`
`
`
`
`
`
`PHOTODETECTOR‘SIZE CONSIDERATIONS IN THE DESIGN OF A
`
`
`
`
`
`NONINVASIVE REFLECTANCE PULSE OXIMETER FOR TELEMEDICINE
`
`APPLICATIONS
`
`
`
`
`
`
`C. Pujary, M. Savage, Y. Mendelson
`
`
`
`
`
`
`Department of Biomedical Engineering, and Bioengineering Institute
`
`
`
`
`
`
`Worcester Polytechnic Institute, Worcester, MA 01609.
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`compromising
`Abstraci—Low power management without
`provided an easy way to simulate various PD areas by
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`signal quality is a key requirement in the optimal design of a
`connecting in parallel multiple PBS to the common
`
`
`
`
`
`
`
`
`
`
`
`
`
`wearable pulse oximeter.
`This paper
`investigates
`the
`summing
`input
`current-to-voltage
`of
`converter.
`a
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`advantage gained by using a photodetector with a larger active
`
`
`
`
`
`
`
`
`Additional circuitry, consisting of amplifiers and band pass
`
`
`
`
`
`
`
`
`
`area. Preliminary in viva experiments have demonstrated that
`
`
`
`
`filters, were used to produce two different signals (a
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`by increasing the area of the photodetector it is possible to
`
`
`
`
`
`
`
`
`
`pulsatile AC component and a non-pulsatile DC component)
`reduce the overall power requirement of a wireless sensor.
`
`
`
`
`
`
`
`
`
`
`
`
`
`from each PPG. Analog data streams were digitized at
`intended for telemedicine applications.
`
`
`
`
`
`
`
`
`
`501—12 for 30s periods by a National Instrument DAQ card
`
`
`
`
`
`
`
`
`
`
`
`Keywords—pulse oxinieter, telemedicine, wearable sensor
`installed in a PC matting LABYIEW 6.0 software.
`
`
`
`
`1. INTRODUCTION
`In- Vitro Experiments
`
`
`
`
`
`
`
`Dark Tests: To test the background noise level generated
`
`
`
`
`
`
`
`
`
`
`by each PD, we performed a series of dark measurements by
`
`
`
`
`
`
`
`
`
`
`switching off the LEDs in the sensor and blocking ambient
`
`
`
`
`
`
`light from reaching the six PDs.
`
`
`
`
`
`
`
`Multiple Photodetectors Tests: Each PD was randomly
`
`
`
`
`
`
`
`connected through the hub: ‘to determine
`the spatial
`
`
`
`
`
`
`
`
`
`
`uniformity of the illuminating field incident on the PDs. To
`
`
`
`
`
`
`
`produce a constant level background illumination, a signal
`
`
`
`
`
`
`
`
`
`
`
`composed of a DC bias voltage modulated by a small lI-Iz
`
`
`
`
`
`
`
`
`
`AC sine wave was generated by a programmable function
`
`
`
`
`
`
`
`generator. The signal from the function generator was
`
`
`
`
`
`
`
`
`
`
`
`applied to a separate LED that was used to simulate a
`
`
`
`
`
`
`
`
`
`
`typical PPG signal. The external LED was attached to a
`
`
`
`
`
`
`
`translucent flat medium serving as an optical difi'user. The
`
`
`
`
`
`
`
`
`
`
`surface of the diffuser was positioned at a distance of 30cm
`
`
`
`
`
`
`
`
`away from the planar surface of the sensor.
`
`
`
`In- Viva Experiments
`
`
`
`
`
`
`
`
`
`A series of in vivo experiments were performed to
`
`
`
`
`
`
`
`
`determine the signal improvement gained by using different
`
`
`
`
`
`
`
`
`
`
`PD areas. The prototype sensor was attached to the base of
`
`
`
`
`
`
`
`
`
`a volunteer’5 finger and the peak currents supplied to the red
`
`
`
`
`
`
`
`
`
`and infrared LEDs were adjusted to 3m and 1.9mA,
`
`
`
`
`
`
`
`respectively As the driving currents were adjusted, ,the
`
`
`
`
`
`
`
`
`
`
`output of each amplifier was monitored to assure that (i) a
`
`
`
`
`
`
`
`
`distinguiShable and stable PPG was observed when a single
`
`
`
`
`
`
`
`
`
`PD was employed, and (ii) maximal PPG signals were
`
`
`
`
`
`
`
`
`produced without causing amplifier saturation when all 6
`
`
`
`
`
`
`
`
`
`
`PDs we're connected in parallel through the hub.
`It
`is
`
`
`
`
`
`
`
`
`
`important
`to note that
`the final currents selected were
`
`
`
`
`
`
`
`
`significantly lower compared to the typical driving currents
`
`
`
`
`
`employed in commercial pulse oximeters.
`
`
`
`
`
`
`
`
`
`
`Commercially available sensors used in reflectance
`
`
`
`
`
`
`
`
`mode pulse oximeters employ a single photodetector (PD)
`
`
`
`
`
`
`
`
`
`element typically with an active area of about 12-15mm2.
`
`
`
`
`
`
`
`
`The light
`intensity detected by the photodetector in a
`
`
`
`
`
`
`
`reflectance pulse oximeter depends on the incident light
`
`
`
`
`
`
`
`intensity, absorption by skin, reflection by bones, tissue
`
`
`
`
`
`
`
`
`
`backseattering and the amount of blood in the vascular bed
`
`
`
`
`
`
`Compared to transmittance mode pulse oximetry,
`-[1].
`
`
`
`
`reflected photoplethysmograms
`(PPGs) have generally
`
`
`lower amplitudes.
`.
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`Low power management without compromising signal
`
`
`
`
`
`
`
`
`
`
`
`quality is a key requirement in optimizing the design of a
`
`
`
`
`
`
`
`
`wearable telesensor. One approach to lowering the power
`
`
`
`
`
`
`
`consumption of a wireless pulse oximcter, which is
`
`
`
`
`
`
`
`
`
`
`
`dominated by the current required to drive the LEDs, is to
`
`
`
`
`
`
`
`
`
`reduce the LED duty cycle [2]. Alternatively, lowering the
`
`
`
`
`
`
`
`
`
`
`current supplied to the red and infiared LEDs can also
`
`
`
`
`
`
`
`reduce power consumption. However, with reduced current
`
`
`
`
`
`
`
`
`
`
`
`drive, there is a direct impact on the quality of the detected
`
`
`
`
`
`
`
`
`
`
`'PPGss Mendelson et a1. [3] showed that a concentric array
`
`
`
`
`
`
`
`
`
`
`of PDs could be used to increase the
`amount of
`
`
`
`
`
`
`
`backseattered light detected-by a reflectance type pulse
`
`
`
`
`
`
`
`
`
`oximeter sensor. In this paper We investigate the advantages
`
`
`
`
`
`
`
`
`
`of increasing the PD area and minimizing LED driving
`
`
`
`
`
`
`
`
`
`currents to optimize the overall power requirement of a
`
`
`
`
`
`
`
`reflectance mode pulse oximeter
`intended for 'future
`
`
`
`applications in telemedicine.
`
`
`II. METHODOLOGY
`
`
`Experimental setup
`
`
`
`
`
`
`
`
`
`To study the effect of different PD areas, we constructed
`
`
`
`
`
`
`
`and tested a prototype reflectance sensor employing 6
`
`
`
`
`
`
`
`
`identical (3mm x 4mm) PDs. The" equally spaced PDs were
`
`
`
`
`
`
`
`' concentrically arranged in a 18mm diameter planar
`
`
`
`
`
`
`
`
`configuration around a pair of red and infrared LEDs. Each
`
`
`
`
`
`
`
`
`PD was individually connected to‘a central hub. The hub
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`III. RESULTS
`
`
`
`
`
`
`
`
`The Rent Mean Square (RMS) values corresponding to
`
`
`
`
`
`
`
`
`the amplitude of the AC components measured frOm each
`
`
`
`
`
`
`
`
`
`PD are plotted1n Fig.1. During darkness the average noise
`
`
`
`
`
`
`
`
`0-7803-7767-21031'31 7.00 ©2003 IEEE
`
`
`
`148
`
`
`
`Page 1
`
`VALENCELL EXHIBIT 2012
`
`IPR2017—00318
`
`Page 1
`
`VALENCELL EXHIBIT 2012
`IPR2017-00318
`
`
`
`
`
`9
`DCortpletedarhtess
`} IConatant light ilbani'tltion i
`
`i
`:
`
`5“:
`
`UCahullIed :5
`IMeesured j;
`
`f
`
`
`
`
`
`PHOTODETECTOROUTPUT(V)
`
`6-PDs
`
`
`i i E !
`
`I l
`
`
`
`PHOTODETECTOROUTPUT(V)
`
`FDl
`
`
`
`
`
`
`5-PDs
`4-PDs
`2-PD!
`PIX
`PDQ
`3-PDs
`1-PD
`P03
`PBS
`F135
`
`
`
`
`
`
`
`
`
`
`
`
`INDIVIDUAL PHOTODETECTORS
`MULTIPLE PHOTODETECTORS
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`Fig. 1. Individual photodetectar perfonnance under complete darkness and
`
`
`
`
`a constant light illumination.
`
`
`
`
`
`
`
`
`In
`individual PBS was 0.114V.
`generated by the
`
`
`
`
`
`
`
`
`comparison, the average PPG amplitude measured in vitro
`
`
`
`
`
`
`
`
`
`
`by the 6 PBS under a spatially uniform illumination field
`
`
`
`
`
`
`
`
`produced by the external LED source was 0.647V.
`
`
`
`
`
`
`
`
`
`
`Fig. 2 shows the signals detected in vitro with multiple
`
`
`
`
`
`
`
`combinations of PBS using the
`simulated uniform
`
`
`
`
`
`
`
`
`illumination produced by the external LED. The right-side
`
`
`
`
`
`
`
`
`bars represent the measured RMS values corresponding to
`
`
`
`
`
`
`
`
`
`different PD areas. For comparison, the left-side bars were
`
`
`
`
`
`
`
`computed
`using
`relationship
`0.647n, where
`the
`n
`
`
`
`
`
`
`
`
`
`
`corresponds to the number of PBS connected in parallel. The
`
`
`
`
`
`
`
`
`trend observed provide sufiicient evidence of a linear
`
`
`
`
`
`
`
`
`
`improvement in signal intensity as a filnction of an increase
`
`
`
`
`
`
`
`in the active area of the PDs.
`
`
`
`
`
`
`
`
`
`Fig. 3 shows the magnitude of the pulsatile red and
`
`
`
`
`
`
`
`infrared PPGs measured
`in-vz'vo
`for
`different PD
`
`combinations.
`
`
`IV. DISCUSSION
`
`
`
`
`
`
`
`
`
`
`Minimizing the current required to drive the LEDs is a
`
`
`
`
`
`
`
`critical design consideration in optimizing the power
`
`
`
`
`
`
`
`consumption of a wearable pulse oximeter. However,
`
`
`
`
`
`
`
`
`
`reduced LED driving currents has a direct impact on the
`
`
`
`
`
`
`
`
`
`incident light intensity produced by the sensor and could
`
`
`
`
`
`
`
`
`
`lead to deterioration in the quality of
`the PPGs.
`
`
`
`
`
`
`
`
`Consequently,
`it could result
`in unreliable and therefore
`
`
`
`
`
`inaccurate calculation of oxygen saturation.
`
`
`
`
`
`
`
`
`
`
`
`From the data presented in Fig. 3, it can be observed that
`
`
`
`
`
`
`
`
`
`
`the overall increase in the reflected PPG signals achieved in
`
`
`
`
`
`
`
`
`
`
`
`viva when all 6 PBS were connected in parallel is smaller
`
`
`
`
`
`
`
`
`
`
`and does not follow the same linear relationship observed in
`
`
`
`
`
`
`
`
`
`
`
`vitra as shown in Fig. 2. This deviation was most likely
`
`
`
`
`
`
`
`caused by the non-uniform backseattered light distribution
`
`
`
`
`emanating from the finger.
`
`
`V. CONCLUSION
`
`
`
`
`
`
`
`
`The data presented in this study demonstrate that the
`
`
`
`
`
`
`
`
`driving currents of the LEDs in a reflectance pulse oximeter
`
`
`
`
`
`
`
`canbe lowered significantly without compromising the
`
`
`
`
`
`
`
`
`
`
`
`
`Fig. 2. Signal improvement observed in-virro with multiple photodetectars.
`
`'i
`iElRedLED
`
`i
`llnfinredlED'i
`
`4-5]
`41
`”J
`,ll
`2.51
`.l
`
`I
`lJ41.
`lI
`
`
`
`PHOTODETECTOROUTPUT(V)
`
`
`
`
`
`3-l’Ds
`é-FDa
`l—PD
`2-?De
`4-FDs
`S‘PDs
`
`
`
`
`
`
`MULTIPLE PHOTODETECTORS
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`Fig. 3. Signal improvement observed in vivo with multiple photodetectors.
`
`
`
`
`
`
`
`
`
`
`
`quality of the PPGs simply by increasing the overall size of
`
`
`
`
`
`
`
`
`
`
`the PD in the sensor. Hence, with reduced LED driving
`
`
`
`
`
`
`
`currents, maximizing the backseattered light collected by
`
`
`
`
`
`
`
`
`the sensor and optimized digital switching techniques, a
`
`
`
`
`
`
`
`
`
`very low power consuming sensor can be developed thereby
`
`
`
`
`
`
`
`
`
`extending the overall battery life of a pulse oximeter
`
`
`
`
`intended for telemedicine applications.
`
`ACKNOWLEDGEMENT
`
`
`
`
`
`
`
`
`
`This research was supported in part by Department of
`
`
`
`
`Defense Cooperative Agreement DAMDl7-03-2-0006.
`
`_
`
`
`
`
`
`
`REFERENCES
`
`
`
`
`
`
`[1] Y. Mendelson, B. Ochs, “Noninvasive Pulse Oximetry
`
`
`
`
`Utilizing Skin Reflectance Photoplethysmography,” IEEE
`
`
`
`
`
`
`
`Transactions on Biomedical Engineering, vol. 35, no. 10,
`
`
`
`
`pp. 798-805, Oct. 1988.
`
`
`
`
`
`
`
`
`
`
`
`
`[2] S. Rhee, B.H. Yang and H. Asada,” The Ring Sensor: A
`
`
`
`
`
`
`
`
`new Ambulatory Wearable Sensor for twenty Four Hour
`
`
`
`
`
`
`
`Patient Monitoring,” Proc. of the 20th Annual International
`
`
`
`
`
`
`
`
`
`Conference ofthe IEE Engineering in Medicine and Biology
`
`
`
`
`
`Society. Hong Kong. Oct. 1998.
`
`
`
`
`
`
`
`
`
`
`[3] Y. Mendelson, 10. Kent, BL. Yocum and MJ. Birle,”
`
`
`
`
`
`
`
`
`Design and Evaluation of new reflectance pulse oximeter
`
`
`
`
`
`
`
`
`sensor,” Medical Instrumentation, Vol.22, NO. 4, pp. 167-
`
`
`
`173, Aug. 1988.
`
`149
`
`
`
`Page 2
`
`Page 2
`
`