throbber
www.uspto.gov
`
`UNITED STATES DEPARTMENT OF COMMERCE
`United States Patent and Trademark Office
`Address: COMMISSIONER FOR PATENTS
`P.O. Box 1450
`Alexandria, Virginia 22313-1450
`
`17/281,668
`
`03/31/2021
`
`Christopher Brian LOCKE
`
`PO01658US02PCT
`
`1000
`
`60402
`
`7590
`
`07/03/2024
`
`KINETIC CONCEPTS, INC.
`c/o Harness Dickey & Pierce
`5445 Corporate Drive
`Suite 20
`Troy, MI 48098
`
`EXAMINER
`
`FLYNN, TIMOTHY LEE
`
`Para NONE
`
`3781
`
`NOTIFICATION DATE
`
`DELIVERY MODE
`
`07/03/2024
`
`ELECTRONIC
`
`Please find below and/or attached an Office communication concerning this application or proceeding.
`
`The time period for reply, if any, is set in the attached communication.
`
`Notice of the Office communication was sent electronically on above-indicated "Notification Date" to the
`following e-mail address(es):
`
`IPDocketing @ Solventum.com
`dgodzisz@hdp.com
`troymailroom @hdp.com
`
`PTOL-90A (Rev. 04/07)
`
`

`

`Application No.
`Applicant(s)
`17/281,668
`LOCKE et al.
`
`Office Action Summary Art Unit|AIA (FITF)StatusExaminer
`TIMOTHY L FLYNN
`3781
`Yes
`
`
`
`-- The MAILING DATEof this communication appears on the cover sheet with the correspondence address --
`Period for Reply
`
`A SHORTENED STATUTORYPERIOD FOR REPLYIS SET TO EXPIRE 3 MONTHS FROM THE MAILING
`DATE OF THIS COMMUNICATION.
`Extensionsof time may be available underthe provisions of 37 CFR 1.136(a). In no event, however, may a reply betimely filed after SIX (6) MONTHSfrom the mailing
`date of this communication.
`If NO period for reply is specified above, the maximum statutory period will apply and will expire SIX (6) MONTHSfrom the mailing date of this communication.
`-
`- Failure to reply within the set or extended period for reply will, by statute, cause the application to become ABANDONED (35 U.S.C. § 133).
`Any reply received by the Office later than three months after the mailing date of this communication, evenif timely filed, may reduce any earned patent term
`adjustment. See 37 CFR 1.704(b).
`
`Status
`
`
`
`1) Responsive to communication(s)filed on 04 April 2024.
`C) A declaration(s)/affidavit(s) under 37 CFR 1.130(b) was/werefiled on
`2a)[¥) This action is FINAL.
`2b) (J This action is non-final.
`3) An election was madeby the applicant in responseto a restriction requirement set forth during the interview
`on
`; the restriction requirement and election have been incorporated into this action.
`4)(2) Since this application is in condition for allowance except for formal matters, prosecution as to the merits is
`closed in accordance with the practice under Exparte Quayle, 1935 C.D. 11, 453 O.G. 213.
`
`Disposition of Claims*
`1-5,7-8,10-11,19-20 and 26 is/are pending in the application.
`)
`Claim(s)
`5a) Of the above claim(s) _ is/are withdrawn from consideration.
`CL] Claim(s)__is/are allowed.
`Claim(s) 1-5,7-8,10-11,19-20 and 26is/are rejected.
`(] Claim(s)__ is/are objectedto.
`C] Claim(s
`are subjectto restriction and/or election requirement
`)
`* If any claims have been determined allowable, you maybeeligible to benefit from the Patent Prosecution Highway program at a
`participating intellectual property office for the corresponding application. For more information, please see
`http://www.uspto.gov/patents/init_events/pph/index.jsp or send an inquiry to PPHfeedback@uspto.gov.
`
`) ) ) )
`
`Application Papers
`10) The specification is objected to by the Examiner.
`11)0) The drawing(s) filedon__ is/are: a)(J accepted or b)( objected to by the Examiner.
`Applicant may not request that any objection to the drawing(s) be held in abeyance. See 37 CFR 1.85(a).
`Replacement drawing sheet(s) including the correction is required if the drawing(s) is objected to. See 37 CFR 1.121(d).
`
`Priority under 35 U.S.C. § 119
`12)7) Acknowledgment is made of a claim for foreign priority under 35 U.S.C. § 119(a)-(d)or (f).
`Certified copies:
`c)Z None ofthe:
`b)() Some**
`a)C All
`1.1.) Certified copies of the priority documents have been received.
`2.2) Certified copies of the priority documents have been received in Application No.
`3.1.) Copies of the certified copies of the priority documents have been receivedin this National Stage
`application from the International Bureau (PCT Rule 17.2(a)).
`*“ See the attached detailed Office action for a list of the certified copies not received.
`
`Attachment(s)
`
`1)
`
`Notice of References Cited (PTO-892)
`
`2) (J Information Disclosure Statement(s) (PTO/SB/08a and/or PTO/SB/08b)
`Paper No(s)/Mail Date
`U.S. Patent and Trademark Office
`
`3)
`
`4)
`
`(LJ Interview Summary (PTO-413)
`Paper No(s)/Mail Date
`(Qj Other:
`
`PTOL-326 (Rev. 11-13)
`
`Office Action Summary
`
`Part of Paper No./Mail Date 20240626
`
`

`

`Application/Control Number: 17/281,668
`Art Unit: 3781
`
`Page 2
`
`DETAILED ACTION
`
`Notice of Pre-AlA orAIA Status
`
`The presentapplication,filed on or after March 16, 2013, is being examined under the first
`
`inventorto file provisions of the AIA.
`
`Response to Arguments
`
`Applicant’s amendments filed 04/04/2024 have been accepted. Claims 12, 14-18, 24, and 24 are
`
`canceled. Claims 1-5, 7-8, 10-11, 19-20, and 26 are pending.
`
`Applicant's argumentsfiled 04/04/2024 have been fully considered but they are not persuasive.
`
`Applicant argues that Wu/Robinson/Weston does not teach the amended limitations of claim 1
`
`wherein the first barrier comprises, “a first layer formed from a film” and “ a second layer sealed to the
`
`first layer to define a fluid path between thefirst layer and the second layer, the second layer formed
`
`froma film,” stating that the upper cover 215 and lower base 210 of Wu, which read on thefirst and
`
`second layers of the first barrier are not formed from a film. However, Dictionary.com defines a film as
`
`“athin sheet of any material.”
`
`|.”
`
`Thus, the upper cover 215 and lower base 210 of Wu meet the
`
`requirements of the claim as set forth below.
`
`Applicant argues that Wu/Robinson/Weston does not teach the amendedlimitations of claim 1
`
`wherein,“a plurality of standoffs formed by at least one of thefirst layer and the second layer,” stating
`
`that the support structures 205 of Wu protrude from a middle layer 207 rather than the first layer or the
`
`second layer (215 and 210 of Wu, respectively). However, Wu [0048] states that the support structures
`
`205 may beintegrated with or coupled to the inner walls of the conduit body 105, which includes layers
`
`215 and 210. Thus, the support structures 205 being integrated or coupled to layer 215 or 210 of Wu
`
`meet the requirements of the claimas set forth below.
`
`Applicant argues that one of ordinary skill would not combine the dressing valve 136 of
`
`Robinson with the pressure indicators of Wu because the connector body 123 of Robinson includes a
`
`

`

`Application/Control Number: 17/281,668
`Art Unit: 3781
`
`Page 3
`
`receptacle 134 that would prevent application of negative pressure, and that the size of connector body
`
`123 would not motivate one to combine Robinson with Wu. However, the combination is drawn to the
`
`dressing valve 136 only, not the entire connector body 123, so the size of connector body 123 would not
`
`prevent the combination from being made. Additionally, the receptacle 134 is merely a connector toa
`
`reduced pressure delivery conduit, and thus would not prevent application of negative pressure when
`
`coupled to a negative pressure source as intended (Robinson 4][0023]). Additionally, since only the
`
`dressing valve 136 of Robinson is combined with thefirst barrier of Wu, the receptacle 134 would not
`
`necessarily be included in the combination. Thus, the rejection is maintained as set forth below.
`
`Applicant again argues that wound cover 40 of Weston cannot be interpreted as having a dual
`
`layer structure. However, the examiner maintains that Weston 4[0055] states that the protrusions 60
`
`may be constructed of a different material than cover 40, which clearly implies a dual layer structure,
`
`since the layer including the protrusions 60 may be constructed of a different material having its own
`
`thickness, pliability, or color that is different than the cover 40, (Weston 4[0055-0058)).
`
`Additionally, Applicant argues that cover 40 cannot be interpreted as having a dual layer
`
`structure because Westonstates in 4 [0054] that the plurality of protrusions 60 are embedded in the
`
`cover 40. However, the examiner maintains that the claimis written broadly enough that the cover 40
`
`and protrusions 60 of Weston arestill capable of being interpreted as having a dual layer structure.
`
`Furthermore, Wuin view of Robinson is relied upon as teaching the dual layer structure, and Westonis
`
`merely relied upon to teachthat the layers may havesimilar profiles and dimensions. In light of the
`
`combination of Wu/Robinson/Weston, it would be obvious to one of ordinary skill to have dual layers
`
`having similar profiles and dimension in order to simplify manufacturing.
`
`Applicant did not specifically argue the dependent claims.
`
`Claim Rejections - 35 USC § 103
`
`

`

`Application/Control Number: 17/281,668
`Art Unit: 3781
`
`Page 4
`
`The following is a quotation of 35 U.S.C. 103 which forms the basis for all obviousness rejections
`
`set forth in this Office action:
`
`A patent for a claimed invention may not be obtained, notwithstanding that the claimed invention is
`not identically disclosed as set forthin section 102, if the differences between the claimed invention
`and the prior art are such that the claimed invention as a whole would have been obvious before the
`effective filing date of the claimed invention to a person having ordinary skillin the art to which the
`claimed invention pertains. Patentability shall not be negated by the manner in which the invention
`was made.
`
`The factual inquiries for establishing a background for determining obviousness under 35 U.S.C.
`
`103 are summarized as follows:
`
`1. Determining the scope and contents of the prior art.
`
`2. Ascertaining the differences between the prior art and the claims at issue.
`
`3. Resolving the level of ordinary skill in the pertinent art.
`
`4. Considering objective evidence present in the application indicating obviousness or
`
`nonobviousness.
`
`This application currently names joint inventors. In considering patentability of the claims the
`
`examiner presumesthat the subject matter of the various claims was commonly ownedas of the
`
`effective filing date of the claimed invention(s) absent any evidence to the contrary. Applicant is advised
`
`of the obligation under 37 CFR 1.56 to point out the inventor and effectivefiling dates of each claim that
`
`was not commonly ownedas of the effective filing date of the later invention in order for the examiner
`
`to consider the applicability of 35 U.S.C. 102(b)(2)(C) for any potential 35 U.S.C. 102(a)(2) prior art
`
`against the later invention.
`
`Claims 1-5, 8, 10-11, 19, and 26 are rejected under 35 U.S.C. 103 as being unpatentable over
`
`Wu (US 20130144230 A1) in view of Robinson (US 20110224633 A1), and furtherin view of Weston
`
`(US 20040073151 A1).
`
`Regarding Claim 1, Wu disclosesa fluid conductor (conduit body 105, Fig 5), comprising:
`
`

`

`Application/Control Number: 17/281,668
`Art Unit: 3781
`
`Page5S
`
`a first barrier (combination of upper cover 215 and base 210, Fig 5) formed fromafirst fluid -
`
`impermeable material (conduit body 105 can be made of polyethylene or polyurethane 4][0046]), the
`
`first barrier comprising:
`
`a first layer formed fromafilm (215 is formed from a film, defined asathin layer of any
`
`material, Fig 5);
`
`a second layer (210, Fig 5) sealedto thefirst layer to define a fluid path between thefirst layer
`
`and the second layer (device 100 may be configured to create and maintain a fluidic pathway,
`
`specifically between layers 215 and 210 4[0041][0050]), the second layer formed from a film (210 is
`
`formed fromafilm, defined as a thin layer of any material, Fig 5);
`
`the fluid path having a first end (distal port 115, Fig 1. Figs 1 and S show the same device, with
`
`Fig 5 illustrating the different layers that make up conduit body 105. Distal port 115, proximal port 125,
`
`and device 100, are presentin both figures, but they are unlabeled in Fig 5), a secondend (proximal port
`
`125, Fig 1), and a longitudinal axis (conduit body 105 has a longitudinal axis extending between 125 and
`
`115, Fig 1);
`
`and a plurality of standoffs (support structures 205, Fig 7A) formed byat least one ofthe first
`
`layer and the second layer (support structures 205 may be integrated with or coupled to the inner walls
`
`of the conduit body 105, which includes layers 215 and 210 4[0048]), the plurality of standoffs disposed
`
`in the fluid path (The support structures 205 prevent or mitigate collapse of the channel 200 upon
`
`application of internal vacuum, external pressure to or compression of the conduit body 105 4[0047)).
`
`and a plurality of pressure indicators (both ports 115 and 125 may include pressure indicators
`
`230, Fig 5. The indicator 230 can be positioned in one or more locations along the device 100 suchas
`
`within the tubing 130, ona region of the conduit body 105, near the distal port 115 or proximal port 125
`
`4[0057]) configured to elastically deform in response to a reduction in pressure (Defor mation of the
`
`exterior wall of the indicator 230 can result in a change fromafirst profile to a second profile. The
`
`

`

`Application/Control Number: 17/281,668
`Art Unit: 3781
`
`Page 6
`
`relative position of the exterior wall between thefirst profile and the second profile can be indicative of
`
`an amount of pressure delivered to the woundsite. [0058]).
`
`Wuis silent regarding the first barrier and second barrier having similar profiles and dimensions,
`
`indicator ports distributed linearly along the first barrier parallel to and in fluid communication with the
`
`fluid path; and asecond barrier coupled to the first barrier, the second barrier formed from asecond
`
`fluid-impermeable material defining a plurality of pressure indicators aligned with the indicator ports
`
`and configured to elastically deform in response to a reduction in pressure through the indicator ports.
`
`However, Robinson teaches a reduced pressure therapy dressing, thus fromthe same field of
`
`endeavor, withindicator ports (port aperture 120, Fig 1), and a second barrier (flexible member 138,Fig
`
`1) coupled to thefirst barrier (sealing member 114, Fig 1), the second barrier (138, Fig 1) formed from a
`
`second fluid-impermeable material (connector body 123 and flexible member 138 may be formed or
`
`molded as an integral member. [0042] Connector body 123 may be made from polyurethane 4[0032]
`
`and thus 138 may be madeof polyurethane as well) defining a pressure indicator (dressing valve 136
`
`may indicate a reduced pressure state. Upon being subjected to at least threshold reduced pressure that
`
`is delivered to the interior of the dressing valve 136, the flexible member 138 collapses and seals the
`
`port aperture 120. 4[0042]) aligned with the indicator ports (dressing valve 136 is aligned with aperture
`
`120, Fig 1) and configuredto elastically deform in response to a reduction in pressure (Upon being
`
`subjected to at least threshold reduced pressure that is delivered to the interior of the dressing valve
`
`136, the flexible member 138 collapses and seals the port aperture 120. 4|[0042]) through the indicator
`
`ports (reduced pressureis delivered to dressing valve 136 through port aperture 120 4|[0040]), in order
`
`to indicate to the user or physician if negative pressureis uniformly distributed and to indicate the
`
`presence andlocation of clogs within the fluid path, and to reduce or eliminate vacuum lock associated
`
`with the delivery conduit and reduced pressure source (4][0041]).
`
`

`

`Application/Control Number: 17/281,668
`Art Unit: 3781
`
`Page 7
`
`Therefore, it would have been obvious to one of ordinary skill in the art before the effective
`
`filing date of the claimed invention to modify the device of Wu to substitute the pressure indicator
`
`structure of Robinson for the pressure indicator structure of Wu. The combination of the conduit having
`
`a first impermeable barrier and a plurality of pressure indicators of Wu with the second impermeable
`
`barrier and aperture of the pressure indicator of Robinson would result in a device having two
`
`impermeable barriers with apertures distributed linearly along the first barrier and in fluid
`
`communication with the fluid path, where each aperture is associated with a single elastically
`
`deformable pressure indicator formed from the second impermeable barrier. The resulting device would
`
`be advantageoussinceaplurality of pressure indicators would indicate to the user or physician if
`
`negative pressure is uniformly distributed and would indicate the presence and location of clogs within
`
`the fluid path, and to reduce or eliminate vacuum lock associated with the deliver conduit and reduced
`
`pressure source (as motivated by Robinson 4[0041]).
`
`Wu/Robinsonis silent regarding the first barrier and second barrier having similar profiles and
`
`dimensions.
`
`However, Weston teaches a reduced pressure treatment system, thus from the same field of
`
`endeavor, witha wound cover that is interpreted as having a dual layer structure wherein both layers
`
`have similar profiles and dimensions (bottom layer of wound cover 40 not including protrusions 60, and
`
`top layer is the portion of the layer including protrusions 60 and remaining surface 43 Fig 5A-D since
`
`having the layers be of similar profiles and dimensions would allow for simpler manufacturing 4][0054)).
`
`Therefore,it would have been obvious to modify the barriers of Wu/Robinsonto have layers
`
`with similar profiles and dimensions a taught by Weston,in order to allow for simpler manufacturing.
`
`Regarding Claim 2, Wu/Robinson/Weston teachesthe claimed limitations as set forth above for
`
`Claim 1.
`
`

`

`Application/Control Number: 17/281,668
`Art Unit: 3781
`
`Page 8
`
`Robinson teachesthat the plurality of pressure indicators comprisesblisters or raised channels
`
`(Fig 1 of Robinson showsa raised configuration of 136) in the second barrier (138, Fig 1), which would
`
`provide a visualor tactile indication whether sufficient reduced pressure is achieved.
`
`Therefore, the combination of Wu/Robinson/Weston alsoteaches that the plurality of pressure
`
`indicators comprises blisters or raised channels in the second barrier, which would provide a visual or
`
`tactile indication whether sufficient reduced pressure is achieved.
`
`Regarding Claim 3, Wu/Robinson/Weston teachesthe claimed limitations as set forth above for
`
`Claim 1.
`
`Robinson teachesthat each of the plurality of pressure indicators (pressure indicators 230 of
`
`Wu,Fig 5) is aligned with only one of the indicator ports (aperture 120 of Robinson, Fig 1) in order to
`
`indicate to the user or physician if negative pressure is uniformly distributed and to indicate the
`
`presence and location of clogs within the fluid path and to reduce or eliminate vacuum lock associated
`
`with the deliver conduit and reduced pressure source (][0041]).
`
`Therefore, the combination of Wu/Robinson/Weston alsoteaches that each of the plurality of
`
`pressureindicatorsis aligned with only one of the indicator ports in order to indicate to the user or
`
`physician if negative pressure is uniformly distributed and to indicate the presence and location of clogs
`
`within the fluid path and to reduce or eliminate vacuum lock associated with the deliver conduit and
`
`reduced pressure source (as motivated by Robinson 4][0041)).
`
`Regarding Claim 4, Wu/Robinson/Weston is silent whether each of the plurality of pressure
`
`indicators has a length of about 10 millimeters to about 30 millimeters, a width of about 2 millimeters to
`
`about 4 millimeters, and a height of about 1 millimeter to about 3 millimeters.
`
`It would have been obvious to cause the device of Wu/Robinson to have each of the plurality of
`
`pressure indicators to have a lengthof about 10 millimeters to about 30 millimeters, a width of about 2
`
`millimeters to about 4 millimeters, and a height of about 1 millimeter to about 3 millimeters since it has
`
`

`

`Application/Control Number: 17/281,668
`Art Unit: 3781
`
`Page 9
`
`been held that “where the only difference between the prior art and the claims was a recitation of
`
`relative dimensions of the claimed device and a device having the claimed relative dimensions would
`
`not perform differently than the prior art device, the claimed device was not patentably distinct from
`
`the prior art device” Gardner v. TEC Syst., Inc., 725 F.2d 1338, 220 USPQ 777 (Fed. Cir. 1984), cert.
`
`denied, 469 U.S. 830, 225 SPQ 232 (1984) (MPEP 2144.04 IV(A)). In the instant case, the device of
`
`Wu/Robinson would not operate differently with the claimed dimensions. Further, applicant places no
`
`criticality on the range claimed, indicating simply that the dimensions “may” be within the claimed
`
`ranges (specification 4][00135]).
`
`Regarding Claim 5, Wu/Robinson/Weston is silent regarding a gap between each ofthe plurality
`
`of pressure indicators, whereinthe gap has a length of about 5 millimeters to about 10 millimeters.
`
`It would have been obvious to cause the device of Wu/Robinson to have a gap between each of
`
`the plurality of pressure indicators, wherein the gap has a length of about 5 millimeters to about 10
`
`millimeters since it has been held that “where the only difference between the prior art and the claims
`
`was a recitation of relative dimensions of the claimed device and a device having the claimed relative
`
`dimensions would not perform differently than the prior art device, the claimed device was not
`
`patentably distinct from the prior art device” Gardner v. TEC Syst., Inc., 725 F.2d 1338, 220 USPQ 777
`
`(Fed. Cir. 1984), cert. denied, 469 U.S. 830, 225 SPQ 232 (1984) (MPEP 2144.04 IV(A)). In the instant
`
`case, the device of Wu/Robinson would not operate differently with the claimed length. Further,
`
`applicant places no criticality on the range claimed, indicating simply that the length “may” be within
`
`the claimed ranges(specification 4][00135]).
`
`Regarding Claim 8, Wu/Robinson/Weston teachesthe claimed limitations as set forth above for
`
`Claim 1.
`
`

`

`Application/Control Number: 17/281,668
`Art Unit: 3781
`
`Page 10
`
`Robinson teachesthat each of the indicator ports (aperture 120, Fig 1) comprises an aperture in
`
`the first barrier, which would indicate to the user or physician if negative pressureis uniformly
`
`distributed and would indicate the presence and location of clogs within the fluid path.
`
`Therefore, Wu/Robinson/Weston teachesthat each of the indicator ports comprises an
`
`aperture in the first barrier, which would indicate to the user or physician if negative pressureis
`
`uniformly distributed and would indicate the presence andlocation of clogs within the fluid path.
`
`Regarding Claim 10, Wu/Robinson/Westonis silent whether the indicator ports are distributed
`
`along the first barrier with a spacing of about 10 millimeters to about 30 millimeters.
`
`It would have been obvious to cause the device of Wu/Robinson to have the indicator ports
`
`distributed along the first barrier with a spacing of about 10 millimeters to about 30 millimeters since it
`
`has been held that “where the only difference betweentheprior art and the claims was a recitation of
`
`relative dimensions of the claimed device and a device having the claimed relative dimensions would
`
`not perform differently than the prior art device, the claimed device was not patentably distinct from
`
`the prior art device” Gardner v. TEC Syst., Inc., 725 F.2d 1338, 220 USPQ 777(Fed. Cir. 1984), cert.
`
`denied, 469 U.S. 830, 225 SPQ 232 (1984) (MPEP 2144.04 IV(A)). In the instant case, the device of
`
`Wu/Robinson would not operate differently with the claimed spacing. Further, applicant places no
`
`criticality on the range claimed, indicating simply that the spacing “may” be within the claimed ranges
`
`(specification 4][00132)).
`
`Regarding Claim 11, Wu further discloses that the first fluid-impermeable material is
`
`polyurethaneor polyethylene (conduit body 105 can be made of polyethylene or polyurethane 4][0046]).
`
`Regarding Claim 19, Wu/Robinson/Weston teachesthe claimed limitations as set forth above
`
`for Claim 1.
`
`

`

`Application/Control Number: 17/281,668
`Art Unit: 3781
`
`Page 11
`
`Robinson teaches an adhesive (bonds 402 may be used to connect the flexible member 138 to
`
`the connector body 123. Fig 4 4 [0049]) between the second barrier (138, Fig 4) and the first barrier
`
`(combination of 215 and 210, Fig 5 of Wu).
`
`Therefore, the combination of Wu/Robinson/Weston teaches anadhesive between the second
`
`barrier and the first barrier.
`
`Regarding Claim 26, Wu teaches an apparatus for treating a tissue site with negative pressure
`
`(device 100, Fig 1), the apparatus comprising:
`
`A source of negative pressure (The proximal port 125 can connect, for example via tubing 130,
`
`to any of a variety of negative pressure sources (not shown) 4][0041]); a dressing (distal port 115 may
`
`connect to a wound dressing positioned at a wound site 4[0041]) configured to be applied to the tissue
`
`site; and a fluid conductor (conduit body 105, Fig 5), comprising:
`
`a first barrier (combination of upper cover 215 and base 210, Fig 5) formed fromafirst fluid -
`
`impermeable material (conduit body 105 can be made of polyethylene or polyurethane 4][0046]), the
`
`first barrier comprising:
`
`a first layer formed from a film (215 is formed fromafilm, defined as a thin layer of any
`
`material, Fig 5);
`
`a second layer (210, Fig 5) sealedto thefirst layer to define a fluid path between thefirst layer
`
`and the second layer (device 100 may be configured to create and maintain a fluidic pathway,
`
`specifically between layers 215 and 210 4[0041][0050]), the second layer formed from a film (210 is
`
`formed fromafilm, defined as a thin layer of any material, Fig 5);
`
`the fluid path having a first end (distal port 115, Fig 1. Figs 1 and 5 show the same device, with
`
`Fig 5 illustrating the different layers that make up conduit body 105. Distal port 115, proximal port 125,
`
`and device 100, are presentin both figures, but they are unlabeled in Fig 5), a secondend (proximal port
`
`

`

`Application/Control Number: 17/281,668
`Art Unit: 3781
`
`Page 12
`
`125, Fig 1), and a longitudinal axis (conduit body 105 has a longitudinal axis extending between 125 and
`
`115, Fig 1);
`
`and a plurality of standoffs (support structures 205, Fig 7A) formed byat least one ofthe first
`
`layer and the second layer (support structures 205 may be integrated with or coupled to the inner walls
`
`of the conduit body 105, which includes layers 215 and 210 4[0048]), the plurality of standoffs disposed
`
`in the fluid path (The support structures 205 prevent or mitigate collapse of the channel 200 upon
`
`application of internal vacuum, external pressure to or compression of the conduit body 105 4[0047)).
`
`and a plurality of pressure indicators (both ports 115 and 125 may include pressure indicators
`
`230, Fig 5. The indicator 230 can be positioned in one or more locations along the device 100 suchas
`
`within the tubing 130, ona region of the conduit body 105, near the distal port 115 or proximal port 125
`
`4[0057]) configured to elastically deform in response to a reduction in pressure (Deformation of the
`
`exterior wall of the indicator 230 can result in a change fromafirst profile to a second profile. The
`
`relative position of the exterior wall between thefirst profile and the second profile can be indicative of
`
`an amount of pressure delivered to the woundsite. 4][0058])
`
`configuredto fluidly couple the dressing to the source of negative pressure (the device 100
`
`may be configured to create and maintain a fluidic pathway or channel for the delivery of negative
`
`pressure to and removal of exudates from the wound site 4[0041)).
`
`Wuis silent regarding the first barrier and second barrier having similar profiles and dimensions,
`
`indicator ports distributed linearly along the first barrier parallel to and in fluid communication with the
`
`fluid path; and asecond barrier coupled to thefirst barrier, the second barrier formed from asecond
`
`fluid-impermeable material defining a plurality of pressure indicators aligned with the indicator ports
`
`and configured to elastically deform in response to a reduction in pressure through the indicatorports.
`
`However, Robinson teaches a reduced pressure therapy dressing, thus fromthe same field of
`
`endeavor, withindicator ports (port aperture 120, Fig 1), and a second barrier (flexible member 138, Fig
`
`

`

`Application/Control Number: 17/281,668
`Art Unit: 3781
`
`Page 13
`
`1) coupled to thefirst barrier (sealing member 114, Fig 1), the second barrier (138, Fig 1) formed froma
`
`second fluid-impermeable material (connector body 123 and flexible member 138 may be formed or
`
`molded as an integral member. 4[0042] Connector body 123 may be made from polyurethane [0032]
`
`and thus 138 may be madeof polyurethane as well) defining a pressure indicator (dressing valve 136
`
`may indicate a reduced pressurestate. Upon being subjected to at least threshold reduced pressure that
`
`is delivered to the interior of the dressing valve 136, the flexible member 138 collapses and seals the
`
`port aperture 120. 4[0042]) aligned with the indicator ports (dressing valve 136 is aligned with aperture
`
`120, Fig 1) and configured to elastically deform in response to a reduction in pressure (Upon being
`
`subjected to at least threshold reduced pressure that is delivered to the interior of the dressing valve
`
`136, the flexible member 138 collapses and seals the port aperture 120. [0042]) through the indicator
`
`ports (reduced pressureis delivered to dressing valve 136 through port aperture 120 4][0040]), in order
`
`to indicate to the user or physician if negative pressureis uniformly distributed and to indicate the
`
`presence and location of clogs within the fluid path and to reduce or eliminate vacuum lock associated
`
`with the delivery conduit and reduced pressure source (4][0041]).
`
`Therefore, it would have been obvious to one of ordinary skill in the art before the effective
`
`filing date of the claimed invention to modify the device of Wu to have substitute the pressure indicator
`
`structure of Robinson for the pressure indicator structure of Wu. The combination of the conduit having
`
`a first impermeable barrier and a plurality of pressure indicators of Wu with the second impermeable
`
`barrier and aperture of the pressure indicator of Robinson would result in a device having two
`
`impermeable barriers with apertures distributed linearly along the first barrier and in fluid
`
`communication with the fluid path, where each apertureis associated with a single elastically
`
`deformable pressure indicator formed from the second impermeable barrier. The resulting device would
`
`be advantageoussinceaplurality of pressure indicators would indicate to the user or physician if
`
`negative pressureis uniformly distributed and would indicate the presence and location of clogs within
`
`

`

`Application/Control Number: 17/281,668
`Art Unit: 3781
`
`Page 14
`
`the fluid path and to reduce or eliminate vacuum lock associated with the delivery conduit and reduced
`
`pressure source (as motivated by Robinson §][0041]).
`
`Wu/Robinsonis silent regarding the first barrier and second barrier having similar profiles and
`
`dimensions.
`
`However, Weston teaches a reduced pressure treatment system, thus from the same field of
`
`endeavor, witha wound cover that is interpreted as having a dual layer structure wherein both layers
`
`have similar profiles and dimensions (bottom layer of wound cover 40 not including protrusions 60, and
`
`top layer is the portion of the layer including protrusions 60 and remaining surface 43 Fig 5A-D since
`
`having the layers be of similar profiles and dimensions would allow for simpler manufacturing 4][0054)).
`
`Therefore,it would have been obvious to modify the barriers of Wu/Robinsonto have layers
`
`with similar profiles and dimensions a taught by Weston,in order to allow for simpler manufacturing.
`
`Claim 7 is rejected under 35 U.S.C. 103 as being unpatentable over Wu/Robinson /Weston in
`
`view of Locke (WO 2016182977 A1).
`
`Regarding Claim 7, Wu/Robinson/Weston is silent regarding open-cell foam or gauze
`
`configured to bias the pressure indicators.
`
`However, Locke teaches a reduced pressure device, thus from the same field of endeavor,
`
`wherein open-cell foam or gauze (foam block 134, Fig 1) configured to bias the pressure indicators
`
`({[0059]). If the foam block remains compressed, a patient or clinician may have anindication that the
`
`therapy pressure has been reached. The compressed foam block may also act as a pressure reservoir.
`
`As negative pressure therapyis provided, there may be a natural leakage or decline of negative
`
`pressure at the tissue site. As the negative pressure decreases inthe sealed therapeutic environment,
`
`the pressure differential across the dressing assembly may decrease and the foam block may gradually

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket