throbber
Dosimetric effect of respiration-gated beam on IMRT delivery
`Jun Duan,a) Sui Shen, John B. Fiveash, Ivan A. Brezovich, Richard A. Popple,
`and Prem. N. Pareek
`Department of Radiation Oncology, University of Alabama Birmingham, 619 South 19th Street,
`Birmingham, Alabama 35233
`~Received 8 January 2003; revised 29 April 2003; accepted for publication 27 May 2003;
`published 25 July 2003!
`Intensity modulated radiation therapy ~IMRT! with a dynamic multileaf collimator ~DMLC! re-
`quires synchronization of DMLC leaf motion with dose delivery. A delay in DMLC communication
`is known to cause leaf lag and lead to dosimetric errors. The errors may be exacerbated by gated
`operation. The purpose of this study was to investigate the effect of leaf lag on the accuracy of
`doses delivered in gated IMRT. We first determined the effective leaf delay time by measuring the
`dose in a stationary phantom delivered by wedge-shaped fields. The wedge fields were generated by
`a DMLC at various dose rates. The so determined delay varied from 88.3 to 90.5 ms. The dosim-
`etric effect of this delay on gated IMRT was studied by delivering wedge-shaped and clinical IMRT
`fields to moving and stationary phantoms at dose rates ranging from 100 to 600 MU/min, with and
`without gating. Respiratory motion was simulated by a linear sinusoidal motion of the phantom. An
`ionization chamber and films were employed for absolute dose and 2-D dose distribution measure-
`ments. Discrepancies between gated and nongated delivery to the stationary phantom were ob-
`served in both absolute dose and 2-D dose distribution measurements. These discrepancies in-
`creased monotonically with dose rate and frequency of beam interruptions, and could reach 3.7% of
`the total dose delivered to a 0.6 cm3 ion chamber. Isodose lines could be shifted by as much as 3
`mm. The results are consistent with the explanation that beam hold-offs in gated delivery allowed
`the lagging leaves to catch up with the delivered monitor units each time that the beam was
`interrupted. Low dose rates, slow leaf speeds and low frequencies of beam interruptions reduce the
`effect of this delay-and-catch-up cycle. For gated IMRT it is therefore important to find a good
`balance between the conflicting requirements of rapid dose delivery and delivery accuracy. © 2003
`@DOI: 10.1118/1.1592017#
`American Association of Physicists in Medicine.
`
`Key words: gated IMRT, respiratory gating, IMRT, dynamic wedge, dosimetric effect
`
`I. INTRODUCTION
`
`Advances in technology have allowed delivery of complex
`radiation fluence patterns with the aid of dynamic multileaf
`collimators ~DMLC!. This technique, known as intensity
`modulated radiation therapy ~IMRT!, is capable of generat-
`ing radiation doses highly conformed to the target while
`sparing surrounding normal tissues. It has been successfully
`applied for irradiation of malignant carcinomas in various
`sites.1– 4 In many instances, patients could not have been
`treated without this sophisticated technique.
`Organ motion due to breathing poses a special challenge
`to IMRT, and may have hindered its wider application. Con-
`ventional
`three-dimensional conformal
`radiation therapy
`~3DCRT! accommodates organ motion by applying wide
`margins around the clinical target volume ~CTV!, usually in
`the range of 1 to 2 cm. While such broad margins inevitably
`include significant volumes of normal tissues, their effect on
`doses to tumors and normal tissues can be estimated. In
`IMRT, however, the simultaneous motions of beam and tar-
`get can lead to significant deviations from the planned doses
`that are difficult to predict.5– 8
`In recent years, respiration-gated radiotherapy has shown
`the potential of reducing dose errors by synchronizing radia-
`tion delivery with the patient’s breathing.9–14 Radiation is
`
`delivered within a small window of each respiratory cycle at
`a phase where organ motion is minimal, e.g., at the end of
`expiration or at
`the end of inspiration. Residual motion
`within the gating window is only a small fraction of the full
`movement.
`The operation of linear accelerators in conjunction with
`DMLCs has been extensively studied.15–22 Litzenberg et al.
`identified the limitations of delivery parameters ~dose rate,
`tolerance, leaf speed, total monitor units, etc.! in realistic
`dynamic delivery that affect MLC leaf position errors that
`triggers beam hold-off.22 By incorporating these limitations
`into the leaf sequencing algorithm, they were able to produce
`leaf sequence of sliding-window IMRT fields deliverable
`with the prescribed constant dose rate, requiring less delivery
`time, and having well-defined, calculable transmission dose
`characteristics. Xia et al. investigated the limitations of com-
`munication and sampling rate in IMRT delivery with dy-
`namic multileaf collimator.17 They observed distortions and
`dose variations on intensity patterns delivered with low MU
`and high dose rate due to insufficient sampling and commu-
`nication lag between treatment console and MLC worksta-
`tion. Several authors have investigated the compatibility of
`dynamic wedges and IMRT with gated delivery. Using film
`dosimetry, Kubo and Wang studied one-dimensional dose
`profiles from gated operations of a Varian 2100C accelerator
`
`2241
`
`Med. Phys. 30 (cid:132)8(cid:133), August 2003
`
`0094-2405(cid:213)2003(cid:213)30(cid:132)8(cid:133)(cid:213)2241(cid:213)12(cid:213)$20.00
`
`© 2003 Am. Assoc. Phys. Med.
`
`2241
`
`Varian (Ex. 1021)
`IPR of U.S. Pat. No. 7,961,843
`
`

`

`2242
`
`Duan etal.: Dosimetric effect of respiration gated beam
`
`2242
`
`pected and actual leaf positions, and other information. The
`acquisition, transfer, processing, and storage of this informa-
`tion requires a significant amount of time, resulting in a re-
`ported 50 to 80 ms delay between the accelerator and DMLC
`response to a given condition.22,25,26
`The mechanical limit of leaf speed is approximately 2.8
`cm/s projected to the isocenter. In our system, the leaf speed
`limit is set at 2.5 cm/s for IMRT delivery by the leaf motion
`calculator software. The dose rate is maintained at the value
`programmed at the linear accelerator console provided that
`the required DMLC leaf motions do not result in any leaf
`exceeding the speed limit. Otherwise, the DMLC application
`reduces the dose rate to fit the speed of leaf movement. In-
`dividual leaf speeds within a segment are constant, but dif-
`ferent leaves travel at different speeds. The accuracy of leaf
`position calibration relative to beam central axis is better
`than 1 mm. The precision of leaf position is 0.002 mm de-
`termined by the counts of the encoder per mm ~512 counts
`per mm for the 0.5 mm leaves used in our study!.27 For
`dynamic delivery, a leaf error tolerance setting permits the
`leaf position to be within a small range of the expected po-
`sition. If the position error exceeds the tolerance, an MLC
`interlock will be activated and the beam turned off. The al-
`lowable leaf position tolerance projected to isocenter can be
`set between 0.05 and 0.50 cm. A tight tolerance tends to
`minimize leaf errors. However, too tight a tolerance will
`cause frequent leaf error interlock that results in an unaccept-
`able number of beam interruptions due to tolerance violation.
`A value of 0.20 cm has been used in our system to compro-
`mise between leaf position accuracy and frequency of
`DMLC leaf error interlock.19 This tolerance setting was used
`in all measurements.
`
`B. Dosimetric method for determining effective leaf
`delay time
`
`equipped with an 80-leaf DMLC. They concluded that gated
`operation was compatible with enhanced dynamic wedges
`and IMRT for dose rates up to 320 MU/min, the maximum
`dose rate provided by their Clinac 2100C.23 Solberg et al.
`found that the Novalis micro-MLC could be used in conjunc-
`tion with respiration gating.24 Hugo et al. investigated the
`effect of gating window size and choice of delivery method
`~segmented and dynamic multileaf collimation! on dosimetry
`of gated IMRT using both single and composite field dosim-
`etry. They found that the error could be reduced by decreas-
`ing gating window size, and that in most cases dynamic de-
`livery generated larger delivery errors than segmented
`delivery.
`A commercial gating system has been installed on one of
`our Clinac 21EX linear accelerators ~Varian Medical Sys-
`tems, Palo Alto, CA!, which is equipped with an IMRT ca-
`pable 120-leaf DMLC. The accelerator can furnish dose rates
`up to 600 MU/min, and thereby offers the potential of faster
`treatment delivery. However, the compatibility of such a high
`dose rate with gated IMRT has not been fully documented. A
`delay in the communications between the DMLC and the
`accelerator is known to cause leaf lag and deviations in the
`delivered radiation dose.17,22,16,25 The problem may be exac-
`erbated by the frequent beam interruptions associated with
`gating. In this study, we investigate the dosimetric effect of
`DMLC leaf lag on gated IMRT in relation to dose rate and
`frequency of beam interruption.
`
`II. MATERIALS AND METHODS
`
`A. Intensity modulating DMLC and leaf lag
`
`The leaves of the Varian Millennium 120-Leaf DMLC are
`5 mm thick within the central 20 cm of the field and 1 cm
`thick over the peripheral 10 cm on each side. IMRT is deliv-
`ered with MLC in two modes, segmented multileaf colli-
`mated IMRT ~SMLC-IMRT!, also known as the step-and-
`shoot mode, and dynamic multileaf collimated IMRT
`~DMLC-IMRT!, also known as the sliding window mode. In
`the SMLC-IMRT mode, the beam is turned off while the
`collimator changes to the next shape in the treatment se-
`quence. In the DMLC-IMRT mode, the beam remains on
`while the collimator changes from one shape to the next. The
`relationship between leaf positions and delivered MU is
`specified in a leaf sequence file. The instantaneous shape of
`the beam during the continuous motion of the treatment is
`determined by linear interpolation of MU and leaf position
`between the previous and the next segment in the sequence.
`The instantaneous accumulated dose value is used as an in-
`dex to keep the relationship constant among continuously
`varying factors during the treatment, such as DMLC beam
`shape, dose rate, and movements of the linear accelerator.
`The DMLC-IMRT mode, which is often used in clinics, was
`used in this study.
`The DMLC on our machine operates through a DMLC
`controller which sends commands to the DMLC head assem-
`bly to move the leaves and the DMLC transmits the infor-
`mation back to the controller. Through a communication
`link, the controller checks the status of the accelerator, ex-
`
`Medical Physics, Vol. 30, No. 8, August 2003
`
`D%5
`
`3100%,
`
`As mentioned above, the delay in accelerator feedback to
`the DMLC controller results in a lag between dose delivery
`and the DMLC awareness of the delivered monitor units
`~MU!. Furthermore, leaf tolerance allows additional leaf lag
`up to the tolerance setting. In a dynamic delivery, these lags
`cause a difference between expected and actually delivered
`dose. The percent dose error, D%, can be calculated using
`dt(cid:149)DR
`MU
`where dt is the effective delay time, DR is the dose rate
`~MU/min!, and MU is the monitor unit setting for the DMLC
`segment. Depending on the dose rate and MU setting for the
`DMLC segment, the relative dose error can be significant.
`However, this formula cannot be directly used to experimen-
`tally determine the delay time. For a wedge field shaped by
`opening or closing leaf gap in DMLC-IMRT mode, the delay
`only affects the doses of the beginning and ending segments
`of the DMLC pattern. The beginning segment corresponds to
`either a fully open gap for the wedge shaped by closing the
`DMLC gap or a fully closed gap for the wedge shaped by
`opening the DMLC gap. The ending segment is opposite to
`
`~1!
`
`

`

`2243
`
`Duan etal.: Dosimetric effect of respiration gated beam
`
`2243
`
`sured at various dose rates ~100 MU/min to 600 MU/min!
`with the ion chamber. Monitor units were set at 12 and 56 for
`the 3 and 14 cm wide wedge fields, respectively, so that the
`leaf speeds for the two wedges were identical for a given
`dose rate, and the leaf speed could reach its maximum of 2.5
`cm/s at 600 MU/min. The average of the two measurements
`at collimator angles 180° apart was used in computing the
`leaf delay time. For each data point, three measurements
`were taken and the average was used. The dose correspond-
`ing to zero dose rate, i.e., no leaf lag effect, was obtained by
`linear extrapolation. Measured doses were then normalized
`to the extrapolated dose value of zero dose rate. Linear re-
`gression was used to fit the D%~DR! curve. Using Eq. ~2!,
`the delay time can be determined from the slope of the linear
`function.
`Leaf transmission was measured using the same measure-
`ment geometry with closed leaf gap. The leaf edges were
`offset 7 cm from central axis to avoid leakage contribution
`due to rounded leaf edge. Five thousand MU were delivered
`and the ion chamber was used to measure the dose on central
`axis with the same position and orientation as in field mea-
`surement. Leaf transmission corrections were applied in pro-
`portion to 50% of the MU in wedge field measurements. In
`wedge factor measurements in which static MLC fields were
`used, leaf transmission correction was applied in proportion
`to 100% of the MU if the leaves crossed the central axis.
`
`C. Respiratory gating system
`The Real-Time Position Management ~RPM! Respiratory
`Gating System ~Varian Medical Systems, Palo Alto, CA! was
`used for gated operation. It consists of an infrared video
`camera, an infrared illuminator ring, an infrared-reflective
`marker block and a PC workstation with system control soft-
`ware. An in-room view finder aids in the adjustment of the
`marker block position with respect to the camera. The video
`camera tracks respiratory motion by monitoring the markers
`that are attached to the patient’s chest or abdomen. The gat-
`ing system sends the gating signal to the linear accelerator to
`trigger beam hold-off when the target volume moves beyond
`the preset limits.
`
`D. IMRT dose delivery and measurement
`
`Various IMRT fields were delivered to a rectangular poly-
`styrene phantom. To simulate respiration-induced organ mo-
`tion, the phantom was placed on a platform that could move
`sinusoidally in the crainiocaudal direction. Sinusoidal motion
`was provided by a steel arm connected eccentrically to a
`motorized wheel rotating at a given angular velocity. The
`frequency of sinusoidal motion could be continuously ad-
`justed from 0 to 1.6 cycles/s ~96 cycles/min! by adjusting the
`angular velocity of the wheel. The amplitude of the motion
`was adjustable from 0 to 2.5 cm by changing the distance
`between the mounting point of the arm and the center of the
`wheel, corresponding to a maximum motion range of 5 cm.
`The infrared reflective marker block was placed on the mov-
`ing platform to generate gating signals. For measurements in
`the moving phantom, a motion amplitude of 1 cm ~2 cm
`
`the beginning segment. Taking the wedge shaped by closing
`the leaf gap as an example, dose contribution to the central
`axis is negligible from the ending segment in which the
`DMLC gap is nearly closed. Therefore, the dose error due to
`leaf lag is predominantly contributed by the leaf delay at the
`beginning segment when the leaf gap is fully open, i.e., dD
`5k(cid:149)dt(cid:149)DR, where k5D/MU ~cGy/MU! is MU-to-dose
`conversion factor for the open field. For a wedge field, the
`wedge factor, WF, is defined as WF5D/D 0 , where D 05k
`(cid:149)MU is the dose of the open field and D is the dose of the
`wedge field delivered without leaf lag. Equation ~1! for a
`wedge field becomes
`dD
`D
`
`D%5
`
`3100%5
`
`3100%.
`
`~2!
`
`dt(cid:149)DR
`MU(cid:149)WF
`To obtain the wedge factor, dose delivered by the wedge field
`without delay effect, D, must be determined. Provided that
`D˙ , the dose rate at central axis without leaf lag effect as a
`function of time, is known, the dose of the wedge field can
`be computed by integration
`D˙ ~t !(cid:149)dt,
`where T is the total delivery time. Note that D˙ denotes the
`dose rate at depth on central axis, or cGy/s, while DR de-
`notes the nominal dose rate of the accelerator, or MU/min.
`Assuming that the leaves move at a constant speed, v, then
`the position of the moving leaves, x, can be determined by
`x5v(cid:149)t2w/2, where t and w are delivery time and wedge
`width, respectively. Equation ~3! thus becomes
`E
`D˙ ~x !(cid:149)dx,
`where D˙ (x) is the dose rate at depth on central axis without
`leaf lag as a function of moving leaf position. To eliminate
`leaf lag effect, static MLC fields of gradually decreasing gap
`size were used to determine D˙ (x) at different leaf positions.
`The static gap size was changed gradually by changing the
`position of one bank of leaves while keeping the other bank
`unchanged. Assuming constant DR, for a given MU, D˙ (x)
`can be determined by
`D~x !
`MU/DR
`
`~3!
`
`~4!
`
`D5E
`
`T
`
`0
`
`w/2
`
`2w/2
`
`1 v
`
`D5
`
`D˙ ~x !5
`
`.
`
`~5!
`
`A 0.6 cc ion chamber was used to measure D(x). The ion
`chamber was positioned at the central axis with the long axis
`perpendicular to the direction of wedge gradient. Similar to
`conventional hard wedge measurement, to minimize errors
`due to leaf and ion chamber position accuracy, measurements
`were performed at two collimator angles 180° apart and the
`average was used for calculation. Once D˙ (x) is determined,
`the dose of the wedge field without leaf delay effect, D, is
`computed using Eq. ~4!. The wedge factor, WF, is then de-
`termined by the ratio of D to D 0 .
`To determine the delay time, dt, central axis doses dy-
`namically delivered by 3 and 14 cm wedge fields were mea-
`
`Medical Physics, Vol. 30, No. 8, August 2003
`
`

`

`2244
`
`Duan etal.: Dosimetric effect of respiration gated beam
`
`2244
`
`FIG. 1. Fluence maps used in the study: ~a! 14 cm wide
`single wedge field, ~b! 14 cm wide double wedge field,
`~c! 14 cm wide quadruple wedge field, ~d! 3 cm wide
`single wedge field, ~e! brain IMRT field, ~f! head-and-
`neck IMRT field, ~g! prostate IMRT field, and ~h! lung
`IMRT field.
`
`motion range! was used. Gating windows centered at the end
`of expiration, where motion is minimum, were used for vari-
`ous gating frequencies.6 For stationary measurements with
`gated delivery, the phantom was placed directly on the treat-
`ment couch while the moving platform served as a surrogate
`to generate gating signals. Since the phantom was not mov-
`ing, gating window size and frequency were adjusted for the
`desired number of beam hold-offs per unit MU without the
`concern of residual motion.
`Figure 1 shows the fluence maps of four wedge-shaped
`fields and four clinical IMRT fields used in the study,
`single wedge, 14314 cm2 double
`namely, 14314 cm2
`wedge, 14314 cm2 quadruple wedge, 3314 cm2 single
`wedge, and IMRT fields for brain, head-and neck, prostate
`and lung. In this study, wedge size is specified by its width
`defined as the field dimension in the direction of the wedge
`gradient. DMLC leaf sequence files for the wedge-shaped
`fluence maps were generated with the Varian ‘‘MLC Shaper’’
`software ~version 6.1! while those for the clinical IMRT
`fields were generated with the Varian Cadplan/Helios Treat-
`ment Planning System. The wedge field can be generated by
`either opening the leaf gap from the closed position or clos-
`ing the leaf gap from the open position. To study dose de-
`viations due to beam gating, doses delivered to the stationary
`phantom with gating were compared to those without gating.
`Kodak X-omat V and X-omat TL films ~Eastman Kodak
`Company, Rochester, NY! were used with a film dosimetry
`system to measure planar dose distributions. Special care
`was taken to ensure that the same batch of films was used,
`and that they were developed under the same processor con-
`ditions. The film dosimetry system used for the study in-
`cluded a Vidar VXR-16 DosimetryPro Film digitizer ~Vidar
`Systems Corporation, Herndon, VA! interfaced with a com-
`puter. In-house software and RIT Radiation Therapy Film
`Dosimetry System Software ~Radiological Imaging Technol-
`ogy, Inc., Colorado Springs, CO! was used for film analysis
`and comparisons. For wedge-shaped fields,
`isodose lines
`from 10% to 100% at 10% increment were used to evaluate
`the dose distributions. Distance-to-agreement ~DTA! and
`gamma ~g! were used for quantitative comparison of dose
`
`Medical Physics, Vol. 30, No. 8, August 2003
`
`distributions of gated and nongated delviery. Gamma is an
`index proposed by Low et al. for quantitative evaluation of
`dose distributions.28,29 It is the minimum vector difference in
`dose-distance space between a reference dose distribution
`and a measurement distribution. For a given set of clinical
`criteria, e.g., 3% dose difference and 3 mm distance, gamma
`~denoted as g3/3) evaluates if a measured dose distribution
`agrees with a reference dose distribution within the clinical
`criteria. A gvalue larger than unity indicates that the refer-
`ence dose does not agree with the measured dose within the
`clinic criteria. For wedge fields, we used mean DTA and
`mean gto compare dose distributions of gated and nongated
`deliveries. The mean DTA and gare defined as the average
`of each, respectively, over the central 60% of a measured
`isodose line.
`Absolute dose measurements were performed with an ion
`chamber. For gated deliveries, dosimetric artifacts were gen-
`erated by leaf lag at each leaf position where a beam hold-off
`was triggered. In ion chamber measurements, the effect of
`the artifacts was detected only if it is projected onto the
`chamber volume when the beam hold-off is triggered. A
`small ion chamber would lead to poor reproducibility and
`partial volume effect in measurements because the location
`of beam hold-off was not perfectly reproducible even with
`identical treatment and gating parameters. To overcome this
`problem, a 0.6 cm3 farmer type ion chamber ~PTW Model
`23333! of 23 mm active chamber length was used in the dose
`measurements. The ion chamber was placed such that the
`chamber long axis was along the leaf moving direction un-
`less otherwise specified. For wedge-shaped fields, this corre-
`sponds to the direction of wedge gradient. For comparison
`between gated and nongated deliveries in a stationary phan-
`tom, the chamber position remained unchanged during mea-
`surements of the two sets of deliveries. Measurements in the
`stationary phantom without beam gating were used as stan-
`dards for comparison with other measurements. All measure-
`ments were done at 5 cm depth. To minimize random errors,
`three readings were taken in all ion chamber measurements
`and the average was used in the results presented in this
`study. Majorities of the measurements presented in the study
`
`

`

`2245
`
`Duan etal.: Dosimetric effect of respiration gated beam
`
`2245
`
`FIG. 2. Normalized dose as a function of MLC leaf position measured with
`static MLC fields to simulate 3314 cm2 and 14314 cm2 wide wedge fields
`without leaf lag effect. The measurement was corrected for leaf leakage if
`the leaves blocked the central axis.
`
`were performed in the stationary phantom with the moving
`platform to generate gating signal to gate the accelerator.
`Measurement results presented hereafter, gated or nongated,
`refer to measurements in the stationary phantom unless ex-
`plicitly specified otherwise.
`
`III. RESULTS
`
`A. Leaf delay time
`
`The dose rate at central axis as a function of leaf position
`is shown in Fig. 2 for 3 and 14 cm wide wedge fields. The
`integral dose of Eq. ~4! was computed and compared to the
`dose of the open field to yield wedge factors. For the 3 and
`14 cm wedges, wedge factors were 0.5413 and 0.5086, re-
`spectively.
`Figure 3 shows the percent dose error, D%, as a function
`of dose rate along with the fitting curves for the 3 and 14 cm
`wide wedge fields generated by opening and closing the
`DMLC leaf gap. As can be seen in the figure, dose errors are
`distributed nearly symmetrically about
`the abscissa for
`wedges generated by opening and closing the leaf gap. Lin-
`ear regression was used for the curve fitting. Fitting param-
`eters are listed in Table I. The near unity R 2 values indicate
`a linear relationship between D% and dose rate. For the 3
`
`FIG. 3. Percent dose error, D%, as a function of dose rate along with the
`fitting curves for the 3 and 14 cm wide wedge fields generated by opening
`and closing the DMLC leaf gap. The dose error was calculated by compar-
`ing to the linearly extrapolated dose value that would be measured without
`any leaf error. Twelve and 56 MU were delivered with the 3 and 14 cm wide
`wedges, respectively, to yield a maximum leaf speed of 2.5 cm/s at the dose
`rate of 600 MU/min. Leaf delay times were calculated from the slopes of the
`fitting curves.
`
`and 14 cm wedges, the dose delivered at 600 MU/min was
`higher than that at 100 MU/min by 10.4% and 2.52% when
`the wedges were generated by closing the leaf gap, but lower
`by 11.8% and 2.59% when the wedges were generated by
`opening the leaf gap, respectively. The fact that measured
`doses increased with dose rate when the wedge was gener-
`ated by closing the leaf gap and decreased when the wedge
`was generated by opening the leaf gap confirmed that DMLC
`leaves were lagging behind dose delivery. The effective de-
`lay time was calculated using Eq. ~2! for each wedge and is
`listed in Table I. The effective delay times for the four
`wedges varied between 88.3 and 90.5 ms. They were consis-
`tent for wedges generated by opening and closing the DMLC
`leaf gap.
`
`B. Dosimetric effect on gated dynamic delivery
`
`Figure 4 shows the dose distributions delivered by a 1 cm
`wide DMLC leaf gap sliding across a 10 cm wide field.
`Kodak X-Omat TL film was used for the measurements. The
`dose was delivered with and without gating at a dose rate of
`600 MU/min and a leaf speed of 2.5 cm/s. Dose level is
`represented by the gray scale where black corresponds to
`zero dose. While a uniform dose distribution was obtained
`across the field in nongated delivery @Fig. 4~a!#, distinctive
`
`TABLE I. Effective leaf lag time and linear regression curve fitting parameters for fitting dose error as a function
`of dose rate in nongated delivery.
`
`3 cm wedge
`
`14 cm wedge
`
`Closing leaf
`
`Opening leaf
`
`Closing leaf
`
`Opening leaf
`
`Slope ~min/MU!
`R 2 value
`Effective delay time ~ms!
`
`2.323E204
`0.9991
`90.5
`
`22.283E204
`0.9979
`89.0
`
`5.184E205
`0.9953
`88.6
`
`25.170E-05
`0.9993
`88.3
`
`Medical Physics, Vol. 30, No. 8, August 2003
`
`

`

`2246
`
`Duan etal.: Dosimetric effect of respiration gated beam
`
`2246
`
`FIG. 4. Dose distribution delivered by 1 cm wide DMLC leaf gap sliding
`across a 10 cm wide field ~a! without and ~b! with gating measured with
`Kodak X-Omat TL films. The dose was delivered at dose rate of 600 MU/
`min while the DMLC leaves traveled at 2.5 cm/s. The brightness of the pixel
`represents the dose level.
`
`dosimetric artifacts appeared in gated delivery @Fig. 4~b!#. A
`cold and a hot stripe appeared, respectively, on the left and
`right side of the DMLC window at the position where a
`beam hold-off took place. Dose profiles along the midlines
`across the two fields are shown in Fig. 5. The dose was
`normalized to central axis (distance50). The cold and hot
`spots are about 5% of the average dose.
`For a typical respiratory rate of 15 cycles/min and a gat-
`ing window of 1.0 s, Table II lists the dose ratios of gated to
`nongated deliveries of four different wedge-shaped IMRT
`fields, namely, 14314 cm2 single wedge, 14314 cm2 double
`wedge, 14314 cm2 quadruple wedge, and 3314 cm2 single
`wedge delivered under various conditions. These data were
`measured in the stationary phantom using the ion chamber.
`Differences between gated and nongated beam were less than
`0.82% for leaf speeds ranging from 0.23 to 2.33 cm/s.
`
`FIG. 5. Dose profile across the two fields shown in Fig. 4. Central axis dose
`value of nongated delivery was used for normalization.
`
`As the gating frequency increases or the gating window
`decreases, the number of beam hold-offs increases. For the
`stationary phantom, Fig. 6 shows the dose ratio of gated to
`nongated delivery as a function of the number of beam hold-
`offs triggered during delivery for ~a! a 3 cm wedge generated
`by opening the leaf gap, ~b! a 3 cm wedge generated by
`closing the leaf gap, and ~c! a 14 cm wedge generated by
`opening the leaf gap. Twelve and 56 MU were delivered at
`dose rates from 100 to 600 MU/min for the 3 and 14 cm
`wedges, respectively, corresponding to DMLC leaf speeds
`between 0.42 and 2.5 cm/s. These ion chamber measure-
`ments show that, for a given dose rate, the dose discrepancy
`between gated and nongated deliveries became greater as the
`number of beam hold-offs was increased. For the 3 cm
`wedge the largest dose difference between gated and non-
`gated deliveries, observed at 600 MU/min with 5 beam hold-
`offs, was 3.7%. With 5 beam hold-offs, the dose measured at
`600 MU/min was higher than that at 100 MU/min by 3.0%.
`The dose measured for the 14 cm wedge showed much
`
`TABLE II. Dose ratio of gated to nongated deliveries measured in a stationary phantom with 1.0 s gating window
`and a respiration rate of 15 cycles/min.
`
`Field
`
`Single wedge (14314 cm2)
`Single wedge (14314 cm2)
`Single wedge (14314 cm2)
`Single wedge (14314 cm2)
`Single wedge (3314 cm2)
`Single wedge (3314 cm2)
`Double wedge (14314 cm2)
`Double wedge (14314 cm2)
`Quadruple wedge (14314 cm2)
`Quadruple wedge (14314 cm2)
`Quadruple wedge (14314 cm2)
`
`DR
`~MU/min!
`
`Leaf speed
`~cm/s!
`
`400
`400
`200
`100
`200
`400
`200
`400
`200
`400
`400
`
`0.93
`1.87
`2.33
`2.33
`0.83
`0.83
`0.47
`0.93
`0.93
`0.93
`0.23
`
`MU
`
`100
`50
`20
`10
`12
`24
`100
`100
`50
`100
`400
`
`Dose ratio
`~Gated/nongated!
`
`1.0015
`1.0031
`1.0048
`1.0063
`1.0082
`1.0080
`1.0004
`1.0032
`1.0019
`1.0021
`1.0004
`
`Medical Physics, Vol. 30, No. 8, August 2003
`
`

`

`2247
`
`Duan etal.: Dosimetric effect of respiration gated beam
`
`2247
`
`FIG. 7. Measured doses of a 3314 cm2 wedge field as a function of nominal
`dose rate for various number of beam hold-offs ~BHOs! over 12 MU. The
`wedge field was generated by opening or closing the DMLC leaf gap.
`
`min without respiratory gating. The dose measured without
`gating is also included for comparison. For wedges generated
`by opening and closing the leaf gap, respectively, dose de-
`creased and increased significantly with dose rate. The larg-
`est change, however, occurred when the dose was delivered
`without gating. The dose discrepancy for the 3 cm wedge
`field between dose rates of 100 and 600 MU/min exceeded
`10%.
`For the four clinical IMRT fields presented in Fig. 1, Fig.
`8 depicts the dose as a function of dose rate measured in the
`stationary phantom without gating. The dose was normalized
`to that measured at 100 MU/min. For the prostate and brain
`fields the dose hardly changed with dose rate. For the head-
`and-neck and the lung fields, however, dose increased or
`decreased in proportion to dose rate. The largest dose differ-
`ence was 1.2% in the lung field. For measurements of gated
`delivery in the stationary phantom, Fig. 9 shows the dose of
`the lung field as a function of gating frequency expressed as
`
`FIG. 6. Dose ratios of gated to nongated deliveries as a function of the
`number of beam hold-offs for ~a! the 3 cm wedge field generated by opening
`the leaf gap, ~b! the 3 cm wedge field generated by closing the leaf gap, and
`~c! the 14 cm wedge field generated by opening the leaf gap. Twelve and 56
`MU were delivered for the 3 and 14 cm wedges, respectively. The dose was
`measured in the stationary phantom with or without gating.
`
`smaller difference between gated and nongated delivery be-
`cause of the larger number of monitor units. The largest dif-
`ference shown in the figure is 1.2%. Figure 7 shows the dose
`of the 3 cm wedge as a function of dose rate for 1, 2, 3 and
`5 beam hold-offs triggered during the delivery of 12 MU.
`Dose values were normalized to that delivered at 100 MU/
`
`Medical Physics, Vol. 30, No. 8, August 2003
`
`FIG. 8. Measured doses of four clinical IMRT fields delivered to the station-
`ary phantom without gating as a function of dose rate.
`
`

`

`2248
`
`Duan etal.: Dosimetric effect of respiration gated beam
`
`2248
`
`FIG. 9. Measured dose of a lung IMRT field delivered to the stationary
`phantom with gating as a function of gating frequency for various dose
`rates.
`
`the number of beam hold-offs per 100 MU. Doses with and
`without beam gating exhibited differences of less than 0.8%.
`Figure 10 shows isodose distributions of the 14 cm wedge
`field delivered to a stationary phantom using gated beam
`overlaying those delivered with nongated beams at dose rates
`of ~a! 200 MU/min and ~b! 600 MU/min. A respiratory rate
`of 15 cycles/min and a gating window of 1.0 s center at the
`end of expiration were used in gated delivery. For the dose
`rate of 200 MU/min, the maximum distance-to-agreement
`was about 1 mm @Fig. 10~a!#. However, for the dose rate of
`600 MU/min, the distance-to-agreement was substantially
`greater. The largest mean distance-to-agreement was 3 mm
`@Fig. 10~b!#.
`Table III compares isodose distributions of 14314 cm2
`wedge field delive

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket