throbber
..
`
`if
`I I
`
`.
`
`.
`
`.. ·~ -. .
`
`. ~ .
`
`.
`
`. ..
`
`' . . .
`
`'"
`
`~ .·. ,•
`
`USP
`NF
`
`i•I
`
`:. I
`I I
`
`The Official Compendia of Standards
`
`Volume 1
`
`1 of 22
`
`Fresenius Kabi
`Exhibit 1028
`
`

`

`2007
`
`USP 30
`
`NF 25
`
`Volume I
`
`THE UNITED STATES PHARMACOPEIA
`
`THE NATIONAL FORMULARY
`
`By authority of The United States Pharmacopeial
`Convention, meeting at Washington, D. C., March 9-13,
`2005. Prepared by the Council of Experts and published
`by the Board of Trustees
`
`Official from May 1, 2007
`
`The designation on the cover of this publication, "USP NF
`2007," is for ease of identification only. The publication
`contains two separate compendia: The United States
`Pharmacopeia, Thirtieth Revision, and the National
`Formulary, Twenty-Fifth Edition.
`
`Goodwin Procter LLP
`Ubrary
`The New York TfmeQ BuUdlng
`620 Eighth Avenue
`New York, NY 10018-1400
`
`THE UNITED STATES PHARMACOPEIAL CONVENTION
`12601 Twinbrook Parkway, Rockville, MD 20852
`
`2 of 22
`
`Fresenius Kabi
`Exhibit 1028
`
`

`

`SIX-MONTH IMPLEMENTATION GUIDELINE
`
`Beginning with USP30-NF25, the United States Pharmacopeid-National Formulary and its Supplements will become official six months
`after being released to the public. The USP-NF, which is released on November 1 of each year, will become official on May 1 of the following
`year.
`This change was adopted to give users more time to bring their methods and procedures into compliance with new and revised USP-NF
`requirements.
`The table below describes the new official dates. The 2006 USP29-NF24, and the Supplements and Interim Revision Announcements (IRAs)
`to that edition, will be official until May 1, 2007, at which time the USP30-NF25 becomes official.
`
`Publication
`
`Release Date
`
`Official Date
`
`Official Until
`
`USP30-NF25
`
`Nov. 1, 2006
`
`May 1, 2007
`
`May 1, 2008 ( except as superceded by Supplements, IRAs, and
`Revision Bulletins)
`May 1, 2008 ( except as superceded by Second Supplement, IRAs,
`and Revision Bulletins)
`May 1, 2008 ( except as superceded by IRAs and Revision
`Bulletins)
`May 1, 2009 ( except as superceded by Supplements, IRAs, and
`Revision Bulletins)
`IRAs will continue to become official on the first day of the second month of the Pharmacopeial Forum (PF) issue in which they are
`published as final. For instance, IRAs published as final in the May-June PF (issue 3) will become official on June 1. This table gives the details
`of the IRAs that will apply to USP29-NF24 and USP30-NF25.
`
`First Supplement
`
`Feb. 1, 2007
`
`Aug. 1, 2007
`
`Second Supplement
`
`June 1, 2007
`
`Dec. 1, 2007
`
`USP31 - NF26
`
`Nov. 1, 2007
`
`May 1, 2008
`
`IRA*
`
`Release Date
`
`Official Date
`
`Revises
`
`USP29-NF24 and its Supplements
`Feb. 1, 2007
`Jan. 1, 2007
`Jan. 1, 2007 IRA, PF 33(1)
`USP29-NF24 and its Supplements
`April 1, 2007
`Mar. 1, 2007
`Mar. 1, 2007 IRA, PF 33(2)
`USP30-NF25
`June 1, 2007
`May 1, 2007
`May 1, 2007 IRA, PF 33(3)
`USP30-NF25 and First Supplement
`Aug. 1, 2007
`July 1, 2007
`July 1, 2007 IRA, PF 33(4)
`USP30-NF25 and First Supplement
`Oct. 1, 2007
`Sept. 1, 2007
`Sept. 1, 2007 IRA, PF 33(5)
`USP30-NF25 and its Supplements
`Dec. 1, 2007
`Nov. 1, 2007
`Nov. 1, 2007 IRA, PF 33(6)
`USP30- NF25 and its Supplements
`Feb. 1, 2008
`Jan. 1, 2008
`Jan. 1, 2008 IRA, PF 34(1)
`USP30-NF25 and its Supplements
`April 1, 2008
`Mar. 1, 2008
`Mar. 1, 2008 IRA, PF 34(2)
`*NOTE-Beginning January 1, 2007, USP will cease identifying IRAs numerically (First, Second, etc.) and instead will designate them by the
`date on which they are published.
`Revision Bulletins published on the USP website will continue to become official immediately upon publication, unless the Revision Bulletin
`specifies otherwise.
`General Chapters, monographs, or monograph revisions that contain a specific official date shall continue to become official upon such
`specified date, which supercedes the general official date for the publication.
`For more information about the change in official dates, please visit the USP website at http://www.usp.org.
`
`NOTICE AND WARNING
`
`Concerning U.S. Patent or Trademark Rights
`
`The inclusion in The United States Pharmacopeia or in the National Formula,y of a monograph on any drug in respect to which patent or
`trademark rights may e?.(i1:,t _sh'clll·not be.deemed, and is not intended as, a grant of, or authority to exercise, any right or privilege protected by
`such patent or trademark. All such rights_.llllcl :privileges are vested in the patent or trademark owner, and no other person may exercise the same
`without express p~rmissi~l).:,:·authority, .or)icense secured from such patent or trademark owner .
`. ,·,. ; .·
`~ •-.... ~- "' :
`
`'
`
`Concerning Use of USP or NF Text
`
`Attention is called to the fact that USP and NF text is fully copyrighted. Authors and others wishing to use portions of the text should request
`permission to do so from the Secretary of the USPC Board of Trustees.
`
`Copyright © 2006 The United States Pharmacopeial Convention
`12601 Twinbrook Parkway, Rockville, MD 20852
`All rights reserved.
`ISSN 0 195-7996
`ISBN 1-889788-47-3
`Printed in the United States by Port City Press, Baltimore
`
`3 of 22
`
`Fresenius Kabi
`Exhibit 1028
`
`

`

`single
`above
`rrts ro;
`:d for a
`sed for
`within
`
`!Ve are
`:;al dry
`ifferent
`
`mercial
`ving. A
`air jet
`as that
`izedair
`1uential
`Jbtain a
`use of
`g. This
`:actions
`
`the test
`hat lifts
`ngs at a
`,werthe
`
`eful for
`1es are
`
`rsion of
`hard to ·
`.grange
`ive, and
`elop an
`1ation is
`that the
`,osed of
`
`the total
`:t values
`Lined on
`convert
`; desired
`1dersize,
`:hall the
`. that the
`i during
`
`USP 30
`
`Physical Tests I (788) Particulate Matter in Injections
`
`321
`
`(788) PARTICULATE MATTER IN
`INJECTIONS
`
`Particulate matter consists of mobile, randomly-sourced, extrane(cid:173)
`ous substances, other than gas bubbles, that cannot be quantitated by
`chemical analysis due to the small amount of material that it
`represents a~d to its_ heteroge~eous composi~ion. ~je~table solu(cid:173)
`tions includmg solut10ns constituted from stenle solids mtended for
`pare~teral use, is essentially free from particulate matter that can be
`observed on visual inspection. The tests described herein are
`physical tests performed for the purpose of enumerating subvisible
`extraneous particles within specific size ranges.
`Microscopic and light obscuration procedures for the determina(cid:173)
`tion of particulate matter are given herein. This chapter provides a
`test approach in two st~ges. The injection is first tested by the light
`obscuration procedure (stage 1). If it fails to meet the prescribed
`limits, it must pass the microscopic procedure (stage 2) with its own
`set of test limits. Where for technical reasons the injection cannot be
`tested by light obscuration, microscopic testing may be used
`exclusively. Documentation demonstrating that the light obscuration
`procedure is incapable of testing the injection or produces invalid
`results is required in each case. It is expected that most articles will
`meet the requirements on the basis of the light obscuration test alone;
`however, it may be necessary to test some articles by the light
`obscuration test followed by the microscopic test to reach a
`conclusion on conformance to requirements.
`All large-volume injections for single-dose infusion and those
`small-volume injections for which the monographs specify such
`requirements are subject to the particulate matter limits set forth for
`the test being applied, unless otherwise specified in the individual
`monograph. Excluded from the requirements of this chapter are
`injections intended solely for intramuscular and subcutaneous
`administration.
`Not all injection formulations can be examined for particles by
`one or both of these tests. Any product that is not a pure solution
`having a clarity and a viscosity approximating those of water may
`provide erroneous data when analyzed by the light obscuration
`counting method. Such materials may be analyzed by the
`microscopic method. Emulsions, colloids, and liposomal prepara(cid:173)
`tions are examples. Similarly, products that produce air or gas
`bubbles when drawn into the sensor, such as bicarbonate-buffered
`formulations, may also require microscopic testing. Refer to the
`specific monographs when a question of test applicability occurs.
`Higher limits are appropriate for certain articles and will be specified
`in the individual monographs.
`In some instances, the viscosity of a material to be tested may be
`sufficiently high so as to preclude its analysis by either test method.
`In this event, a quantitative dilution with an appropriate diluent may
`be made to decrease viscosity, as necessary, to allow the analysis to
`be performed.
`In the tests described below for large-volume and small-volume
`injections, the results obtained in examining a discrete unit or group
`of units for particulate matter cannot be extrapolated with certainty
`to other units that remain untested. Thus, statistically sound
`sampling plans based upon known operational factors must be
`developed if valid inferences are to be drawn from observed data to
`characterize the level of particulate matter in a large group of units.
`Sampling plans should be based on consideration of product volume,
`numbers of particles historically found to be present in comparison
`to limits, particle size distribution of particles present, and variability
`of particle counts between units.
`
`LIGHT OBSCURATION PARTICLE COUNT
`TEST
`
`USP Reference Standards ( 11 )-USP Particle Count RS.
`The test applies to large-volume injections labeled as containing
`more than 100 mL, unless otherwise specified in the individual
`mo_nograph. It counts suspended particles that are solid or liquid.
`This test applies also to single-dose or multiple-dose small-volume
`
`injections labeled as containing 100 mL or less that are either in
`solution or in solution constituted from sterile solids, where a test for
`particulate matter is specified in the individual monograph. Products
`for which the individual monograph specifies that the label states
`that the product is to be used with a final filter are exempt from these
`requirements.
`
`Test Apparatus
`
`The apparatus is an electronic, liquid-borne particle counting
`system that uses a light-obscuration sensor with a suitable sample(cid:173)
`feeding device. A variety of suitable devices of this type are
`commercially available. It is the responsibility of those performing
`the test to ensure that the operating parameters of the instrumentation
`are appropriate to the required accuracy and precision of the test
`result, and that adequate training is provided for those responsible
`for the technical performance of the test.
`It is important to note that for Pharmacopeial applications the
`ultimate goal is that the particle counter reproducibly size and count
`particles present in the injectable material under investigation. The
`instruments available range from systems where calibration and
`other components of standardization must be carried out by manual
`procedures to sophisticated systems incorporating hardware- or
`software-based functions for the standardization procedures. Thus, it
`is not possible to specify exact methods to be followed for
`standardization of the instrument, and it is necessary to emphasize
`the required end result of a standardization procedure rather than a
`specific method for obtaining this result. This section is intended to
`emphasize the criteria ·that must be met by a system rather than
`specific methods to be used in their determination. It is the
`responsibility of the user to apply the various methods of
`standardization applicable to a specific instrument. Critical opera(cid:173)
`tional criteria consist of the following.
`Sensor Concentration Limits-Use an instrument that has a
`concentration limit (the maximum number of particles per mL)
`identified by the manufacturer that is greater than the concentration
`of particles in the test specimen to be counted. The vendor-certified
`concentration limit for a sensor is specified as that count level at
`which coincidence counts due to simultaneous presence of two or
`more particles in the sensor view volume comprise less than 10% of
`the counts collected for 10-µm particles.
`Sensor Dynamic Range-The dynamic range of the instrument
`used (range of sizes of particles that can be accurately sized and
`counted) must include the smallest particle size to be enumerated in
`the test articles.
`
`Instrument Standardization
`The following discussion of instrument standardization empha(cid:173)
`sizes performance criteria rather than specific methods for calibrating
`or standardizing a given instrument system. This approach is
`particularly evident in the description of calibration, where
`allowance must be made for manual methods as well as those
`based on firmware, software, or the use of electronic testing
`instruments. Appropriate instrument qualification is essential to
`performance of the test according to requirements. Since different
`brands of instruments may be used in the test, the user is responsible
`for ensuring that the counter used is operated according to the
`manufacturer's specific instructions; the principles to be followed to
`ensure that instruments operate within acceptable ranges are defined
`below.

`The following information for instrument standardization helps
`ensure that the sample volume accuracy, sample flow rate, particle
`size response curve, sensor resolution, and count accuracy are
`appropriate to performance of the test. Conduct these procedures at
`intervals of not more than six months.
`
`SAMPLE VOLUME ACCURACY
`
`Since the particle count from a sample aliquot varies directly with
`the volume of fluid sampled, it is important that the sampling
`accuracy is known to be within a certain range. For a sample volume
`
`4 of 22
`
`Fresenius Kabi
`Exhibit 1028
`
`

`

`322
`
`(788) Particulate Matter in Injections / Physical Tests
`
`USP 30
`
`[}SP 3l
`
`determination, determine the dead (tare) volume in the sample feeder
`with filtered distilled or deionized water that has been passed
`through a filter having a 1.2-µm or finer porosity. Transfer a volume
`of filtered distilled or deionized water that is greater than the sample
`volume to a container, and weigh. Withdraw through the sample
`feeding device a volume that is appropriate for the specific sampler,
`and again weigh the container. Determine the sample volume by
`subtracting the tare volume from the combined sample plus tare
`volumes. Verify that the value obtained is within 5% of the
`appropriate sample volume for the test. Alternatively, the sample
`volume may be determined using a suitable Class A graduated
`cylinder (see Volumetric Apparatus (31) ). [NOTE-Instruments of this
`type require a variable tare volume. This is the amount ot: sample
`withdrawn prior to counting. This volume may be determmed for
`syringe-operated samplers by setting the sample vol~me to zer? and
`initiating sampling, so that the only volume of solut10n drawn 1s the
`tare. Subtract the tare volume from the total volume of solution
`drawn in the sampling cycle to determine the sample volume.]
`
`SAMPLE FLOW RATE
`
`Verify that the flow rate is within the manufacturer's specifications
`for the sensor used. This may be accomplished by using a calibrated
`stopwatch to measure the time required for the instrument to
`withdraw and count a specific sample volume (i.e., the time between
`beginning and ending of the count cycle as denoted by instrument
`indicator lights or other means). Sensors may be operated accurately
`over a range of flow rates. Perform the Test Procedure at the same
`flow rate as that selected for calibration of the instrument.
`
`CALIBRATION
`
`Use one of the following methods.
`Manual Method-Calibrate the instrument with a minimum of
`three calibrators, each consisting of near-monosize polystyrene
`spheres having diameters of about 10, 15, and 25 µ1?1, in an aqu~o~s
`vehicle.* The calibrator spheres must have a mean diameter ofw1thm
`5% of the 10-, 15-, and 25-µm nominal diameters and be
`standardized against materials traceable to NIST standard reference
`materials. The number of spheres counted must be within the
`sensor's concentration limit. Prepare suspensions of the calibrator
`spheres in water at a concentration of 1000 to 5000 particles per mL,
`and determine the channel setting that corresponds to the highest
`count setting for the sphere distribution. This is determined by using
`the highest count threshold setting to split the distribution into two
`bins containing equal numbers of counts, with the instrument set in
`the differential count mode (moving window half-count method).
`Use only the central portion of the distribution in this calculat_ion to
`avoid including asymmetrical portions of the peak. The port10n of
`the distribution, which must be divided equally, is the count window.
`The window is bounded by threshold settings that will define a
`threshold voltage window of ± 20% around the mean diameter of
`the test spheres. The window is intended to include all single
`spheres, taking into account the standard deviation of the spheres
`and the sensor resolution, while excluding noise and aggregates of
`spheres. The value of 20% was chosen based on the W?r~t-case
`sensor resolution of 10% and the worst-case standard deviation of
`the spheres of 10%. Since the thresholds are proportional to the area
`of the spheres rather than the diameter, the lower and upper voltage
`settings are determined by the equations:
`Vr = 0.64V5
`in which Vr is the lower voltage setting and V5 is the voltage at the
`peak center, and
`
`in which Vu is the upper voltage setting.
`
`Vu= 1.44Vs
`
`Once the center peak thresholds are determined, use these
`thresholds for the standards to create a regression of log voltage
`versus log particle size, from which the instrument settings for the
`10- and 25-µm sizes can be determined.
`Automated Method-The calibration (size response) curve may
`be determined for the instrument-sensor system by the use of
`validated software routines offered by instrument vendors; these may
`be included as part of the instrument software or used in conjunction
`with a microcomputer interfaced to the counter. The use of these
`automated methods is appropriate if the vendor supplies written
`certification that the software provides a response curve equivalent to
`that attained by the manual method and if the automated calibration
`is validated as necessary by the user.
`Electronic Method-Using a multichannel peak height analyzer,
`determine the center channel of the particle counter pulse response
`for each standard suspension. This peak voltage setting becomes the
`threshold used for calculation of the voltage response curve for the
`instrument. The standard suspensions to be used for the calibration
`are run in order, and median pulse voltages for each are determined.
`These thresholds are then used to generate the size response curve
`manually or via software routines. The thresholds determined from
`the multichannel analyzer data are then transferred to the counter to
`complete the calibration. If this procedure is used with a comparator(cid:173)
`based instrument, the comparators of the counter must be adjusted
`accurately beforehand.
`
`SENSOR RESOLUTION
`
`The particle size resolution of the instrumental particle counter is
`dependent upon the sensor used and may vary with individual
`sensors of the same model. Determine the resolution of the particle
`counter for 10-µm particles using the 10-µm calibrator spheres. The
`relative standard deviation of the size distribution of the standard
`particles used is not more than 5%. Acceptable methods of
`determining particle size resolution are (1) manual determination
`of the amount of peak broadening due to instrument response; (2)
`using an electronic method of measuring and sorting particle sensor
`voltage output with a multichannel analyzer; and (3) automated
`methods.
`Manual Method-Adjust the particle counter to operate in the
`cumulative mode or total count mode. Refer to the calibration curve
`obtained earlier, and determine the threshold voltage for the 10-µm
`spheres. Adjust 3 channels of the counter to be used in the
`calibration procedure as follows:
`Channel 1 is set for 90% of the threshold voltage.
`Channel 2 is set for the threshold voltage.
`Channel 3 is set for 110% of the threshold voltage.
`Draw a sample through the sensor, observing the count in Channel
`2. When the particle count in that channel has reached approximately
`1000, stop counting, and observe the counts in Channels 1 and 3.
`Check to see if the Channel 1 count and the Channel 3 count are
`1.68 ± 10% and 0.32 ± 10%, respectively, of the count in Channel
`2. If not, adjust Channel 1 and Channel 3 thresholds to meet these
`criteria. When these criteria have been satisfied, draw a sample of
`suspension through the counter until the counts in Channel 2 have
`reached approximately 10,000, or until an appropriate volume (e.g.,
`10 mL) of the sphere suspension has been counted. Verify that
`Channel 1 and Channel 3 counts are 1.68 ± 3% and 0.32 ± 3%,
`respectively, of the count in Channel 2.
`Record the particle size for the thresholds just determined for
`Channels 1, 2, and 3. Subtract the particle size for Channel 2 frorn
`the size for Channel 3. Subtract the particle size for Channel 1 fro~
`the size for Channel 2. The values so determined are the observe
`standard deviations on the positive and negative side of the m~an
`count for the 10-µm standard. Calculate the percentage of resolution
`of the sensor by the formula:
`
`· ASTM standard F658-87 provides useful discussions pertaining to
`calibration procedures applying near-monosize latex spheres.
`
`in which S0 is the highest observed standard deviation determined
`for the sphere; S5 is the supplier's reported standard deviation for the
`
`spheres;
`the supp
`Autor
`deterrnin
`This sol
`conjunct
`of these
`written
`determin
`resolutio
`Elect1
`the parti
`suspensi
`resolutio
`down th
`identify
`approxir
`Use oft
`these rn
`within 1
`to calcu:
`
`Deter
`Method
`volume
`Meth
`ProC€:.
`Particle
`(total) n
`µm and
`25 time:
`80 to 1
`Remov(
`by han
`introduc
`the anal
`volume:
`discard
`procedu
`suspens
`resultin:
`greater
`portiorn
`number
`
`in whi
`suspem
`and Vi:
`the calc
`than 15
`Inter
`Particli
`or equa
`or equa
`confon
`the inst
`Accura
`blank.
`above
`Counti;
`meettb
`the fail
`Met
`Proc
`diamet
`SO and
`80 to
`Proper
`counts
`counte
`
`5 of 22
`
`Fresenius Kabi
`Exhibit 1028
`
`

`

`'P 30
`
`USP 30
`
`Physical Tests I (788) Particulate Matter in Injections
`
`323
`
`these
`oltage
`'or the
`
`e may
`1se of
`:emay
`nction
`~ these
`vritten
`lent to
`>ration
`
`alyzer,
`;ponse
`1es the
`for the
`Jration
`mined.
`curve
`i from
`nter to
`arator(cid:173)
`ijusted
`
`mter is
`ividual
`,article
`!S. The
`andard
`Jds of
`ination
`tse; (2)
`sensor
`Jmated
`
`in the
`1 curve
`10-µm
`in the
`
`:hannel
`imately
`and 3.
`unt are
`:hannel
`~t these
`nple of
`2 have
`1e (e.g.,
`ify that
`± 3%,
`
`ned for
`2 from
`1 from
`bserved
`e mean
`mlution
`
`ermined
`1 for the
`
`spheres; and D is the diameter, in µm, of the spheres as specified by
`the supplier. The resolution is not more than 10%.
`Automated Method-Software that allows for the automated
`determination of sensor resolution is available for some counters.
`This software may be included in the instrument or used in
`conjunction with a microcomputer interfaced to the counter. The use
`of these automated methods is appropriate if the vendor supplies
`written certification that the software provides a resolution
`determination equivalent to the manual method and if the automated
`resolution determination is validated as necessary by the user.
`Electronic Method-Record the voltage output distribution of
`the particle sensor, using a multichannel analyzer while sampling a
`suspension of the 10-µm particle size standard. To determine
`resolution, move the cursor of the multichannel analyzer up and
`down the electric potential scale from the median pulse voltage to
`identify a channel on each side of the 10-µm peak that has
`approximately 61 % of the counts observed in the center channel.
`Use of the counter size response curve to convert the m V values of
`these two channels to particle sizes provides the particle size at
`within 1 standard deviation of the 10-µm standard. Use these values
`to calculate the resolution as described under Manual Method.
`
`PARTICLE COUNTING ACCURACY
`
`Determine the particle counting accuracy of the instrument, using
`Method 1 (for small-volume injections) or Method 2 (for large(cid:173)
`volume injections).
`Method 1-
`Procedure-Prepare the suspension and blank using the USP
`Particle Count RS. With the instrument set to count in the cumulative
`(total) mode, collect counts at settings of greater than or equal to 10
`µm and greater than or equal to 15 µm. Mix the blank by inverting
`25 times within 10 seconds, and degas the mixture by sonicating ( at
`80 to 120 watts) for about 30 seconds or by allowing to stand.
`Remove the closure from the container, and gently stir the contents
`by hand-swirling or by mechanical means, taking care not to
`introduce air bubbles or contamination. Stir continuously throughout
`the analysis. Withdraw directly from the container three consecutive
`volumes of not less than 5 mL each, obtain the particle counts, and
`discard the data from the first portion. [NOTE-Complete the
`procedure within five minutes.] Repeat the procedure, using the
`suspension in place of the blank. From the averages of the counts
`resulting from the analysis of the two portions of the suspension at
`greater than or equal to 10 µm and from the analysis of the two
`portions of the blank at greater than or equal to 10 µm, calculate the
`number of particles in each mL by the formula:
`
`(Ps - Po)! V
`in which Ps is the average particle count obtained from the
`suspension; P0 is the average particle count obtained from the blank;
`and Vis the average volume, in mL, of the 4 portions tested. Repeat
`the calculations, using the results obtained at the setting of not less
`than 15 µm.
`Interpretation-The instrument meets the requirements for
`Particle Counting Accuracy if the count obtained at greater than
`or equal to 10 µm and the ratio of the counts obtained at greater than
`or equal to 10 µm to those obtained at greater than or equal to 15 µm
`conform to the values that accompany the USP Particle Count RS. If
`the instrument does not meet the requirements for Particle Counting
`Accuracy, repeat the procedure using the remaining suspension and
`blank. If the results of the second test are within the limits given
`above, the instrument meets the requirements of the test for Particle
`Counting Accuracy. If on the second attempt the system does not
`meet the requirements of the test, determine and correct the source of
`the failures, and retest the instrument.
`Method 2-
`. Procedure-Using standard calibrator spheres having a nominal
`diameter of 15 to 30 µm, prepare a suspension containing between
`50 and 200 particles per mL. Degas the suspension by sonicating (at
`80 to 120 watts) for about 30 seconds or by allowing to stand.
`Properly suspend the particles by stirring gently, and perform five
`counts on 5-mL volumes of the suspension, using the particle
`counter 10-µm size threshold. Obtain the mean cumulative particle
`
`count per mL. Pipet a volume of this suspension containing 250 to
`500 particles into a filter funnel prepared as described for Filtration
`Apparatus under Microscopic Particle Count Test. After drying the
`membrane, count the total number of standard spheres collected on
`the membrane filter. This count should be within 20% of the mean
`instrumental count per mL for the suspension.
`
`Test Environment
`Perform the test in an environment that does not contribute any
`significant amount of particulate matter. Specimens must be cleaned
`to the extent that any level of extraneous particles added has a
`negligible effect on the outcome of the test. Preferably, the test
`specimen, glassware, closures, and other required equipment are
`prepared in an environment protected by high-efficiency particulate
`air (HEPA) filters, and nonshedding garments and powder-free
`gloves are worn throughout the preparation of samples.
`Cleanse glassware, closures, and other required equipment,
`preferably by immersing and scrubbing in warm, nonionic detergent
`solution. Rinse in flowing tap water, and then rinse again in flowing
`filtered distilled or deionized water. Organic solvents may also be
`used to facilitate cleaning. [NOTE-These steps describe one way to
`clean equipment; alternatively, particulate-free equipment may be
`obtained from a suitable vendor.] Finally, rinse the equipment in
`filtered distilled or deionized water, using a hand-held pressure
`nozzle with final filter or other appropriate filtered water source, such
`as distilled or deionized water passed through a capsule filter having
`a 1.2-µm or finer porosity.
`To collect blank counts, use a cleaned vessel of the type and
`volume representative of that to be used in the test. Place a 50-mL
`volume of filtered distilled or deionized water in the vessel, and
`agitate the water sample in the cleaned glassware by inversion or
`swirling. [NOTE-A smaller volume, consistent with the article to be
`counted, can be used.] Degas by sonicating (at 80 to 120 watts) for
`about 30 seconds or by allowing to stand. Swirl the vessel containing
`the water sample by hand or agitate by mechanical means to suspend
`particles. Withdraw and obtain the particle counts for three
`consecutive samples of not less than 5 mL each, disregarding the
`first count. If more than 10 particles of 10-µm or greater size, or
`more than 2 particles of 25-µm or greater size are observed in the
`combined 10-mL sample, the environment is not suitable for
`particulate analysis: the filtered distilled or deionized water and
`glassware have not been properly prepared or the counter is
`generating spurious counts. In this case, repeat the preparatory steps
`until conditions of analysis are suitable for the test.
`
`Test Procedure
`
`TEST PREPARATION
`
`Prepare the test specimens in the following sequence. Outside of
`the laminar enclosure, remove outer closures, sealing bands, and any
`loose or shedding paper labels. Rinse the exteriors of the containers
`with filtered distilled or deionized water as directed under Test
`Environment. Protect the containers from environmental contamina(cid:173)
`tion until analyzed. Withdraw the contents of the containers under
`test in a manner least likely to generate particles that could enter the
`sample. Contents of containers with removable stoppers may be
`withdrawn directly by removing the closures. Sampling devices
`having a needle to penetrate the unit closure may also be employed.
`Products packaged in flexible plastic containers may be sampled by
`cutting the medication or administration port tube or a comer from
`the unit with a suitably cleaned razor blade or scissors.
`Dry or lyophilized products may be constituted either by
`removing the closure to add diluent or by injecting diluent with a
`hypodermic syringe having a 1.2-µm or finer syringe filter. If test
`specimens are to be pooled, remove the closure and empty the
`contents into a clean container.
`The number of test specimens must be adequate to provide a
`statistically sound assessment of whether a batch or other large group
`of units represented by the test specimens meets or exceeds the
`limits. If the volume in the container is less than 25 mL, test a
`solution pool of 10 or more units. Single small-volume injection
`units may be tested if the individual unit volume is 25 mL or more.
`
`6 of 22
`
`Fresenius Kabi
`Exhibit 1028
`
`

`

`324
`
`(788) Particulate Matter in Injections / Physical Tests
`
`USP 30 USP 30
`
`For large-volume injections, individual units are tested. For large(cid:173)
`volume injections or for small-volume injections where the
`individual unit volume is 25 mL or more, fewer than 10 units may
`be tested, based on the definition of an appropriate sampling plan.
`
`per mL would be multiplied by 10 to obtain the test result based on
`the 10-mL maximum dose. [NOTE-For the calculations of test
`results, consider this maximum dose portion to be the equivalent of
`the contents of one full container.]
`
`PRODUCT DETERMINATION
`
`Calculations
`
`Depending upon the dosage form being tested, proceed as directed
`under the appropriate category below.
`Liquid Preparations-
`Volume in Container Less than 25 ml-Prepare the containers a

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket