throbber
Trials@uspto.gov
`571-272-7822
`
`Paper 31
`Entered: April 12, 2022
`
`UNITED STATES PATENT AND TRADEMARK OFFICE
`
`BEFORE THE PATENT TRIAL AND APPEAL BOARD
`
`APPLE INC.,
`Petitioner,
`v.
`MASIMO CORPORATION,
`Patent Owner.
`
`IPR2020-01526
`Patent 6,771,994 B2
`
`
`
`
`
`
`
`
`
`Before JOSIAH C. COCKS, ROBERT L. KINDER, and
`AMANDA F. WIEKER, Administrative Patent Judges.
`KINDER, Administrative Patent Judge.
`
`
`
`JUDGMENT
`Final Written Decision
`Determining the Challenged Claim Unpatentable
`35 U.S.C. § 318(a)
`
`
`
`
`
`

`

`IPR2020-01526
`Patent 6,771,994 B2
`
`
`I.
`
`INTRODUCTION
`
`Background
`A.
`Apple Inc. (“Petitioner”) filed a Petition requesting an inter partes
`review of claim 15 of U.S. Patent No. 6,771,994 B2 (Ex. 1001, “the ’994
`patent”). Paper 2 (“Pet.”). Masimo Corporation (“Patent Owner”) waived
`filing a Preliminary Response. Paper 6. We instituted an inter partes review
`of the challenged claim on all asserted grounds of unpatentability, pursuant
`to 35 U.S.C. § 314. Paper 7 (“Institution Decision” or “Inst. Dec.”).
`After institution, Patent Owner filed a Response (Paper 12, “PO
`Resp.”) to the Petition, Petitioner filed a Reply (Paper 17, “Pet. Reply”), and
`Patent Owner filed a Sur-reply (Paper 19, “Sur-reply”). An oral hearing was
`held on January 19, 2022, and a transcript of the hearing is included in the
`record. Paper 30 (“Tr.”).
`We issue this Final Written Decision pursuant to 35 U.S.C. § 318(a)
`and 37 C.F.R. § 42.73. Based on the record before us and for the reasons set
`forth below, Petitioner has met its burden of showing, by a preponderance of
`the evidence, that claim 15 of the ’994 patent is unpatentable.
`
`Related Matters
`B.
`The parties identify the following matters related to the ’994 patent:
`Masimo Corporation v. Apple Inc., Civil Action No. 8:20-cv-00048
`(C.D. Cal.) (filed Jan. 9, 2020);
`Apple Inc. v. Masimo Corporation, IPR2020-01520 (PTAB
`Aug. 31, 2020) (challenging claims of U.S. Patent No. 10,258,265 B1);
`Apple Inc. v. Masimo Corporation, IPR2020-01521 (PTAB
`Sept. 2, 2020) (challenging claims of U.S. Patent No. 10,292,628 B1);
`
`2
`
`

`

`IPR2020-01526
`Patent 6,771,994 B2
`
`
`Apple Inc. v. Masimo Corporation, IPR2020-01523 (PTAB
`Sept. 9, 2020) (challenging claims of U.S. Patent No. 8,457,703 B2);
`Apple Inc. v. Masimo Corporation, IPR2020-01524 (PTAB
`Aug. 31, 2020) (challenging claims of U.S. Patent No. 10,433,776 B2);
`Apple Inc. v. Masimo Corporation, IPR2020-01536 (PTAB
`Aug. 31, 2020) (challenging claims of U.S. Patent No. 10,588,553 B2);
`Apple Inc. v. Masimo Corporation, IPR2020-01537 (PTAB
`Aug. 31, 2020) (challenging claims of U.S. Patent No. 10,588,553 B2);
`Apple Inc. v. Masimo Corporation, IPR2020-01538 (PTAB
`Sept. 2, 2020) (challenging claims of U.S. Patent No. 10,588,554 B2); and
`Apple Inc. v. Masimo Corporation, IPR2020-01539 (PTAB
`Sept. 2, 2020) (challenging claims of U.S. Patent No. 10,588,554 B2).
`Pet. 68; Paper 3, 2.
`
`The parties further identify certain pending patent applications, as
`well as other issued applications, that claim priority to, or share a priority
`claim with, the ’994 patent. Paper 3, 1.
`
`The ’994 Patent
`C.
`The ’994 patent is titled “Pulse Oximeter Probe-Off Detection
`System,” and issued on August 3, 2004, from U.S. Patent Application
`No. 10/374,303, filed February 24, 2003. Ex. 1001, codes (21), (22), (45),
`(54). The ’994 patent claims priority through a series of applications to
`Provisional Application No. 60/140,000, filed June 18, 1999. Id. at
`codes (60), (62).
`The ’994 patent relates to a pulse oximeter probe that detects when a
`probe has become dislodged from a patient or that acts to prevent a potential
`probe-off condition. Ex. 1001, code (57). The ’994 patent relies on
`
`3
`
`

`

`IPR2020-01526
`Patent 6,771,994 B2
`
`electrical contacts that contact the skin of a patient when the probe is
`properly attached and a number of louvers placed in front of a sensor’s
`photodetector to filter out oblique light rays that do not originate from a
`point in front of the detector. Id. According to one aspect of the invention,
`if the emitter and photodetector are not properly aligned, the photodetector
`will not produce a signal within the valid operating range of the pulse
`oximeter, which may trigger an alarm or warning. Id.
`As depicted in Figure 1 below, pulse oximeter 140 is attached through
`connector 142 to probe 110 and probe 110 comprises LEDs 112, 114, which
`are “preferably configured to produce different wavelengths of light.” Id. at
`3:21–55, Fig. 1.
`
`
`Figure 1 illustrates a schematic of a pulse oximeter system. Id. at 2:30–31.
`The wavelengths of light pass through the flesh of a patient to be detected by
`photodetector 116. Id. at 3:34–37, Fig. 1. The ’994 patent describes certain
`embodiments where the photodetector is placed opposite the light emitters to
`detect transmitted light as it emerges from the user’s body tissue. See id. at
`
`4
`
`

`

`IPR2020-01526
`Patent 6,771,994 B2
`
`1:41–43 (describing the configuration of known pulse oximetry probes as
`positioning the detector “opposite the LED”), 4:19–25 (“the emitters located
`within the probe are spaced opposite the detector assembly 235 . . . such that
`the light from the emitters passes . . . through the finger 250 and is incident
`upon the detector assembly 235”), Figs. 2A–B, 4, 5A–B.
`As illustrated in Figure 5B below, if probe 202 is properly attached
`emitter aperture 220 will be directly in front of detector assembly 235 and
`light rays will pass directly through louvers 502 along direct path 510. Id. at
`6:29–33.
`
`
`Figure 5B illustrates a properly attached probe wherein a number of
`louvers (502) are placed in front of the detector assembly. Ex. 1001, 2:60–
`62.
`
`
`
`
`5
`
`

`

`IPR2020-01526
`Patent 6,771,994 B2
`
`
`D. Claim 15
`Claim 15 is the only challenged claim and it is reproduced below.
`15. [pre] A sensor which generates at least first and second
`intensity signals from a light-sensitive detector which detects
`light of at least first and second wavelengths transmitted through
`body tissue carrying pulsing blood; the sensor comprising:
`[a] at least one light emission device;
`[b] a light sensitive detector; and
`[c] a plurality of louvers positioned over the light sensitive
`detector to accept light from the at least one light emission device
`originating from a general direction of the at least one light
`emission device and then transmitting through body tissue
`carrying pulsing blood, wherein the louvers accept the light when
`the sensor is properly applied to tissue of a patient.
`Ex. 1001, 8:21–35 (bracketed identifiers pre–c added).
`
`Applied References
`E.
`Petitioner relies upon the following references:
`Diab et al., U.S. Patent No. 5,638,818, filed November 1, 1994,
`issued June 17, 1997 (Ex. 1006, “Diab”);
`Benjamin et al., U.S. Patent No. 4,015,595, filed
`September 15, 1975, issued April 5, 1977 (Ex. 1007, “Benjamin”);
`Melby et al., U.S. Patent No. 5,254,388, filed December 20,
`1991, issued October 19, 1993 (Ex. 1008, “Melby”);
`Fine, WO Pub. No. 1996/41566, filed June 6, 1995, published
`December 27, 1996 (Ex. 1009, “Fine”); and,
`Webster, Excerpts from Design of Pulse Oximeters, J.G.
`Webster; Institution of Physics Publishing, 1997 (Ex. 1010,
`“Webster”).
`Pet. 3–4.
`
`6
`
`

`

`IPR2020-01526
`Patent 6,771,994 B2
`
`
`Petitioner also submits, inter alia, the Declaration of Brian W.
`Anthony, Ph.D. (Ex. 1003). Patent Owner submits, inter alia, the
`Declaration of Vijay K. Madisetti, Ph.D. (Ex. 2001). The parties also
`provide deposition testimony from Dr. Anthony and Dr. Madisetti.
`Exs. 1038, 2003.
`
`Asserted Grounds
`F.
`We instituted an inter partes review based on the following grounds.
`Inst. Dec. 7, 29.
`Claim Challenged
`15
`15
`15
`15
`
`Reference(s)/Basis
`Diab, Benjamin, Melby
`Webster, Melby
`Fine
`Fine, Benjamin, Melby
`
`35 U.S.C. §
`103
`103
`103
`103
`
`II. DISCUSSION
`Claim Construction
`A.
`For petitions filed on or after November 13, 2018, a claim shall be
`construed using the same claim construction standard that would be used to
`construe the claim in a civil action under 35 U.S.C. § 282(b), including
`construing the claim in accordance with the ordinary and customary
`meaning of such claim as understood by one of ordinary skill in the art and
`the prosecution history pertaining to the patent. 37 C.F.R. § 42.100(b)
`(2019).
`Petitioner submits that the entirety of clause 15[c] requires
`construction. Pet. 8. Clause 15[c] (set forth above), requires in part, “a
`plurality of louvers positioned over the light sensitive detector to accept light
`from the at least one light emission device originating from a general
`
`7
`
`

`

`IPR2020-01526
`Patent 6,771,994 B2
`
`direction of the at least one light emission device and then transmitting
`through body tissue.” Ex. 1001, 8:29–35 (emphases added). Petitioner
`contends that this limitation requires that the light sensitive detector must
`“be positioned opposite the at least one light emission device.” Pet. 8–9.
`Petitioner seemingly bases this interpretation on one embodiment in the
`Specification, “which only depict[s] and describe[s] the placement of the
`body tissue carrying pulsing blood between the at least one light emission
`device and the light sensitive detector.” Pet. 9 (citing Ex. 1001, 1:41–43
`(“[T]he ‘photodiode is positioned opposite the LED so as to detect the LED
`transmitted light as it emerges from the [body] tissue.’” (second alteration in
`original))).
`Although Petitioner describes one embodiment of the invention that
`may be encompassed by the claim scope, this one embodiment is narrower
`in scope than the claim language of clause 15[c]. For example, the claim
`requires the louvers positioned over the detector to accept light originating
`from a general direction of the at least one light emission device, while the
`light also passes through tissue. Petitioner contends that for this to occur the
`detector must be positioned opposite the light emission device, as depicted
`in Figure 5B. Pet. 9.
`In the Institution Decision, we noted that Petitioner does not explain
`why it is even necessary to adopt its proposed claim construction. Inst.
`Dec. 8–9. We noted that for purposes of our analysis, we accept that the
`embodiment Petitioner relies upon (positioned opposite) is within the claim
`scope of clause 15[c], but we were not prepared to limit the claim to just this
`embodiment without further explanation and evidence. Id. We determined
`that clause 15[c] provides sufficient detail as to the positioning of elements
`
`8
`
`

`

`IPR2020-01526
`Patent 6,771,994 B2
`
`such that further defining the limitation did not seem necessary. We invited
`the parties to further explain why these terms would require construction for
`this proceeding. Id. at 9.
`In its Response, Patent Owner argued that further defining the
`limitations of clause 15[c] is not necessary. PO Resp. 22. Patent Owner
`contends that “Apple’s attempt to inject the non-claimed requirement of the
`detector being opposite the light emission device is unnecessary and
`inappropriate.” Id. Patent Owner provides several reasons in support of this
`position. Id. at 22–26. Likewise, at oral argument Patent Owner maintained
`it was unnecessary to reach the claim construction issue in this proceeding.
`See Tr. 38:3–19.
`In its Reply, Petitioner did not provide any justification as to why it
`would be necessary in this proceeding to construe the limitations of clause
`15[c]. See generally Pet. Reply.
`Based on the final record, it is unnecessary for us to construe the
`limitations of clause 15[c] as requested by Petitioner. We determine that
`clause 15[c] provides sufficient detail as to the positioning of elements such
`that further defining the limitation is not necessary to resolve this
`proceeding. Further, no other terms require construction.
`
`Principles of Law
`B.
`A claim is unpatentable under 35 U.S.C. § 103 if “the differences
`between the subject matter sought to be patented and the prior art are such
`that the subject matter as a whole would have been obvious at the time the
`invention was made to a person having ordinary skill in the art to which said
`subject matter pertains.” KSR Int’l Co. v. Teleflex Inc., 550 U.S. 398, 406
`(2007). The question of obviousness is resolved on the basis of underlying
`
`9
`
`

`

`IPR2020-01526
`Patent 6,771,994 B2
`
`factual determinations, including (1) the scope and content of the prior art;
`(2) any differences between the claimed subject matter and the prior art;
`(3) the level of skill in the art; and (4) objective evidence of non-
`obviousness.1 Graham v. John Deere Co., 383 U.S. 1, 17–18 (1966). When
`evaluating a combination of teachings, we must also “determine whether
`there was an apparent reason to combine the known elements in the fashion
`claimed by the patent at issue.” KSR, 550 U.S. at 418 (citing In re Kahn,
`441 F.3d 977, 988 (Fed. Cir. 2006)). Whether a combination of prior art
`elements would have produced a predictable result weighs in the ultimate
`determination of obviousness. Id. at 416–417.
`In an inter partes review, the petitioner must show with particularity
`why each challenged claim is unpatentable. Harmonic Inc. v. Avid Tech.,
`Inc., 815 F.3d 1356, 1363 (Fed. Cir. 2016); 37 C.F.R. § 42.104(b). The
`burden of persuasion never shifts to Patent Owner. Dynamic Drinkware,
`LLC v. Nat’l Graphics, Inc., 800 F.3d 1375, 1378 (Fed. Cir. 2015).
`We analyze the challenges presented in the Petition in accordance
`with the above-stated principles.
`
`Level of Ordinary Skill in the Art
`C.
`Petitioner identifies the appropriate level of skill in the art as that
`possessed by a person having “a Bachelor of Science degree in an academic
`discipline emphasizing the design of electrical, computer, or software
`technologies, in combination with training or at least one to two years of
`related work experience with capture and processing of data or information,
`including but not limited to physiological monitoring technologies.” Pet. 7–
`
`
`1 Neither party has introduced objective evidence of non-obviousness.
`
`10
`
`

`

`IPR2020-01526
`Patent 6,771,994 B2
`
`8 (citing Ex. 1003 ¶¶ 1–15, 36–37). “Additional education in a relevant field
`or industry experience may compensate for one of the other aspects of the
`[person of ordinary skill in the art] POSITA characteristics stated above.”
`Id.
`
`Patent Owner makes several observations regarding Petitioner’s
`identified level of skill in the art but, “for this proceeding, Masimo
`nonetheless applies Apple’s asserted level of skill.” PO Resp. 21.
`We adopt Petitioner’s assessment as set forth above, which appears
`consistent with the level of skill reflected in the Specification and prior art.
`
`D. Obviousness over Webster and Melby
`Petitioner contends that claim 15 of the ’994 patent would have been
`obvious over the teachings of Webster and Melby. Pet. 29–44; see also Pet.
`Reply 4–5. Patent Owner disagrees. PO Resp. 42–46; see also Sur-reply
`12–15.
`Based on our review of the parties’ arguments and the cited evidence
`in the final record, we determine that Petitioner has met its burden of
`showing by a preponderance of the evidence that claim 15 is unpatentable.
`
`Overview of Webster (Ex. 1010)
`1.
`Webster is a book titled “Design of Pulse Oximeters” that provides an
`overview of the history and design of pulse oximeters, stating that “[p]ulse
`oximetry was introduced in 1983 as a noninvasive method for monitoring
`the arterial oxygen saturation of a patient’s blood.” Ex. 1010, xv2. Webster
`
`
`2 Although Petitioner has added page numbering to the bottom of
`Exhibit 1010, Petitioner cites to the original page numbers at the top of the
`book. We adopt Petitioner’s citations for clarity.
`
`11
`
`

`

`IPR2020-01526
`Patent 6,771,994 B2
`
`details “both the hardware and software required to fabricate a pulse
`oximeter as well as the equations, methods, and software required for
`effective functioning” in addition to “testing methods.” Id. Webster
`provides a comprehensive summary of the state of the art as of 1997, and
`also describes the purpose of components of a pulse oximeter as well as the
`design process and selection criteria for particular components. See
`generally Ex. 1010, Chs. 2, 5–7.
`As shown in FIG. 7.1 of Webster, reproduced below, Webster
`explains that the “probe of a pulse oximeter consists of two LEDs of selected
`wavelengths and a detector.” Ex. 1010, 86.
`
`
`Figure 7.1 of Webster depicts a probe using a transmittance principle such
`that light from two LEDs passes alternately through the tissue of the finger
`and is detected by the photodiode. Id. at 87. The LEDs can emit
`wavelengths of, for example, “660 nm and 940 nm,” and “the detector used
`is a photodiode.” Id. at 86. Webster explains that its “photodiode has to
`detect the light transmitted through the tissue” and is “placed in line with the
`LEDs so that the maximum amount of transmitted light is detected.” Id. at
`
`12
`
`

`

`IPR2020-01526
`Patent 6,771,994 B2
`
`87. Because light from the LEDs “is partially reflected, transmitted,
`absorbed, and scattered by the skin and other tissues and the blood before it
`reaches the detector,” reducing the amount of light that eventually reaches
`the detector, the assembly “must be protected from the ambient light for the
`wavelengths to which the photodiode is sensitive.” Id. at 86. Because the
`pulse oximeter “uses two specific wavelengths, spectral response, or the
`relative response of the device to different wavelengths must be considered.”
`Id. at 71 (emphasis omitted).
`Webster describes how the LEDs can be “powered alternately so that
`light of one particular wavelength will pass through the tissue, and the
`transmitted light will be detected by the photodiode” before light of the other
`wavelength is emitted. Ex. 1010, 86. Webster notes that the “intensity of
`the light emerging from the tissue is attenuated by the amount of blood
`present in the tissue.” Id. Because this intensity “varies with the arterial
`pulse,” it can be used “as a measure to indicate the pulse rate.” Id.
`As seen in Figure 7.13 below, Webster describes the importance of
`preventing light other than the actual signals of interest (e.g., light that has
`passed through the tissue carrying pulsing blood) from reaching the detector.
`See Ex. 1010, 95 (describing optical shunting as the condition when light
`from the sensor’s LEDs reaches the photodiode without passing through the
`tissue).
`
`13
`
`

`

`IPR2020-01526
`Patent 6,771,994 B2
`
`
`
`
`Figure 7.13 of Webster depicts light that does not pass through arterioles
`causing optical shunting. Id. Webster states that optical shunting can occur
`“when light from the sensor’s LEDs reaches the photodiode without passing
`through the tissue” and “leads to erroneous readings in the value of SaO2 as
`the amount of light detected by the photodiode is greatly increased by
`optical shunting.” Id.
`Webster proposes that, in order to “minimize the effects from light
`other than the optical signals of interest,” “some type of light filter” can be
`placed over the detector. Id. at 79. Using a light filter allows “light of
`wavelengths of interest to pass through the filter but does not allow light of
`other wavelengths to pass through the filter.” Id. Webster contemplates that
`in order for the pulse oximeter device to function effectively, “most of the
`light being transmitted from the LEDs must not reach the photodiode unless
`it has passed through tissue containing arterial blood.” Id. Indeed, Webster
`discloses that in order to “minimize errors, the pulse oximeter designer must
`
`14
`
`

`

`IPR2020-01526
`Patent 6,771,994 B2
`
`attempt to limit the light reaching the photodiode to that which has traveled
`through tissue containing arterial blood” and suggests that “[l]ight
`impervious barriers should be placed between LEDs and the photodiode in
`all areas where the emitted light could reach the photodiode without passing
`through tissue” containing arterial blood. Id.
`
`Overview of Melby (Ex. 1008)
`2.
`Melby is titled “Light Control Film with Reduced Ghost Images.”
`Ex. 1008, code (54). Melby discloses a light control film, or a “louvered
`plastic film.” Id. at code (57). Melby describes a film that includes “louver
`elements,” and Melby acknowledges that it was known to cant louver
`elements with respect to a louvered plastic film, which produces a film that
`transmits light in a direction other than perpendicular to the surface of the
`film. Id. at 1:9–22, 3:46–62. Melby describes using louvers having an outer
`portion with a relatively low optical density and an inner portion having a
`relatively high optical density. Id. at 3:21–22.
`Melby describes a “louvered plastic film [that] has louvers including
`central regions with a relatively high coefficient[] of extinction and outer
`regions with relatively low coefficients of extinction.” Ex. 1008, code (57).
`As shown in Figures 1 and 2 of Melby, reproduced below, Melby discloses
`“a louvered plastic film has a plurality of clear regions separated by
`louvers,” where each louver has “a central region with a relatively high
`coefficient of extinction and outer regions, adjacent said clear regions,
`having relatively low coefficients of extinction.” Id. at 3:1–8.
`
`15
`
`

`

`IPR2020-01526
`Patent 6,771,994 B2
`
`
`
`Figure 1 of Melby depicts a schematic cross section of a louvered
`plastic film and Figure 2 depicts an enlarged drawing of a portion of the
`louvered plastic film. Ex. 1008, 3:10–14. Melby describes its louvered film
`as including “cover sheets 11 provided for clarification and includes
`alternating clear layers, such as layer 12 and louvers, such as louver 14,” that
`“includes outer layers 16 and 18 and inner layer 20.” Id. at 3:46–4:25.
`Referring to Figure 2, Melby describes the operation of the louvers:
`A light ray 22 enters transparent layer 12. It then strikes layer 16
`at surface 24. Because layer 16 has only a low concentration of
`carbon black, there is not a large difference in index of refraction
`between it and layer 12. Therefore very little of the light is
`reflected by layer 24. Most of the light will enter layer 16 and be
`refracted. As the light traverses layer 16, some will be absorbed.
`Some, however, will strike layer 20 on surface 26. Some of light
`beam 22 will enter layer 20 where, due to the relatively high
`concentration of carbon black, it will be absorbed. Some of light
`
`16
`
`

`

`IPR2020-01526
`Patent 6,771,994 B2
`
`
`beam 22 will be reflected at surface 26 due to the large difference
`in index of refraction between layer 16 and layer 20.
`Ex. 1008, 3:63–4:10.
`
`Independent Claim 15
`3.
`Petitioner contends that claim 15 would have been obvious over
`Webster and Melby. Pet. 29–44. Below, we set forth how the combination
`of prior art references teaches or suggests the claim limitations that are not
`disputed by the parties. For those limitations and reasons for combining the
`references that are disputed, we examine each of the parties’ contentions and
`then provide our analysis.
`
`i. [15 pre] “A sensor which generates at least first and
`second intensity signals from a light-sensitive detector
`which detects light of at least first and second wavelengths
`transmitted through body tissue carrying pulsing blood;
`the sensor comprising:”
`The cited evidence and argument supports Petitioner’s undisputed
`contention that Webster and Melby satisfy the subject matter of the
`preamble. Pet. 30–32. According to Petitioner, Webster describes devices
`known as “pulse oximeters,” which provide “an empirical measure of
`arterial saturation.” Pet. 30 (quoting Ex. 1010, 13). Webster describes the
`operation of the devices as shining “light of two wavelengths through a
`tissue bed such as the finger or earlobe and measures the transmitted light
`signal.” Id. (quoting Ex. 1010, 13).
`Petitioner relies on Webster’s Figure 7.1 as providing detail as to the
`probe of a pulse oximeter.
`
`17
`
`

`

`IPR2020-01526
`Patent 6,771,994 B2
`
`
`
`Petitioner’s annotated and highlighted Figure 7.1 of Webster depicts a probe
`using transmittance of light from two LEDs (blue labeled as “Light Emission
`Device”) passing alternately through the tissue of the finger such that it is
`detected by a photodiode (green labeled as “Light Sensitive Detector”).
`Pet. 31; Ex. 1010, 87. Petitioner persuasively relies on Webster’s teaching
`of how the “probe of a pulse oximeter consists of two LEDs of selected
`wavelengths and a detector.” Pet. 31; Ex. 1010, 86. Webster describes how
`the LEDs can emit wavelengths of, for example, “660 nm and 940 nm,” and
`“the detector used is a photodiode.” Ex. 1010, 86.
`Petitioner contends that “[b]ecause light from the LEDs ‘is partially
`reflected, transmitted, absorbed, and scattered by the skin and other tissues
`and the blood before it reaches the detector,’ reducing the amount of light
`that eventually reaches the detector, the assembly ‘must be protected from
`the ambient light for the wavelengths to which the photodiode is sensitive.’”
`Pet. 31–32 (quoting Ex. 1010, 86). Petitioner additionally relies on
`Webster’s teaching that the pulse oximeter “uses two specific wavelengths,
`spectral response, or the relative response of the device to different
`
`18
`
`

`

`IPR2020-01526
`Patent 6,771,994 B2
`
`wavelengths must be considered,” as well as Webster’s teaching that the
`intensity of light emerging from the tissue is attenuated by the amount of
`blood present in the tissue. Ex. 1010, 71; Pet. 32.
`
`Relying on the testimony of Dr. Anthony (Ex. 1003 ¶¶ 76–82),
`Petitioner contends that a person of ordinary skill in the art “would have
`recognized Webster to teach a sensor that generates first and second
`intensity signals from a detector, such as a photodiode,” and further, “[t]he
`photodiode is able to detect, and therefore is sensitive to, light of a first
`wavelength, such as 660 nm, and light of a second wavelength, such as
`940 nm, and such light has been ‘attenuated by the amount of blood present
`in the tissue’ of a subject[].” Pet. 32 (quoting Ex. 1010, 86). Petitioner
`argues, and we agree, that Webster explains the advantages of using two
`different wavelengths to allow pulse oximeters to “determine the oxygen
`saturation of arterial blood” by “using the arterial pulsation to differentiate
`between absorbance of arterial blood and other absorbers.” Pet. 32–33
`(quoting Ex. 1010, 13–14, 44).
`
`We find Petitioner’s contentions set forth above supported by a
`preponderance of the evidence on the final record.
`
`ii.“[a] at least one light emission device;”
`Based on the final record, the cited evidence supports Petitioner’s
`undisputed contention that Webster discloses this limitation. Pet. 33–34.
`Specifically, Webster teaches a sensor that measures first and second
`intensity signals using a photodetector, such as a diode, to detect light from
`at least two LEDs emitting at two different wavelengths. Id. (citing
`Ex. 1010, 13–14, 44, 86, Fig. 7.1).
`
`19
`
`

`

`IPR2020-01526
`Patent 6,771,994 B2
`
`
`As shown in Figure 7.1 of Webster, reproduced above, Webster
`teaches two light emitting diodes (LEDs highlighted in blue) that emit light
`at two different wavelengths. Thus, we agree with Petitioner that Webster
`teaches at least one light emission device. Ex. 1010, 79 (Fig. 7.1), 13–14;
`Ex. 1003 ¶ 84. We find Petitioner’s contentions supported by the final
`record.
`
`iii. “[b] a light sensitive detector; and;”
`Based on the final record, the cited evidence supports Petitioner’s
`undisputed contentions regarding this limitation. Pet. 35–36. Specifically,
`Petitioner contends that Webster describes a sensor that measures first and
`second intensity signals using a photodetector, such as a diode, to detect
`light from at least two LEDs emitting at two different wavelengths. Pet. 35
`(citing Ex. 1010, 13–14, 44, 56, 86, Fig. 7.1; Ex. 1003 ¶¶ 76–87).
`As shown in Figure 7.1 of Webster, reproduced above, Webster
`teaches a light sensitive photodiode detector (highlighted in green). Pet. 36
`(citing Ex. 1003 ¶ 88). Petitioner relies on Webster’s disclosure that the
`“photodiode has to detect the light transmitted through the tissue” and it is
`“placed in line with the LEDs so that the maximum amount of transmitted
`light is detected.” Ex. 1010, 87; Pet. 36–37 (also discussing Webster’s
`disclosure of “optical shunting” to increase the amount of light detected by
`the photodiode (citing Ex. 1010, 95)).
`
`iv.“[c] a plurality of louvers positioned over the light
`sensitive detector to accept light from the at least one light
`emission device originating from a general direction of the
`at least one light emission device and then transmitting
`through body tissue carrying pulsing blood, wherein the
`
`20
`
`

`

`IPR2020-01526
`Patent 6,771,994 B2
`
`
`louvers accept the light when the sensor is properly
`applied to tissue of a patient.”
`Petitioner’s Contentions
`Petitioner contends that Webster and Melby teach these limitations.
`Pet. 38–45. Petitioner first notes that Webster proposes the use of a light
`filter to be placed over the detector to minimize the effects from light other
`than the optical signals of interest. Pet. 39 (“Webster proposes that, in order
`to ‘minimize the effects from light other than the optical signals of interest,’
`some type of light filter’ can be placed over the detector.”) (quoting
`Ex. 1010, 79). Specifically, “[u]sing a light filter allows ‘light of
`wavelengths of interest to pass through the filter but does not allow light of
`other wavelengths to pass through the filter.’” Id. (quoting Ex. 1010, 79).
`Petitioner contends that Webster teaches the importance of ensuring
`that light reaching the photodiode must pass through tissue containing
`arterial blood in order for the pulse oximeter to function effectively. Id.
`Petitioner relies on Webster’s disclosure that in order to “minimize errors,
`the pulse oximeter designer must attempt to limit the light reaching the
`photodiode to that which has traveled through tissue containing arterial
`blood” and suggests the use of “[l]ight impervious barriers [that] should be
`placed between LEDs and the photodiode in all areas where the emitted
`light could reach the photodiode without passing through tissue.” Pet. 39
`(quoting Ex. 1010, 79).
`Based on these disclosures, Dr. Anthony testifies that a person of
`ordinary skill in the art would have recognized the advantages of using a
`light filter, such as the type taught by Melby and as suggested by Webster.
`Ex. 1003 ¶¶ 92–96. Dr. Anthony testifies that Melby’s louvered plastic film
`could be used with the combined Webster system as the “light filter” to
`
`21
`
`

`

`IPR2020-01526
`Patent 6,771,994 B2
`
`control light such that only light originating from the direction of the light
`emitters reaches the detector. Id. ¶¶ 96–100; Ex. 1010, 79. Petitioner relies
`on Melby’s teaching of “a ‘louvered plastic film [that] has louvers including
`central regions with a relatively high coefficient[] of extinction and outer
`regions with relatively low coefficients of extinction.’” Pet. 40 (alterations
`in original) (quoting Ex. 1008, code (57)). Further, “Melby teaches ‘a
`louvered plastic film has a plurality of clear regions separated by louvers,’
`where each louver has ‘a central region with a relatively high coefficient of
`extinction and outer regions, adjacent said clear regions, having relatively
`low coefficients of extinction.’” Id. (quoting Ex. 1008, 3:1–8). Petitioner
`points out that Melby’s louvered film includes “cover sheets 11 provided for
`clarification and includes alternating clear layers, such as layer 12 and
`louvers, such as louver 14,” that “includes outer layers 16 and 18 and inner
`layer 20.” Pet. 41 (quoting Ex. 1008, 3:46–4:25).
`Petitioner relies on Melby’s teaching that when “[a] light ray 22
`enters transparent layer 12[,] [i]t then strikes layer 16 at surface 24.”
`Pet. 41 (quoting Ex. 1008, 3:66–67). Petitioner also notes that “very little of
`the light is reflected by layer 24,” instead, “[m]ost of the light will enter
`layer 16 and be refracted.” Id. (quoting Ex. 1008, 3:66–67). According to
`Petitioner, Melby’s “advantage lies with ‘the effective optical density of a
`medium is directly proportional to the distance that the light must travel
`through that

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket