throbber
www.uspto.gov
`
`UNITED STATES DEPARTMENT OF COMMERCE
`United States Patent and Trademark Office
`Address: COMMISSIONER FOR PATENTS
`P.O. Box 1450
`Alexandria, Virginia 22313-1450
`
`17/281,668
`
`03/31/2021
`
`Christopher Brian LOCKE
`
`PO01658US02PCT
`
`1000
`
`60402
`
`7590
`
`01/04/2024
`
`KINETIC CONCEPTS, INC.
`c/o Harness Dickey & Pierce
`5445 Corporate Drive
`Suite 20
`Troy, MI 48098
`
`EXAMINER
`
`FLYNN, TIMOTHY LEE
`
`Para NONE
`
`3781
`
`NOTIFICATION DATE
`
`DELIVERY MODE
`
`01/04/2024
`
`ELECTRONIC
`
`Please find below and/or attached an Office communication concerning this application or proceeding.
`
`The time period for reply, if any, is set in the attached communication.
`
`Notice of the Office communication was sent electronically on above-indicated "Notification Date" to the
`following e-mail address(es):
`
`dgodzisz@hdp.com
`troymailroom @hdp.com
`
`PTOL-90A (Rev. 04/07)
`
`

`

`
`
`Disposition of Claims*
`1-5,7-8,10-12,14-20,23-24 and 26 is/are pending in the application.
`)
`Claim(s)
`5a) Of the above claim(s) _ is/are withdrawn from consideration.
`CL] Claim(s)__is/are allowed.
`Claim(s) 1-5,7-8,10-12, 14-20,23-24 and 26 is/are rejected.
`(] Claim(s)__ is/are objectedto.
`C] Claim(s
`are subjectto restriction and/or election requirement
`)
`* If any claims have been determined allowable, you maybeeligible to benefit from the Patent Prosecution Highway program at a
`participating intellectual property office for the corresponding application. For more information, please see
`http://www.uspto.gov/patents/init_events/pph/index.jsp or send an inquiry to PPHfeedback@uspto.gov.
`
`) ) ) )
`
`Application Papers
`10) The specification is objected to by the Examiner.
`11)0) The drawing(s) filedon__ is/are: a)(J accepted or b)( objected to by the Examiner.
`Applicant may not request that any objection to the drawing(s) be held in abeyance. See 37 CFR 1.85(a).
`Replacement drawing sheet(s) including the correction is required if the drawing(s) is objected to. See 37 CFR 1.121(d).
`
`Priority under 35 U.S.C. § 119
`12)7) Acknowledgment is made of a claim for foreign priority under 35 U.S.C. § 119(a)-(d)or (f).
`Certified copies:
`c)Z None ofthe:
`b)() Some**
`a)C All
`1.1.) Certified copies of the priority documents have been received.
`2.2) Certified copies of the priority documents have been received in Application No.
`3.1.) Copies of the certified copies of the priority documents have been receivedin this National Stage
`application from the International Bureau (PCT Rule 17.2(a)).
`*“ See the attached detailed Office action for a list of the certified copies not received.
`
`Attachment(s)
`
`1)
`
`Notice of References Cited (PTO-892)
`
`2) (J Information Disclosure Statement(s) (PTO/SB/08a and/or PTO/SB/08b)
`Paper No(s)/Mail Date
`U.S. Patent and Trademark Office
`
`3)
`
`4)
`
`(LJ Interview Summary (PTO-413)
`Paper No(s)/Mail Date
`(Qj Other:
`
`PTOL-326 (Rev. 11-13)
`
`Office Action Summary
`
`Part of Paper No./Mail Date 20231227
`
`Application No.
`Applicant(s)
`17/281,668
`LOCKE et al.
`
`Office Action Summary Art Unit|AIA (FITF)StatusExaminer
`TIMOTHY L FLYNN
`3781
`Yes
`
`
`
`-- The MAILING DATEof this communication appears on the cover sheet with the correspondence address --
`Period for Reply
`
`A SHORTENED STATUTORYPERIOD FOR REPLYIS SET TO EXPIRE 3 MONTHS FROM THE MAILING
`DATE OF THIS COMMUNICATION.
`Extensionsof time may be available underthe provisions of 37 CFR 1.136(a). In no event, however, may a reply betimely filed after SIX (6) MONTHSfrom the mailing
`date of this communication.
`If NO period for reply is specified above, the maximum statutory period will apply and will expire SIX (6) MONTHSfrom the mailing date of this communication.
`-
`- Failure to reply within the set or extended period for reply will, by statute, cause the application to become ABANDONED (35 U.S.C. § 133).
`Any reply received by the Office later than three months after the mailing date of this communication, evenif timely filed, may reduce any earned patent term
`adjustment. See 37 CFR 1.704(b).
`
`Status
`
`1) Responsive to communication(s) filed on 07 December 2023.
`C) A declaration(s)/affidavit(s) under 37 CFR 1.130(b) was/werefiled on
`
`2a)() This action is FINAL. 2b)¥)This action is non-final.
`3) An election was madeby the applicant in responseto a restriction requirement set forth during the interview
`on
`; the restriction requirement and election have been incorporated into this action.
`4)(2) Since this application is in condition for allowance except for formal matters, prosecution as to the merits is
`closed in accordance with the practice under Exparte Quayle, 1935 C.D. 11, 453 O.G. 213.
`
`

`

`Application/Control Number: 17/281,668
`Art Unit: 3781
`
`Page 2
`
`DETAILED ACTION
`
`Notice of Pre-AlA or AIA Status
`
`The present application, filed on or after March 16, 2013, is being examined under the first
`
`inventor to file provisions of the AIA.
`
`Continued Examination Under 37 CFR 1.114
`
`A request for continued examination under 37 CFR 1.114, including the fee set forth in 37 CFR
`
`1.17(e), was filed in this application after final rejection. Since this application is eligible for continued
`
`examination under 37 CFR 1.114, and the fee set forth in 37 CFR 1.17(e) has been timely paid, the
`
`finality of the previous Office action has been withdrawn pursuant to 37 CFR 1.114. Applicant's
`
`submission filed on 12/07/2023 has been entered.
`
`Response to Arguments
`
`Applicant’s amendmentsfiled 12/07/2023 have been accepted.
`
`Applicant's argumentsfiled 12/07/2023 have been fully considered but they are not persuasive.
`
`Applicant argues that wound cover 40 of Weston cannotbeinterpreted as having a dual layer structure.
`
`However, Weston §][0055] states that the protrusions 60 may be constructed of a different material
`
`than cover 40, whichclearly implies a dual layer structure, since the layer including the protrusions 60
`
`may be constructed of a different material having its own thickness and pliability than the cover 40
`
`(Weston 4[0055-0056]).
`
`Additionally, Applicant argues that cover 40 cannotbe interpreted as having a dual layer
`
`structure because Weston states in §[0054] that the plurality of protrusions 60 are embeddedin the
`
`cover 40. However, the claim is written broadly enough that the cover 40 and protrusions 60 of Weston
`
`are still capable of being interpreted as having a dual layer structure. Furthermore, Wu in view of
`
`Robinson is relied upon as teaching the dual layer structure, and Weston is merely relied upon to teach
`
`that the layers may havesimilar profiles and dimensions. In light of the combination of
`
`

`

`Application/Control Number: 17/281,668
`Art Unit: 3781
`
`Page 3
`
`Wu/Robinson/Weston,it would be obvious to one of ordinary skill to have dual layers having similar
`
`profiles and dimension in order to simplify manufacturing.
`
`Applicant did not specifically argue the dependentclaims.
`
`Claim Rejections - 35 USC § 103
`
`The following is a quotation of 35 U.S.C. 103 which forms the basis for all obviousness rejections
`
`set forth in this Office action:
`
`A patent for a claimed invention may not be obtained, notwithstanding that the claimed invention is
`not identically disclosed as set forth in section 102, if the differences between the claimed invention
`and the prior art are such that the claimed invention as a whole would have been obvious before the
`effective filing date of the claimed invention to a person having ordinaryskill in the art to which the
`claimed invention pertains. Patentability shall not be negated by the manner in which the invention
`was made.
`
`The factual inquiries for establishing a background for determining obviousness under 35 U.S.C.
`
`103 are summarized as follows:
`
`1. Determining the scope and contentsofthe prior art.
`
`2. Ascertaining the differences between the prior art and the claims at issue.
`
`3. Resolving the level of ordinary skill in the pertinentart.
`
`4. Considering objective evidence present in the application indicating obviousness or
`
`nonobviousness.
`
`This application currently names joint inventors. In considering patentability of the claims the
`
`examiner presumesthat the subject matter of the various claims was commonly owned as of the
`
`effective filing date of the claimed invention(s) absent any evidence to the contrary. Applicant is advised
`
`of the obligation under 37 CFR 1.56 to point out the inventor and effectivefiling dates of each claim that
`
`was not commonly ownedas of the effective filing date of the later invention in order for the examiner
`
`to consider the applicability of 35 U.S.C. 102(b)(2)(C) for any potential 35 U.S.C. 102(a)(2) prior art
`
`against the later invention.
`
`

`

`Application/Control Number: 17/281,668
`Art Unit: 3781
`
`Page 4
`
`Claims 1-5, 8, 10-12, 14-19, and 26 are rejected under 35 U.S.C. 103 as being unpatentable
`
`over Wu (US 20130144230 A1)in view of Robinson (US 20110224633 A1), and further in view of
`
`Weston (US 20040073151 A1).
`
`Regarding Claim 1, Wu discloses a fluid conductor (conduit body 105, Fig 5), comprising:
`
`a first barrier (combination of upper cover 215 and base 210, Fig 5) formed fromafirst fluid-
`
`impermeable material (conduit body 105 can be made of polyethylene or polyurethane 4[0046]) and
`
`defining a fluid path (device 100 may be configured to create and maintain a fluidic pathway 4][0041])
`
`having a first end (distal port 115, Fig 1. Figs 1 and 5 show the same device, with Fig 5 illustrating the
`
`different layers that make up conduit body 105. Distal port 115, proximal port 125, and device 100, are
`
`presentin both figures, but they are unlabeled in Fig 5), a second end (proximal port 125, Fig 1), anda
`
`longitudinal axis (conduit body 105 has a longitudinal axis extending between 125 and 115,Fig 1);
`
`and a plurality of pressure indicators (both ports 115 and 125 may include pressure indicators
`
`230, Fig 5. The indicator 230 can be positioned in one or morelocations along the device 100 such as
`
`within the tubing 130, on a region of the conduit body 105, near the distal port 115 or proximal port 125
`
`4[0057]) configured to elastically deform in response to a reduction in pressure (Deformation of the
`
`exterior wall of the indicator 230 can result in a change fromafirst profile to a second profile. The
`
`relative position of the exterior wall between the first profile and the second profile can be indicative of
`
`an amountof pressure delivered to the wound site. 4[0058)]).
`
`Wuis silent regarding the first barrier and second barrier having similar profiles and dimensions,
`
`indicator ports distributed linearly along the first barrier parallel to and in fluid communication with the
`
`fluid path; and a second barrier coupled to the first barrier, the second barrier formed from a second
`
`fluid-impermeable material defining a plurality of pressure indicators aligned with the indicator ports
`
`and configured to elastically deform in response to a reduction in pressure throughthe indicator ports.
`
`

`

`Application/Control Number: 17/281,668
`Art Unit: 3781
`
`Page 5
`
`However, Robinson teaches a reduced pressure therapy dressing, thus from the same field of
`
`endeavor, with indicator ports (port aperture 120, Fig 1), and a second barrier (flexible member 138, Fig
`
`1) coupled to the first barrier (sealing member 114, Fig 1), the second barrier (138, Fig 1) formed from a
`
`second fluid-impermeable material (connector body 123 and flexible member 138 may be formed or
`
`molded as an integral member.{][0042] Connector body 123 may be made from polyurethane 4][0032]
`
`and thus 138 may be made of polyurethane as well) defining a pressure indicator (dressing valve 136
`
`may indicate a reduced pressure state. Upon being subjected to at least threshold reduced pressure that
`
`is delivered to the interior of the dressing valve 136, the flexible member 138 collapses and seals the
`
`port aperture 120. 4[0042]) aligned with the indicator ports (dressing valve 136is aligned with aperture
`
`120, Fig 1) and configured to elastically deform in response to a reduction in pressure (Upon being
`
`subjected to at least threshold reduced pressurethat is delivered to the interior of the dressing valve
`
`136, the flexible member 138 collapses and seals the port aperture 120. 4[0042]) through the indicator
`
`ports (reduced pressureis delivered to dressing valve 136 through port aperture 120 4[0040)), in order
`
`to indicate to the user or physician if negative pressure is uniformly distributed and to indicate the
`
`presenceand location of clogs within the fluid path, and to reduceor eliminate vacuum lock associated
`
`with the delivery conduit and reduced pressure source (4][0041)]).
`
`Therefore, it would have been obvious to one ofordinaryskill in the art before the effective
`
`filing date of the claimed invention to modify the device of Wu to substitute the pressure indicator
`
`structure of Robinson for the pressure indicator structure of Wu. The combination of the conduit having
`
`a first impermeable barrier and a plurality of pressure indicators of Wu with the second impermeable
`
`barrier and aperture of the pressure indicator of Robinson would result in a device having two
`
`impermeable barriers with apertures distributed linearly along the first barrier and in fluid
`
`communication with the fluid path, where each aperture is associated with a single elastically
`
`deformable pressure indicator formed from the second impermeable barrier. The resulting device would
`
`

`

`Application/Control Number: 17/281,668
`Art Unit: 3781
`
`Page 6
`
`be advantageous since a plurality of pressure indicators would indicate to the user or physician if
`
`negative pressure is uniformly distributed and would indicate the presence and location of clogs within
`
`the fluid path, and to reduceor eliminate vacuum lock associated with the deliver conduit and reduced
`
`pressure source (as motivated by Robinson 4][0041]).
`
`Wu/Robinson is silent regarding the first barrier and second barrier having similar profiles and
`
`dimensions.
`
`However, Weston teaches a reduced pressure treatment system, thus from the same field of
`
`endeavor, with a wound cover that is interpreted as having a dual layer structure wherein both layers
`
`have similar profiles and dimensions (bottom layer of wound cover 40 not including protrusions 60, and
`
`top layer is the portion of the layer including protrusions 60 and remaining surface 43 Fig 5A-D since
`
`having the layers be of similar profiles and dimensions would allow for simpler manufacturing 4[0054)).
`
`Therefore, it would have been obvious to modify the barriers of Wu/Robinson to havelayers
`
`with similar profiles and dimensions a taught by Weston,in order to allow for simpler manufacturing.
`
`Regarding Claim 2, Wu/Robinson/Weston teachesthe claimed limitations as set forth above for
`
`Claim 1.
`
`Robinson teaches that the plurality of pressure indicators comprise blisters or raised channels
`
`(Fig 1 of Robinson showsa raised configuration of 136) in the second barrier (138, Fig 1), which would
`
`provide a visual or tactile indication whether sufficient reduced pressureis achieved.
`
`Therefore, the combination of Wu/Robinson/Weston also teaches that the plurality of pressure
`
`indicators comprise blisters or raised channels in the second barrier, which would provide a visual or
`
`tactile indication whether sufficient reduced pressureis achieved.
`
`Regarding Claim 3, Wu/Robinson/Weston teachesthe claimed limitations as set forth above for
`
`Claim 1.
`
`

`

`Application/Control Number: 17/281,668
`Art Unit: 3781
`
`Page 7
`
`Robinson teachesthat each of the plurality of pressure indicators (pressure indicators 230 of
`
`Wu,Fig 5) is aligned with only one of the indicator ports (aperture 120 of Robinson, Fig 1) in order to
`
`indicate to the user or physician if negative pressureis uniformly distributed and to indicate the
`
`presence and location of clogs within the fluid path and to reduceor eliminate vacuum lock associated
`
`with the deliver conduit and reduced pressure source (4][0041)).
`
`Therefore, the combination of Wu/Robinson/Weston also teaches that each of the plurality of
`
`pressureindicatorsis aligned with only one of the indicator ports in order to indicate to the user or
`
`physician if negative pressure is uniformly distributed and to indicate the presence and location of clogs
`
`within the fluid path and to reduceor eliminate vacuum lock associated with the deliver conduit and
`
`reduced pressure source (as motivated by Robinson 4][0041]).
`
`Regarding Claim 4, Wu/Robinson/Weston is silent whether each of the plurality of pressure
`
`indicators has a length of about 10 millimeters to about 30 millimeters, a width of about 2 millimeters to
`
`about 4 millimeters, and a height of about 1 millimeter to about 3 millimeters.
`
`It would have been obvious to cause the device of Wu/Robinson to have eachofthe plurality of
`
`pressure indicators to have a length of about 10 millimeters to about 30 millimeters, a width of about 2
`
`millimeters to about 4 millimeters, and a height of about 1 millimeter to about 3 millimeters since it has
`
`been held that “where the only difference between the prior art and the claims was a recitation of
`
`relative dimensions of the claimed device and a device having the claimed relative dimensions would
`
`not perform differently than the prior art device, the claimed device was not patentably distinct from
`
`the prior art device” Gardner v. TEC Syst., Inc., 725 F.2d 1338, 220 USPQ 777 (Fed. Cir. 1984), cert.
`
`denied, 469 U.S. 830, 225 SPQ 232 (1984) (MPEP 2144.04 IV(A)). In the instant case, the device of
`
`Wu/Robinson would not operate differently with the claimed dimensions. Further, applicant places no
`
`criticality on the range claimed, indicating simply that the dimensions “may” be within the claimed
`
`ranges(specification 4[00135]).
`
`

`

`Application/Control Number: 17/281,668
`Art Unit: 3781
`
`Page 8
`
`Regarding Claim 5, Wu/Robinson/Weston is silent regarding a gap between each ofthe plurality
`
`of pressure indicators, wherein the gap has a length of about 5 millimeters to about 10 millimeters.
`
`It would have been obvious to cause the device of Wu/Robinson to have a gap between each of
`
`the plurality of pressure indicators, wherein the gap has a length of about 5 millimeters to about 10
`
`millimeters since it has been held that “where the only difference between the prior art and the claims
`
`was a recitation of relative dimensions of the claimed device and a device having the claimed relative
`
`dimensions would not perform differently than the prior art device, the claimed device was not
`
`patentably distinct from the prior art device” Gardner v. TEC Syst., Inc., 725 F.2d 1338, 220 USPQ 777
`
`(Fed. Cir. 1984), cert. denied, 469 U.S. 830, 225 SPQ 232 (1984) (MPEP 2144.04 IV(A)). In the instant
`
`case, the device of Wu/Robinson would not operate differently with the claimed length. Further,
`
`applicant places no criticality on the range claimed, indicating simply that the length “may” be within
`
`the claimed ranges (specification 4[00135]).
`
`Regarding Claim 8, Wu/Robinson/Weston teachesthe claimed limitations as set forth above for
`
`Claim 1.
`
`Robinson teachesthat each of the indicator ports (aperture 120, Fig 1) comprises an aperture in
`
`the first barrier, which would indicate to the user or physician if negative pressure is uniformly
`
`distributed and would indicate the presence and location of clogs within the fluid path.
`
`Therefore, Wu/Robinson/Weston teachesthat each of the indicator ports comprises an
`
`aperturein the first barrier, which would indicate to the user or physician if negative pressureis
`
`uniformly distributed and would indicate the presence and location of clogs within the fluid path.
`
`Regarding Claim 10, Wu/Robinson/Weston is silent whether the indicator ports are distributed
`
`along the first barrier with a spacing of about 10 millimeters to about 30 millimeters.
`
`It would have been obvious to cause the device of Wu/Robinson to havethe indicator ports
`
`distributed along the first barrier with a spacing of about 10 millimeters to about 30 millimeters since it
`
`

`

`Application/Control Number: 17/281,668
`Art Unit: 3781
`
`Page 9
`
`has been held that “where the only difference betweenthe prior art and the claims was a recitation of
`
`relative dimensions of the claimed device and a device having the claimed relative dimensions would
`
`not perform differently than the prior art device, the claimed device was not patentably distinct from
`
`the prior art device” Gardner v. TEC Syst., Inc., 725 F.2d 1338, 220 USPQ 777 (Fed. Cir. 1984), cert.
`
`denied, 469 U.S. 830, 225 SPQ 232 (1984) (MPEP 2144.04 IV(A)). In the instant case, the device of
`
`Wu/Robinson would not operate differently with the claimed spacing. Further, applicant places no
`
`criticality on the range claimed, indicating simply that the spacing “may” be within the claimed ranges
`
`(specification 4[00132]).
`
`Regarding Claim 11, Wu further discloses that the first fluid-impermeable material is
`
`polyurethane or polyethylene (conduit body 105 can be made of polyethylene or polyurethane 4][0046]).
`
`Regarding Claim 12, Wu further discloses the first barrier (combination of 215 and 210, Fig 5)
`
`comprisesa first layer (215, fig 7A) and a second layer (210, Fig 7A) sealed to the first layer (conduit
`
`body 105 can have a multilayer structure as shownin Fig 5, where two or more components may be
`
`bonded together by adhesive 4][0048]) to define the fluid path (channel 200, Fig 7A) betweenthe first
`
`layer (215, Fig 7A) and the second layer (210, Fig 7A).
`
`Regarding Claim 14, Wu further discloses a manifold (plurality of support structures 205, Fig
`
`7A) disposed between the first layer (215, Fig 7A) and the second layer (210, Fig 7A).
`
`Regarding Claim 15, Wu further discloses at least one of the first layer (215, Fig 7A) and the
`
`second layer (210, Fig 7A) comprise a plurality of bubbles or blisters (support structures 205, Fig 7A can
`
`be rounded or semi-spherical, 4[0048]) configured to support the fluid path (The support structures
`
`205 prevent or mitigate collapse of the channel 200 upon application of internal vacuum, external
`
`pressure to or compression of the conduit body 105 4[0047)).
`
`

`

`Application/Control Number: 17/281,668
`Art Unit: 3781
`
`Page 10
`
`Regarding Claim 16, Wu further discloses open-cell foam or a textile (The support structure
`
`205 can include a fabric or other material having open channels running therethrough 4[0048])
`
`disposed between the first layer (215, Fig 7A) and the second layer (210, Fig 7A).
`
`Regarding Claim 17, Wufurther discloses that the first fluid-impermeable material is a polymer
`
`(conduit body 105 can be made of polyethylene or polyurethane 4][0046]).
`
`Regarding Claim 18, Wu further discloses that the first barrier (combination of 215 and 210,
`
`Fig 7A) comprises a tube (Fig 7A showsa flattened tube configuration).
`
`Regarding Claim 19, Wu/Robinson/Weston teachesthe claimed limitations as set forth above
`
`for Claim 1.
`
`Robinson teaches an adhesive (bonds 402 may be used to connectthe flexible member 138 to
`
`the connector body 123. Fig 4 §[0049]) between the second barrier (138, Fig 4) and the first barrier
`
`(combination of 215 and 210, Fig 5 of Wu).
`
`Therefore, the combination of Wu/Robinson/Weston teaches an adhesive between the second
`
`barrier and the first barrier.
`
`Regarding Claim 26, Wu teaches an apparatus for treating a tissue site with negative pressure
`
`(device 100, Fig 1), the apparatus comprising:
`
`A source of negative pressure (The proximal port 125 can connect, for example via tubing 130,
`
`to any of a variety of negative pressure sources (not shown) 4[0041]); a dressing (distal port 115 may
`
`connect to a wound dressing positioned at a wound site [0041]) configured to be applied to the tissue
`
`site; and a fluid conductor (conduit body 105, Fig 5), comprising:
`
`a first barrier (combination of upper cover 215 and base 210, Fig 5) formed fromafirst fluid-
`
`impermeable material (conduit body 105 can be made of polyethylene or polyurethane 4[0046]) and
`
`defining a fluid path (device 100 may be configured to create and maintain a fluidic pathway 4][0041])
`
`having a first end (distal port 115, Fig 1. Figs 1 and 5 show the same device, with Fig5illustrating the
`
`

`

`Application/Control Number: 17/281,668
`Art Unit: 3781
`
`Page 11
`
`different layers that make up conduit body 105. Distal port 115, proximal port 125, and device 100, are
`
`presentin both figures, but they are unlabeled in Fig 5), a second end (proximal port 125, Fig 1), anda
`
`longitudinal axis (conduit body 105 has a longitudinal axis extending between 125 and 115,Fig 1);
`
`and a plurality of pressure indicators (both ports 115 and 125 may include pressure indicators
`
`230, Fig 5. The indicator 230 can be positioned in one or more locations along the device 100 such as
`
`within the tubing 130, on a region of the conduit body 105, near the distal port 115 or proximal port 125
`
`4[0057]) configured to elastically deform in response to a reduction in pressure (Deformation of the
`
`exterior wall of the indicator 230 can result in a change fromafirst profile to a second profile. The
`
`relative position of the exterior wall between the first profile and the second profile can be indicative of
`
`an amount of pressure delivered to the wound site. 4][0058])
`
`configuredto fluidly couple the dressing to the source of negative pressure (the device 100
`
`may be configured to create and maintain a fluidic pathway or channel for the delivery of negative
`
`pressure to and removal of exudates from the wound site 4][0041]).
`
`Wuis silent regarding the first barrier and second barrier having similar profiles and dimensions,
`
`indicator ports distributed linearly along the first barrier parallel to and in fluid communication with the
`
`fluid path; and a second barrier coupled to the first barrier, the second barrier formed from a second
`
`fluid-impermeable material defining a plurality of pressure indicators aligned with the indicator ports
`
`and configured to elastically deform in response to a reduction in pressure throughthe indicator ports.
`
`However, Robinson teaches a reduced pressure therapy dressing, thus from the same field of
`
`endeavor, with indicator ports (port aperture 120, Fig 1), and a second barrier (flexible member 138, Fig
`
`1) coupled to the first barrier (sealing member 114, Fig 1), the second barrier (138, Fig 1) formed from a
`
`second fluid-impermeable material (connector body 123 and flexible member 138 may be formed or
`
`molded as an integral member.4][0042] Connector body 123 may be made from polyurethane 4][0032]
`
`and thus 138 may be made of polyurethane as well) defining a pressure indicator (dressing valve 136
`
`

`

`Application/Control Number: 17/281,668
`Art Unit: 3781
`
`Page 12
`
`may indicate a reduced pressure state. Upon being subjected to at least threshold reduced pressure that
`
`is delivered to the interior of the dressing valve 136, the flexible member 138 collapses and seals the
`
`port aperture 120. 4[0042]) aligned with the indicator ports (dressing valve 136 is aligned with aperture
`
`120, Fig 1) and configured to elastically deform in response to a reduction in pressure (Upon being
`
`subjected to at least threshold reduced pressurethat is delivered to the interior of the dressing valve
`
`136, the flexible member 138 collapses and seals the port aperture 120. 4[0042]) through the indicator
`
`ports (reduced pressureis delivered to dressing valve 136 through port aperture 120 4[0040)), in order
`
`to indicate to the user or physician if negative pressure is uniformly distributed and to indicate the
`
`presenceand location of clogs within the fluid path and to reduceor eliminate vacuum lock associated
`
`with the delivery conduit and reduced pressure source (4][0041)]).
`
`Therefore, it would have been obvious to one ofordinaryskill in the art before the effective
`
`filing date of the claimed invention to modify the device of Wu to have substitute the pressure indicator
`
`structure of Robinson for the pressure indicator structure of Wu. The combination of the conduit having
`
`a first impermeable barrier and a plurality of pressure indicators of Wu with the second impermeable
`
`barrier and aperture of the pressure indicator of Robinson would result in a device having two
`
`impermeable barriers with apertures distributed linearly along the first barrier and in fluid
`
`communication with the fluid path, where each apertureis associated with a single elastically
`
`deformable pressure indicator formed from the second impermeable barrier. The resulting device would
`
`be advantageous since a plurality of pressure indicators would indicate to the user or physician if
`
`negative pressure is uniformly distributed and would indicate the presence and location of clogs within
`
`the fluid path and to reduce or eliminate vacuum lock associated with the delivery conduit and reduced
`
`pressure source (as motivated by Robinson 4][0041)).
`
`Wu/Robinson is silent regarding the first barrier and second barrier having similar profiles and
`
`dimensions.
`
`

`

`Application/Control Number: 17/281,668
`Art Unit: 3781
`
`Page 13
`
`However, Weston teaches a reduced pressure treatment system, thus from the same field of
`
`endeavor, with a wound cover that is interpreted as having a dual layer structure wherein both layers
`
`have similar profiles and dimensions (bottom layer of wound cover 40 not including protrusions 60, and
`
`top layer is the portion of the layer including protrusions 60 and remaining surface 43 Fig 5A-D since
`
`having the layers be of similar profiles and dimensions would allow for simpler manufacturing 4[0054)).
`
`Therefore, it would have been obvious to modify the barriers of Wu/Robinson to have layers
`
`with similar profiles and dimensions a taught by Weston,in order to allow for simpler manufacturing.
`
`Claim 7 is rejected under 35 U.S.C. 103 as being unpatentable over Wu/Robinson/Weston in
`
`view of Locke (WO 2016182977 A1).
`
`Regarding Claim 7, Wu/Robinson/Weston is silent regarding open-cell foam or gauze
`
`configured to bias the pressure indicators.
`
`However, Locke teaches a reduced pressure device, thus from the same field of endeavor,
`
`wherein open-cell foam or gauze (foam block 134, Fig 1) configured to bias the pressure indicators
`
`([0059]). If the foam block remains compressed, a patient or clinician may have an indication that the
`
`therapy pressure has been reached. The compressed foam block may also act as a pressure reservoir.
`
`As negative pressure therapy is provided, there may be a natural leakage or decline of negative
`
`pressureat the tissue site. As the negative pressure decreasesin the sealed therapeutic environment,
`
`the pressure differential across the dressing assembly may decrease and the foam block may gradually
`
`expand, reapplying negative pressure at the tissue site ({][0067]).
`
`Therefore, it would have been obvious to modify the device of Wu/Robinson to include open-
`
`cell foam or gauze configured to bias the pressure indicators, as taught by Locke, becauseif the foam
`
`block remains compressed, a patient or clinician may have an indication that the therapy pressure has
`
`been reached. The compressed foam block may also act as a pressure reservoir. As negative pressure
`
`therapy is provided, there may be a natural leakage or decline of negative pressure at the tissue site.
`
`

`

`Application/Control Number: 17/281,668
`Art Unit: 3781
`
`Page 14
`
`As the negative pressure decreases in the sealed therapeutic environment, the pressure differential
`
`across the dressing assembly may decrease and the foam block may gradually expand, reapplying
`
`negative pressureat the tissue site (as motivated by Locke 4[0067]).
`
`Claim 20 is rejected under 35 U.S.C. 103 as being unpatentable over Wu/Robinson/Weston in
`
`view of Bogie (US 20200061379 A1).
`
`Regarding Claim 20, Wu/Robinson/Weston is silent regarding a diagnostic port in the first
`
`barrier; and a diagnostic indicator operatively coupled to the fluid path through the diagnostic port and
`
`at least partially visible through the second barrier.
`
`However, Bogie teaches a device for remotely monitoring and treating wounds, thus from the
`
`same field of endeavor, with a diagnostic port (holes 126, Fig 1A 4[0043]) in the first barrier (absorbent
`
`layer 106 has holes 126 therethrough, Fig 1A 4[0043]); and a diagnostic indicator (plurality of
`
`temperature sensors 124, Fig 1 4[0043]) operatively coupled to the fluid path through the diagnostic
`
`port (temperature sensors 124 can be positioned within the holes 126 and attached to the underside of
`
`the top layer 104, Fig 1A 4[0043]) and at least partially visible through the second barrier (top layer 104
`
`is transparent 4[0041] with temperature sensors 124 visible underneath) to determine an infection
`
`status of the wound based on a temperaturedifference (][0021]).
`
`Therefore, it would have been obvious to modify the device of Wu/Robinson/Weston to include
`
`a diagnostic port in the fir

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket