throbber
United States Patent [191
`Kenna et al.
`
`[541
`[151
`
`[73]
`[2 1]
`[22]
`[51]
`[52]
`
`[58]
`
`[56]
`
`PROSTHETIC KNEE IMPLANTATION
`Inventors: Robert V. Kenna, Saddle River, N.J.;
`David S. Hungerford, Cockeysville;
`Kenneth A. Krackow, Baltimore, both
`of Md.
`Assignee: Howmedica, Inc., Wilmington, Del.
`App]. No.: 811,020
`Filed:
`Dec. 19, 1985
`
`Int. Cl.4 .............................................. .. A61F 5/04
`US. Cl. ...................... .. 128/92 VW; 128/92 VD;
`128/92 VL
`Field of Search ........... .. 128/1 R, 303, 305, 92 E,
`128/92 EB, 92 C, 92 H
`
`References Cited
`U.S. PATENT DOCUMENTS
`
`I
`
`[11]
`[45]
`
`Patent Number:
`Date of Patent:
`
`4,646,729
`Mar. 3, 1987
`
`OTHER PUBLICATIONS
`David S. Hungerford, M. D., Kenneth A. Krackow, M.
`D., Robert V. Kenna, The Howmedica, “Universal
`Total Knee Instrument System”, 1981.
`Nas. S. Eftekhar, M. D., “Total Knee-Replacement
`Arthroplasty”, Journal of Bone and Joint Surgery, vol.
`65-A, No. 3, Mar. 1983, pp. 283-309.
`Primary Examiner-Stephen C. Pellegrino
`Attorney, Agent, or Firm—-Harold Pezzner
`
`ABSTRACT‘
`[57]
`A prosthetic knee is implanted after cutting the femor
`and tibia in the proper manner with the aid of instru
`ments which include axial alignment guides and a series
`of cutting jigs.
`
`4,211,228 9/1980 Cloutier ........................ .. 128/303 R
`
`8 Claims, 101 Drawing Figures
`
`Smith & Nephew Ex. 1032
`IPR Petition - USP 8,377,129
`
`

`
`U.S.Patent Mar. 3, 1987
`U.S.Patent Mar. 3, 1987
`
`%
`-
`
`Sheet1of23
`Sheetlof23
`
`4,646,729
`4,646,729
`
`‘ 127.1
`. V
`
`1;
`
`- Mclza'nl'cal
`
`

`
`U.S.IPa-tent Mar.3, 1987
`
`Sheet20f23
`
`4,646,729
`
`W245.
`
`21?. W20.
`
`ANTERIOR FEMUR
`
`TRANSVERSEDISTAL FEMUR msmlmmnrmmrmn POSTERIORIFEMUR
`
`J/gZd. 179.26. 175/. g.
`ANTERIOR CWFER
`m
`A;
`
`mmsvmiss Mimi-1m POSTERIOR 111mm}:
`
`TIBIAL FZIXATION
`
`

`
`US. Patent Mar.3, 1987
`
`Sheet30f23
`
`4,646,729
`
`>..
`
`Emma
`
`

`
`U.S.Patent Mar. 3, 1987
`U.S.Patent Mar. 3, 1987
`
`Sheet4of23
`Sheet4of23
`
`4,646,729
`4,646,729
`
`

`
`U.SQ Patent Mar. 3, 1987
`
`Sheet50f23' 4,646,729
`
`4-7-71 759M
`
`

`
`US. Patent Mar. 3, 1987
`
`Sheet6of23 ' 4,646,729 '
`
`

`
`US. Patent Mar. 3, 1987
`
`Sheet7of23 - 4,646,729
`
`

`
`US. Patent Mar. 3, 1987
`
`Sheet80f23 - 4,646,729 ‘
`
`

`
`US. Patent Mar. 3,1987
`
`Sheet90f23 4,646,729 '
`
`Wé’OA.
`
`

`
`U.S. Patent Mar. 3, 1987
`US. Patent Mar. 3, 1987
`
`Sheet 10 01923 4,646,729
`Sheet 10 of>23 4,646,729
`
`

`
`U.S. Patent Mar. 3, 1987
`US. Patent Mar‘.3, 1987
`
`‘Sheet 11 of 23 4,646,729
`Sheet 11 of 23 4,646,729
`
`~58
`
`32.
`
`

`
`U.S. Patent Mar. 3, 1987
`US. Patent Mar. 3, 1987
`
`Sheet 12 of 23 4,646,729
`Sheet 12 of 23 4,646,729
`
`

`
`US. Patent Mar. 3, 1987
`
`Sheet 13 of 23 4,646,729
`
`mlmw
`
`16 "’./ v, v,
`
`75
`
`6g
`
`, , 16
`
`2%
`
`76
`
`

`
`U.S.Patent Mar. 3, 1987
`
`Sheet 14 of23‘ 4,646,729 '
`
`w
`
`?lm“
`
`..
`
`.
`
`

`
`U.S.Patent Mar. 3, 1987
`
`Sheet 15 of 23 4,6465729
`
`49.
`
`$ 4
`
`

`
`U.S.Patent Mar. 3, 1987 I
`
`Sheet 16 of 23 '4,646§72v9
`
`W": ~
`100
`104.
`
`1'
`
`I
`
`

`
`U.S.Patent' Mar. 3, 1987
`
`Sheet 17 of23 4,646,:729
`
`w _
`52.4
`M
`
`W60.
`
`1
`
`Hlhw"
`32
`
`

`
`U.S. Patent Mar.1987
`US. Patent Mar. 1987
`
`Sheet 18 of 23 4,646,729
`Sheet 18 of 23 4,646,729
`
`

`
`U.S.'Patent’ Mar. 3, 1987
`
`Sheet 19 of23 4,646,729
`
`B6
`
`

`
`U.S.Patent
`
`‘Mar.3,_1987
`
`% Sheet 20 of23 4,646,729
`
`

`
`U.S. Patent Mar. 3, 1987
`
`Sheet 21 of 23 4,646,729
`
`

`
`U.S.L Patent Mar. 3, 1937
`
`Sheet 22 of23 4,646,729
`
`

`
`U.S. Patent Mar. 3, .1987
`
`
`
`Sheet 23 of23‘ 4,646,729 %
`
`
`
`
`
`
`
`‘fsv
`
`
`
`17" .93.
`
`
`
`\““~\
` x
`
`
`
`

`
`H .156
`
`‘in
`
`
`
`
`
`
`
`. y....'l///£57/III]
`
` 4
`
`4.148
`
`

`
`1
`
`4,646,729
`
`2
`
`PROSTHETIC KNEE IMPLANTATION
`
`BACKGROUND OF THE INVENTION
`
`Prosthetic knee components have long been known
`and used in the art. In order to prepare the femor and
`tibia for the components, it is necessary to make a series
`of cuts from these bones to conform to the size and
`shape of the prosthetic components. Generally these
`cuts are made by visual reliance on where the cuts
`should be sometimes with the aid of simplified jigs and-
`/or score lines. Such techniques are necessarily impre-
`cise which limits the ability to provide the anatomically
`most desirable prosthetic knee.
`filed February
`Applicant’s copending application,
`1982, the details of which are incorporated herein by
`reference thereto, describes such an anatomically desir-
`able prosthetic knee system.
`The aforenoted prosthesis is designed to reproduce
`anatomic movement of the knee without compromising
`stability. This prosthetic knee, as the natural knee, pro-
`vides a “screw home” mechanism which increases sta-
`bility in extension. As flexion proceeds, the femoral
`condyles initially roll posteriorly. Through asymmetric
`condylar and tibial compartments,
`the natural and
`changing axes of rotation are preserved, thereby pre-
`venting the development of abnormal
`tension in re-
`tained ligaments. When abnormal ligamentous tension
`develops, it either restricts flexion and increases shear
`stress at fixation interfaces and/or leads to eventual
`progressive ligament attenuation and joint instability.
`The design also allows natural internal and external
`rotation of the knee in flexion. Furthermore, contouring
`of the posterior margins of the tibial plateaus facilitates
`stability in flexion and provides a broad Contact surface.
`The combination of these factors, which balance soft
`tissue elements in the joint, minimizes shear stress at the
`fixation interfaces, enhancing the potential for long
`term function of the replaced knee.
`In order to make such prosthetic knee system feasible,
`the necessary bone cuts must be precisely accom-
`plished. This, in turn, requires a set of proper instru-
`ments including guides and jigs. Such instruments
`should assure reproducibly accurate bone cuts, pros-
`thetic seating, and lower limb alignment.
`SUMMARY OF THE INVENTION
`
`An object of this invention is to provide techniques
`which permit the above-noted knee prosthesis to be
`implanted.
`A further object of this invention is to provide instru-
`ments including guides and cutting jigs which assure the
`necessary accurate bone cuts.
`In accordance with this invention, a set of instru-
`ments is provided which allows for variations in the
`anatomical axis of the femor. The instruments include
`cutting jigs which are selectively locked directly to the
`bone. By use of these instruments, eight basis cone cuts
`-are made to align and seat the femoral and tibial compo-
`nents of the prosthesis.
`
`THE DRAWINGS
`
`FIG. 1 illustrates the anatomical considerations of the
`legs which are taken into account in accordance with
`the invention;
`FIGS. 2a through 2f illustrate the eight basis bone
`cuts which are made in accordance with the invention;
`
`FIGS. 3-30 illustrate the various steps and instru-
`ments used therein from incision to final tibial prepara-
`tion in the practice of the invention;
`FIGS. 31-35 are side, top, front, bottom and rear
`views, respectively, of the distal femoral cutting jig
`shown in FIGS. 7-13;
`FIGS. 36-40 are side, front, rear, bottom and plan
`views, respectively, of the pin holder alignment guide
`shown in FIGS. 8-9;
`FIGS. 41-44 are cross-sectional views taken through
`FIG. 37 along the lines 41—41, 42-42, 43-43 and
`44-44;
`FIG. 45 is a front elevation view partly in section of
`the distal femoral cutting jig shown in FIGS. 10-13;
`FIGS. 46-49 are top, bottom, side and rear views,
`respectively, of the femoral cutting jig shown in FIG.
`45;
`
`FIG. 50 is a cross-sectional view taken through FIG.
`48 along the line 50--50;
`FIGS. 51-55 are bottom, front, rear, side and top
`views, respectively, of the femoral drill jig shown in
`FIGS. 13-14;
`FIG. 56 is a cross-sectional View taken through FIG.
`52 along the line 56-56;
`FIGS. 57-61 are top, front, rear, bottom and side
`views, respectively, of the transverse femoral cutting jig
`shown in FIGS. 15-16;
`FIG. 62 is a cross-sectional view taken through FIG.
`57 along the line 62-62;
`FIGS. 63-67 are side, top, front, rear and bottom
`views, respectively, of the femoral spacer/tensor jig
`shown in FIGS. 17-22;
`FIG. 68 is a cross-sectional view taken through FIG.
`66 along the line 68-68;
`FIGS. 69-72 are top, rear, front and bottom elevation
`views, respectively, of the transverse tibial cutting jig
`shown in FIGS. 17-23;
`FIGS. 73-74 are end elevation views of the trans-
`verse tibial cutting jig shown in FIGS. 29-72;
`FIG. 75 is a cross-sectional view taken through FIG.
`69 along the line 75-75;
`FIGS. 76-80 are bottom, front, rear, top and end
`views, respectively, of the femoral chamfer cutting jig
`shown in FIG. 24;
`FIG. 81 is a cross-sectional view taken through FIG.
`76 along the line 81-81;
`FIGS. 82-86 are bottom, front, rear, top and end
`views, respectively, of a modified form of femoral
`chamfer cutting jig;
`FIG. 87 is a cross-sectional view taken through FIG.
`82 along the line 87-87;
`FIGS. 88-92 are side, top, bottom, front and rear
`views, respectively, of the tibial positioning/fixation jig
`shown in FIGS. 25-30;
`FIG. 93 is a cross-sectional view taken through FIG.
`89 along the line 93-93; and
`FIG. 94 is a cross-sectional view taken through FIG.
`88 along the line 94-94.
`
`DETAILED DESCRIPTION
`
`The instruments utilized in the practice of the inven-
`tion consist of long axial alignment guides and cutting
`jigs. The jigs may be sequentially numbered relative to
`their order of use to simplify the procedure. The align-
`ment guides are designed to assist the surgeon in posi-
`tioning all primary cutting jigs prior to making the bone
`cuts. All cutting jigs lock onto their respective bones to
`insure the accuracy of the cuts.
`
`l0
`
`15
`
`20
`
`25
`
`30
`
`35
`
`40
`
`45
`
`50
`
`55
`
`60
`
`65
`
`

`
`4,646,729
`
`3
`The inventive instrument system is designed to seat
`the femoral and tibial components parallel to the ana-
`tomic transverse axis of the knee. Since this axis is paral-
`lel to the ground and perpendicular to the vertical in
`two-legged stance, this positioning achieves: (l) uni-
`form stress distribution at fixation interfaces, (2) optimal
`alignment; and (3) physiological ligamentous balance of
`the knee. It is also important to recognize that the ankles
`remain closer to the midline vertical axis of the body
`that either the knees or hips throughout noral gait. To
`achieve the goals of total knee arthroplasty, these align-
`ment features must be preserved or reconstituted.
`The inventive instrument system uses the femoral
`shaft axis (S), the center of the knee (K), the center of
`the ankle joint (A), and the transverse axis of the knee
`(T) as its alignment references (FIG. 1). The mechanical
`axis of the lower libm, which runs from the center of the
`hip (H) thorugh the center of the knee (K) to the center
`of the ankle (A), generally forms an angle of 3° with the
`vertical (V), because the hips are wider apart than the
`ankles in both normal stance and gait (FIG. 1).
`Since the femoral head and neck overhang the shaft,
`the axis of the femoral shaft does not coincide with that
`of the leg, but forms with the leg a more acute valgus
`angle of 6° (SKH, FIG. 1). In total, the femoral shaft
`axis averages 9° of valgus with the vertical. The valgus
`angle of the femur varies relative to body build. The
`specific femoral valgus for a given individual can be
`determined by measuring angle HKS (FIG. 1) on a long
`X-ray which includes both the hip and the knee, and
`adding 3° (the mechanical axis). This method is valid for
`reconstituting a mechanical axis of 3° regardless of the
`-~ degree of pre-operative axial deformity at the knee.
`For tibial alignment, the center of the knee and the
`center of the ankle are used as reference points. Instru-
`ments which rely on the proximal tibial shaft as their
`key alignment reference tend to be inaccurate due to the
`frequent occurrence of tibial bowing. Recognizing that
`. the center of the ankle is closer to the midline vertical
`,; axis than the center of the knee in two-legged stance
`and throughout gait, this system uniquely requires a
`small angle at the proximal transverse tibal cut (TKA,
`FIG. 1). This angle keeps the transverse axis of the
`prosthesis parallel to the ground while the mechanical
`axis of the entire lower extremity remains in valgus
`(HKA, FIG. 1).
`FIG. 2 illustrates the eight basic bone cuts required to
`align and seat the femoral and tibial components of the
`prosthesis. As later described in detail, the instrument
`system consists of seven sequentially numbered cutting
`jigs and a femoral/tibial alignment guide. These are
`designed to insure the accuracy and alignment of all
`femoral and tibial bone cuts.
`the knee is ap-
`In the practice of the invention,
`proached through a longitudinal skin incision, followed
`by a medical parapatellar capsular incision. The quadri-
`ceps tendon is incised longitudinally, allowing eversion
`and dislocation of the patella laterally (FIG. 3).
`Orientation
`
`With the knee flexed to 90°, a 5/16 inch drill hole is
`made by drill bit 10 in the distal femur. It is placed
`roughly in the center 12 of the intercondylar notch just
`anterior to the femoral attachment of the posterior cru'-
`ciate ligament (FIG. 4). The position of this hole 13
`(FIG. 6) is not critical to the orientation of any femoral
`bone cuts—it is simply a point of purchase for the distal
`femoral cutting jig IA. The laterally protruding handles
`
`4
`the posterior
`22 are used to rotate the jig so that
`rounded eminences parallel
`to the posterior femoral
`condyles and the anterior femur is seen as on a sunrise
`view (FIG. 5). The jig is then hammered into place. Jig
`IA includes a pair of locking pins on the side hidden
`from view in FIG. 5. If there has been significant preop-
`erative deformity, the short locking pin facing the most
`prominent condyle is hammered into place bringing the
`face of the jig IA flush to that condyle only.
`
`The Transverse Distal Femoral Cut: Varus-Valgus And
`Flexion-Extension Alignment
`
`The long axial alignment guide 16 is used to establish
`proper varus-valgus and flexion extension alignment of
`the distal femoral cuts. The guide pin 18 is positioned
`into the pin holder in the appropriate right or left 7°, 9°
`or 11° hole 20 (FIG. 37). This angle is chosen relative to
`the pre-operative X-ray measurement technique previ-
`ously described in the alignment rationale section. Most
`often 9° will be appropriate. The guide 16 is then placed
`into the anterior holes of the distal femoral cutting jig
`IA. Correct axial alignment is achieved when the long
`alignment guide pin 18 is parallel to the femoral shaft
`axis in both the anterior and lateral views (FIGS. 7 and
`8). An examining finger can be slipped proximally under
`the quadriceps to get a better idea of the direction of the
`femoral shaft during this alignment procedure. If the
`alignment pin 18 does not parallel the femoral shaft, a
`mallet is used to tap the medial handle 20 or lateral
`handle 22, advancing the jig IA away from the de-
`formed condyle until the axial guide pin 18 parallels the
`femoral shaft (FIG. 9). With the alignment completed,
`one can easily visualize how much bone is missing from
`the deformed condyle. The IA jig is not stable at this
`point but must be manually held during insertion and
`removal of the alignment pinholder and also while plac-
`ing the IB jig.
`A short alignment pin is available to facilitate align-
`ment in two special circumstances. The shorter pin
`avoids impingement with the tourniquet on an obese
`thigh or with the abdomen of a short patient.
`
`Locking The Distal Femoral Cutting Jig In Position
`
`The axial alignment guide 16 is removed and the
`tongue of the distal femoral cutting jig IA and jig IB is
`slid into the first part of the distal femoral jig IA (FIG.
`10). With the teeth of the IB jig resting lightly on the
`anterior femur, recheck the alignment for both varus-
`valgus and flexion-extension prior to hammering it in
`place. When proper alignment is assured, first gently
`tap the IB jig with a mallet so that the teeth engage the
`bone without sliding down an oblique surface, thereby
`changing alignment. Then remove the alignment guide
`and hammer the IB jig firmly in place. There are also
`drill holes in the cutting bar of the IB jig through which
`% inch drill pins can be passed into the condyles if addi-
`tional stability is necessary.
`Final position of the distal femoral cutting jig IA is
`rechecked with the alignment guide. Measure twice, cut
`once (FIG. 12).
`
`Cutting The Distal Femur
`
`Once the distal femoral cutting jig IA is locked on to
`the femur, the surgeon _is free to direct his full attention
`to the cutting of the distal condyles. Pistol grip, endcut-
`ting oscillating saws are most effective for these cuts. It
`is important that the surgeon pay strict attention to
`maintaining the saw blade flat against the proximal
`
`l0
`
`15
`
`20
`
`25
`
`30
`
`35
`
`45
`
`50
`
`55
`
`60
`
`65
`
`

`
`5
`
`4,646,729
`
`cutting surface of the jig in order to achieve a precise
`cut (FIG. 13). Care should be taken to avoid the central
`intramedullary fixation peg of the jig. Following the
`initial transverse cut, the saw blade is passed back and
`forth across the cutting bar to shave the condylar cuts
`level with the plane of the cutting jig. This step is im-
`portant since the saw blade tends to scive away from
`the desired plane, particularly in more sclerotic bone
`and toward the deeper portions of the cut. The jigs are
`now removed and the distal femoral cuts are completed.
`In completing these cuts,
`the anterior aspect of the
`distal femoral cut will serve as the “cutting block” for
`the remaining posterior aspect. It is, therefore, impor-
`tant that the saw blade be inserted to the full depth of
`the initial cut before the oscillation is started. Other-
`wise, one runs the risk of starting a new plane. Once the
`posterior part has been completed, the broadest blade
`should be passed over the surface to be sure that the cut
`is a single-flat plane. This can also be checked with a
`cutting block.
`
`Femoral Corponent Rotational, Medial-Lateral And
`Anterior-Posterior Alignment
`
`Rotational, medial-lateral and anterior-posterior ori-
`entation of the femoral prosthesis is determined by the
`femoral drill jig II. This jig has two posterior skids 24
`which are slid between the posterior femoral condyles
`and tibial plateaus. These skids automatically position
`the instrument in 0° of rotation relative to the coronal
`plane of the distal femur (A,A, FIG. 14). The jig should
`firstbe centered in the medial-lateral position on the flat
`cut distal femoral surface, ignoring the initial keying
`hole for the IA jig. The jig II is now hammered flush
`with the flat surface of the distal femoral condyles. The
`jig’s anterior projection contains two holes 26 marked
`respectively for right and left knees. When a 3% inch drill
`pin 28, placed in the appropriate hole, is aligned with
`the center of the patello-femoral groove, correct medi-
`al-lateral and rotational positioning is assured (B,B,
`FIG. 14). When correct positioning has been assured, a
`5/16 inch drill 30 is used to make the holes for the femo-
`ral prosthesis fixation studs (C,C, FIG. 14).
`
`Anterior And Posterior Femoral Cuts
`
`Following the removal of the drill jig II, the two 5/ 16
`inch locking studs 32 of the anterior-posterior femoral
`cutting jig III, are inserted into the distal femoral fixa-
`tion holes. The jig III is hammered flush with the flat
`cut surface of the distal femoral condyles. The anterior
`plane of the cutting jig should intersect the anterior
`cortex of the femur at the proximal margin of the patel-
`lar facets (FIG. 15). If this plane appears too deep or too
`anterior, the next most appropriate size jig should be
`chosen. If there is any question of which size jig is to be
`used, always start with the larger jig. The anterior and
`posterior femoral bone cuts are now completed (FIG.
`16). Once again, care must be taken to rest the saw blade
`flush against the flat surface of the jig. With the poste-
`rior condyles removed, complete access to the posterior
`compartment allows removal of the menisci and ante-
`rior cruciate ligament. The tibial attachment of the
`posterior cruciate is identified and carefully avoided
`during the next step. Also, all remaining marginal osteo-
`phytes on the tibia and femur must be removed so they
`do not shorten or constrict ligaments or block full ex-
`tension.
`
`6
`
`Tibial Alignment And Cuts
`
`5
`
`10
`
`15
`
`20
`
`25
`
`30
`
`35
`
`40
`
`45
`
`50
`
`55
`
`60
`
`65
`
`The femoral spacer/tensor jig IV and transverse tib-
`ial cutting jig V are now assembled and positioned. The
`mortise cut out 34 in the transverse tibial cutting jig V
`is slipped over the tongue 36 of the femoral spacer/ten-
`sor jig and slid as far proximally as it will go (FIG. 17).
`Following insertion of the jig’s studs into the femoral
`fixation holes, the leg is brought into full extension. A
`folded towel or sheet is placed behind the knee to pre-
`vent
`inadvertent hyperextension at
`this stage of the
`procedure. Next the axial alignment guide 16 with both
`its femoral and tibial alignment pins in place is posi-
`tioned into‘ the femoral spacer/tensor jig IV (FIG. 18).
`
`Axial And Rotational Alignment
`
`To achieve correct tibial axial alignment prior to
`making the transverse tibial cut, longitudinal traction
`and manipulation are applied from the foot. The goal is
`to bring the center of the ankle joint directly under the
`distal dip of the alignment pin. This alignment will
`produce a slight tibial angle of 25° (A, FIG. 19). Ana-
`tomically, the centers of the ankle joints are closer to-
`gether than the centers of the knee joints. Overall leg
`alignment will still be in valgus. Correct extension
`alignment is achieved when the tibial shaft parallels the
`alignment pin when viewed from the side. Rotational
`alignment is correct when the medial malleolus is ap-
`proximately 30° anterior to the lateral malleolus in the
`coronal plane (B, FIG. 19).
`If correct axial alignment cannot be achieved at this
`point with the jig IV in place, then one of the special
`techniques for dealing with the severe deformity will
`have to be employed. The jig V cannot be locked onto
`the tibia until correct alignment is achieved.
`
`Soft Tissue/Joint Tension/Alignment
`
`tissue stability is established following initial
`Soft
`manual tibial axial alignment. Each side of the femoral
`spacer/tensor jig IV has expandable arms 38 which
`spread when its appropriate thumb screw 40 is tight-
`ened (FIG. 20). The tensor arms are extended to stabi-
`lize the joint in the correct alignment. Do not over-
`tighten the femoral spacer/tensor jig IV. The transverse
`tibial cutting jig V, pushed as far proximally as it will go
`ensures that only the minimal amount of the tibial pla-
`teaus will be removed. When one plateau is considera-
`bly more depressed than the other, the transverse cut-
`ting jig should be slid distally so that the plane of the
`tibial cut will rerove enough bone from the depressed
`plateau to provide a sufficiently flat surface for seating
`the tibial prosthesis.
`'
`
`Check Of Overall Alignment
`
`Prior to locking the transverse tibial cutting jig V in
`place, overall lower limb alignment should be checked.
`For correct alignment, the femoral alignment pin 18
`will parallel the femoral shaft in both the anterior and
`lateral planes. The tibial alignment pin 19 will extend
`from the center of the knee to the center of the ankle
`and be parallel to the tibial shaft, in the lateral plane.
`Rotation is correct when the medial malleolus is ap-
`proximately 30° anterior to the lateral malleolus (FIG.
`19).
`
`The Transverse Tibial Cut
`
`While still under visual control of the axial alignment
`guide pins 518, 19, the transverse tibial cutting jig V is
`
`

`
`7
`
`4,646,729
`
`locked in place by drilling two 5 inch pins through the
`appropriate holes in the jig (FIG. 21). The alignrent
`guide 16 is now removed and tension is removed from
`the jig IV by loosening the thumb screws 40. As the
`knee is flexed to 90°, the femoral spacer/tensor jig IV
`will pull out of its anchoring holes in the femur. The jig
`IV is then slipped proximally, disengaging it from the
`tibial cutting jig V (FIG. 22). The transverse tibial pla-
`teau cut is then made by resting the saw blade flush
`against the broad flat surface of the cutting jig V (FIG.
`23). The cut is made as deeply as the saw blade will
`allow, while care is taken to protect collateral liga-
`ments. Once again, following the initial cut of the oscil-
`lating saw blade should be run back and forth across the
`flat surface of the transverse cutting jig to shave off any
`prominent bone that may be left posteriorly due to
`sciving of the saw blade in sclerotic bone. The jig V is
`then slipped off the locking pins and the cut is com-
`pleted, making sure that the posterior rims of the pla-
`teaus are level with the plane of the transverse cut.
`Additional care should be taken to preserve the poste-
`rior cruciate ligament. Since the transverse tibial cut is
`made parallel to the ground for optimal stress distribu-
`tion as the prosthesis bone interface and because the
`normal tibial plateau slopes 7° to 10° posteriorly, more
`bone will be removed anteriorly than posteriorly.
`Femoral Chamfer Cuts
`
`The femoral chamfer cutting jig V1 is inserted then
`into the femoral fixation holes. With the saw blade flush
`against the jig’s cutting planes 42, the anterior and pos-
`terior cuts are made (FIG. 24). As with other cutting
`jigs, it is important to maintain the saw blade perfectly
`flush with the cutting surfaces 42 of the jig to assure
`precise cuts, otherwise the femoral component will not
`fully seat.
`
`Tibial Component Rotational, Medial-Lateral And
`Anterior-Posterior Alignment
`
`'
`
`I The knee is extended and traction is applied from the
`foot to open the joint space. The posterior tabs of the
`appropriate sized tibial positioning/fixation jigs VII are
`hooked behind the cut proximal tibia (FIG. 25). The
`two posterior tabs 44 of the jig position behind the
`posterior rims of the tibial plateau, assuring correct
`posterior position of the tibial prosthesis. The knee is
`then flexed and the jig VII is centrally positioned. Since
`the posterior margins of the tibial plateaus are nearly
`parallel to the transverse axis of the tibia, the posterior
`tabs 44 will position the jig in correct rotation. Rota-
`tional and medial-lateral positioning are checked by
`slipping an axial alignment guide pin 19 through the
`appropriate right or left alignment hole in the anterior
`flange of the jig. The two anterior thumb screws 46 of
`the jig are then lightly tightened, securing the jig in
`place. If alignment is correct, the distal tip of the align-
`ment pin should center over the ankle joint with the
`medial malleolus 30° anterior to the lateral malleolus
`(FIG. 26). If this is not the case, the jig is manipulated
`into proper alignment. Rotational malalignment tends
`to be toward external rotation of the tibia.
`An appropriate size tibial trial prosthesis 48 is inserted
`onto the jig (FIG. 27). The trial femoral prosthesis is
`then positioned onto the femur.
`Initial Trial Reduction
`
`Range of motion and stability are now tested. If the
`» joint is too lax, the next thickest tibial trial is slipped
`
`8
`onto the tibial positioning/fixation jig VII. Once flex-
`ion, rotation and stability are satisfactory, the overall
`alignment is checked in full extension (FIGS. 28 and
`29). The alignment guide is removed, and the knee is
`flexed to 90°. Next, the femoral trial and tibial spacer
`are removed. The thickness marked on the tibial trial
`spacer indicates the thickness of the prosthesis to be
`implanted.
`
`l0
`
`15
`
`20
`
`25
`
`30
`
`35
`
`40
`
`45
`
`50
`
`55
`
`60
`
`65
`
`Final Tibial Preparation
`
`When the stemmed tibial prosthesis is being used, the
`window 50 in the jig VII is used, as a cutting guide. A
`é inch osteotome 52 is used to prepare the fixation slot
`while the jig is still locked ontothe proximal tibia (FIG.
`30).
`For the aforenoted prosthesis resurfacing tibial pros-
`thesis which uses medial and lateral fixation studs, a
`5/16 inch drill 54 is used to make the stud holes through
`the drill guides in the jig VII (FIG. 30). With fixation
`peg or stud holes completed, the jig is removed. A final
`check of the joint is made for posterior loose bodies, and
`soft tissue debridement is completed.
`Instruments
`
`FIGS. 31-94 illustrate in full scale various instru-
`ments used in accordance with this invention.
`FIGS. 31-35 illustrate the distal femoral cutting jig
`IA which is used for achieving axial alignment when
`used with the axial adjustment guide 16. Distal femoral
`cuts are made when jig IA is used with cut jig IB. As
`indicated therein, the upper surface of the central por-
`tion 56 of jig IA has attached thereto a pair of lateral
`handles 22, 22 while a medial handle 20 also extends
`from the front face of central section 56. A positioning
`pin 58 is disposed on the rear face of central section 56
`as previously described. Attaching means are provided
`on the upper surface of the central section 56 for selec-
`tive attachment of the pin holder alignment guide 16.
`The attaching means is in the form of a pair of holes 60,
`62 which are of different diameter corresponding to the
`different diameter locking pins 64, 66 (FIG. 37) of the
`pin holder 16. In this manner there is assurance that the
`pin holder can be mounted in only the correct position.
`The rear surface of central section 56 also includes a
`pair of short alignment pins 68 as previously described.
`A vertical cut-out 70 extends completely through
`central section 56 for receiving the tongue 72 (FIG. 45)
`of the distal femur cutting jig IB. An adjustable locking
`screw 74 is movable into cut-out 70 to lock tongue 72 in
`place.
`-
`FIGS. 36-44 illustrate the details of pin holder 16
`which is used with guide pins 18, 19 (FIG. 19) to assure
`correct axial alignment throughout the surgical proce-
`dure. Pin holder 16 is in the form of an elongated bar
`and includes bifurcated pins 64, 66 to complement the
`holes 60, 62 in jig IA. The lower portion of pin holder
`16 is provided with a pair of mirror image tibial holes
`76, 76 for selectively receiving tibial alignment pin 19.
`The holes are angled oppositely each other, as previ-
`ously described, with one holder being for the right
`knee and the other being for the left knee. Similarly,
`three sets of holes 78, 80, 82 are provided for the femo-
`ral alignment pin 18. The holes are disposed at the most
`likely angle required such as 7°, 9° and 11" with one set
`being for the right knee and the other for the left knee.
`FIGS. 45-50 illustrate the details of distal femoral
`cutting jig IB which is in the form of a plate or bar
`having a downwardly extending tongue 72 with non-
`
`

`
`9
`symmetric cross-section of complementary size and
`shape to fit within the mortise cut-out 70 of jig IA. Jig
`IB is used for cutting the distal femoral condyles. Plate
`84 includes a pair of femoral securing pins 86 for attach-
`ment to the femur with two pairs of positioning pins 88
`spaced inwardly thereof. The upper surface of plate 84
`is also provided with a pair of non-identical holes 90, 92
`for receiving the locking pins 64, 66 of pin holder 16.
`Plate 84 is also provided with a generally vertical guide
`surface 94 on each wing portion thereof to act as a saw
`cutting guide (see FIG. 13). A pair of vertical holes 96
`are provided in plate 84.
`FIGS. 51-56 illustrate the femoral cutting jig II
`which is used to determine the rotational, medial-lateral
`and anterior-posterior orientation of the femoral com-
`ponent and allows drilling of holes for prosthesis fixa-
`tion stud. As shown therein, jig II is in the form of a
`generally vertical plate 98 having a flat inner surface
`100. A pair of posterior skids 24 extend outwardly from
`inner surface 100 at the lower portion thereof. A pair of
`drill holes 26 extend through plate 98 as previously
`described. If the posterior condyles are intact, a hole
`may be drilled therein through the aid of one of the drill
`holes 102, 102 with holes 102, 102 being inclined for the
`right and left knee. Plate 98 also includes a pair of posi-
`tioning oins 104 on its inner surface 100.
`FIGS. 57-62 illustrate the anterior-posterior cutting
`jig III. As indicated therein, jig III is in the form of a bar
`or plate 106 which has a flat vertical surface 108. A pair
`of distal femoral fixation pins 32 extend from surface
`108. A generally horizontal lower cutting guide surface
`110 is also provided as well as an upwardly inclined
`cutting guide surface 112. A cut-out 114 is located cen-
`trally of bar 106. Preferably jig III would come in small,
`medium and large sizes.
`FIGS. 63-68 illustrate the details of femoral spacer/-
`tensor jig IV which assembles and positions transverse
`tibial cutting jig V to determine correct soft tissue bal-
`ance and tibial axial and rotational alignment prior to 40
`making the transverse tibial cut. Jig IV comprises a pair
`of spaced fixed legs 116 with each leg having a lower
`vertical section and an inwardly inclined upper section.
`A transverse br

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket