throbber
Am J Nucl Med Mol Imaging 2015;5(5):527-547
`www.ajnmmi.us /ISSN:2160-8407/ajnmmi0014068
`
`Review Article
`Dosage optimization in positron emission tomography:
`state-of-the-art methods and future prospects
`
`Nicolas A Karakatsanis1,2, Eleni Fokou3, Charalampos Tsoumpas1,3
`
`1Translational and Molecular Imaging Institute and Department of Radiology, Icahn School of Medicine at Mount
`Sinai, New York, USA; 2Division of Nuclear Medicine and Molecular Imaging, School of Medicine, University of
`Geneva, Geneva, Switzerland; 3Division of Biomedical Imaging, University of Leeds, Leeds, United Kingdom
`Received August 6, 2015; Accepted September 4, 2015; Epub October 12, 2015; Published October 15, 2015
`
`Abstract: Positron emission tomography (PET) is widely used nowadays for tumor staging and therapy response in
`the clinic. However, average PET radiation exposure has increased due to higher PET utilization. This study aims
`to review state-of-the-art PET tracer dosage optimization methods after accounting for the effects of human body
`attenuation and scan protocol parameters on the counting rate. In particular, the relationship between the noise
`equivalent count rate (NECR) and the dosage (NECR-dosage curve) for a range of clinical PET systems and body at-
`tenuation sizes will be systematically studied to prospectively estimate the minimum dosage required for sufficiently
`high NECR. The optimization criterion can be determined either as a function of the peak of the NECR-dosage curve
`or as a fixed NECR score when NECR uniformity across a patient population is important. In addition, the systematic
`NECR assessments within a controllable environment of realistic simulations and phantom experiments can lead
`to a NECR-dosage response model, capable of predicting the optimal dosage for every individual PET scan. Unlike
`conventional guidelines suggesting considerably large dosage levels for obese patients, NECR-based optimization
`recommends: i) moderate dosage to achieve 90% of peak NECR for obese patients, ii) considerable dosage reduc-
`tion for slimmer patients such that uniform NECR is attained across the patient population, and iii) prolongation of
`scans for PET/MR protocols, where longer PET acquisitions are affordable due to lengthy MR sequences, with mo-
`tion compensation becoming important then. Finally, the need for continuous adaptation of dosage optimization to
`emerging technologies will be discussed.
`
`Keywords: PET, PET/CT, PET/MR, dosage, NECR, optimization, Monte-Carlo
`
`Introduction
`
`Imaging is nowadays considered an essential
`part of modern nuclear medicine procedures
`as it provides the means to obtain an abun-
`dance of multi-dimensional information for the
`diagnosis, prognosis and therapy response
`monitoring of a variety of diseases in the clinic
`[1]. In particular, positron emission tomography
`(PET) is an established imaging modality for the
`in-vivo inspection of a wide spectrum of diseas-
`es in oncology [2-4], cardiology [5] and neurol-
`ogy [6]. PET has gained considerable attention
`in the medical imaging community in the last
`few decades, over other nuclear imaging tech-
`niques, as most of the chemically relevant to
`human body radioisotopes are positron emit-
`ters (11C, 13N, 15O, 18F) and thus can more effi-
`ciently label pharmaceutical substances of
`
`physiological interest. Meanwhile, the signifi-
`cant evolution of PET scanner technology over
`the last decades enabled, for the first time, 3D
`multi-functional imaging of numerous molecu-
`lar and cellular biochemical processes [4].
`
`Despite the advantages of PET as a metabolic
`in-vivo imaging modality, acquisition of accu-
`rate and high-resolution anatomical informa-
`tion is currently not possible by PET alone due
`to underlying physical limitations, even when
`prolonging scan time through transmission
`acquisitions or when state-of-the-art time-of-
`flight (TOF) technology is employed. This has led
`to the introduction of hybrid scanners, such as
`PET/Computed Tomography (PET/CT) and PET-
`Magnetic Resonance (PET/MR) systems where
`the CT or MR modalities can be utilized to
`obtain the missing anatomical information.
`
`Petitioner GE Healthcare – Ex. 1037, p. 527
`
`

`

`Dosage optimization strategies in PET
`
`Additionally, the development of fully integrated
`multi-modality PET systems nowadays enables
`acquisition of anatomical maps that are per-
`fectly co-registered with PET images, thus
`allowing for the precise localization and attenu-
`ation correction of the PET signal, which is very
`important for quantitative imaging tasks and
`image-guided therapy [2].
`
`However, radiation exposure can be a serious
`concern for adult and particularly children
`patients, especially in the case of PET/CT hybrid
`systems, due to the ionizing nature of both PET
`and CT radiation, with the latter contributing to
`relatively higher absorbed doses than the for-
`mer modality. Previously, studies assessed the
`risk of radiation dose absorbed by the human
`tissue during whole-body PET and CT scans and
`proposed efficient PET dosage regimens to
`reduce the total exposure as much as possible,
`although exposure from CT remains significant
`[7, 8]. Fortunately with the advent of hybrid
`PET/MR imaging, the total radiation absorbed
`dose in clinic can be now significantly reduced
`thanks to the zero contribution of ionizing radia-
`tion by MR modality. Therefore, as PET/MR
`imaging becomes widely applicable, the main
`effort in further reducing absorbed dose to
`patients and personnel can be concentrated
`on the injected dosage minimization without
`degrading image quality (IQ). The decay rate, or
`activity, of a positron-emitting radiopharma-
`ceutical is a random process following the well-
`characterized Poisson distribution [9]. Thus,
`the acquisition of the PET data is a statistical
`imaging process characterized by a level of
`inherent uncertainty, which can be quantified
`with the statistical index of standard deviation
`[10]. According to the principles of Poisson dis-
`tribution, the relative percentage uncertainty in
`the detected coincidence rate is reduced with
`higher number of counts and thus, from a sta-
`tistical point of view, it is highly desirable to be
`able to achieve the maximum possible counts
`for a given amount of scan time or equivalently
`the highest count rate.
`
`However, the type of counts recorded during a
`PET acquisition can be classified as either true,
`scattered or random coincidence events, as it
`will be explained later, of which only the trues
`accurately reflect the actual underlying radio-
`nuclide activity distribution, while scattered
`and random events contaminate the total sig-
`
`nal with misleading information. Thus, a high
`number of overall detected coincidence events,
`also known as prompts, do not necessarily
`translate to higher statistical quality and signal
`to noise ratio (SNR) for the acquired PET raw
`data and respective reconstructed images.
`Instead, the index of noise-equivalent counts
`could be employed to quantify the net positive
`contribution or SNR of the prompts. Therefore,
`retaining a sufficiently high noise-equivalent
`count rate (NECR), as opposed to prompt
`counts rate, could be a more reliable strategy
`to ensure a high SNR for the measured PET
`data and the respective images for a given
`scan period.
`
`PET systems detect coincidence events at a
`particular rate for a given underlying activity
`distribution in the FOV [9, 11]. The measured
`total coincidences count rate per unit of total
`activity distribution in the FOV characterizes
`the sensitivity of a PET scanner [10]. Higher
`scanner sensitivity can indeed be one of the
`factors ensuring high NECR and SNR perfor-
`mance, but it is usually fixed for each PET sys-
`tem, thus leaving no margins of data quality
`optimization. However, another parameter that
`can drastically affect NECR for a given scanner
`and could be used for its optimization is the
`amount of administered radiopharmaceutical
`dosage [12-14]. Therefore, NECR and conse-
`quently SNR, can be significantly regulated for
`a given scanner or sensitivity, either to a maxi-
`mum or a sufficiently high value by administer-
`ing a proper amount of radiopharmaceutical
`dosage.
`
`Furthermore, the relationship between the
`administered dosage and the “statistically use-
`ful” count rate, i.e. the NECR, of a PET scanner
`may depend on a range of parameters which
`can be either scanner-related such as the geo-
`metric efficiency, detection efficiency, energy
`window and dead-time effect [11, 13] or
`patient-related, such as the degree of radiation
`attenuation caused by the body and the rela-
`tive position between the targeted source dis-
`tribution and the current bed field-of-view (FOV)
`of the scanner [13]. By systematically and
`quantitatively analyzing the effects of these
`factors on NECR for a range of dosage levels,
`an accurate NECR-dosage response model can
`be designed allowing for the prediction of the
`minimum possible amount of dosage required
`
`528
`
`Am J Nucl Med Mol Imaging 2015;5(5):527-547
`
`Petitioner GE Healthcare – Ex. 1037, p. 528
`
`

`

`Dosage optimization strategies in PET
`
`to sufficiently maintain NECR, or statistically
`useful counts, at a quantitatively acceptable
`level. In the past, many studies [10, 14] system-
`atically examined the effect of these factors on
`different types of PET coincidence count rates
`(true, random, scatter) and the NECR.
`
`As a result, a possible strategy for the optimiza-
`tion of the administered dosage prior to every
`individual PET scan could consist of two essen-
`tial steps: i) Initially, a systematic assessment
`of the NECR response as a function of the
`administered dosage needs to be conducted
`for a family of clinical PET scanners and a vari-
`ety of patient- and scanner-related study
`parameters, as described above. This will
`enable the design of a parameterized model
`that can adequately describe the NECR-dosage
`relationship for a wide set of acquisition sce-
`narios often present in clinical routine. Then, ii)
`such model could be utilized to predict, for a
`given set of patient attenuation properties and
`scanner settings, an optimal range of dosage-
`levels for which a nearly maximal NECR is
`expected by the model. The terms “range of
`dosage” and “nearly maximal NECR” imply that
`the predicted outcome will initially be a range of
`dosages as opposed to a single value while
`the final dosage recommendation may corre-
`spond to a NECR value close to the peak,
`expressed as a user-defined percentage (%) of
`the latter, e.g. at least 90% of the peak NECR
`for the given set of parameters.
`
`However, the previous dosage regimen scheme
`may result in highly variable NECR scores
`among patient scans, as the criterion is a per-
`centage of the peak which may vary significant-
`ly across the scan populations, e.g. between
`slim and obese patients. Alternatively, a dose
`optimization strategy that would ensure NECR
`equivalency over a population of patient scans
`could utilize the same NECR-dosage response
`model to predict the amount of dosage that
`would induce a fixed NECR score for all scans,
`regardless of the patient and scanner factors
`involved. In that case, the optimal dosage
`would obviously not correspond to a maximal
`NECR for all cases, as the previous scheme, but
`would ensure a sufficient and uniform NECR
`performance across the population. The losses
`in NEC counts could then be compensated by
`accordingly elongating the PET scan duration.
`This could be applicable i) to some extent, to
`
`longer PET/CT protocols, but ii) more efficiently
`to simultaneous PET-MR protocols, where the
`lengthy MR acquisitions permit equally long
`PET acquisitions.
`
`Therefore, the aim of this review is to present a
`collection of studies demonstrating the feasibil-
`ity of NECR-based dosage optimization in indi-
`vidual PET studies, after systematically evaluat-
`ing the dosage effect on NECR for a representa-
`tive set of patient- and scanner- parameters.
`Thus, individualized prediction of optimal dos-
`age may become possible prior to every PET
`scan, paving the way for personalized multimo-
`dality PET clinical imaging protocols across a
`wide range of patients and scanner configu-
`rations.
`
`Critical analysis
`
`Coincidence types and their contribution to the
`actual PET information
`
`PET utilizes trace amounts of positron emitting
`radiopharmaceuticals (radio-tracers), such as
`18F-fluorodeoxyglucose (FDG) [2, 4, 15]. Each
`emitted positron travels a short distance
`(range) before it annihilates with an electron
`causing the emission of two gamma photons,
`each of 511 keV energy, at nearly opposite
`directions [16, 17]. Radiotracers emitting posi-
`trons of relatively low kinetic energy are pre-
`ferred to ensure minimum energy (dose) depo-
`sition at the point of emission and small trav-
`elled distance of the positron (positron range)
`before its annihilation. Subsequently, the two
`emitted gamma photons are detected in coinci-
`dence by two opposing PET detectors. The line
`connecting that pair of detectors is known as
`the line of response (LOR) for that coincident
`event and defines the geometrical space where
`the annihilation occurred [16, 17]. This type of
`coincidence event, called ‘true coincidence’, is
`the desired type of count in PET imaging, as it
`defines an LOR that includes the actual point of
`annihilation and thus positively contributes to
`the true or net PET signal [18]. However, there
`are also two other possible types of detection
`events, which are contaminating the overall
`PET signal with erroneous information: the
`scattered and random coincidences.
`
`Scattered coincidence events, simply denoted
`also as “scatter”, occur when at least one of the
`two emitted gamma photons undergoes
`
`529
`
`Am J Nucl Med Mol Imaging 2015;5(5):527-547
`
`Petitioner GE Healthcare – Ex. 1037, p. 529
`
`

`

`Dosage optimization strategies in PET
`
`Figure 1. Realistic digital framework of multiple simulated FDG-PET studies.
`The presented framework enables comparisons of realistic SUV values from
`reconstructed simulated PET images against the ground truth. Published with
`permission of the author and the Publisher. Originally published in Silva-Ro-
`driguez J, Aguiar P, Dominguez-Prado I, Fierro P, Ruibal A. Simulated FDG-PET
`studies for the assessment of SUV quantification methods. Rev Esp Med Nucl
`Imagen Mol 2015; 34(1): 13-8. Copyright © 2014 Elsevier España, S.L.U. and
`SEMNIM. All rights reserved [22].
`
`result, randoms are more
`likely observed at high-count
`rates where
`interference
`between different but near-
`ly simultaneous annihilations
`is more probable [20, 21].
`Therefore, random coinciden-
`ce detection rates depend on
`the rate of individual gamma
`photon detections, known as
`singles rate, which in turn,
`depends on
`the
`rate of
`gamma photon emissions
`(activity) within the field-of-
`view (FOV) [16]. In fact ran-
`doms rate is proportional to
`the square of singles rate,
`and
`thus
`it can become
`considerable at high activity
`concentration levels. Further-
`more, randoms are expected
`to increase with the coinci-
`dence time window (CTW)
`width of the scanner, which
`determines
`the maximum
`allowed difference in the esti-
`mated detection time bet-
`ween two singles event to be considered as a
`coincidence event [16]. The slower the detec-
`tion system is responding to a given rate of inci-
`dent gamma photons, the larger the uncertain-
`ty in estimating the exact time of detection and,
`thus the larger the CTW width has to be to mini-
`mize rejection of true coincidences. Con-
`sequently, for a given activity level, wider CTWs
`are expected to yield higher fraction of random
`events [16, 17, 19]. As a result, both scatter
`and random events reduce image quality (IQ) by
`degrading spatial resolution and contrast due
`to inaccuracies caused by LOR mispositioning
`[18].
`
`Standardized uptake value (SUV) metric and
`its limitation for image-based PET data quality
`assessments
`
`One important feature of PET imaging is its abil-
`ity to acquire quantitative estimates of in-
`vivo concentration of physiologically relevant
`radiopharmaceuticals which can later be used
`for a wide range of quantitative imaging tasks,
`e.g. for tumor characterization, differentiation
`and response to therapy in oncology [22]. The
`activity concentration measurements at every
`image pixel can be normalized with respect
`to the total administered PET tracer dosage
`
`Am J Nucl Med Mol Imaging 2015;5(5):527-547
`
`Compton scattering when interacting with the
`patient’s body or a detector crystal before its
`actual detection, causing it to change direction
`and thus divert from the true LOR while losing
`part of its kinetic energy deposited at detection
`point [16]. Scatter events may still be detect-
`able as coincidences, provided the energy of
`each scattered photon deposited in PET detec-
`tors still lies within the energy window account-
`ed by the PET system. However, the trajectories
`of two photons, of which at least one has under-
`gone Compton scattering, are no longer aligned
`with respect to each other, thus violating the
`basic assumption of 180-degree angle between
`them and resulting in mispositioned LORs [16,
`19]. The fraction of scatter to the total number
`of detected events, known as scatter fraction,
`depends on the attenuation properties of the
`imaging subject and detectors as well as the
`energy window of the system, the latter deter-
`mined by the finite energy resolution of the
`detection system.
`
`Similarly, random coincidence events, also
`known simply as “randoms”, may also trigger
`LORs mispositioning but through a different
`mechanism: the nearly simultaneous but acci-
`dental (random) detection of two gamma pho-
`tons produced by different annihilations. As a
`
`530
`
`Petitioner GE Healthcare – Ex. 1037, p. 530
`
`

`

`Dosage optimization strategies in PET
`
`and the lean body mass or weight of the patient
`to obtain the surrogate metric of standardized
`uptake value (SUV), a well-established index in
`clinical nuclear medicine [22]. However, SUV
`only reflects the normalized radioactivity con-
`centration present in the tissue of interest at a
`certain time. As a result, it should be pointed
`out that SUV may not be considered fully quan-
`titative, as it is inherently time-dependent and
`can be also influenced by various patient- and
`scanner-related parameter [23, 24].
`
`Recently, Silva-Rodriguez et al. (2015) present-
`ed a PET simulation framework (Figure 1) con-
`sisting of multiple FDG-PET realistic digital
`phantoms to enable systematic comparative
`SUV evaluations at various regions of interest
`(ROIs) between true input and reconstructed
`simulated PET images for three different SUV
`calculation methods [22]. Although the study
`showed the usefulness of SUV metric in evalu-
`ating clinically relevant tumor characteristics
`from PET images, it also demonstrated its high
`sensitivity to the partial volume effect and sta-
`tistical noise which can considerably affect
`detectability and quantification of
`tumor
`regions in the reconstructed PET images [22].
`
`The ALARA principle
`
`The general human population is daily exposed
`to low levels of ionizing radiation, relative to the
`safety limits, originated from the universe and
`long-lived radioisotopes in certain earth loca-
`tions. However, radiation exposure of the gen-
`eral population due to medical procedures has
`recently increased on average due to the more
`extensive usage of radiation
`in medicine,
`including nuclear medicine exams, mainly for
`the purpose of i) diagnosing diseases at earlier
`stages and assessing their progress, ii) evalu-
`ating treatment response and iii) designing
`optimal image-guided therapeutic schemes. In
`particular, absorbed radiation dose itself is an
`important parameter to study, because it may
`lead to deterministic or stochastic damaging
`radiobiological effects in living tissues if the
`safety limits are exceeded [25]. In particular,
`PET is associated with a relatively low radiation
`exposure to patients due to internal administra-
`tion of beta (positron) emitting radiopharma-
`ceuticals. The injected dosage is then distrib-
`uted over time across the human tissue while
`emitting gamma ionizing radiation due to the
`
`positron annihilation effect [26]. The amount of
`energy or dose deposited as a result of the
`interaction of beta and gamma particles with
`human tissue is proportionally related to the
`amount of administered dosage. The radiobio-
`logical effects of beta and gamma radiation
`may potentially be harmful at any age, if a very
`large dosage is administered, although typical
`PET dosage levels are within safety limits.
`However, dosage regularization may be particu-
`larly important for children and the fetus, which
`are considered highly radiosensitive [8, 27,
`28-31], primarily because their cells are divid-
`ing faster and, secondly, because they have
`longer life-time ahead to potentially develop
`long-term stochastic effects, such as carcino-
`genesis [8]. For instance, young children (5
`years old) exposed to radiation from an atomic
`bomb have 1.7-2.1 greater risk of developing
`cancer compared to young adults (20 years old)
`[32]. In addition, unlike dosage-radiobiological
`risk
`relationship described above, NECR-
`dosage curve is not linear, but instead exhibits
`a peak at moderate dosage levels, as it will be
`demonstrated later. Thus, very high dosages
`not only increase exposure, if applied repeat-
`edly, e.g. in longitudinal studies, but also are
`undesirable as they may in fact diminish statis-
`tical quality of the acquired PET data. On the
`other hand, extremely low dosage levels may as
`well impact quality, except if proportionally lon-
`ger PET acquisitions can be afforded, e.g. in the
`case of PET/MR acquisitions, to compensate
`for the low NECR [33]. Therefore, radiation
`exposure should be restricted to levels as low
`as reasonably achievable (ALARA principle) or
`avoided if not absolutely necessary for special
`radiosensitive population members such as
`children and pregnant women [15, 31, 34, 35].
`
`Thus, the ALARA concept encourages adminis-
`tration of reasonably low or moderate amounts
`of administered PET dosage after accounting
`for the available scan time and any other factor
`that could help further reduce exposure with-
`out affecting data quality. It essentially demon-
`strates both the need for dosage minimization
`as well as the importance of retaining sufficient
`or reasonable statistical quality. Therefore, any
`dosage
`regularization strategy, as
`those
`described in this review article, should follow
`the ALARA principles. Table 1 presents a
`collection of human population average
`exposure levels due to characteristic nuclear
`
`531
`
`Am J Nucl Med Mol Imaging 2015;5(5):527-547
`
`Petitioner GE Healthcare – Ex. 1037, p. 531
`
`

`

`Dosage optimization strategies in PET
`
`Table 1. Radiation exposure for characteristic medical exams and cosmic radiation levels
`Radiation exposure (mSv)
`Medical Imaging sources
`3-4
`FDG (185 MBq) PET scan
`7.0
`CT-chest
`25
`PET/CT (non-modified protocol)
`0.1
`Chest X-ray
`7.6
`PET scan (15 yrs old)-modified protocol
`5.9
`CT scan (15 yrs old)-modified protocol
`13.5
`PET/CT scan (15 yrs old)-modified protocol
`Exposure from PET and CT vary upon scan purpose and local protocols. Data collected from literature [2, 32, 36-38].
`
`Daily activities and occupations (Cosmic radiation)
`Flight: Montreal to London
`0.0478
`Dentistry
`0.06 (annual)
`Living at sea level
`0.24
`Average dose/person (UK)
`2.5 (annual)
`Aircrew
`4.5 (annual)
`Metal mines
`2.7 (annual)
`
`randoms are contaminating
`rements.
`
`the measu-
`
`Thus, an alternative acquisition scheme result-
`ing in similar total number of counts but
`acquired from a smaller amount of dosage and
`for longer time period may be more appropriate
`from a statistical and dosimetric point of view
`as it will be associated with a larger trues frac-
`tion and less radiation exposure for the patient
`and the personnel. On the other hand, pro-
`longed scan periods may not be clinically feasi-
`ble, given the high patient throughput require-
`ments in the PET clinic.
`
`NECR
`
`=
`
`true
`
`2
`
`true
`
`scatter
`
`random
`
`A suitable metric to quantify the overall statisti-
`cal usefulness of the total count rates by taking
`into account their synthesis is NECR [3, 10] and
`is defined as follows [9]:
`R
`+
`+
`aR
`R
`bR
`where Rtrue, Rscatter and Rrandom denote the true,
`scatter and random coincidences count rates,
`respectively. Also ‘α’ represents the occupied
`area in the FOV by the patient while ‘b’ depends
`on the method used to estimate random events
`[33]. As its name implies, NECR can be consid-
`ered as the equivalent trues count rate that
`would yield to the same level of statistical noise
`as the one currently present in the measured
`prompt counts [39].
`
`medicine exams (left column) and cosmic radi-
`ation (right column), as compiled from pub-
`lished reports [2, 32, 36-38]. Although the
`occupational dose limit for an adult is only 50
`mSv, the ALARA principles should always be fol-
`lowed in medical procedures for the maximum
`reasonable limitation of patient and personnel
`exposure.
`
`The importance of thenoise-equivalent count-
`ing rate (NECR) metric
`
`An important parameter when evaluating the
`performance of a PET system is the count rate
`as a function of the total activity present in the
`FOV because it can considerably affect the sta-
`tistical quality and SNR of projection data and
`IQ of reconstructed images for a given imaging
`protocol [3]. The inherent uncertainty in coinci-
`dence rate measurements is expressed with
`the presence of noise in the projection and
`image space. The Poisson properties of the
`counts distribution in PET measurements sug-
`gest lower percentage noise and enhanced
`SNR when higher numbers of true coincidence
`counts are acquired. Thus, studies have sug-
`gested increasing the amount of injected dos-
`age as an attempt to limit the presence of noise
`in the acquired PET data [3]. However, this
`strategy may have negative implications for
`both the data quality in the projection and
`image data as well as the patient. In particular,
`the administration of very large amounts of
`dosage is expected to i) considerably rise the
`number of randoms and the associated arti-
`facts, ii) trigger counting losses due to dead-
`time effect and iii) increase radiation exposure
`[13]. In other words, not only the total number
`of counts (prompts) but also their synthesis in
`trues, scatter and randoms is also important
`from a statistical point of view, as scatter and
`
`Random coincidences can be estimated by
`employing a delayed coincidence window (δ +
`ΔT) which is time-shifted relative to the stan-
`dard coincidence window (ΔT) [9]. The use of
`delayed window allows for the exclusive esti-
`mation of random events spatial distribution,
`thus providing an automated and practical
`method for online randoms subtraction from
`
`532
`
`Am J Nucl Med Mol Imaging 2015;5(5):527-547
`
`Petitioner GE Healthcare – Ex. 1037, p. 532
`
`

`

`Dosage optimization strategies in PET
`
`Figure 2. Random (sub-figure A), true (sub-figure B) and scatter (sub-figure C) coincidences counts as a function of
`injected dosage activity for six patients (A, B,…, F) of different weights (legend of sub-figure A). Obtained with permis-
`sion from Monte Carlo simulation of PET images for injection dose optimization/Jiri Boldys, Jiri Dvorak, Magdalena
`Skopalova and Otakar Belohlavek/ International Journal for Numerical Methods in Biomedical Engineering 2013;
`29: 988-999. Copyright © 2012 John Wiley & Sons, Ltd [45].
`
`the prompts distribution, the latter estimated
`from the original window [9]. However, random
`online subtraction can significantly increase
`the statistical noise in the data. On the other
`hand, estimation of scatter coincidences is
`possible by applying a scatter simulation algo-
`rithm on registered anatomical images, when
`the activity distribution is entirely within the
`scanner FOV [9]. The estimated scatter count
`profile is then subtracted from the prompts pro-
`jection profiles. Overall, the true coincidences
`rate can be estimated by Rtrue = Rprompt - Rscatter -
`Rrandom [10].
`NECR has been evaluated by many studies [10,
`14, 40] as a surrogate metric of statistical qual-
`ity of acquired projection data. The objective
`quantitative assessment of the statistical qual-
`ity of acquired PET raw data offered by the
`NECR metric could help towards a more effi-
`cient optimization of the administered dosage
`which,
`in
`turn, determines
`the
`trade-off
`between competing scanner and patient fac-
`tors. By modeling the relation between the indi-
`vidual components of NECR response (trues,
`scatter and randoms) and various patient- and
`scanner-related factors of PET studies, it is
`possible to predict the minimum administered
`dosage required to acquire sufficient noise
`equivalent counts within a clinically acceptable
`scan period [13].
`
`The detected coincidence counts at every line
`of response (LOR) bin in the raw data follow the
`Poisson distribution. Thus, the standard devia-
`tion (δ) of each estimated count rate is expect-
`ed to be equal to the square root of the mea-
`sured mean count rates N, i.e.
`, suppos-
`=v
`N
`ing the same activity distribution was measured
`
`multiple times. Then, the ratio of the standard
`deviation of counts to their mean, known as %
`statistical noise,
`can be derived by
`^
`h
`, where N represents the mean
`1 N
`100
`#
`counts in each LOR [41]. Moreover, NECR per
`pixel denotes the metric of NECR density
`(NECRD) [42] and is obtained by dividing NECR
`by the number of image pixels defining the
`attenuating volume of the scanned patient. On
`the other hand, the SNR of the projection data
`is defined as the ratio of the mean pixel value to
`
`the standard deviation, i.e. / NN
` and it
`=
`N
`is equivalent to the square root of NEC counts
`data equiv. to
`for a fixed acquisition time Δt (SNR2
`NEC = NECR × Δt) [10, 11, 43]. A recent study
`investigated the relationship between NEC and
`SNR2 in an experimental phantom set-up as
`well as in clinical PET scans and found satisfac-
`tory correlation between the two metrics in all
`phantom cases while, in the clinical cases, high
`correlation was observed for low and moderate
`body mass index (BMI) indices (< 28 kg/m2).
`The authors note that further studies are
`required to determine this relationship for
`patients of higher BMI indices [44].
`
`There are strong indications that NECR can be
`a suitable criterion for determining the optimal
`amount of administered dosage in a PET clini-
`cal study. Nevertheless, as explained above,
`NECR refers directly to the statistical quality
`(SNR2) of the projection data (sinogram) mea-
`surements only. Although the average quality of
`the reconstructed
`images
`is expected to
`increase with NECR for the same reconstruc-
`tion algorithm, image regions of different level
`of counts may be affected differently. Moreover,
`each iteration of various reconstruction algo-
`rithms may exercise a different impact in the
`
`533
`
`Am J Nucl Med Mol Imaging 2015;5(5):527-547
`
`Petitioner GE Healthcare – Ex. 1037, p. 533
`
`

`

`Dosage optimization strategies in PET
`
`final quality of the images for the same set of
`projection data. Thus, despite the usefulness
`of the NECR in quantifying the statistical quality
`of PET projection data, its value as a metric of
`quality of the images is limited and depends on
`a large set of reconstruction parameters. Thus,
`NECR may not be a 100% reliable indicato
`when the target is quality of a specific image
`ROI, as the choice of a reconstruction algorithm
`may considerably affect IQ to an extent that
`cannot be predicted by NECR alone [14].
`
`Noise-equivalent counting rate and injected
`dosage: The administration of amounts of activ-
`ity below or above the recommended optimal
`dosage levels can be problematic for various
`reasons. Figure 2 presents the effect of within
`FOV activity distributions on true, scatter and
`random coincidence events. Very large amounts
`of administered dosage may degrade SNR in
`projection data and quality of images due to
`the relatively higher increase rate of random
`(Figure 2A) compared to true (Figure 2B) and
`scatter (Figure 2C) coincidences [45, 46].
`
`This is attributed to the exponential relation-
`ship between randoms rate and activity, unlike
`true and scatter coincidence rates that are
`exhibiting an approximately linear dependence
`with activity. As a result, NECR will increase
`with dosage at low injected levels but then
`is expected to level-off or even drop as the dos-
`age becomes moderate o

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket