throbber
Int. J. Radiation Oncology Biol. Phys., Vol. 53, No. 2, pp. 453–463, 2002
`Copyright © 2002 Elsevier Science Inc.
`Printed in the USA. All rights reserved
`0360-3016/02/$–see front matter
`
`PII S0360-3016(02)02777-3
`
`PHYSICS CONTRIBUTION
`
`CLINICAL IMPLEMENTATION OF INTENSITY-MODULATED ARC THERAPY
`
`CEDRIC X. YU, D.SC., X. ALLEN LI, PH.D., LIJUN MA, PH.D., DONGJUN CHEN, PH.D.,
`SHAHID NAQVI, PH.D., DAVID SHEPARD, PH.D., MEHRDAD SARFARAZ, PH.D.,
`TIMOTHY W. HOLMES, PH.D., MOHAN SUNTHARALINGAM, M.D., AND CARL M. MANSFIELD, M.D.
`Department of Radiation Oncology, University of Maryland School of Medicine, Baltimore, MD
`
`Purpose: Intensity-modulated arc therapy (IMAT) is a method for delivering intensity-modulated radiation
`therapy (IMRT) using rotational beams. During delivery, the field shape, formed by a multileaf collimator
`(MLC), changes constantly. The objectives of this study were to (1) clinically implement the IMAT technique, and
`(2) evaluate the dosimetry in comparison with conventional three-dimensional (3D) conformal techniques.
`Methods and Materials: Forward planning with a commercial system (RenderPlan 3D, Precision Therapy
`International, Inc., Norcross, GA) was used for IMAT planning. Arcs were approximated as multiple shaped
`fields spaced every 5–10° around the patient. The number and ranges of the arcs were chosen manually. Multiple
`coplanar, superimposing arcs or noncoplanar arcs with or without a wedge were allowed. For comparison,
`conventional 3D conformal treatment plans were generated with the same commercial forward planning system
`as for IMAT. Intensity-modulated treatment plans were also created with a commercial inverse planning system
`(CORVUS, Nomos Corporation). A leaf-sequencing program was developed to generate the dynamic MLC
`prescriptions. IMAT treatment delivery was accomplished by programming the linear accelerator (linac) to
`deliver an arc and the MLC to step through a sequence of fields. Both gantry rotation and leaf motion were
`enslaved to the delivered MUs. Dosimetric accuracy of the entire process was verified with phantoms before
`IMAT was used clinically. For each IMAT treatment, a dry run was performed to assess the geometric and
`dosimetric accuracy. Both the central axis dose and dose distributions were measured and compared with
`predictions by the planning system.
`Results: By the end of May 2001, 50 patients had completed their treatments with the IMAT technique. Two to
`five arcs were needed to achieve highly conformal dose distributions. The IMAT plans provided better dose
`uniformity in the target and lower doses to normal structures than 3D conformal plans. The results varied when
`the comparison was made with fixed gantry IMRT. In general, IMAT plans provided more uniform dose
`distributions in the target, whereas the inverse-planned fixed gantry treatments had greater flexibility in
`controlling dose to the critical structures. Because the field sizes and shapes used in the IMAT were similar to
`those used in conventional treatments, the dosimetric uncertainty was very small. Of the first 32 patients treated,
`the average difference between the measured and predicted doses was ⴚ0.54 ⴞ 1.72% at isocenter. The
`80%–95% isodose contours measured with film dosimetry matched those predicted by the planning system to
`within 2 mm. The planning time for IMAT was slightly longer than for generating conventional 3D conformal
`plans. However, because of the need to create phantom plans for the dry run, the overall planning time was
`doubled. The average time a patient spent on the table for IMAT treatment was similar to conventional
`treatments.
`Conclusion: Initial results demonstrated the feasibility and accuracy of IMAT for achieving highly conformal
`dose distributions for different sites. If treatment plans can be optimized for IMAT cone beam delivery, we expect
`IMAT to achieve dose distributions that rival both slice-based and fixed-field IMRT techniques. The efficient
`delivery with existing linac and MLC makes IMAT a practical choice. © 2002 Elsevier Science Inc.
`
`Intensity-modulated radiation therapy (IMRT), Dynamic radiation therapy, Intensity modulation, Arc therapy.
`
`INTRODUCTION
`
`Stemming from the increasing evidence that improved local
`tumor control may enhance long-term survival (1, 2) and
`reduce the cost of cancer treatments (3), intensity-modu-
`
`Reprint requests to: Cedric X. Yu, D.Sc., Department of Radi-
`ation Oncology, University of Maryland School of Medicine, 22 S.
`Greene Street, Baltimore, MD 21201. Tel: (410) 328-0324; Fax:
`(410) 328-2618; E-mail: Cyu002@umaryland.edu
`This work was
`supported in part by NIH Grant No.
`R29CA66075. The authors appreciate the support and technical
`
`lated radiation therapy (IMRT) is receiving increasing in-
`terest and acceptance in radiation oncology. Presently, sev-
`eral IMRT techniques have been proposed. One method is
`to use multiple coplanar and noncoplanar beams at different
`orientations, each beam having spatially modulated intensi-
`
`assistance from Elekta Oncology Systems, Inc.
`Acknowledgment—The authors would like to thank Dr. Matt Earl
`and Dr. Lance Weems for their contributions.
`Received Mar 2, 2001, and in revised form Feb 8, 2002. Ac-
`cepted for publication Feb 15, 2002.
`
`Varian (Ex. 1007)
`IPR of U.S. Pat. No. 7,961,843
`
`453
`
`

`

`454
`
`I. J. Radiation Oncology ● Biology ● Physics
`
`Volume 53, Number 2, 2002
`
`ties (4–15, 19). Another approach, referred to as tomo-
`therapy, delivers the treatment in multiple slices, with each
`slice of the target volume treated with temporally modulated
`fan beams rotating around the patient (16, 17). Each of these
`two methods has its advantages and disadvantages in dose
`conformity and in efficiency of dose delivery, as discussed
`by Webb (18), Brahme (19), and Yu (20). In general, the
`tomotherapy approach spreads the normal tissue dose over a
`greater volume and produces a tighter dose conformation to
`the target.
`We have implemented a new technique, intensity-modu-
`lated arc therapy, or IMAT, to deliver highly conformal
`dose distributions by combining gantry rotation and dy-
`namic multileaf collimation. Instead of delivering intensity-
`modulated beams with fixed gantry angles, IMAT delivers
`optimized dose distributions by rotating the radiation beam
`around the patient. During delivery, the field shape, which is
`formed by a multileaf collimator (MLC), changes continu-
`ously as determined by the treatment plan. Intensity distri-
`butions at all angles around the patient are achieved with
`multiple overlapping arcs, with each arc having a different
`set of field apertures. The weight of the arcs, or total MUs
`delivered in different arcs, are typically different. Therefore,
`IMAT is also different from tomotherapy (16, 17), which
`uses intensity-modulated fan beams rotating around the
`patient, delivering the treatment slice by slice. As with
`tomotherapy, IMAT combines intensity modulation and ro-
`tational delivery. A detailed description of the technique
`was reported by Yu in a previous article (20).
`A Phase I clinical trial using the dynamic MLC and
`rotational delivery technique was approved by the institu-
`tional review board to assess the feasibility and safety of the
`technique. From November 1999 to May 2001, 50 patients
`with cancers of the central nervous system, head and neck,
`and prostate were treated in our clinic using the IMAT
`technique. This article describes the issues in the implemen-
`tation and clinical usage of this technique. Clinical exam-
`ples of IMAT treatments will be presented to illustrate the
`dosimetric advantages of rotational delivery.
`
`METHODS AND MATERIALS
`As with conventional treatment techniques, IMAT in-
`volves treatment planning and delivery. It has been demon-
`strated that
`treatment plans developed for tomotherapy
`treatment delivery can be converted into multiple arcs and
`delivered with IMAT (20). The same inverse treatment
`planning system has also been adapted to MLC delivery
`(CORVUS, NOMOS Corp., Sewickley, PA) (17). However,
`because the treatment plans are optimized with a simulated
`annealing algorithm with little constraint on smoothness of
`the beam intensities, the beam intensity distributions are
`overly modulated. Because we approximate an arc delivery
`as equally spaced beams, the number of beams is usually
`large. Intuitively, as the number of beams increases, the
`degree of intensity modulation required to meet the dosi-
`metric constraints should be reduced. With simulated an-
`
`nealing algorithm and without any constraint on the smooth-
`ness of the intensity maps, however, the result is just the
`opposite. More beams generally increase the randomness of
`the intensity patterns. As a result, a plan with two or three
`intensity levels would typically require more than 10 arcs to
`deliver.
`To overcome such inefficiency, and as the first step in
`using rotational delivery with dynamic MLC, we imple-
`mented IMAT into clinical use with forward planning. From
`simulation CT images, the target and surrounding normal
`structures are delineated on a commercial three-dimensional
`(3D) treatment planning system (RenderPlan 3D, Precision
`Therapy, Inc., Norcross, GA). Arcs are approximated as
`multiple shaped fields spaced every 5–10° around the pa-
`tient. The ranges of the arcs are chosen manually to give the
`desired dose distributions. Multiple coplanar or noncoplanar
`arcs are allowed. Wedges are often used in combination
`with dynamic field shaping to achieve a more uniform dose
`distribution in the planning target volume. At each beam
`angle,
`irregular field shapes are defined based on the
`beam’s-eye–view (BEV) of the planning target and normal
`critical structures. Depending on the normal structure toler-
`ance, the regions in the BEV where the projection of the
`target and the normal critical structure overlap may be
`blocked at some or all beam angles. When such overlap
`region is in the center of the BEV of the target and blocking
`is desired, the MLC-shaped fields cover only the part of the
`target on one side of the critical structure. The other side
`will be irradiated with another arc. Superimposing arcs are
`often used. For example, one arc may cover the BEV of the
`target, including the region where the projections of the
`target and critical structure overlap, and a second overlap-
`ping arc that excludes the overlap region may be used to
`provide the required sparing for the critical structure. Typ-
`ically, two to five arcs spanning an angular range 40–180°
`are used. For dose calculation, each arc is approximated
`with fixed beams equally spaced at 10° intervals. The MLC
`field shapes of these fixed beams are arranged in the order
`of delivery to form the MLC leaf sequence. To keep the
`gantry speed constant for smooth delivery, the weights of
`the beams, i.e., the relative contributions of different beams
`to the dose prescription point, are determined for each arc
`such that each beam angle delivers the same number of
`MUs. This automatically allows the beams with shallower
`radiologic depth to the prescription point to have greater
`dose contributions. For most of the treatment plans, the
`weights of different arcs are adjusted manually to achieve
`acceptable target uniformity and critical structure sparing.
`Once a satisfactory dose distribution is generated, the plan
`is analyzed, as with conventional 3D conformal plans.
`For all patients intended to receive IMAT treatment, a
`conventional 3D conformal plan was independently gener-
`ated by a different planner. Comparisons of dose distribu-
`tions and dose–volume histograms (DVHs) were made by
`the physician. The IMRT technique was used only when the
`physician determined that there was an advantage of IMAT
`treatment over 3D conformal treatment. This comparison
`
`

`

`Clinical implementation of IMAT ● C. X. YU et al.
`
`455
`
`also allowed us to gain experience in the types of cases that
`were more suitable for rotational delivery. Once IMAT
`technique was chosen over conventional 3D conformal
`technique, the plan was read by a leaf-sequencer developed
`at our institution. Because the plan already contained the
`field shapes at all angles, the leaf sequencer simply con-
`verted the shapes into MLC field segments. Because of the
`ways the field shapes at each beam angle were determined,
`the field shapes at neighboring angles typically did not
`differ significantly. As a result, MLC leaves were not re-
`quired to travel large distances from one angle to the next.
`For most cases, gantry rotation speed, rather than leaf trav-
`eling speed, was the factor limiting the dose rate. The MLC
`prescriptions generated by the leaf sequencer were then sent
`to the MLC controller for dynamic delivery through a local
`network link.
`The IMAT delivery is implemented on an MLC system
`equipped on a digitally controlled linear accelerator (SL-20
`linear accelerator with MLCi, Elekta Oncology Systems,
`Inc., Norcross, GA) (21). It consists of 40 pairs of opposing
`leaves, each free to move along its length and projecting 1
`cm in width in the isocenter plane at 100 cm from the
`source. Complementary to the 80 leaves are two pairs of
`backup diaphragms (solid tungsten jaws) in the x and y
`directions, respectively. Both the leaves and the backup
`diaphragms are used for defining the dynamic MLC seg-
`ments. During beam delivery, the linac is programmed to
`deliver arc treatments, and the MLC is programmed to
`dynamically step through a sequence of field shapes. Both
`gantry rotation and leaf motion are coupled to the delivered
`MUs. As a result, although fluctuations in machine dose rate
`can cause the gantry to rotate with changing speed, the
`effect on dose delivery is minimal. It is important to under-
`stand that the field shapes are only defined at a set of beam
`angles spaced 10° apart. In between two successive beam
`angles, the MLC controller linearly interpolates the leaf
`positions. Therefore, the leaves are moving continuously
`throughout the delivery, unless the leaf positions of two
`successive segments are the same.
`To verify the clinical value of such a simplified process,
`we compared the treatment plans using two to five forward
`planned arcs with the plans generated by conventional tech-
`niques and by a commercial inverse planning system (COR-
`VUS by NOMOS Corp., Inc., Sewickley, PA) for treatment
`of head-and-neck cancers, central nervous system tumors,
`and the prostate. Both the conventional plans and the IMAT
`plans are presented to the physician.
`An anthropomorphic phantom (Alderson Rando Phan-
`tom, Alderson Research Laboratories, Inc., Stamford, CT)
`was modified for dosimetric verification. Multiple original
`slices of the phantom at different sites were replaced with
`two sheets of water-equivalent plastic material of the same
`shape, each with half the original slice thickness. Grooves
`were made on each of the two plastic sheets, so that an ion
`chamber could be placed at various positions along the
`horizontal axis. Radiographic film (XV-2, Kodak, Roches-
`ter, NY) cut to the exact shape of the plastic sheets could
`
`also be placed between the two plastic sheets without the
`chamber groove for relative dose measurement. The modi-
`fied phantom was scanned on our CT-simulator unit, and the
`images were imported to the planning system. The use of
`the anthropomorphic phantom for the dry run allowed us not
`only to verify the absolute dose quantitatively, but also to
`make a visual comparison of the dose pattern to the patient
`plan and to identify setup problems or difficulties before
`treatment.
`Dosimetric accuracy of the entire process was verified
`with phantoms before IMAT was used clinically. Shaped
`fields ranging from 4 cm ⫻ 4 cm to 30 cm ⫻ 30 cm spaced
`5–20° were used for approximating an arc. The sequence of
`shaped fields with drastically changing field shapes was
`delivered to a cubical phantom both individually as calcu-
`lations were carried out in the plan and in arc fashion.
`When approximating an arc with multiple fixed fields,
`each field was essentially a sample within the range of the
`arc. Each sample should be treated as though it were at the
`center of the interval. That is, if a 10° interval is used, an arc
`should start 5° ahead of the first field and end 5° beyond the
`last field. Because the field shapes were not defined beyond
`the angles at both ends of the arc, we chose not to have the
`arc extending beyond the angles of the first and last fields.
`To keep the plan and delivery consistent, we set the MUs of
`the first beam and the last beam to be one-half of those of
`the other beams within the same arc.
`Because the planning system was not designed for IMAT
`planning, the MUs provided by the planning system had to
`be adjusted to achieve accurate delivery. Although the pri-
`mary arc commonly used large fields, the overlapping arcs
`were generally small and possibly off axis. A dual source
`model that accounted for the 3D geometry of the collimat-
`ing system (22) was used to calculate the head scatter. The
`ratio of our calculation result to the value predicted by the
`planning system was used to adjust the total MUs. For
`IMAT treatment, a plan might include more than 50 beams.
`The number of MUs for the beams in the overlapping arcs
`could be very small, in the range 3–5 MUs per beam.
`Because no fractional MU is allowed in the planning sys-
`tem, the rounding error for each beam could cause an arc to
`give a dose contribution that differs from the intended value.
`The result would be a total dose to the prescription point
`that deviates from the prescription dose by a small percent-
`age (1%–2%). To correct for such a rounding error, the total
`MU of an arc was adjusted also by the ratio of the intended
`contribution to the contribution used by the plan.
`For all IMAT treatments, a dry run was conducted to
`assess the geometric and dosimetric accuracy and to elim-
`inate possible technical problems, such as setup difficulties
`and MLC movement constraints. The dry run was per-
`formed by copying the patient plan parameters, including
`field shapes and MUs, to the same site of the modified
`humanoid phantom. When it was required to measure dose
`distributions in sagittal planes, common rectangular phan-
`toms consisting of water-equivalent plastic slabs were also
`used. Both the central axis dose and the dose distribution on
`
`

`

`456
`
`I. J. Radiation Oncology ● Biology ● Physics
`
`Volume 53, Number 2, 2002
`
`a plane were measured and compared to those generated by
`the planning system. The accelerator output variation at the
`time of verification was factored out by either performing an
`output reading with the same ion chamber or by using the
`output reading obtained in daily quality assurance. Because
`the speed of leaf travel is limited, it is important that the
`field shapes of adjacent beam angles not differ too much.
`For most clinical cases, the field shapes varied slowly be-
`tween angles. Treatments could be delivered with the high-
`est machine dose rate determined by the linac based on the
`maximum speed of gantry rotation, resulting in very short
`delivery time. There were cases where large leaf travel was
`required, and the leaf motion lagged radiation delivery.
`Once the leaf was behind the desired position for the deliv-
`ered MU by a preset amount, 1 mm in our case, either
`radiation pause, which could recover automatically if leaf
`reaches position within a second, or termination, which
`could not recover by itself, would occur. In such cases, a
`reduced nominal accelerator dose rate setting was used.
`Although treatment could be resumed after an interruption,
`such radiation pause or termination could increase delivery
`time. Although we could predict such occurrences and se-
`lect a dose rate by using the maximum leaf speed, the total
`MU, and the fastest gantry speed, we did not do so in the
`trial because of the rarity of such occurrence and because of
`the use of dry runs for dosimetric verification. If the need for
`a reduced machine dose rate was observed during the dry
`run quality assurance procedure, the right machine dose rate
`setting would be selected for treatment delivery.
`Detailed linac prescriptions for treatment were entered in
`the linac control after the dry run. The linac was pro-
`grammed to deliver an arc treatment with or without
`wedges, and the MLC was programmed to step through a
`sequence of field shapes. Each arc was treated as a separate
`beam and delivered separately.
`
`RESULTS
`Between November 1999 and May 2001, a total of 50
`patients were treated using the IMAT technique. Of the 50
`patients, 13 had cancers of the central nervous system, 16
`had prostate cancer, two had thoracic cancer, three had
`gastrointestinal cancers, and 16 patients had head-and-neck
`cancers. For 31 of the 50 patients, IMAT was used for the
`final boost, with the total number of fractions ranging from
`8 to 12. For 19 of the 50 patients, IMAT was used for the
`entire course of treatment.
`For complex cases, the time needed to create a satisfac-
`tory treatment plan was found to be longer than for conven-
`tional 3D conformal plans. This is largely because of the
`number of fields that the planner must specify and outline
`for IMAT treatment plans. For cases where a precalculated
`treatment plan template could be used, such as for the
`treatment of prostate cancer, the planning times were similar
`to those for conventional planning. However, because of the
`need to create phantom plans for the dry run and 3D
`conformal plans for comparison, the overall planning time
`
`tripled. The dry run quality assurance procedure also takes
`an additional 1–2 h per course of treatment. We have been
`continually modifying our quality assurance procedures to
`speed up the dry runs. Alternative quality assurance proce-
`dures, such as the use of electronic portal imaging systems,
`are being investigated.
`To determine the acceptable spacing of fields used for
`approximating an arc, we performed measurements using
`different field shapes and different angular spacing between
`fields. It was found that spacing of the fields from 5° to 20°
`did not change the central axis dose or the target dose
`coverage for the same total MUs. However, dose distribu-
`tions outside the target, especially at low isodose levels near
`the surface, differed between calculations with fixed fields
`and those delivered with arc beams as angular spacing
`increased. Figures 1a–d illustrate planning results using a
`fixed field width of 5 cm but different angular spacing to
`approximate a 150° arc. In all four figures, the isodose
`levels from the center outward are 95%, 80%, 50%, 30%,
`20%, and 10%, respectively. Figure 1a is with fields spaced
`every 3°, which most closely approximates a continuous arc
`delivery. All isodose lines from 10% to 95% are smooth, as
`expected in an arc delivery. Figures 1b–d are dose distri-
`butions with fields spaced 5°, 10°, and 15°, respectively. As
`the angular spacing increased, lower isodose levels started
`to show ripples. However, for all angular spacing used, the
`isodose lines of 80% and 95% remained the same. The
`rippling appearance on isodose lines near the surface can be
`explained by the gaps in geometric overlap of the fixed
`beams. We found that a spacing of 10° represents a good
`compromise. If the target is small and the lower dose areas
`coincide with critical structures, a finer angular spacing of
`5° should be used.
`Preclinical dose verifications were conducted, starting
`with simple spherical targets in a water-equivalent cubic
`phantom. Plans were generated to test the dosimetric accu-
`racy of the entire process from planning to delivery. For
`these simple and well-controlled cases, we expected perfect
`agreement between the calculations and measurements. It
`was quickly realized that the treatment planning system did
`not properly model the MLC for head scatter. Because the
`MLC, which replaces the upper jaw of the secondary col-
`limator, is used for shaping the fields, the head scatter
`should be determined based on the irregular MLC-shaped
`fields. However, the treatment planning system uses the
`rectangle circumscribing the irregular field to estimate the
`head scatter factor. For small field sizes, an overestimation
`of the equivalent field size by 2 cm can cause 2%–3%
`errors. To correct for the modeling deficiency of the plan-
`ning system, we used a dual source model with consider-
`ation of leaf thickness and shape (22) to calculate the head
`scatter factors of all the fields approximating an arc. The
`total MU of the arc was then adjusted upward by the ratio of
`the head scatter factors predicted by the planning system to
`that obtained with our dual source model. After the correc-
`tions were made for all field shapes, the agreements between
`plan calculation and measurements were all within 1% for
`
`

`

`Clinical implementation of IMAT ● C. X. YU et al.
`
`457
`
`Fig. 1. An illustration of the effect of angular spacing on the accuracy of using multiple fixed fields to approximate an
`arc delivery. The dose distributions show 95%, 80%, 50%, 30%, 20%, and 10% levels from the innermost to the
`outermost isodose lines. (a–d) The isodose distributions shown were obtained with the angular spacing of 3°, 5°, 10°,
`and 15°, respectively.
`
`the simple test cases. For patient-specific verifications, all
`absolute dose measurements were found to be within ⫾3%
`of the calculated values, except for one case, where one of
`the arcs was at sharp angles to the stem of the ion chamber,
`and the effect was not corrected. Figure 2 shows the scat-
`tered plot of the discrepancies between the plan predicted
`
`and the absolute dose measurements for the first 32 patients
`treated with the IMAT technique. The quantities are ex-
`pressed as (Measured ⫺ Predicted)/Measured ⫻ 100%. The
`mean error was found to be ⫺0.54%, and the standard
`deviation was 1.72%.
`Isodose distributions were also measured with films. Fig-
`
`

`

`458
`
`I. J. Radiation Oncology ● Biology ● Physics
`
`Volume 53, Number 2, 2002
`
`ures 3a–d show a clinical example of the planned and
`measured isodose distributions for the treatment of an
`ependymoma. Figure 3a is the dose distribution predicted
`by the planning system in the patient. Unlike the traditional
`treatment with anterior-posterior setup, IMAT uses two
`wedged posterior arcs and one anterior arc overlapped with
`two fixed fields. By spreading lower doses to a greater
`volume, the 90% isodose covers the tumor uniformly and
`the dose dropoff from 90% to 80% within 4 mm, except at
`the superior and inferior ends. The plan is applied to a
`plastic phantom for the dry run. Figure 3b shows the isodose
`distribution calculated for the phantom by applying the
`same beam arrangements and weightings as for the patient
`plan. It shows a dosimetric pattern similar to that of the
`patient plan. The white area is encompassed by the 90%
`isodose line. Figure 3c is the grayscale image of the film
`used for relative isodose measurements. The film was ex-
`posed by sandwiching it between two slices of the plastic
`phantom and delivering the same treatment as for the patient
`with scaled-down total MUs. The grayscale image for the
`film was converted to dose with a precalibrated relationship.
`
`Fig. 2. Absolute dose discrepancies between plan predictions and
`measurements for the first 32 patients. The quantities are expressed
`in (Measured ⫺ Predicted)/Measured ⫻ 100%.
`
`Fig. 3. A clinical example of the planned and measured dose distributions for the treatment of an ependymoma showing
`(a) dose distribution predicted by the planning system in the patient, (b) dose distribution calculated for the phantom by
`applying the same beam arrangements and weightings as for the patient plan, (c) the inverted grayscale image of the film
`used for relative isodose measurements, and (d) resultant measured dose distribution after converting grayscales to dose.
`
`

`

`Clinical implementation of IMAT ● C. X. YU et al.
`
`459
`
`Fig. 4. An example comparing (a) the conventional 3D plan, (b) the IMAT plan, and (c) the inverse plan generated with
`the CORVUS system for an adenoid cystic carcinoma of the maxillary sinus. The DVHs generated by the three plans
`are shown in (d), where the solid lines are for the IMAT plan, the dashed lines are for the 3D plan, and the dotted lines
`are for the inverse plan. DVHs for four structures, including the PTV, the brainstem, and the left and right eyes, are
`compared.
`
`The resultant measured isodose distribution is shown on
`Fig. 3d). As compared with what was predicted by the
`planning system (Fig. 3b), for isodose lines at levels higher
`than 80%, the agreement with calculation is within 2 mm.
`Because of the simplifications in scatter dose calculation
`used by the treatment planning system, the spacing of the
`isodose lines from 80% to 100% is consistently tighter than
`that measured with film. For lower dose levels, such as
`those at or below the 50% level, which typically lie in the
`shallow gradient region with rotational delivery, the dis-
`agreement can be larger than 5 mm in some areas. This is
`caused by both the simplified scatter dose calculation algo-
`rithms and the intrinsic inaccuracy of film dosimetry for
`photon beams. We set an acceptance criterion that
`the
`predicted and measured isodose lines at levels greater than
`
`90% in the phantom must match within 2 mm. When doses
`to critical structures needed to be verified more accurately,
`additional point dose measurements had to be carried out.
`An example comparing the IMAT plan, the inverse plan
`by the CORVUS system, and the conventional 3D plan for
`an adenoid cystic carcinoma of the maxillary sinus is shown
`in Fig. 4. Isodose distributions are shown on the transverse
`and sagittal planes with the regions covered by the 95%
`isodose contour highlighted (darker). Figure 4a is the con-
`ventional plan with five fields, including anterior, left lateral
`and posterior, and vertex fields. Wedges were used in all the
`fields. Figure 4b is the IMAT plan with five coplanar arcs.
`They include a primary arc from 355° to 175°, two over-
`lapping arcs from 45° to 95°, one overlapping arc from 115°
`to 175°, and a primary arc from 225° to 275°. In the primary
`
`

`

`460
`
`I. J. Radiation Oncology ● Biology ● Physics
`
`Volume 53, Number 2, 2002
`
`Fig. 4. (Cont’d)
`
`arcs, the radiation field was shaped as the beam’s-eye–view
`of the target, excluding the beam’s-eye–view of the left eye.
`In all other arcs, the field shapes excluded the beam’s-eye–
`view of both eyes and brainstem, as well as areas in the
`target that already received a hot spot from the primary arcs.
`Figure 4c shows the inverse plan with seven coplanar in-
`tensity-modulated fields, each with 20 intensity levels.
`The DVHs for four structures, including the planning
`target volume (PTV), the brainstem, and the left and right
`eyes, generated by the three plans are shown as Fig. 4d. The
`solid lines represent results of the IMAT plan. The dashed
`lines represent results of the conventional 3D plan. The
`dotted lines are the results of the inverse plan by the COR-
`VUS system. All three plans showed similar dose coverage
`
`to the PTV, with the inverse plan showing a slightly more
`uniform dose. The inverse plan showed an overall lower
`dose to the brainstem but a maximum dose slightly higher
`than that found in the other two plans, because of a slight
`increase in the dose to both eyes. The 3D plan was not able
`to spare the left eye. Both the IMAT plan and the inverse
`plan were clinically acceptable. The IMAT plan was chosen
`because the delivery of five coplanar arcs was much less
`time-consuming than the delivery of seven intensity-modu-
`lated fields with our mixed dynamic and step-and-shoot
`sliding window technique.
`Figure 5 shows the same type of comparison of DVHs as
`in Fig. 4d for a prostate cancer treatment. For the IMAT
`plan and the 3D conformal plan, a 1-cm margin was used
`
`

`

`Clinical implementation of IMAT ● C. X. YU et al.
`
`461
`
`Fig. 5. Comparison of DVHs generated by conventional 3D plan (thin solid lines), the IMAT plan (thick solid lines),
`and the inverse plan by CORVUS (dashed lines) for the treatment of prostate cancer. Two structures, the GTV and
`rectum, are compared.
`
`around the gross tumor volume to derive the planning
`target volume. In the inverse plan using a commercial
`inverse planning system (CORVUS, NOMOS Corp.,
`Sewickley, PA), a 1-cm margin was assigned at
`the
`“prescription” stage, so the optimizer could get better
`target coverage. The IMAT treatment used four arcs. In
`two of the four arcs, an internal 60° wedge was used,
`together with dynamic MLC field shaping. At each gantry
`angle, the field covers only part of the planning target
`BEV on one side of the rectum. Therefore, the rectum
`was shielded in these two arcs. The other two arcs use the
`MLC-shaped irregular fields to cover the BEV of the
`entire planning target at all beam angles. The weighting
`between the partial target BEV arcs with wedge and the
`open arcs is 1:4. By blocking the BEV projection of the
`critical organ during beam rotation, it is

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket