throbber
www.uspto.gov
`
`UNITED STATES DEPARTMENT OF COMMERCE
`United States Patent and Trademark Office
`Address: COMMISSIONER FOR PATENTS
`P.O. Box 1450
`Alexandria, Virginia 22313-1450
`
`16/918,682
`
`07/01/2020
`
`Christopher Brian LOCKE
`
`P001182US04CON
`
`5599
`
`60402
`
`7590
`
`08/27/2024
`
`KINETIC CONCEPTS, INC.
`c/o Harness Dickey & Pierce
`5445 Corporate Drive
`Suite 20
`Troy, MI 48098
`
`EXAMINER
`
`BASET, NASHEHA
`
`Para NONE
`
`3771
`
`NOTIFICATION DATE
`
`DELIVERY MODE
`
`08/27/2024
`
`ELECTRONIC
`
`Please find below and/or attached an Office communication concerning this application or proceeding.
`
`The time period for reply, if any, is set in the attached communication.
`
`Notice of the Office communication was sent electronically on above-indicated "Notification Date" to the
`following e-mail address(es):
`
`IPDocketing @ Solventum.com
`dgodzisz@hdp.com
`troymailroom @hdp.com
`
`PTOL-90A (Rev. 04/07)
`
`

`

`
`
`Disposition of Claims*
`1-28 is/are pending in the application.
`)
`Claim(s)
`5a) Of the above claim(s) _ is/are withdrawn from consideration.
`C} Claim(s)__ is/are allowed.
`Claim(s) 1-28is/are rejected.
`(] Claim(s)__ is/are objectedto.
`C] Claim(s
`are subjectto restriction and/or election requirement
`)
`* If any claims have been determined allowable, you maybeeligible to benefit from the Patent Prosecution Highway program at a
`participating intellectual property office for the corresponding application. For more information, please see
`http://www.uspto.gov/patents/init_events/pph/index.jsp or send an inquiry to PPHfeedback@uspto.gov.
`
`) ) ) )
`
`Application Papers
`10)¥] The specification is objected to by the Examiner.
`11){¥ The drawing(s) filed on July 1, 2020 is/are: a) accepted or b)(j objected to by the Examiner.
`Applicant may not request that any objection to the drawing(s) be held in abeyance. See 37 CFR 1.85(a).
`Replacement drawing sheet(s) including the correction is required if the drawing(s) is objected to. See 37 CFR 1.121(d).
`
`Priority under 35 U.S.C. § 119
`12)£) Acknowledgment is made of a claim for foreign priority under 35 U.S.C. § 119(a)-(d)or (f).
`Certified copies:
`_—_c)L) None ofthe:
`b)L) Some**
`a)Q) All
`1.1) Certified copies of the priority documents have been received.
`2.1.) Certified copies of the priority documents have been received in Application No. |
`3.2.) Copies of the certified copies of the priority documents have been receivedin this National Stage
`application from the International Bureau (PCT Rule 17.2(a)).
`*“ See the attached detailed Office action for a list of the certified copies not received.
`
`Attachment(s)
`
`1)
`
`Notice of References Cited (PTO-892)
`
`Information Disclosure Statement(s) (PTO/SB/08a and/or PTO/SB/08b)
`2)
`Paper No(s)/Mail Date
`U.S. Patent and Trademark Office
`
`3)
`
`4)
`
`(LJ Interview Summary (PTO-413)
`Paper No(s)/Mail Date
`(Qj Other:
`
`PTOL-326 (Rev. 11-13)
`
`Office Action Summary
`
`Part of Paper No./Mail Date 20240722
`
`Application No.
`Applicant(s)
`16/918,682
`LOCKE et al.
`
`Office Action Summary Art Unit|AIA (FITF)StatusExaminer
`NASHEHA BASET
`3771
`Yes
`
`
`
`-- The MAILING DATEof this communication appears on the cover sheet with the correspondence address --
`Period for Reply
`
`A SHORTENED STATUTORYPERIOD FOR REPLYIS SET TO EXPIRE 3 MONTHS FROM THE MAILING
`DATE OF THIS COMMUNICATION.
`Extensionsof time may be available underthe provisions of 37 CFR 1.136(a). In no event, however, may a reply betimely filed after SIX (6) MONTHSfrom the mailing
`date of this communication.
`If NO period for reply is specified above, the maximum statutory period will apply and will expire SIX (6) MONTHSfrom the mailing date of this communication.
`-
`- Failure to reply within the set or extended period for reply will, by statute, cause the application to become ABANDONED (35 U.S.C. § 133).
`Any reply received by the Office later than three months after the mailing date of this communication, evenif timely filed, may reduce any earned patent term
`adjustment. See 37 CFR 1.704(b).
`
`Status
`
`1) Responsive to communication(s) filed on June 13, 2024.
`C} A declaration(s)/affidavit(s) under 37 CFR 1.130(b) was/werefiled on
`
`2a)() This action is FINAL. 2b)¥)This action is non-final.
`3) An election was madeby the applicant in responseto a restriction requirement set forth during the interview
`on
`; the restriction requirement and election have been incorporated into this action.
`4)(2) Since this application is in condition for allowance except for formal matters, prosecution as to the merits is
`closed in accordance with the practice under Exparte Quayle, 1935 C.D. 11, 453 O.G. 213.
`
`

`

`Application/Control Number: 16/918,682
`Art Unit: 3771
`
`Page 2
`
`DETAILED ACTION
`
`Notice of Pre-AlA or AIA Status
`
`The present application, filed on or after March 16, 2013, is being examined under the first
`
`inventor to file provisions of the AIA.
`
`Continued Examination Under 37 CFR 1.114
`
`A request for continued examination under 37 CFR 1.114, including the fee set forth in 37 CFR
`
`1.17(e), was filed in this application after final rejection. Since this application is eligible for continued
`
`examination under 37 CFR 1.114, and the fee set forth in 37 CFR 1.17(e) has been timely paid, the
`
`finality of the previous Office action has been withdrawn pursuant to 37 CFR 1.114. Applicant's
`
`submission filed on June 13, 2024 has been entered.
`
`Response to AmendmentandStatus of the Claims
`
`This office action is in response to the remarksfiled on June 13, 2024.
`
`In response to the amendmentfiled June 13, 2024, claim 28 is new,claims 1-27 are pending.
`
`Examiner notes that the double patenting rejection of the office action dated 09/21/2023 should not be
`
`held in abeyance.
`
`Specification
`
`The specification is objected to asfailing to provide proper antecedent basis for the claimed
`
`subject matter. See 37 CFR 1.75(d)(1) and MPEP § 608.01(0). Correction of the following is required:
`
`Claim 7 lines 7-8 recites “wherein the perforation shape factor of the hole is less than about 1. Although
`
`values supporting more than about 1 is present in paragraph 0062, values of less than about 1, or
`
`regarding this language, for the perforation shape factor of the hole is not disclosed in the instant
`
`

`

`Application/Control Number: 16/918,682
`Art Unit: 3771
`
`Page 3
`
`specifications. The specification lacks antecedentbasis for the claimed limitation. The word “about”
`
`broadens the interpretation for how any value below1is interpreted. It can be seen that there is a
`
`range for more than about 1 is present in paragraph 0062 in the specifications, but not for “less than
`
`about 1”. For examination purposes, the examiner is interpreting this as any value less than 1 .
`
`Claim Rejections - 35 USC § 112
`
`The following is a quotation of 35 U.S.C. 112(b):
`(b) CONCLUSION.—The specification shall conclude with one or more claims particularly pointing out
`and distinctly claiming the subject matter which the inventor or a joint inventor regards as the
`invention.
`
`The following is a quotation of 35 U.S.C. 112 (pre-AlA), second paragraph:
`The specification shall conclude with one or more claims particularly pointing out and distinctly
`claiming the subject matter which the applicant regards as his invention.
`
`Claims 2-8 are rejected under 35 U.S.C. 112(b) or 35 U.S.C. 112 (pre-AlA), second paragraph, as
`
`being indefinite for failing to particularly point out and distinctly claim the subject matter which the
`
`inventor or a joint inventor (or for applications subject to pre-AlA 35 U.S.C. 112, the applicant), regards
`
`as the invention.
`
`Claim 2 recites “wherein the plurality of holes have a perforation shape factor and a strut angle
`
`configured to cause the plurality of holes to collapse in a direction substantially perpendicular to the
`
`opening”. The terms “perforation shape factor” and “strut angle” are not standard terms in the art and
`
`have not been defined by the disclosure, but rather merely exemplified. Therefore, the metes and
`
`bounds of the claim as written are unclear. For example, the specification paragraph 0062 states that
`
`“the perforation shape factor may represent an orientation of the hole 128 relative to the first
`
`orientation line 127 and the second orientation line 129. Generally, the perforation shape factor is a
`
`ratio of a ratio of % a maximum length of the hole 128 that is parallel to the second orientation line to %
`
`a maximum length of the hole 128 that is parallel to the first orientation line 127”.
`
`It is noted that this
`
`definition is based on generalities and examples. Further, the examples and generalities given are based
`
`

`

`Application/Control Number: 16/918,682
`Art Unit: 3771
`
`Page 4
`
`on first orientation line 127 and second orientation line 129.
`
`In the specification paragraph 0060 various
`
`examples of what the orientation lines are also given. As the claim does not clearly identify if the shape
`
`perforation factor claimedis the same as exemplified in paragraph 0062, nor doesit clearly identify
`
`which orientation lines are used for its definition, the scope of the claim is unclear.
`
`It is noted that the
`
`term “strut angle”, also recited in claim 2 is also unclear for substantially the same reasons. The term is
`
`discussedin the specification paragraph 0063 and is exemplified as being relative to the first orientation
`
`line. For the purpose of examination, the term will be considered as encompassing at least one distance
`
`from the center of a hole towards the perimeter in a ratio to another distance from the center of the
`
`hole to the perimeter as this is consistent with the disclosure (see for example Paragraphs 0060 and
`
`0062, and Figure 3, line 144 and 146). The “strut angle” will be considered as a line connecting the
`
`center of adjacent rows, relative to any orientation line, as this is consistent with the disclosure
`
`(paragraphs 0060 and 0063).
`
`It is noted that although claims 3 and 6 appear to define one term or the
`
`other, both terms have not been defined in one claimed limitation.
`
`Claim Rejections - 35 USC § 102
`
`In the event the determination of the status of the application as subject to AIA 35 U.S.C. 102
`
`and 103 (or as subject to pre-AlA 35 U.S.C. 102 and 103) is incorrect, any correction of the statutory
`
`basis (i.e., changing from AIA to pre-AlA) for the rejection will not be considered a new ground of
`
`rejection if the prior art relied upon, and the rationale supporting the rejection, would be the same
`
`under either status.
`
`The following is a quotation of the appropriate paragraphs of 35 U.S.C. 102 that form the basis
`
`for the rejections under this section made in this Office action:
`
`A person shall be entitled to a patent unless —
`
`(a)(1) the claimed invention was patented, described in a printed publication, or in public use, on sale,
`or otherwise available to the public before the effective filing date of the claimed invention.
`
`

`

`Application/Control Number: 16/918,682
`Art Unit: 3771
`
`Page 5
`
`(a)(2) the claimed invention was described in a patent issued under section 151, or in an application
`for patent published or deemed published under section 122(b), in which the patent or application, as
`the case may be, namesanother inventor and was effectively filed before the effective filing date of
`the claimed invention.
`
`Note: The term “tissue site” has been defined in the instant specification, paragraph 0035, as
`
`referring to a wound or a defect located on or within tissue; and may also refer to areas of any tissue
`
`that are not necessarily wounded or defective but are instead areas which it may be desirable to add or
`
`promote growth of additional tissue.
`
`Claims 1-6, 9, 11, 14, 17-18, 21 and 28 is/are rejected under 35 U.S.C. 102(a){1) and 35 U.S.C.
`
`102(a)(2) as being anticipated by Dunn et al (WO 2014014922 A1), herein after Dunn.
`
`It is noted that Dunn teaches several different embodiments which anticipate the claims.
`
`In a first embodiment, system 100, paragraph 0081-0086, Figure 1, Dunn anticipatesclaims1,
`
`9, 11, and 28 as follows:
`
`Claim 1. Dunn teachesa dressing for closing an opening through a surface ofa tissue site, the
`
`dressing comprising: a tissue interface (system 100, comprising wound packer 102 figure 1 paragraph
`
`0081), comprising: a contracting layer (stabilizing structure or wound packer 102 described as
`
`comprising porous material such as foam and may comprise one or more embodimentsof the wound
`
`closure device, figure 1 paragraphs 0081, 0083), the contracting layer comprising a plurality of holes
`
`(porous material such as foam or polymers, see paragraph 0081) extending through the contracting
`
`layer to form a void space (interior space of the wound packer as the pores would create the void
`
`space), and a protective layer configured to be positioned between the contracting layer and at least a
`
`portion of the surface of the tissue site surrounding the opening (organ protection layer paragraph
`
`
`0083, “...a non- or minimally-adherent organ protection layer (notillustrated) may be applied over any
`
`
`exposedviscera.” Being placed over any exposed viscera is considered to be a portion of a surface of
`
`tissue surrounding the openingin the broadest reasonable interpretation of the word surface in
`
`

`

`Application/Control Number: 16/918,682
`Art Unit: 3771
`
`Page 6
`
`application to surrounding the opening, if the openingis the viscera and viscerais tissue. If the organ
`
`protection layer is placed over the wound,it is contacting surface tissue surrounding the opening and
`
`not within the wounditself as the wound packer would beinserted within the wound and sealed with
`
`the drape, see paragraph 0083); and a sealing drape (drape 104; paragraph 0081 and 0083) configured
`
`to coverthe tissue interface to form a sealed space (paragraphs 0081 and 0083). It is noted that the
`
`claimed limitation in the preamble “for closing an opening through a surfaceofa tissue site” is intended
`
`use and thus is only considered as it breatheslife and meaning into the claims, and would only require a
`
`product capable of performing the claimed use. As discussed above, the product of Dunn would be
`
`capable of performing the use as claimed, and thus the product of Dunn encompassesthe product as
`
`claimed.
`
`Claim 9. Dunn teachesthe device as taughtin the rejection of claim 1 above. Dunn continues to
`
`teach wherein the plurality of holes are formed in two or moreparallel rows(as in the perimeter region,
`
`one row would be the wound packer and another parallel row to this would be the perimeter or on
`
`top of any wound closure device, creating two parallel rowsof plurality of holes; paragraph 0081).
`
`Claim 11. Dunn teaches the device as taughtin the rejection of claim 1 above. Dunn teaches
`
`wherein the contracting layer is formed from a foam material having a firmness factor (the foam used
`
`by Dunn inherently hasa firmnessfactor as it collapses when negative pressure is applied; paragraph
`
`0085), wherein the firmnessfactor is a ratio of density of the foam material in a compressed state to a
`
`density of the foam material in an uncompressed state (as this would be calculated as the ratio of the
`
`density of the foam material used by Dunn in a compressedstate to a density of the foam material in
`
`used by Dunn in an uncompressedstate, as all foam would have an inherent density depending on the
`
`compression factors, as Dunn recites the application of multiple levels of pressure applied to the foam
`
`in paragraph 0085).
`
`

`

`Application/Control Number: 16/918,682
`Art Unit: 3771
`
`Page 7
`
`Claim 28. Dunn teaches the device as taughtin the rejection of claim 1 above. Dunn continues
`
`to teach wherein the protective layer is configured to be positioned over the opening (organ protection
`
`layer paragraphs 0083, “...a non- or minimally-adherentorgan protection layer (notillustrated) may
`
`be applied overany exposed viscera.”; as the instant invention and claims are regarding an apparatus
`
`and not a methodofuse, one of ordinary skill could position the protective layer over the opening).
`
`In a second embodiment, system of the structure 1701, Figures 3A-3D, paragraphs 0087-0092,
`
`Dunn anticipates claims 1-6, 9, 14, and 28 as follows:
`
`Claim 1. Dunn teachesa dressing for closing an opening through a surface ofa tissue site, the
`
`dressing comprising: a tissue interface (stabilizing structure 1701 comprising the porous material such
`
`as foam of paragraph 0092), comprising: a contracting layer (stabilizing structure 1701 as it is capable
`
`of collapsing; paragraph 0087) the contracting layer comprising a plurality of holes extending through
`
`the contracting layer to form a void space ( as the holes would bethe voids of space within the beams
`
`that create the stabilizing structure 1701, and the plurality of holes can be seen to be present
`
`extending through thelayer in figures 3A-3D; paragraph 0087), and a protective layer (“organ
`
`protection layer” paragraph 0092) configured to be positioned between the contracting layer and at
`
`least a portion of the surface of the tissue site surrounding the opening (as one could place the organ
`
`protection layer between the tissue, as it states that it may be placed into contact with at least the
`
`bottom portion of the wound,and the contracting layer; paragraph 0092); and a sealing drape
`
`(“drape” paragraph 0092) configured to cover the tissue interface to form a sealed space (as it may be
`
`sealed over the skin surrounding the woundandeffectuate woundclosure; paragraph 0092).
`
`Claim 2. Dunn teaches the device as taughtin the rejection of claim 1 above. Dunn teaches
`
`wherein the plurality of holes (see cells or holes in between the stabilizing structures of 1701 created
`
`by the struts 1703 and 1705in figure 3A-3B; paragraph 0087) have a perforation shape factor and a
`
`

`

`Application/Control Number: 16/918,682
`Art Unit: 3771
`
`Page 8
`
`strut angle (all holes inherently have a perforation shape factor and a strut angle which would be
`
`created by the struts 1703 and 1705 and multiple rowsof this embodiment, but Dunn discloses as the
`
`struts meet, they maybein right angles to each otherto create planar support structures 1702, orin
`
`other configurations such as honeycomb, and multiple planar support structures 1702 maybe joined
`
`together to form thestabilizing structure; paragraphs 0087 and 0092) configured to causethe plurality
`
`of holes to collapse in a direction substantially perpendicular to the opening (as the structure’s holes’
`
`openings can be from anydirection, any direction, best seen in figure 3A-3D,it could collapse in a
`
`direction perpendicular to the horizontal openings; see figures 3A-3D and paragraphs 0087 and 0092),
`
`and wherein the contracting layer generates a closing force substantially parallel to the surface of the
`
`tissue site to close the opening in response to an application of negative pressure (as the negative
`
`pressureis applied, the closing force would be parallel as it would collapse inward, causing the tissue
`
`of the wound to cometogetherin a parallel direction and close the opening, see paragraphs 0087,
`
`0088, and 0092 regarding direction of the collapse and the contracting layer, porous foam,
`
`surroundingthetissuesite).
`
`Claim 3. Dunn teaches the device as taughtin the rejection of claim 2 above. Dunn continues to
`
`teach wherein the strut angle (see claim 2, as Dunn’s strut angle is interpreted in accordance with the
`
`specifications paragraph 0063) is an angle (see interpreted “strut angle” in the annotated figure
`
`below) between a line (see interpreted “first line” in the annotated figure below for a diagonal line
`
`created by the holes diagonal from one another in adjacent rows OR “second line” for holes
`
`orientated directly above one another in adjacent rows) connecting centers of holes (represented as
`
`the exaggerated black dots in the centersof the holes in the annotatedfigure below) in adjacent rows
`
`and an orientation line (see interpreted “orientation line” in the annotated figure below)
`
`perpendicular to a direction of the closing force (see interpreted “direction of closing force” in the
`
`annotatedfigure below,as the structure 1701 would compress so that the planar support structures
`
`

`

`Application/Control Number: 16/918,682
`Art Unit: 3771
`
`Page 9
`
`1702 come closer to one another, paragraphs 0087 and 0092)(as the struts would meetat right
`
`angles, and other configurations, and two or more support structures which are planes of 2D holes
`
`create 3D holes when they are joined together adjacently, see figure 3A for example, using the beams
`
`at the angles wherein the holes would be perpendicular, see figure 3D, when the direction of the
`
`closing force is on top and arrangedto collapse downward, the orientation line is interpreted as the
`
`axis that would be interpreted as going horizontally through the device as that would create the
`
`perpendicular angle to the closing force best seen by the general orientation ofthe struts in figure 3D;
`
`see paragraphs 0087 and 0092 regarding arrangementofstruts, planes, and collapse when closing
`
`force is applied).
`
`Line trut Angle of the
`
`
`bottom hele and
`
`
`“the diagonal hole
`
`
`:
`
`
`
`Sing Angie of
`the bottom
`hole and the
`
`
`
`hole directly
`above it
`
`Orientation
`
`Claim 4. Dunn teachesthe device as taughtin the rejection of claim 3 above. Dunn continues to
`
`teach wherein the strut angle is about 90 degrees (best seen in annotated figure 3D above,as the strut
`
`

`

`Application/Control Number: 16/918,682
`Art Unit: 3771
`
`Page 10
`
`angle is created between the interpreted “second line” and the “orientation line”, as the angle
`
`created by these lines can be seen to be about 90 degrees).
`
`Claim 5. Dunn teachesthe device as taughtin the rejection of claim 3 above. Dunn continues to
`
`teach wherein the strut angle is less than about 90 degrees (best seen in annotatedfigure 3D above, as
`
`the strut angle is created between the interpreted “first line” and the “orientation line”, as the angle
`
`created by these lines can be seen to beless than 90 degrees).
`
`Claim 6. Dunn teaches the device as taughtin the rejection of claim 2 above. Dunn continues to
`
`teach wherein the perforation shape factor (interpreted as how the hole is shaped of Dunn, see the
`
`limitation regardingthe ratio thatis following, see figure 3D below) of a hole of the plurality of holes is
`
`a ratio of a first line segment to a second line segment (interpreted as the ratio between the twoline
`
`segmentsofthe hole seen in figure 3D below), the first line segment extending from a center of the
`
`hole to a perimeter of the hole in a direction perpendicular to a direction of the closing force (see
`
`Interpreted direction of the Closing Force and Interpreted First Line Segment, as they are arranged ina
`
`perpendicular manner in annotated figure 3D below) and the second line segment extending from a
`
`center of the hole to a perimeter of the hole in a direction parallel to the direction of the closing force
`
`(see Interpreted Direction of the Closing Force and Interpreted Second Line Segment, as they are
`
`arrangedin a perpendicular manner in annotated figure 3D below) (See paragraphs 0087 regarding
`
`the struts 1703 and 1705; please note that the word “segment”is presentin “line segment”, therefore
`
`boththe first and the second line segments are segmentsof the greater lines of the holes, see zoomed
`
`and annotatedfigure 3D below).
`
`

`

`Application/Control Number: 16/918,682
`Art Unit: 3771
`
`Page 11
`
`
`
`interpreted First
`
`
`interpreted
`Line Segment
`Direction of the
`
`
`seat at ae
`
`ne tat a
`Closing Farce
`
`
` Seeece aoe erelee agen nr ee
`
`
`Reta
`nescenceee
`ana
`Serene Ate
`
`
`
`
`
`interpreted
`Second Line
`
`segment
`
`melee
`
`eat
`
`araey
`neineee
`
`
`
`Fig. 3D Zoomed
`and Annotated ents ¥
`epee ne
`
`roe
`. ¥
`
`Claim 9. Dunn teachesthe device as taughtin the rejection of claim 1 above. Dunn continues to
`
`teach wherein the plurality of holes are formed in two or moreparallel rows (best seen in the full figure
`
`3D, as two or moreparallel rowsof the plurality of holes can be seen goingin the x, y, and z axis).
`
`Claim 14. Dunn teaches the device as taught in the rejection of claim 1 above. Dunn teaches
`
`wherein a shapeof each hole (see cells or holes in between the stabilizing structures of 1701 created
`
`by the struts 1703 and 1705in figure 3A-3B; paragraph 0087)of the plurality of holes is hexagonal
`
`(paragraph 0087 states honeycomb configurations are possible; merriam-webster.com defines
`
`honeycomb as “a mass of hexagonal waxcells built by honeybeesin their nest to contain their brood
`
`and stores of honey; something that resembles a honeycomb in structure or appearance especially: a
`
`strong lightweight cellular structural material” therefore honeycomb configurations of the holes
`
`created by the beams would be hexagonal).
`
`Claim 28. Dunn teaches the device as taughtin the rejection of claim 1 above. Dunn continues
`
`to teach wherein the protective layer is configured to be positioned over the opening(as the instant
`
`

`

`Application/Control Number: 16/918,682
`Art Unit: 3771
`
`Page 12
`
`invention and claims are regarding an apparatus and not a methodof use, oneofordinary skill could
`
`position the protective layer over the opening; as one could place the organ protection layer between
`
`the tissue, as it states that it may be placed into contact with at least the bottom portion of the
`
`wound,and the contracting layer; paragraph 0092).
`
`In a third embodiment, system of the structure 1101, Figures 8A-8E, paragraphs 0098-0108,
`
`Dunn anticipates claims 1-6, 9, 17-18, 21, and 28 as follows:
`
`Claim 1. Dunn teachesa dressing for closing an opening through a surface ofa tissue site, the
`
`dressing comprising: a tissue interface (stabilizing structure 1100 comprising porous material such as
`
`“felted” reticulated foam; paragraph 0098-0099), comprising: a contracting layer (stabilizing structure
`
`1100 asit is configured to collapse; paragraph 0098), the contracting layer comprising a plurality of
`
`holes extending through the contracting layer to form a void space (as the holes would bethe voids of
`
`space within the strips 1102 and 1104 thatcreate the stabilizing structure 1100, and the plurality of
`
`holes can be seen to be present extending throughthe layer best in figure 8C-8D; paragraphs 0099-
`
`0101), and a protective layer (“organ protection layer” paragraph 0107) configured to be positioned
`
`between the contracting layer and at least a portion of the surface of the tissue site surrounding the
`
`opening (as one could place the organ protection layer between thetissue,as it states that it may be
`
`placed into contact with at least the bottom portion of the wound, and the contracting layer;
`
`paragraph 0107); and a sealing drape configured to cover the tissue interface to form a sealed space
`
`(drape, as the drapeis disclosed to be covering the wound andissue interface in order for the
`
`pressure to be exerted onit; paragraph 0099).
`
`Claim 2. Dunn teachesthe device as taughtin the rejection of claim 1 above. Dunn teaches
`
`wherein the plurality of holes (see cells or holes in between the stabilizing structure 1101 created by
`
`the strips 1102 and 1104 bestin figure 8C-8D; paragraphs 0099-0101) have a perforation shape factor
`
`

`

`Application/Control Number: 16/918,682
`Art Unit: 3771
`
`Page 13
`
`and a strut angle (all holes inherently have a perforation shape factor and a strut angle which would
`
`be created by each ofthe strip portions 1102 and 1104 of the respective holes and multiple rows of
`
`this embodiment, but Dunn disclosesas the strips meet at the notches 1106 and 1108, they may be in
`
`right angles to each other to create planar support structures 1101; paragraphs 0099-0102) configured
`
`to cause the plurality of holes to collapse in a direction substantially perpendicular to the opening
`
`(parallel collapse against the tissue and perpendicular collapse to the opening of this embodimentis
`
`disclosed in paragraph 0105), and wherein the contracting layer generates a closing force substantially
`
`parallel to the surface of the tissue site to close the opening in response to an application of negative
`
`pressure (as the negative pressure Is applied, the closing force would be parallel in order for the
`
`collapse to be oriented to go in a direction generally parallel with the orientation of the collagen and
`
`other fibrous tissue, causing the tissue of the wound to come together in a parallel direction and close
`
`the opening, see paragraphs 0103-0105 regardingdirection of the collapse and thestabilizing
`
`structure, porous foam, surroundingthe tissue site).
`
`Claim 3. Dunn teaches the device as taughtin the rejection of claim 2 above. Dunn continues to
`
`teach wherein the strut angle (see claim 2, as Dunn’s strut angle is interpreted in accordance with the
`
`specifications paragraph 0063)is an angle (see interpreted “strut angle” in the annotated figure
`
`below) between a line (see interpreted “first line” in the annotated figure below for a diagonal line
`
`created by the holes adjacent from one another in adjacent rows OR “second line” for holes
`
`orientated directly above one another in adjacent rows) connecting centers of holes (represented as
`
`the exaggerated black dots in the centersof the holesin the annotated figure below) in adjacent rows
`
`(the first line is created by the dot directly in the adjacent row to the bottom most dot, and the
`
`secondline is created by the dot diagonally in the adjacent row to the bottom mostdot; see
`
`annotatedfigure below) and an orientation line (see interpreted “orientation line” in the annotated
`
`figure below) perpendicular to a direction of the closing force (see interpreted “direction of closing
`
`

`

`Application/Control Number: 16/918,682
`Art Unit: 3771
`
`Page 14
`
`force” in the annotated figure below, as the structure 1101 would compress effecting planar support
`
`structures 1102 and 1104, paragraphs 0099-0102) (as the strips would meetat right angles, and the 2D
`
`strips 1102 and 1104 would create 3D holes when they are joined together, see figure 8A and 8Dfor
`
`example, the segmentsofthe strips 1102 and 1104 would be perpendicular where they meet, see
`
`figure 8A and 8C-8D, when the direction of the closing force is on top and arrangedto collapse
`
`downward,parallel to the collagen and fibrous tissue, paragraph 0105,the orientationline is
`
`interpreted as the axis that would be interpreted as going horizontally through the device as that
`
`would create the perpendicular angle to the closing force best seen by the general orientation of the
`
`struts in figures 8A and 8D; see paragraphs 0099-0102 regarding arrangementof struts, planes, and
`
`collapse when closing force is applied).
`
`1
`I
`
`III
`
`—
`Directian of
`Closing Farce
`
`First Line
`
`
`Line
`
`m Second
`
`
`
`
`
`
`FS, SA
`
`
`Strut Angle of
`the bottom
`hole and the
`hole directly
`above it
`
`
`
`
`
`
`Claim 4. Dunn teachesthe device as taughtin the rejection of claim 3 above. Dunn continues to
`
`teach wherein the strut/strip angle is about 90 degrees (best seen in annotated figure 8A above, as the
`
`strut angle is created between the interpreted “first line” and the “orientation line”, as the angle
`
`created by these lines can be seen to be about 90 degrees).
`
`

`

`Application/Control Number: 16/918,682
`Art Unit: 3771
`
`Page 15
`
`Claim 5. Dunn teachesthe device as taughtin the rejection of claim 3 above. Dunn continues to
`
`teach wherein the strut angle is less than about 90 degrees (best seen in annotatedfigure 8A above, as
`
`the strut angle is created between the interpreted “second line” and the “orientation line”, as the
`
`angle created by these lines can be seen to beless than 90 degrees).
`
`Claim 6. Dunn teachesthe device as taughtin the rejection of claim 2 above. Dunn continues to
`
`teach wherein the perforation shape factor (interpreted as how the hole is shaped of Dunn, see the
`
`limitation regardingthe ratio thatis following, see figure 8D below) of a hole of the plurality of holes is
`
`a ratio of a first line segment to a second line segment (interpreted as the ratio between the twoline
`
`segmentsofthe hole seen in figure 8D below), the first line segment extending from a center of the
`
`hole to a perimeter of the hole in a direction perpendicular to a direction of the closing force (see
`
`Interpreted direction of the Closing Force and Interpreted First Line Segment, as they are arranged in a
`
`perpendicular manner in annotated figure 8D below) and the second line segment extending from a
`
`center of the hole to a perimeter of the hole in a direction parallel to the direction of the closing force
`
`(see Interpreted Direction of the Closing Force and Interpreted Second Line Segment,as they are
`
`arranged in a perpendicular manner in annotated figure 8D below) (See paragraphs 0087 regarding
`
`the arrangementand anglesofstrips 1102 and 1104; please note that the word “segment”is present
`
`in “line segment”, therefore both the first and the second line s

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket