throbber
View Article Online / Journal Homepage / Table of Contents for this issue
`
`Integrated flow-cells for novel adjustable sheath flows
`
`J. H. Nieuwenhuis,a J. Bastemeijer,b P. M. Sarroc and M. J. Vellekoopa
`
`a Industrial Sensor Systems, Institute IEMW, Vienna University of Technology, Austria
`b Electronic Instrumentation Lab. – DIMES, Delft University of Technology, The Netherlands
`c Laboratory of ECTM – DIMES, Delft University of Technology, The Netherlands
`
`Received 30th October 2002, Accepted 26th February 2003
`First published as an Advance Article on the web 13th March 2003
`
`In this paper two integrated flow-cells are presented that can generate novel sheath flows. The flow-cells allow for
`dynamic orthogonal control of the sample flow dimensions. In addition to this, the sample flow can be freely
`positioned inside the channel. The flow-cells are attractive, because they are very simple to fabricate and are
`compatible with the integration of sensors. Experiments have been carried out demonstrating that the sample flow
`dimensions can be controlled over a wide range; also the results show good agreement with finite element
`simulation results.
`
`Introduction
`
`In this paper two integrated flow-cells are presented that can
`generate novel sheath flows. The flow-cells presented here
`allow dynamic orthogonal control of the sample flow dimen-
`sions. Also the sample flow can be freely positioned inside the
`flow-channel so that the sample flow touches or nearly touches
`one side of the channel; this assures a good contact with any
`integrated sensor interface. Finally, another attractive feature of
`the flow-cells presented here is that they are very simple to
`fabricate, consisting of a two layer structure and requiring only
`2 etch-steps to fabricate.
`Due to the limited interaction between the sample flow and
`sheath flow the sample flow can be considered a virtual flow
`channel with adaptable dimensions (e.g. diameter 10% of the
`physical flow-channel). As a result the presented sheath flows
`combine the advantages of a large diameter channel with those
`of a small diameter channel: due to the small dimensions of the
`sample flow the low detection limits and modest sample
`consumption of a small channel device can be achieved; at the
`same time the large physical dimensions of the channel alleviate
`many problems such as clogging, air bubbles and strict
`fabrication tolerances. Most of these advantages are obvious,
`but the final advantage might need some further explanation.
`The imperfections in the physical flow channel are smoothed
`out by the sheath liquid, which is in direct contact with the wall.
`The sample flow, which is screened from the wall by the sheath
`liquid, will not be disturbed as much. Therefore, the imperfec-
`tions of the channel are not as critical as without the application
`of a sample flow.
`Because of the advantages mentioned above, a number of
`realisations of sheath flow on a chip have been presented in the
`literature. Most of them are limited to a semi-sheath flow,
`consisting of a layered-flow configuration. Here the sample
`liquid is confined in one dimension between the sheath liquid
`and the wall of the physical channel1,2 or between two layers of
`sheath liquid.3,4,5 In the other dimension the sample flow is
`fixed by the physical channel. Therefore these are not sheath
`flows in the classical sense. However, a few flow-cells have
`been presented where a classical, coaxial sheath flow was
`realised on chip (in a coaxial sheath flow the sheath liquid
`surrounds the sample flow completely). One of the first
`realisations consists of a miniaturised cytometer configuration,
`which resulted in a rather complex 5-layer device.6 More
`recently less complicated devices were demonstrated where a
`special ‘U-shaped’ sample inlet7 or a circular ‘chimney-shaped’
`
`56
`
`Lab Chip, 2003, 3, 56–61
`
`sample inlet8 was used to lift the sample liquid from the bottom
`of the channel. In these flow-cells the shape of the sample flow
`is fixed by the geometry of the flow-cell and therefore these
`flow-cells offer little control over the sample flow dimensions.
`The flow-cells described above use a sheath liquid to confine
`the sample liquid, but also a rare example is known where a gas
`was used to confine the sample flow.9 The main problem with
`this device is that special measures are required, such as
`controlling the surface properties of the channel walls, to create
`a stable flow.
`The flow-cells presented in this paper distinguish themselves
`from the flow-cells known in the literature by their versatility.
`Both layered sheath flows and coaxial sheath flows can be
`realised within the same device. What is even more important,
`due to a number of orthogonal control mechanisms both the
`dimensions and the position of the sample flow can accurately
`be controlled. As a consequence the flow-cells can create an
`optimal sample flow for each specific application. And because
`they can be fabricated by adding a few post-processing steps to
`a standard IC-process, many known sensors can be integrated
`into the device.
`Some examples of applications are: an integrated Coulter
`counter,10 sensors for particle shape analysis11 and sensors for
`cell analysis and (bio)chemical analysis.
`
`Sheath flows for microfluidics systems
`
`Traditionally, sheath flows have been applied in cytometers12 to
`confine the sample liquid into a very narrow sample stream for
`optical analysis, mainly by fluorescence and light scattering
`techniques. By integrating the flow-channel on a chip it has
`become possible to include the flow-channel and the sensors in
`the same device. The flexibility of the flow-cells presented here
`allows optimisation of the sample flow dimensions and sample
`flow position to the integrated sensor. For example, for an
`optical evanescent wave sensor an optimal sample flow would
`have a small height and would be positioned on the same side as
`the sensor interface,2 whereas for an integrated Coulter counter
`a sample flow with equal horizontal and vertical dimensions
`will be optimal.10
`Due to the small channel dimensions of microfluidics
`systems the Reynolds number in these devices is usually very
`low. As a consequence the flow behaviour in these systems is
`typically completely laminar and no turbulence occurs. There-
`
`DOI: 10.1039/b210724d
`
`This journal is © The Royal Society of Chemistry 2003
`ABS Global, Inc. and Genus plc – Ex. 1014, p. 56
`
`
`
`Published on 13 March 2003. Downloaded by University of California - Los Angeles on 04/10/2015 20:18:06.
`
`

`

`View Article Online
`
`the pictures from top to bottom the vertical sample flow control
`is shown.
`The horizontal dimensions of the sample flow are controlled
`by two horizontal control ports that are located on the sides of
`the flow-channel, downstream of the sample inlet (see Fig. 1).
`By adding or removing sheath liquid through these control ports
`at an equal rate the already present sheath flow is horizontally
`compressed or expanded respectively which leads to a narrow
`or wider sample flow. In Fig. 2 from left to right this effect is
`shown. Notice that the height of the sample flow is not affected
`by this control mechanism.
`
`Flow-cell 2: additional control of sample flow
`position
`
`Flow-cell 2 looks quite similar to flow-cell 1, but there is one
`main difference. Flow-cell 2 has an additional inlet, located in
`between the focussing section and the sample inlet (see Fig. 3).
`This additional inlet gives flow-cell 2 the added functionality to
`freely position the sample flow anywhere inside the flow-
`channel. Two control mechanisms are required to achieve
`this.
`The vertical position of the sample flow is controlled by the
`additional vertical control inlet. When sheath liquid is added
`through this inlet the entire sheath flow in the channel is lifted
`up from the channel bottom. As a result a coaxial sheath flow is
`formed that no longer has any contact with the channel bottom.
`The more liquid is added through this inlet the higher the sample
`flow is positioned. This vertical position control inlet has a
`narrow shape (625 μm 3 50 μm) to create a flow profile through
`this inlet that is as uniform as possible. This means that the
`shape of the sample flow is hardly influenced by the vertical
`position control, except for some vertical compression. Vertical
`position control of the sample flow is demonstrated in Fig. 4 in
`the pictures from top to bottom.
`The same horizontal control ports that are used to control the
`width of the sample flow can also be used to control its position.
`In this type of operation the direction of flow through both inlets
`is now opposite. By adding sheath liquid through one of the
`control ports and removing it from the other inlet at the same
`flow-rate the sample flow is shifted in the horizontal plane. The
`results can be seen in Fig. 4 in the pictures from left to right.
`Apart from the additional inlet the configuration of the inlets
`of flow-cell 2 is similar to that of flow-cell 1, therefore the
`sample dimension control mechanisms described for that flow-
`cell work for flow-cell 2 as well. The coaxial flow that can be
`achieved with this flow-cell is very suitable for sensors that do
`not require any contact with the sample liquid and especially
`those sensors in which the interface might get polluted such as
`optical sensors.
`
`fore, the main interaction between liquids in a microfluidics
`system is by diffusion,13 which is a relative slow process. The
`laminar flow behaviour and the limited interaction between
`different liquids make microfluidics systems very suitable to be
`used with sheath flow. In the following sections two flow-cells
`using sheath flows are described that have been designed using
`finite element modelling.
`
`Flow-cell 1: control of sample flow dimensions
`
`Flow-cell 1 has been developed to allow dynamic control of the
`dimensions of the sample flow (see Fig. 1). In this flow-cell a
`non-coaxial sheath flow is formed by vertically injecting a
`sample liquid into a channel through which sheath liquid is
`flowing. By hydrodynamic focussing a smooth flow of sample
`liquid is formed that still touches the bottom of the channel. A
`focussing section brings the channel width down from 625 μm
`to a width of 160 μm. The application of such a focussing
`section allows the use of fairly large inlets which are convenient
`in handling the liquid connections to the chip. The less critical
`alignment of the inlets and the lower pressure drop over the
`wider sections are additional advantages. The non-coaxial type
`of sheath flow that is formed with this flow-cell is suitable for
`sensors that require contact with the sample liquid such as
`impedance sensors. The dimensions of the sample flow can
`dynamically be adapted by 2 orthogonal control mechanisms.
`The vertical dimensions of the sample flow are controlled by
`the relative flow-rate at which the sample liquid is injected in
`relation to the flow-rate of the sheath flow. At higher relative
`flow-rates of the sample liquid, it penetrates further into the
`sheath liquid thereby increasing the sample flow height.
`Lowering the relative flow-rate of the sample liquid will result
`in a sample flow with less height. In Fig. 2 simulation results are
`shown, that were obtained using the model shown in Fig. 1. In
`
`Fig. 1 Model of flow-cell 1, this flow-cell creates a non-coaxial sheath
`flow and allows control of the sample flow dimensions; here the control
`ports are used to widen the sample flow.
`
`Published on 13 March 2003. Downloaded by University of California - Los Angeles on 04/10/2015 20:18:06.
`
`Fig. 2 Overview of the control of the sample flow dimensions in a cross-
`section of the channel.
`
`Fig. 3 Model of flow-cell 2, the additional vertical position inlet of this
`flow-cell also allows positioning of the sample flow in the channel; here the
`control ports are used to position the sample flow vertically in the centre of
`the channel with a horizontal shift.
`
`Lab Chip, 2003, 3, 56–61
`ABS Global, Inc. and Genus plc – Ex. 1014, p. 57
`
`57
`
`

`

`View Article Online
`
`Due to the planar nature of the device it is only possible to
`analyse it from the top. Performing measurements on the width
`of the sample flow is straightforward in this configuration, but
`analysing the vertical dimensions of the sample flow is more
`complicated. However by analysing the intensity of the dye,
`also quantitative information about the sample height can be
`obtained. Before each series of experiments photographs were
`taken of the channel without any dye present and photographs
`were taken when the channel was completely filled with the dye.
`After the experiment the height of the sample flow was
`calculated by regression between these two references. To
`obtain a reliable result each measurement point was taken from
`the average of three photographs. The noise was reduced by
`averaging the intensity of the dye over 100 adjacent image lines.
`The area of the photographs that was used in the experiments is
`marked with the dotted white squares in Fig. 7, 9 and 11.
`Unfortunately analysing the vertical position of the sample flow
`from the top is not possible with this method; therefore it was
`decided to base the experiments only on flow-cell 1.
`The finite element simulations were performed with the
`Netflow-module of the finite element package Coventorware.
`Detailed models of up to 100 000 elements with a well designed
`grid were required to obtain a high-enough accuracy. Besides
`the geometry of the model and the flow-rates also the diffusion
`constant of the dye was included in the simulations. From the
`simulation results the height of the sample flow could easily be
`obtained by just summing the concentrations of the sample
`distribution along the vertical axis for all data points in the
`measurement area. Finally the experimental results and the
`simulation results were compared by plotting them in the same
`graph.
`
`Fabrication of the flow-cells
`
`Both types of flow-cells can be fabricated using the same,
`relatively simple process. In a glass wafer by isotropic etching
`the channel is defined with a depth of 100 μm and a minimum
`width of 160 μm. In a silicon wafer by isotropic etching
`through-holes are defined that form the liquid inlets of the
`device. The glass wafer and the silicon wafer are then anodically
`bonded together to form the complete devices. A photograph of
`a flow-cell chip is depicted in Fig. 5.
`There are a number of reasons to make the devices like this.
`Firstly, the dimensions of the sample-inlet need to be accurate
`and etching of silicon can be easier controlled than glass
`etching. Secondly, with the liquid inlets in the silicon part the
`glass side of the device is still available for optical inspection
`during operation. Finally, since the channel was etched in the
`glass wafer the surface of the silicon wafer is still smooth and
`suitable for future integration of sensors to form a complete
`integrated analysis system.
`
`Experimental verification
`
`A series of experiments was carried out to verify the simulation
`results. The measurement setup is shown in Fig. 6. The chip
`with the flow-cell is placed in a custom made aluminium holder
`which is used to make the connections to the tubing. The liquid
`connections to the chip are sealed by O-rings. Two syringe
`pumps (kdScientific model 200 series) are used to control the
`flow-rates at the inlets. The flow-cell is positioned underneath a
`microscope (Zeiss Stemi SV11) and a diluted red dye (standard
`plotter dye) is used as a sample liquid. The sheath liquid is
`purified water. A digital still camera (Sony Cybershot DSC-75)
`captures images of the flow-cell for quantitative analysis and
`comparison with the simulation results.
`
`Published on 13 March 2003. Downloaded by University of California - Los Angeles on 04/10/2015 20:18:06.
`
`Fig. 4 Overview of the position control of the sample flow in a cross-
`section of the channel.
`
`Fig. 6 Overview of the experiment set-up.
`
`Fig. 5 The flow-cell chip with dimensions of 2 cm by 1 5 cm (the
`electrical contacts are not used in the experiments).
`
`Fig. 7 Illustrative results from the diffusion experiment.
`
`58
`
`Lab Chip, 2003, 3, 56–61
`
`ABS Global, Inc. and Genus plc – Ex. 1014, p. 58
`
`

`

`View Article Online
`
`In Fig. 9 it can be seen that the width of the sample is constant
`for relative flow-rates of the sample flow up to 10%, at the
`highest flow-rate the sample flow becomes slightly wider. For
`the height control of the sample flow there is a good match
`between the finite element simulations and the experimental
`results. This means that the model can be used to reliably predict
`the vertical control of the sample flow. The slight mismatch at
`the highest vertical dimension is due to saturation of the camera
`for the red colour.
`
`Horizontal control of sample flow dimensions
`
`In the second control experiment the horizontal dimensions of
`the sample flow were controlled. During this experiment the
`sheath liquid and sample liquid flow-rates were kept constant at
`10 μl min21 and 1 μl min21 respectively. The flow-rate through
`each control port was varied from 23 to +25 μl min21. The dye
`was measured at the same location as in the previous
`experiments; here it was even more important to have a visible
`area that is maximally wide. Some illustrative results are
`depicted in Fig. 11 and the quantitative results are shown in Fig.
`12.
`The results in Fig. 12 show that the width of the sample flow
`can be controlled over a wide range: for the flow-rates used in
`the experiment the width of the sample flow varies from 40 to
`
`Fig. 9 Illustrative results from the sample height control experiment.
`
`Sample diffusion
`
`First a reference experiment was carried out to determine the
`influence of diffusion and to see how well it can be modelled. In
`this experiment the ratios of the flow-rate of the sample liquid,
`the flow-rate through the control ports and the flow-rate of the
`sheath-liquid were kept constant at 1+5+10. During the
`experiment the total flow-rate was varied from 1 to 50 μl min21.
`The intensity of the dye was measured in the wider section
`following the narrow section (see Fig. 7), since in this location
`the visible area is not blocked so much by the rounded corners
`of the isotropic etching of the channel. A detailed finite element
`model similar to the one in Fig. 1 was made, only the model
`used here has a wide section following the narrow section so
`that the quantitative data from the simulations can be obtained
`in the same location as in the experiment. In Fig. 7 some
`illustrative results of the experiments are shown.
`In Fig. 8 the quantitative results of both the simulation and the
`experiment are shown. They show that there is a very good
`agreement between experimental and simulation results. The
`entire shapes of the dye distributions match very well. This
`means that the finite element model represents the real flow-cell
`very well and that this model can be used for the further
`experiments. The influence of diffusion is realistically taken
`into account in the experiments and for a flow-rate higher than
`10 μl min21 the influence of diffusion is small, so measure-
`ments can be carried out at this flow-rate.
`
`Vertical control of sample flow dimensions
`
`In the first control experiment the vertical dimensions of the
`sample flow were controlled. Since for this measurement the
`width control of the sample was not necessary, a device without
`horizontal control ports was used similar to flow-cell 1. During
`the experiment the flow-rate of the sheath liquid was kept
`constant at 10 μl min21 and six different flow-rates for the
`sample liquid were applied in a range from 0.05 to 2 μl min21,
`which correspond to relative sample flow-rates of 0.5% to 20%
`of the sheath flow-rate. Again the intensity of the dye was
`measured in the wide section of the channel, downstream of the
`narrow section. Some illustrative results are depicted in Fig. 9
`and the quantitative results are shown in Fig. 10.
`The results in Fig. 10 show that the height of the sample can
`be controlled over a wide range. For the relative flow-rates used
`in the experiment the height of the sample flow can be
`controlled from 3 to 60 μm in a channel with a depth of 100 μm.
`
`Published on 13 March 2003. Downloaded by University of California - Los Angeles on 04/10/2015 20:18:06.
`
`Fig. 8 Quantitative comparison of the simulation (blue) and the experimental (red) results for the diffusion experiment, for sheath flow-rates of 1, 2.5, 5,
`10, 25 and 50 μl min21.
`
`Lab Chip, 2003, 3, 56–61
`ABS Global, Inc. and Genus plc – Ex. 1014, p. 59
`
`59
`
`

`

`View Article Online
`
`292 μm in the wide part of the channel, which has a width of 625
`μm. As a result the width of the sample flow in the narrow
`section where the sensor should be positioned (width 160 μm)
`is varied in the range from 10 to 73 μm. The width of the sample
`flow is defined here as the width of the sample at 50% of the
`maximum dye intensity. So also for the horizontal control of the
`sample flow dimension the results show a good match between
`simulation and experimental results.
`
`Sample flow splitting and switching
`
`The flow-cell used in the experiment turned out to be a very
`flexible device. Besides the horizontal and vertical control of
`the sample also other useful functions can be realised. The first
`example is sample splitting. For this application only one
`control port is used. Through this inlet 50% of the total flow of
`liquid is removed. This results in an equal splitting of the sample
`(see Fig. 13). By removing more or less liquid through the
`control port an unequal splitting of the sample is also possible of
`course.
`A second example of the flexibility of the flow-cell is sample
`switching. Here again only one control port needs to be active.
`
`This time the liquid is removed through the control port at a
`flow-rate that equals the flow-rate of the total flow of liquid. As
`a result the complete sheath flow including the sample liquid is
`switched (see Fig. 14). When a coaxial sheath flow is used such
`as can be generated with flow-cell 2 this is even possible
`without the sample liquid touching any wall of the channel.
`
`Conclusions
`
`Two new flow-cells have been developed that are very versatile.
`The flow-cells can generate both non-coaxial and coaxial sheath
`flows. Using an orthogonal control mechanism the flow-cells
`allow dynamic control of the sample flow dimensions. With one
`additional vertical position inlet the sample flow can also be
`freely positioned inside the channel. The processing of the flow-
`cells is very simple, and can be easily combined with the
`integration of sensors to form a complete integrated analysis
`system.
`Experiments were carried out with one of the flow-cells and
`a comparison of experimental results and simulation results
`shows that the complicated flow behaviour of the device can be
`modelled very accurately. Also the influence of diffusion is
`
`Fig. 10 Comparison of the simulation (Finite Element Modelling, FEM)
`results and the experimental (Dye) results for the sample height control
`experiment.
`
`Fig. 12 Comparison of the simulation (Finite Element Modelling, FEM)
`results and the experimental (Dye) results for the sample width control
`experiment.
`
`Published on 13 March 2003. Downloaded by University of California - Los Angeles on 04/10/2015 20:18:06.
`
`Fig. 11 Illustrative results from the sample width control experiment.
`
`60
`
`Lab Chip, 2003, 3, 56–61
`
`ABS Global, Inc. and Genus plc – Ex. 1014, p. 60
`
`

`

`View Article Online
`
`The authors also would like to thank Dr. A. Bossche and Mr. P.
`Turmezei of the Electronic Instrumentation Laboratory, Delft
`University of Technology for fruitful cooperation and useful
`discussions. Finally, the authors thank the Hochschuljubiläums-
`stiftung der Stadt Wien for financial support.
`
`References
`
`1 M. Tidare, H. Roos and U. Lindberg, Hydrodynamic Addressing; A
`Flow Cell for Improved Signal Quality in Bioanalytical Systems, , in
`Proc. of Transducers ’01, Munich, Germany, June 10–14, 2001, pp.
`1154–1157.
`2 O. Hofmann, P. Niedermann and A. Manz, Modular Approach to
`Fabrication of Three-Dimensional Microchannel Systems in PDMS –
`Application to Sheath Flow Microchips, Lab Chip, 2001, 1(2),
`108–114.
`3 U. D. Larsen, G. Blankenstein and J. Branebjerg, Microchip Coulter
`particle counter, , in Proc. of Transducers ’97, Chicago, June 16–19,
`1997, pp. 1319–1322.
`4 K. Roberts, M. Parameswaran, M. Moore and R. S. Muller, A Silicon
`Microfabricated Aperture for Counting Cells using the Aperture
`Impedance Technique, , Proc. of 1999 IEEE Canadian Conference on
`Electrical and Computer Engineering, Edmonton, Alberta, Canada,
`May 9–12, 1999, pp. 1668–1673.
`5 G. Lee, B. Hwei and G. Huang, Microfluidic Chips with MxN
`Continuous Sample Introduction Function Using Hydrodynamic
`Flow Switching, , in Proc. of Transducers ’01, Munich, Germany,
`June 10–14, 2001, pp. 1158–1161.
`6 R. Miyake, H. Ohki, I. Yamazaki and R. Yabe, A Development of
`Micro Sheath Flow Chamber, , in Proc. of MEMS ’91, 1991, pp.
`265–270.
`7 K. Tashiro, T. Sekiguchi, S. Shoji, T. Funatsu, W. Masumoto and H.
`Sato, Design and Simulation of Particles and Biomolecules Handling
`Micro Flow Cells with Three Dimensional Sheath Flow, , in Proc. of
`μTAS 2000, Enschede, The Netherlands, May 14–18, 2000, pp.
`209–212.
`8 A. Wolff, U. D. Larsen, P. Friis and P. Telleman, Chip-Integrated
`Microfluidic System for Cell Sorting and Cell Culturing, , in Proc. of
`Eurosensors XIV, Copenhagen, Denmark, August 27–30, 2000, pp.
`235–238.
`9 D. Huh, Y. Tung, H. Wei, J. B. Grotberg, S. J. Skerlos, K.
`Kurabayashi and S. Takayama, Use of Air-Liquid Two-Phase Flow in
`Hydrophobic Microfluidic Channels for Disposable Flow Cyt-
`ometers, Biomed. Microdev., 2002, 4(2), 121–149.
`10 J. H. Nieuwenhuis, J. Bastemeijer, P. M. Sarro and M. J. Vellekoop,
`Integrated Coulter counter with non-coaxial sheath-flow and dy-
`namic aperture control, , in Proc. of Eurosensors XVI, Prague, Czech
`Republic, September 15–18, 2002, pp. 1194–1197.
`11 J. H. Nieuwenhuis, J. Bastemeijer, A. Bossche and M. J. Vellekoop,
`Dynamic Particle-Shape Measurements Using a Near-Field Optical
`Sensor, , in Proc. of 1st IEEE Sensors Conference, June 12–14, 2002,
`Orlando, Florida, USA.
`12 J. V. Watson, Introduction to flow cytometry, 1st edn., Cambridge
`University Press, Cambridge 1991, pp. 2–4.
`13 M. Koch, A. Evans and A. Brunnschweiler, Microfluidic Technology
`and Applications, 1st edn., Research Studies Press Ltd., Hertfordshire,
`2000, pp. 227–228.
`
`Fig. 13 The flow-cell applied for sample splitting.
`
`Fig. 14 The flow-cell applied for flow-switching.
`
`realistically taken into account. The results on the horizontal
`and vertical control of the sample flow dimensions show that
`both mechanisms work well and that they can be predicted from
`simulation results. The control ports make the flow-cells very
`versatile so that they can also be used for other applications such
`as sample splitting and flow switching.
`The flow-cells can be used in a broad range of applications
`such as an integrated Coulter counter,10 sensors for particle
`shape analysis11 and sensors for cell analysis and (bio)chemical
`analysis.
`
`Acknowledgement
`
`The authors would like to thank the DIMES Technology Centre
`of the Delft University of Technology for the fabrication of the
`device, especially Mr. C. R. de Boer and Mr. W. van der Vlist.
`
`Published on 13 March 2003. Downloaded by University of California - Los Angeles on 04/10/2015 20:18:06.
`
`Lab Chip, 2003, 3, 56–61
`ABS Global, Inc. and Genus plc – Ex. 1014, p. 61
`
`61
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket