throbber
NIH Public Access
`Author Manuscript
`Brain Stimul. Author manuscript; available in PMC 2013 February 09.
`Published in final edited form as:
`Brain Stimul. 2013 January ; 6(1): 1–13. doi:10.1016/j.brs.2012.02.005.
`
`Electric field depth–focality tradeoff in transcranial magnetic
`stimulation: simulation comparison of 50 coil designs
`
`Zhi-De Deng1,2, Sarah H. Lisanby1,3, and Angel V. Peterchev1,4,*
`Zhi-De Deng: zd2119@columbia.edu; Sarah H. Lisanby: sarah.lisanby@duke.edu; Angel V. Peterchev:
`angel.peterchev@duke.edu
`1Department of Psychiatry and Behavioral Sciences, Duke University, Durham, NC, USA
`2Department of Electrical Engineering, Columbia University, New York, NY, USA
`3Department of Psychology and Neuroscience, Duke University, Durham, NC, USA
`4Department of Biomedical Engineering and Department of Electrical and Computer Engineering,
`Duke University, Durham, NC, USA
`
`Abstract
`Background—Various transcranial magnetic stimulation (TMS) coil designs are available or
`have been proposed. However, key coil characteristics such as electric field focality and
`attenuation in depth have not been adequately compared. Knowledge of the coil focality and depth
`characteristics can help TMS researchers and clinicians with coil selection and interpretation of
`TMS studies.
`Objective—To quantify the electric field focality and depth of penetration of various TMS coils.
`Methods—The electric field distributions induced by 50 TMS coils were simulated in a spherical
`human head model using the finite element method. For each coil design, we quantified the
`electric field penetration by the half-value depth, d1/2, and focality by the tangential spread, S1/2,
`defined as the half-value volume (V1/2) divided by the half-value depth, S1/2 = V1/2/d1/2.
`Results—The 50 TMS coils exhibit a wide range of electric field focality and depth, but all
`followed a depth–focality tradeoff: coils with larger half-value depth cannot be as focal as more
`superficial coils. The ranges of achievable d1/2 are similar between coils producing circular and
`figure-8 electric field patterns, ranging 1.0–3.5 cm and 0.9–3.4 cm, respectively. However,
`figure-8 field coils are more focal, having S1/2 as low as 5 cm2 compared to 34 cm2 for circular
`field coils.
`Conclusions—For any coil design, the ability to directly stimulate deeper brain structures is
`obtained at the expense of inducing wider electrical field spread. Novel coil designs should be
`benchmarked against comparison coils with consistent metrics such as d1/2 and S1/2.
`
`*Corresponding author. Department of Psychiatry and Behavioral Sciences, Duke University, Box 3950 DUMC, Durham, NC 27710,
`tel. +1 919 684 0383, fax +1 919 681 9962, angel.peterchev@duke.edu.
`Disclosure
`Mr. Deng is inventor on patent applications and invention disclosures on TMS technology. Dr. Lisanby has served as Principal
`Investigator on industry-sponsored research grants to Columbia/RFMH or Duke (Brainsway, ANS/St. Jude Medical, NeoSync);
`equipment loans to Columbia or Duke (Magstim, MagVenture); is co-inventor on a patent application on TMS technology; is
`supported by grants from NIH (R01MH091083-01, 5U01MH084241-02, 5R01MH060884-09), Stanley Medical Research Institute,
`and National Alliance for Research on Schizophrenia and Depression; and has no consultancies, speakers bureau memberships, board
`affiliations, or equity holdings in related industries. Dr. Peterchev is inventor on patents and patent applications on TMS technology,
`including TMS technology licensed to Rogue Research; is Principal Investigator on a research grant to Duke from Rogue Research
`and equipment donations to Columbia (Magstim, ANS/St. Jude Medical); has received patent royalties from Rogue Research through
`Columbia; and is supported by NIH grant R01MH091083 and Wallace H. Coulter Foundation Translational Partners grant.
`
`NIH-PA Author Manuscript
`
`NIH-PA Author Manuscript
`
`NIH-PA Author Manuscript
`
`LUMENIS EX1045
`Page 1
`
`

`

`Deng et al.
`
`Page 2
`
`Keywords
`transcranial magnetic stimulation; electric field; depth; focality; simulation
`
`Introduction
`Transcranial magnetic stimulation (TMS) is a non-invasive technique that is used or
`investigated for numerous research and therapeutic applications, including the study of
`normal and pathological brain function and the treatment of neurological and psychiatric
`disorders.1,2 TMS uses brief, intense pulses of electric current delivered to a coil placed on
`the subject’s head to generate an electric field in the brain via electromagnetic induction.
`The induced electric field modulates the neural transmembrane potentials and, thereby,
`neural activity. The locus of activation in the brain is approximately in the area where the
`induced electrical field is maximal3,4; this location, in turn, depends on the stimulating coil
`geometry and placement. For the purpose of determining a map of direct neural activation
`by TMS, the induced electric field distributions generated by different coil types have been
`characterized by theoretical calculations,5–13 numerical simulation models,14–21 and
`measurements of the electric currents induced in phantoms22,23 or in vivo.24,25 However, the
`analytic studies use idealized circular and/or figure-8 coil geometries, and only a handful of
`commercial coils have been modeled in computational studies.21,26–28 Thus, electric field
`distribution data for many commercial or experimental TMS coils are still lacking.
`Knowledge of the electric field spatial distribution of specific coils and how it compares to
`other coils is valuable in the design and interpretation of basic research and clinical studies,
`as well as in the development of novel coils.
`
`Two electric field spatial features of particular interest are depth of penetration and focality.
`Proposed or implemented coil designs have often been developed with the objective of
`improving one or both of these field characteristics. For example, in applications where the
`stimulation target is small (e.g., hand muscle representations in the primary motor cortex),
`localization of the induced electric field is important in order to minimize the stimulation of
`non-target regions. There has also been substantial interest in direct, non-invasive
`stimulation of brain regions deeper than the superficial cortex. For example, the subcallosal
`cingulate cortex, which lies approximately 6 cm from the head surface, is a putative target
`for the treatment of depression.29,30 However, the design of TMS coils to stimulate such
`deep brain targets is limited by the rapid attenuation of the electric field in depth. It has been
`mathematically proven that at the quasi-static frequencies used in TMS the electric field is
`always stronger on the surface than inside of a spherically symmetric volume conductor.31
`Further, the electric field in the center of a uniformly conducting sphere is always
`zero.10,13,32,33 Therefore, direct stimulation with TMS of regions near the center of the head
`appears impossible.
`
`It is known that coils with larger dimensions generate an electric field that penetrates deeper
`but is less focal than that of smaller coils.3,33–35 However, this electric field depth–focality
`tradeoff is not always acknowledged when novel TMS coil designs are proposed and has not
`been characterized with a uniform set of metrics. This is a serious limitation to clinical and
`basic neuroscience applications because stimulation of non-target brain regions may affect
`clinical outcomes, and certainly affects the degree to which any observed changes in
`behavior can be attributed to stimulation of the deep target alone rather than the neighboring
`regions that are also affected by the broader field. Furthermore, the electric field depth and
`focality may relate to risk of accidental seizure and other adverse side effects. Consequently,
`the safety guidelines for the use of repetitive TMS which are based on studies with ~70 mm
`
`Brain Stimul. Author manuscript; available in PMC 2013 February 09.
`
`NIH-PA Author Manuscript
`
`NIH-PA Author Manuscript
`
`NIH-PA Author Manuscript
`
`LUMENIS EX1045
`Page 2
`
`

`

`Deng et al.
`
`Page 3
`
`figure-8 coils36 should be applied with caution to coil designs generating significantly
`different electric field characteristics.
`
`The first TMS system used a circular coil (see, e.g., Figure 1(2, 4, 5)) due to its simple
`geometry and, hence, ease of construction.37,38 However, the circular coil induces a
`nonfocal ring-shaped electric field maximum potentially stimulating a swath of brain regions
`under the coil perimeter. Attempts have been made to focalize the stimulation site with
`circular coils by introducing an angulated extension in the winding39 or by modifying the
`winding density40,41 or concavity.42 These approaches, however, only marginally enhance
`the circular coil’s ability to produce a focal field.43 Thus far, the most significant
`improvement of TMS focality has been the introduction of the figure-8 coil (see, e.g., Figure
`1(30, 31)), which has generally been credited to Ueno and colleagues,44 although this coil
`configuration had been proposed earlier for the purpose of more localized sensing of
`magnetoencephalographic sources,45 which is the inverse problem of focal TMS. The
`figure-8 configuration consists of a pair of adjacent circular loops with current flow in
`opposite directions, producing a relatively focal electric field maximum under the center of
`the coil where the two loops meet.44 Cohen and Cuffin studied the focality of the figure-8
`coil and found improvements with decreasing loop diameter down to 2.5 cm, at which point
`heating and stress limitations would necessitate sophisticated coil fabrication
`techniques.32,46 Nevertheless, the search for even more focal TMS coils continues. The
`“cloverleaf” coil design consists of four sets of nearly circular windings (Figure 1(45)) and
`has been shown to be more efficient in stimulating long fibers compared to the figure-8
`coil.34,47–51 The “slinky” coil design consists of multiple circular or rectangular loops joined
`together at one edge and fanned out to form a half toroid (Figure 1(23, 24)).12,52–55
`Knäulein and Weyh introduced a figure-8 coil with eccentric windings producing higher
`density of winding turns toward the center of the coil than in the periphery (Figure 1(22)),56
`which has been shown to have better focality compared to the regular figure-8 and slinky
`coils of the same outer loop diameter.57–59 The 3-D differential coil (Figure 1(41)) consists
`of a small figure-8 coil with a third loop positioned perpendicular to the center of the
`figure-8 coil and flanked by two additional loops to restrict the area of stimulation.60 This 3-
`D differential coil has been shown to provide more focal stimulation compared to the
`figure-8 and slinky coils.60,61 Other efforts to control the TMS electric field focality include
`the use of litz wire to construct small diameter figure-8 coils (Figure 1(21)),62,63 conductive
`shielding plates (Figure 1(32)),64–68 active shields (Figure 1(33)),68,69 and ferromagnetic
`cores (Figure 1(3, 10, 11, 34, 40)).28,70
`
`In parallel with the attempts to optimize the focality of magnetic stimulation, there have
`been efforts to increase the depth of stimulation. With conventional circular or figure-8
`coils, the TMS-induced electric field is restricted to superficial cortical targets due to its
`rapid attenuation in depth. The double cone coil, formed by two large adjacent circular
`windings fixed at an angle (Figure 1(37)), has been shown to induce a more deeply
`penetrating and less focal electric field compared to a planar 70-mm figure-8 coil.27 The
`double cone coil has been used for activation of the pelvic floor and lower limb motor
`representation at the interhemispheric fissure,71 as well as transsynaptic activation of the
`anterior cingulate cortex via stimulation of the medial frontal cortex.72 This coil is also
`highly efficient at seizure induction,73 which is an advantage in the context of magnetic
`seizure therapy, but a potential source of risk in subconvulsive applications.
`
`A family of coil designs called Hesed (H) coils has been proposed to achieve effective
`stimulation of deep brain structures.74–77 The H coils have complex winding patterns and
`larger dimensions compared to conventional TMS coils, and consequently can be expected
`to have slower electric field attenuation with depth, at the expense of reduced focality.
`Measurements of the electric field induced by some H coils in a saline-filled phantom brain
`
`Brain Stimul. Author manuscript; available in PMC 2013 February 09.
`
`NIH-PA Author Manuscript
`
`NIH-PA Author Manuscript
`
`NIH-PA Author Manuscript
`
`LUMENIS EX1045
`Page 3
`
`

`

`Deng et al.
`
`Page 4
`
`reveal circular field patterns that are similar to those produced by large circular coils.76 It
`has been proposed to use high-permeability ferromagnetic cores to improve the electric
`efficiency, field penetration, and focality of H coils.78,79 The magnetic field associated with
`a particular coil design can be influenced by placing a ferromagnetic core within the field.
`However, the reported improvements from placing ferromagnetic cores in H coils were
`minor.78,79 H coils are currently being evaluated for the treatment of a variety of psychiatric
`and neurological disorders, including major depression,80–87 schizophrenia,88,89 dystonia,90
`autism,91 pain relief,92 and chronic migraine.93 However, the comparative advantage of H
`coils over conventional 70-mm figure-8 coils with regard to depth of stimulation has been
`disputed.94–96 In addition, a randomized, sham-controlled, observer-blinded study in
`patients with benign essential blepharospasm has shown that the H coil used in that study
`has similar clinical effects to a 90-mm circular coil.90
`
`Other designs for deep brain TMS coils have been proposed, such as the stretched C-shaped
`ferromagnetic core coil,3,68,97 circular crown coil,3 and large halo coil.98 The halo coil has
`been shown to elicit lower limb and trunk muscle contractions in a cynomolgus monkey,
`suggesting that this coil design is capable of directly stimulating relatively deep brain
`regions.99 The effects on the brain of low-field magnetic stimulation using MRI gradient
`coils have also been investigated.100–102 These deep TMS and MRI coils have larger
`dimensions than conventional coils and H coils, and are expected to provide slower decay
`rate of the electric field with distance, at the expense of reduced focality.
`
`Another line of work on improving magnetic stimulation targeting concerns the use of
`independently-controlled multi-channel coil arrays.103–115 The major advantage of such a
`system is the flexibility of controlling the spatial electric field distribution and the ability to
`stimulate several brain areas simultaneously.111 However, multi-channel arrays do not
`circumvent the fundamental limitations of electric field targeting.31 Furthermore, high
`resolution targeting with multi-channel TMS requires small coil windings. Unfortunately,
`decreasing coil size is associated with higher electric losses, mechanical stress, and driving
`current.9 The need for sophisticated control electronics further complicates the design.
`Another putative advantage of the independently-controlled multi-channel system is
`temporal summation of spatially distinct electric field pulses. Specifically, it has been
`proposed that the neuronal activation threshold in depth can be reduced relative to the
`superficial neural threshold by applying several consecutive pulses from a set of coils at
`different spatial locations.77,116 However, our previous study showed theoretically that
`synchronous firing of all TMS coil channels is more effective for stimulating deep neurons
`than sequential firing.3
`
`Understanding the electric field depth–focality tradeoff could help TMS clinicians and
`researchers with coil selection and interpretation of TMS studies, and can inform the
`selection of coil designs for magnetic seizure therapy. In the present paper we systematically
`compare the electric field penetration and focality for 50 commercial and experimental TMS
`coil designs in a spherical head model. Even though more anatomically accurate head
`models have been developed to reveal local detail of the TMS electric field and current
`density distribution,17,19,117–121 the spherical model remains a useful reduction for
`parametric and comparative studies of global TMS coil characteristics such as electric field
`penetration and focality.3,9,13,15,32–34,122 The results obtained with the spherical model are
`in reasonable agreement with a variety of quantitative and qualitative observations of
`TMS,121,123–126 and are not limited to a particular subject’s head anatomy or coil placement,
`which is an advantage for comparative studies of global TMS coil characteristics.
`Furthermore, the spherical model provides a standardized measure that can be easily
`replicated by various researchers evaluating coil designs.
`
`Brain Stimul. Author manuscript; available in PMC 2013 February 09.
`
`NIH-PA Author Manuscript
`
`NIH-PA Author Manuscript
`
`NIH-PA Author Manuscript
`
`LUMENIS EX1045
`Page 4
`
`

`

`Deng et al.
`
`Page 5
`
`This study has been presented in part in abstract form.127
`
`Methods
`Electric field simulation
`The TMS coil and human head models, and the electric field solution were implemented
`with the finite element method package MagNet (Infolytica, Inc., Canada). The human head
`was modeled by a homogeneous sphere with 8.5 cm radius and isotropic conductivity of
`0.33 S m−1. The cortical surface was assumed to be at a depth of 1.5 cm from the surface of
`the head.128 The distinct head tissue layers (scalp, skull, corticospinal fluid, and brain) were
`not differentiated, since magnetically induced electric field in a sphere is insensitive to radial
`variations of conductivity.13
`
`We modeled 50 TMS coil configurations shown in Figure 1. In addition to modeling
`commercial coils from Brainsway (Jerusalem, Israel), Cadwell (Kennewick, WA, USA),
`Magstim (Whitland, Wales, UK), MagVenture (Farum, Denmark), and Neuronetics
`(Malvern, PA, USA), we have also included various coil designs proposed for enhancing the
`field focality or penetration. The MagVenture twin coil (Figure 1(36) and Figure 5) was
`modeled for various opening angles between the centers of the two loops to examine the
`effect of this parameter on the electric field depth–focality profile. The coils were modeled
`based on published data, manufacturer’s specifications, coil X-rays, and inductance
`measurements. Detailed descriptions of the various coils are provided in Table S1 in the
`Supplementary Material. The minimum spacing between the coil windings and the surface
`of the head model was 5 mm to account for the thickness of the coil insulation.
`Ferromagnetic cores were modeled with a linear, homogeneous, isotropic material, with
`relative permeability of 1000 and electrical conductivity of 1 S m−1.79 Finally, we simulated
`the “flux ball” coil whose windings are parallel to the circles of latitude of the spherical
`model and cover the whole head (Figure 1, coil #0).129 While this coil configuration is not
`physically realizable for brain stimulation, it illustrates the limiting case for large coils
`where the magnetic field intensity within the sphere becomes uniform, resulting in linear—
`the slowest possible—electric field attenuation in depth.3,129
`
`Some clinical studies using H coils have not provided details of the coil geometry90–93 or
`were published after the initial submission of this paper.130 Since geometric details of these
`coils were not available to us for this study, these H coils were not modeled.
`
`The electric field was computed with the 3-D electromagnetic time-harmonic solver of
`MagNet. MagNet first solves for the magnetic field H via the edge-element version of the T-
`Ω method.131,132 Fields are assumed to be time-harmonic with angular frequency ω =2π×5
`kHz. MagNet solves the equation
`
`V x [ (a-+jwtT 1V x u] +jwµH=O
`
`(1)
`
`where σ is the electrical conductivity, and ε and μ are electric permittivity and magnetic
`permeability of the medium, respectively. The tissue permeability was set to that of free
`space since head tissues are non-magnetic. The tissue permittivity was also set to that of free
`space since the quasistatic approximation is applicable to TMS121 and since permittivity will
`not affect the normalized electric field depth and focality metrics used in this study, as
`discussed in the next section. The electric field was subsequently computed using Ampère’s
`and Ohm’s Laws
`
`Brain Stimul. Author manuscript; available in PMC 2013 February 09.
`
`NIH-PA Author Manuscript
`
`NIH-PA Author Manuscript
`
`NIH-PA Author Manuscript
`
`LUMENIS EX1045
`Page 5
`
`

`

`Deng et al.
`
`Page 6
`
`Electric field characterization
`We compared the overall electric field distribution features of the various coils by plotting
`the field magnitude and direction on the brain surface (Figure 2).
`
`(2)
`
`We quantified the electric field penetration by the half-value depth, d1/2, defined as the
`radial distance from the cortical surface to the deepest point where the electric field strength
`E is half of its maximum value on the cortical surface, Emax.133 The d1/2 metric has units of
`length and quantifies the extent of the field along the radial brain dimension.
`
`Focality is traditionally defined as the area where the electric field strength exceeds a certain
`value relative to the maximum at a given depth (e.g., the cortical surface). For example, the
`half-value area, A1/2, is defined as the area of the cortex where the electric field exceeds half
`of the maximum electric field strength. Thus, focality quantifies the electric field spread in
`the two dimensions tangential to the brain surface. However, focality metrics such as A1/2
`are sensitive to the chosen surface depth where the area is calculated. Therefore, we adopted
`a more robust quantification of the electric field tangential spread, defined as
`
`(3)
`
`where V1/2 is the half-value volume—the volume of the brain region that is exposed to an
`electric field as strong as or stronger than half of the maximum electric field.123,133 Figure 3
`illustrates the definitions of d1/2 and V1/2 for three different coils. Like A1/2, the spread
`metric S1/2 has units of area. The lower S1/2, the more focal the electric field is. Collectively,
`d1/2 and S1/2 characterize the 3-D spatial extent of the TMS electric field.
`
`The fraction of Emax relative to which the various depth and focality metrics are defined,
`1/ Y2
`,60,61,65,66,79,134 1/2,33,122 or 1–1/ e,135 is arbitrary and can be set to any
`typically
`value between 0 and 1 depending on the objectives of the evaluation. Furthermore, since the
`depth and focality metrics are normalized to the peak electric field, Emax, they describe
`purely the relative spatial characteristics of the electric field and depend only on the coil
`geometry and placement. The peak electric field strength, as well as the pulse shape and
`width affect the focality of neural activation with TMS, but these parameters can be
`controlled by the pulse generation circuit of the TMS device and are therefore excluded
`from this characterization of the intrinsic spatial features of the coil electric field.136
`
`Finally, the direction of the induced electric field in the brain is another important
`determinant of the effective stimulation strength and focality.137–139 However, most TMS
`coils can be flexibly positioned and oriented on the head; therefore, the field direction
`relative to the brain is not an intrinsic property of the coil, but rather depends on the specific
`coil positioning. Therefore, we did not characterize the field direction as part of the intrinsic
`coil focality properties.
`
`Figure 2 shows the induced electric field distribution on the brain surface. Symmetric
`circular type coils (#1–7, 15, 16, 19, and 20) induce a single loop of eddy current with axial
`symmetry around the coil central axis. The H coils (#8–14), asymmetric crown coils (#17,
`
`Brain Stimul. Author manuscript; available in PMC 2013 February 09.
`
`Results
`
`NIH-PA Author Manuscript
`
`NIH-PA Author Manuscript
`
`NIH-PA Author Manuscript
`
`LUMENIS EX1045
`Page 6
`
`

`

`Deng et al.
`
`Page 7
`
`18), circular coil array (#46), and reverse current figure-8 coil (#47) produce electric field
`distributions similar to the single loop pattern of the circular coils. Symmetric figure-8 type
`coils (#21–26, 28, 29, 31, 32, 34–40) induce two loops of eddy current with reflection
`symmetry with respect to the x-z and y-z planes. The asymmetric figure-8 coils (#27, 30),
`coils arrays #42–44, and active shield figure-8 coil (#33) produce electric field distributions
`similar to those of the symmetric figure-8 coils. The 3-D differential coil (#41), cloverleaf
`coil (#45), active shield figure-8 coil (#48), MRI z-gradient coil in opposing-current mode
`(#49), and MRI x- (or y-) gradient coil (#50) produce more complex electric field
`distributions with multiple eddy current loops. Finally, the flux ball coil (#0) produces a
`broad, circular, symmetric electric field distribution.
`
`Figure 4 shows the TMS electric field half-value spread S1/2 as a function of the half-value
`depthd1/2 for the 50 simulated TMS coils. The solid and dashed lines are curves of best fit of
`the points corresponding to the symmetric circular (#1–7, 15, 16, 19, and 20) and figure-8
`(#21–26, 28, 29, 31, 32, 34–40) type coils, respectively. Coils #34* and #40*—are the air-
`core counterparts (core relative permeability set to 1) to ferromagnetic-core coils #34 and
`40, respectively. The TMS coils exhibit a wide range of S1/2 and d1/2, but, as expected, are
`all subject to a depth–focality tradeoff. The range of d1/2 is similar between the simulated
`circular and figure-8 type coils, 1.0–3.5 cm and 0.9–3.4 cm, respectively. However, figure-8
`type coils are more focal compared to circular type coils, with S1/2 range of 5–261 cm2 and
`34–273 cm2, respectively. For either circualr or figure-8 type coils with d1/2 < 2 cm, S1/2 and
`d1/2 follow approximately a power law demonstrated by their near linear relationship on the
`logarithmic axes plot in Figure 4. Coils with deeper electric field penetration cannot be as
`focal as more superficial coils. Figure S1 in the Supplementary Material shows analogous
`plots of V1/2 and A1/2 versus d1/2, demonstrating consistent depth–focality tradeoff for these
`alternative focality metrics as well.
`
`The focality advantage of figure-8 over circular type coils diminishes as the d1/2 increases.
`The depth–focality tradeoff curves for both coil types converge to the flux ball (coil #0)
`characteristics which correspond to linear electric field decay from the circles of latitude to
`the sphere axis where the field is zero.3 Therefore, the maximum achievable d1/2 is half the
`brain radius, 3.5 cm in our model (see Figure 4). The maximum S1/2 is 308 cm2,
`representing the most unfocal coil configuration possible. The most focal circular type coil
`is the mini-coil (#1) designed for stimulation in non-human primates, with d1/2 and S1/2 of
`1.0 cm and 34 cm2, respectively. The most focal figure-8 type coil is the three-layered
`double coil (#21), with d1/2 and S1/2 of 0.9 cm and 5 cm2, respectively.
`
`Developed for deep brain TMS, the H coils (#8–14) have relatively nonfocal, near circular
`electric field distribution. The H coils have deeper electric field penetration (d1/2 = 1.7–2.4
`cm) compared to conventional circular coils (e.g., #4–7, 15; d1/2 = 1.4–1.9 cm). Larger
`diameter circular coils (e.g., #16–19, d1/2 = 2.5–2.7 cm), in turn, produce more deeply
`penetrating electric field than H coils, at the expense of reduced focality. Large double-cone
`type coils (#36, 39) provide deeper stimulation (d1/2 = 2.5–3.1 cm) than H coils with
`comparable or better focality.
`
`The insertion of a ferromagnetic core in a coil can alter the focality and depth of the electric
`field, but is not able to circumvent the depth–focality tradeoff limits observed for air core
`coils. The effects of inserting a cylindrical ferromagnetic core (#3) through the center of a
`50-mm circular coil (#2) are minor, reducing d1/2 and S1/2 by 3% and 5%, respectively.
`Placing a ferromagnetic core lateral to a circular coil can increase the field focality without
`sacrificing half-value depth. For example, adding a partial cylindrical core frontal (#10) or
`lateral (#11) to the H1 coil (#9) reduces S1/2 by 6% and 0.8% while increasing d1/2 by 5%
`and 1.4%, respectively. However, the resultant depth and focality are still inferior to the
`
`Brain Stimul. Author manuscript; available in PMC 2013 February 09.
`
`NIH-PA Author Manuscript
`
`NIH-PA Author Manuscript
`
`NIH-PA Author Manuscript
`
`LUMENIS EX1045
`Page 7
`
`

`

`Deng et al.
`
`Page 8
`
`performance of comparable figure-8 type coils. When a C-shaped ferromagnetic core (#34)
`is added to a figure-8 coil (#34*), both d1/2 and S1/2 are increased by 13% and 34%,
`respectively. On the other hand, adding a stretched C-core (#40) to a partial toroidal coil
`(#40*) reduces both d1/2 and S1/2 by 11% and 46%, respectively.
`
`Finally, Figure 5 shows the electric field S1/2 and d1/2 profile for the twin coil for opening
`angles ranging from 90° to 180°. The field characteristics for a circular coil identical to one
`of the twin coil windings are plotted for comparison. The spread S1/2 of the twin coil
`increases monotonically as the inter-loop opening angle widens. However, d1/2 increases for
`opening angles from 90° to 110°, and decreases for opening angles from 110° to 180°. The
`focality locus in Figure 5 is replicated with a dotted line in Figure 4 for comparison with the
`other coils.
`
`Discussion
`Electric field depth–focality tradeoff
`Among the TMS coil designs, there is a tradeoff between electric field depth of penetration
`and focality, as illustrated in Figure 4 and Figure S1 in the Supplementary Material. In
`general, coils with larger dimensions produce electric field with greater d1/2 and S1/2. In
`contrast, smaller coils produce electric field that is more localized and superficial. For
`conventional and smaller coil sizes (d1/2 < 2 cm), S1/2 and d1/2 are related approximately by
`a power law for either circular or figure-8 type coils. It is noteworthy that none of the coil
`designs was able to overcome the depth–focality tradeoff set by the figure-8 type coils—no
`TMS coil performance lies significantly to the right of the dashed curve in Figure 4. This
`finding supports the claim that the figure-8 coil focality can only be modified somewhat but
`not improved substantially.32 Hence, no TMS coil can achieve deep and focal stimulation
`simultaneously, consistent with previous findings.31,33
`
`In this work we used a fixed size spherical model with radius of 8.5 cm. Variation in head
`anatomy could affect the electric field depth of penetration and focality. We have previously
`investigated the sensitivity of the induced electric field to anatomical variability (variation in
`head diameter, scalp and skull thickness, brain volume, and tissue electrical properties) in
`the context of magnetic seizure therapy.140 The metrics d1/2 and S1/2 are most sensitive to
`the head diameter, but considered as percentage of their maximum values, which are head
`diameter dependent, they both scale similarly with head size. Furthermore, variability in
`head size is equivalent to variation in coil size for fixed head size, which is already
`illustrated in Figure 4. Therefore, variation in head anatomy would not change qualitatively
`the relative performance of the different coils.
`
`This study focused on a comparison of the geometric characteristics of the electric field
`induced by various TMS coils. It did not characterize the effect of coil current waveform
`and amplitude, coil placement relative to anatomical head landmarks, and neural activation
`thresholds. Clearly these aspects of stimulation have to be considered in a more complete
`analysis of the extent of neural activation by TMS. In previous modeling studies of TMS we
`have considered the coil current waveform and amplitude and the neural response
`threshold.123,140 Anatomically realistic TMS models120,121 can be particularly useful in
`studying the effect of the complex tissue structure of the head on the electric field for
`various coil placements.
`
`Strategies for controlling electric field focality
`Several strategies have been deployed for making the electric field more focal. As illustrated
`in Figure 4 and Figure S1, a figure-8 type coil is always more focal than a circular type coil
`with the same d1/2. The figure-8 type coil maintains its focality advantage over the circular
`
`Brain Stimul. Author manuscript; available in PMC 2013 February 09.
`
`NIH-PA Author Manuscript
`
`NIH-PA Author Manuscript
`
`NIH-PA Author Manuscript
`
`LUMENIS EX1045
`Page 8
`
`

`

`Deng et

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket