attainable efficiency. The
`constant flip angle approach is clearly less efficient, es-
`pecially at low flip angles. Amplitude stabilized echo
`trains, on the other hand, are highly efficient. An RF
`pulse train designed to produce a constant echo ampli-
`tude generates a signal very close to the optimal value at
`moderate to high asymptotic flip angles but the efficiency
`starts to drop at low asymptotic flip angles. The relatively
`high efficiency of the amplitude stabilized train can be
`attributed to the gradual decrease of the flip angle from
`angles near 180' at the first echo to the lower, constant
`flip angle used at large echo numbers. An example of an
`optimized RF pulse train is plotted in Fig. 2.
`Intuitively, one would expect that increasing the flip
`angle of the first few RF pulses would further improve
`the efficiency of the echo train because the transition
`from transverse to pseudosteady state angles would be
`even more gradual. This can be achieved by specifying
`larger echo amplitudes for the first few echoes. RF pulse
`
`i
`
`- b- - Stab\lhzed, w l t h Ramp
`0 Stabilized, no Ramp
`
`a
`.- -
`L L
`
`0
`
`5
`
`10
`Echo Number
`
`15
`
`20
`
`FIG. 2. Examples of echo amplitude stabilized RF pulse trains
`designed to achieve amplitudes of 0.4 Mo. The constant echo
`amplitude RF pulse train, open squares, is slightly less efficient
`than the RF pulse train designed for an initial downward ramp in
`echo amplitude, open triangles, so a slightly larger asymptotic flip
`angle is required to achieve the same echo amplitude. Because of
`the initial downward ramp in echo amplitude, the ramp stabilized
`RF pulse train begins with a larger flip angle.
`
`2 0.4
`Ln
`x
`U
`
`. I
`
`o . o & , ,
`0
`
`,
`
`.
`
`,
`
`,
`
`,
`
`,
`
`~
`
`,
`
`- - - - - -
`-
`,
`,
`50
`100
`Steody State Flip Angle (degrees)
`
`Steady State Theory
`Stabilized, w i t h Romp
`StobiLzed. no Romp
`Constant Flip Angle
`.
` , 1
`L
`,
`150
`
`,
`
`,
`
`FIG. 1. Pseudosteady state RARE echo amplitudes as a function
`of the asymptotic refocusing flip angle. The theoretical pseudoste-
`ady state amplitude, solid line, represents the maximum possible
`amplitude achievable. The asymptotic echo amplitudes for a con-
`stant flip angle RF pulse train, a train optimized for constant echo
`amplitude, and one optimized for a short downward ramp in echo
`amplitude followed by constant echo amplitude are also plotted.
`The optimized RF pulse trains come closest to producing the
`optimal echo amplitude.
`
`= M o 1 + i i
`
`sina sin ( e ~ l ) j - '
`
`(1 -cosa)
`The average of this complex expression over all 4 is
`simply given by (14)
`
`in agreement with the earlier empirical result (9) .
`
`Optimization of Echo Amplitudes
`In the previous paragraph it was shown that the ineffi-
`ciency of the constant flip angle RF pulse train reflects
`the difference in angular orientation of spins in the
`pseudosteady state condition compared with the spins
`placed in the transverse plane by the excitation pulse.
`One way to improve this efficiency is to slowly lower the
`flip angle of the RF pulse train from 180°, for which the
`pseudosteady state solution is in the transverse plane, to
`the desired asymptotic angle so the difference between
`the angle of the actual spins and the pseudosteady state
`solution is minimized. Because the flip angles used at the
`beginning of the echo train also have an impact on vari-
`ations in the spin echo amplitude from echo to echo, it is
`
`General Electric Co. 1004 - Page 3
`
`

`
`The Sensitivity of Low Flip Angle RARE Imaging
`
`179
`
`trains were therefore also calculated to produce a linear
`decrease in signal amplitude from echo one to echo three
`and an echo amplitude equal to the third echo amplitude
`for subsequent echoes. The slope of this amplitude ramp
`was chosen such that the amplitude of echo zero, if it
`existed, would be M,,. An example of such a ramp opti-
`mized RF pulse train is plotted in Fig. 2. The pseudoste-
`ady state echo amplitude achieved with this optimiza-
`tion approach is also plotted in Fig. 1 but it is so nearly
`identical to the theoretical optimum that it is partially
`obscured. Ramp optimized RF pulse trains are more ef-
`ficient than constant optimized RF pulse trains but the
`higher echo amplitudes of the first two echoes must be
`magnitude corrected prior to image reconstruction.
`
`Effects of T2 and T, Decay
`The echo trains produced with multiple spin echoes are
`not truly in a steady state because T, and T, ultimately
`cause the decay of the signal. For perfect 180' pulses,
`only T, affects the amplitude but when lower flip angles
`are used, TI decay also plays a role because stimulated
`echoes are present. Once an RF pulse train has been
`selected, it is straightforward to insert T, and T, decay
`into the calculation of echo amplitudes. The echo ampli-
`tudes of the resulting echo trains do not decay as a
`perfect exponential. The effective T, for the decay, Tzeff,
`was therefore defined as
`
`-\-, *
`
`where S(t, T2, T,, a ) is the echo amplitude as a function
`of time, T2, T, and the refocusing flip angle, a. This
`definition for TZeff was chosen as the most appropriate
`for the discussion of signal-to-noise ratio which follows,
`and also yields the correct relaxation rate for an expo-
`nential decay. In Fig. 3, the ratio of TZeff to T, is plotted
`versus the ratio of T, to T2 for several constant optimized
`RF pulse trains with different asymptotic flip angles.
`When T , is long compared with T,, TZeff is longer than T,
` '
`' 1
`
`5 L "
`
`.
`
`'
`
`"
`
`'
`
`"
`
`"
`
`'
`
`
`
`" 1
`
`'
`
`'
`
`0
`
`50
`100
`Flip Angle ( degrees )
`
`150
`
`FIG. 4. The calculated signal-to-noise ratio versus asymptotic flip
`angle calculated by using Eq. [ll] and the TZe,;s plotted in Fig. 3.
`Signal-to-noise ratio decreases only very slowly with the refocus-
`ing flip angle.
`
`but still relatively independent of T, except at very small
`asymptotic flip angles. When T, becomes comparable
`with T,, the contribution of T, to TZeff is considerable.
`
`Signal-to-Noise Ratio of Low Flip Angle RARE
`Although the use of lower flip angles in RARE imaging
`necessarily reduces the acquired echo amplitude, the
`signal-to-noise ratio need not decrease proportionately.
`Because the TZeff of low flip angle RARE is longer than
`that of T,, there is more time to acquire signal. Conse-
`quently, signal averaging of echoes or, alternatively, nar-
`rower bandwidth acquisitions can be used to reduce the
`measured noise. Employing the matched filter approach
`of Ernst and Anderson (11), the relative signal-to-noise
`ratio of low flip angle RARE is given by
`
`1
`
`The definition of Tzeff was chosen so that this expression
`would be true. This function is plotted in Fig. 4 for
`several different T, values. For objects with TI to T2
`ratios of 10, which is typical for tissues at moderate to
`high field, the signal-to-noise ratio decreases only very
`slowly with asymptotic flip angle. This argues that low
`flip angle RARE may be very competitive with other
`acquisition methods.
`Equation 1111 has neglected the effect of the acquisition
`time on the time available for signal recovery prior to
`another excitation. Many of the applications for low flip
`angle RARE, including fast T, and proton density imag-
`ing, require TR to be much longer than T I . In these
`applications, the effect of the slightly longer acquisition
`on signal-to-noise ratio should be small but will depend
`on the TR selected for the application
`
`84 degrees
`
`-
`
`1
`0
`
`5
`
`.
`
`I
`10
`T 1 / 7 2
`
`.
`
`.
`
`.
`
`,
`15
`
`.
`
`.
`
`.
`
`.
`
`
`
`20
`
`FIG. 3. The decay rate of the echo amplitudes, TZeff, is plotted
`versus the ratio of T, to T2 for several RF pulse trains with different
`asymptotic flip angles. The TZeff was calculated from simulations
`of constant echo amplitude optimized RF pulse trains.
`
`METHODS
`To experimentally verify the theory and evaluate the
`image quality, the low flip angle RARE sequence was
`
`General Electric Co. 1004 - Page 4
`
`

`
`180
`
`Alsop
`
`implemented on a clinical GE SIGNA 1.5 Tesla scanner
`equipped with a prototype gradient system. The gradient
`system could achieve gradient amplitudes of 23 mT/m
`and gradient switching speeds of 100 ps. The strong and
`fast gradients made possible very compact echo trains.
`To further compress the echoes, a Hamming windowed
`1.6 ms sinc pulse that was truncated at the first zero
`crossing was employed. With a 64 kHz acquisition band-
`width and a 256 frequency matrix, echo spacings of
`slightly under 5 ms were achievable.
`Single-shot images were acquired by using ramp opti-
`mized RF pulse trains. An acquisition matrix of 256 X 80,
`an FOV of 24 X 15 cm, and a slice thickness of 5 rnm
`were employed. To avoid artifacts in the reconstruction
`due to the higher amplitudes of the first two echoes, a
`total of 82 echoes were acquired and the first two were
`discarded. The phase encode ordering of Melki et 01. (2)
`was used for all imaging. The acquisition of a single
`image required 410 ms.
`To confirm the theoretical calculations, images of a
`uniform long T, phantom were acquired with the phase
`encoding gradient turned off. The amplitude of each echo
`was then calculated. Such data were obtained for trains
`with asymptotic flip angles ranging from 180" to 1 7 O . For
`comparison to the nonselective pulse theory, the ampli-
`tude of the slice select gradient for the refocusing pulses
`was reduced by a factor of four to achieve a more uniform
`flip angle across the slice. To avoid artifacts, the crusher
`gradients in the slice select direction were increased.
`To assess the image quality, in vivo single-shot images
`were obtained in the brain of a normal volunteer. Flip
`angles from 90" to 17" were evaluated. At these flip
`angles, the images could be acquired with a TR of 500 ms
`without exceeding manufacturers safety limits on heat-
`ing. Higher flip angles were not evaluated in vivo.
`
`RESULTS
`Phantom measurements confirmed the validity of the
`theory. Echo amplitudes are plotted in Fig. 5 for several
`different asymptotic flip angles. The initial amplitudes of
`
`0.0
`0
`
`10
`
`20
`Echo Number
`
`30
`
`40
`
`FIG. 5. Experimental echo amplitudes from a long T2 phantom
`using RF pulse trains optimized for a ramp down to a constant
`echo amplitude. Echo amplitudes from RF pulse trains with as-
`ymptotic flip angles of 1 B O O , go", 60", 30", and 17" are plotted.
`
`the echo trains are in excellent agreement with the the-
`ory. Also apparent in the figure is the excellent stability
`of echo amplitude that is possible with optimized RF
`pulse trains.
`Axial single-shot images of the brain of a normal vol-
`unteer are shown in Fig. 6. The images demonstrate high
`signal-to-noise ratio and good spatial resolution. As ex-
`pected, the image signal intensity for short TE images
`decreases with refocusing flip angle but signal-to-noise
`ratio is still fairly high with 30" refocusing pulses. T2-
`weighted images are also shown, demonstrating the flex-
`ibility in selecting T2 contrast with the RARE sequence.
`Differences in the degree of T2 weighting are apparent in
`the images. This is best illustrated in Fig. 7 where Tzeff
`decay is shown for both 90" and 17" asymptotic refocus-
`ing flip angles. The short TE images were windowed for
`identical appearance but longer TE images show much
`more T, weighting for the 90" flip angle than the 17"
`images. This is consistent with the increase in TZoff ex-
`pected in tissues with T, much longer than T2.
`
`DISCUSSION
`A theory for the pseudosteady state signal amplitude in
`spin echo trains has been presented that provides a the-
`oretical optimum signal intensity against which practical
`implementations can be compared. This theory sheds
`new light on the stability of CPMG (13, 15) echo trains
`and allows derivation of the asymptotic signal amplitude
`as a function of refocusing flip angle for such echo trains.
`For the specific purpose of optimizing low flip angle
`RARE imaging, the theory has shown that constant am-
`plitude low flip angle echo trains are inefficient at trans-
`ferring signal intensity into the pseudosteady state con-
`dition. This inefficiency adds to the problems associated
`with the gradual approach of the echo amplitudes to the
`pseudosteady state. Flip angle optimized RF pulse trains
`are clearly superior both at producing a pseudosteady
`state rapidly and converting transverse magnetization
`into the pseudosteady state. RF pulse trains optimized
`for constant echo amplitude produce nearly the theoret-
`ical maximum signal amplitude at intermediate to high
`asymptotic flip angles but at lower asymptotic flip an-
`gles, the efficiency begins to drop. The efficiency at lower
`flip angles can be increased to nearly 100% by optimiz-
`ing RF pulse trains for a downward ramp in echo ampli-
`tude from the 180" echo amplitude to the desired asymp-
`totic echo amplitude. Experimental implementation of
`this approach (Fig. 5) confirmed the exceptionally high
`efficiency of these RF pulse trains. While this approach
`does improve sensitivity, it requires magnitude correc-
`tion of the first few echoes prior to image reconstruction
`to avoid image artifact.
`The use of low flip angles in the RF pulse train mixes
`both spin and stimulated echo components in the echo
`amplitudes. The decay rate of the echo amplitudes is
`thus determined by an effective T,, Tzeff, which is a
`mixture of T, and T2. For many tissues, Tl is much
`greater than T2. In these tissues, TZeff is longer than T, but
`is still relatively insensitive to the value of Tl. For very
`low flip angle RF pulse trains or in tissues such as fat
`where TI is not much longer than T,, T, also has an
`
`General Electric Co. 1004 - Page 5
`
`

`
`The Sensitivity of Low Flip Angle RARE Imaging
`
`181
`
`FIG. 6. Axial, single-shot RARE
`images through the brain of a nor-
`mal volunteer. Images were ob-
`tained with ramp optimized RF
`pulse trains with asymptotic flip an-
`gles of 90" (left), 60" (center), and
`30" (right). The images in the upper
`row were obtained with an effective
`TE of 18 ms, the lower row with an
`effective TE of 83 ms.
`
`impact on contrast. In many pathologies, T, and TI
`change in the same direction. In such pathologies, the
`mixture of TI and I", decay should not strongly influence
`contrast, as long as a longer TEeff is used to compensate
`for the longer TZeff. In other pathologies, it is possible that
`Tzeff contrast will be poorer than T, contrast.
`Although the use of low refocusing flip angles does
`reduce the amplitude of the measured signal, the signal-
`to-noise ratio need not necessarily reduce proportion-
`ately. Accounting for the longer time available for imag-
`ing due to the longer TZeff shows that the signal-to-noise
`ratio is nearly unchanged until asymptotic angles of un-
`der 60" are reached (Fig. 4). Practical applications of low
`flip angle RARE will have an even slower decrease in
`signal-to-noise ratio than in the figure because the anal-
`ysis does not include the variation of data acquisition
`efficiency with bandwidth. For example, in the present
`implementation, a 64 kHz bandwidth was used that ac-
`quired a line of k space in 2 ms. The remaining 3 ms of
`the echo spacing were required for applying gradients
`and the pulse. At a lower flip angle where TZoff is twice
`Tz, a 10-ms echo spacing could instead be used. Since the
`RF and gradient timing is unchanged, the entire addi-
`tional 5 ms could be used for data acquisition by decreas-
`ing the bandwidth to 18 kHz. The practical reduction of
`noise is a factor of 1.9, rather than the 1.4 assumed in the
`theoretical considerations above, because the data acqui-
`sition efficiency increases from 40-70%.
`Single-shot images in the brain of a normal volunteer
`demonstrate excellent image quality and the expected
`changes in signal amplitude and contrast with refocusing
`flip angle. Long echo-train imaging with reduced refocus-
`ing flip angles reduces the power deposition of the se-
`
`quence dramatically. By employing 30" flip angles, for
`example, the signal amplitude is decreased by a factor of
`2, the signal-to-noise ratio by 30%, yet the power depo-
`sition is reduced by a factor of 36 relative to a 180" train.
`In practical application, and in the in vivo images
`shown in Figs. 6 and 7, slice selective refocusing pulses
`must be used and the imperfect slice profile will have
`implications for sensitivity and contrast. The methods of
`LeRoux and Hinks (10) for stabilizing the echo ampli-
`tudes when slice selective pulses are used can easily be
`applied to lower asymptotic flip angle RF pulse trains. As
`can be seen in Figs. 6 and 7, naive use of the nonselective
`pulse theory to tailor the amplitudes of selective pulses
`can produce images of excellent quality even though
`poorly selective pulses were used. Accurately accounting
`for the effects of slice selection will likely slightly alter
`the optimized RF pulse amplitudes and slightly increase
`the TZnff relative to the nonselective pulse theory. Since
`the RF power in low flip angle RARE is lower than when
`180" pulses are used, RF pulses with better slice profiles
`but that deposit more RF power might be employed. This
`would lead to more uniformity of contrast across the
`slice.
`The variation of RF amplitude, which typically occurs
`within a practical RF coil, has been neglected in this
`work. Experience with the standard head coil, in which
`the amplitude variations are of the order of lo%, and
`with the body coil, in which amplitude variation can be
`considerably higher than the head coil, suggests that RF
`inhomogeneity is not a major problem for this method.
`Errors in the amplitude of the RF pulse train will cause
`slight amplitude fluctuations in the first few echoes.
`These fluctuations could produce subtle artifacts, espe-
`
`General Electric Co. 1004 - Page 6
`
`

`
`182
`
`Alsop
`
`FIG. 7. Comparison of apparent T,
`decay in images obtained using
`ramp optimized RF pulse trains
`with asymptotic flip angles of 90"
`(top), and 17" (bottom). The differ-
`ence of T, decay in the images ob-
`tained at TFs of 18 ms (left), 83 rns
`(center), and 157 m s (right) is
`clearly apparent.
`
`cially in proton density images where the center of k-
`space is scanned in the first echoes. Discarding the first
`few echoes may be sufficient to eliminate these artifacts.
`Contrast will also be slightly affected by RF inhomoge-
`neity because TZeff will be longer when the RF amplitude
`is weaker.
`The major disadvantage of long echo-train imaging is
`the increased acquisition time required for each slice.
`This reduces the number of slices that can be acquired in
`a given repetition time and can lead to inefficiency in
`data acquisition. In this respect, long echo-train RARE
`imaging is similar to fast gradient-echo imaging. Long
`echo-train RARE imaging will most likely be optimal in
`applications where fast or motion insensitive imaging is
`required or in 3-dimensional sequences. The extension of
`the number of echoes made possible by low flip angle RF
`pulse trains may make it possible to acquire %dimen-
`sional T, images without the use of multiple slab acqui-
`sitions (16) that can suffer from slab boundary artifacts.
`Reduced flip angle echo trains need not be limited to
`long echo train acquisitions. In multiple slice interleaved
`RARE acquisitions at high field, power deposition can
`restrict the number of slices that can be acquired. In such
`circumstances, reduced flip angles can be used, although
`the signal-to-noise ratio will decrease more like the echo
`amplitude plotted in Fig. 1 rather than the signal-to-noise
`ratio plotted in Fig. 4 because the extended acquisition
`time available in each slice was not used for data acqui-
`sition.
`The single-shot images in Figs. 6 and 7 invite compar-
`ison with other single shot techniques including echo
`planar (17), GRASE (18), and prepared gradient echo
`sequences (19). Single-shot RARE does not suffer from
`
`the distortion and ghosting artifacts that plague echo-
`planar imaging so it may be competitive with echo-pla-
`nar imaging for motion insensitive T, and proton density
`imaging. However, echo-planar imaging can be operated
`in low flip angle gradient echo mode, allowing for higher
`temporal resolution than is feasible with RARE, and echo
`planar acquisitions are generally a factor of 2 to 4 times
`faster than RARE acquisitions. Because the CPMG con-
`dition is required, the RARE acquisition is also harder to
`prepare with contrasts such as T,* and diffusion, al-
`though approaches to overcoming these problems have
`been discussed (7). Low flip angle RARE is directly com-
`patible with the acquisition of multiple gradient echoes
`per echo train as are used in GRASE. Optimized RF pulse
`trains have previously been suggested as an approach to
`reduce the T2 decay artifacts in GRASE (20). As dis-
`cussed earlier, the use of low refocusing flip angles in
`RARE does make possible data acquisition efficiencies of
`70%, which are comparable with those of GRASE (21).
`with the better point spread function of RARE (22).
`The existence of a pseudosteady state and the use of
`low flip angle RF pulse trains in a RARE sequence seem
`to blur the distinction between fast gradient-echo and
`RARE imaging. In truth, there is little difference between
`RARE and prepared gradient echo sequences where the
`signal due to T , recovery during the acquisition is
`crushed by applied gradients (23) except that the theory
`for RARE explicitly takes into account the multiple spin
`and stimulated echoes that can occur in such sequences.
`There is a major difference between RARE and equilib-
`rium fast gradient echo sequences, however, because T,
`recovery of magnetization plays a major role in such
`gradient echo sequences. For long echo train, low flip
`
`General Electric Co. 1004 - Page 7
`
`

`
`i i
`
`If it is assumed that all of the equilibrium magnetization,
`M,, has been transferred to the pseudosteady state then
`
`183
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket