throbber

`
`wmawwwmn—Kurt-ESE.
`
`>m02w2
`
`moto<m<o
`
`
`
`><u_mn_ws.:.—.
`
`._<>mm:.z_
`
`Io._._>>m
`
`mmajmsa_m
`
`
`
`
`
`
`
`goodE0522.5550mumaomW5...J“=03823;2.55301588550.
`
`
`
`
`muomoommllllllJilllllllllllllIIhummmChummpmmmma
`
`8.18.26mo<bo>-
`
`E?
`
`
`
`
`
`\ufilui1!.‘«.5?
`
`on
`
`
`
`N?
`
`
`
`Iota/m§<moomm
`
`m>._<>aozmjom_
`
`0¢\\_D
`
`mmammmmm
`
`mumDom
`
`Aug. 28, 1962
`
`Filed Oct. 14. 1959
`
`A.
`D.KOMPEUEN
`APPARATUS FOR AUTOMATICALLY MEASURING BLOOD
`PRESSURE AND PULSE RATE
`
`3,051
`
`,
`
`165
`
`DOOJm
`
`030._.m>m
`
`o_|_o._.m>m
`
`>mozw2
`
`2 Sheets—Sheet 1
`
`wIut>>mmeum
`
` N”5:053m/%EozmzMmmommumm«midiAW”
`
`
`z.mo<50>o:1K.(KmozmmmmmmooD
`.MmM82598”.ozEaEmmMéN0A
`
`
`
`.mn‘muMN592....I1II2.5
`
`|PR2018—01093
`
`Apple Inc. Ex. 1006 Page 1
`
`IPR2018-01093
`Apple Inc. Ex. 1006 Page 1
`
`
`
`
`
`
`
`
`
`
`
`
`

`

`Aug. 23, 1962
`
`A. D. KOMPELIEN
`APPARATUS FOR AUTOMATICALLY MEASURING BLOOD
`PRESSURE AND PULSE RATE
`
`3,051,165
`
`Filed Oct. 14. 1959
`
`2 Sheets-Sheet 2
`
`
`
`DIASTOLIC POINT-
`DIASTOLIC AMPLIFIER
`CEASES OPERATION
`
`TYPICAL AV TRIGGERING
`
`DIASTOLIC AMPLIFIER
`I
`71?—
`
`Fiyflg
`
`SYSTOUC POINT_
`VOLTAGE PEAK T00
`SMALL TO TRIGGER
`
`
`PREAMPLIFIEROUTPUT
`M"
`II
`SYSTOLICAMPLIFIER
`zozoxoxozozoxoxoxoz:
`
`PRESSURE AND VOLTAGE VARIATION ON TRANSDUCER
`
`CYCLE MOTOR RUN TIME
`
`PULSE RATE
`
`—V////////////////////////////////A
`
`DIASTOLIC BLOOD PRESSURE ._
`
`SYSTOLIC BLOOD PRESSURE
`
`'
`
`.
`
`.
`
`—V//////////////////m
`
`—w,,/,/////,.
`
`PROGRAM SWITCHES 40 a 59 >10331.332.31.33319301033333333?
`
`RESET SWITCH 60
`
`0 IO 20 30 40 so so 70 so 90'IOO IIO
`SECONDS
`
`,
`
`PROGRAM CYCLE
`I:I INDICATES INTERVAL
`DURING WHICH FUNCTION F' 4
`'3 35"“ TAKEN
`1?”
`m INDICATES INTERVAL DURING
`WHICH READOUT MAY OCCUR
`XXX SWITCHES 4O, 59, a 60 CLOSED
`
`INVENTOR.
`ARLON D. KOMPELIEN
`f
`-A 7' ORA/EY
`
`yB
`‘ W‘Wf’l‘
`
`|PR2018—01093
`
`Apple InC. Ex. 1006 Page 2
`
`IPR2018-01093
`Apple Inc. Ex. 1006 Page 2
`
`

`

`w
`United States Patent Office
`3,®5l,lfi5j
`Patented Aug. 28, 1962
`
`1
`
`3,051,165
`APPARATUS FOR AUTOMATICALLY MEASURING
`BLOOD PRESSURE AND PULSE RATE
`Arlon D. Kompelien, Richfield, Minn., assignor to Minne-
`apolis-Honeywell Regulator Company, Minneapolis,
`'
`., a corporation of Delaware
`Filed Oct. 14, 1959, Ser. No. 846,468
`10 Claims.
`(Cl. 128—105)
`
`u
`
`My invention relates to apparatus for sensing and
`recording blood pressure and pulse rate and more par-
`ticularly to an improved apparatus for automatically
`sensing such physical functions on a human being.
`Medical apparatus which performs sensing and record-
`ing functions automatically and simply has become of
`increased importance as a means for replacing the man-
`ual operations performed by those working in the med-
`ical field in connection with operative and post-operative
`patient care and in connection with research of a med-
`ical nature.
`Increased accuracy and repeatability of
`measurement are significant both from a research stand-
`point and from the operational standpoint of modern
`hospital facilities. The present invention is directed to
`a photoelectric pneumatic process of sensing the human
`function of diastolic and systolic blood pressure and
`pulse rate in an improved automatic apparatus which
`provides for extreme accuracy of reading or measurement
`and continuing records relative to patient care.
`In the
`research field, physical body functions are being ex-
`amined in the field of aviation or avionics and continuous
`records of the physical characteristics of patients or sub-
`jects of research investigation .are becoming essential.
`The photoelectric method of detecting the presence of
`blood in a portion of the human body has been previ-
`ously utilized in connection with ox gen analysis of
`blood and has also been utilized in connection with the
`measurement of blood pressure. The present apparatus
`permits rapid and numerous measurements of these body
`functions accurately and rapidly with a continuous indi-
`cation and record of these functions without requiring
`any manual operation.
`It is therefore an object of this
`invention to provide an improved electrical apparatus for
`automatically sensing and recording a plurality of body
`functions.
`It is further an object of this invention to
`provide an improved electrical apparatus for the meas-
`urement of the physical function of systolic and diastolic
`blood pressure.
`Another object of this invention is to provide an im-
`proved apparatus for measuring blood pressure which
`performs the measurement of a rise in cuff pressure. A
`still further object of this invention is to provide an im-
`proved apparatus of this type which permits rapid meas-
`urement of blood pressure and pulse rate in a continu-
`ous sequence Without efiecting physical response or dis-
`comfort of the body to which it is applied. These and
`other objects of this invention will become apparent
`from a reading of the attached description together with
`the drawings wherein:
`FIGURE 1
`is a schematic drawing of the improved
`apparatus,
`
`FIGURE 2 is a diagrammatic view of an ear cuff used
`in the improved measurement apparatus,
`FIGURE 3 is a graph relating pressure to electrical
`response in the apparatus, and
`FIGURE 4 is a graph indicating the on .and off periods
`for the switching circuitry of the programming portion
`of the apparatus.
`One of the most variable and difficult to measure of
`
`30
`
`35
`
`4O
`
`45
`
`50
`
`55
`
`60
`
`65
`
`10
`
`15
`
`20
`
`25
`
`2
`all human functions with repeatability of results and
`accuracy is that of blood pressure. While accuracy and
`measurement may be obtained through arterial punc-
`tures, this method is normally not feasible in patient care.
`The sphygmomanometer or arm cult which is used in
`the measurement of blood pressure on human beings is
`further not a feasible apparatus for the continuous meas-
`urement of blood pressure because of inaccuracies caused
`by the effect of continuous pressure on the portion of
`the body to which it is applied. The present apparatus
`utilizes an ear piece or cuff in which a light source photo-
`cell combination or photoelectric pickup is utilized to
`determine the amount of light transmitted through the
`tissue of the ear as a means for determining blood con-
`tent of the organ and hence the effect of pressure on the
`same. This approach and apparatus were basically used
`in connection with studies of oxygen content and blood
`and has also been used in connection with measurement
`of blood pressure in medical research. However the
`approach utilized to date has not provided for continu-
`ous and automatic measurement of blood pressure.
`Therefore the arrangement and apparatus for performing
`this automatic measurement is the subject of the present
`invention.
`The ear piece as shown in FIGURE 2 includes basi-
`cally a two part structure indicated generally at 10 as
`including a first part ll: having a cylindrical member 12
`mounted thereon and the upstanding portion 13 to which
`the second part 15 of the ear piece is connected and
`pivoted. The cylindrical portion 12 includes a light
`source indicated schematically in FIGURE 1 at 18 and
`an expansible chamber or pressure member 20 which
`chamber has a transparent surface 19 such that the light
`may be directed therethrough. The part 15 mounts
`thereon a photocell 25 so associated that the parts 15
`and 11 may be pivoted together with a portion of the
`ear positioned therebetween and clamped in this relation-
`ship such that pressure may be applied to the chamber
`20 and allowing expansion of the extremity of the cham-
`ber against the ear lobe forcing it against the piece 15
`in which this photocell 25 is located forcing blood from
`the ear therebetween. A pin or locking mechanism on
`the ear piece (not shown) prevents the device from open;
`ing when pressure is applied to the chamber 29.
`In this
`position the light and photocell are directly in line and
`enough space exists between these parts so that there is
`no pressure applied to the ear in addition to that of the
`bellows.
`
`Variation in opacity of the ear or portion of the body
`to which the ear piece is applied is caused by variation
`of blood content of the same to provide a varying amount
`of light from the source to the photocell to produce a
`signal output. With the transducer or ear piece on the ear
`and no pressure applied to the bellows the blood content
`in the ear or the capillaries pulses with normal blood
`flow causing the photocell to put out an alternating cur.
`rent type signal. As pressure in the bellows is slowly
`increased the alternating current signal from the photo-
`cell increases and reaches a maximum. This is caused
`by the pressure in the bellows forcing much of the blood
`from the ear except during a period when the blood is
`forced during heart pumping operation back through the
`capillaries. The pressure in the belllows corresponding
`to this point where the photocell signal is a maximum is
`the diastolic blood pressure. This signal is maximum since
`the pressure of the bellows is just able to squeeze out the
`blood in the capillaries at the minimum part of the blood
`pulsing pressure allowing the greatest chafigeRiéolifigt01093
`Apple Inc. Ex. 1006 Page 3
`
`IPR2018-01093
`Apple Inc. Ex. 1006 Page 3
`
`

`

`3,051,165
`
`3
`variations. As pressure in the chamber is increased still
`further,
`the signal decreases and approaches zero level
`indicating the least opacity or variation in density. The
`point where the pressure drops to zero is the systolic pres-
`sure point since the maximum part of the pulsing blood
`pressure just reaches a point where it can no longer push
`- blood through the pressurized portion of the ear to vary
`the light of the photocell and hence the density of the ear
`remains uniform.
`The apparatus for automatically measuring these func-
`tions in addition to the ear piece 10 includes an air source
`for varying the air pressure applied to the ear piece. A
`conduit or pipe indicated at 36 has connected thereto a
`valve 31 to vent the conduit, a variable pressure reservoir
`32 operated by a cam 33 to vary the pressure in the con-
`duit, and a pressure to voltage transducer 34 connected in
`the conduit 3% with a connection leading to the pressure
`chamber 20 of the ear piece. Valve 31 is a solenoid op-
`erated valve which is energized through a switch 40 op-
`era-ted from a programmer 45. The programmer also
`operates the .cam 33 and other switches to be later identi-
`fied. A broken line connection in FIGURE 1 at 42, 43
`indicates the coupling to these switches.
`In addition, valve
`31 is operated by a second control switch 41 which, as
`will be later noted, is controlled by a time delay interval
`switch. The pneumatic or pressure portion of the ap-
`paratus is initiated with operation of the valve 31 to close
`the conduit which is normally vented to the atmosphere.
`Air is then trapped in the conduit 30 and chamber 32 such
`that .upon rotation of the cam 33 a pressure build up will
`be experienced in the chamber 20 of the ear piece to
`press against the ear causing the blood therein to be
`forced therefrom. The switch 40 or switch 41 will release
`or de-energize the valve 31 to allow the air in the conduit
`to rapidly vent to the atmosphere for release of pressure
`on the ear. The pressure to voltage transducer 34 pro-
`vides an electrical signal output in accordance with the
`pressure in the chamber 20 of the ear piece, which signal
`will be indicative of the actual pressure applied to the
`portion of the ear. Programmer or master control 45
`is controllably energized from a power source through a
`switching means
`(not shown) and will operate con—
`tinuously for a given cycle to perform the blood pressure
`measurement and pulse measurement functions to be
`later described.
`- The photocell .25 of the ear piece is energized from a
`light source and will provide a variable signal output of
`the alternating current type which is connected through
`conductor means 48 to a preamplifier unit indicated in
`block form at 59. The details of the preamplifier together
`with the amplifiers and relays to be later defined are
`omitted herein for simplicity since they are conventional
`components. The output of the preamplifier section 50
`is coupled through a condenser indicated at 51 to a voltage
`limiting network which permits the passage or flow of
`signals of pulsed voltage output only when the succeeding
`voltage signal is larger than the preceding voltage signal.
`This network includes a condenser 52 and a rectifier or
`diode 53 connected through a limiting resistor 54 to the
`diastolic amplifier indicated in block form at 55 with a
`restoring resistor 56 connected to ground connection and
`having a diode 57 in parallel circuit therewith. A further
`resistor 58 is also connected to ground connection beyond
`the diode 53. With this circuit arrangement the capaCitor
`52 charges to the peak output voltage from the pream-
`plrfier 50 each time the signal is higher than any preceding
`one. No signal will flow through the resistor 58 when
`signals of lesser amplitude occur since diode 53 remains
`back biased by the voltage on capacitor 52 during the time
`of these signals. Hence the diastolic amplifier 55 will
`respond onlyto a signal larger than the preceding charge
`on the condenser 52. Diodes 57 and 53 limit direction of
`current flow through this network.
`Connected in parallel with the condenser 52 is a short-
`ingor reset switch 60 operated through a mechanical
`
`10
`
`15
`
`20
`
`25
`
`3O
`
`35
`
`40
`
`45
`
`50
`
`60
`
`4
`connection of the step controller or programmer 45, the
`mechanical connection being schematically indicated at
`42. Diastolic amplifier 55 provides a pulsed output which
`controls a conventional relay mechanism 65 having a
`contact or switch 66 which controls or connects the pres-
`sure to voltage transducer 34 to a memory capacitor or
`condenser indicated at 70. The schematic diagram in
`FIGURE 1 shows an electrical connection 68 schematical-
`ly connecting the transducer 34 to the relay 65 to be con-
`trolled by the contacts '66 of the relay 65 and being con-
`nected through a conductor 72 and an interval switch con-
`tact 75, to be later defined, to the memory capacitor 70.
`The output of the memory capacitor 70 controls the en-
`ergization of a diastolic memory amplifier 80 which re-
`sponds to the charge on the condenser 70 to provide a
`signal output to a recorder indicated generally at 100. A
`read-out switch mechanism 85 is included in the output
`connection from the amplifier 80 to the recorder 100 to
`selectively connect the output of the amplifier 80 to the
`recorder. This read-out switch, as shown by the me-
`chanical connection 42,
`is operated by the programmer
`45 or as a part of the same.
`It should be recognized,
`however, that this portion of the apparatus may be in-
`cluded in the recorder drive to sequence with the opera-
`tion of the programmer 45. A parallel circuit from the
`preamplifier 50 and coupling condenser 51' includes a
`conductor or electrical connection 90 and a limiting re-
`sistor 91 connected to a systolic amplifier indicated in
`block form at 95 whose output controls the operation of
`a pulsing relay 96 having a contact or switch 97 connecting
`the output of the pressure to voltage transducer 34 through
`the conductors or connection indicated at 99 to a circuit
`including an interval switch 101 and a memory capacitor
`102. The output of the memory capacitor controls the
`energization of a systolic memory amplifier 104 whose
`output is coupled to the recorder 100 through a read-out
`The read-out switch ‘105 like
`, switch indicated at 105.
`switch 85 may be operated through the mechanical con-
`nection 42 from the programmer or be included in the
`recorder.
`Also connected to the output of the diastolic amplifier
`is a stepping relay or switch 110 which is connected in an
`electrical circuit indicated at 111 through a contact 112
`of programmer 45 and a contact 114 of relay 65 controlled
`by diastolic amplifier 55, the stepping relay being opera-
`tively connected to actuate a resistance divider or signal-
`ling device 116 which is energized from a reference source
`117 to provide a signal output through a circuit "118 and
`a read-out switch 120 to the recorder 100. The stepping
`relay pulses with each pulse of output of the diastolic
`amplifier 55 to provide a stepping action which adjusts
`the wiper of the potentiometer type device or divider 116
`to provide a signal output in accordance with the number
`of pulses of relay 65. This will provide a voltage output
`in accordance with the number of blood pulses sensed by
`the photocell over a given time interval. The stepping
`relay 110 and resistance divider 116 act as an analogue
`converter to count and convert the number of pulses to
`an analogue signal output or a measure of the number of
`pulses sensed. An interval switch 131 of the time delay
`type which operates the contacts 75, 101, and 41, pre-
`viously indicated, is connected to and controlled by the
`operation of the relay 96 of the systolic amplifier 95
`through a contact 98 of the relay and connections indi-
`cated at 1138 and 139. A reset contact 59 for the interval
`switch is operated by the programmer 45 to insure that
`the interval switch will operate only during a period when
`measurements are being taken and will prevent extraneous
`signals from operating the switch after the systolic blood
`pressure measurement is taken until a new cycle of meas-
`urement has commenced. The contacts 112 connecting
`the stepping relay 110 to the pulse relay 65 of diastolic
`amplifier 55 are closed by the programmer 45 during the
`initial portion of the measurement cycle while reset con-
`tact 136 for the stepping relay “335310 1%ft8'1tbe9psplse
`
`70
`
`75
`
`Apple Inc. Ex. 1006 Page 4
`
`IPR2018-01093
`Apple Inc. Ex. 1006 Page 4
`
`

`

`3,051,165
`
`5
`measurement period is complete, contact 112 opens and
`contact 136 closes to reset the stepping relay and the
`signalling device 116 back to the starting position for a
`new cycle of measurement.
`In addition, the time delay
`interval switch includes the switch contact 41 which is
`connected in a series circuit with switch 40 of programmer
`45 to control the energization of solenoid valve 31. Nor-
`mal blood flow in the ear without air pressure in the
`clamp of the ear piece is suflicient to produce a signal
`output from the photocell
`to periodically energize the
`systolic amplifier and keep the internal switch closed once
`it has been closed by contact 59. Since the programmer
`45 at the outset closes switch 59 pulling in interval switch
`131, the contact 41 will be closed at the outset of meas—
`urement. Contact 98 included in a time delay holding
`circuit (not shown) keeps interval switch 131 operated
`as long as it is not left unoperated for a period of over
`three or four seconds. With the series circuit for solenoid
`31 complete, the valve wil close, permitting a pressure
`build up in the clamp of the ear piece while contacts 75
`and 101 are also closed during the pulse measurement
`period and relays 65 and 96 are operating, read-out con-
`tacts 85 and 105 are open so that no blood pressure sig-
`nals reach the recorder.
`As indicated above, reset switch 60 which is operated
`by the programmer 45, operates to shunt condenser 52
`and under these conditions the diastolic amplifier 55 will
`respond directly to the pulsed output of the photocell.
`These conditions maintain during measurement of pulse
`rate and during this period the pressure source 32 does not
`alter the pressure in the chamber 20 of the ear piece sig—
`nificantly since contact 40 keeps the pressure system vented
`to the atmosphere. Following the measurement of pulse
`rate the programmer closes contact 40 and opens contacts
`60 and 59. The pressure in chamber 20 then increases
`to vary blood flow in the ear and the photocell 25 provides
`an increasing voltage output to the preamplifier 50. Dur-
`ing the pulse rate measurement period the contacts 59 and
`112 will be closed by the programmer and the interval
`switch 131 will remain energized, keeping contacts 41, 75
`and 101 closed. The reset contact 136 of the programmer
`will be open. During a larger portion of this time period
`the remaining contacts of the programmer will be open
`and no signals will be applied to the recorder. During
`this period the counting operation of the stepping relay,
`which is converted into an electrical analogue output,
`will provide a pulse count for the recorder which after a
`predetermined time period will be read out to the recorder
`through operation of switch 120. The interval switch
`which as indicated above is of the time delay type will
`remain closed between each pulse of signal output since
`the time delay circuit (shown in part) is being constantly
`reset through switch contact 98 of the systolic pulse relay.
`The interval switch had been initially energized and sus-
`tained in energization through contact 59 which is open
`at this time. Upon reaching the systolic blood pressure
`point,
`the signal output from the photocell will go to
`zero so the systolic relay will cease pulsing and the interval
`switch will open, dropping out contacts 41, 75 and 1011
`thereby isolating the memory capacitors 70 and 102 from
`the pressure to voltage transducer 34 and releasing or
`opening valve 31. During this period programmer 45
`will have also opened contact 59 so the interval switch
`can drop out. A contact in the interval switch (not
`shown) only allows contact 59 to re—energize it once
`it has dropped out.
`In this event should any extraneous
`signals be sensed by the amplifier 95 corresponding error
`voltages will not be placed on the memory capacitors be—
`fore they can be read out to the recorder. Prior to this
`time, the programmer will have opened switch 112 and
`closed switch 136. The latter switch establishes a reverse
`energization circuit through the structure of the stepping
`relay 110 which Will cause it
`to operate back to its
`original position. As indicated above, the amplifiers, pre-
`amplifier, relays and stepping relay together with the sig-
`
`6
`nal transducers are all of the conventional type and their
`details have been omitted for simplicity.
`In considering the operation of the apparatus refer-
`ence is made to the graph 4 showing the time sequence
`of operation of the switches associated with the program
`controller 45. It will be noted that valve ‘31 of the pressure
`source is initially energized through operation of switches
`40 and 41 to close the conduit to the atmosphere and
`seal the pneumatic system so that the cam may operate
`on the pressure bellows source 32 to raise the pressure
`in chamber 20. During this period the program reset
`switch 60 is closed for the pulse counting operation and
`the contact 112 of the programmer will be closed for
`operation of the stepping relay. The read-out switch 120
`wil-l be closed at the end portion of this period during
`which time it connects the count of the pulse rate from
`the potential signalling device 116 to the recorder. Dur-
`ing this period of time the cam 33 Will be so positioned
`that it will be opening the bellows of the pressure source
`and there will be no pressure in chamber 20 since valve
`31 is venting and the photocell output will be at the level
`indicated in the graph 3 of a small but uniform mag-
`nitude. As pressure is applied to the bellows 32 from
`the cam 33 the pressure in chamber 20 will
`increase
`when valve 31 is energized. At this portion, the reset
`switch 60 is open and the blood in the capillaries of the
`ear pulsing because of normal heart operation Will vary
`the optical density of the ear and hence the variation in
`light from the light source to the photocell to cause an
`increasing alternating current output. As the pressure
`is slowly increased by the bellows, the alternating cur-
`rent signal from the photocell increases and reaches a
`maximum pressure corresponding to the point of diastolic
`blood pressure. During this period of time the systolic
`and diastolic amplifiers 55 and 95 have been receiving
`a pulsed voltage output with increasing voltage pulses.
`The limiting network including the condenser 52 and
`rectifiers 53, 57 leading to the diastolic amplifier will
`respond only to increasing voltage pulses caused by the
`build up in pressure and the change in optical density
`in the ear as blood is forced out of the capillaries.
`It
`is during this period of time that the read-out switch
`120 is closed and the pulse rate signal is fed to the re-
`corder. As the diastolic and systolic amplifiers respond
`to varying voltage pulses, they operate the relays which
`connect through their switches the pressure to voltage
`transducer 34 to the memory capacitors 70 and 102.
`Each time one of the relays operates the pressure to
`voltage transducer is connected to the respective capac—
`itors. Since the diastolic relay makes its last pulse when
`the blood pressure is at the diastolic pressure point and
`the systolic relay makes its last pulse when the pressure
`in chamber 20 and hence on the transducer 34 is at the
`systolic pressure point the proper value can be left on
`the memory capacitors. The associated memory read-
`out amplifiers then feed the corresponding outputs to
`the recorder without discharging the capacitors.
`.As in-
`dicated by the graph in FIGURE 3 this diastolic pres-
`sure point is reached when the change in optical density
`in the ear and hence the photocell output reach a maxi-
`mum at which point the charge on the memory condenser
`70 will correspond to the pressure in the clamp of the
`ear piece. The continuing pressure applied from bellows
`32 acting as a pressure source will force more and more
`blood from the capillaries to a point where the pulsing
`blood Will not be able to enter the portion of the ear
`which is pressurized between the light source and photo-
`cell at which point the optical density variation will be
`at a minimum and the photocell output will be zero.
`However, at this point the voltage signal from the pres-
`sure transducer 34 will be at a maximum leaving the
`maximum charge on the condenser l1024. Slightly after
`these points, the switches 85, 105 respectively close to
`connect the outputs of the amplifiers 80, 104 to the re-
`ocrder, recording systolic and diastolic prifiufi28%;01 093
`Apple Inc. Ex. 1006 Page 5
`
`10
`
`15
`
`20
`
`25
`
`30
`
`35
`
`40
`
`50
`
`55
`
`60
`
`65
`
`7O
`
`75
`
`IPR2018-01093
`Apple Inc. Ex. 1006 Page 5
`
`

`

`3,051,165
`
`7
`The valve 31 is opened rapidly after the systolic point
`is reached through de-energization of interval switch 41
`and the switch 40 is also opened to exhaust the pressure
`chamber 20 and prevent any further pressure application
`to this car. The cam '36 also rotates towards its original
`position, releasing the bellows 32. The program selec-
`tor or controller rotates back to its original position to
`renew the sequence starting with the shorting of con-
`denser 60 and the closing of the switch contacts initiating
`a new sequence of measurement.
`While I have shown in this apparatus only two or
`three measurements of physical functions of a being it
`should be recognized that other physical properties can
`be sensed and recorded in a similar manner.
`It should
`be recognized, however, that this apparatus provides a
`simple, and accurate means of measuring systolic and
`diastolic blood pressure while at the same time utiliz-
`ing these measurements for determining pulse rate. Fur-
`ther. the measurement is taken on a rise in pressure at
`which point the pressure is released from the portion
`of the body in which blood is occluded to provide for
`a minimum of period of time during which the clamp-
`ing or occluding process takes place. This minimizes
`the physical discomfort and abnormal efiects on the
`portion of the body involVed such that repeated measure-
`ments may be taken with repeatability and accuracy of
`results.
`In considering this invention it should be remembered
`that the present disclosure is intended to be illustrative
`only and I wish to be limited only by the appended
`claims.
`Iclaim:
`1. Apparatus for automatically sensing and recording
`a plurality of body functions comprising a primary sensor
`including a light source and photocell combination and
`a variable pressure clamping means, a source of variable
`pressure connected to said clamping means, means for
`cyclically adjusting said pressure source, signalling means
`connected to said clamping means for converting varia-
`ble pressures therein to an electrical signal output propor-
`tional
`to the same, said sensor being adapted to be,
`mounted on a body and so aligned that when said sensor
`is associated with a body said clamping means can alter
`the flow of blood through a portion of the body affected
`by said clamping means to vary the amount of light
`from said light source to said photocell, said photocell
`when so clamped on a body being adapted to produce
`an output of pulsed signals in accordance with blood
`flow providing a maximum output in accordance with
`a lack of blood in the portion of the body when clamped
`and a minimum output under normal conditions of
`blood flow in the body, a first circuit means connected
`to said photocell and including a voltage peak detecting
`circuit means, an amplifier, a relay connected to and
`controlled by said amplifier connected to said peak
`detecting circuit and providing pulsed operation of said
`relay in accordance with pulsed input signals thereto
`from said photocell as long as each succeeding pulsed
`signal is larger than the preceding signal, switching cir-
`cuit means including a part of said relay of said amplifier
`and cennected to the signalling means, a memory circuit
`means including a Capacitor Connected to said relay
`switching circuit means and adapted to [receive the signal
`output from said signalling means with each pulsed oper-
`ation of said relay means, first Output circuit means con-
`nected 'to said memory circuit means and responsive to
`the signal on said capacitor, second circuit means con-
`nected to said photocell and responsive to the pulsating
`signal output thereof,'a second amplifier and relay means
`connected to said second circuit means and operated in
`accordance with said pulsed signal output,
`second
`switching means included in said second relay means con-
`nected to said signalling means, memory circuit means
`including asecond capacitor connected to said second
`relay switching means and responsive to the output of
`
`8
`said signalling means, second output circuit means con-
`nected to and responsive to the signal on said second
`capacitor, third circuit means connected to said first on-
`cuit means and including a stepping relay respousive to»
`the pulsed output of said photocell, transducer means
`providing an analogue signal output connected to said
`stepping relay means, a recorder, and a motor driven
`switching circuit means cyclically connecting said first
`and second output circuit means and said analogue
`transducer means to said recorder.
`2. Apparatus for automatically sensing and recording
`a plurality of body functions comprising a primary
`sensor including a light source and photocell combina-
`tion and a variable pressure clamping means, a source
`of variable pressure connected to said clamping means,
`means for cyclically adjusting a pressure source, signal-
`ling means connected to said clamping means for convert-
`ing variable pressures therein to an electrical signal out-
`put proportional to the same, said sensor being adapted
`to be mounted on a body and so aligned that when said
`sensor is associated with -a body said clamping means
`can alter the flow of blood through a portion of the
`body affected by said clamping means to vary the amount
`of light from said light source to said photocell, said
`photocell when so clamped on a body being adapted to
`produce an output of pulsed signals in accordance with
`blood flow providing a maximum output in accordance
`with a lack of blood in the portion of the body when
`clamped and a minimum output under normal condi-
`tions of blood flow in the body, a first circuit means
`connected to said photocell and including a voltage
`peak detecting circuit means and an amplifier, a relay
`connected to and controlled by said amplifier connected
`to said peak detecting circuit and providing pulsed op-
`eration of said relay 'in accordance With pulsed input
`signals thereto from said photocell as long as each suc-
`ceeding pulsed signal is larger than the preceding signal,
`switching circuit means including a part of said relay
`of said amplifier and connected to the signalling means,
`a memory circuit means including a capacitor connected
`to said relay switching circuit means and adapted to re-
`ceive the signal output from said signalling means with
`each pulsed operation of said relay means, first output
`circuit means connected to said memory circuit means
`and responsive to the signal on said capacitor, second
`circuit means connected to said photocell and responsive
`to the pulsating signal output thereof, a second amplifier
`and relay means connected to said second of circuit
`means and operated in accordance with said jpulsed signal
`output, second switching means included in said second
`relay means, connected to si

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket