throbber

`
`«» UK Patent Application .»GB «. 2077 946 A
`
`
`
`~
`
`(21)
`(22)
`(30)
`(37)
`(32)
`(33)
`(43)
`
`(51)
`
`{52)
`
`(56)
`
`(58)
`
`(71)
`
`(72)
`
`{74}
`
`Application No 81187189
`Date of filing 12 Jun 1981
`Priority data
`80/19259
`12 Jun 1980
`United Kingdom (GB)
`Application published
`23 Dec 1981
`INT CL?
`AGIB 3/12
`Domestic classification
`G20 13A5
`Documents cited
`GB 1401664
`Field of search
`G2J
`Applicants
`Robin Devonshire,
`87, Hallam Grange
`Crescent, Fullwacd,
`Sheffield $10 4BB,
`Gholam Ali Peyman,
`535 North Michigan,
`Chicago,IHinois 60671,
`United States of America,
`Robert L. Epstein,
`912, Pawnee Road,
`Wilmette, Illinois 60091,
`United States of America
`Inventors
`Robin Devonshire,
`Gholam Ali Peyman,
`Robart L. Epstein
`Agents
`Marks & Cterk,
`57-60 Lincoln's Inn
`Fields, London WC24 3L5
`
`
`
`(54) Indirect aphthalmescope
`
`(57) An indirect ophthalmoscope
`consists of a single unit incorporating
`a light source §, collimating lens 6,
`and a beam splitter 7 reflecting the
`light through a lens 8 serving as a
`projection lens and a viewing lens. The
`
`user views the aerial image produced
`by the lens 8, through a viewing
`window 13 close to the image plane.
`The light source and collimating lens
`are mounted in a handle 4. The
`ophthalmoscope can be used with an
`infra-red image converterfor infra-red
`examination.
`
`7
`
`okidandeaak
`
`
`
`V976LL0Z4D
`
`HAAG-STREIT AG - EXHIBIT 1032
`Page 1 of 6
`
`HAAG-STREIT AG - EXHIBIT 1032
`Page 1 of 6
`
`

`

`
`
`2077946
`
`Wt
`
`FrreaOracleelgametes
`
`+
`re)yj
`
`. yANy
`
`5
`
`‘
`
`hs
`
`‘a
`
`r
`
`et
`EF
`
`4
`DS
`
`Te
`4
`iM
`A
`
`b
`
`TY
`A
`ty
`
`q
`
`1d i
`
`:
`
`HAAG-STREIT AG - EXHIBIT 1032
`Page 2 of 6
`
`HAAG-STREIT AG - EXHIBIT 1032
`Page 2 of 6
`
`
`
`

`

`
`
`ray
`
`10
`
`15
`
`20
`
`25
`
`30
`
`35
`
`40
`
`45
`
`60
`
`SPECIFICATION
`Indirect ophthalmoscope
`
`65
`
`70
`
`75
`
`80
`
`B5
`
`90
`
`95
`
`100
`
`This invention relates to indirect
`ophthalmoscopes.
`Conventional indirect ophthalmoscopes
`comprise a head band which carries a light-
`projection system and viewing optics, usually a
`binocular aye piege. The projection system forms a
`beam of light which is directed into the eye of a
`patient by way of a hand-held condensing lens,
`and the retina thusilluminated is viewed through
`the same condensinglens, and the viewing
`optics. Such instruments are widely used but have
`a numberof disadvantages. Thelight projector
`and viewing optics are quite heavy, and a power
`supply lead for the light source must be provided.
`The head band is consequently awkward and
`uncomfortable. It takes a substantial ammountof
`practice, to learn to use such an instrument
`effectively. Careful co-ordination is needed to
`position correctly the hand-held fens and the
`observer's eye. There is usually a large amount of
`stray light which impairs the clarity of the image
`seen by the obsarver, and this makes it necessary
`to use relatively intense illumination, which may
`be uncomfortable or even dangerousfor the
`patient. Because of the difficulty in eliminating
`stray light, inconvenient corneal reflexes may be
`produced.
`According to the presentinvention, tha fight
`projection system, and a converging lens used to
`form an aerial image of the fundus of the eye
`illuminated by the projection system, are
`combinedin a single unit, which can be designed
`to be hand-held. The viewing line of sight andthe
`axis of the light beam are made substantially
`coincident, for example by the use of a beam
`splitter. With such an instrument, the aerial image
`formed can be viewed directly, or through auxiliary
`optics for example a binocular eye piece. Because 105
`the light source and the viewing and image-
`forming lens are combinedin a single unit,
`accurate alignment of the light beam axis and the
`line of sight is ensured and the amount of stray
`light is very muchless than in the case ofa
`conventional indirect ophthalmoscope using a
`separate hand-held lens. The present instrumentis
`therefore much easier to use than a conventional
`indirect ophthalmoscope,requiring little practice to
`secure a clear and well iNuminated image, andthe 115
`accurate controlof the illuminating beam whichit
`makes possible greatly reduces discomfort of the
`patient, both directly and because it becomes
`possible to use less Intenseillumination than ina
`conventionalindirect ophthalmoscope.
`Ina preferred arrangement, the converging lens
`is placed in the light path fram the light source to
`the patient as wall as in theline of sight between
`the patient and the observer, so that this lens
`serves both to direct the light into the eye ofthe
`patient and to form an aerial image of the fundus,
`in a manner analogous to the hand-held lens of a
`conventional indirect ophthalmoscope.
`Preferably, the aarial image is formed
`
`110
`
`120
`
`125
`
`GB 2 077 946 A
`
`1
`
`substantially at a viewing aperture of the
`instrument. This makes it possible to place
`graticules and other optical elaments in or near
`the image plane, for measuring or mapping
`features of the retina.
`The light sourceis preferably a light-emitting
`diode or similar source, or a fibre optic light guide.
`The accompanying drawing show, in section,
`an indirect ophthalmoscope embodying the
`present invention. The eye of a patient is shown
`schematically at 1.
`The instrument has a housing 2 whichis
`generally T-shaped and comprises a cylindrical
`tube 3 mountedacross the wider end of 2 tapered
`tube 4. The tube 4 has at its narrow end a fibre
`optic light guide 5 which illuminates a plano-
`convex Isns 6 to form a parallel beam oflight
`whichis projected into the tube 3.In the latter is a
`partly reflecting and partly transmitting plate 7
`mounted obliquely so as to reflect the incident
`light along the tube 3 onto a plano-convexlens 8
`at one end of the tube 3. This lens forms an image
`of the fibre optic light source, of unit
`magnification, and this image in use is made to
`coincide with the iris of the patient's eye. The lens
`8 also forms an image of the fundus of the eye
`‘thus illuminated, in a plane 9 at the opposite end
`of the tube 3.
`A fitter wheel 10 is mounted on the outer end
`of the light source tube 4, so thatdifferentfilters
`and graticules carried by the wheel can be moved
`into the path of the light from the fibre optic fight
`guide 5. The light guideis longitudinally adjustable
`for focusing.
`:
`On the output side of the lens 6 is a polariser
`11, and another polariser 12 is mounted in the
`viewing window 13 at the end of the tube 3, close
`ta the image plane 9. The polarisers are to
`eliminate unwanted reflections.
`in use, the instrument is held by the observer or
`examiner, so that the light beam is focused at the
`iris of the patient's eye. The examiner observes the
`aerial image formed at the image plane 9. This
`image can be viewed directly or by way of further
`optics for example a binocular optical system (for
`stereoscopic viewing) mounted on a headband or
`on a spectacle frame. The instrumentis very easy
`to use. It is only necessary to ensure that thelight
`beam {s accurately placed in the patient's iris. The
`optical system will then be necessarily correctly
`aligned to produce the desired aerial image, which
`is easily found by the examiner.
`An important advantageof the instrumentis
`the accurate imaging of the light source onto the
`desired position in the eye [usually the plane of the
`pupil). The object-image relation is one to onein
`the case of the lenses 6 and 8 being of the same
`power.
`The iens 8 may be mounted so that lenses of
`different powers can be fitted. In this case, 4 dust
`plate consisting of an optically clear disc is
`provided behind the interchangeable lens 8.
`Typically, a set of lenses of powers 16D, 20D, 300
`and 36D is provided. The lenses 8 preferably have
`projecting hoods, which whenthelens is attached
`
`HAAG-STREIT AG - EXHIBIT 1032
`Page 3 of 6
`
`HAAG-STREIT AG - EXHIBIT 1032
`Page 3 of 6
`
`

`

`70
`
`75
`
`80
`
`85
`
`90
`
`95
`
`to the body of the instrumant projects forwards
`towards the patient. Tha hoods preferably have
`Jengths which are in inverse relation to the tens
`power. This serves to identify the particular lens,
`and to establish an appropriate viewing distance
`from the instrumentto the patient's head.
`The instrument can be held by the userin
`whatever wayis convenient for example by the
`tube 4 or by meansof the region containing the
`lens 8 or a lens hood. Holding the instrument by
`the lens hood or lens-containing region has the
`advantage that the user's fingers, other than the
`index finger and thumb) which grip the instrument,
`are free to rest against the forehead of the patient
`in order to steady the instrument.
`The tube 4 may be attached to or form part of a
`hand grip which can contain a famp bulb and
`- battery, and optionally, a dimmercontrol,
`Alternatively, light may be conducted from a
`remote lamp by the light guide.
`Fibre optic light sources have excellent
`characteristics for use in the present instrument.
`The light outputfalls within a well defined cone
`whichfills the aperture of the collimating lans 6
`andthe resulting parallel beam can be focused to
`give a smail imags at the patient's iris. For
`example, a 2 mm diameterfibre bundle, with an
`optical system of unit magnification, produces an
`image of 2 mm diameter which results in very
`efficient and uniformly distributed illumination of
`the fundus. Because of the small size of the beam
`image, the eye can be examined with a small
`pupil. There is very little stray light in the focal
`plane of the fibre optic source, with the result that
`the reflex from the cornea is spatially well defined 100
`and can be easily avoided by the examiner.
`The position of the light source is adjustable
`towards and awayfrom the lens 6, so that the
`effective focal length of the illumination system
`can be continuously varied whereas the focal
`length for viewing temains unchanged. The
`instrumentis normally set for viewing an eye with
`a “standard”iris-retina distance. To view an object
`closer to the front of the eye, the viewing lens 8
`must be moved awayfrom the eye. This would
`increase the diameter of the light beam attheiris,
`reducing theillumination with the eye and causing
`reflexes. By adjustment ofthe light saurce towards
`the lens 6, the narrowest part of the beam can be
`moved to coincide withtheiris. Such focusing
`adjustmentis also desirable to compensate for
`changesin the focal length of the lens when the
`instrument is used in different regionsof the
`spectrum, for example by meansof filters in the
`filter wheel 10.
`Thefibre optic bundle may have the simplest
`end profile namely a circle, or other bundle end
`profiles can be used for example further to
`alleviate problems of reflexes originating at
`refractive index boundaries on the surface of or
`within the eye. Any desired bundle end profile can
`be selected for example a rectangular or “slit”
`profile forming a corresponding image which can
`be rotated about the axis of the eye by rotation of
`the fibre optic bundle &. Two or more fibre optic
`
`105
`
`110
`
`115
`
`120
`
`125
`
`130
`
`10
`
`15
`
`20
`
`25
`
`30
`
`35
`
`40
`
`45
`
`50
`
`55
`
`60
`
`65
`
`.
`
`GB 2 O77 346 A
`
`2
`
`bundles may be provided to give a desired light
`distribution in the eye.
`Alternative light sources are solid state
`emitters, for example of gallium arsenide. These *
`also have.the advantage thatthe lightis directed-,
`forwards, Incandescentlight sources are inferior,
`because they radiatelight in all directions, and aré
`therefore inefficient and cause straylight.
`Conveniently, the collimating lens 6 is an
`aspheric lens of plastics material which is well
`corrected and has good optical transmission
`throughout thevisible and near infra red regions of
`the spectrum.
`The dual-function liens 8 can also be a plastics
`aspheric lens, but the optical layout enables a
`wide variety of lenses to be used. This lens should
`exhibit no birefringence. Since it is iluminated by
`parallel light, substitution of a lens ofdifferent
`power does not require any other changesin the
`system. The iHuminating light will stili be brought
`to a focus close to the plane of the iris, and will
`Hluminate 4 larger or smailer area of the fundus as
`the powerof the fens is increased or decreased.lf
`the lens 8 is a high-power, small-aperture lens,
`light could be scattered from the lens mountor
`retaining ring which would cause a halo around
`the observed image. To eliminate this, the
`instrument may include an adjustable aperture 14
`so that the illuminating beam can be stopped
`downso asjust to fill the lens 8.
`The polarisers are also to reduce stray light. The
`first polariser 11 preferably has its plane of
`polarisation perpendicular to the plane of
`incidence of the partial reflector 7. The second
`polariser 12 has its plane of polarisation set to
`give minimum transmission to the examinerof
`light reflected from the surface of the cornea and
`from the lens 8. The instrument described has a
`very low aura of stray light. The only significant
`light outside the instrumentis that in the
`converging forward beam, which can be made to
`pass wholly throughtheiris of the patient as
`already described. The high background
`illumination typical of conventional indirect
`ophthalmoscopesIs eliminated and the image
`quality is therefore greatly improved.It is therefore
`possible considerably to reduce the level of
`ilumination of the retina without any loss of
`subjective brightness in the image compared with
`a conventional indirect ophthalmoscope.
`The partial reflector plate 7 has a reflectivity
`‘chosen to maximise the brightness of the
`observed image.If the rear surface of the plate has
`an anti-reflection coating, the front surface should
`have a reflectance of 50%. Hf there is no coating,
`the front surface should have areflectancecf
`_
`approximately 47%.The plate will in general be at
`45° to the axis of the tube 3, providing a fixed
`lens/light source angle ofzero. It may be
`advantageous to have a small fixed angie between
`the lens and light source, for example by mounting
`the reflector plate at 46° to the tube axis. The
`viewing/illummination angle is continuously variable
`from zero upwards, in contrast to conventional
`indirect ophthalmoscopesin which the angle is
`
`HAAG-STREIT AG - EXHIBIT 1032
`Page 4 of 6
`
`
`
`HAAG-STREIT AG - EXHIBIT 1032
`Page 4 of 6
`
`

`

`
`
`3
`GB 2 077 946 A
`Bo
`
`fixed at about 1° or 2°:
`To enable severa! observers to view the image,
`a biprism orthelike can be clipped or hinged ta
`the instrument at the viewing window 13.
`The instrument described has parallel surfaces
`at opposite endsof the tube 3. These can be used
`for accurately known alignment orrelative
`orientation of the axis of the instrument and the
`examiner's eye, using graduation marks on the
`twosurfaces. For example, a smail reflecting or
`juminous point can be provided at the centre of
`the inner face of the viewing lens as a target, to ba
`used in conjunction with a graticule of concentric
`rings.
`‘
`The fact that the imageis within the instrument
`makes it easier for the user to locate the image,
`without howeverrestricting viewing, and makes
`the image accessible to viewing screens,
`graticules and otherarticle devices, or the face
`plate of an image convertor or image intensifier. it
`should be noted that the imageis not plane, and
`such auxiliary devices may have to be shaped
`accordingly to match the shape of the image.
`itis not necessary to use visibie light. One of
`the advantages of the instrument described is
`that, sinceit is not necessary for the user to wear
`the usual heavy light projector and viewing
`system on a headband, he can wear other
`components on a headband,for example an image
`converterfor viewing an-image producedbyinfra-
`red light. In another possible arrangement, an
`image converter may be placed at the viewing
`window 73.
`For infra-red use, a focusing target may be
`provided in or near the retinal image plane at the
`polariser 12. This target can for example be an
`illuminatedring, the illumination being channeled
`from the light source of the hand unit, whether a
`fibre optic bundle or a built-in source such as a
`gallium arsenide emitter.
`The imageintensification properties of an infra- 105
`red image converter, used with the instrument,
`permit examinationof interior portions of the eye
`at Hlurmination lavels significantly lower than those
`used at presentin indirect ophthalmoscopy.
`Conventional illumination levels are being
`increasingly recognised as hazardousto patients.
`Because the image intensifier output depends on
`the properties of the screen phosphor, such
`devices are particularly suitable for low
`illumination level examination of the eye using
`monochromatic radiation of various wavelengths,
`infra-red or visible. The use of monochromatic
`radiation is a recognised technique in
`ophthalmoscopy but at present requires relatively
`high intensities of ittumination. When the present 120
`instrumentis used with an imageintensification
`device, dilation of the pupil is not needed, which Is
`an important practical advantage.
`it will be understood that if the instrumentis
`intended for use with infra-red illumination,its
`optical components must be made of materials
`transparentto infra-red wavelengths and must be
`appropriately designed for operation at such
`wavelengths.
`
`70
`
`76
`
`80
`
`85
`
`90
`
`95
`
`100
`
`110
`
`115
`
`125
`
`10
`
`15
`
`20
`
`25
`
`30
`
`35
`
`40
`
`45
`
`50
`
`55
`
`60
`
`65
`
`
`
`-
`
`The instrument can be used, with an
`appropriate accessory,for fluorescence
`angiography, with visible radiation (using for
`example fiuorescein) or infra-red (using
`cardiogreen), using suitable filters and, in the case
`of infra-red, an Image converter as already
`mentioned, preferably mounted on the examiner's
`head by meansof a head band.The fluorescence
`excitation wavelength Is selected by an
`appropriatefilter in the filter wheel 10, and the
`fluorescence emission is selected bya filter for
`example clipped over the output face of the
`instrument.
`The instrument described combines the ease of
`use of a direct ophthalmoscope, with the wider
`field of view, betterillumination and possibility of
`stereoscopic vision, of an indirect
`ophthalmescope, and has substantial advantages
`not found in either form of conventional
`instrument.
`
`CLAIMS
`1. An indirect ophthalmoscope comprising a
`light projection system for projecting a beam af
`light, and a converging lens arranged te form an
`aerial image of the region illuminated by the light
`projection system, the said light projection system
`and converging lens belng combined in a single
`unit in which the viewing line of sight through the
`lens is substantially coincident with the axis of the
`projected light beam.
`2. An Indirect ophthalmoscopeas claimed in
`claim 1
`in which the light projection system
`includes a beam splitter and a light source, the
`beam splitter being arranged to directlight from
`the source along the saidline of sight.
`3. An Indirect ophthalmoscope as claimed in
`claim 2 having 2 generally T-shaped housing of
`which the stem portion contains the light source,
`and the cross pieceIs a viewing tube which
`contains the beam splitter and converging lens.
`4, An indirect ophthalmoscope as claimed in
`claim 1, 2 or 3 in which the said jens is placed in
`the light path from the light source to the region
`illuminated by the light beam as well as in the line
`of sight from an observerto the said region.
`5. An indirect ophthalmoscope according to
`claim 1,2, 3 or 4 having a viewing aperture and in
`which the lens is arranged to form the said aerial
`image substantially at the viewing aperture.
`6. An indirect ophthalmoscope as claimed in
`any of claims 1 to 5 in the form of a hand-held
`instrument.
`7. An indirect ophthalmoscope according to
`elaim 6 having a handle which incorporates the
`light projection system.
`8. An indirect ophthalmoscope according to
`any of claims 1 to 7 in which thelight projection
`system includes a fibre optic bundle constituting a
`light source.
`9. An indirect ophthalmoscope according to
`claim 8 in which thefibre optic bundle has a non-
`circular end profile.
`10. An indirect ophthalmoscope according to
`any of claims 17 to 7 in which the light projection
`
`HAAG-STREIT AG - EXHIBIT 1032
`Page 5 of 6
`
`
`
`HAAG-STREIT AG - EXHIBIT 1032
`Page 5 of 6
`
`

`

`ted radiation, and viewing means including an
`system includes a gallium arsenide light source.
`infra-red image converter.
`11. In combination, an Indirect
`12. An indirect ophthalmoscope substantiaily_
`ophthalmoscopeas claimed in any of clairns 1 to
`as herein described with reference to the
`.
`10 incorporating an infra-redlight source and
`10 accompanying drawing.
`5 having optical components transparentto infra-
`
`
`
`
`5*
`
`BGR277BABA
`4
`GR 2077 946A
`4
`
`Printed for Her Majesty's Stationery Office by the Courler Press, Leamington Spa, 1981. Published by tha Patert Office,
`25 Southampton Buildings, London, WC2A (AY, from which coples may be obtained.
`
`HAAG-STREIT AG - EXHIBIT 1032
`Page 6 of 6
`
`HAAG-STREIT AG - EXHIBIT 1032
`Page 6 of 6
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket