throbber
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
`Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
`
`Designation: D2240 − 15
`
`Standard Test Method for
`Rubber Property—Durometer Hardness1
`
`This standard is issued under the fixed designation D2240; the number immediately following the designation indicates the year of
`original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
`superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
`
`This standard has been approved for use by agencies of the U.S. Department of Defense.
`
`1. Scope
`1.1 This test method covers twelve types of rubber hardness
`measurement devices known as durometers: Types A, B, C, D,
`DO, E, M, O, OO, OOO, OOO-S, and R. The procedure for
`determining indentation hardness of substances classified as
`thermoplastic elastomers, vulcanized (thermoset) rubber, elas-
`tomeric materials, cellular materials, gel-like materials, and
`some plastics is also described.
`1.2 This test method is not equivalent to other indentation
`hardness methods and instrument
`types, specifically those
`described in Test Method D1415.
`1.3 This test method is not applicable to the testing of
`coated fabrics.
`1.4 All materials, instruments, or equipment used for the
`determination of mass, force, or dimension shall have trace-
`ability to the National Institute for Standards and Technology,
`or other internationally recognized organizations parallel in
`nature.
`1.5 The values stated in SI units are to be regarded as
`standard. The values given in parentheses are for information
`only. Many of the stated dimensions in SI are direct conver-
`sions from the U. S. Customary System to accommodate the
`instrumentation, practices, and procedures that existed prior to
`the Metric Conversion Act of 1975.
`1.6 This standard does not purport to address all of the
`safety concerns,
`if any, associated with its use. It
`is the
`responsibility of the user of this standard to establish appro-
`priate safety and health practices and determine the applica-
`bility of regulatory limitations prior to use.
`
`2. Referenced Documents
`2.1 ASTM Standards:2
`
`1 This test method is under the jurisdiction of ASTM Committee D11 on Rubber
`and is the direct responsibility of Subcommittee D11.10 on Physical Testing.
`Current edition approved Aug. 1, 2015. Published January 2016. Originally
`approved in 1964. Last previous edition approved in 2010 as D2240 – 05 (2010).
`DOI: 10.1520/D2240-15.
`2 For referenced ASTM standards, visit the ASTM website, www.astm.org, or
`contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
`Standards volume information, refer to the standard’s Document Summary page on
`the ASTM website.
`
`D374 Test Methods for Thickness of Solid Electrical Insu-
`lation (Withdrawn 2013)3
`D618 Practice for Conditioning Plastics for Testing
`D785 Test Method for Rockwell Hardness of Plastics and
`Electrical Insulating Materials
`D1349 Practice for Rubber—Standard Conditions for Test-
`ing
`D1415 Test Method for Rubber Property—International
`Hardness
`D4483 Practice for Evaluating Precision for Test Method
`Standards in the Rubber and Carbon Black Manufacturing
`Industries
`F1957 Test Method for Composite Foam Hardness-
`Durometer Hardness
`2.2 ISO Standard:4
`ISO/IEC 17025: 1999 General Requirements for the Com-
`petence of Testing and Calibration Laboratories
`
`3. Summary of Test Method
`3.1 This test method permits hardness measurements based
`on either initial indentation or indentation after a specified
`period of time, or both. Durometers with maximum reading
`indicators used to determine maximum hardness values of a
`material may yield lower hardness when the maximum indi-
`cator is used.
`3.2 The procedures for Type M, or micro hardness
`durometers, accommodate specimens that are, by their dimen-
`sions or configuration, ordinarily unable to have their durom-
`eter hardness determined by the other durometer types de-
`scribed. Type M durometers are intended for the testing of
`specimens having a thickness or cross-sectional diameter of
`1.25 mm (0.050 in.) or greater, although specimens of lesser
`dimensions may be successfully accommodated under the
`conditions specified in Section 6, and have a Type M durometer
`hardness range between 20 and 90. Those specimens which
`have a durometer hardness range other than specified shall use
`another suitable procedure for determining durometer hard-
`ness.
`
`3 The last approved version of
`www.astm.org.
`4 Available from International Organization for Standardization (ISO), 1 rue de
`Varembé, Case postale 56, CH-1211, Geneva 20, Switzerland.
`
`this historical standard is referenced on
`
`Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
`
`1
`
`
`
`Copyright by ASTM Int'l (all rights reserved); Tue May 9 13:08:59 EDT 2017
`Downloaded/printed by
`Bill Bouchard (Plan Tech Inc) pursuant to License Agreement. No further reproductions authorized.
`
`ClearCorrect Exhibit 1059, Page 1 of 13
`
`

`

`D2240 − 15
`
`FIG. 1 (a) Type A and C Indentor
`
`4. Significance and Use
`4.1 This test method is based on the penetration of a specific
`type of indentor when forced into the material under specified
`conditions. The indentation hardness is inversely related to the
`penetration and is dependent on the elastic modulus and
`viscoelastic behavior of the material. The geometry of the
`indentor and the applied force influence the measurements
`such that no simple relationship exists between the measure-
`ments obtained with one type of durometer and those obtained
`with another type of durometer or other instruments used for
`measuring hardness. This test method is an empirical test
`intended primarily for control purposes. No simple relationship
`exists between indentation hardness determined by this test
`method and any fundamental property of the material tested.
`For specification purposes, it is recommended that Test Method
`D785 be used for materials other than those described in 1.1.
`
`5. Apparatus
`5.1 Hardness Measuring Apparatus, or Durometer, and an
`Operating Stand, Type 1, Type 2, or Type 3 (see 5.1.2)
`consisting of the following components:
`5.1.1 Durometer:
`5.1.1.1 Presser Foot, the configuration and the total area of
`a durometer presser foot may produce varying results when
`there are significant differences between them. It is recom-
`mended that when comparing durometer hardness determina-
`tions of the same type (see 4.1), that the comparisons be
`between durometers of similar presser foot configurations and
`total area, and that the presser foot configuration and size be
`noted in the Hardness Measurement Report (see 10.2.4 and
`5.1.1.3).
`5.1.1.2 Presser Foot, Types A, B, C, D, DO, E, O, OO,
`OOO, and OOO-S, with an orifice (to allow for the protrusion
`of the indentor) having a diameter as specified in Fig. 1 (a, b,
`c, d, e, f, and g), with the center a minimum of 6.0 mm (0.24
`in.) from any edge of the foot. When the presser foot is not of
`a flat circular design, the area shall not be less than 500
`mm2 (19.7 in.2).
`
`NOTE 1—The Type OOO and the Type OOO-S, designated herein,
`
`differ in their indentor configuration, spring force, and the results
`obtained. See Table 1 and Fig. 1 (e and g).
`5.1.1.3 Presser Foot—flat circular designs designated as
`Type xR, where x is the standard durometer designation and R
`indicates the flat circular press foot described herein, for
`example, Type aR, dR, and the like. The presser foot, having a
`centrally located orifice (to allow for the protrusion of the
`indentor) of a diameter as specified in Fig. 1 (a through g). The
`flat circular presser foot shall be 18 6 0.5 mm (0.71 6 0.02 in.)
`in diameter. These durometer types shall be used in an
`operating stand (see 5.1.2).
`(a) Durometers having a presser foot configuration other
`than that
`indicated in 5.1.1.3 shall not use the Type xR
`designation, and it is recommended that their presser foot
`configuration and size be stated in the Hardness Measurement
`Report (see 10.2.4).
`5.1.1.4 Presser Foot, Type M, with a centrally located
`orifice (to allow for the protrusion of the indentor), having a
`diameter as specified in Fig. 1 (d), with the center a minimum
`of 1.60 mm (0.063 in.) from any edge of the flat circular
`presser foot. The Type M durometer shall be used in a Type 3
`operating stand (see 5.1.2.4).
`5.1.1.5 Indentor, formed from steel rod and hardened to 500
`HV10 and shaped in accordance with Fig. 1 (a, b, c, d, e, or g),
`polished over the contact area so that no flaws are visible under
`20× magnification, with an indentor extension of 2.50 6 0.04
`mm (0.098 6 0.002 in.).
`5.1.1.6 Indentor, Type OOO-S, formed from steel rod and
`hardened to 500 HV10, shaped in accordance with Fig. 1 (f),
`polished over the contact area so that no flaws are visible under
`20× magnification, with an indentor extension of 5.00 6 0.04
`mm (0.198 6 0.002 in.).
`5.1.1.7 Indentor, Type M, formed from steel rod and hard-
`ened to 500 HV10 and shaped in accordance with Fig. 1 (d),
`polished over the contact area so that no flaws are visible under
`50× magnification, with an indentor extension of 1.25 6 0.02
`mm (0.049 6 0.001 in.).
`5.1.1.8 Indentor Extension Indicator, analog or digital
`electronic, having a display that is an inverse function of the
`indentor extension so that:
`
`2
`
`
`
`Copyright by ASTM Int'l (all rights reserved); Tue May 9 13:08:59 EDT 2017
`Downloaded/printed by
`Bill Bouchard (Plan Tech Inc) pursuant to License Agreement. No further reproductions authorized.
`
`ClearCorrect Exhibit 1059, Page 2 of 13
`
`

`

`D2240 − 15
`
`FIG. 1 (b) Type B and D Indentor (continued)
`
`FIG. 1 (c) Type O, DO, and OO Indentor (continued)
`
`FIG. 1 (d) Type M Indentor (continued)
`
`(1) The display shall indicate from 0 to 100 with no less
`than 100 equal divisions throughout the range at a rate of one
`hardness point for each 0.025 mm (0.001 in.) of indentor
`movement,
`(2) The display for Type OOO-S durometers shall indicate
`from 0 to 100 with no less than 100 equal divisions throughout
`the range at a rate of one hardness point for each 0.050 mm
`(0.002 in.) of indentor movement,
`(3) The display for Type M durometers shall indicate from
`0 to 100 with no less than 100 equal divisions at a rate of one
`hardness point for each 0.0125 mm (0.0005 in.) of indentor
`movement, and
`
`(4) In the case of analog dial indicators having a display of
`360°, the points indicating 0 and 100 may be at the same point
`on the dial and indicate 0, 100, or both.
`5.1.1.9 Timing Device (optional), capable of being set to a
`desired elapsed time, signaling the operator or holding the
`hardness reading when the desired elapsed time has been
`reached. The timer shall be automatically activated when the
`presser foot is in contact with the specimen being tested, for
`example, the initial indentor travel has ceased. Digital elec-
`tronic durometers may be equipped with electronic timing
`
`3
`
`
`
`Copyright by ASTM Int'l (all rights reserved); Tue May 9 13:08:59 EDT 2017
`Downloaded/printed by
`Bill Bouchard (Plan Tech Inc) pursuant to License Agreement. No further reproductions authorized.
`
`ClearCorrect Exhibit 1059, Page 3 of 13
`
`

`

`D2240 − 15
`
`FIG. 1 (e) Type OOO Indentor (continued)
`
`FIG. 1 (f) Type OOO-S Indentor (continued)
`
`devices that shall not affect the indicated reading or determi-
`nations attained by more than one-half of the calibration
`tolerance stated in Table 1.
`5.1.1.10 Maximum Indicators (optional), maximum indicat-
`ing pointers are auxiliary analog indicating hands designed to
`remain at the maximum hardness value attained until reset by
`the operator. Electronic maximum indicators are digital dis-
`plays electronically indicating and maintaining the maximum
`value hardness valued achieved until reset by the operator.
`5.1.1.11 Analog maximum indicating pointers have been
`shown to have a nominal effect on the values attained,
`however, this effect is greater on durometers of lesser total
`mainspring loads; for example,
`the effect of a maximum
`indicating pointer on Type D durometer determinations will be
`less than those determinations achieved using a Type A
`durometer. Analog style durometers may be equipped with
`maximum indicating pointers. The effect of a maximum
`indicating pointer shall be noted at the time of calibration in the
`calibration report (see 10.1.5), and when reporting hardness
`
`determinations (see 10.2.4). Analog Type M, OO, OOO, and
`Type OOO-S durometers shall not be equipped with maximum
`indicating pointers.
`5.1.1.12 Digital electronic durometers may be equipped
`with electronic maximum indicators that shall not affect the
`indicated reading or determinations attained by more than one
`half of the spring calibration tolerance stated in Table 1.
`for applying force to the
`5.1.1.13 Calibrated Spring,
`indentor, in accordance with Fig. 1 (a through g) and capable
`of applying the forces as specified in Table 1.
`5.1.2 Operating Stand (Fig. 2):
`5.1.2.1 Type 1, Type 2, and Type 3 shall be capable of
`supporting the durometer presser foot surface parallel to the
`specimen support table (Fig. 3) throughout the travel of each.
`The durometer presser foot to specimen support table parallel-
`ism shall be verified each time the test specimen support table
`is adjusted to accommodate specimens of varying dimensions.
`This may be accomplished by applying the durometer presser
`foot to the point of contact with the specimen support table and
`
`4
`
`
`
`Copyright by ASTM Int'l (all rights reserved); Tue May 9 13:08:59 EDT 2017
`Downloaded/printed by
`Bill Bouchard (Plan Tech Inc) pursuant to License Agreement. No further reproductions authorized.
`
`ClearCorrect Exhibit 1059, Page 4 of 13
`
`

`

`D2240 − 15
`
`FIG. 1 (g) Type E Indentor (continued)
`
`TABLE 1 Durometer Spring Force CalibrationA
`All Values are in N
`Type C, D, DO
`Type M
`0
`0.324
`4.445
`0.368
`8.89
`0.412
`13.335
`0.456
`17.78
`0.5
`22.225
`0.544
`26.67
`0.589
`31.115
`0.633
`35.56
`0.677
`40.005
`0.721
`44.45
`0.765
`0.4445
`0.0044
`± 0.4445 N
`± 0.0176 N
`
`Type A, B, E, O
`0.55
`1.3
`2.05
`2.8
`3.55
`4.3
`5.05
`5.8
`6.55
`7.3
`8.05
`0.075
`± 0.075 N
`
`Type OO, OOO
`0.203
`0.294
`0.385
`0.476
`0.566
`0.657
`0.748
`0.839
`0.93
`1.02
`1.111
`0.00908
`± 0.0182 N
`
`Type OOO-S
`0.167
`0.343
`0.520
`0.696
`0.873
`1.049
`1.226
`1.402
`1.579
`1.755
`1.932
`0.01765
`± 0.0353 N
`
`Indicated Value
`0
`10
`20
`30
`40
`50
`60
`70
`80
`90
`100
`N/durometer unit
`Spring Calibration
`Tolerance
`
`A Refer to 5.1.1.3 for the Type xR designation.
`
`making adjustments by way of the durometer mounting assem-
`bly or as specified by the manufacturer.
`5.1.2.2 Operating Stand, Type 1 (specimen to indentor
`type), shall be capable of applying the specimen to the indentor
`in a manner that minimizes shock.
`5.1.2.3 Operating Stand, Type 2 (indentor to specimen
`type), shall be capable of applying the indentor to the specimen
`in a manner that minimizes shock.
`5.1.2.4 Operating Stand, Type 3 (indentor to specimen
`type), hydraulic dampening, pneumatic dampening, or electro-
`mechanical (required for the operation of Type M durometers)
`shall be capable of controlling the rate of descent of the
`indentor to the specimen at a maximum of 3.2 mm/s (0.125
`in./s) and applying a force sufficient to overcome the calibrated
`spring force as shown in Table 1. Manual application, Type 1
`or Type 2 operating stands are not acceptable for Type M
`durometer operation.
`5.1.2.5 The entire instrument should be plumb and level,
`and resting on a surface that will minimize vibration. Operating
`the instrument under adverse conditions will negatively affect
`the determinations attained.
`5.1.2.6 Specimen Support Table, (Fig. 3) integral to the
`operating stand, and having a solid flat surface. The specimen
`
`support platform may have orifices designed to accept various
`inserts or support fixtures (Fig. 3) to provide for the support of
`irregularly configured specimens. When inserts are used to
`support test specimens, care must be taken to align the indentor
`to the center of the insert, or the point at which the indentor is
`to contact the specimen. Care should be exercised to assure that
`the indentor does not abruptly contact the specimen support
`table as damage to the indentor may result.
`
`6. Test Specimen
`6.1 The test specimen, herein referred to as “specimen” or
`“test specimen” interchangeably, shall be at least 6.0 mm (0.24
`in.) in thickness unless it is known that results equivalent to the
`6.0-mm (0.24-in.) values are obtained with a thinner specimen.
`6.1.1 A specimen may be composed of plied pieces to obtain
`the necessary thickness, but determinations made on such
`specimens may not agree with those made on solid specimens,
`as the surfaces of the plied specimens may not be in complete
`contact. The lateral dimensions of the specimen shall be
`sufficient to permit measurements at least 12.0 mm (0.48 in.)
`from any edge, unless it is known that identical results are
`obtained when measurements are made at a lesser distance
`from an edge.
`
`5
`
`
`
`Copyright by ASTM Int'l (all rights reserved); Tue May 9 13:08:59 EDT 2017
`Downloaded/printed by
`Bill Bouchard (Plan Tech Inc) pursuant to License Agreement. No further reproductions authorized.
`
`ClearCorrect Exhibit 1059, Page 5 of 13
`
`

`

`D2240 − 15
`
`FIG. 2 Durometer Operating Stand
`
`6.1.2 The surfaces of the specimen shall be flat and parallel
`over an area to permit the presser foot to contact the specimen
`
`over an area having a radius of at least 6.0 mm (0.24 in.) from
`the indentor point. The specimen shall be suitably supported to
`
`FIG. 3 Small Specimen Support Table
`
`6
`
`
`
`Copyright by ASTM Int'l (all rights reserved); Tue May 9 13:08:59 EDT 2017
`Downloaded/printed by
`Bill Bouchard (Plan Tech Inc) pursuant to License Agreement. No further reproductions authorized.
`
`ClearCorrect Exhibit 1059, Page 6 of 13
`
`

`

`D2240 − 15
`
`7.1.2 Indentor extension and shape shall be in accordance
`with 5.1.1.5, 5.1.1.6, or5.1.1.7, respective to durometer type.
`See Fig. 1 (a through g). Examination of the indentor under
`20× magnification, 50× for Type M indentors, is required to
`examine the indentor condition. Misshapen or damaged inden-
`tors shall be replaced.
`7.1.3 A combination of dimensional gage blocks shall be
`used to achieve a difference of 2.54 + 0.00/–0.0254 mm (0.100
`+ 0.00/–0.001 in.) between them. For Type OOO-S durometers,
`the gage block dimensions are 5.08 + 0.00/–0.0508 mm (0.200
`+ 0.00/–0.002 in.). For Type M durometers, the gage block
`dimensions are 1.27 + 0.0/–0.0127 mm (0.050 + 0.00/–0.0005
`in.) between them (Fig. 4).
`7.1.4 Carefully lower the durometer presser foot until it
`contacts the largest dimensional block(s),
`the indentor tip
`should just contact the smaller block, verifying full indentor
`extension.
`7.1.5 Adjust the indentor extension to 2.50 6 0.04 mm
`(0.098 6 0.002 in.). For Type OOO-S durometers, adjust the
`indentor extension to 5.0 6 0.04 mm (0.198 6 0.002 in.). For
`Type M durometers, adjust the indentor extension to 1.25 6
`0.02 mm (0.049 6 0.001 in.), following the manufacturer’s
`recommended procedure.
`7.1.5.1 When performing the procedures in 7.1, care should
`be used so as not to cause damage to the indentor tip. Fig. 4
`depicts a suitable arrangement for gaging indentor extension.
`to the
`7.1.6 Parallelism of the durometer presser foot
`support surface, and hence the dimensional gage blocks, at the
`time of instrument calibration, may be in accordance with Test
`Methods D374, Machinist’s Micrometers, or otherwise accom-
`plished in accordance with the procedures specified by the
`manufacturer.
`7.2 Indentor Display Adjustment:
`7.2.1 After adjusting the indentor extension as indicated in
`7.1, use a similar arrangement of dimensional gage blocks to
`verify the linear relationship between indentor travel and
`indicated display at two points: 0 and 100. Following the
`manufacturer’s recommendations, make adjustments so that:
`7.2.2 The indicator displays a value equal to the indentor
`travel measured to within:
`–0.0 +1.0 durometer units measured at 0;
`60.50 durometer units measured at 100;
`61 durometer units at all other points delineated in 7.4.
`7.2.3 Each durometer point indicated is equal to 0.025 mm
`(0.001 in.) of indentor travel, except for:
`7.2.3.1 Type M Durometers, each indicated point is equal to
`0.0125 mm (0.0005 in.) of indentor travel;
`7.2.3.2 Type OOO-S Durometers, each indicated point is
`equal to 0.050 mm (0.002 in.) of indentor travel.
`7.2.4 The indicator shall not display a value greater than 100
`or less than 0 at the time of calibration.
`7.2.5 Other means of determining indentor extension or
`indentor travel, such as optical or laser measurement methods,
`are acceptable. The instrumentation used shall have traceability
`as described in 1.4.
`7.2.6 The durometer shall be supported in a suitable fashion
`when performing the procedures described in 7.1 and 7.2.
`7.3 Calibration Device:
`
`FIG. 4 Detail of Indentor Extension and Display Adjustment
`
`provide for positioning and stability. A suitable hardness
`determination cannot be made on an uneven or rough point of
`contact with the indentor.
`6.2 Type OOO, OOO-S, and M test specimens should be at
`least 1.25 mm (0.05 in.) in thickness, unless it is known that
`results equivalent to the 1.25-mm (0.05-in.) values are obtained
`with a thinner specimen.
`6.2.1 A Type M specimen that is not of a configuration
`described in 6.2.2 may be composed of plied pieces to obtain
`the necessary thickness, but determinations made on such
`specimens may not agree with those made on solid specimens
`because the surfaces of the plied specimens may not be in
`complete contact. The lateral dimensions of the specimen
`should be sufficient to permit measurements at least 2.50 mm
`(0.10 in.) from any edge unless it is known that identical results
`are obtained when measurements are made at lesser distance
`from an edge. A suitable hardness determination cannot be
`made on an uneven or rough point of contact with the indentor.
`6.2.2 The Type M specimen, when configured as an o-ring,
`circular band, or other irregular shape shall be at least 1.25 mm
`(0.05 in.) in cross-sectional diameter, unless it is known that
`results equivalent to the 1.25-mm (0.05-in.) values are obtained
`with a thinner specimen. The specimen shall be suitably
`supported in a fixture (Fig. 3) to provide for positioning and
`stability.
`6.3 The minimum requirement for the thickness of the
`specimen is dependent on the extent of penetration of the
`indentor into the specimen; for example, thinner specimens
`may be used for materials having higher hardness values. The
`minimum distance from the edge at which measurements may
`be made likewise decreases as the hardness increases.
`
`7. Calibration
`7.1 Indentor Extension Adjustment Procedure:
`7.1.1 Place precision ground dimensional blocks (Grade B
`or better) on the support table and beneath the durometer
`presser foot and indentor. Arrange the blocks so that
`the
`durometer presser foot contacts the larger block(s) and the
`indentor tip just contacts the smaller block (Fig. 4). It is
`necessary to observe the arrangement of the blocks and the
`presser foot/indentor under a minimum of 20× magnification to
`assure proper alignment.
`
`7
`
`
`
`Copyright by ASTM Int'l (all rights reserved); Tue May 9 13:08:59 EDT 2017
`Downloaded/printed by
`Bill Bouchard (Plan Tech Inc) pursuant to License Agreement. No further reproductions authorized.
`
`ClearCorrect Exhibit 1059, Page 7 of 13
`
`

`

`D2240 − 15
`
`7.7 Spring Force Combinations:
`7.7.1 For Type A, B, E, and O durometers:
`Force, N = 0.55 + 0.075 HA
`Where HA = hardness reading on Type A, B, E, and O
`durometers.
`7.7.2 For Type C, D, and DO durometers:
`Force, N = 0.4445 HD
`Where HD = hardness reading on Type C, D, and DO
`durometers.
`7.7.3 For Type M durometers:
`Force, N = 0.324 + 0.0044 HM
`Where HM = hardness reading on Type M durometers.
`7.7.4 For Type OO and OOO durometers:
`Force, N = 0.203 + 0.00908 HOO
`Where HOO = hardness reading on Type OO durometers.
`7.7.5 For Type OOO-S durometers:
`Force, N = 0.167 + 0.01765 HOOO-S
`Where HOOO-S = hardness reading on Type OOO-S
`durometers.
`
`7.8 The rubber reference block(s) provided for verifying
`durometer operation and state of calibration are not to be relied
`upon as calibration standards. The calibration procedures
`outlined in Section 7 are the only valid calibration procedures.
`7.8.1 The use of metal reference blocks is no longer
`recommended (see Note 2).
`
`7.9 Verifying the state of durometer calibration, during
`routine use, may be accomplished by:
`7.9.1 Verifying that the zero reading is no more than 1
`indicated point above zero, and not below zero (on durometers
`so equipped), when the durometer is positioned so that no
`external force is placed upon the indentor.
`7.9.2 Verifying that the 100 reading is no more than 100 and
`no less than 99 when the durometer is positioned on a flat
`surface of a non-metallic material so that the presser foot is in
`complete contact, causing the indentor to be fully retracted.
`7.9.2.1 It is important that when performing the verification
`of 100, as described in 7.9.2, that extreme care be taken so as
`to not cause damage to the indentor. Verification of the 100
`value is not recommended for durometers having a spring force
`greater than 10 N (Types C, D, and DO).
`7.9.2.2 When performing the verification of 100, as de-
`scribed in 7.9.2,
`the non-metallic material shall be of a
`hardness value greater than 100 of the type (scale) of the
`durometer being employed. Tempered glass of a thickness
`greater than 6.35 mm (0.25 in.) has been found satisfactory for
`this application.
`7.9.3 Verifying the displayed reading at any other point
`using commercially available rubber reference blocks which
`are certified to a stated value of the type (scale) of the
`durometer being employed. The displayed value of the durom-
`eter should be within 62 durometer points of the reference
`block’s stated value.
`7.9.4 Verification of the zero and 100 readings of a durom-
`eter provide reasonable assurance that the linear relationship
`between the indicated display and the durometer mechanism
`remain valid.
`
`FIG. 5 Example of Durometer Calibration Apparatus
`
`7.3.1 The durometer spring shall be calibrated by support-
`ing the durometer in a calibrating device, see Fig. 5, in a
`vertical position and applying a measurable force to the
`indentor tip. The force may be measured by means of a balance
`as depicted in Fig. 5, or an electronic force cell. The calibrating
`device shall be capable of measuring applied force to within
`0.5 % of the maximum spring force necessary to achieve 100
`durometer units.
`7.3.2 Care should be taken to ensure that the force is applied
`vertically to the indentor tip, as lateral force will cause errors
`in calibration. See 7.1.5.1 and 7.1.6.
`7.4 Spring Calibration—The durometer spring shall be
`calibrated at displayed readings of 10, 20, 30, 40, 50, 60, 70,
`80, and 90. The measured force (9.8× mass in kilograms) shall
`be within the spring calibration tolerance specified in Table 1.
`Table 1 identifies the measured force applied to the indentor for
`the entire range of the instrument, although it is necessary only
`to verify the spring calibration at points listed herein.
`7.5 Spring Calibration Procedure:
`7.5.1 Ensure that the indentor extension has been adjusted
`in accordance with 7.1, and the linear relationship between
`indentor travel and display is as specified in 7.2.
`7.5.2 Place the durometer in the calibration device as
`depicted in Fig. 5. Apply the forces indicated in Table 1 so that
`forces applied are aligned with the centerline of the indentor in
`a fashion that eliminates shock or vibration and adjust the
`durometer according to manufacturers’ recommendations so
`that:
`7.5.3 At the points enumerated in 7.4, the display shall
`indicate a value equal to 0.025 mm (0.001 in.) of indentor
`travel. For Type OOO-S durometers, the display shall indicate
`a value equal to 0.05 mm (0.002 in.) of indentor travel. For
`Type M durometers, the display shall indicate a value equal to
`0.0125 mm (0.0005 in.) of indentor travel within the spring
`calibration tolerances specified in 7.6.
`7.6 Spring calibration tolerances are 61.0 durometer units
`for Types A, B, C, D, E, O, and DO, 62.0 durometer units for
`Types OO, OOO, and OOO-S, and 64.0 durometer units for
`Type M, while not indicating below 0 or above 100 at the time
`of calibration (see Table 1).
`
`8
`
`
`
`Copyright by ASTM Int'l (all rights reserved); Tue May 9 13:08:59 EDT 2017
`Downloaded/printed by
`Bill Bouchard (Plan Tech Inc) pursuant to License Agreement. No further reproductions authorized.
`
`ClearCorrect Exhibit 1059, Page 8 of 13
`
`

`

`D2240 − 15
`
`7.9.5 Verification of points between zero and 100 provide
`reasonable assurance that the curvilinear relationship between
`the indicated display and the durometer mechanism remain
`valid.
`7.9.6 This is not a calibration procedure, it is a means by
`which a user may routinely verify that the durometer may be
`functioning correctly. (See Note 2.)
`
`8. Laboratory Atmosphere and Test Specimen
`Conditioning
`8.1 Tests shall be conducted in the standard laboratory
`atmosphere, as defined in Practice D618, Section 4.2.
`8.2 The instrument shall be maintained in the standard
`laboratory atmosphere, as defined in Practice D618, Section
`4.1, for 12 h prior to performing a test.
`8.3 The specimen shall be conditioned in accordance with
`condition 40/23 exclusive of humidity control, as described in
`Practice D618, Section 8.1, Procedure A and tested under the
`same conditions, exclusive of humidity control.
`8.4 These procedures may be modified if agreed upon
`between laboratories or between supplier and user and are in
`accordance with alternative procedures identified in Practice
`D618.
`8.5 No conclusive evaluation has been made on durometers
`at temperatures other than 23.0 6 2.0°C (73.4 6 3.6°F).
`Conditioning at temperatures other than the above may show
`changes in calibration. Durometer use at temperatures other
`than the above should be decided locally (see Practice D1349).
`
`9. Procedure
`9.1 Operating Stand Operation (Type 3 Operating Stand
`Required for Type M):
`9.1.1 Care shall be exercised to minimize the exposure of
`the instrument to environmental conditions that are adverse to
`the performance of the instrument, or adversely affect test
`results.
`9.1.2 Adjust the presser foot to support table parallelism as
`described in 5.1.2.1. It is necessary to make this adjustment
`each time the support table is moved to accommodate speci-
`mens of varying dimensions.
`9.1.3 Prior to conducting a test, adjust the vertical distance
`from the presser foot to the contact surface of the test specimen
`to 25.4 6 2.5 mm (1.00 6 0.100 in.), unless it is known that
`identical results are obtained with presser foot at a greater or
`lesser vertical distance from the test specimen contact surface,
`or if otherwise stipulated by the manufacturer.
`9.1.4 Place the specimen on the specimen support table, in
`a manner that the contact point of the indentor is in accordance
`with Section 6, unless it is known that identical results are
`obtained when measurements are made with the indentor at a
`lesser distance from the edge of the test specimen.
`9.1.5 Actuate the release lever (Fig. 2) of the operating
`stand or activate the electromechanical device, allowing the
`durometer to descend at a controlled rate and apply the presser
`foot to the specimen in accordance with 5.1.2. In the case of
`“specimen to indentor” type operating stands, operate the lever
`or other mechanism to apply the specimen to the indentor in a
`
`manner that assures parallel contact of the specimen to the
`durometer presser foot without shock and with just sufficient
`force to overcome the calibrated spring force as shown in Table
`1.
`
`9.1.6 An operating stand that applies the mass at a con-
`trolled rate of descent, without shock is mandatory for Type M
`durometers. Hand-held application or the use of a Type 1 or
`Type 2 operating stand for the Type M durometer is not an
`acceptable practice, see 5.1.2.4.
`9.1.7 For any material covered in 1.1, once the presser foot
`is in contact with the specimen, for example, when the initial
`indentor travel has ceased, the maximum indicated reading
`shall be recorded. The time interval of 1 s, between initial
`indentor travel cessation and the recording of the indicated
`reading, shall be considered standard. Other time intervals,
`when agreed upon among laboratories or between supplier and
`user, may be used and reported accordingly. The indicated
`hardness reading may change with time.
`9.1.7.1 If the durometer is equipped with an electronic
`maximum indicator or timing device (refer to 5.1.1.9) the
`indicated reading shall be recorded within 1 6 0.3 s of the
`cessation of indentor travel and reported (refer to 10.2.9

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket