throbber
ABS Global, Inc. and Genus plc – Ex. 1020, cover 1
`
`

`

`ABS Global, Inc. and Genus plc – Ex. 1020, cover 2
`
`

`

`ABS Global, Inc. and Genus plc – Ex. 1020, cover 3
`
`

`

`RESULTS OBTAINED USING A PROTOTYPE
`
`MlCROFLUlDICS-BASED HEMATOLOGY ANALYZER
`
`Eric Altendorfl', Diane Zebert‘, Mark Hollz, Anthony Vannelli‘, Caicai Wu',
`Thomas Schulte1
`
`1) Micronics, lnc., Redmond. WA 98052, USA
`
`2) Department of Bioengineering, University of Washington, Box 352141,
`
`Seattle, WA 98195, USA
`
`Abstract
`
`Microfluidic laminate-based structures incorporating hydrodynamic focusing and flow
`channels with dimensions much less than 1 mm were fabricated and used to transport and
`analyze blood samples. Optically transparent windows integral to the flow channels were
`used to intercept the sample streams with a tightly focused diode laser probe beam. The size
`and structure of the blood cells passing through the laser beam determined the intensity and
`directional distribution of the scattered light generated. Forward and small angle light
`scattering channels were used to count and differentiate platelets, red blood cells, and various
`populations of white blood cells. All the blood samples used were characterized using a
`commercial hematology analyzer for comparison and validation purposes.
`
`Keywords: cytometry, light scattering, microfluidics, clinical diagnostics
`
`1. Introduction
`
`flow cytometry is an established means of counting and classifying particles
`Optical
`contained within a fluidic sample [1]. One application involves the analysis of a blood
`sample for the purposes of determining the numbers of platelets, red blood cells (RBCs), and
`white blood cells per unit volume. This is a common clinical measurement, and optical
`cytometers have been incorporated into a number of commercial hematology analyzers.
`Recently, microfluidic techniques have been employed for the purposes of developing
`cytometers which require smaller sample and reagent volumes [2-4]. Analytical instruments
`based on these efforts will be smaller and more portable than conventional devices. This
`paper reports results obtained with a microcytometer utilizing laminate-based microfluidic
`structures and diode laser light scattering optics. A small portable hematology analyzer
`incorporating this microcytometer is presently being designed.
`
`Blood cell counting and differentiation by laser light scattering: Blood cells flowing through
`a tightly focused laser beam can be identified by the resultant scattered light. At small angles
`relative to the incident beam, the scattered light is largely a function of blood cell size [5]. At
`larger angles,
`the scattered light
`intensity is also dependent on the degree of cellular
`complexity and internal structure [5]. Terstappen et al. [6] demonstrated that
`the scattered
`light at relatively small scattering angles also contains structural as well as size dependent
`information. In particular, a combination of forward angle light scattering (FALS), which
`deviates from the incident beam direction by less than 3 degrees, and small angle light
`scattering (SALS) which is collected at slightly larger angles (3 to 11 degrees from the
`incident beam direction), can be used to differentiate white blood cells. In addition. either
`FALS or SALS signals can be used to count and differentiate RBCs and platelets.
`
`73
`
`ABS Global, Inc. and Genus plc – Ex. 1020, p. 73
`ABS Global, Inc. and Genus plc — EX. 1020, p. 73
`
`

`

`74
`
`Hydrodynamic focusing: In order to count and classify blood cells in the manner described
`above,
`the cells must be made to pass one at a time through the incident laser beam.
`Traditionally this is done by using hydrodynamic focusing [1], where the blood cell sample
`stream is encircled and combined in appropriate proportions with a sheath fluid stream, and
`both fluids are made to pass through a tapered orifice or flow channel. The hydrodynamic
`forces of the flowing and constricting sheath fluid then cause the sample stream to form a
`narrow thread of flowing cells.
`
`Microfluidic flow cytometry: Microfluidic structures can be used to spatially confine blood
`cells in very narrow and precisely formed channels [3] without the need for a sheath fluid.
`Such microstructures cannot, however, typically accommodate high flow rates due to the
`shear stresses generated. The microcytometer described in this paper spatially confines both
`the sample and sheath streams within a microchannel of appropriate dimensions for the
`assays intended.
`
`2. Experimental
`
`All the experiments were performed in laminate-based planar microstructures manufactured
`by Micronics, Inc. Two representative structures are shown in Fig.
`l and Fig. 2. The device
`shown in Fig.
`l was comprised of two thin glass windows separated by a 100 micron thick
`adhesive layer into which flow channels were cut. This device produced a focusing of the
`sample stream in one plane only, and hence will be referred to as the 2-D flow cell. The
`cross-sectional dimensions of the channel containing the focused cell stream was 100
`microns by 1mm. The structure shown in Fig. 2 was formed from Mylar and Mylar-laminate
`sheets (3M, Austin, TX). The channels were cut with a C02 laser system (ULS—25E, Universal
`Laser Systems, Scottsdale, AZ). This device produced a focusing of the sample stream by
`fully encircling the sample stream with a sheath fluid stream, and will be referred to as the 3-
`D flow cell. The cross-sectional dimensions of the channel containing the focused cell stream
`in this device was 100 microns by 500 microns. A custom control station consisting of two
`computer controlled syringe pumps (Kloehn Company, Ltd, Las Vegas. NV) was used to
`provide constant sheath and sample flow rates in the micro-structures. The sample flow rate
`was 95nl/s in the 2-D flow cells and
`440nl/s in the 3-D flow cells. A 9mw, 685nm
`
`wavelength diode laser module (Melles Griot, Boulder, CO), which produced a near circular
`collimated 4mm diameter beam, was used as the light source. The laser beam passed through
`two crossed cylindrical lenses in order to produce a focused elliptical beam with dimensions
`perpendicular to the direction of propagation of 13 microns by 105 microns at the sample
`stream. Two high speed narrow format seaming photodiode detectors (Centro Vision,
`Newbury Park, CA) were used to collect the PALS (1.4 degrees to 2.2 degrees) and SALS
`(2.2 degrees to 8.4 degrees) signals. An obscuration bar was used to block the direct laser
`beam from impinging on the detectors, and three lenses and a custom mirror with a central
`aperture (hole) were used to direct the scattered light to the detectors. Electronics signals
`were collected with an AT-MIO-l6E-l data acquisition board (National Instruments, Austin,
`TX) when using the 2-D flow cell, and with a with a custom high speed data acquisition
`system when using the 3-D flow cell.
`
`All blood samples (human) were collected with vacutainer (Becton Dickinson, Franklin
`Lakes. NJ) tubes containing the anticoagulant EDTA, prior to preprocessing and dilution. For
`the RBC and platelet assays, dilution of the samples with phosphate buffered saline was
`carried out by external manual mixing prior to introduction to the flow system. For the white
`blood cell assays, external mixing of the blood sample with a commercial soft lysing reagent
`(Snack-Sheath, Streck Laboratories. Omaha, NE) was carried out prior to introduction to the
`instrument. Times between mixing and analysis in both cases were kept to a minimum to
`
`ABS Global, Inc. and Genus plc – Ex. 1020, p. 74
`ABS Global, Inc. and Genus plc — EX. 1020, p. 74
`
`

`

`avoid excess lysing of white cells or osmotic distortion of the RBCs and platelets. Aliquots of
`all the original samples, as well as aliquots of the mixed and diluted samples where possible,
`were analyzed with a commercial hematology analyzer
`(Cell-Dyn 3500R, Abbott
`Laboratories, IL) for comparison and control purposes.
`
`75
`
`
`m1
`‘(—""'"'—'—\,"‘.'
`'»
`;: ¥ (Ti-w
`:39 %l_\
`j»;
`.u V l:
`1 WL “1"].
`i
`PS) i
`f
`
`Fig.1
`
`:3
`
`10mm
`
`3. Results and discussion
`
`Fig. 3 displays results obtained using a 1:400 prediluted sample of whole blood, and a 2-D
`flow cell similar to that shown in Fig. 1. A histogram of the SALS light scattering pulse
`amplitudes is shown indicating a bimodal distribution corresponding to platelets and RBCs.
`Integration of the area under each of the peaks was used to determine the relative percentages
`of RBCs and platelets in the sample. These results along with data obtained from three
`additional samples are plotted in Fig. 4 against RBC and platelet percentages obtained using
`the Cell-Dyn 3500R, Fig. 4 indicates a good correlation between the two methods.
`
`
`
`a” h
`
`3°“ ’
`
`Platelet
`
`Rees
`
`W 20“ -—
`
`E
`
`ton —
`
`“i
`
`1 HL.
`U
`50
`
`.4
`
`.J
`100
`
`A
`150
`
`in
`g
`2 "
`gQ)CL
`on
`E)O
`Q a0
`gg.
`§0
`i
`
`20
`
`n
`
`
`
`1”
`
`I RBC counts
`o Pmm count:
`-—L|M' "suntan
`
`ll
`
`20
`
`.0
`
`W
`
`.0
`
`no
`
`SALS lntendty
`
`Fig. 3
`
`Cell-dyn cell percentages
`
`Fig. 4
`
`Fig. 5 displays results obtained using a whole blood sample diluted 1:50 in the commercial
`sofi lyse reagent, and a 3-D flow cell similar to the one shown in Fig. 2. A 2-D histogram of
`the FALS and SALS data is shown in Fig. 5 and an analysis of an aliquot of the same sample
`by the Cell-Dyn 3500K in shown in Fig. 6. The 3 dominant clusters in each plot correspond
`to lymphocytes (L), monocytes (M), and granulocytes (G). The microcytometer, and
`
`ABS Global, Inc. and Genus plc – Ex. 1020, p. 75
`ABS Global, Inc. and Genus plc — EX. 1020, p. 75
`
`

`

`ll1I1 I
`
`76
`
`
`
`._ i
`(an)9......4___HA....‘l
`
`FALSIntensity
`
`...
`
`U
`
`7' S-\lS intensity (au)
`
`255
`
`CUHPLEX 1T?
`
`Fig. 5
`
`Fig. 6
`
`commercial analyzer produce similar degrees of cluster resolution. The relative cell counts
`(to the total white cell count) obtained with the microcytometer are 27.0% (L), 9.31% (M),
`and 63.7% (G), which are also in good agreement with the Cell-Dyn percentages of 26.9%
`(L), 9.68% (M), and 63.4% (G).
`
`4. Conclusions
`
`The microcytometer described in this paper has demonstrated the ability of counting and
`classifying platelets, RBCs, and various white cell populations by means of laminate-based
`microfluidic flow channels, and light scattering optics. Additional light scattering and data
`analysis channels will be used to extend the capabilities of the microcytometer toward a
`complete blood cell assay including a five-part white cell differential. This microcytometer
`will then be incorporated into a prototype hematology analyzer for field testing.
`
`References
`
`1.
`
`2.
`
`3.
`
`4.
`
`5.
`
`6.
`
`HM. Shapiro, Practical flow cytometry, Third Edition, Wiley-Liss publishers, New
`York, 1995.
`
`E. Altendorf, D. Zebert, M. H011, and P. Yager, Differential blood cell counts
`obtained using a microchannel based flow cytometer, Sensors and Actuators, 1 (1997)
`531-534.
`
`D. Sobek, S.D. Senturia, and M.L. Gray, Microfabricated fused silica flow chambers
`for flow cytometry, Solid-State Sensor and Actuator Workshop, Hilton Head Island,
`South Carolina (1994).
`
`R. Miyake, H. Ohki, I. Yamazaki, and R. Yabe, A development of micro sheath flow
`chamber, Proceedings of the IEEE micro electro mechanical systems workshop, Nara,
`Japan, (1991) 265-270.
`
`M. Kerker, Elastic and inelastic light scattering in flow cytometry, Cytometry, 4
`(1983) 1-10.
`
`L.W.M.M Terstappen, 3.6. de Grooth, K. Visscher, F.A. van Kouterik, and J. Greve,
`Four parameter white
`cell differential
`counting based on
`light
`scattering
`measurements, Cytometry, 9 (1988) 39-43.
`
`ABS Global, Inc. and Genus plc – Ex. 1020, p. 76
`ABS Global, Inc. and Genus plc — EX. 1020, p. 76
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket