throbber
Communication and sampling rate limitations in IMRT delivery
`with a dynamic multileaf collimator system
`Ping Xia,a) Cynthia F. Chuang, and Lynn J. Verhey
`The Department of Radiation Oncology, University of California at San Francisco, San Francisco,
`California 94143
`~Received 13 July 2001; accepted for publication 4 December 2001; published 21 February 2002!
`The delivery of an intensity modulated radiation field with a dynamic multileaf collimator ~MLC!
`requires precise correlation between MLC positions and cumulative monitor units ~MUs!. The
`purpose of this study is to investigate the precision of this correlation as a function of delivered
`MUs and dose rate. A semi-Gaussian shaped intensity profile and a simple geometric intensity
`pattern consisting of four square segments were designed to deliver a total of 1, 4, 16, 64, and 100
`MUs at three different dose rates of 100, 400, and 600 MU/min. The semi-Gaussian intensity
`pattern was delivered using both sliding window and step and shoot techniques. The dose profiles
`of this intensity pattern were measured with films. The four square intensity pattern was delivered
`using step and shoot and conventional delivery techniques for comparison. Because of geometrical
`symmetry, the dose to each segment in this intensity pattern is expected to be the same when the
`same MU is assigned to each segment. An ionization chamber was used to measure the dose in the
`center of each of the four square segments. For the semi-Gaussian shaped profile, significant
`artifacts were observed when the profile was delivered with small MUs and/or at a high dose rate.
`For the four square intensity pattern, the dose measured in each segment presented a large variation
`when delivered with small MUs and a high dose rate. The variation increases as the MU/segment
`decreases and as the dose rate increases. These MU and dose rate dependencies were not observed
`when the intensity pattern was delivered using a conventional delivery technique. The observed
`distortion of the semi-Gaussian profile and dose variations among the segments of the four square
`intensity pattern are explained by considering the sampling rate and the communication time lag
`between the control systems. Finally, clinical significance is discussed. © 2002 American Asso-
`@DOI: 10.1118/1.1449496#
`ciation of Physicists in Medicine.
`
`Key words: intensity modulated radiotherapy, dynamic multileaf collimator, step and shoot
`
`I. INTRODUCTION
`
`Computer controlled multileaf collimator ~MLC! systems
`have made intensity modulated radiation therapy ~IMRT!
`clinically practical, using either dynamic or static delivery
`techniques.1–10 The fundamental difference between these
`two delivery methodologies is that with dynamic delivery,
`the radiation and the MLC leaf motions can be executed
`simultaneously,2–5,7 whereas with static delivery, the radia-
`tion and leaf motions are executed sequentially.6,8 –10 The lat-
`ter delivery method resembles conventional delivery, except
`that many segments are included in each given field. Due to
`the use of many small sized segments with associated small
`monitor units ~MUs!, the dose accuracy of static delivery can
`be affected by the accuracy of leaf positioning, and by dose
`nonlinearity for small MU delivery.
`In dynamic delivery, the key factors that affect dose ac-
`curacy include the accuracy of the leaf positions and the
`correlation between the leaf positions and the accumulated
`dose, similar to the situation of the dynamic wedge.11 Unlike
`the dynamic wedge, in which only one pair of jaws is used,
`the dynamic MLC-IMRT delivery employs many pairs of
`MLC leaves, and each has a different intensity profile. In
`order to let all leaves move to their designated positions,
`both leaf speed modulation and dose rate modulation are
`
`needed for implementation of dynamic MLC-IMRT delivery.
`In leaf speed modulation, the leaf speed for each pair of
`MLC leaves is different, but constant within each segment.
`In dose rate modulation, a maximum dose rate is used when-
`ever it is possible to achieve efficient delivery, but when the
`required leaf speed exceeds the maximum mechanical leaf
`speed, the dose rate is reduced.
`A special dynamic MLC delivery technique is considered
`similar to the static MLC delivery. In this delivery method,
`within each segment, the MLC leaf speed is zero, and the
`leaf speed is set to infinity between two segments, thus forc-
`ing the dose rate to be zero, i.e., the beam is off. It should be
`noted that this special dynamic MLC delivery is different
`from static MLC delivery, because in this special dynamic
`MLC delivery, the radiation and the leaf motion are still
`correlated. Despite their difference, both static MLC-IMRT
`delivery and special dynamic MLC-IMRT delivery are often
`referred to as step and shoot delivery. To distinguish these
`two delivery methods, we call the static MLC-IMRT delivery
`method mechanism I step and shoot delivery, in which each
`segment is considered as an individual field. The special dy-
`namic MLC-IMRT delivery method is called mechanism II
`step and shoot delivery, in which the beam off/on command
`is sent from the MLC control station to the machine console
`
`412
`
`Med. Phys. 29 (cid:132)3(cid:133), March 2002
`
`0094-2405(cid:213)2002(cid:213)29(cid:132)3(cid:133)(cid:213)412(cid:213)12(cid:213)$19.00
`
`© 2002 Am. Assoc. Phys. Med.
`
`412
`
`Varian (Ex. 1020)
`IPR of U.S. Pat. No. 7,961,843
`
`

`

`413
`
`Xia, Chuang, and Verhey: Communication and sampling rate limitations
`
`413
`
`FIG. 1. A semi-Gaussian shaped intensity pattern, consisting of five intensity
`levels.
`
`through a periodic comparison between the programmed MU
`to each segment and the cumulative MU controlled by the
`machine console.
`The purpose of this paper is to investigate how the preci-
`sion of the MU and the leaf position correlation varies as a
`function of the delivered MUs and dose rate in a commercial
`dynamic MLC system ~Varian Oncology System, Palo Alto,
`CA!.
`
`II. MATERIALS AND METHODS
`
`A. General description
`
`The multileaf collimator in Varian’s linear accelerator is a
`single-focused MLC,12,13 in which a MLC field follows the
`beam divergence along the direction perpendicular to the leaf
`motion, but not along the direction of the leaf motion. In
`other words, the leaves move along straight lines in a plane
`perpendicular to the central axis of the beam. This design
`simplifies the mechanics of the MLC system, but may cause
`variations in the width of the penumbra when leaves move to
`different locations.12,14 A rounded leaf end is used in this
`MLC system to minimize this effect.15 Due to the rounded
`leaf end, however, the leakage between two leaves when they
`are closed is significant.13 In conventional treatment, these
`closed leaf pairs are normally shielded under the primary and
`the secondary jaws to reduce the leakage between two leaf
`ends. In addition, to minimize leakage radiation between two
`adjacent leaves, a tongue and groove arrangement is used.
`
`B. MLC control system
`
`The MLC control system controls the movement of each
`MLC leaf, including verifying the correspondence of each
`leaf position with its programmed position stored in an
`ASCII file, referred to as a MLC file. The control system can
`be operated either in a static mode ~for a conventional MLC
`field! or a dose mode ~for an intensity modulated MLC field!.
`In the static mode, there is only one position for each leaf for
`a given field. Once all leaves have moved to their pro-
`grammed positions ~within given tolerance!, the associated
`
`Medical Physics, Vol. 29, No. 3, March 2002
`
`FIG. 2. A simple intensity pattern consisting of four 434 cm2 square seg-
`ments, located at an equal distance from the iso-center.
`
`MU is delivered. In the dose mode, there are serial MLC
`positions for a given field stored in a MLC file, in which the
`positions of each MLC leaf are described as a function of a
`dose index ranging from 0.0 to 1.0. The dose index is a
`fraction of the total monitor units for the entire intensity
`modulated ~IM! field.
`
`C. Dynamic MLC delivery
`
`In this system, the machine console and the MLC control
`station separately control the MU delivery and MLC posi-
`tions, respectively. In conventional delivery, since only one
`MLC shape is associated with the total MU, the radiation
`beam is turned on only when the machine console receives a
`‘‘ready’’ signal from the MLC control station. The commu-
`nication between the MLC station and the machine console
`is sequential. Any delay in this communication would have
`no effect on the accuracy of the dose delivered to the field. In
`dynamic delivery, however, the leaf positions and the accu-
`mulated MUs are correlated. This correlation is established
`by communication between the MLC station and the ma-
`chine console every 50 ms, independent of the complexity of
`the intensity profiles. In other words, the sampling frequency
`is fixed at 20 Hz.
`Experiment A: Semi-Gaussian profile. Two experiments
`were designed to investigate the dose accuracy in IMRT de-
`livery using the dynamic MLC. The first experiment was
`designed to study how a semi-Gaussian shaped intensity pro-
`file varies as a function of the dose rate and the delivered
`MU, using both the dynamic and mechanism II step and
`shoot delivery techniques. This semi-Gaussian shape pro-
`duced a simple dose profile and semicontinuous dose inten-
`sity modulation across the field. Therefore, it is particularly
`suitable for dynamic delivery. The intensity map consists of
`five nonzero intensity levels, ranging from 20%, 40%, 60%,
`80%, and 100% of the total MU, as shown in Fig. 1. This
`intensity map was manually input into the planning system
`~CORVUS 3.0, NOMOS Corp., Sewickley, PA!, using the beam
`utility module. The MLC delivery files were created by the
`CORVUS system using the static and dynamic modes for
`Varian linear accelerators, respectively. At dose rates of 100
`and 400 MU/min, total MUs of 1, 4, 16, 64, 100 MUs were
`delivered to the IM field. Films were used to measure the
`dose profiles. In order to avoid the saturation of optical den-
`
`

`

`414
`
`Xia, Chuang, and Verhey: Communication and sampling rate limitations
`
`414
`
`FIG. 3. ~a!–~e! Cross plane intensity profiles for the semi-Gaussian shaped
`IM field, delivered with mechanism II step and shoot method at 100 MU/
`min dose rate for total MUs of 1, 4, 16, 64, 100, respectively.
`
`sity, the enhanced contrast localization films ~EC-L, Kodak,
`Rochester, NY! were used in a cassette for the measurement
`of 1 MU delivery. XTL films were used for the measurement
`of 4 and 16 MU delivery, and XV films were used for the
`measurement of 64 and 100 MU delivery. All films were
`exposed at 1.5 cm depth, 100 cm SSD, except the top surface
`of the EC-L film cassette was set at this depth, and irradiated
`at 6 MV. All films were scanned with a film scanner
`~VXR-12, Vidar Systems Corp., VA, using the Wellhofer
`software—Wellhofer North American, Bartlett, TN!.
`
`Experiment B: Four square intensity pattern. The second
`experiment was designed to deliver an intensity modulated
`field consisting of four 434 cm2 segments located at an
`equal distance ~5.66 cm! from the isocenter, as shown in Fig.
`2. These four segments were first delivered with the mecha-
`nism I step and shoot method, in which the conventional
`delivery mode ~or static mode in the VARIS system, Varian
`Oncology System, Palo Alto, CA! was used, with the same
`MU to each segment. This pattern was also delivered with
`the mechanism II step and shoot method ~using the dose
`
`Medical Physics, Vol. 29, No. 3, March 2002
`
`

`

`415
`
`Xia, Chuang, and Verhey: Communication and sampling rate limitations
`
`415
`
`FIG. 4. ~a!–~e! Cross plane intensity profiles for the semi-Gaussian shaped
`IM field, delivered with mechanism II step and shoot method at 400 MU/
`min dose rate for total MUs of 1, 4, 16, 64, 100, respectively.
`
`mode in the VARIS system!, with the same dose index to
`each segment. The MLC file for the mechanism II step and
`shoot delivery was created manually, following instructions
`in the DMLC manual ~DMLC implementation Guide, Varian
`Oncology System, Palo Alto, CA!. The MLC file consisted
`of a total of eight fields to describe the four segments, since
`each segment requires two MLC fields for the mechanism II
`step and shoot delivery. A Varian Clinac 2300 C/D, 6 MV
`photon beam energy was used. An ionization chamber ~IC10,
`Wellhofer, Wellhofer North American, Bartlett TN! was used
`for measurements, located at 1.5 cm depth inside a solid
`
`water phantom with 100 cm source to surface distance. Be-
`cause of geometric symmetry, the dose delivered to each
`segment should be the same, regardless of the delivery
`method. The relative dose difference D i for each segment is
`defined as
`D i5~D i2D 0!/D 0 ,
`where D i is the dose measured in the ith segment and D 0 is
`the average dose of the four segments. This pattern was de-
`livered at three different dose rates of 100, 400 and 600
`MU/min with total MUs of 1, 4, 16, 64, and 100, respec-
`
`~1!
`
`Medical Physics, Vol. 29, No. 3, March 2002
`
`

`

`416
`
`Xia, Chuang, and Verhey: Communication and sampling rate limitations
`
`416
`
`FIG. 5. ~a!–~e! Cross plane intensity profiles for the semi-Gaussian shaped
`IM field, delivered with the dynamic delivery method at 100 MU/min dose
`rate for total MUs of 1, 4, 16, 64, 100, respectively.
`
`tively. Since there were four segments in this IM field and
`each segment was assigned with the same MU, the MU per
`segment in these measurements was 0.25, 1, 4, 16, and 25
`MU/seg, respectively.
`
`III. RESULTS
`Figures 3~a!–3~e! show cross plane profiles for the semi-
`Gaussian shaped IM field shown in Fig. 1, delivered with the
`mechanism II step and shoot method at 100 MU/min dose
`rate for total MUs of 1, 4, 16, 64, and 100, respectively. All
`profiles in Figs. 3~a!–3~e! were normalized to their maxi-
`
`mum intensities. Since these profiles were obtained from
`three different kinds of films with different sensitivities, only
`the shapes of these profiles are important. The scales of the
`relative intensities, therefore, are not shown in Figs. 3~a!–
`3~e!. Figures 4~a!– 4~e! are cross plane profiles for the same
`IM field as in Fig. 3, but delivered at 400 MU/min dose rate.
`In Figs. 3 and 4, severe distortions were observed in profiles
`delivered with small MUs, such as in Figs. 3~a! and 3~b!, and
`4~a! and 4~b!. The shorter the beam-on time, the more seri-
`ous the observed distortions were. In Fig. 3~a!, there is a
`somewhat semi-Gaussian shape in the profile, which was de-
`
`Medical Physics, Vol. 29, No. 3, March 2002
`
`

`

`417
`
`Xia, Chuang, and Verhey: Communication and sampling rate limitations
`
`417
`
`FIG. 6. ~a!–~e! Cross plane intensity profiles for the semi-Gaussian shaped
`IM field, delivered with the dynamic delivery method at 400 MU/min dose
`rate for total MUs of 1, 4, 16, 64, 100, respectively.
`
`livered with a lower dose rate of 100 MU/min, whereas the
`profile is strongly distorted, as shown in Fig. 4~a!, when it
`was delivered at the higher dose rate of 400 MU/min. When
`the total MU is increased to more than 16 MU as shown in
`Figs. 3~c!–3~e!, and Figs. 4~c!– 4~e!, no obvious distortions
`were observed in these profiles. Similarly, Figs. 5~a!–5~e!
`and 6~a!– 6~e! show cross plane profiles for the IM field de-
`livered with the dynamic mode at 100 and 400 MU/min,
`respectively, with the same total MUs as in Fig. 3. Again,
`
`severe distortions were observed in profiles delivered with
`short beam-on times, such as in Figs 5~a! and 6~a!.
`Figure 7 shows the relative dose variations among the
`four segments delivered with the mechanism I step and shoot
`method, with 1 MU/seg at three different dose rates of 100,
`400, and 600 MU/min, respectively. The relative dose varia-
`tions were calculated according to Eq. ~1!. The observed
`dose variations among the segments were about 1%–3%,
`almost independent of the dose rate. As mentioned in Sec. II,
`
`Medical Physics, Vol. 29, No. 3, March 2002
`
`

`

`418
`
`Xia, Chuang, and Verhey: Communication and sampling rate limitations
`
`418
`
`sampling theory, any arbitrary intensity profile can be recon-
`structed if the sampling frequency is at least twice as large as
`the maximum spatial variation frequency ~Nyquist condi-
`tion!. Some of the results of these two experiments can be
`explained by sampling theory. The sampling frequency used
`in this IMRT dynamic delivery system is fixed at 20 Hz. At
`this sampling rate, the total delivery MU and the dose rate
`determine the number of sampling points taken for each de-
`livery. Smaller MUs with a higher dose rate include fewer
`sampling points, resulting in the observed distortions, as
`shown clearly in Figs. 3~a!, 4~a!, 5~a!, and 6~a!. Table II
`shows the number of sampling points calculated in the deliv-
`ery of a total MU of 1, 4, 16, 64, and 100 MU at dose rates
`of 100, 400, and 600 MU/min, respectively. The formula
`used for this calculation is
`
`N560
`
`T MU
`R
`
`r,
`
`~2!
`
`where N is the number of sampling points, T MU is the total
`MU, R is the delivery dose rate in MU/min, and r is the
`sampling rate in s21. The coefficient 60 converts minutes
`T MU5100 MU,
`into
`seconds.
`For
`example,
`if
`R
`5100 MU/min, and r520 s21, then N51200.
`According to sampling theory, more complex intensity
`profiles require more frequent sampling to accurately repro-
`duce the intensity profile. In other words, to deliver complex
`intensity profiles accurately, larger MU and/or lower dose
`rates are necessary to have sufficient beam-on time to in-
`clude an adequate number of sampling points. It should be
`noted that the concept of sampling points is different from
`the concept of control points introduced in other papers16,17
`in studies of leaf sequencers or interpreters to deliver inten-
`sity modulated beams using dynamic MLC systems. To real-
`ize a continuous intensity profile using a dynamic MLC, the
`MLC control system requires the leaf setting at a series of
`control points. These control points approximate a continu-
`ous intensity profile with a discrete profile. In step and shoot
`IMRT delivery, an intensity map is decomposed into multiple
`segments, and two control points define a segment, espe-
`cially with mechanism II step and shoot delivery. It is rea-
`sonable to argue that for dynamic or mechanism II step and
`shoot delivery, the number of sampling points should at least
`be equal to that of the control points, provided that the sam-
`pling points are synchronized with the control points.
`The semi-Gaussian shaped IM field consists of 9 seg-
`ments for step and shoot delivery and 18 control points for
`dynamic delivery. As shown in Table II, if the IM field is
`delivered with a total of 1 or 4 MU at 400 MU/min dose rate,
`and 1 MU at 100 MU/min dose rate, the included number of
`the sampling points in these deliveries is less than the num-
`ber of control points or segment defining points. With a total
`of 4 MU delivered at 100 MU/min dose rate, even though the
`number of the sampling points included is 48, greater than
`the control points or segment defining points, a small distor-
`tion at the tail of the profile is still observed.
`For the simple IM field shown in Fig. 2, at least 8 sam-
`pling points are needed since it consists of four segments. 8
`
`FIG. 7. The relative dose variations among the four square segments deliv-
`ered with mechanism I step and shoot method, with 1 MU /seg at three
`different dose rates of 100, 400, and 600 MU/min.
`
`the doses to these segments should be identical, but the im-
`perfect field flatness and symmetry and other measurement
`uncertainties may contribute to this small 1%–3% dose
`variation.
`Figures 8~a!– 8~e! show the dose variations of each seg-
`ment for the same IM field ~Fig. 2! delivered with the
`mechanism II step and shoot method, with a total of 1, 4, 16,
`64, and 100 MUs at dose rates of 100, 400, and 600 MU/
`min, respectively. A clear trend observed in these figures is
`that the magnitude of the dose variations among the seg-
`ments increases as the dose rate increases. At a given dose
`rate ~400 MU/min!, as shown in Fig. 9, the magnitude of
`dose variations decreases as the total MU increases. These
`results indicate that the longer the beam-on time for each
`segment, the smaller the relative dose variations. If one sets
`the acceptable dose variation to be about 3%, similar to dose
`variations in the mechanism I step and shoot delivery, Table
`I lists the minimum MUs required to achieve ,3% dose
`variation among segments for this simple intensity pattern
`delivered with three dose rates. From Table I, a total
`beam-on time of about 10 s is needed. It should be noted that
`the required total MUs depend on the complexity of an in-
`tensity pattern.
`Another result observed from Figs. 8~a! to 8~e! is that the
`first segment always received more dose than expected,
`while the last segment always received less dose than ex-
`pected. For a given total MU, the average from each segment
`is almost a constant, as shown in Fig. 10, independent of
`delivery dose rate, or beam-on time. This result can be ex-
`plained that because the total MU is controlled only by the
`machine console, the radiation beam will be terminated when
`the expected total MU is delivered, regardless of the MLC
`control system.
`
`IV. DISCUSSION
`In IMRT delivery using a dynamic MLC, the key is to
`establish a precise correlation between the MLC positions
`and the cumulative MUs. This relationship is analogous to
`the relationship between the spatial domain ~MLC positions!
`and the frequency domain ~cumulative MUs!. According to
`
`Medical Physics, Vol. 29, No. 3, March 2002
`
`

`

`419
`
`Xia, Chuang, and Verhey: Communication and sampling rate limitations
`
`419
`
`FIG. 8. ~a!–~e! The relative dose variations among the four square segments
`delivered with mechanism II step and shoot method, with total MUs of 1, 4,
`16, 64, and 100 MU at dose rates of 100, 400, and 600 MU/min.
`
`sampling points at a sampling rate of 50 ms is equal to 400
`ms, corresponding to 0.67, 2.67, and 4 MU at dose rates of
`100, 400, and 600 MU/min, respectively. If the sampling
`points and the segment defining points are out of synchroni-
`zation, more than 2 sampling points will be needed for each
`segment. The result of our experiment, however, indicates
`that a beam-on time of about 10 s is necessary in order to
`obtain about 3% dose accuracy for each segment. This
`beam-on time consists of 200 sampling points, i.e., 50 sam-
`pling points per segment. For this simple intensity pattern, it
`is, however, difficult to explain why 50 sampling points per
`segment are needed even if the sampling points and segment
`defining points are out of synchronization. In addition, sam-
`pling theory cannot explain why the first segment is always
`
`overdosed and the last segment is always underdosed. The
`communication time lag between the two control systems,
`i.e., the machine console and the MLC station may explain
`this result. It has been confirmed by the vendor ~Varian On-
`cology System, Palo Alto, CA! that this communication time
`lag can be 50 ms. Even with sufficient sampling points and
`depending on the coherence between the sampling points and
`the segment defining points, the communication lag between
`the two control systems causes the machine console to ter-
`minate 50 ms more or less later than it should for each seg-
`ment. The MU delivered during this time lag can be esti-
`mated as follows:
`
`DD5tR,
`
`~3!
`
`Medical Physics, Vol. 29, No. 3, March 2002
`
`

`

`420
`
`Xia, Chuang, and Verhey: Communication and sampling rate limitations
`
`420
`
`TABLE III. Time delay in milliseconds ~ms! calculated for the first segment
`in the four segment IM field according to the measured dose error shown in
`Fig. 8 using Eq. ~1!.
`
`Total
`
`1 MU
`
`4 MU
`
`16 MU
`
`64 MU
`
`100 MU
`
`MU/seg
`100 MU/min
`400 MU/min
`600 MU/min
`
`0.25
`40.3 ms
`68.9 ms
`50.3 ms
`
`1
`43.6 ms
`50.3 ms
`62.5 ms
`
`4
`86.5 ms
`63.7 ms
`66.7 ms
`
`16
`78.5 ms
`59.3 ms
`66.6 ms
`
`25
`92.6 ms
`64.0 ms
`77.3 ms
`
`where t is the time lag, and R is the dose rate in MU/s. If the
`programmed MU to the first segment is D 1 , the relative dose
`error caused by this time delay is
`~4!
`d5DD/D 1 .
`R5400 MU/min
`D 151 MU,
`if
`For
`example,
`t550 ms50.05 s, d533%. From Fig. 8~b!,
`56.67 MU/s,
`the dose variation for this first segment is about 33%. The
`agreement between the dose error estimated from Eqs. ~3!
`and ~4! and the measured dose difference indicates that the
`dose error for this segment is mainly from the communica-
`tion delay, not from undersampling. For the first segment
`delivered with three different dose rates and the five given
`total MUs, Table III shows the calculated time delays accord-
`ing to Eqs. ~3! and ~4!. The relative dose errors used in the
`calculation were from the measurement data shown in Figs
`8~a!– 8~e!. Only the time delays for the first segment were
`calculated because the measured relative dose errors in the
`other segments may not be completely attributed to the time
`delay. The calculated time delays in Table III include not
`only the communication lag between the two controllers, but
`also the time between the two sampling points, which is also
`50 ms. The added communication lag changes the effective
`sampling rate such that it becomes somewhat irregular. The
`middle sampling points force the MLC leaves to catch up
`with the MU delivery, and the middle segments may some-
`times need to be skipped in order to catch up with the MU
`delivery. The machine console terminates the total MU
`whenever it reaches the programmed MU, regardless of
`whether the later segments receive the intended MU or not
`since the total MU is controlled by the machine console,
`independent of the MLC control station.
`In IMRT delivery, due to the use of small MU, the dose
`accuracy may also be affected by changes in dose linearity
`and symmetry while delivering these small MU fields. The
`details of machine-related quality assurance issues have been
`discussed in our previously published paper.18 Briefly, the
`dose linearity measured down to 0.1 MU/segment delivered
`with mechanism II step and shoot method for the Clinac
`2300C/D used in this study was less than 1% different from
`
`FIG. 9. The relative dose variations among the four square segments deliv-
`ered with a dose rate of 400 MU/min, at a total of 1, 4, 16, 64, 64, and 100
`MU.
`
`FIG. 10. The average dose of the four square segments delivered with
`mechanism II step and shoot method, is nearly a constant, independent of
`the delivery dose rates.
`
`TABLE I. Total MUs and beam-on time required to achieve about 3% dose
`variation among the four square segments.
`
`Dose rate
`~MU/min! Segment 1 Segment 2 Segment 3 Segment 4
`
`Total
`MU
`
`Time
`~s!
`
`100
`400
`600
`
`13.6%
`12.5%
`13.1%
`
`20.3%
`0%
`20.3%
`
`21.0%
`20.4%
`20.5%
`
`22.3%
`22.1%
`22.2%
`
`16
`64
`100
`
`9.6
`9.6
`10
`
`TABLE II. Number of sampling points calculated according to the sampling
`rate of 20 Hz at three different dose rates for five given total MUs.
`
`TABLE IV. Average standard deviations ~%! for experiment B.
`
`1 MU
`
`4 MU
`
`16 MU
`
`64 MU
`
`100 MU
`
`1 MU
`
`4 MU
`
`16 MU
`
`64 MU
`
`100 MU
`
`100 MU/min
`400 MU/min
`600 MU/min
`
`12
`3
`2
`
`48
`12
`8
`
`192
`48
`32
`
`768
`192
`128
`
`1200
`300
`200
`
`100 MU/min
`400 MU/min
`600 MU/min
`
`7.3%
`8.8%
`13.4%
`
`2.3%
`4.6%
`7.2%
`
`1.2%
`2.2%
`1.3%
`
`0.24%
`0.44%
`0.94%
`
`0.12%
`0.19%
`0.73%
`
`Medical Physics, Vol. 29, No. 3, March 2002
`
`

`

`421
`
`Xia, Chuang, and Verhey: Communication and sampling rate limitations
`
`421
`
`FIG. 11. A histogram of the number of segments as a function of MU/seg for
`a nasopharyngeal plan, using 15 gantry angles, and 10 intensity levels with
`a total of 491 segments.
`
`the dose delivered by the conventional method with the same
`total MUs. An ionization chamber was used to measure point
`doses at several symmetric locations. The field symmetry
`and flatness for the Clinac 2300C/D using mechanism II step
`and shoot delivery were found comparable with that in con-
`ventional delivery.
`Since the measurement conducted in this study involved
`very small MUs, the issue of measurement uncertainty needs
`to be addressed, especially for experiment B. To measure the
`dose in each segment with different total MUs at different
`dose rates, three readings were taken in each measurement.
`The average standard deviations of doses measured from all
`four segments delivered with mechanism II step and shoot
`delivery are listed in Table IV. The greatest measurement
`uncertainty is found to be 13.4% for the shortest beam-on
`time for the delivery of a total of 1 MU at 600 MU/min. This
`uncertainty decreases as the beam-on time increases, as
`shown in Table IV. The relatively larger measurement uncer-
`tainties for shorter beam-on time may be due to the limited
`sensitivity in the electrometer, or the stability of the linear
`accelerator. These measurement uncertainties, however,
`should not alter the results of this study.
`
`V. CLINICAL SIGNIFICANCE
`With increasing numbers of beam angles and intensity
`levels, a clinical treatment plan may include many small MU
`segments and the accuracy of dose delivered to these seg-
`ments may be affected. Figure 11 shows a histogram of the
`number of segments as a function of MU/seg for a nasopha-
`ryngeal plan using 15 gantry angles, and 10 intensity levels
`with a total of 491 segments. This plan was used for a patient
`treatment, and the total delivery time for this plan was about
`20 min. Due to the efficiency of Varian IMRT delivery, com-
`plex plans such as this are used in clinics. In Fig. 11, the 1–2
`MU/seg bin has the highest incidence. Since these small MU
`segments are spread out through the entire IMRT fields, our
`phantom plan measurement indicates that the effects of sam-
`pling error and timing error are not significant in this clinical
`case. Specifically, ionization chamber measurements for a
`phantom plan created for this patient yield a dose difference
`within 2% between the measured and the calculated doses in
`a relatively uniform high dose region of 92% of the maxi-
`
`Medical Physics, Vol. 29, No. 3, March 2002
`
`FIG. 12. Cross plane profiles delivered with 96 MU at three different rates of
`100, 400, and 600 MU/min compared for three different MLC leaf pairs
`shown in ~a!, ~b!, and ~c!.
`
`

`

`422
`
`Xia, Chuang, and Verhey: Communication and sampling rate limitations
`
`422
`
`FIG. 13. ~a! A lateral portal film image for a head and neck case. The
`intensity pattern of this field is superimposed with the patient’s anatomy. ~b!
`The planned intensity pattern for this lateral field.
`
`mum dose while at the low dose region of 30% of the maxi-
`mum dose, the measured dose was 5% higher than the pre-
`dicted iso-dose line.
`Furthermore, we delivered the AP field intensity pattern of
`this plan with the planned 96 MU to XV films using three
`different dose rates. The films were positioned at 1.5 cm
`depth in a solid water phantom, and the exposed films were
`scanned with a film scanner. Cross plane profiles of a se-
`lected MLC leaf pair were compared for 100, 400, and 600
`MU/min delivery, as shown in Fig. 12~a!. The profiles were
`normalized to the maximum dose, and the spatial location of
`the maximum dose was manually matched. Figures 12~b!
`and 12~c! show cross plane profiles of two other MLC pairs.
`Again, Figs. 12~a!–12~c! show that with a clinical dose of
`180 cGy, the effects of undersampling and communication
`time lag are not significant.
`At a very low dose, 10 cGy, for example, the effects of
`undersampling and communication time lag are significant.
`One scenario related to such a low dose delivery is when an
`intensity pattern is verified using a portal film. To avoid over-
`exposure of the film, a special port film plan with a total dose
`of 10 cGy was created for the patient. The average MU to
`each IMRT field was about 5–7 MU, and an open field port
`with 1–2 MU was added to the IMRT pattern to image the
`patient anatomy surrounding the treated region. Figure 13~a!
`shows a port film image for the patient with the intensity
`pattern superimposed. Figure 13~b! shows the intensity pat-
`tern printed from the treatment planning system. Figure
`13~b! is a fluence pattern while Fig. 13~a! is an attenuation
`pattern, but their correlation should be straightforward. In
`Fig. 13~a!, arrows are used to point out the differences be-
`tween the two intensity patterns. To eliminate possible at-
`tenuation differences due to anatomic structures, Fig. 14~a!
`shows the same pattern for Fig. 12 delivered to a solid water
`phantom with an EC-L port film embedded in a cassette,
`positioned at 1.5 cm depth, with a total of 5 MU and deliv-
`ered at 400 MU/min. Under the same setup, Fig. 14~b! shows
`the same intensity pattern delivered to a XV film with a total
`of 96 MU when the plan is prescribed to 180 cGy. Figure
`14~c! shows an intensity pattern printed from the treatment
`planning system. With the higher MU, the agreement be-
`tween the delivered intensity pattern ~Fig. 14~b!! and the
`expected pattern ~Fig. 14~c!! is nearly perfect by visual in-
`
`Medical Physics, Vol. 29, No. 3, March 2002
`
`FIG. 14. ~a! An intensity pattern from a patient plan delivered to a EC-L
`portal film placed under 1.5 cm solid wat

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket