throbber
INTERNATIONAL
`
`STANDARD
`
`ISO
`
`10555-1
`
`First edition
`1995-06-15
`
`Sterile, single-use intravascular
`catheters -—
`
`Part 1:
`
`General requirements
`
`Catheters intravasculaires stéri/es, non réuti/isables —
`
`Fame 7: Prescriptions géne’ra/es
`
`gm?
`IESO
`=‘W:___::
`
`Reference number
`ISO 10555-1:1995(E)
`
`B. Braun 2022
`
`1
`
`B. Braun 2022
`
`

`

`ISO 10555-1:1995(E)
`
`Contents
`
`Page
`
`1
`
`Scope ..............................................................................................
`
`2 Normative references .....................................................................
`
`3 Definitions .......................................................................................
`
`1
`
`1
`
`1
`
`4 Requirements
`
`............................................................................ 3
`
`5 Designation of nominal size
`
`...................................................... 3
`
`6
`
`Information to be supplied by manufacturer
`
`............................ 3
`
`Annexes
`
`A Test method for corrosion resistance
`
`..................................... 5
`
`B Method for determining force at break
`
`................................... 6
`
`C
`
`D
`
`E
`
`Test method for liquid leakage under pressure
`
`....................... 7
`
`Test method for air leakage into hub assembly during aspiration
`
`8
`
`Bibliography
`
`.............................................................................. 9
`
`© ISO 1995
`All rights reserved. Unless otherwise specified, no part of this publication may be reproduced
`or utilized in any form or by any means, electronic or mechanical including photocopying and
`microfilm, without permission in writing from the publisher,
`International Organization for Standardization
`Case Postale 56 ° CH-1211 Geneve 20 0 Switzerland
`Printed in Switzerland
`
`ii
`
`2
`
`2
`
`

`

`© ISO
`
`ISO 10555-1:1995(E)
`
`Foreword
`
`is a worldwide
`ISO (the International Organization for Standardization)
`federation of national standards bodies (ISO member bodies). The work
`of preparing International Standards is normally carried out through ISO
`technical committees. Each member body interested in a subject for
`which a technical committee has been established has the right to be
`represented on that committee. International organizations, governmental
`and non-governmental, in liaison with ISO, also take part in the work. ISO
`collaborates closely with the International Electrotechnical Commission
`(IEC) on all matters of electrotechnical standardization.
`
`Draft International Standards adopted by the technical committees are
`circulated to the member bodies for voting. Publication as an International
`Standard requires approval by at least 75 % of the member bodies casting
`a vote.
`
`International Standard ISO 10555-1 was prepared by Technical Committee
`ISO/IC 84, Medical devices for injections, Subcommittee SC I, Syringes,
`needles and intravascular catheters for single use.
`
`ISO 10555 consists of the following parts, under the general title Sterile,
`single-use intravascu/ar catheters:
`
`— Part 7: General requirements
`
`— Part 2: Angiographic catheters
`
`— Part 3: Central venous catheters
`
`—— Part 4: Balloon dilatation catheters
`
`— Part 5: Over-needle peripheral catheters
`
`Attention is drawn to ISO 11070, which will specify requirements for ac—
`cessory devices for use with intravascular catheters.
`
`Annexes A, B, C and D form an integral part of this part of ISO 10555.
`Annex E is for information only.
`
`'
`
`3
`
`iii
`
`3
`
`

`

`
`
`ISO 10555-1:1995(E)
`INTERNATIONAL STANDARD © ISO
`
`
`Sterile, single-use intravascular catheters —
`
`Part 1:
`
`General requirements
`
`1 Scope
`
`This part of ISO 10555 specifies general requirements
`for
`intravascular catheters,
`supplied in the sterile
`condition and intended for single use, for any appli-
`cation.
`
`It does not apply to intravascular catheter accessories,
`which will be covered by a separate standard.
`
`ISO 7886—1:1993, Sterile hypodermic syringes for sin-
`gle use — Part 1: Syringes for manual use.
`
`3 Definitions
`
`For the purposes of this part of ISO 10555, the fol-
`lowing definitions apply.
`
`intravascular catheter: Tubular device, single
`3.1
`or multilumen, designed to be partially or totally in-
`serted or implanted into the cardiovascular system for
`diagnostic and/or therapeutic purposes.
`
`2 Normative references
`
`3.2 distal end: End of the catheter inserted furthest
`
`The following standards contain provisions which,
`through reference in this text, constitute provisions
`of this part of ISO 10555. At the time of publication,
`the editions indicated were valid. All standards are
`
`subject to revision, and parties to agreements based
`on this part of ISO 10555 are encouraged to investi-
`gate the possibility of applying the most recent edi-
`tions of the standards indicated below. Members of
`
`IEC and ISO maintain registers of currently valid
`International Standards.
`
`ISO 594-1 :1986, Conical fittings with a 6 % (Luer) ta-
`per for syringes, needles and certain other medical
`equipment — Part 7: General requirements.
`
`ISO 594-221991, Conical fittings with a 6 % (Luer) ta-
`per for syringes, needles and certain other medical
`equipment — Part 2: Lock fittings.
`
`into the patient.
`
`3.3 proximal end; access end: End of the catheter
`to which connection can be made.
`
`3.4 hub: Connectoris) at the proximal end of the
`catheter which may either be integral with the
`catheter or be capable of being securely fitted to the
`proximal end of the catheter.
`
`3.5 effective length, 1: Length of the catheter that
`can be inserted into the body. (See figure 1 .)
`
`3.6 outside diameter: Maximum diameter of that
`
`part of the catheter that can be inserted into the
`vessel.
`
`junction: That portion of the catheter that joins
`3.7
`one tube to multiple tubes.
`
`d
`
`4
`
`

`

`ISO 10555-1:1995(E)
`
`© ISO
`
`
`
`
`
`
`
`
`
`Key
`
`1 = effective length
`1.
`catheter hub
`
`2.
`
`3.
`
`4.
`
`catheter strain reinforcement
`
`length mark
`
`junction
`
`Figure 1 — Examples of effective length of catheters
`
`5
`
`

`

`© ISO
`
`ISO 10555-1:1995(E)
`
`Table 1 — Force at break of catheter test pieces
`
`Smallest outside
`diameter of tubular
`
`portion of test piece
`
`mm
`
`Minimum force at break
`
`
`
`4 Requirements
`
`4.1 General
`
`The catheter shall have been sterilized by a validated
`method, and shall comply with 4.2 to 4.7 in the sterile
`condition.
`
`ISO 11135 and ISO 11137 for
`See ISO 11134,
`NOTE1
`appropriate methods of sterilization.
`
`4.2 Biocompatibility
`
`The catheter shall be free from biological hazard.
`
`See ISO 10993-1 for the selection of appropriate
`NOTE 2
`test methods.
`
`4.3 Surface
`
`When examined by normal or corrected to normal
`vision with x 2,5 magnification, the external surface
`of the effective length of the catheter shall appear
`free from extraneous matter.
`
`The external surface of the effective length of the
`catheter, including the distal end, should be free from
`process and surface defects and should cause mini-
`mum trauma to vessels during use.
`
`If the catheter is lubricated, the lubricant should not
`be visible as drops of fluid on the external surface
`when the catheter is examined under normal or cor-
`rected to normal vision.
`
`4.4 Corrosion resistance
`
`When tested in accordance with the method given in
`annex A, metallic components of the catheter shall
`show no signs of corrosion.
`
`4.5 Force at break
`
`When tested in accordance with the method given in
`annex B, the force at break of each test piece shall
`be as given in table 1.
`
`4.6 Freedom from leakage
`
`4.6.1 The hub or connection fitting assembly or any
`other part of the catheter shall not leak liquid when
`tested in accordance with the method given in
`annex C.
`
`NOTE -— This part of ISO 10555 does not specify re-
`quirements for force at break for tubing of less than
`0,55 mm outside diameter.
`
`4.7 Hubs
`
`If the catheter is supplied with either an integral or a
`separate hub, it shall be a female hub and shall com-
`ply with ISO 594-1 and ISO 594—2.
`
`5 Designation of nominal size
`
`The nominal size of the catheter shall be designated
`as specified in 5.1 and 5.2.
`
`5.1 Outside diameter
`
`Unless otherwise specified in the International Stan-
`dard for a particular type of catheter, the outside di-
`ameter shall be expressed in millimetres,
`rounded
`upwards to the nearest 0,05 mm for outside diam-
`eters of less than 2 mm, or to the nearest 0,1 mm for
`outside diameters of 2 mm and greater.
`
`5.2 Effective length
`
`The effective length shall be expressed in a whole
`number of millimetres for effective lengths of less
`than 99 mm and in either a whole number of milli-
`metres or a whole number of centimetres for effec-
`
`tive lengths of 99 mm and greater.
`
`Information to be supplied by
`6
`manufacturer
`
`4.6.2 Air shall not leak into the hub assembly during
`aspiration when tested in accordance with the
`method given in annex D.
`
`The manufacturer shall supply at least the following
`information. All dimensions given shall be expressed
`in SI units of measurement.
`
`6
`
`

`

`ISO 10555-1:1995(E)
`
`© ISO
`
`Units of other measurement systems may additionally
`be used.
`
`g)
`
`any special storage and handling instructions;
`
`a) description of the product;
`
`bl outside diameter;
`
`0)
`
`effective length;
`
`h)
`
`indication of sterility;
`
`i) method of sterilization;
`
`j)
`
`indication for single use;
`
`d) name or tradename and address of manufacturer;
`
`e)
`
`lot designation;
`
`1‘)
`
`expiry date or use by date;
`
`k)
`
`incompati-
`any known chemical and/or physical
`bilities with substances likely to be used with the
`catheter;
`
`l)
`
`instructions for use and warnings, as appropriate.
`
`7
`
`

`

`© I50
`
`I80 10555-1:1995(E)
`
`Annex A
`
`(normative)
`
`Test method for corrosion resistance
`
`A.1 Principle
`
`The catheter is immersed in sodium chloride solution,
`
`then in boiling distilled water, and afterwards exam—
`ined visually for evidence of corrosion.
`
`A.2 Reagents
`
`solution of
`A.2.1 Saline solution, comprising a
`analytical reagent grade sodium chloride in freshly
`prepared distilled water [C(NaCI) = 0,15 mol/I].
`
`A.2.2 Distilled or deionized water.
`
`in boiling
`move the test specimen and immerse it
`distilled water (A22) for 30 min. Allow the water and
`the test specimen to cool to 37 ”C, and maintain them
`at this temperature for 48 h. Remove the test speci-
`men and allow it to dry at room temperature. Disas—
`semble
`specimens
`that
`have
`two
`or more
`components which are intended to be separable in
`use. Do not strip away or cut open any coatings on
`metallic components. Inspect the specimen visually
`for signs of corrosion.
`
`A.3 Apparatus
`
`A.5 Test report
`
`A.3.1 Borosilicate glass beakers.
`
`The test report shall include the following information:
`
`A.4 Procedure
`
`a)
`
`identity of the catheter;
`
`Immerse the catheter in the saline solution (A.2.1) in
`
`b)
`
`a glass beaker (A3) at room temperature for 5 h. Re-
`
`statement as to whether corrosion occurred dur-
`ing the test.
`
`8
`
`

`

`ISO 10555-1:1995(E)
`
`© ISO
`
`Annex B
`
`(normative)
`
`Method for determining force at break
`
`B.1 Principle
`
`Test pieces of a catheter are chosen so that each
`tubular portion, each junction between hub or con-
`nector and tubing, and each junction between tubular
`portions is tested. A tensile force is applied to each
`test piece until the tubing breaks or the junction sep-
`arates.
`
`in
`the applied tensile force,
`Note the value of
`newtons, at which separation occurs, and record this
`value as the force at break.
`
`3.3.6 If testing a catheter that consists of a single
`tubular portion having regions of different outside di—
`ameter, repeat 8.3.2 to 8.3.5 on test pieces of each
`different diameter.
`
`3.2 Apparatus
`
`If testing a catheter that has a sidearm or
`3.3.7
`sidearms,
`
`8.2.1 Tensile testing apparatus, capable of exert-
`ing a force of greater than 15 N.
`
`a)
`
`repeat 8.3.2 to 8.3.5 on each sidearm;
`
`8.3 Procedure
`
`B.3.1 Assemble the catheter in accordance with the
`
`manufacturer's instructions. Select a test piece from
`the catheter to be tested. Include in the test piece the
`hub or connector,
`if present, and the junction be-
`tween segments, eg. between the tubing and the tip,
`if present. Exclude distal
`tips of
`lengths less than
`3 mm from the test piece.
`
`332 Condition the test pieces in an atmosphere
`of 100 % relative humidity or water and a temperature
`of (37 i 2) °C for 2 h. Test
`immediately after con—
`ditioning.
`
`Fix the test piece in the tensile testing ap-
`3.3.3
`If a hub or connector is present, use an ap-
`paratus.
`propriate fixture to avoid deforming the hub or
`connector.
`
`B.3.4 Measure the gauge length of the test piece,
`i.e.
`the distance between the jaws of the tensile
`testing apparatus or the distance between the hub or
`connector and the jaw holding the other end of the
`test piece, as appropriate.
`
`B.3.5 Apply a tensile strain at a unit strain rate of
`20 mm/min/mm of gauge length (see table 8.1) until
`the test piece separates into two or more pieces.
`
`b)
`
`repeat 8.3.2 to 8.3.5 on a test piece that includes
`the joint between a sidearm and the adjacent part
`of that portion of the catheter intended to be in-
`troduced into the body;
`
`c)
`
`repeat 83.7 b) for each joint.
`
`3.3.8 Do not perform more than one test on any
`test piece.
`
`Table 3.1 -— Examples of conditions for
`20 mm/min/mm strain rate
`
`Gauge length
`
`Test speed
`
`mm
`
`mm/min
`
`
`
`3.4 Test report
`
`The test report shall include the following information:
`
`a)
`
`identity of the catheter;
`
`b)
`
`the force at break, in newtons, and outside diam—
`eter of each test piece.
`
`9
`
`

`

`© ISO
`
`ISO 10555-1 :1 995(El
`
`Annex C
`(normative)
`
`Test method for liquid leakage under pressure
`
`0.1 Principle
`
`The catheter is connected, via a leakproof connection,
`to a syringe. A hydraulic pressure is applied to the
`catheter and to the hub assembly, if present, and the
`catheter tube inspected for leakage.
`
`the hub to the reference fitting (0.3.1) by
`nect
`applying an axial force of 27,5 N for 5 5 while applying
`a twisting action to a value of torque not exceeding
`0,1 N-m to give rotation not exceeding 90°, both
`components being dry. Connect the reference fitting
`(0.3.1) via
`the connector
`(0.3.2)
`to the syringe
`(0.3.4).
`
`0.2 Reagent
`
`0.2.1 Distilled or deionized water.
`
`0.3 Apparatus
`
`0.3.1 Reference steel fitting, having a male 6 %
`(Luer) taper as specified in ISO 594—1.
`
`to connect reference
`0.3.2 Leakproof connector,
`fitting (0.3.1) to syringe (0.3.4), fitted with a gauge
`capable of measuring up to 350 kPa pressure and
`having a small internal volume.
`
`0.3.3 Connector, to make leakproof connection be-
`tween syringe (0.3.4) and catheters which do not
`have hubs.
`
`0.3.4 10 ml syringe, which has passed the tests for
`leakage past the piston and nozzle as specified in
`ISO 7886-1.
`
`0.4.2 When testing catheters which do not have
`hubs, connect the catheter to the syringe (0.3.4) by
`means of a connector (0.3.3).
`
`at
`(0.2)
`syringe with water
`the
`0.4.3 Fill
`(22 i 2) °C and expel the air. Adjust the volume of
`water in the syringe to the nominal graduated ca-
`pacity. Occlude (0.3.5) the test specimen as near the
`distal end as possible.
`
`0.4.4 Position the apparatus so that the axis of the
`connection between syringe and catheter is horizon-
`tal. Apply an axial force to the syringe so that a
`pressure of 300 kPa to 320 kPa is generated by the
`relative action of the piston and barrel. Maintain the
`pressure for 30 3. Examine the hub assembly,
`if
`present, and catheter tube for liquid leakage, i.e. the
`formation of one or more falling drops of water, and
`record whether or not leakage occurs.
`
`0.3.5 Means for occluding test specimen, eg. a
`clamp.
`
`0.5 Test report
`
`0.4 Procedure
`
`The test report shall include the following information:
`
`0.4.1 When testing catheters which have a hub or
`hubs,
`if necessary assemble detachable hubs in ac-
`cordance with the manufacturer's instructions. Con—
`
`a)
`
`identity of the catheter;
`
`b)
`
`statement as to whether leakage occurred from
`the hub assembly, if present, or catheter tube.
`
`10
`
`10
`
`

`

`ISO 10555-1z1995iE)
`
`© ISO
`
`Annex D
`
`(normative)
`
`Test method for air leakage into hub assembly during aspiration
`
`D.1 Principle
`
`The hubls) of the catheter is (are) connected via a
`reference male conical
`fitting,
`to a partially filled
`syringe. A reduced pressure is applied to the interface
`of the hub and the reference fitting by withdrawing
`the syringe plunger, and visual inspection made for
`the ingress of air bubbles to the syringe.
`
`D.2 Reagent
`
`D.2.1 De-aerated distilled water or de-aerated
`deionized water.
`
`D.3 Apparatus
`
`D.3.1 Reference steel fitting, as specified in (3.3.1.
`
`D.3.2 Leakproof connector, as specified in C.3.2,
`but without tapping and pressure gauge.
`
`D.3.3 Syringe, as specified in C.3.4.
`
`D.3.4 Means for occluding test specimen, e.g. a
`clamp.
`
`0,1 N-m to give rotation not exceeding 90°, both
`components being dry.
`
`D.4.2 Connect the reference fitting (D.3.1) via the
`connector (D.3.2) to the syringe (D.3.3). Seal all valves
`that are intended to open during aspiration.
`
`through the test
`D.4.3 Draw into the syringe,
`specimen and reference fitting, a volume of water
`(D.2) at (22 i 2) °C exceeding 25 % of the graduated
`capacity
`of
`the
`syringe. Avoid wetting
`the
`hub/reference fitting union.
`
`D.4.4 Expel the air from the apparatus except for a
`small air bubble. Adjust the volume of the water in the
`syringe to 25 % of the graduated capacity. Occlude
`(D.3.4) the test specimen as close as practicable to
`the hub.
`
`D.4.5 With the nozzle of the syringe downward,
`withdraw the plunger to the maximum graduated ca-
`pacity mark. Hold for 15 s and examine the water in
`the syringe for the formation of air bubbles,
`ignoring
`bubbles formed during the first 5 3.
`
`D.4 Procedure
`
`D5 Test report
`
`in accordance
`D.4.1 Assemble detachable hubs
`with the manufacturer's instructions. Connect the hub
`
`to be tested to the reference fitting (D.3.1) by applying
`an axial force of 27,5 N for 53 while applying a
`twisting action to a value of torque not exceeding
`
`The test report shall include the following information:
`
`a)
`
`identity of the catheter;
`
`b)
`
`statement as to whether leakage of air occurred
`from the hub assembly after the first 5 s.
`
`11
`
`11
`
`

`

`© ISO
`
`ISO 10555-1:1995(E)
`
`Annex E
`
`(informative)
`
`Bibliography
`
`of
`evaluation
`Biological
`[1] ISO 1099311992,
`medical devices — Part 1: Guidance on selec-
`tion of tests.
`
`[3] ISO 11135:1994, Medical devices — Validation
`and
`routine
`control
`of
`ethylene
`oxide
`sterilization.
`
`[2] ISO 11134:1994, Sterilization of health care
`products — Requirements for validation and
`routine
`control — Industrial moist heat
`sterilization.
`
`[4] ISO 11137:1995, Sterilization of health care
`products — Requirements for validation and
`routine control — Radiation sterilization.
`
`12
`
`12
`
`

`

`ISO 10555-1:1995(E)
`
`© ISO
`
`
`
`ICS 11.040.20
`
`sterile equipment, disposable equipment, vascular system, catheters,
`Descriptors: medical equipment,
`designation, consumer information.
`Price based on 9 pages
`
`specifications,
`
`tests,
`
`13
`
`13
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket