throbber
I N T U r T I V E
`S U R G I C A L
`
`SureForm™ 60
`
`S T A P L E R U S E R M A N U A L
`
`I N S T R U M E N T S A N D A C C E S S O R I E S
`U S E R M A N U A L A D D E N D U M
`
`Intuitive Surgical, inc.
`1266 Kifer R°ad
`Sunnyvale,CA 94086 USA
`www.intuitivesurgical.com
`
`| EC I REP I Intuitive Surgical, SSrl
`1, chemin des MCiriers,
`_i
`r •„
`i
`1170 Aubonne Switzerland
`
`0543 C€
`
`PN 553910-03 Rev. A 2018.03
`
`1
`
`IS 1027
`Intuitive Surgical v. Ethicon
`IPR2019-00880
`
`

`

`2
`
`Copyright
`© 2018 Intuitive Surgical, Inc. All rights reserved.
`
`Trademarks
`Intuitive, Intuitive Surgical, Beyond the Limits of the Human Hand, da Vinci, da Vinci S, da Vinci Si,
`da Vinci Xi, da Vinci X, ProGrasp, HotShears, EndoWrist, EndoPass, ClearField, CardioVac, Vacuum
`Source, SureForm, TilePro, and InSite are trademarks or registered trademarks of Intuitive
`Surgical, Inc. Other parties'trademarks are the property of their respective owners and should
`be treated as such.
`
`Rx only
`
`da Vinci Xi and da Vinci X SureForm 60 User Manual Addendum
`
`2
`
`

`

`Contents
`
`1 Introduction
`• 1.1 Contact Information
`• 1.2 About This Manual
`• 1.3 Indications for Use
`J
`• 1.4 Contraindications
`• 1.5 Compatibility Information
`• 1.6 General Warnings and Cautions
`• 1.7 Device Description
`• Articulation Overview
`• 1.8 Stapler Components Defined
`
`2 Preparation and Use
`• 2.1 Open Packaging
`• 2.2 Device Inspection
`• 2.3 Spare Items and Emergency Equipment
`• 2.4 Prepare for Intraoperative Use
`• Install Reload
`• 2.5 Intraoperative Use
`• Precautions for Intraoperative Use
`• Install onto Instrument Arm
`• Calibration Sequence
`• Clamp and Fire Overview
`• Clamp and Fire Instructions
`• SmartFire™
`• Instrument Removal
`• Remove and Replace SureForm 60 Reload
`• Stapler Guided Tool Change
`• Disposal
`
`3 Troubleshooting
`• 3.1 Stapler Initialization Failed
`• 3.2 Missing or Used Reload
`• Missing or Used Reload
`• 3.3 SureForm 60 Reload Recognition
`• SureForm 60 Reload Type Not Detected
`• Dual Console
`• 3.4 Wrong Reload Selected or Reload Shown Mismatch
`• 3.5 Fire Unable to Complete
`• 3.6 Mid-Procedure Restart
`• 3.7 Blade Exposed
`• 3.8 Force Unclamp
`
`da VinciXi and da Vinci X SureForm 60 User Manual Addendum
`
`Contents
`
`3
`
`5
`5
`5
`6
`6
`6
`6
`9
`9
`10
`
`14
`14
`14
`14
`14
`15
`16
`16
`17
`18
`19
`19
`22
`24
`24
`26
`26
`
`27
`27
`27
`27
`28
`28
`30
`30
`30
`31
`31
`32
`
`3
`
`

`

`• 3.9 Manual Stapler Release Instructions
`
`A Appendix A: MR) Safety Information
`• RF Heating
`• MR Artifact
`
`B Appendix B: Symbols and Icons Reference Table
`
`C Appendix C: Natural Rubber Latex
`
`D Appendix D: Sterilization Methods
`
`da VinciXi and da Vinci X SureForm 60 User Manual Addendum
`
`4
`
`

`

`Introduction
`
`Introduction
`
`This addendum to the da Vinci® System Instruments and Accessories User Manual provides
`details specific to the SureForm 60 and SureForm 60 reloads. This document is not a reference
`for surgical techniques. Refer to the appropriate da Vinci System User Manual, for full system
`information and warnings.
`Failure to properly follow all instructions, including instructions supplied with the
`da Vinci System, may lead to injury and ^result in improper functioning of the device.
`
`1.1 Contact Information
`For Customer Service and Reporting of Complaints or Adverse Events
`Use the following information for customer service, including ordering, reporting complaints
`or adverse events, and general information regarding Intuitive Surgical or our products and
`services.
`
`In the U.S.
`Intuitive Surgical, Inc.
`1266 KiferRoad
`Sunnyvale, CA 94086 USA
`Toll free: 1.800.876.1310
`Direct: 408.523.2100
`Fax: 408.523.2377
`
`In Europe
`Intuitive Surgical
`1, chemin des Muriers,
`1170 Aubonne, Switzerland
`Toll free: +800.0821.2020
`Direct: +41.21.821.2020
`Fax: +41.21.821.2021
`
`For Technical Support
`If the system requires maintenance or service, please call ourTechnical Support line. In the US,
`call 1.800.876.1310, where phones are staffed 24 hours a day, seven days a week. In Europe,
`call+41.21.821.2020.
`
`Manufacturer
`
`_
`
`Intuitive Surgical, Inc.
`
`•aaI 1266 Kifer Road
`mi
`
`Sunnyvale, CA 94086 USA
`www.intuitivesurgical.com
`
`1.2 About This Manual
`This manual presents representative images of the stapler instrument. The representative
`images provide instruction on the use of the product. In some cases, the aesthetics or length
`of the device shown in the representative images may be different from the product in use.
`
`da VinciXi and da Vinci X SureForm 60 User Manual Addendum
`
`5
`
`

`

`Introduction
`
`1.3 Indications for Use
`SureForm 60
`The Intuitive Surgical SureForm 60, SureForm 60 reloads, and other stapler accessories are
`intended to be used with da Vinci Surgical Systems for resection, transection, and, or creation
`of anastomoses in General, Thoracic, Gynecologic, Urologic, and Pediatric surgery. The device
`can be used with staple line or tissue buttressing material (natural or synthetic).
`
`Additional Considerations - Precautions for Use of the Stapler in Smaller Patients
`Performance in pediatric surgical procedures is based on similarity of tasks and target tissues
`encountered in adult surgical procedures. As is appropriate with any surgical procedure,
`consideration must be given to patient size and workspace volume when using the stapler.
`
`1.4 Contraindications
`• Do not use the SureForm 60 or SureForm 60 reloads on the aorta.
`
`1.5 Compatibility Information
`The stapler referenced in this manual is compatible for use with the:
`
`• da Vinci Xi Surgical System, model IS4000 (referred to in this manual as the da Vinci Xi
`System, or the system)
`
`• da Vinci X Surgical System, model IS4200 (referred to in this manual as the da Vinci X
`System, or the system)
`
`It is not compatible for use with the da Vinci, da Vinci S, or da Vinci Si Systems.
`The stapler is compatible for use with the accessories listed in Table 1-1 on page 10.
`
`The SureForm 60 is only compatible with the SureForm 60 reloads referenced in this manual.
`The stapler is not compatible with third-party reloads. Do not use the stapler with
`incompatible reloads.
`
`1.6 General Warnings and Cautions
`/i\ WARNING: Proper surgical procedures and techniques are the responsibility of the
`medical professional. Procedures described herein are furnished for information
`purposes only. Each surgeon must evaluate the appropriateness of the procedure based
`on their own medical training and experience, the type of surgical procedure, and the
`type of systems utilized.
`
`^ WARNING: Only surgeons who have developed adequate robotic skills to perform the
`tasks associated with each procedure and who have received specific training provided
`by Intuitive Surgical, Inc., in the use of the da Vinci Xi System and, or the da Vinci X
`System should use the system. Training provided by Intuitive Surgical is limited to the
`use of the da Vinci Xi System and, or the da Vinci X System and does not replace the
`necessary medical training and experience required to perform surgery.
`
`da Vinci Xi and da Vinci X SureForm 60 User Manual Addendum
`
`6
`
`

`

`Introduction
`
`a WARNING: Only surgeons who have received specific training provided by
`Intuitive Surgical, Inc., in the use of the stapler should use the instrument. Training
`provided by Intuitive Surgical is limited to the use of the stapler and does not replace the
`necessary medical training and experience required to perform surgery.
`
`^ WARNING: Be sure to read and understand all information, particularly the caution and
`warning information, found in the applicable user manuals before using these products.
`Failure to properly follow all instructions, including those in the da VinciXi System User
`Manual or the da Vinci X System User Manual, and instructions supplied with accessory
`devices like generators, may lead to injury and result in improper functioning of the
`device.
`
`^ WARNING: Do not use the stapler where adequacy of hemostasis cannot be verified
`visually after staple line formation.
`
`^ WARNING: Do not use the white reloads on any tissue that can be easily compressed to
`less than 1.0 mm in thickness, or on any tissue that cannot comfortably compress to
`1.0 mm.
`
`^ WARNING: Do not use the blue reloads on any tissue that can be easily compressed to
`less than 1.5 mm in thickness, or on any tissue that cannot comfortably compress to
`1.5 mm.
`
`^ WARNING: Do not use the green reloads on any tissue that can be easily compressed to
`less than 2.0 mm in thickness, or on any tissue that cannot comfortably compress to
`2.0 mm.
`A WARNING: Do not use the black reloads on any tissue that can be easily compressed to
`less than 2.3 mm in thickness, or on any tissue that cannot comfortably compress to
`2.3 mm.
`
`^WARNING:
`• Always inspect the tissue thickness and select an appropriate stapler reload before
`application of a staple line using the stapler.
`
`•
`
`Improper reload color selection (staple size) can result in over-compression or
`under-compression of tissue and improper staple formation.
`
`^ WARNING: Preoperative radiotherapy may result in changes to tissue. These changes
`may, for example, cause the tissue thickness to exceed the indicated range for the
`selected staple size. Careful consideration should be given to any pre-surgical treatment
`the patient may have undergone.
`
`^ WARNING: Never activate an electrocautery instrument while in contact with the stapler.
`
`^ WARNING: Eliminate tension when stapling on vasculature, as the seal may be
`compromised.
`
`^ WARNING: After firing, always inspect the staple line for hemostasis. Minor bleeding
`may be controlled by electrocautery or manual sutures.
`
`^ WARNING: Use instruments with care. Avoid contact between instruments inside the
`patient and do not use any instrument to apply force to the stapler inside the patient.
`
`da Vinci Xi and da Vinci X Sure Form 60 User Manual Addendum
`
`7
`
`

`

`Introduction
`
`a WARNING: Remove any loose staples, needles or metal clips from between instrument
`jaws and from the surrounding target tissue before and after stapling.
`
`^WARNING: Do not use an instrument to clean debris from another instrument
`intraoperatively. This may result in damage to the instruments and other unintended
`consequences, such as disengagement of the stapler reload.
`
`A WARNING: The staples are made of titanium alloy. Ensure that patients do not have
`metal allergies.
`
`^CAUTION: The stapler is shipped sterile. Do not reuse or re-sterilize the stapler if it has
`been removed from its package, even if unused. Attempts to do so could cause device
`malfunction and serious injury to the patient.
`
`^CAUTION: Stapler reloads are shipped sterile. Do not reuse or re-sterilize stapler reloads
`that have been removed from their package, even if unused. Attempts to do so could
`cause device malfunction and serious injury to the patient.
`
`^CAUTION: Endoscopic instruments are designed and manufactured for a specific
`surgical function. Use of an instrument for a task other than that for which it is intended
`may result in a damaged or broken instrument.
`
`/j\ CAUTION: Handle the instruments with care. Avoid mechanical shock or stress that can
`cause damage to the instruments.
`
`/f\ CAUTION: As a general precaution, alternative methods to establish adequate
`hemostasis or tissue approximation should be held readily available in the operating
`room.
`
`^CAUTION: Always close the jaws and straighten the wrist of the stapler before
`introducing or removing the instrument through the cannula.
`
`Caution for single-use devices
`
`A breach in the sterile packaging of a single-use device indicates possible
`contamination. Do not use the device if the packaging is not intact.
`
`Caution for MR Environment Compatibility
`/^CAUTION: The RF heating behavior does not scale with static field strength. Devices that
`do not exhibit detectable heating at one field strength may exhibit high values of
`localized heating at another field strength. See Appendix A: MRI Safety Information on
`page 35 for more information.
`
`da VinciXi and da Vinci X SureForm 60 User Manual Addendum
`
`8
`
`

`

`Introduction
`
`1.7 Device Description
`The SureForm 60, when used with the compatible SureForm 60 reloads, delivers multiple rows
`of staples and transects the tissue along the middle of the staple line. The SureForm 60 and
`the SureForm 60 reloads (available in various sizes) are single-use/disposable devices.
`
`Setup and intraoperative use of the stapler may require sterile and non-sterile tasks
`performed by the following users: surgeon, circulating nurse (non-sterile user) or scrub nurse
`(sterile user).
`
`Articulation Overview
`The stapler allows the da Vinci surgeon to position, clamp, and fire (simultaneously staple and
`cut) target tissue using the da Vinci System.
`
`The stapler features fully-wristed articulation with grip, full roll and extensive up-and-down
`and left-to-right range of motion up to 60 degrees in each direction (Figure 1.1). First-time
`users should become familiar with the stapler's long jaws and their full range of motion
`characteristics before clinical use.
`
`Up
`
`A
`
`Right v m
`
`t r
`
`Down
`
`Figure 1.1 Stapler range of motion
`
`da VinciXi and da Vinci X SureForm 60 User Manual Addendum
`
`9
`
`

`

`Introduction
`
`1.8 Stapler Components Defined
`Table 1-1 shows reusable and disposable components, and compatible accessories for the
`stapler.
`
`Table 1-1 Stapler and Accessories
`Part Number (PN)
`Reusable Component Name
`PN 470375
`12 mm & Stapler Cannula, Standard
`PN 470389
`12 mm & Stapler Cannula, Long
`PN 470376
`12 mm & Stapler Blunt Obturator, Standard
`PN 470390
`12 mm & Stapler Blunt Obturator, Long
`PN 470395
`12 mm & Stapler Bladeless Obturator, Standard
`PN 470396
`12 mm & Stapler Bladeless Obturator, Long
`Disposable Component Name
`Part Number (PN)
`PN 480460
`SureForm 60
`See Table 1-2
`SureForm 60 reloads
`PN 470380
`12 mm & Stapler Cannula Seal
`PN 470403
`12 mm Bladeless Optical Obturator, Standard
`PN 470404
`12 mm Bladeless Optical Obturator, Long
`PN 470381
`1 2 - 8 m m R e d u c e r
`
`Caution for single-use devices
`
`A CAUTION:
`
`DO NOT RE-STERILIZE. (^) 00 N0T RE-USE.
`
`Reprocessing and, or reuse of products intended for single use may result in degraded
`instrument performance or loss of functionality, and in exposure to viral, bacterial,
`fungal, or prionic pathogens.
`
`Stapler: The stapler includes the jaws (anvil and reload channel), shaft, and instrument
`housing (Figure 1.2). The functions of the stapler are initiated by the hand controls (master
`grips) and the associated pair of foot pedals on the Surgeon Console. The stapler provides
`fully-wristed articulation with grip, full roll, and extensive up-and-down and left-to-right range
`of motion used to position the jaws across the target tissue (Figure 1.1). Additionally, the
`clamp feature, which is the same motion as grip, provides higher compression on the tissue
`than grip before staple firing. The fire feature simultaneously activates stapling and cutting of
`tissue with a knife blade. The Reload Channel holds the SureForm 60 reload, and the anvil
`contains features that form the staples to capture tissue.
`
`da VinciXi and da VinciXSureForm 60 User Manual Addendum
`
`10
`
`

`

`Introduction
`
`Stapler Name
`
`Release Button
`
`Manual Release
`Knob
`
`Figure 1.2 Stapler (reload installed)
`
`Etching on Jaws Explained
`Lines are etched on the reload channel to indicate the cut and staple boundaries. Thick lines
`indicate the cut boundaries and thin lines indicate staple boundaries and increments.
`Numbers (where included with staple boundaries or increments) indicate length in
`millimeters (mm). The 0 line closest to the tip shows the distal boundary of the staple line. The
`thick line closest to the hinge shows the proximal boundary of the cut line (Figure 1.3).The cut
`boundaries are also shown on the anvil.
`
`_ Distal staple
`boundary
`
`SIDE
`
`ANVIL
`
`CHANNEL
`
`l-Beam Home
`Position
`
`with SmartFIrem
`
`CUT: Distal
`Boundary
`
`CUT: Proximal
`boundary
`
`Figure 1.3 Etching on jaws explained
`
`da VinciXi and da Vinci X SureForm 60 User Manual Addendum
`
`11
`
`

`

`Introduction
`
`^ Note: The stapler is designed for tissue to be loaded only as far as the CUT boundaries
`on the reload channel. It will not cut tissue past the CUT boundaries shown on the reload
`channel and anvil. Load target tissue between the CUT boundaries shown on the
`instrument jaws. More than one SureForm 60 reload may be necessary if a longer staple
`line is desired.
`
`Fires Remaining
`The SureForm 60 is programmed to complete 12 fires. Once the stapler is installed, if five or
`fewer fires remain, a message displays the number of fires left (Figure 1.4). In addition,
`immediately following the last fire, a message appears indicating the stapler has expired.
`Discard expired staplers immediately.
`
`FIRES REMAINING: 3
`
`R 3
`
`SUREFORM 60 (BLUE)
`
`R
`
`A
`
`Figure 1.4 Fires remaining message example
`For further information on instrument expiration, such as how to view remaining uses or fires,
`see the da Vinci System User Manual.
`Disposable SureForm 60 Reloads: The SureForm 60 reloads are sterile, disposable, single-use
`cartridges that contain staggered rows of unformed titanium alloy staples and a cutting blade.
`SureForm 60 reloads come in various staple sizes, distinguished by color. Each is packaged
`with a staple retainer.
`The staple line produced by the SureForm 60 reload is approximately 60 mm long and the cut
`line is approximately 55 mm long.
`A staple retainer holds the staples in place during installation of the reload into the instrument
`jaws. Remove the staple retainer only after the reload has been installed, and before inserting
`the loaded stapler into the cannula.
`^ Note: Use only da VinciXi or da Vinci X SureForm 60 reloads in the da VinciXi or da VinciX
`SureForm 60. Table 1-2 describes the SureForm 60 reload specifications.
`
`da Vinci Xi and da Vinci X SureForm 60 User Manual Addendum
`
`12
`
`

`

`Introduction
`
`Table 1-2 SureForm 60 Reload Specifications
`
`Reload
`
`Component Name
`
`Part
`Number
`(PN)
`
`Reload
`Length
`(mm)
`
`Staple Leg
`Length
`(mm)
`
`Closed
`Staple
`Height (mm)
`
`Number of
`Staples
`
`Total
`Number of
`Staggered
`Staple Rows
`
`SureForm 60 White
`Reload
`
`SureForm 60 Blue
`Reload
`
`SureForm 60 Green
`Reload
`
`SureForm 60 Black
`Reload
`
`PN 48360W
`
`60
`
`2.5
`
`1.0
`
`90
`
`PN 48360B
`
`60
`
`3.5
`
`1.5
`
`90
`
`PN 48360G
`
`60
`
`4.3
`
`2.0
`
`90
`
`PN 48360T
`
`60
`
`4.6
`
`2.3
`
`90
`
`End of section
`
`6
`
`6
`
`6
`
`6
`
`da VinciXi and da Vinci X SureForm 60 User Manual Addendum
`
`13
`
`

`

`14 Preparation and Use
`
`2 Preparation and Use
`
`2.1 Open Packaging
`1. Inspect the outer box, inner Tyvek lid, and clear plastic blister pack for any potential
`damage or break in the sterile barrier. Make sure to inspect the expiration date on the
`packaging and do not use the device or reload if expired.
`2. Using proper sterile technique, remove the stapler from its packaging and transfer to the
`sterile field. Always use 2 people for transfer (one sterile and one non-sterile). Do not let
`the stapler fall from its packaging on to a surface as damage may possibly occur.
`
`2.2 Device Inspection
`Proper care and handling is essential for satisfactory performance of surgical instruments and
`accessories. Examine the stapler thoroughly - including all of its components, reloads and
`accessories - before and after each use. Do not use the device if any abnormality is found.
`Examples of damage include misaligned or loose jaws and deformed or bent shaft
`components. Use the device for Its Intended purpose only.
`
`2.3 Spare Items and Emergency Equipment
`^ WARNING: The following spare item should be ready for use during the procedure:
`
`• Stapler
`
`2.4 Prepare for Intraoperative Use
`
`^CAUTION: Do not lift the stapler by the jaws when the stapler is not installed on an
`instrument arm. Support the housing or the shaft of the stapler instrument and do not
`manipulate it by the jaws, as this may damage the instrument.
`
`Follow these steps to prepare the stapler for use.
`
`da VinciXi and da Vine/XSureForm 60 User Manual Addendum
`
`14
`
`

`

`Preparation and Use
`
`Install Reload
`
`/j\ WARNING: Make sure the reload channel on the stapler is free of foreign material or
`debris before installing a reload.
`
`^CAUTION: Keep the stapler wrist straight when removing and loading reloads or
`otherwise handling the instrument when it is not installed on an instrument arm.
`
`1. Before installing a reload, ensure that the stapler jaws are fully open. If jaws are closed,
`gently turn the Manual Release Knob in a counterclockwise direction until jaws are fully
`open (Figure 2.1).
`
`Figure 2.1 Open the jaws with Manual Release Knob
`2. Select a reload with the appropriate staple size and closed staple height for the tissue
`thickness of the target tissue. See Table 1-2 on page 13 for reference.
`3. Using sterile technique, remove the reload from its packaging, but do not remove the
`staple retainer.
`4. With the retainer in place, hold the reload by the distal (tapered) end (Figure 2.2).
`
`rr
`
`A
`
`**
`
`3.
`
`Figure 2.2 Reload with staple retainer in place
`
`5. Insert the reload into the reload channel
`(bottom jaw) of the stapler: place the
`proximal end into the reload channel at
`a slight angle.
`
`6. Slide the reload firmly into the reload
`channel as far as it will go.
`
`Figure 2.3 Insert at an angle, then slide in
`
`da VinciXi and da Vinci X SureForm 60 User Manual Addendum
`
`15
`
`

`

`16 Preparation and Use
`
`7. Once firmly inserted, press down on the
`tapered end of the reload until it snaps
`into the reload channel with an audible
`click.
`
`Figure 2.4 Snap reload into reload channel
`
`8. After installation, remove the retainer
`and inspect the reload surface for
`damage before use. Dispose of the
`retainer.
`
`fa WARNING: After installing the reload, remove the staple retainer and inspect the stapler
`reload surface for damage before use. Examples of damage include an exposed blade,
`protruding staple tips, or missing staples. Replace the reload if damage is observed.
`
`Figure 2.5 Remove retainer
`
`2.5 Intraoperative Use
`
`Precautions for Intraoperative Use
`
`^WARNING: Be deliberate in selecting tissue to clamp, as clamp compression may
`damage tissue.
`
`fa WARNING: Make sure that no obstructions (such as clips, staples, bone or other hard
`objects) are between the jaws. Firing over an obstruction may result in incomplete
`cutting action and, or improperly formed staples.
`
`^WARNING: To avoid tissue damage, visually check for pinching of tissue at the
`instrument wrist.
`
`fa WARNING: Ensure the tissue lies flat and is positioned properly between the jaws. Any
`bunching of tissue within the jaws may result in an incomplete staple line.
`
`^ WARNING: While working inside the body cavity, do not collide the stapler with other
`instruments or hard surfaces. Doing so may cause the reload to dislodge or damage the
`instrument.
`
`fa WARNING: If the reload dislodges during use, remove the instrument from the body and
`reseat the reload by hand. Do not attempt to reseat the reload by closing the jaws on the
`dislodged reload.
`
`^WARNING: To prevent patient injury, if any stapler abnormality or malfunction is
`suspected during use, stop using the
`instrument and
`immediately contact
`Intuitive Surgical Customer Service.
`
`da VinciXi and da Vinci X SureForm 60 User Manual Addendum
`
`16
`
`

`

`Preparation and Use
`
`a CAUTION: Do not use excessive force while inserting or removing instruments through
`the cannula because it may damage the seal. If there is significant resistance, stop and
`check the seal for damage. Replace the seal if damaged.
`
`a Note: The stapler will not fire until the instrument is fully clamped.
`
`install onto Instrument Arm
`Q Note: For visibility of system hardware, this section presents images of the Patient Cart
`arm without sterile drapes.
`
`Always use two hands when installing the stapler; one hand holding the jaws closed and the
`other supporting the housing of the stapler. Insert the instrument tip into the cannula and
`slide the instrument housing into the sterile adapter.
`
`Figure 2.6 Inserting the instrument into the cannula and sterile adapter
`
`^Note: For further details and instructions on instrument installation, insertion, and
`removal, see the da Vinci System User Manual.
`
`da VinciXi and da Vinci X SureForm 60 User Manual Addendum
`
`17
`
`

`

`18 Preparation and Use
`
`Calibration Sequence
`To ensure full functionality of the stapler, a calibration/initialization sequence occurs each
`time the stapler is successfully engaged on the instrument arm. If the calibration/initialization
`sequence does not initiate, the instrument may not be properly installed into the sterile
`adapter. Remove and re-install the instrument. While calibrating, the affected instrument LED
`blinks and an on-screen message appears (Figure 2.7).
`
`CALIBRATING. DO NOT REMOVE
`INSTRUMENT: 30%
`R
`3
`
`SUREFORM 60
`
`A
`
`| PiRt
`
`Figure 2.7 Calibration in progress
`
`A WARNING: Always insert the instrument into the patient under visualization.
`
`/K CAUTION: Do not remove the stapler while calibrating.
`
`During calibration, the system can detect and report various issues. Refer to Troubleshooting
`on page 27 for more information.
`When calibration completes, an audio tone sounds and a message appears on the stereo
`viewer and touchscreen (Figure 2.8). The patient-side assistant must clutch the arm and
`advance the instrument tip through the cannula, after which the surgeon can control the
`instrument.
`
`INSTRUMENT TIP IN CANNULA.
`CLUTCH AND ADVANCE.
`
`3
`
`SUREFORM 60 [BLUE]
`
`CLAMP
`
`Figure 2.8 Advance tip through cannula
`
`da VinciXi and da Vinci X SureForm 60 User Manual Addendum
`
`18
`
`

`

`Preparation and Use
`
`Clamp and Fire Overview
`The stapler is activated from the associated pair of Surgeon Console foot pedals by applying
`two controls in succession: first, the associated blue pedal applies clamp, which compresses
`the tissue; then the associated yellow pedal fires the instrument to staple and cut target
`tissue, and provide additional compression. You must clamp before firing. Activate each
`control as described in this section. See Figure 2.9 for reference.
`
`Yellow: Fire
`
`Blue: Clamp
`
`Figure 2.9 Pedal activation
`
`Ciamp and Fire Instructions
`When ready to clamp, follow these instructions:
`1. Position the stapler Jaws across the target tissue. Load target tissue between the cut
`boundaries. Squeeze the hand control to grasp the target tissue. Make sure the jaws are
`not in contact with non-target tissue. If necessary, open the hand control to reposition
`the jaws before applying clamp.
`2. Clamp: To clamp tissue, press and hold the associated blue pedal while holding the hand
`control closed. Figure 2.10 shows on the 3D viewer and touchscreen that clamping is
`occurring. To complete clamping, hold down the associated blue pedal and hold the
`hand control closed until an audible tone sounds, and the "Ready to Fire" message is
`displayed (Figure 2.11). Before the audible tone sounds and the "Ready to Fire" message
`appears, lifting foot off the associated blue pedal or opening the hand control will cause
`the stapler to unclamp.
`
`CLAMPING...
`
`vR
`
`":-i
`
`SUREFORM 60 (BLUE)
`
`R CLAMP
`
`Figure 2.10 Clamping in progress
`
`READY TO FIRE.
`
`iUREFORM 60 (BLUE)
`
`EL FIRE
`
`R CLAMP
`
`Figure 2.11 Ready to Fire
`
`da VinciXi and da VinciX SureForm 60 User Manual Addendum
`
`19
`
`

`

`Preparation and Use
`
`The stapler remains clamped on tissue when foot is lifted off the associated blue pedal. While
`clamped, the instrument has limited range of motion.
`To unclamp without firing, tap the associated blue pedal once. After unclamping is complete,
`match grips to control the instrument as needed.
`If during the clamp cycle the device detects that there is either too much tissue in the crotch
`of the jaws or the tissue is simply too thick to begin firing, the message in Figure 2.12 will
`display. Unclamp the device and either reposition the tissue or upsize the reload.
`
`Figure 2.12 Unable to Clamp
`
`da Vinci Xi and da Vinci X SureForm 60 User Manual Addendum
`
`20
`
`

`

`Preparation and Use
`
`3. Fire (Staple and Cut): Press and hold the associated yellow pedal for three seconds to
`staple and cut the target tissue. The system emits an audible tone and counts down on
`screen "3-2-1" before firing starts (Figure 2.13). This provides the chance to release the
`associated yellow pedal during countdown and prevent firing. When the 3-2-1
`countdown completes, the instrument fires; at this point, firing - both staple and cut -
`continues to completion even if the associated yellow pedal is released. The progress bar
`shows firing progress and the system plays the firing sound through both the Surgeon
`Console and the Patient Cart.
`
`FIRING IN 3...
`
`fl 3

`
`SUREFORM 60 (BLUE)
`
`1 R FIRE
`
`"3-2-1"
`countdown
`while holding
`the
`associated
`yellow pedal
`
`FIRING IN 2...
`
`B
`
`O
`
`SUREFORM 60 (BLUE)
`
`R FIRE
`
`F I R I N G I N 1
`
`o
`
`SUREFORM 60 (BLUE)
`
`R FIRE
`
`1
`
`Fire Progress bar
`corresponds to
`markings on the —
`reload channel
`
`Firing starts
`and cannot
`be stopped
`
`Figure 2.13 3-2-1 Fire sequence
`^)Note: During clamping and while the fire progress bars fill, the hand controls cannot be
`used to move, articulate, or roll the stapler.
`
`da VinciXi and da VinciXSureForm 60 User Manual Addendum
`
`21
`
`

`

`Preparation and Use
`
`When firing completes, the instrument automatically unclamps, as shown on screen (Figure
`2.14) and a message displays to straighten the wrist prior to removal (Figure 2.15). An audible
`tone sounds when unclamping completes.
`
`Figure 2.14 Automatic unclamp
`
`Figure 2.15 Straighten wrist before removing stapler
`4. When unclamping completes, match grips to take control of the Instrument again.
`5. Under endoscopic vision, gently remove the jaws from tissue. Check the site for tissue
`approximation and hemostasis. Minor bleeding can be controlled using electrocautery,
`clips, or sutures.
`6. After the stapler is undamped, straighten the wrist and remove the device from the
`instrument arm.
`QNote: For problems with unclamping, see Manual Stapler Release Instructions on
`page 32.
`
`SmartFire™
`The stapler uses SmartFire to create the best opportunity for a properly formed staple line
`while avoiding potential tissue damage, such as serosal tearing, due to an excessive clamp
`force. SmartFire takes into account multiple variables in order to deliver the appropriate
`compression for the targeted tissue for the SureForm 60 reload selected. The SmartFire
`algorithm can cause the system to pause during firing to allow better tissue compression prior
`to continuing the fire, this happens automatically without the need for user interaction.
`
`da VinciXi and da Vinci X SureForm 60 User Manual Addendum
`
`22
`
`

`

`Preparation and Use
`
`During the firing cycle if additional compression is required, the system pauses, displays the
`"Pausing For Compression" message (Figure 2.16) and plays audio feedback. Once the
`appropriate compression is gained on the target tissue, the system resumes firing. Additional
`pauses can happen during a single firing cycle as SmartFire continuously monitors the
`appropriate variables that influence staple line formation.
`
`PADS ;\G (-OR 110MPRESS'SN
`
`«1
`
`O
`O
`
`SUREFORM 60 [BLUE]
`
`R FIRE
`
`Figure 2.16 Pausing For Compression
`If during the firing cycle the system detects that the tissue simply cannot be compressed
`based on the reload selected without leading to malformed staples or causing damage to
`the tissue, the system displays the "Tissue Too Thick to Continue" message (Figure 2.17).
`
`'TISSUE ICQ THICK TO CONTINU E.
`
`AO i • Mm
`TAP 8LUE PEDAL TO UNCLAMP
`CONSIDER CHAN SING RELOAD
`1 COLOR.
`
`
`
`
`
`3 3 3
`
`
`
`
`
`SUREFORM 60 [BLUE] SUREFORM 60 [BLUE] SUREFORM 60 [BLUE]
`
`R CLAMP
`
`J
`1
`
`Figure 2.17 Tissue Too Thick To Continue
`The progress bar will display how far the firing cycle had progressed before being unable to
`proceed. The tissue is transected to the distance indicated on the firing progress bar. There will
`be at least one fully formed set of staples just beyond the cut progress to ensure proper tissue
`apposition up to this point. Unclamp the device and inspect the area that could not be
`transected to ensure there are no elements that would prevent continuation of the staple line
`with a reload designed for thicker tissue. Consider changing reload colors before continuing.
`
`da VinciXi and da VinciXSureForm 60 User Manual Addendum
`
`23
`
`

`

`Preparation and Use
`
`Instrument Removal
`
`A WARNING: Do not remove the stapler, or any

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket