throbber
U.S. PATENT NO. 9,561,314
`Petition for Inter Partes Review
`
`UNITED STATES PATENT AND TRADEMARK OFFICE
`____________________________________________
`
`
`
`BEFORE THE PATENT TRIAL AND APPEAL BOARD
`____________________________________________
`
`
`
`
`
`
`
`Abiomed, Inc. and Abiomed R&D, Inc.
`Petitioner
`
`v.
`
`Maquet Cardiovascular, LLC
`Patent Owner
`
`Case No. IPR2017-01204
`
`
`
`PETITION FOR INTER PARTES REVIEW OF U.S. PATENT NO. 9,561,314
`
`
`
`
`CLAIMS 1-8, 10-23, and 25-26
`
`
`
`
`
`

`

`U.S. PATENT NO. 9,561,314
`Petition for Inter Partes Review
`
`TABLE OF CONTENTS
`
`Introduction ...................................................................................................... 1
`
`
`
`I.
`
`II. Mandatory Notices ........................................................................................... 1
`
`A. Real Party-in-Interest ................................................................................ 1
`
`B. Related Matters ......................................................................................... 1
`
`C. Counsel ...................................................................................................... 1
`
`D. Service Information ................................................................................... 1
`
`III. Grounds for Standing ....................................................................................... 2
`
`IV. Relief Requested .............................................................................................. 2
`
`A. The Challenged Claims Are Invalid in View of the Following Prior
`
`Art: ........................................................................................................... 2
`
`B. Grounds for Challenge .............................................................................. 3
`
`V.
`
`Conventional Technology ................................................................................ 4
`
`A. Conventional Intravascular Blood Pumps ................................................. 4
`
`B. Placing Intravascular Blood Pumps .......................................................... 5
`
`1. Over-the-Wire .................................................................................... 5
`
`2. Rapid-Exchange ................................................................................ 6
`
`3. Guide Catheter ................................................................................... 8
`
`i
`
`

`

`U.S. PATENT NO. 9,561,314
`Petition for Inter Partes Review
`4. Interchangeability of Over-the-Wire and Rapid Exchange ............... 8
`
`VI. Overview of the ’314 Patent ............................................................................ 9
`
`A. Summary of the ’314 Patent ...................................................................... 9
`
`B. Prosecution History ................................................................................. 13
`
`C. The Earliest Possible Priority Date for the ’314 Patent is September
`
`1, 2000 .................................................................................................... 14
`
`VII. Overview of the Prior Art .............................................................................. 15
`
`A. Overview of Aboul-Hosn ........................................................................ 16
`
`B. Overview of Siess .................................................................................... 21
`
`C. Overview of Jegaden ............................................................................... 21
`
`D. Overview of Wampler ............................................................................. 23
`
`E. Overview of Wampler_712 ................................................................... 24
`
`F. Overview of Crowley ............................................................................. 26
`
`VIII. Claim construction ......................................................................................... 27
`
`A. “distal” ..................................................................................................... 27
`
`B. “proximal” ............................................................................................... 28
`
`IX. Person having ordinary skill in the art ........................................................... 29
`
`X.
`
`Specific Grounds for Petition ........................................................................ 29
`
`ii
`
`

`

`U.S. PATENT NO. 9,561,314
`Petition for Inter Partes Review
`A. Ground I: Claims 1-8, 14-20, and 25-26 are obvious over Aboul-
`
`Hosn in view of Siess and Wampler ...................................................... 30
`
`1. Claim 1 ............................................................................................ 30
`
`2. Claim 2 ............................................................................................ 61
`
`3. Claim 3 ............................................................................................ 62
`
`4. Claim 4 ............................................................................................ 62
`
`5. Claim 5 ............................................................................................ 64
`
`6. Claim 6 ............................................................................................ 67
`
`7. Claim 7 ............................................................................................ 70
`
`8. Claim 8 ............................................................................................ 70
`
`9. Claim 14 .......................................................................................... 73
`
`10. Claim 16 .......................................................................................... 73
`
`11. Claim 17 .......................................................................................... 75
`
`12. Claim 18 .......................................................................................... 75
`
`13. Claim 19 .......................................................................................... 76
`
`14. Claim 20 .......................................................................................... 78
`
`15. Claim 25 .......................................................................................... 86
`
`16. Claim 26 .......................................................................................... 86
`
`iii
`
`

`

`U.S. PATENT NO. 9,561,314
`Petition for Inter Partes Review
`B. Ground II: Claims 10-11, 13, 21, and 23 are obvious over Aboul-
`
`Hosn in view of Siess and Wampler, and further in view of Jegaden. .. 86
`
`1. Claim 10 .......................................................................................... 86
`
`2. Claim 11 .......................................................................................... 88
`
`3. Claim 13 .......................................................................................... 89
`
`4. Claim 21 .......................................................................................... 91
`
`5. Claim 23 .......................................................................................... 91
`
`C. Ground III: Claim 12 and 22 are obvious over Aboul-Hosn in view
`
`Siess and Wampler, and further in view of Crowley. ............................ 91
`
`1. Claim 12 .......................................................................................... 91
`
`2. Claim 22 .......................................................................................... 92
`
`D. Ground IV: Claim 15 is obvious over Aboul-Hosn in view of Siess
`
`and Wampler, and further in view of Wampler_712. ............................ 92
`
`1. Claim 15 .......................................................................................... 93
`
`XI. Conclusion ..................................................................................................... 94
`
`
`
`iv
`
`

`

`U.S. PATENT NO. 9,561,314
`Petition for Inter Partes Review
`TABLE OF AUTHORITIES
`
`Page(s)
`
`
`
`FEDERAL CASES
`Dynamic Drinkware, LLC. V. Nat’l Graphics, Inc.,
`800 F.3d 1375 (Fed. Cir. 2015) .......................................................................... 14
`
`In re ICON Health & Fitness, Inc.,
`496 F.3d 1374 (Fed. Cir. 2007) .......................................................................... 28
`
`FEDERAL STATUTES
`
`35 U.S.C. § 102(a) ................................................................................................... 15
`
`35 U.S.C. § 102(b) ............................................................................................... 4, 15
`
`35 U.S.C. § 103 .......................................................................................................... 1
`
`35 U.S.C. § 103(a) ................................................................................................. 4, 5
`
`35 U.S.C. § 112 ¶ 1 .................................................................................................. 14
`
`35 U.S.C. § 312 ......................................................................................................... iv
`
`35 U.S.C. § 314(a) ..................................................................................................... 4
`
`FEDERAL RULES
`
`Rule 42.104(a) ............................................................................................................ 3
`
`Rule 42.104(b)(4)-(5) ............................................................................................... 31
`
`Rules 42.22(a)(1) and 42.104(b)(1)-(2) ..................................................................... 3
`
`FEDERAL REGULATIONS
`
`37 C.F.R. § 42.8(b)(4) ................................................................................................ 2
`
`37 C.F.R. § 42.100(b). ............................................................................................. 28
`
`37 C.F.R. § 42.104 .................................................................................................... iv
`
`v
`
`

`

`U.S. PATENT NO. 9,561,314
`Petition for Inter Partes Review
`
`I.
`
`INTRODUCTION
`
`Petitioners Abiomed, Inc. and Abiomed R&D, Inc. (collectively,
`
`“Petitioner”) petition for inter partes review (“IPR”) of claims 1-8, 10-23, and 25-
`
`26 (the “Challenged Claims”) of U.S. Patent No. 9,561,314 (the “’314 patent”).
`
`The Challenged Claims add nothing new to the art and should be found
`
`unpatentable and canceled.
`
`II. MANDATORY NOTICES
`A. Real Party-in-Interest
`The real parties in interest are Abiomed, Inc. and Abiomed R&D, Inc.
`
`B. Related Matters
`Petitioner has filed, or will file, concurrently with the present Petition: (1) a
`
`second petition for inter partes review of claims 27 and 29-30 of the ’314 patent;
`
`(2) petitions for IPR of U.S. Patent Nos. 9,545,468 and 9,597,437; and (3) petitions
`
`for IPR of U.S. Patent Nos. 7,022,100 (IPR2017-01025), 8,888,728 (IPR2017-
`
`01026 and IPR2017-01027), and 9,327,068 (IPR2017-01028 and IPR2017-01029)
`
`(the “related patents”) which are related to the ’314 patent.
`
`C. Counsel
`Lead Counsel: David M. Tennant (Reg. No. 48,362)
`
`Backup Counsel: Charles D. Larsen (Reg. No. 48,533); Christopher Carroll
`
`(Reg. No. 55,776)
`
`D. Service Information
`
`1
`
`

`

`U.S. PATENT NO. 9,561,314
`Petition for Inter Partes Review
`Pursuant to 37 C.F.R. § 42.8(b)(4), papers concerning this matter should be
`
`served on the following. Petitioner consents to electronic service.
`
`David M. Tennant (Reg. No. 48,362)
`
`E-mail: WCAbiomedIPR@whitecase.com
`
`Post and hand delivery: White & Case LLP
`
`701 Thirteenth Street, NW
`
`Washington, DC 20005
`
`Telephone: (202) 626-3684
`
`
`
`Fax: (202) 639-9355
`
`III. GROUNDS FOR STANDING
`Petitioner certifies the ’314 patent is available for IPR and that Petitioner is
`
`not barred or estopped from requesting IPR of the Challenged Claims.
`
`IV. RELIEF REQUESTED
`Petitioner requests review of the Challenged Claims and a ruling that the
`
`Challenged Claims are unpatentable.
`
`A. The Challenged Claims Are Invalid in View of the Following
`Prior Art1:
`1. WO 99/02204 to Aboul-Hosn (EX1004,“Aboul-Hosn”), published January
`
`21, 1999, is prior art under 35 U.S.C. § 102(b).
`
`2.
`
`U.S. Patent No. 5,921,913 to Siess (EX1005,“Siess”), filed June 24, 1997
`
`and issued July 13, 1999, is prior art under 35 U.S.C. § 102(b).
`
`1 The pre-AIA statutory framework applies to the ’314 Patent.
`
`2
`
`

`

`U.S. PATENT NO. 9,561,314
`Petition for Inter Partes Review
`Jegaden, Clinical results of Hemopump support in surgical cases, Published
`
`3.
`
`in Temporary Cardiac Assist with an Axial Pump System, p.61-65 (Springer 1991)
`
`(EX1033,“Jegaden”), is prior art under 35 U.S.C. § 102(b).2
`
`4. Wampler et al., Clinical Experience with the Hemopump Left Ventricular
`
`Support Device, published in Supported Complex and High Risk Coronary
`
`Angioplasty, ch. 14, 231-49 (Springer 1st ed. 1991) (EX1007,“Wampler”),
`
`published in 1991, is prior art under 35 U.S.C. § 102(b).3
`
`5.
`
`U.S. Patent No. 5,421,338 to Crowley (EX1047,“Crowley”), filed June 3,
`
`1994 and issued June 6, 1995, is prior art under 35 U.S.C. § 102(b).
`
`6.
`
`U.S. Patent No. 4,625,712 to Wampler (EX1008, “Wampler_712”). Filed
`
`September 28, 1983 and issued December 2, 1986, is prior art under 35 U.S.C. §
`
`102(b).
`
`B. Grounds for Challenge
`Petitioner respectfully requests review of the Challenged Claims and
`
`cancellation of those claims under the following statutory grounds:
`
`2 Jegaden bears a copyright date of 1991 and was publicly available from 1992.
`
`EX1033; Declaration of Leupold (EX1046).
`
`3 Wampler bears a copyright date of 1991 and was publicly available from 1991.
`
`See EX1007; Declaration of Kiersten Batzli (EX1035); Library of Congress Card
`
`Catalog (EX1036).
`
`3
`
`

`

`U.S. PATENT NO. 9,561,314
`Petition for Inter Partes Review
`• Ground 1: Claims 1-8, 14, 16-20, and 25-26 are rendered obvious by
`
`Aboul-Hosn in view of Siess and Wampler, under 35 U.S.C. § 103(a).
`
`• Ground 2: Claims 10-11, 13, 21, and 23 are rendered obvious by Aboul-
`
`Hosn in view of Siess and Wampler, and further in view of Jegaden
`
`under 35 U.S.C. § 103(a).
`
`• Ground 3: Claims 12 and 22 are rendered obvious by Aboul-Hosn in
`
`view of Siess and Wampler, and further in view of Crowley under 35
`
`U.S.C. § 103(a).
`
`• Ground 4: Claims 15 is rendered obvious by Aboul-Hosn in view of Siess
`
`and Wampler, and further in view of Wampler_712 under 35 U.S.C. §
`
`103(a).
`
`V. CONVENTIONAL TECHNOLOGY
`A. Conventional Intravascular Blood Pumps
`The blood pump features of the Challenged Claims were well known: (1) a
`
`cannula formed as a tube, connected at its proximal end to an axial flow pump and
`
`with a distal end to be disposed in a heart chamber (Collins ¶¶52, 55-59;
`
`EX1004[Aboul-Hosn] 30:20-28; U.S. Patent No. 4,625,712 to Wampler
`
`(EX1008[Wampler_712] 3:40-51); EX1013[Siess] 5:28-61); (2) a pump having a
`
`tapered rotor with a distally extending tip and multiple blades disposed within a
`
`shroud, to pump blood axially along the pump and through the cannula (Collins
`
`4
`
`

`

`U.S. PATENT NO. 9,561,314
`Petition for Inter Partes Review
`¶¶61-65; EX1004[Aboul-Hosn] 12:28-13:31, 16:30-17:26;
`
`EX1008[Wampler_712] 3:26-39; U.S. Patent No. 4,846,152 to Wampler et al.
`
`(EX1009,[Wampler_152] 2:63-3:23); EX1005[Siess] 6:26-7:21); (3) a purge fluid
`
`system to prevent blood from entering the pump motor and to lubricate the pump
`
`motor (Collins ¶¶66-72; EX1004[Aboul-Hosn] 20:16-21:3;
`
`EX1008[Wampler_712] 3:40-51; EX1005[Seiss913] 8:31-44); and (4) techniques
`
`for monitoring blood pressure near the pump (Collins ¶¶73-79; EX1004[Aboul-
`
`Hosn] 29:16-25; EX1005[Siess] 11:25-56). The few other minor details of the
`
`Challenged Claims were also well-known in intravascular blood pumps in the prior
`
`art—I.e., purge fluid pump and fittings (in Wampler), or pressure sensor
`
`configurations (in Aboul-Hosn or Siess). (Collins ¶¶70-72, 76-78.)
`
`Placing Intravascular Blood Pumps
`
`B.
`The guide wire mechanism of the Challenged Claims was well-known in
`
`catheterization as the “over-the-wire” technique. Along with “rapid-exchange” and
`
`“guide catheters,” over-the-wire was used routinely to position blood pumps
`
`intravascularly. (Collins ¶¶81-82.)
`
`Over-the-Wire
`
`1.
`POSITAs used “over-the-wire” guide mechanism to place intravascular
`
`blood pumps. (Collins ¶¶86-88.) For example, as shown below in FIG. 3, U.S.
`
`5
`
`

`

`U.S. PATENT NO. 9,561,314
`Petition for Inter Partes Review
`Patent No. 6,248,091 to Voelker4 (EX1011,“Voelker”) applied the “over-the-wire”
`
`guide mechanism to an axial flow intravascular blood pump with the guide wire
`
`extending coaxially through the flexible shaft 23, the shaft 19, and the impeller
`
`wheel 20 so that the pump may be slipped over the guide wire. (Collins ¶88;
`
`EX1011[Voelker] 3:56-60; EX1004[Aboul-Hosn] 22:10-16, FIG. 3.)
`
`
`
`(Collins ¶88; EX1011[Voelker] FIG. 3, annotated.)
`
`As explained in further detail in Sections VII.A and X below, Aboul-Hosn
`
`disclosed that same well-known “over-the-wire” catheterization technique and
`
`used it in delivering intravascular blood pumps into the heart.
`
`Rapid-Exchange
`
`2.
`“Rapid-exchange” was a well-known catheterization technique. (Collins
`
`¶90.) Yock disclosed placing a conventional “rapid-exchange” catheter by sliding
`
`4 Voelker is also published as PCT Publication WO97/46270 on Dec. 11, 1997.
`
`6
`
`

`

`U.S. PATENT NO. 9,561,314
`Petition for Inter Partes Review
`it along a guide wire extending through a sleeve secured to the exterior of the
`
`cannula or embedded within the cannula wall itself. (Id. ¶91; EX1006[Yock] FIG.
`
`10;7:64-8:2.)
`
`(Collins ¶91; EX1006[Yock] FIG. 10, annotated.)
`
`
`
`Voelker, at Fig. 2 (below) applied this rapid exchange approach to an
`
`intravascular blood pump—a guide wire 25 extended through a side channel for
`
`positioning a blood pump as illustrated below. (Collins ¶93; EX1011[Voelker]
`
`
`
`3:34-43.)
`
`(Collins ¶93; EX1011[Voelker] FIG. 2, annotated.)
`
`
`
`7
`
`

`

`U.S. PATENT NO. 9,561,314
`Petition for Inter Partes Review
`
`Guide Catheter
`
`3.
`Yock also discloses using a guide catheter to position a guide wire. (Collins
`
`¶83; EX1006[Yock] 3:56-4:50.) The same technique as disclosed by Yock has had
`
`been adapted to place axial flow intravascular blood pumps. (Collins ¶84.) In fact,
`
`the ’314 Patent acknowledges this conventional technique. (EX1001[’314 Patent]
`
`2:34-55.)
`
`Interchangeability of Over-the-Wire and Rapid Exchange
`
`4.
`Over-the-wire and rapid-exchange techniques have long been used
`
`interchangeably, with minimal differences in design, to deliver cardiac assist
`
`devices, including blood pumps. (Collins ¶96; EX1023[Faxon] 58-59;
`
`EX1006[Yock] 8:16-25; EX1011[Voelker] FIGS. 2 and 3.) Over-the-wire and
`
`rapid-exchange were part of a limited set of delivery techniques. (Collins ¶97.)
`
`For example, U.S. Patent No. 5,928,181 to Coleman et al. (EX1042, “Coleman”)
`
`discloses that catheters coupled to a cardiopulmonary bypass pump may be
`
`suitably guided into position using either technique. (EX1042[Coleman] 34:14-
`
`39.)
`
` The interchangeability of over-the-wire and rapid-exchange was also well
`
`understood for intravascular blood pump applications. (Collins ¶98.) For
`
`example, Voelker’s blood pump could be configured to use either technique. (Id.;
`
`EX1011[Voelker] FIG. 2 (over-the-wire), and FIG. 3 (rapid-exchange).) Of
`
`8
`
`

`

`U.S. PATENT NO. 9,561,314
`Petition for Inter Partes Review
`course, deploying an intravascular blood pump to use over-the-wire or rapid-
`
`exchange involves certain design choices, but such design choices were also well-
`
`known within the prior art. (Collins ¶99; EX1001[Voelker] 3:34-55.)
`
`VI. OVERVIEW OF THE ’314 PATENT
`A. Summary of the ’314 Patent
`The ’314 patent describes placement of a conventional intravascular blood
`
`pump system using the same three conventional guide wire delivery techniques
`
`discussed above. (EX1001[’314 patent] 9:13-32, 13:62-14:10, 15:5-21; Collins
`
`¶102-104.) The background of the ’314 patent openly admits that it is not the first
`
`to use “guide mechanism[s]” to place an intravascular pump. (EX1001[’314
`
`patent] 2:35-45.)
`
`FIG. 1 of the ’314 patent illustrates a conventional over-the-wire placement
`
`technique. (Id. 5:25-30.)
`
`9
`
`

`

`U.S. PATENT NO. 9,561,314
`Petition for Inter Partes Review
`
`Pump
`
`Cannula
`
`“Over-the-wire”
`guide mechanism
`
`Guide wire
`
`(Collins ¶105; EX1001[’314 patent] FIG. 1, annotated.)
`
`
`
`The conventional intravascular blood pump system includes an intravascular
`
`blood pump 12 rotor hub, cannula 14, and over-the-wire guide mechanism 16 with
`
`a guide wire lumen that passes through the center of the rotor hub and the cannula
`
`14. (Id. 9:13-24; Collins ¶105.) The blood pump 12 provides heart support in the
`
`same manner as conventional axial-flow intravascular blood pumps – by
`
`“deliberately re-rout[ing] through and past the right and/or left ventricle in an
`
`10
`
`

`

`U.S. PATENT NO. 9,561,314
`Petition for Inter Partes Review
`effort to reduce the volume of blood to be pumped by the particular ventricle.”
`
`(Compare EX1001[’314 patent] 20:43-48 with id. 2:16-19; Collins ¶103.)
`
`FIG. 6 shows the conventional “rapid-exchange” or “side-rigger” guide
`
`mechanism of the prior art. (Id. 5:25-30.) The guide mechanism 122 “includes a
`
`guide carriage 124 formed along at least a portion of the cannula 14, and a suitable
`
`… guide wire 22 … dimensioned to pass slideably through a lumen (not shown)
`
`extending through the guide carriage 124.” (Id. 14:15-21; Collins ¶107-108.) As
`
`with the prior art (Collins ¶108), the guide carriage 124 may be formed as “an
`
`integral extension of the wall of the cannula 14.” (EX1001[’437 patent] 14:37-39.)
`
`11
`
`

`

`U.S. PATENT NO. 9,561,314
`Petition for Inter Partes Review
`
`(Collins ¶107; EX1001[’314 patent] FIG. 6, annotated.)
`
`
`
`Finally, the ’314 patent at FIG. 10 shows a “guide catheter” mechanism 132
`
`as in the prior art where the rotor and shroud are placed in two different steps.
`
`(EX1001[’314 patent] 5:65-6:3; Collins ¶110)
`
`12
`
`

`

`U.S. PATENT NO. 9,561,314
`Petition for Inter Partes Review
`
`
`
`(Collins ¶110; EX1001[’314 patent] FIG. 10, annotated.)
`
`Prosecution History
`
`B.
`During prosecution of the ’314 patent, in the sole office action the Examiner
`
`indicated that eight co-pending and patented applications were relevant to the
`
`Challenged Claims, but found the Challenged Claims to be patentably distinct—
`
`not because of a specific combination of elements, but because “a guide wire not
`
`passing through the rotor hub and a housing and catheter with a purge lumen” were
`
`not disclosed in those applications. (EX1003[’314 PH] 238.) There is no
`
`13
`
`

`

`U.S. PATENT NO. 9,561,314
`Petition for Inter Partes Review
`patentable synergy between the recited guide wire configuration and the purge
`
`system feature that the Examiner found missing. (Collins ¶112.) As with the other
`
`conventional features of the Challenged Claims, a “guide wire not passing through
`
`the rotor hub” and a “housing and catheter with a purge lumen” were conventional
`
`elements that were well-known and disclosed by prior art references not relied on
`
`by the Examiner. (Id. ¶113.)
`
`C. The Earliest Possible Priority Date for the ’314 Patent is
`September 1, 2000
`
`The earliest possible priority date (as defined above, the “EPD”) for the
`
`Challenged Claims is September 1, 2000, which is the date of PCT Application
`
`No. PCT/US00/24515, to which the ’314 patent claims priority. The subject matter
`
`of the Challenged Claims is not supported by its claimed earlier-filed provisional
`
`application. (EX1012, the “’249 provisional application”).
`
`Independent claims 1 and 20 require, for example, “a purge lumen extending
`
`through the catheter and operatively arranged to deliver purge fluid,” “a pressure
`
`sensing element”, “an elongate lumen arranged coaxially with at least a portion of
`
`the cannula,” “an end of the elongate lumen is adjacent an end of the cannula,” and
`
`“the elongate lumen sized to slidably receive the guide wire.” (EX1001[’314
`
`patent] 34:11-14; 34:22; 34:17-18; 34:15-21; 34:18-21; 36:2-3; 36:12, 36:5-9.)
`
`14
`
`

`

`U.S. PATENT NO. 9,561,314
`Petition for Inter Partes Review
`None of these limitations are supported in the ’249 provisional application.5
`
`(Collins ¶124-125; Dynamic Drinkware, LLC. v. Nat’l Graphics, Inc., 800 F.3d
`
`1375, 1378 (Fed. Cir. 2015) (“[T]he specification of the provisional must contain a
`
`written description of the invention and the manner and process of making and
`
`using it, in such full, clear, concise, and exact terms, 35 U.S.C. § 112 ¶1, to enable
`
`an ordinarily skilled artisan to practice the invention claimed in the non-
`
`provisional application.”) (emphasis in original).)
`
`Accordingly, the EPD for the Challenged Claims is September 1, 2000.6
`
`(Collins ¶127.)
`
`VII. OVERVIEW OF THE PRIOR ART
`
`
`5 During prosecution of the related ’728 Patent, the Examiner found that the claims
`
`directed to “a blood pressure detection mechanism,” and “a guide wire and an
`
`elongate lumen” were not entitled to the priority date of the ’249 provisional
`
`application, and the Patent Owner never challenged in any subsequent response.
`
`(EX1043[’728 PH] 259-280.)
`
`6 If the Board finds that one or more Challenged Claims is entitled to the
`
`September 3, 1999 filing date of the ’249 provisional application, the cited prior art
`
`would still qualify as prior art under 35 U.S.C. § 102(a) or 102(b).
`
`15
`
`

`

`U.S. PATENT NO. 9,561,314
`Petition for Inter Partes Review
`
`A. Overview of Aboul-Hosn7
`Aboul-Hosn discloses an axial flow intravascular blood pump for heart
`
`support, that is delivered intravascularly to a desired location within the heart
`
`using the same well-known guide mechanisms as noted in the ’314 patent.
`
`(Collins ¶128; EX1004[Aboul-Hosn] 11:9-14; 30:1-2.) Aboul-Hosn further
`
`discloses both percutaneous and surgical approaches for delivering the blood
`
`pump. (Collins ¶147-151; EX1004[Aboul-Hosn] FIG. 21; 21:19-22:30; 11:8-12.)
`
`Annotated FIGS. 21 and 23, below, show a percutaneous approach for
`
`delivering the pump using a guide wire. (Collins ¶148; EX1004[Aboul-Hosn]
`
`30:1-2, 20-27.) FIG. 21 shows how the blood pump (green) passes along the guide
`
`wire up the femoral artery, so the cannula 4118 (blue) goes through the aorta and
`
`into the left ventricle. In FIG. 23, the cannula then also continues into the left
`
`7 Aboul-Hosn and Siess were cited in an Information Disclosure Statement dated
`
`August 19, 2016 but there is no record the Examiner relied upon them.
`
`(EX1003[’314 PH] 271, 280). There is no record of Jegaden, Yock, or Wampler
`
`being disclosed during prosecution of the ’314 patent. (EX1003[’314 PH] 552-
`
`558.)
`
`8 FIG. 20 is a zoomed-in view of the stabilization system 410 of FIG. 23; it
`
`identifies element 411 as the stabilization cannula that passes through the
`
`stabilization balloon 440 (red). (EX1004[Aboul-Hosn] 28:23-27.)
`
`16
`
`

`

`U.S. PATENT NO. 9,561,314
`Petition for Inter Partes Review
`atrium, where it is positioned to pump blood from the left atrium to the aorta.
`
`(Collins ¶¶148-149; EX1004[Aboul-Hosn] 29:17-28, 30:1-2, 30:20-27.) In
`
`addition, the cannula’s inflow tip may also be placed in “the left ventricle, … or
`
`any of the left heart vessels” to provide left heart support. (Id. 26:10-13.)
`
`
`
`(Collins ¶148; EX1004[Aboul-Hosn] FIGS. 21, 23, annotated.)
`
`FIGS. 1-13 show a surgical approach with details about the interior of the
`
`pump and cannula. Numerous conventional features of intravascular blood pumps
`
`are disclosed, including a “commercially available” cannula (EX1004[Aboul-
`
`Hosn] 11:14, together with a “reverse flow” feature that reverses the direction of
`
`blood flow as it exits the pump. (Collins ¶¶135-137.)
`
`17
`
`

`

`U.S. PATENT NO. 9,561,314
`Petition for Inter Partes Review
`As shown below in FIGS. 1 and 2, and similar to the ’314 patent, the pump
`
`system has a conventional drive motor 80 (purple) connected to a rotor and
`
`associated blades 70 (red), within a housing body 52 (green) and a housing cap 60
`
`(green)9. (EX1004[Aboul-Hosn] 12:12-13:13.) The inner cannula 20 (blue) is
`
`coupled to the housing cap 60 (green), and extends beyond the distal opening 32 of
`
`the outer conduit 30 (dark grey). (Id.)
`
`(Collins ¶135; EX1004[Aboul-Hosn] FIGS. 1, 2, annotated.)
`
`
`9 The housing body 52 and the housing cap 60 may form “a unitary body.”
`
`(EX1004[Aboul-Hosn] 12:22-23.)
`
`
`
`18
`
`

`

`U.S. PATENT NO. 9,561,314
`Petition for Inter Partes Review
`As shown below, to apply the percutaneous approach in the reverse flow
`
`configuration, the pump (green) in the system of FIGS. 1-13 (i.e. an intravascular
`
`pump “provided for by the present invention”) would be readily connected to the
`
`multilumen catheter 428 (yellow). (Collins ¶¶144-145.) In this configuration, the
`
`catheter 428 would be used to advance the pump 420 and stabilization cannula 411
`
`over a guide wire to the desired location within the patient’s heart through the
`
`femoral artery. (Collins ¶145; EX1004[Aboul-Hosn] 29:18-25.)
`
`(Collins ¶138; EX1004[Aboul-Hosn] FIGS. 1, 23, annotated.)
`
`The pump 420 could also be configured without the reverse flow feature of
`
`the pump system of FIGS. 1-13. (Collins ¶139-141; see also EX1004[Aboul-
`
`
`
`19
`
`

`

`U.S. PATENT NO. 9,561,314
`Petition for Inter Partes Review
`Hosn] 31:6-9.) In this configuration, the pump 420 would include the components
`
`of the pump system of FIGS. 1-13 that generate the axial flow of blood through the
`
`pump (i.e. rotor 70 and inlet tube 55, connected to drive unit 80), without the
`
`components that cause the blood flow to reverse course (i.e. housing body 52,
`
`housing cap 60, and outer cannula 30). (Collins ¶139.) Instead, the blood
`
`(represented by the red arrows) discharges axially over the drive unit and out the
`
`pump 420 (green). (Id.)
`
`(Collins ¶139; EX1004[Aboul-Hosn] FIGS. 1, 23, annotated.)
`
`As explained in greater detail in Section X.A.1(i) below, Aboul-Hosn uses a
`
`guide wire to deploy the pumps intravascularly using the over-the-wire technique.
`
`
`
`20
`
`

`

`U.S. PATENT NO. 9,561,314
`Petition for Inter Partes Review
`(Collins ¶129; EX1004[Aboul-Hosn] 11:26-28, 14:13-16, 14:20-24, 21:22-24,
`
`22:10-16.) FIG. 3 shows the conventional over-the-wire technique. (Collins ¶129;
`
`EX1004[Aboul-Hosn] 14:17-15:18, 17:19-22, FIG. 12.)
`
`B. Overview of Siess
`Siess also discloses an intravascular blood pump configured to be delivered
`
`to a desired location within the patient’s vasculature using a guide wire. (Collins
`
`¶¶168-172; EX1005[Siess] 5:55-58.) The pump is connected to a catheter that
`
`includes a lumen to deliver a “biocompatible purge fluid[,] … that is pressurized so
`
`as to maintain a slow purge rate over the seals of about 1-5ml/hr[,]” to the
`
`microaxial pump 10. (Id. 8:31-44.). Additionally, it can have “electrical conduits
`
`extending therethrough to allow the operation of the drive unit to be monitored and
`
`controlled.” (Id. 3:15-18, 11:23-40.)
`
`(EX1005[Siess] FIG. 4, annotated.)
`
`C. Overview of Jegaden
`
`
`
`21
`
`

`

`U.S. PATENT NO. 9,561,314
`Petition for Inter Partes Review
`Jegaden discloses placing a conventional axial flow intravascular pump
`
`system (i.e. the Hemopump) to a desired location using the conventional rapid-
`
`exchange technique. (Collins ¶¶152-157; EX1033[Jegaden] 61-62.) As shown in
`
`FIG. 2, the five French (“5F”) catheter having a guide wire extending coaxially
`
`through its lumen and exiting its distal end is “passed through the distal hole of the
`
`cannula and introduced into the femoral artery up to the aorta,” and “[t]hen the
`
`cannula is introduced into the femoral artery and is pushed into the aorta, guided
`
`by the vasculature catheter” (i.e. the 5F catheter). (Collins ¶153; EX1033[Jegaden]
`
`62.) The 5F catheter and cannula’s distal hole function as a rapid-exchange guide
`
`mechanism for the pump. (Collins ¶154-156.)
`
`22
`
`

`

`U.S. PATENT NO. 9,561,314
`Petition for Inter Partes Review
`
`Guide wire
`extending out in a
`distal direction
`
`Distal End of 5F
`Catheter
`
`Distal Hole of
`Canula
`
`5F Catheter
`
`Cannula
`
`Guide wire
`
`
`
`(Collins ¶154; EX1033[Jegaden] FIG. 1, annotated.)
`D. Overview of Wampler
`Wampler discloses the Hemo

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket